Model 2290-10-SHV
Keithley Instruments, Inc.
28775 Aurora Road
Cleveland, Ohio 44139
1-888-KEITHLEY
http://www.keithley.com
Description
The Model 2290-10-SHV can contain hazardous voltages. Be certain that the high volta g e is
completely discharged before rem o ving or connecting the high-voltage cable. High-voltage
cables can store charge if they are dis connected from the supply while high voltage is turned
on and can cause personal injury or d eath if not handled properly.
Do not connect the HIGH-VOLTAGE output to exposed circuitry. Any load connected to the
HIGH-VOLTAGE output should be enclosed in a metal shield that is connected to safety earth
ground using #18AWG or larger wire.
The high-voltage output connector is a SHV male connector that connects the Model 2290-10 source output to
external equipment.
High-voltage cable installation:
1. Turn off power and allow all voltage to discharge.
2. Install Model 2290-10-SHV high-voltage cable (see next figure).
10 kV SHV Male to SHV Male Cable
Figure 1: Model 2290-10 kV Male to 10 kV Male Cable
*077084600*

Model-10-SHV 10 kV SHV Male to SHV Male Cable
Electrical char ac teristics
Length 10 feet (3.05 meters)
Type RG-213/U coax cable
Maximum operating voltage (RMS) 10,000 volts