Tektronix 222PS Operator's Manual

Page 1
222PS PowerScout Power Systems Oscilloscope Operator Manual
070-8097-02
Please check for change information at the rear of this manual.
Second Edition: December 1993 Last Revised: September 29, 1994
Page 2
Instrument Serial Numbers
Each instrument manufactured by Tektronix has a serial number on a panel insert or tag, or stamped on the chassis. The first letter in the serial number designates the country of manufacture. The last five digits of the serial number are assigned sequentially and are unique to each instrument. Those manufactured in the United States have six unique digits. The country of manufacture is identified as follows:
B010000 Tektronix, Inc., Beaverton, Oregon, USA E200000 Tektronix United Kingdom, Ltd., London J300000 Sony/Tektronix, Japan H700000 Tektronix Holland, NV, Heerenveen, The Netherlands
Instruments manufactured for Tektronix by external vendors outside the United States are assigned a two digit alpha code to identify the country of manufacture (e.g., JP for Japan, HK for Hong Kong, IL for Israel, etc.).
Tektronix, Inc., P.O. Box 500, Beaverton, OR 97077
Printed in U.S.A.
Copyright are covered by U.S. and foreign patents, issued and pending. The following are registered trademarks: TEKTRONIX, TEK, TEKPROBE, and SCOPEĆMOBILE.
E Tektronix, Inc., 1991, 1993. All rights reserved. Tektronix products
Page 3
WARRANTY
Tektronix warrants that this product will be free from defects in materials and workmanship for a period of three (3) years from the date of shipment. If any such product proves defective during this warranty period, Tektronix, at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period and make suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the defective product to the servicecenter designated by Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting from attempts by personnel other than Tektronix representatives to install, repair or service the product; b) to repair damage resulting from improper use or connection to incompatible equipment; or c) to service a product that has been modified or integrated with other products when the effect of such modification or integration increases the time or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THIS PRODUCT IN LIEU OF ANY OTHER WARRANTIES, EXPRESSED OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX' RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.
Page 4
Certificate of the Manufacturer/Importer
We hereby certify that the
SYSTEMS OSCILLOSCOPE AND ALL INSTALLED OPTIONS
222PS PowerScout POWER
complies with the RF Interference Suppression requirements of Amtsbl.ĆVfg 1046/1984.
The German Postal Service was notified that the equipment is being marketed.
The German Postal Service has the right to reĆtest the series and to verify that it complies.
TEKTRONIX
Bescheinigung des Herstellers/Importeurs
Hiermit wird bescheinigt, daß der/die/das
POWER SYSTEMS OSCILLOSCOPE AND ALL INSTALLED OPTIONS
222PS PowerScout
in Übereinstimmung mit den Bestimmungen der AamtsblattĆ Verfugüng 1046/1984 funkentstört ist.
Der Deutschen Bundespost wurde das Inverkehrbringen dieses Gerätes angezeigt und die Berechtigung zur ÜberĆ prufüng der Serie auf Einhalten der Bestimmungen einĆ geräumt.
TEKTRONIX
Page 5
NOTICE to the user/operator:
The German Postal Service requires that Systems assembled by the operator/user of this instrument must also comply with Postal Regulation, Vfg. 1046/1984, Par. 2, Sect. 1.
HINWEIS für den Benutzer/Betreiber:
Die vom Betreiber zusammengestellte Anlage, innerhalb derer dies Gerät eingesetzt wird, muß ebenfalls den Voraussetzungen nach Par. 2, Ziff. 1 der Vfg. 1046/1984 genugen.
NOTICE to the user/operator:
The German Postal Service requires that this equipment, when used in a test setup, may only be operated if the requirements of Postal Regulation, Vfg. 1046/1984, Par. 2, Sect. 1.7.1 are complied with.
HINWEIS für den Benutzer/Betreiber:
Dies Gerät darf in Meßaufbauten nur betrieben werden, wenn die Voraussetzungen des Par. 2, Ziff. 1.7.1 der Vfg. 1046/1984 eingehalten werden.
Page 6
Page 7

Welcome

This manual contains the following sections:
H Overview describes the 222PS PowerScout and provides safety
information.
H At A Glance describes the controls and connectors for the 222PS.
H In Detail provides further detail on some aspects of the 222PS,
building on the information contained in At A Glance. The 15 topics of this chapter are in alphabetical order for your conveĆ nience:
H Acquisition Modes
H Auto Setup
H Calibration
H Capturing Random Events
H Channels
H The Display
H Horizontal Operation
H Maintenance and Repair
H Power
H Probes
H Saving and Recalling Data
H Store Mode
H Triggering
H Vertical Operation
H XY Mode
H Tutorial: Measuring Signals provides stepĆbyĆstep instructions to
get you started making measurements quickly.
222PS Operator Manual
i
Page 8
H Remote Communication provides information on RSĆ232 commuĆ
nication procedures between the 222PS and a PC.
H Performance Verification describes the procedures necessary to
verify that the 222PS is performing according to specifications.
H Specifications provides complete specifications for the 222PS
PowerScout.
H Accessories describes the standard and optional accessories
available for the 222PS.
H Glossary defines various words used in the text.
H The Index helps you locate information quickly.
NOTE
If you have never used an oscilloscope before, please read the tutorial in Appendix A before using the 222PS.
ii
Welcome
Page 9
List of Illustrations ix.....................................
List of Tables xiii..........................................
Overview
About the 222PS PowerScout 1Ć1......................
Safety 1Ć2............................................
At a Glance
Front Panel Controls 2Ć2...............................
The Display 2Ć6.......................................
Top Panel Controls 2Ć11................................
Side Connectors 2Ć13..................................
Rear Panel Controls and Connectors 2Ć14................
The Tilt Stand 2Ć16.....................................

Contents

Symbols and Terms 1Ć2.............................
Specific Precautions 1Ć3............................
Vertical Controls 2Ć3................................
Trigger Controls 2Ć4................................
Horizontal Controls 2Ć5.............................
Readouts 2Ć6......................................
Menus and Menu Buttons 2Ć9.......................
In Detail
Acquisition Modes 3Ć1.....................................
Kinds of Acquisition Modes 3Ć1........................
Normal Acquisition Mode 3Ć1........................
Average Acquisition Mode 3Ć1.......................
222PS Operator Manual
iii
Page 10
Envelope Acquisition Mode 3Ć2......................
Continuous Envelope Acquisition Mode 3Ć2...........
Selecting an Acquisition Mode 3Ć3.....................
Auto Setup 3Ć5............................................
Procedure 3Ć5........................................
Parameter Effects 3Ć5.................................
Channels 3Ć5......................................
Vertical Scaling 3Ć6.................................
Horizontal Scaling 3Ć6..............................
Low and High Frequencies 3Ć6......................
Calibration 3Ć9............................................
Running the SelfĆCalibration Routine 3Ć9................
Capturing Random Events 3Ć15.............................
Channels 3Ć17.............................................
Selecting a Channel 3Ć17...............................
Displaying a Channel 3Ć18..............................
Setting Channel Coupling 3Ć19..........................
The Display 3Ć21...........................................
Readouts 3Ć21.........................................
Displaying and Clearing the Readouts 3Ć21.............
Inverting the Display 3Ć22...............................
Varying the Brightness 3Ć23.............................
Horizontal Operation 3Ć25..................................
Horizontal Positioning 3Ć25.............................
Seconds per Division 3Ć26..............................
Aliasing 3Ć26.......................................
TimeĆBase Mode 3Ć27...............................
Magnifying the Signal 3Ć29..............................
Maintenance 3Ć31..........................................
Identifying the Firmware Version 3Ć31....................
Repackaging for Shipment 3Ć33.........................
Power 3Ć35................................................
Battery Operation 3Ć35.................................
Connecting the Battery 3Ć35..........................
Charging the Battery 3Ć37............................
iv
Contents
Page 11
Charging the Battery Externally 3Ć37...................
Time Out 3Ć37......................................
Replacing the Battery 3Ć39...........................
Storing the Instrument 3Ć40..........................
Deep Discharge 3Ć40................................
External Power Operation 3Ć41..........................
AC Line Operation 3Ć41..............................
Other Sources of External Power 3Ć42.................
Probes 3Ć43...............................................
Connecting the Probes 3Ć44............................
Configuring the Probes 3Ć45............................
Probe Accessories 3Ć47................................
Saving and Recalling Data 3Ć51.............................
Waveforms 3Ć51........................................
Saving a Waveform 3Ć51.............................
Recalling a Saved Waveform 3Ć53.....................
Setups 3Ć55............................................
Saving a Setup 3Ć55.................................
Recalling a Saved Setup 3Ć57........................
Erasing a Saved Setup 3Ć58..........................
Store Mode 3Ć61...........................................
Triggering 3Ć63.............................................
The Trigger Light 3Ć63..................................
Trigger Source 3Ć64....................................
External Triggering 3Ć65.............................
Motor Trigger Function 3Ć67.............................
Triggering on Motor Drive Signals 3Ć67................
Triggering on 50/60 Hz Line Signals 3Ć67...............
Trigger Coupling 3Ć69..................................
Trigger Slope 3Ć69......................................
Trigger Level 3Ć69......................................
AutoĆLevel 3Ć70.....................................
Trigger Position 3Ć70...................................
Trigger Modes 3Ć71.....................................
Normal 3Ć72........................................
AutoĆBaseline 3Ć72..................................
AutoĆLevel 3Ć72.....................................
222PS Operator Manual
v
Page 12
Vertical Operation 3Ć75.....................................
Vertical Positioning 3Ć75................................
Volts per Division 3Ć76..................................
XY Mode 3Ć79..............................................
Entering and Exiting XY Mode 3Ć79......................
Positioning XY Waveforms 3Ć80.........................
Appendices
Appendix A: Tutorial AĆ1...................................
The Screen AĆ1.......................................
Measuring Voltage AĆ2.................................
Measuring Time AĆ5...................................
Appendix B: Remote Communication AĆ9...................
Introductory Information AĆ9...........................
Setting the Baud Rate AĆ11..............................
Command Formats AĆ13................................
FrontĆPanel Encoding AĆ20..............................
RSĆ232 Error Codes AĆ28...............................
Transfer Options AĆ30..................................
RSĆ232 Interface Specifications AĆ32....................
SingleĆSequence 3Ć73...............................
Setting the Trigger Mode 3Ć73........................
Variable Volts per Division 3Ć76.......................
Measuring PeakĆtoĆPeak Voltages AĆ3................
Using a Ground Reference Point AĆ4.................
Measuring RiseĆ or FallĆtime AĆ7.....................
Commands, Queries, and Responses AĆ13.............
Vertical Settings AĆ20................................
SEC/DIV Setting AĆ22................................
Trigger Positions, Slope, Source, and Mode Settings AĆ24
Acquisition Mode and Miscellaneous Settings AĆ26......
Status Messages AĆ28...............................
Diagnostic Error Codes AĆ28.........................
Transfer to a Local PC AĆ30...........................
Transfer via Modem AĆ30.............................
vi
Contents
Page 13
RSĆ232 Communication Parameters AĆ32..............
Baud Rates AĆ32....................................
Levels AĆ33.........................................
Maximum Applied Voltage AĆ33.......................
Plotter/Printer Support AĆ33..........................
Messages AĆ33.....................................
Optional Accessories AĆ33..............................
CAT200 AĆ33.......................................
RSĆ232 Interconnection Cables AĆ34...................
Appendix C: Quick Checks AĆ37.............................
General Instructions AĆ37...............................
Conventions AĆ38......................................
Initial Setup Procedure AĆ39............................
Self Cal Tests AĆ39.....................................
Autoset Tests AĆ41.....................................
Appendix D: Performance Tests AĆ43........................
Prerequisites AĆ43......................................
Equipment Required AĆ44...............................
Preparation AĆ47.......................................
Preliminaries AĆ47......................................
Vertical Checks AĆ48...................................
Horizontal Check AĆ57..................................
Trigger Checks AĆ60....................................
Appendix E: Specifications AĆ65.............................
Nominal Traits AĆ65.....................................
Warranted Characteristics AĆ71..........................
Typical Characteristics AĆ76.............................
Appendix F: Accessories AĆ79...............................
Standard Accessories AĆ79.............................
Optional Accessories AĆ80..............................
Instrument Options AĆ81................................
External Power AC Adapter Options AĆ81.................
222PS Operator Manual
vii
Page 14
Glossary & Index
Glossary GĆ1..............................................
Index IĆ1.................................................
viii
Contents
Page 15

