222PS
PowerScout
Power Systems Oscilloscope
Operator Manual
070-8097-02
Please check for change information at the rear
of this manual.
Second Edition: December 1993
Last Revised: September 29, 1994
Page 2
Instrument Serial Numbers
Each instrument manufactured by Tektronix has a serial number on a panel
insert or tag, or stamped on the chassis. The first letter in the serial number
designates the country of manufacture. The last five digits of the serial number
are assigned sequentially and are unique to each instrument. Those
manufactured in the United States have six unique digits. The country of
manufacture is identified as follows:
B010000Tektronix, Inc., Beaverton, Oregon, USA
E200000Tektronix United Kingdom, Ltd., London
J300000Sony/Tektronix, Japan
H700000Tektronix Holland, NV, Heerenveen, The Netherlands
Instruments manufactured for Tektronix by external vendors outside the United
States are assigned a two digit alpha code to identify the country of manufacture
(e.g., JP for Japan, HK for Hong Kong, IL for Israel, etc.).
Tektronix, Inc., P.O. Box 500, Beaverton, OR 97077
Printed in U.S.A.
Copyright
are covered by U.S. and foreign patents, issued and pending. The following are
registered trademarks: TEKTRONIX, TEK, TEKPROBE, and SCOPEĆMOBILE.
E Tektronix, Inc., 1991, 1993. All rights reserved. Tektronix products
Page 3
WARRANTY
Tektronix warrants that this product will be free from defects in materials
and workmanship for a period of three (3) years from the date of
shipment. If any such product proves defective during this warranty
period, Tektronix, at its option, either will repair the defective product
without charge for parts and labor, or will provide a replacement in
exchange for the defective product.
In order to obtain service under this warranty, Customer must notify
Tektronix of the defect before the expiration of the warranty period and
make suitable arrangements for the performance of service. Customer
shall be responsible for packaging and shipping the defective product to
the servicecenter designated by Tektronix, with shipping charges prepaid.
Tektronix shall pay for the return of the product to Customer if the
shipment is to a location within the country in which the Tektronix service
center is located. Customer shall be responsible for paying all shipping
charges, duties, taxes, and any other charges for products returned to any
other locations.
This warranty shall not apply to any defect, failure or damage caused by
improper use or improper or inadequate maintenance and care. Tektronix
shall not be obligated to furnish service under this warranty a) to repair
damage resulting from attempts by personnel other than Tektronix
representatives to install, repair or service the product; b) to repair
damage resulting from improper use or connection to incompatible
equipment; or c) to service a product that has been modified or integrated
with other products when the effect of such modification or integration
increases the time or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THIS
PRODUCT IN LIEU OF ANY OTHER WARRANTIES, EXPRESSED OR
IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY IMPLIED
WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A
PARTICULAR PURPOSE. TEKTRONIX' RESPONSIBILITY TO REPAIR
OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND
EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH
OF THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE
LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR
CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER
TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE
POSSIBILITY OF SUCH DAMAGES.
Page 4
Certificate of the Manufacturer/Importer
We hereby certify that the
SYSTEMS OSCILLOSCOPE AND ALL INSTALLED OPTIONS
222PS PowerScout POWER
complies with the RF Interference Suppression requirements
of Amtsbl.ĆVfg 1046/1984.
The German Postal Service was notified that the equipment is
being marketed.
The German Postal Service has the right to reĆtest the series
and to verify that it complies.
TEKTRONIX
Bescheinigung des Herstellers/Importeurs
Hiermit wird bescheinigt, daß der/die/das
POWER SYSTEMS OSCILLOSCOPE AND ALL INSTALLED OPTIONS
222PS PowerScout
in Übereinstimmung mit den Bestimmungen der AamtsblattĆ
Verfugüng 1046/1984 funkentstört ist.
Der Deutschen Bundespost wurde das Inverkehrbringen
dieses Gerätes angezeigt und die Berechtigung zur ÜberĆ
prufüng der Serie auf Einhalten der Bestimmungen einĆ
geräumt.
TEKTRONIX
Page 5
NOTICE to the user/operator:
The German Postal Service requires that Systems assembled
by the operator/user of this instrument must also comply with
Postal Regulation, Vfg. 1046/1984, Par. 2, Sect. 1.
HINWEIS für den Benutzer/Betreiber:
Die vom Betreiber zusammengestellte Anlage, innerhalb
derer dies Gerät eingesetzt wird, muß ebenfalls den
Voraussetzungen nach Par. 2, Ziff. 1 der Vfg. 1046/1984
genugen.
NOTICE to the user/operator:
The German Postal Service requires that this equipment,
when used in a test setup, may only be operated if the
requirements of Postal Regulation, Vfg. 1046/1984, Par. 2,
Sect. 1.7.1 are complied with.
HINWEIS für den Benutzer/Betreiber:
Dies Gerät darf in Meßaufbauten nur betrieben werden, wenn
die Voraussetzungen des Par. 2, Ziff. 1.7.1 der Vfg. 1046/1984
eingehalten werden.
