Tektronix 2290-10 User manual

Model 2290-10-SHVBH
077084700
1
28775 Aurora Road Cleveland, Ohio 44139 1-888-KEITHLEY
http://www.keithley.com

Description

The Model 2290-10 can contain hazardous voltages. Be certain that the high voltage is completely discharged before rem o ving or connecting the high-voltage cable. High-voltage cables can store charge if they are dis connected from the supply while high voltage is turned on and can cause personal injury or d eath if not handled properly. Do not connect the HIGH-VOLTAGE output to exposed circuitry. Any load connected to the HIGH-VOLTAGE output should be enclosed in a metal shield that is connected to safety earth ground using #18AWG or larger wire.
The high-voltage 10 kV SHVBH SHV female bulkhead connector connects to the Model 2290-10-SHVUC male to unterminated cable (refer to document number 077084500 for more information).
High-voltage ins tallation:
1. Turn off power and allow all voltage to discharge.
2. Install the Model 2290-10-SHVBH female bulkhead connector to the Model 2290-10-SHVUC male to unterminated cable (see next figure).
SHV Female Bulkhead Connector
Figure 1: Model 2290-20 SHV female bulkhead connector

Electrical char ac ter i stics

Nominal impedance non-constant Voltage rating 10 kV DC Insulation resistance
10E
6
M
*077084700*
Model-10-SHVBH Female Bulkhead Connector
2 077084700

Material characteristics

A. Insulator: high-density polyethylene, ASTM-D4976, natural B. O-rings: silicone rubber, A-A-59588 C. Lock-washer: phosphor bronze D. A ll other parts: brass, ASTM-B16 E. Center contact: gold plate per ASTM-B488 F. Leakage may not exceed 10x-7 cc/second of tracer gas at atmospheric pressure