List of Illustrations

Figure 2Ć1:ă222PS Front Panel 2Ć2...........................
Figure 2Ć2:ăVertical Controls 2Ć3.............................
Figure 2Ć3:ăTrigger Controls 2Ć4.............................
Figure 2Ć4:ăHorizontal Controls 2Ć5..........................
Figure 2Ć5:ăVertical Readouts 2Ć7............................
Figure 2Ć6:ăSaved Waveform Readouts 2Ć8...................
Figure 2Ć7:ăTrigger and Horizontal Readouts 2Ć9...............
Figure 2Ć8:ăParts of a Menu 2Ć10..............................
Figure 2Ć9:ă222PS Top Panel 2Ć11............................
Figure 2Ć10:ăSide of 222PS, Storage Pouch Removed 2Ć13......
Figure 2Ć11:ă222PS Rear Panel 2Ć14..........................
Figure 2Ć12:ăThe 222PS With Tilt Stand 2Ć16...................
Figure 3Ć1:ăNormal, Average, and Envelope Signals 3Ć3........
Figure 3Ć2:ăThe Acquisition Menu 3Ć3........................
Figure 3Ć3:ăThe Auxiliary Functions Menu 3Ć10.................
Figure 3Ć4:ăThe Calibration Menu 3Ć10........................
Figure 3Ć5:ăThe Ground External Trigger Calibration Menu 3Ć12...
Figure 3Ć6:ăChannel One Selected 3Ć17.......................
Figure 3Ć7:ăThe Channel Menu 3Ć18...........................
Figure 3Ć8:ăThe Display Menu 3Ć21...........................
Figure 3Ć9:ăNormal and Inverted Waveforms 3Ć22...............
Figure 3Ć10:ăThe Horizontal Controls 3Ć25.....................
Figure 3Ć11:ăAn Aliased Waveform 3Ć27.......................
Figure 3Ć12:ăThe Auxiliary Functions Menu 3Ć32................
Figure 3Ć13:ăXY Alignment Menu 3Ć32.........................
Figure 3Ć14:ăThe Firmware Version 3Ć33.......................
Figure 3Ć15:ăSide View Without Battery Cover 3Ć36..............
Figure 3Ć16:ăThe Auxiliary Functions Menu 3Ć38................
Figure 3Ć17:ăThe Configuration Menu 3Ć38.....................
Figure 3Ć18:ăSide View Showing Probe Connectors
(Pouch Removed) 3Ć44......................................
Figure 3Ć19:ăThe Auxiliary Functions Menu 3Ć45................
222PS Operator Manual
ix
Page 16
Figure 3Ć20:ăThe Configuration Menu 3Ć46.....................
Figure 3Ć21:ăThe Probe Menu 3Ć46............................
Figure 3Ć22:ăProbe and Accessories 3Ć48......................
Figure 3Ć23:ăSaved Waveforms Menu 3Ć52.....................
Figure 3Ć24:ăSaved Waveform Parameters 3Ć53.................
Figure 3Ć25:ăThe Recall Waveforms Menu 3Ć54.................
Figure 3Ć26:ăThe Setup Menu 3Ć56............................
Figure 3Ć27:ăThe Save Setup Menu 3Ć56.......................
Figure 3Ć28:ăThe Recall Setup Menu 3Ć57......................
Figure 3Ć29:ăThe Erase Setups Menu 3Ć59.....................
Figure 3Ć30:ăHorizontal Readouts in Store Mode 3Ć61...........
Figure 3Ć31:ăTrigger Controls 3Ć63............................
Figure 3Ć32:ăThe Trigger Source Menu 3Ć64....................
Figure 3Ć33:ăRear Panel External Trigger Connectors 3Ć65.......
Figure 3Ć34:ăMotor Drive Signals Separated by a Low Region 3Ć68 Figure 3Ć35:ăMotor Drive Signals Separated by a High Region 3Ć68
Figure 3Ć36:ăTrigger Position Menu 3Ć71.......................
Figure 3Ć37:ăThe Trigger Mode Menu 3Ć74.....................
Figure 3Ć38:ăThe Vertical Controls 3Ć75........................
Figure 3Ć39:ăThe Display Menu 3Ć79..........................
Figure AĆ1:ăThe Screen Markings AĆ2........................
Figure AĆ2:ăMeasuring PeakĆtoĆPeak Voltages AĆ3.............
Figure AĆ3:ăMeasuring With Respect to Ground AĆ5............
Figure AĆ4:ăMeasuring Time AĆ6.............................
Figure AĆ5:ăMeasuring RiseĆTime AĆ8.........................
Figure AĆ6:ăRSĆ232 Communications Port AĆ10.................
Figure AĆ7:ăCommunication Between the 222PS and a PC AĆ10..
Figure AĆ8:ăThe Auxiliary Functions Menu AĆ11.................
Figure AĆ9:ăThe Configuration Menu AĆ12......................
Figure AĆ10:ăThe Baud Settings Menu AĆ12....................
Figure AĆ11:ăCommands, Queries, and Responses AĆ14.........
Figure AĆ12:ăRSĆ232 Interconnection Cable Pin Wiring AĆ34......
Figure AĆ13:ăInitial Setup for Vertical Checks AĆ49...............
Figure AĆ14:ăSetup for Input Coupling Check AĆ52..............
Figure AĆ15:ăSetup for Probe Compensation Check AĆ55........
Figure AĆ16:ăSetup for Analog Bandwidth Check AĆ56...........
x
Contents
Page 17
Figure AĆ17:ăSetup for Horizontal Test AĆ58....................
Figure AĆ18:ăSetup for Trigger Checks AĆ61....................
Figure AĆ19:ăSetup for External Trigger Sensitivity Check AĆ63....
Figure AĆ20:ăMaximum NormalĆMode Voltage Versus
Frequency Derating Curve AĆ67...............................
Figure AĆ21:ăMaximum CommonĆMode Voltage Versus
Frequency Derating Curve AĆ68...............................
222PS Operator Manual
xi
Page 18
xii
Contents
Page 19

List of Tables

TableĂ3Ć1:ă Auto Setup Settings 3Ć7..........................
TableĂ3Ć2:ă TimeĆBase Modes 3Ć28............................
TableĂ3Ć3:ă AutoĆLevel Trigger Interval Time Limits 3Ć73..........
TableĂAĆ1:ăBUT Button Codes AĆ15...........................
TableĂAĆ2:ăDAC Code and Value Data AĆ17.....................
TableĂAĆ3:ăChannel Settings AĆ21.............................
TableĂAĆ4:ăSecondsĆperĆDivision and Misc. Settings AĆ23........
TableĂAĆ5:ăTrigger Position, Slope, Source, and Mode Settings AĆ25
TableĂAĆ6:ăAcquisition Mode and Misc. Settings AĆ27............
TableĂAĆ7:ă222PS Interconnect Pin Assignments AĆ34...........
TableĂAĆ8:ăTest Equipment AĆ45..............................
TableĂAĆ9:ăVoltsĆperĆDivision Accuracy Settings AĆ53............
TableĂAĆ10:ăNominal TraitsĊVertical System AĆ65..............
TableĂAĆ11:ăNominal TraitsĊHorizontal System AĆ68............
TableĂAĆ12:ăNominal TraitsĊTriggering System AĆ69............
TableĂAĆ13:ăNominal TraitsĊDisplay System AĆ69..............
TableĂAĆ14:ăNominal TraitsĊPower System AĆ70................
TableĂAĆ15:ăNominal TraitsĊCommunications Interface AĆ70.....
TableĂAĆ16:ăNominal TraitsĊMechanical AĆ71..................
TableĂAĆ17:ăWarranted CharacteristicsĊVertical System AĆ72....
TableĂAĆ18:ăWarranted CharacteristicsĊHorizontal System AĆ74. TableĂAĆ19:ăWarranted CharacteristicsĊTriggering System AĆ74.. TableĂAĆ20:ăWarranted CharacteristicsĊ
ąąąEnvironmental, Safety, and Reliability AĆ75...............
TableĂAĆ21:ăTypical CharacteristicsĊVertical System AĆ76.......
TableĂAĆ22:ăTypical CharacteristicsĊHorizontal System AĆ77.....
TableĂAĆ23:ăTypical CharacteristicsĊTriggering System AĆ77.....
TableĂAĆ24:ăTypical CharacteristicsĊPower System AĆ77........
TableĂAĆ25:ăStandard Accessories AĆ79.......................
TableĂAĆ26:ăOptional Accessories AĆ80........................
222PS Operator Manual
xiii
Page 20
TableĂAĆ27:ăInstrument Options AĆ81..........................
TableĂAĆ28:ăExternal Power AC Adapter Options AĆ81...........
xiv
Contents
Page 21
Page 22
Page 23
This section summarizes the product features and safety precautions of the 222PS PowerScout.

