Female to Female Cable Information
Description
The Model 2290-5 SHV cable can contain hazardous voltages. Be certain that the high voltage
is completely discharged before removing or connecting the high-voltage cable. High-voltage
cables can store charge if they are dis connected from the supply while high voltage i s turned
on and can cause personal injury or d eath if not handled properly.
Do not connect the HIGH-VOLTAGE output to exposed circuitry. Any load connected to the
HIGH-VOLTAGE output should be enclosed in a metal shield that is connected to safety earth
ground using #18AWG or larger wire.
The high-voltage output cable connects the Model 2290-5 source output to external equipment.
High-voltage cable installation:
1. Turn off power and allow all voltage to discharge.
2. Install the Model 2290-5-SHV cable (see next figure).
Figure 1: 2290-5-SHV cable
077084301 *P077084301* 1

Model 2290-5-SHV
Characteristics
Length
Type
Maximum operating voltage (RMS)
Impedance
3.05 m (10')
RG-8A/U coaxial
5,000 V
50 Ω
2 077084301