Description

The Model 2290-5 can contain hazardous voltages. Be certain that the high voltage is completely discharged before rem o ving or connecting the high-voltage cable. High-voltage cables can store charge if they are dis connected from the supply while high voltage is turned on and can cause personal injury or d eath if not handled properly. Do not connect the HIGH-VOLTAGE output to exposed circuitry. Any load connected to the HIGH-VOLTAGE output should be enclosed in a metal shield that is connected to safety earth ground using #18AWG or larger wire.
The high-voltage output cable connects the Model 2290-5 source output to external equipment. High-voltage cable installation:
1. Turn off power and allow all voltage to discharge.
2. Install the Model 2290-5-MHV high-voltage cable (see next figure).
Figure 1: Model 2290-5 MHV cable
SHV Plug to MHV Male Cable
*077084200*
Model 2290-5-MHV SHV plug to MHV male cable
2 077084200

Characteristics

Length 10 feet (3.05 meters) Type RG-8A/U coax Maximum operating voltage (RMS) 5,000 volts Impedance 50 Ohms