About the 222PS PowerScout

The Tektronix 222PS PowerScout is a portable digitizing oscilloscope. It has two fully isolated, independently floatable channels rated to 600 VAC oscilloscopes in the world with this feature.
The 222PS also has these features.
H light weight with a battery power source for field operations
H automatic setup button
H automatic triggering modes
H averaging and enveloping acquisition modes
H remote operationcapabilities via theRSĆ232 communications port
H internal memory for saving up to four waveforms and four instruĆ
ment setups
. The 222PS/224 digitizing oscilloscopes are the only
RMS

Overview

H DCĆtoĆ10ĂMHz signal bandwidth
H 10ĂMS/s digitizing rate
H 1 MHz single pass storage bandwidth
H Motor trigger
If you need more information about your Tektronix 222PS PowerScout or other Tektronix products, contact the nearest Tektronix sales office or distributor, consult the Tektronix product catalog, or, in the U. S., call the Tektronix National Marketing Center tollĆfree at 1Ć800Ć426Ć2200.
222PS Operator Manual
1Ć1
Page 24
Overview

Safety

You may be eager to begin using your Tektronix 222PS but please take a moment to review these safety precautions. We provide them for your protectionand to prevent damage to the 222PS PowerScout. This safety information applies to all operators and service personnel.
WARNING
To avoid personal injury or damage to the 222PS, do not apply more than 850ĂV peak between probe tip and earth ground, between probe tip and probe common, or beĆ tween probe common and earth ground.
WARNING
To avoid injury, use caution when working on equipment with voltages above 42ĂV peak. Such voltages pose a shock hazard.
1Ć2
WARNING
Do not float the external trigger common connector, the RSĆ232 communications port, or the external power input above 42ĂV peak. These inputs are not electrically isolated from each other.
Symbols and Terms
These two terms appear in manuals:
H
result in damage to the equipment or other property.
H
result in personal injury or loss of life.
These two terms appear on equipment:
statements identify conditions or practices that could
statements identify conditions or practices that could
Overview
Page 25
Overview
H CAUTION indicates a personal injury hazard not immediately
accessible as one reads the marking or a hazard to property including the equipment itself.
H DANGER indicates a personal injury hazard immediately accessiĆ
ble as one reads the marking.
This symbol appears in manuals:
StaticĆSensitive Devices
These symbols appear on equipment:
DANGER
High Voltage
Protective
ground (earth)
terminal
ATTENTION
Refer to
manual
Specific Precautions
Observe all these precautions to ensure your personal safety and to prevent damage either to the 222PS or to equipment connected to it.
Power Source Ċ The 222PS can use its selfĆcontained sealed lead
acid battery as a power source. It can also operate using power supplied to the external power input. Power supplied to this input must be 12 to 28ĂVDC or 16 to 20ĂVAC power conductor negative by more than 0.5ĂV with respect to chassis ground. Both conductors of the external power input are fused interĆ nally. These fuses are not user accessible.
You can operate the instrument with external power operation from local 110ĂV or 240ĂV power supply using the appropriate external power AC adapter. Use only external power AC adapters specified for this instrument.
. Do not force either external
RMS
222PS Operator Manual
1Ć3
Page 26
Overview
Grounding the PowerScout Ċ The channel 1 and channel 2
measurement inputs of the 222PS are doubly insulated from each other and all other accessible portions of the instrument cabinet. It is not necessary to ground the instrument to avoid electric shock.
Fuse Ċ The 222PS has no userĆreplaceable fuses.
Do Not Disassemble the Cabinet Ċ To avoid personal injury, do
not operate the instrument without a properly assembled cabinet. The cabinet of the instrument should be disassembled only by qualified service personnel.
Do Not Operate in Explosive Atmospheres Ċ The 222PS
provides no explosion protection from static discharges or arcing components. Do not operate the 222PS in an atmosphere of explosive gasses.
Electric Overload Ċ Never apply a voltage to a probe or connector
on the 222PS that is outside the range specified for that probe or connector.
1Ć4
Overview
Page 27
Page 28
Page 29

At a Glance

This chapter describes the controls, connectors, and display readouts of the 222PS PowerScout. It is intended to help orient you and to provide basic information. For more detailed operating instructions for various features, see the appropriate section in the chapter entitled In
Detail.
This section provides page references to the In Detail chapter for further information.
222PS Operator Manual
2Ć1
Page 30
At a Glance

Front Panel Controls

The front panel for the 222PS PowerScout appears as shown in Figure 2Ć1.
1
2
3
4
65
2Ć2
Figure 2Ć1:ă222PS Front Panel
1. The display area includes the screen and associated buttons. The screen shows signal traces, readouts, and menu items. Buttons along the side of the screen allow you to manipulate menus. See page 2Ć6 for a more complete description of the display area.
2. The vertical controls allow you to manipulate the vertical aspects of your signal. See page 3Ć77 for a more complete description of the vertical controls.
3. The trigger controls allow you to manipulate the aspects of your signal having to do with triggering. See page 3Ć63 for a more complete description of the trigger controls.
4. The horizontal controls allow you to manipulate the horizontal aspects of your signal. See page 3Ć25 for more information on the horizontal controls.
At a Glance
Page 31
At a Glance
5. AUTO SETUP allows you to set up the instrument with the push of a single button. Press this button for a quick, informative display of any signal between 20 Hz and 1 MHz. See page 3Ć5 for more information on the AUTO SETUP button.
6. The ON button toggles the instrument on or off. The instrument beeps when it is turned on.
Vertical Controls
Figure 2Ć2 shows the vertical controls that are located on the front panel.
321
Figure 2Ć2:ăVertical Controls
1. The outer knob sets the volts per division, which is the vertical scale of your display. Turn the knob clockwise to decrease the volts per division and counterclockwise to increase the volts per division. See page 3Ć78 for more information on setting the volts per division.
2. The inner knob sets the vertical position of the signal. Turn the knob clockwise to move the signal towards the top of the screen and counterclockwise to move the signal towards the bottom of the screen. For more information on setting the vertical position see page 3Ć77.
222PS Operator Manual
2Ć3
Page 32
At a Glance
3. The 222PS can display signals acquired through either or both of
Trigger Controls
The trigger controls are on the front panel of the 222PS PowerScout. They appear as shown in Figure 2Ć3.
You can also use this knob to change the size of a signal to an arbitrary number of divisions. To do so, see page 3Ć78.
its two channels. The channel selector buttons allow you to select the channel that is affected by changes to the controls. With these buttons you can also select channel coupling or turn a channel off so that the signal it is measuring is not displayed. For more information on channels, see page 3Ć17.
12
2Ć4
3456
Figure 2Ć3:ăTrigger Controls
1. When you push the button labeled SLOPE, you toggle between a positiveĆ and negativeĆtrigger slope. For more details, see page 3Ć70.
2. When you push the button labeled MODE, you invoke a menu that allows you to specify the trigger mode. For a complete explanation of trigger modes, see page 3Ć72.
At a Glance
Page 33
At a Glance
3. When you push the button labeled SOURCE, you invoke a menu that allows you to specify the trigger source. For a complete explanation of trigger sources, see page 3Ć64.
4. The light labeled TRIG'D turns on when the instrument is trigĆ gered. See page 3Ć63 for more details.
5. The inner button, labeled AUTOLVL: PUSH, sets the trigger level automatically. When you push it, it determines the peak values and sets the trigger level to the midpoint of the signal. For more information on this button, see page 3Ć71.
6. The outer knob sets the trigger level Ċ the threshold voltage the signal must cross in order to trigger the instrument. Turn it clockĆ wise to raise the trigger level; turn it counterclockwise to lower the trigger level. For more information on the trigger level see page 3Ć70.
Horizontal Controls
The horizontal controls are on the front panel of the 222PS PowerScĆ out. They appear as shown in Figure 2Ć4.
Figure 2Ć4:ăHorizontal Controls
1. The outer knob sets the seconds per division. This is the horizonĆ tal scale of your display. Turn the knob clockwise to decrease the seconds per division; turn it counterclockwise to increase the seconds per division. Setting the seconds per division is disĆ cussed in more detail on page 3Ć26.
222PS Operator Manual
1
2
2Ć5
Page 34
At a Glance
2. The inner knob sets the horizontal position of the signal. Turn the knob clockwise to move the signal to the right. Turn it counterĆ clockwise to move the signal to the left. For more information on setting the horizontal position see page 3Ć25.
You can also use this knob to magnify the signal by ten times. To do so, see page 3Ć29.

