Tektronix 2290-5 User manual

Model 2290-5-SHVBH
077084400
1
28775 Aurora Road Cleveland, Ohio 44139 1-888-KEITHLEY
http://www.keithley.com

Description

The Model 2290-5 can contain hazardous voltages. Be certain that the high voltage is completely discharged before removing or connecting the high-voltage cable. High-voltage cables can store charge if they are dis connected from the supply while high voltage is turned on and can cause personal injury or d eath if not handled properly. Do not connect the HIGH-VOLTAGE output to exposed circuitry. Any load connected to the HIGH-VOLTAGE output should be enclosed in a metal shield that is connected to safety earth ground using #18AWG or larger wire.
The high-voltage 5 kV SHV male bulkhead connector is used to co nnec t exter na l equ ipment.
High-voltage ins tallation:
1. Turn off power and allow all voltage to discharge.
2. Install the Model 2290-5-SHVBH SHV male bulkhead connector (see next figures).
Figure 1: Model 2290-5 SHV male bulkhead connector
SHV Male Bulkehead Connector
Figure 2: Model 2290-5 SHV male bulkhead connector mounting dimensions
*077084400*
Model 2290-5-SHVBH Connector
2 077084400

Characteristics

Nominal impedance non-constant Maximum operating voltage (RMS) 3500 RMS
5000 V DC Insulation resistance Temperature range
6
10E
M
-65° C to 165° C
(-85° F to +329° F)

Materials

A. Bodies and fittings: brass, QQ-B-626, alloy 360, 1/2 hard B. Crimp sleeves: commercial bronze, alloy 220 C. Male center contact: brass, QQ-B-626, alloy 360, 1/2 hard D. Outer contact: beryllium copper, ASTM-B-196 E. Insulator: PTFE, ASTM D1457 F. Gaskets and seals: silicone rubber, ZZ-R-765A, class IIA, grade 50 G. Contact finish: .0001 inch minimum, hard gold plate, MIL-G-45204B