Tektronix 2280 User manual

Keithley Instruments
Model 2280 and 2281S
28775 Aurora Road Cleveland, Ohio 44139 1-800-833-9200
tek.com/keithley
Declassification and Security Instructions

Letter of Volatility

If you have data security concerns, this document tells you how to clear or sanitize instrument memory devices in Model 2280 and 2281 instruments. It also explains how to declassify an instrument that is not functioning.
The procedures in this document are written to meet the requirements specified in:
NISPOM, DoD 5220.22-M, chapter 8
ISFO Process Manual for Certification and Accreditation of Classified Systems under NISPOM

Contact information

If you have any questions after you review the information in this documentation, please contact your local Keithley Instruments office, sales partner, or distributor. You can also call the Tektronix corporate headquarters (toll-free inside the U.S. and Canada only) at 1-800-833-9200. For worldwide contact numbers, visit
tek.com/contact-us
.

Products

Series 2280 Precision Measurement DC Power Supplies
2280S-32-6
2280S-60-3
Series 2281 Dynamic Model Battery Simulator
2281S-20-6

Terminology

The following terms may be used in this document:
Clear: Removes data on media or in memory before reusing it in a secured area. Clears all reusable
memory to deny access to previously unsecured information.
Demo setups: Demonstration applications that come loaded on the instrument; you cannot modify them.
Direct method of modification: You can modify data directly.
Erase: Equivalent to clear (see above).
Indirect method of modification: The instrument system resources modify the data; you
cannot modify the data directly.
077172100 July 2021 *P077172100* 1
Model 2280 and 2281S Declassification and Security Instructions
Instrument declassification: Procedures that must be completed before an instrument can be removed
from a secure environment. Declassification procedures include memory sanitization and memory removal.
Media storage and data export device: Devices that can be used to store or export data from the
instrument, such as a USB port.
Nonvolatile memory: Data is retained when the instrument power is turned off.
Protected user data area: Contains data that is protected by a password.
Remove: Clears instrument data by physically removing the memory device from the instrument.
Sanitize: Eradicates instrument data from media and memory so it cannot be recovered by other means
or technology. This is typically used when the device will be moved (temporarily or permanently) from a secured area to a nonsecured area.
Scrub: Directly retrieve and clear the contents of the memory device.
User accessible: You can directly retrieve the contents of the memory device.
User data: Measurement data that represents signals that you connect to the instrument.
User modifiable: You can write to the memory device during normal instrument operation using the
front-panel interface or remote control.
User settings: Instrument settings that you can change.
Volatile memory: Temporary memory; data is lost when the instrument is turned off.

Description of memory

These instructions will work for all instruments listed in the products (on page 1) section above.
All Model 2280 and 2281S instruments contain the following volatile and nonvolatile memory:
(8) 2 GB nonvolatile NAND FLASH chips
(8) 128 KB volatile DDR2 SRAM chips