The Display

The 222PS display shows waveforms that represent electrical signals. However, it also shows two other kinds of informationĊreadouts and menus.
Readouts
Readouts are numeric or symbolic information associated with a signal.
The 222PS displays readouts at three places on the screen: along the top, along the bottom, and slightly above the bottom.
Readouts along the top show information associated with the vertical controls. Readoutsalong the bottom show informationassociated with the triggerand horizontal controls. The readoutsjust above them show information associated with saved waveforms. Figures 2Ć5, 2Ć6, and 2Ć7 show these readouts.
2Ć6
Vertical Readouts Ċ Figure 2Ć5 shows the vertical readouts along
the top of the display. The readouts on the left refer to channel 1. The readouts on the right refer to channel 2.
At a Glance
Page 35
At a Glance
213
.... .... ................ ............ ....
6754
Figure 2Ć5:ăVertical Readouts
1. The > indicates that the waveform is uncalibrated. For more information on uncalibrated waveforms, see page 3Ć78.
2. The ~ indicates AC coupling. For more information on coupling, see page 3Ć19.
A
indicates ground coupling.
3. This is the channel 2 coupling. The = indicates DC coupling.
4. The box around the channel information indicates that this chanĆ nel is selected.
5. This number is the volts per division for channel 2 Ċ its vertical scaling. For more details on vertical scaling, see page 3Ć78.
6. This number is volts per division for channel 1.
If either channel is off, the voltsĆperĆdivision number is replaced by an OFF.
7. The downwardĆpointing arrow indicates that the channel is inverted. For more information on inverting a channel, see page 3Ć22.
222PS Operator Manual
NOTE
2Ć7
Page 36
At a Glance
Saved Waveform Readouts Ċ Figure 2Ć6 shows the saved
waveform readouts above the bottomof the display. The readouts refer to the last saved waveform displayed. For more information on saved waveforms, see page 3Ć51.
21
.... .... ................ ............ ....
54 3
Figure 2Ć6:ăSaved Waveform Readouts
2Ć8
1. This indicates that the waveform is uncalibrated. For more inforĆ mation on uncalibrated waveforms see page 3Ć78.
2. This is the channel coupling for the saved waveform. For more information on coupling see page 3Ć19.
3. This is the seconds per division setting for the saved waveform.
4. This is the volts per division setting for the saved waveform.
5. This is the memory location to which the waveform was saved. In this case, the waveform is saved in memory location 1. See page 3Ć51.
Trigger and Horizontal Readouts Ċ Figure 2Ć7 shows the
horizontal and trigger readouts along the bottom of the display.
At a Glance
Page 37
At a Glance
12
54 3
Figure 2Ć7:ăTrigger and Horizontal Readouts
1. The second from the left readout is the trigger slope. A + indiĆ cates that triggering occurs on a rising edge. A - indicates that triggering occurs on a falling edge. For more information on trigger slope, see page 3Ć70.
2. The second from the right readout is the seconds per division Ċ the horizontal scale factor. For more information on horizontal scaling, see page 3Ć26.
3. At the right is the magnification indicator. For more details on the magnification feature, see page 3Ć29.
4. The middle readout indicates that the instrument is in store mode. For more information on store mode, see page 3Ć61.
5. At the left is the trigger level in volts. For more details on trigger level see page 3Ć70.
Menus and Menu Buttons
Menus are lists of choices that you can select in order to perform some action, such as placing the instrument in XY mode or turning off the timeĆout feature.
A number of buttons on the front and top panels of the 222PS invoke menus when pressed. When a menu is on the display, you can select one of its items to perform an action. Figure 2Ć8 illustrates the parts of a menu.
222PS Operator Manual
2Ć9
Page 38
At a Glance
1
4
CHOICE 1NAME:
3
CHOICE 2
CHOICE 3
CHOICE 4
2
Figure 2Ć8:ăParts of a Menu
1. The button labeled CLEAR erases the menu from the display.
2. The menu buttons are next to the screen, along its right edge. Pressing the button next to a menu item performs the action represented by that item.
3. The menu items appear along the right edge of the display. Up to four items can appear on a menu. Each represents a possible action you can perform.
4. The name of the menu appears at the top left of the display, followed by a colon.
2Ć10
At a Glance
Page 39

Top Panel Controls

The top panel for the 222PS PowerScout appears as shown in FigĆ ure 2Ć9.
At a Glance
1
8
7
23
56
4
Figure 2Ć9:ă222PS Top Panel
1. Pressing this button invokes a menu that allows you to recall saved waveforms. See page 3Ć53.
2. Pressing this button toggles the instrument in or out of store mode. For more information about store mode, see page 3Ć61.
3. Pressing this button invokes a menu that allows you to specify the acquisition mode of the instrument. For more information about acquisition modes, see page 3Ć1.
222PS Operator Manual
2Ć11
Page 40
At a Glance
4. Pressing this button invokes a menu that allows you to specify
5. Pressing this button invokes a menu that allows you to execute a
whether to invert a channel, display signals in XY mode, or display signal readouts. For more information about inverting channels, see page 3Ć22. For more information about XY mode, see page 3Ć81. For more information about displaying or clearing signal readouts, see page 3Ć21.
variety of special functions. In some cases, you may execute items from two or three layers of menus.
H You can check the display alignment and view the instrument
identification and firmware number. See page 3Ć9.
H You can start selfĆcalibration routines for either channel or the
external trigger input. See page 3Ć9.
H You can enable or disable the timeĆout feature, set the baud
rate, activate the modem, or select different probe types. For more information about the timeĆout feature, see page 3Ć37. For more information about setting the baud rate or activating the modem, see page AĆ11. For more information about configuring the 222PS for the correct probes, see page 3Ć45.
H You can enable or disable MOTOR TRIG, which places a
special filter in the trigger path to allow triggering on motor drive signals that are pulseĆwidth modulated and on 50/60 Hz line signals. See page NO TAG.
2Ć12
6. Pressing this button invokes a menu that allows you to specify the trigger position. See page 3Ć71.
7. Pressing this button invokes a menu that allows you to save or recall frontĆpanel setups. See page 3Ć55.
8. Pressing this button invokes a menu that allows you to save waveforms. See page 3Ć51.
At a Glance
Page 41

Side Connectors

The right side of the 222PS PowerScout appears as shown in FigĆ ure 2Ć10.
Figure 2Ć10:ăSide of 222PS, Storage Pouch Removed
At a Glance
21
3
1. This is the probe connector for channel 1.
2. This is the probe connector for channel 2.
3. This is thebattery connection. See page 3Ć35 for more information on connecting the battery.
The 222PScomes with an attached storage pouch. Store the probes in the pouch when you are not using them. You do not need to disconĆ nect the probes before you store them.
222PS Operator Manual
2Ć13
Page 42
At a Glance
Rear Panel Controls and Connectors
The rear panel for the 222PS PowerScout appears as shown in FigĆ ure 2Ć11.
1
876
23
Figure 2Ć11:ă222PS Rear Panel
4
5
2Ć14
1. This panel points to the probe inputs on the instrument's side and indicates the 222PS maximum input voltage rating.
2. This is the RSĆ232 connection port for remote communications. For more information on remote communications, see page AĆ9.
3. This is the instrument serial number. You will need it if you must ever arrange to ship the instrument back for maintenance. For more maintenance information, see page 3Ć31.
4. This knob focuses the 222PS screen. Insert a small screwdriver into the slot and turn it to adjust the focus.
At a Glance
Page 43
At a Glance
5. This knob varies the brightness of the 222PS screen. See page 3Ć23 for more information on screen brightness.
6. This is the external trigger common reference connector. To use a grounded reference with your external trigger source, connect the reference signal here. See page 3Ć65.
WARNING
To avoid possible injury, do not connect the trigger comĆ mon reference input to voltages greater than 42ĂV peak. The trigger common reference input is not insulated.
7. This is the external trigger input connector. To use an external signal as a trigger source, connect the external trigger signal here. For more information on external triggering, see page 3Ć65.
WARNING
To avoid possible injury or damage to the 222PS or equipment connected to it, do not float the external trigger common connector, the RSĆ232 communications port, or the external power input above 42ĂV peak. These inputs are not electrically isolated from each other.
8. This is the external power input. Connect the External Power AC Adapter to the input to run the instrument from line voltage. See page 3Ć41 for more information on external power.
222PS Operator Manual
2Ć15
Page 44
At a Glance

The Tilt Stand

The 222PS PowerScout comes with a tilt stand so that you can view the frontĆpanel and screen more easily. The tilt stand folds under the instrument when not in use. To use it, lift the instrument and pull the tilt stand forward until the instrument rests on it.
Figure 2Ć12:ăThe 222PS With Tilt Stand
2Ć16
At a Glance
Page 45
Page 46
Page 47

Acquisition Modes

Acquiring signals involves accepting an analog electrical signal, sampling it, and producing a waveform. The 222PS allows you to specify how the instrument acquires the signal and constructs a waveform. This is the acquisition mode of the instrument.

Kinds of Acquisition Modes

The 222PS provides four acquisition modes: normal, average, enveĆ lope, and continuous envelope. These modes help you to examine and manage noisy signals.
Normal Acquisition Mode
Normal acquisition mode is the most common, and it is the instrument default.
In normal acquisition mode, the instrument displays a waveform with one sample point for each acquired point.
Average Acquisition Mode
Average acquisition mode displays a waveform that is the average of the last four waveforms acquired. This mode is useful for reducing random noise and displaying a cleaner signal.
Average acquisition mode works only when the instrument is triggered. In auto level and autoĆbaseline trigger modes, untriggered displays appear identical to those using normal acquisition mode.
In auto baseline trigger mode, the instrument displays the last waveform acquired; it is unaveraged.
222PS Operator Manual
NOTE
3Ć1
Page 48
Acquisition Modes
Envelope Acquisition Mode
In envelope mode, the instrument displays the positive and negative peak signal values that occur during a display sample interval. This mode is useful for detecting glitches such as unwanted peaks in a signal.
Envelope acquisition mode functions only for time base settings between 20Ăms and 20Ăs per division. If you set the instrument at a faster setting, it functions as if it were in normal acquisition mode.
Envelope mode samples the signal at 10ĂMHz, thereby acquiring many samples for each point it displays.
Because of the 10ĂMHz sampling rate, the instrument cannot detect glitches that last less than 100Ăns.
NOTE
NOTE
3Ć2
Continuous Envelope Acquisition Mode
Continuous envelope mode is similar to envelope mode. The differĆ ence is that continuous mode accumulates and displays peak values until you press the button labeled INIT.
NOTE
Continuous envelope acquisition mode functions only for time base settings between 20Ăms and 20Ăs per division. If you specify a faster setting, the instrument functions as if it were in normal acquisition mode.
Changes to most frontĆpanel control settings also act like the INIT button, discarding the old waveform data and starting the envelope sequence anew. The only frontĆpanel controls that do not affect continuous envelope mode are the horizontal and vertical position knobs and the trigger level knob.
In Detail
Page 49
Figure 3Ć1 summarizes the effects of averaging and envelope modes on a signal.
Figure 3Ć1:ăNormal, Average, and Envelope Signals

Selecting an Acquisition Mode

To choose an acquisition mode, follow these steps.
ĂĂStep 1:ăPress the ACQ button on the top panel to invoke the
acquisition menu.
The normal acquisition mode is boxed (as in Figure 3Ć2) unless you have previously selected another acquisition mode.
Acquisition Modes
ĂĂStep 2:ăPress the button next to the acquisition mode you wish to
select. The acquisition mode takes effect, and the menu disapĆ pears.
Figure 3Ć2:ăThe Acquisition Menu
222PS Operator Manual
NORMACQ:
ENV
AVG
CONT ENV
3Ć3
Page 50
Acquisition Modes
3Ć4
In Detail
Page 51
The 222PS allows you to obtain a readable display of a waveform by pushing a single button. You can also set up the instrument in a specific way and then save the setting in memory (see page 3Ć55). This section discusses various automatic ways to set up the instruĆ ment.

Procedure

If you wish to view a signal quickly, follow these steps.
ĂĂStep 1:ăConnect the probe tip to the signal you wish to see.
ĂĂStep 2:ăPress the button marked AUTO SETUP on the front
panel.
Pressing AUTO SETUP gets a quick, informative display of any signal between 20 Hz and 1 MHz.

Parameter Effects

Auto Setup

The AUTO SETUP button has these effects on setup parameters.
Channels
When you press AUTO SETUP, the instrument first determines which channels to display. It checks both probe tips for a signal on either channel.
H If a channel is receiving a signal, the 222PS turns the channel on.
H If neither channel is receiving a signal, the 222PS turns on chanĆ
nel 1 and turns off channel 2.
222PS Operator Manual
3Ć5
Page 52
Auto Setup
Vertical Scaling
The 222PS then determines the characteristics of the signal, so it can produce a useful display.
The 222PS sets the vertical position to display the signal in the center of the screen. If both channels have a signal, the 222PS displays both signals.
The instrument also sets the volts per division to display each signal with several divisions of amplitude.
Horizontal Scaling
If only one channel is receiving a signal, the 222PS sets the horizontal position to display 1-5 waveform cycles. It determines the peak values and sets the trigger level at the midpoint.
If both channels have a signal, the instrument uses the channel 1 signal to set the seconds per division and trigger level. If the two signals are synchronized, they both appear stable. Otherwise, the channel 2 signal is untriggered.
Low and High Frequencies
3Ć6
AUTO SETUP avoids the timeĆbase modes used with the slower time scales. It therefore does not set the seconds per division to 0.1 s or slower and does not produce a readable display for signals slower than 20ĂHz.
For signals with frequencies above 100ĂkHz, AUTO SETUP always uses a secondsĆperĆdivision setting of 5Ă The 222PS may display such highĆfrequency signals with too many or too few cycles on the screen; therefore, they may appear confusing. You may need to make small corrections to the seconds per division knob to get a useful display.
Table 3Ć1 shows all AUTO SETUP actions.
ms to minimize search time.
In Detail
Page 53
TableĂ3Ć1:ă Auto Setup Settings
Auto Setup
Control
VOLTS/DIV As determined by signal
VOLTS/DIV VAR Calibrated
Coupling AC if in AC before, and if a signal exists; otherwise
Acquisition mode Normal
STORE/NONSTORE NONSTORE
Invert Off (not inverted)
XY Display Off
SEC/DIV
Trigger source Vertical
Trigger mode Auto level
Trigger position Post
Trigger slope Plus
Trigger level Midpoint of signal
X10 MAG Off
Readouts On
Setting After Auto Setup
DC if a signal exists; otherwise OFF
As determined by signal; 5Ăms for signals above 100ĂkHz
Selected channel Channel 1 if signal exists, or if channel 2 has no
signal
NOTE
AUTO SETUP does not disable the motor trigger selection.
222PS Operator Manual
3Ć7
Page 54
Auto Setup
3Ć8
In Detail
Page 55

Calibration

The 222PS has a vertical channel selfĆcalibration routine to maintain best balance and accuracy with temperature variations.
Running the SelfĆCalibration Routine
To achieve the specified performance, you should recalibrate the 222PS any time the ambient temperature has changed by more than 5° C. If the trace jumps when you rotate the volts per division knob with no signal applied, the instrument probably needs recalibration.
To run the selfĆcalibration routine, follow these steps.
ĂĂStep 1:ăDisconnect both probes from the signal source.
NOTE
To ensure that the selfĆcalibration routine produces accuĆ rate results, do not run the selfĆcalibration routine while either probe is connected to a signal source.
ĂĂStep 2:ăPress the AUX FUNCT button on the top panel to display
the auxiliary functions menu. The display appears as shown in Figure 3Ć3.
222PS Operator Manual
3Ć9
Page 56
Calibration
ALIGNAUX:
SELF CAL
CONFIG
MOTOR TRIG
Figure 3Ć3:ăThe Auxiliary Functions Menu
ĂĂStep 3:ăPress the menu button next to the SELF CAL menu item
to access the calibration menu. The display now appears as shown in Figure 3Ć4.
3Ć10
CH1SELF CAL:
CH2
EXT TRIG
PROBE
Figure 3Ć4:ăThe Calibration Menu
NOTE
The PROBE selection on the SELF CAL menu is for use by qualified service personnel only.
In Detail
Page 57
Calibration
ĂĂStep 4:ăTo begin the selfĆcalibration routine for channel 1, press
the menu button next to CH1.
This routine takes about a minute to perform. When it is finished, the instrument displays PASS or FAIL.
If PASS appears on the screen, channel 1's recalibration was successful.
If FAIL appears on the screen, run the calibration routine again. If the failure persists, refer the instrument to a qualified service person.
If the selfĆcalibration routine fails, the instrument sends a coded error message. To capture this message, connect the instrument to a terminal (or a PC emulating a terminal) through the RSĆ232 port on the rear panel and run the failed routine again. See Appendix B: Remote Communications for a discussion of the error codes.
ĂĂStep 5:ăNow calibrate channel 2. Invoke the menu again, and
press themenu button next to CH2. Follow the procedure outlined in Step 4.
ĂĂStep 6:ăNowrecalibrate the external trigger input. Before you start
the calibration, connect the external trigger input connector to the trigger common reference connector on the rear panel. Use a jumper cable with a banana plug connector on each end.
222PS Operator Manual
3Ć11
Page 58
Calibration
NOTE
To calibrate the external trigger input, the TRIG COM connector must be connected to the EXT TRIG INPUT connector.
ĂĂStep 7:ăInvoke the menu again, and press the menu button next
to EXT TRIG. A new menu appears on the display, as shown in Figure 3Ć5.
GND EXT TRIG:
CAL
3Ć12
Figure 3Ć5:ăThe Ground External Trigger Calibration Menu
In Detail
Page 59
Calibration
ĂĂStep 8:ăPress the menu button next to CAL to begin the calibraĆ
tion. Follow the procedure outlined in Step 4.
ĂĂStep 9:ăWhen you are done, press the CLEAR button above the
menu buttons to return the instrument to normal operation.
222PS Operator Manual
3Ć13
Page 60
Calibration
3Ć14
In Detail
Page 61
Capturing Random
Events
Capturing random electric events in circuits is difficult and timeĆconĆ suming. You can use the 222PS's single sequence mode to make the task easier.
SingleĆsequence mode is one of four trigger modes you can use with the 222PS. In singleĆsequence mode, the instrument acquires one triggered signal. It then displays the signal and holds it until you press the button labeled INIT to start the sequence all over again. Changing a control that affects the signal acquisition also starts the sequence again.
The following procedure will help you use singleĆsequence mode to capture a random event.
ĂĂStep 1:ăPreset the instrument controls to display a baseline
signal.
ĂĂStep 2:ăApply a test signal to the channel 1 input to set the trigger
level. Make sure the test signal is the same amplitude and general type (negative or positive pulse or sinusoidal) as the signal you want to trigger on.
ĂĂStep 3:ăPress AUTO SETUP to obtain a quick frontĆpanel control
setup for the test signal. If the resulting vertical or horizontal scaling result is not precisely what you want, adjust the voltsĆperĆ division and secondsĆperĆdivision controls as you wish. If necesĆ sary, reposition the trace vertically.
ĂĂStep 4:ăSet the trigger mode to normal.
ĂĂStep 5:ăSet the trigger source to channel 1.
ĂĂStep 6:ăAchieve a stable display by adjusting the trigger level
control.
ĂĂStep 7:ăTo set the trigger mode to singleĆsequence, Press the
button labeled MODE on the front panel. Press the menu button next to the menu item SSEQ.
222PS Operator Manual
3Ć15
Page 62
Capturing Random Events
ĂĂStep 8:ă Check that the sweep triggers each time the INIT button
is pressed. If it does not, readjust the trigger level control slightly until the sweep triggers each time you press INIT. The TRIG'D indicator lights when the instrument triggers.
ĂĂStep 9:ăDisconnect the test signal from the oscilloscope and
apply the random signal to the input.
ĂĂStep 10:ăPress INIT to arm the trigger system. The instrument
then waits for the trigger event. The TRIG'D indicator lights when the instrument triggers.
ĂĂStep 11:ăAfter the instrument triggers and completes the single
sequence, press INIT again to acquire another signal.
In scroll mode, the singleĆsequence trigger mode is useful for capturĆ ing an event that occurs either randomly or infrequently. If the event is also very narrow, use the envelope mode as well.
Initialize the singleĆsequence function and let the oscilloscope watch for the event. The 222PS acquires data continuously up to the trigger point. When the trigger event occurs, the instrument acquires the data it needs to fill the rest of the display. It then halts the acquisition and displays the waveform (with the captured trigger event) until you press the INIT button again.
3Ć16
In Detail
Page 63
The 222PS has two fully isolated input channels with which you can make floating measurements. With signals of up to 600 VAC you can make measurements as you would with a volt meter.
This section explains how to select a channel, display the signal it acquires, and choose the right channel coupling.

Selecting a Channel

You must select a channel before you can change its settings using the frontĆpanel controls. You can select only one channel at a time. The display indicates which channel is currently selected.
To select a channel, push the appropriate channel button (CH1 or CH2) on the front panel.
Figure 3Ć6 shows the 222PS displaying signals for two channels. Channel 1 is selected.

Channels

input,
RMS
-46.7mV+
Figure 3Ć6:ăChannel One Selected
222PS Operator Manual
5mV=.5V~
mS
2
3Ć17
Page 64
Channels

Displaying a Channel

When a channel is on, the 222PS displays any signal it acquires through that channel. When a channel is off, the 222PS does not display its signal. If either channel is off, the screen displays OFF" instead of its voltsĆperĆdivision value.
Even when a channel is off, the instrument can still use it as a trigger source.
By default, the 222PS displays both channels. To turn a channel off, follow these steps.
ĂĂStep 1:ăIf the channel is not already selected, select it by pushing
the appropriate channel button (CH1 or CH2) on the front panel.
ĂĂStep 2:ă Press the channel button again to display the channel
menu. The screen appears as shown below.
DCCH 2 CPLG:
3Ć18
AC
GND
CH 2 OFF
Figure 3Ć7:ăThe Channel Menu
ĂĂStep 3:ăPress the button next to the bottom menu item,
CH 2 OFF. This turns the selected channel off and clears the
menu from the display. However, until you select channel 1 for another purpose, channel 2 is still the selected channel.
In Detail
Page 65
To display a channel you have turned off, repeat the above procedure until you see the menu on the screen. Then select the type of coupling you wish to use for the channel signal. The following section describes how to select the channel coupling.
When a channel is off, you can still change its vertical settings with the POS, VOLTS/DIV,orVAR VOLTS/DIV controls. First, select the channel. Then make the changes you wish. The changes take effect when you turn the channel back on.

Setting Channel Coupling

There are three possible couplings for each channel:
H DC coupling passes all frequencies of the input signal up to the
useful bandwidth of the instrument.
H AC coupling blocks any DC component of the signal and is the
most commonly used.
Channels
NOTE
H Ground coupling disconnects the input signal and grounds the
input for the selected channel.
To select the coupling for a channel, follow these steps.
ĂĂStep 1:ăIf the channel is not already selected, select it by pushing
the appropriate channel button (CH1 or CH2) on the front panel.
ĂĂStep 2:ăPress the channel button again to invoke the channel
menu. The screen appears as in Figure 3Ć7.
ĂĂStep 3:ăPress the button next to the menu item representing the
type of coupling you wish to use. For example, press the second button from the top to select AC coupling.
If the type of coupling you wish already appears boxed, it is already the selected coupling. Press the CLEAR button to remove the menu from the display.
222PS Operator Manual
3Ć19
Page 66
Channels
3Ć20
In Detail
Page 67
The 222PS display shows you menus, signals, and readouts. This section explains how you can control the information displayed to you.

Readouts

Readouts include volts per division for each signal, seconds per division, and trigger level. The instrument also displays additional information, depending on the characteristics of the signal you display and the mode of the instrument.
Displaying and Clearing the Readouts
Unless you explicitly turn off the readouts, the 222PS displays them. If you wish to turn the readouts off and view only the signals, follow these steps.

The Display

ĂĂStep 1:ăPress the DISPL button on the top panel to access the
display menu. Figure 3Ć8 shows the display menu.
Figure 3Ć8:ăThe Display Menu
222PS Operator Manual
INV1DISPLAY:
INV2
XY
RO OFF
3Ć21
Page 68
The Display
ĂĂStep 2:ăPress the menu button next to the bottom menu item RO
OFF. The menu disappears and the screen displays only the
signal.
ĂĂStep 3:ăIf you wish to turn the readouts back on again, repeat the
above procedure. The next time you access the display menu, the RO OFF item appears boxed, indicating that it is selected and that readouts have therefore been turned off. Press the button again to deselect the menu item. The menu disappears and readouts again appear.

Inverting the Display

You can invert the waveforms displayed for either channel. Figure 3Ć9 shows an example of a normal and an inverted waveform.
Normal
Waveform
Inverted
Waveform
3Ć22
Figure 3Ć9:ăNormal and Inverted Waveforms
Some signal processingcircuits,such as circuits thatprocesscomposĆ ite video signals, automatically invert waveforms. While working on these circuits you may want to invert these signals to view them normally.
To invert a signal, follow these steps.
ĂĂStep 1:ăPress the DISPL button on the top panel to invoke the
display menu. See Figure 3Ć8.
In Detail
Page 69
ĂĂStep 2:ăPress the button next to the menu item corresponding to
the channel you wish to invert (INV1 OR INV2).
After you press the menu button, the menu disappears and the inverted signal (if any) reappears on the screen. A small downĆ arrow appears in front of the volts per division readout for an inverted channel.
ĂĂStep 3:ăIf you wish to display the channel normally, repeat the
above procedure. The next time you access the display menu, the inverted channel item appears boxed, indicating that it is selected. Press the button again to deselect the menu item. The menu disappears, and the channel is no longer inverted.

Varying the Brightness

Use the intensity control on the back panel (labeled INTEN) to adjust the display's brightness.
To changethe brightness, insert a small screwdriver into thecenter slot of the knob and rotate the knob until the display appears as you wish.
The Display
222PS Operator Manual
3Ć23
Page 70
The Display
3Ć24
In Detail
Page 71

Horizontal Operation

This section discussescontrollingthe horizontal aspects ofyoursignal. The knobs used to do this are at the bottom left of the front panel, as shown in Figure 3Ć10.
Figure 3Ć10:ăThe Horizontal Controls
The horizontal controls affect signals acquired through both channels.

Horizontal Positioning

To position the signals horizontally, rotatethe inner knob, labeled POS.
You can position the trace to the right or the left up to oneĆhalf the length of the screen. Readouts are not affected by horizontal positionĆ ing.
For information on the positioning of XY traces, see page 3Ć81.
When X10 magnification is on, the horizontal position control scrolls through the magnified waveform. For more information about magnifyĆ ing waveforms, see page 3Ć29.
The horizontal position control affects the position of a saved waveĆ form. The 222PS displays a saved waveform at the horizontal position presently in effect, not the horizontal position at which it was saved.
222PS Operator Manual
3Ć25
Page 72
Horizontal Operation

Seconds per Division

The 222PS can display a signal with a range of 20Ăs to 50Ăns. If magniĆ fication is on, the instrument's range is 2Ăs to 5Ăns. This means that the 222PS can display a waveform that represents as long a period as 200Ăseconds or as short a period as 50Ăns (in the latter case, with magnification on).
To change the secondsĆperĆdivision setting, turn the outer knob (labeled SEC/DIV). The instrument displays the resulting seconds per division at the bottom of the screen. The 222PS uses a 1-2-5 switchĆ ing sequence: this means that each click of the knob changes the time scale from, for example, 1Ăms to 2Ăms and then to 5Ăms, before going to 10Ă ms.
If you are using battery power at slow time bases, turn off the timeĆout feature described on page 3Ć37. Otherwise, the instrument may time out and turn itself off before it can completely acquire the signal.
NOTE
3Ć26
Aliasing
Aliased waveforms are waveforms that appear to have a frequency much lower than is accurate.
Aliasing can occur when the seconds per division setting, and thereĆ fore the sample rate, is too low to display a high frequency waveform accurately. When this occurs, the instrument does not sample the signal often enough. The resulting waveform it displays is misleading. Figure 3Ć11 illustrates an aliased waveform.
A common symptom of aliasing is an unstable display even when the TRIG'D light is on.
In Detail
Page 73
Horizontal Operation
HighĆFrequency
Apparent LowĆFrequency
Waveform Due to Aliasing
Sampled Points
To avoid aliasing, make sure that your sample rate is fast enough for the frequency of the signal you wish to view Ċ usually, at least twice as fast as the highest frequency component of the signal.
Actual
Waveform
Figure 3Ć11:ăAn Aliased Waveform
TimeĆBase Mode
When you choose a seconds per division setting, the instrument automatically selects the timeĆbase mode required to display the resulting signal. The timeĆbase mode can also depend on the trigger and acquisition modes. The 222PS uses four timeĆbase modes: record, equivalentĆtime, scroll, and scrollĆscan.
Record Ċ For most seconds per division settings, the instrument
uses the timeĆbase mode you are most accustomed to: record. When a trigger occurs, the 222PS acquires and displays a full screen of the waveform in one pass.
EquivalentĆtime Ċ However, when the seconds per division setting
is too fast, the instrument cannot sample fast enough to capture all 512 samples and display them in one pass. Therefore, the instrument depends on successive repetitions of the same waveform to fill the display with samples.
Scroll and ScrollĆscan Ċ When the seconds per division setting is
very slow, the display takes too long to fill using record timeĆbase mode. Therefore, the instrument uses one of scrolling timeĆbase modes.
222PS Operator Manual
3Ć27
Page 74
Horizontal Operation
The instrument uses scroll mode for slow time bases when either autoĆlevel or autoĆbaseline trigger mode is in effect.
The instrument uses scrollĆscan mode for slow time bases when normal or single sequence trigger mode is in effect.
Averaging and continuous envelope acquisitionmodesacquireseveral records of data before displaying a waveform. Therefore, if either of these acquisition modes is in effect, the instrument uses record timeĆbase mode at slow time bases instead of a scrolling mode. Under these circumstances, the display updates slowly.
Table 3Ć2 shows these dependencies and the actual secondsĆperĆdiviĆ sion settings associated with each timeĆbase mode. The boundary between the medium and the slow ranges differs according to whether the instrument is in store or nonstore mode.
TableĂ3Ć2:ă TimeĆBase Modes
Store Mode
On or off
On Off
On
Off
Seconds per Division
Fast: 2Ăms to 50Ăns
Normal: 50Ăms to 5Ăms 20Ăms to 5Ăms
Slow: 20Ăs to 0.1Ăs
20Ăs to 50Ăms
Acquisition Mode
Any EquivalentĆtime
Any Record
Normal or Envelope
Averaging or Continuous Envelope
TimeĆbase Mode
Scroll or ScrollĆscan
RecordĊ slow update
3Ć28
In Detail
Page 75

Magnifying the Signal

You can magnify waveforms by ten times. To do so, push the inner horizontal control knob, labeled POS.
When magnification is on, each division contains five data points horizontally instead of the normal 50. The secondsĆperĆdivision readĆ outs show corresponding values. Magnification also affects saved waveforms on the display and their secondsĆperĆdivision readouts.
To view the rest of the magnified waveform, turn the horizontal position knob. This action allows you to pan through the magnified waveform from side to side.
When you pan through a magnified waveform, you may also be moving the trigger position. The trigger position indicator, shown as a +, cannot move off the screen. Therefore, if you pan the trigger position off the screen to either side, the trigger position indicator remains at the edge of the screen to show the direction of the trigger point.
When magnification is on, a 10X indicator appears at the bottom of the screen, to the right of the seconds per division readout. Figure 2Ć7 shows the bottom readouts.
Horizontal Operation
Displays in XY mode cannot be magnified.
When you magnify displays in scroll and scrollĆscan timeĆbase modes, they update only after the instrument completely acquires the waveĆ form.
To turn off magnification, push the horizontal POS knob again.
222PS Operator Manual
NOTE
3Ć29
Page 76
Horizontal Operation
3Ć30
In Detail
Page 77

Maintenance

The 222PS is covered by a standard Tektronix threeĆyear warranty. If it fails during the warranty period, return it to Tektronix for free servicing (subject to the conditions of the warranty statement).
To arrange for warranty service or get an estimate for outĆofĆwarranty repairs, call 1Ć800ĆTEKĆWIDE (1Ć800Ć835Ć9433).
To help diagnose the problem, have the instrument serial number and firmware version number available. The serial number is located at the top right of the rear panel. To get the firmware identification number, follow the steps below.
If your instrument must be returned for servicing, package it as deĆ scribed on page 3Ć33.

Identifying the Firmware Version

To identify the 222PS firmware version, follow these steps.
ĂĂStep 1:ăPress the AUX FUNCT button on the top panel to access
the auxiliary functions menu. The display now appears as shown in Figure 3Ć12.
222PS Operator Manual
3Ć31
Page 78
Maintenance
ĂĂStep 2:ăPress the menu button next to ALIGN. This calls up the
ALIGNAUX:
SELF CAL
CONFIG
MOTOR TRIG
Figure 3Ć12:ăThe Auxiliary Functions Menu
XY Alignment Menu as shown in figure 3Ć13.
3Ć32
XYALIGN:
IDENT
Figure 3Ć13:ăXY Alignment Menu
ĂĂStep 3:ăPress the menu button next to IDENT. A message apĆ
pears in the middle of the display, identifying the instrument, as shown in Figure 3Ć14. The version number you see may differ.
In Detail
Page 79
TEKĆ222PS VER: x.xx
Figure 3Ć14:ăThe Firmware Version

Repackaging for Shipment

Remember to put the instrument in its carry case before repacking. If the original packing materials are unfit or unavailable, then repackage the instrument in the following manner:
Maintenance
1. Use a corrugated cardboard shipping carton with a test strength of at least 125 kg (275 lb) and an interior size at least 15 cm (6 in) greater than the instrument size in all dimensions. See AppenĆ dix D: Specifications for instrument dimensions.
2. Enclose the following information:
H the owner's name and address
H the name and phone number of a contact person
H the serial number of the instrument
H the reason for returning the instrument
H a complete description of the service required
3. Disconnect the battery before packing the instrument.
4. Completely wrap the instrument with polyethylene sheeting or its equivalent to protect the outside finish and keep harmful subĆ stances out of the instrument.
222PS Operator Manual
3Ć33
Page 80
Maintenance
5. Cushion the instrument on all sides with three inches of padding
6. Seal the shipping carton with an industrial stapler or strapping
7. Call 1Ć800ĆTEKĆWIDE (1Ć800Ć835Ć9433) for shipping instructions.
material or urethane foam, tightly packed between the carton and the instrument.
tape.
3Ć34
In Detail
Page 81
You can operate the 222PS by using the internal battery or by plugĆ ging it into external power. This section explains how to do both. It also explains how to charge and change the battery when necessary.

Battery Operation

The 222PS comes supplied with a battery for use when portable operation is convenient.
Completely recharge the battery as soon as possible after each use of the instrument under battery power.
The instrument is shipped from the factory with the battery charged. However, the battery may not retain its charge while in transit to you. Therefore, we recommend that you charge the battery for three hours before operating the 222PS for the first time.

Power

NOTE
Even when the instrument is off, current trickles slowly from the battery. If the current drawn off in this way deĆ pletes the battery below 7.32ĂV, the instrument cannot start on battery power. If this condition occurs, recharge the battery immediately. Instructions for charging the battery are on pageĂ3Ć37.
Connecting the Battery
The battery is charged at the factory. It is shipped disconnected to prolong its shelf life. To connect the battery, follow these steps.
ĂĂStep 1:ăPlace the oscilloscope on its left side as viewed from the
front panel.
222PS Operator Manual
NOTE
3Ć35
Page 82
Power
ĂĂStep 2:ăOpen the probe pouch and disconnect the probes.
ĂĂStep 3:ăSlide the battery compartment cover and probe pouch
toward the rear of the instrument to unlock the locking tabs.
ĂĂStep 4:ăLift the battery compartment cover and remove it. The
instrument now appears as shown in Figure 3Ć15.
Figure 3Ć15:ăSide View Without Battery Cover
ĂĂStep 5:ăConnect the threeĆwire battery connector to the pins at
the rear of the instrument. The orientation of the connector does not matter.
3Ć36
ĂĂStep 6:ăReplace the battery compartment cover. Position the
cover locking tabs into the matching slots in the battery compartĆ ment. Make sure the locking tabs are all the way in the slots on both the top and the bottom. If the tabs do not seat easily, first seat the top tabs and then press on the bottom of the cover to seat the bottom tabs.
ĂĂStep 7:ăPush forward on the rear of the battery compartment
cover to lock the tabs.
ĂĂStep 8:ăAs soon as possible, charge the battery for three hours.
See the following instructions for charging the battery.
In Detail
Page 83
Power
Charging the Battery
During periods of heavy use in a harsh environment, you will need to recharge the battery after three hours of operation. Under better circumstances, it may need recharging less often. The battery will last longer if you recharge the instrument after each use.
When the battery charge is low, a lowĆbattery indicator the upper right corner of the display. If the battery voltage drops below
7.32ĂV, the instrument automatically turns itself off.
In order to recharge the battery, plug in the External Power AC Adapter and leave the instrument turned off for three hours.
appears in
NOTE
The 222PS battery recharges whenever you plug it into external power. However, it recharges faster if the instruĆ ment is off.
Charging the Battery Externally
You can charge the battery outside the instrument using the external battery charger accessory. See the OptionalAccessories information in Appendix E. You can also use any other charger that supplies 9.8ĂVDC at 20
_C with the supply current limited to 1ĂA. For best results in
various temperatures, thermally compensate the charging voltage by
-10 mV per degree C.
For example, at 50
9.80 V + [(50 - 20)
To charge the battery, follow the steps on the data sheet that comes with the battery charger unit.
_C, the charging voltage is
-10 mV] = 9.50 V
Time Out
An automatic timeĆout feature prevents the battery from losing power when the instrument is unattended for a long period. When you enable the timeĆout feature, the 222PS turns itself off after two minutes of operating under battery power with no changes to the controls.
222PS Operator Manual
3Ć37
Page 84
Power
You can disable the timeĆout feature so that the instrument stays on for the life of the battery's charger. You can also enable the timeĆout feature again when you wish. To do so, follow these steps.
ĂĂStep 1:ăPress the AUX FUNCT button on the top panel to display
the auxiliary functions menu. Figure 3Ć16 shows the display.
ALIGNAUX:
SELF CAL
CONFIG
MOTOR TRIG
Figure 3Ć16:ăThe Auxiliary Functions Menu
3Ć38
ĂĂStep 2:ăPress the menu button next to CONFIG to access the
configuration menu. Figure 3Ć17 shows the configuration menu.
TIME OUTCONFIG:
BAUD
MODEM ON
PROBE
Figure 3Ć17:ăThe Configuration Menu
In Detail
Page 85
Power
ĂĂStep 3:ăThe TIME OUT menu item appears boxed when the
timeĆout feature is enabled. To disable the timeĆout feature press the top menu button.
This menu item is a toggle. Repeating this procedure reĆenables the timeĆout feature and boxes the menu item again.
ĂĂStep 4:ăTo clear the menu from the display, press the button
labeled CLEAR.
Replacing the Battery
If you use the instrument on battery power often, you may wish to obtain and charge extra batteries to take with you. Then, when the battery charge inside the instrument gets low, you can switch to one of the fully charged spare batteries.
NOTE
Because the 222PS loses saved data after 30 s without power, have the spare battery handy before beginning this procedure.
To replace the battery, follow these steps:
ĂĂStep 1:ăOpen the battery compartment. Follow the procedure on
pageĂ3Ć35.
ĂĂStep 2:ăDisconnect the battery from the threeĆwire battery conĆ
nector.
ĂĂStep 3:ăLift the battery pack out of the battery compartment.
ĂĂStep 4:ăPlace the charged replacement battery into the battery
compartment with the battery leads on the bottom facing toward the threeĆwire battery connector.
ĂĂStep 5:ăConnect the battery to the threeĆwire battery connector.
ĂĂStep 6:ăClose the battery compartment. Follow the procedure on
pageĂ3Ć35.
ĂĂStep 7:ăRecharge the low battery as soon as possible. See the
procedure on page 3Ć37.
222PS Operator Manual
3Ć39
Page 86
Power
Storing the Instrument
When storing the instrument for a period shorter than two months, leave the battery connected. When the instrument is off, the current drawn from the battery is less than 1ĂmA. With the battery in place, waveform settings and front panel setups remain in memory; they are available when you turn the oscilloscope on again.
When storing the instrument for a period longer than two months, follow these steps to extend the life of your battery and instrument.
ĂĂStep 1:ăCharge the battery fully. Follow the instructions on
pageĂ3Ć37.
ĂĂStep 2:ăRemove the battery from the instrument. Use the proceĆ
dure on pageĂ3Ć35 to remove the battery cover.
ĂĂStep 3:ăStore the fully charged battery in a cool place.
Deep Discharge
Under certain circumstances, the battery can become deeply disĆ charged. When in this state, the battery accepts a charge very slowly. In some cases, it may not accept a charge at all.
3Ć40
A deep discharge condition is caused by three situations:
H using the instrument until the battery charge is low and then
storing it without recharging it
H storing the battery in a discharged state
H storing the instrument for over two months without removing the
battery
If the battery becomes deeply discharged, you may be able to recover it with the following procedure.
ĂĂStep 1:ăCharge the battery for 24Ăhours. Follow the instructions
on pageĂ3Ć37.
ĂĂStep 2:ăIf the battery does not accept the charge, remove it from
the instrument and try again to charge it using a 20ĂV power supply that is currentĆlimited to 100ĂmA.
In Detail
Page 87
ĂĂStep 3:ăDuring this operation, check the power supply frequently
for a currentĆlimited state. If the battery recovers from its state of deep discharge, it will cause the power supply to currentĆlimit. Do not leave the battery connected to the external power supply without checking it frequently.
ĂĂStep 4:ăIf the power supply shows that it is currentĆlimited,
reinstall the battery in the instrument.
ĂĂStep 5:ăContinue to recharge the battery.
ĂĂStep 6:ăIf the battery does not recover, return it to Tektronix for
safe disposal or dispose it in accordance with local environmental regulations.

External Power Operation

The 222PS has an external power input connector so that it need not use the battery power. You can connect the 222PS to a wall socket using the External Power AC Adapter or you can use your own exterĆ nal power source.
Power
You can also operate the instrument on external power without the battery present. For instructions on removing the battery, see pageĂ3Ć39.
AC Line Operation
The 222PS comes with an External Power AC Adapter. This adapter converts AC line voltage to the 16-20ĂVAC input voltage that the instrument requires. The adapter also recharges the 222PS's battery.
In order to maintain the battery charge for times when you require portable operation, we recommend that you use the External Power AC Adapter whenever practical.
In order to operate the instrument from line power, follow these steps.
ĂĂStep 1:ăPlug the jack end of the External Power AC Adapter into
the external power input on the rear panel of the instrument.
222PS Operator Manual
NOTE
3Ć41
Page 88
Power
ĂĂStep 2:ăPlug the prong end of the External Power AC Adapter into
an AC power source.
ĂĂStep 3:ăPress the ON button.
When the oscilloscope is operating under external power, an external power indicator
the display.
appears in the upper right corner of
Other Sources of External Power
You can operate the 222PS from your own external power source. The power source must supply at least 15ĂW or 16ĂvoltĆamperes.
H An AC power source must provide 16-20ĂVAC at 47-400 ĂHz.
H A DC power source must provide 12-28ĂVDC.
The external power input connector has two contacts. DC power of either polarity can be between contacts.
NOTE
To prevent blowing internal fuses, do not force either pin lower than .5 volts more negative than the instrument chassis. The instrument chassis is connected to the ground pin of the RSĆ232 communications port and to the external trigger connector common.
3Ć42
WARNING
To avoid possible injury or damage to the 222PS or equipment connected to it, do not float the external trigger common connector, the RSĆ232 communications port, or the external power input above 42ĂV peak. These inputs are not electrically isolated from each other.
In Detail
Page 89

Probes

The 222PS comes with two P850 10X probes. An additional 1X probe, the P400 probe, is available as an optional accessory.
The P850 probes provide high attenuation so that you can scale signals approaching 600 VAC useful for measuring sensitive highĆimpedance electronic circuits or highĆvoltage divider circuits. The optional P400 probes measure lowĆlevel signals requiring high sensitivity.
To prevent improper operation and the risk of electric shock, use only Tektronix P400 or P850 probes with this instrument.
Actual probe attenuation factors are 3X for the P400 probe and 30X for the P850 probe. The instrument is calibrated to compensate for these attenuation factors. Other probes or input devices will therefore give incorrect amplitude displays.
for better display. They are also
RMS
CAUTION
The optional P400 1X probe limits the maximum deflection factor of the 222PS to 50Ăvolts per division. It decreases the probe tip input impedĆ ance to 1ĂMW and is sufficient for minimal loading of sensitive circuits. The P400 probe also limits input frequency to 20 MHz.
To avoid personal injury or damage to the 222PS or the probes, do not apply more than 850ĂV peak between probe tip and earth ground, between probe tip and probe common, or between probe common and earth ground.
222PS Operator Manual
WARNING
3Ć43
Page 90
Probes

Connecting the Probes

The input connectors for the probes are inside the pouch over the battery compartment, on the right side of the oscilloscope as you face the screen. You must unzip or remove the pouch to access the conĆ nectors.
Figure 3Ć18:ăSide View Showing Probe Connectors
(Pouch Removed)
Probe Connectors
3Ć44
To connect a probe, place its jack end into a channelĂinput connector. Press until you feel the probe is firmly seated.
You do not need to disconnect the probes before storing them in the pouch.
CAUTION
The exposed probe tips are sharp for probing through solderĆresin and oxide layers. When placing the probes in the side pouch, store them with the retractable hook tip attached to prevent unnecessary damage to the pouch.
In Detail
Page 91

Configuring the Probes

To ensure that the 222PS is operating with the correct settings, confiĆ gure the instrument to match the probes you are using.
To set the probe configuration, follow these steps.
ĂĂStep 1:ăPress the AUX FUNCT button on the top panel to display
the auxiliary functions menu, as shown in Figure 3Ć19.
Probes
ALIGNAUX:
SELF CAL
CONFIG
MOTOR TRIG
Figure 3Ć19:ăThe Auxiliary Functions Menu
ĂĂStep 2:ăPress the menu button next to the CONFIG menu item to
access the configuration menu. The display now appears as shown in Figure 3Ć20.
222PS Operator Manual
3Ć45
Page 92
Probes
TIME OUTCONFIG:
BAUD
MODEM ON
PROBE
Figure 3Ć20:ăThe Configuration Menu
ĂĂStep 3:ăPress the menu button next to the PROBE menu item to
access the probe menu, as shown in Figure 3Ć21.
3Ć46
CH 1 1XPROBE:
CH 1 10X
CH 2 1X
CH 2 10X
Figure 3Ć21:ăThe Probe Menu
In Detail
Page 93
Probes
ĂĂStep 4:ăBoxes appear around the items that represent the current
probe configuration. The default configuration for the 222PS assumes 10X probes on both channels. Therefore, those menu items appear boxed unless you have already changed the probe configuration.
Press the menu button or buttons next to the menu items correĆ sponding to the configuration you need. If the current probe settings are appropriate, you need not press any buttons.
ĂĂStep 5:ăAfter you have configured the 222PS for the correct
probes, press the CLEAR button to remove the menu from the display.
NOTE
The probe configurations are in the 222PS memory. They remain there until you change them again or until the memory loses power. If the probe configuration is lost, it returns to the default value of 10X probes for both chanĆ nels.

Probe Accessories

The P850 probes come with four accessories (shown in Figure 3Ć22):
H a retractable hook tip
H an IC lead protection shroud
H a detachable probe common lead
H two cableĆmarker rings
When removing the hookĆtip assembly from the probe, you can accidentally disconnect the probe body from the probe cable. If this occurs, no signal can pass from the probe to the oscilloscope. To reconnect the probe body to the cable, insert the connector at the end of the cable into the probe body until it seats firmly.
222PS Operator Manual
NOTE
3Ć47
Page 94
Probes
1
2
3Ć48
Probe
34
Figure 3Ć22:ăProbe and Accessories
In Detail
Page 95
Probes
1. You can use the detachable probe common lead shown in FigĆ ure 3Ć22 to connect the oscilloscope input common to the referĆ ence point of the circuit being tested. The probe common lead is not chassis ground and you can connect it to an active circuit component. You can therefore make a floating measurement across a component, with neither point connected to ground potential.
WARNING
To avoid personal injury or damage to the 222PS or the probes, do not apply more than 850 V peak between probe tip and earth ground, between probe tip and probe common, or between probe common and earth ground.
2. You can snap the colored cable marker rings shown in FigĆ ure 3Ć22 into the grooves on the probe cable to distinguish between the probes connected to channel 1 and channel 2.
3. When testing integrated circuit devices, remove the hook tip and use the IC lead protection shroud on the probe tip. The shroud shown in Figure 3Ć22 exposes the sharp probe tip, but it prevents the probe from creating a short circuit across adjacent IC leads.
4. The hook tip shown in Figure 3Ć22 can connect to accessible test points such as a component lead or test point connector. This accessory frees your hands for other tasks.
222PS Operator Manual
3Ć49
Page 96
Probes
3Ć50
In Detail
Page 97
Saving and
Recalling Data
The 222PS has memory to store saved waveforms and frontĆpanel setups. Data in this memory remains as long as the instrument has access to one of these power sources:
H a battery with a charge greater than 7.32ĂV (see page 3Ć35 for a
discussion of battery operation)
H line voltage coming through a power cord plugged into a wall
socket through the external AC adapter (see page 3Ć41 for a discussion of line operation)
H power coming in from another external power source (see
pageĂ3Ć42 for a description of acceptable external power sources)
The instrument does not need to be on for it to access power for the memory. However, if you turn off the instrument, disconnect it from any external power source, and remove its battery, it will lose any data in memory after 30 s.

Waveforms

A saved waveform is a record of a single acquisition cycle. Think of it as a snapshot of a waveform. The 222PS lets you save up to four waveforms in memory. It can recall these at any time.
Saving a Waveform
Use these steps to save a waveform.
ĂĂStep 1:ăSelect the channel whose waveform you wish to save or,
if you wish, put the instrument in XY mode.
ĂĂStep 2:ăUsing the selected channel, acquire and display the
waveform you wish to save.
ĂĂStep 3:ăPosition the waveform where you wish it to be saved.
222PS Operator Manual
3Ć51
Page 98
Saving and Recalling Data
ĂĂStep 4:ăPress the SAVE button on the top panel. Acquisition
stops, the screen freezes, and a menu appears as shown in Figure 3Ć23.
Figure 3Ć23:ăSaved Waveforms Menu
If the waveform you are saving uses channel 1, the menu name is SAVE CH1 WV:. If the waveform you are saving uses channel 2, the menu name is SAVE CH2 WV:. If the waveform you are saving uses XY mode, the menu name is SAVE XY WV:.
1SAVE CH1 WV:
2
3
4
3Ć52
The example in Figure 3Ć23 assumes a waveform acquired using channel 1.
ĂĂStep 5:ăPress the button next to the memory location in which
you wish to save the waveform.
NOTE
You can save only one waveform to each memory location; the 222PS can store only four waveforms. If you choose a memory location that already holds a waveform, the instrument replaces it with the one you are presently saving.
The 222PS displays readouts of the parameters of the saved waveform at the bottom of the screen, as shown in Figure 3Ć24.
In Detail
Page 99
Saving and Recalling Data
In Figure 3Ć24, the waveform has been saved to memory locaĆ tion 1 and therefore named W1. Its vertical setting is 10 mV per division, its coupling is DC (as indicated by the =), and its horiĆ zontal setting is 10 ms per division.
1SAVE CH1 WV:
2
3
W1 10 mV = 10 ms
4
Figure 3Ć24:ăSaved Waveform Parameters
A waveform saved in XY mode shows the scale factor for the x axis (channel 1)in the position of thevolts per division setting. The scale factor for the y axis (channel 2) appears in the position of the seconds per division setting.
ĂĂStep 6:ăTo clear the menu, press the button labeled CLEAR or
invoke another menu. Changing one of the following frontĆpanel controls also clears the menu: volts per division, seconds per division, X10 magnification, or autolevel.
ĂĂStep 7:ăAfter clearing the menu, the 222PS continues to display
the saved trace and its readout. To clear them, press the button labeled CLEAR again.
Recalling a Saved Waveform
To display a saved waveform, follow these steps.
ĂĂStep 1:ăPress the RCL button on the top panel. A menu appears
as shown in Figure 3Ć25.
222PS Operator Manual
3Ć53
Page 100
Saving and Recalling Data
Figure 3Ć25:ăThe Recall Waveforms Menu
ĂĂStep 2:ăPress the button next to the memory location that holds
the waveform you wish to view. The menu disappears and the 222PS displays the waveform at the same position in which it was saved, with the same parameters in effect. The instrument also displays readouts of these parameters at the bottom of the screen.
1RECALL WVFRM:
2
3
4
3Ć54
If the instrument displays two saved waveforms, the last one you recall is the one whose parameters appear on the screen.
If the 222PS is already displaying a saved waveform, the memory location that holds it appears boxed in the menu. If you press a button next to a memory location holding an already displayed waveform, the waveform disappears from the screen.
If you press a button corresponding to an empty memory locaĆ tion, the instrument beeps and the menu remains displayed.
ĂĂStep 3:ăTo clear the menu press the CLEAR button or invoke
another menu.
ĂĂStep 4:ăAfter you clear the menu, the instrument continues to
display the recalled waveform. To clear the recalled waveform, press the CLEAR button again. This action clears all recalled waveforms.
ĂĂStep 5:ăTo erase a waveform from a memory location, save
another waveform to the same location.
In Detail
Loading...