Page 6
Page 7
Welcome
This manual contains the following sections:
HOverview describes the 222PS PowerScout and provides safety
information.
HAt A Glance describes the controls and connectors for the 222PS.
HIn Detail provides further detail on some aspects of the 222PS,
building on the information contained in At A Glance. The 15
topics of this chapter are in alphabetical order for your conveĆ
nience:
HAcquisition Modes
HAuto Setup
HCalibration
HCapturing Random Events
HChannels
HThe Display
HHorizontal Operation
HMaintenance and Repair
HPower
HProbes
HSaving and Recalling Data
HStore Mode
HTriggering
HVertical Operation
HXY Mode
HTutorial: Measuring Signals provides stepĆbyĆstep instructions to
get you started making measurements quickly.
222PS Operator Manual
i
Page 8
HRemote Communication provides information on RSĆ232 commuĆ
nication procedures between the 222PS and a PC.
HPerformance Verification describes the procedures necessary to
verify that the 222PS is performing according to specifications.
HSpecifications provides complete specifications for the 222PS
PowerScout.
HAccessories describes the standard and optional accessories
available for the 222PS.
HGlossary defines various words used in the text.
HThe Index helps you locate information quickly.
NOTE
If you have never used an oscilloscope before, please
read the tutorial in Appendix A before using the 222PS.
ii
Welcome
Page 9
List of Illustrationsix.....................................
List of Tablesxiii..........................................
Overview
About the 222PS PowerScout1Ć1......................
TableĂAĆ28:ăExternal Power AC Adapter OptionsAĆ81...........
xiv
Contents
Page 21
Page 22
Page 23
This section summarizes the product features and safety precautions
of the 222PS PowerScout.
About the 222PS PowerScout
The Tektronix 222PS PowerScout is a portable digitizing oscilloscope.
It has two fully isolated, independently floatable channels rated to
600 VAC
oscilloscopes in the world with this feature.
The 222PS also has these features.
Hlight weight with a battery power source for field operations
Hautomatic setup button
Hautomatic triggering modes
Haveraging and enveloping acquisition modes
Hremote operationcapabilities via theRSĆ232 communications port
Hinternal memory for saving up to four waveforms and four instruĆ
ment setups
. The 222PS/224 digitizing oscilloscopes are the only
RMS
Overview
HDCĆtoĆ10ĂMHz signal bandwidth
H10ĂMS/s digitizing rate
H1 MHz single pass storage bandwidth
HMotor trigger
If you need more information about your Tektronix 222PS PowerScout
or other Tektronix products, contact the nearest Tektronix sales office
or distributor, consult the Tektronix product catalog, or, in the U. S., call
the Tektronix National Marketing Center tollĆfree at 1Ć800Ć426Ć2200.
222PS Operator Manual
1Ć1
Page 24
Overview
Safety
You may be eager to begin using your Tektronix 222PS but please take
a moment to review these safety precautions. We provide them for
your protectionand to prevent damage to the 222PS PowerScout. This
safety information applies to all operators and service personnel.
WARNING
To avoid personal injury or damage to the 222PS, do not
apply more than 850ĂV peak between probe tip and earth
ground, between probe tip and probe common, or beĆ
tween probe common and earth ground.
WARNING
To avoid injury, use caution when working on equipment
with voltages above 42ĂV peak. Such voltages pose a
shock hazard.
1Ć2
WARNING
Do not float the external trigger common connector, the
RSĆ232 communications port, or the external power input
above 42ĂV peak. These inputs are not electrically isolated
from each other.
Symbols and Terms
These two terms appear in manuals:
H
result in damage to the equipment or other property.
H
result in personal injury or loss of life.
These two terms appear on equipment:
statements identify conditions or practices that could
statements identify conditions or practices that could
Overview
Page 25
Overview
HCAUTION indicates a personal injury hazard not immediately
accessible as one reads the marking or a hazard to property
including the equipment itself.
HDANGER indicates a personal injury hazard immediately accessiĆ
ble as one reads the marking.
This symbol appears in manuals:
StaticĆSensitive Devices
These symbols appear on equipment:
DANGER
High Voltage
Protective
ground (earth)
terminal
ATTENTION
Refer to
manual
Specific Precautions
Observe all these precautions to ensure your personal safety and to
prevent damage either to the 222PS or to equipment connected to it.
Power Source Ċ The 222PS can use its selfĆcontained sealed lead
acid battery as a power source. It can also operate using power
supplied to the external power input. Power supplied to this input must
be 12 to 28ĂVDC or 16 to 20ĂVAC
power conductor negative by more than 0.5ĂV with respect to chassis
ground. Both conductors of the external power input are fused interĆ
nally. These fuses are not user accessible.
You can operate the instrument with external power operation from
local 110ĂV or 240ĂV power supply using the appropriate external
power AC adapter. Use only external power AC adapters specified for
this instrument.
. Do not force either external
RMS
222PS Operator Manual
1Ć3
Page 26
Overview
Grounding the PowerScout Ċ The channel 1 and channel 2
measurement inputs of the 222PS are doubly insulated from each
other and all other accessible portions of the instrument cabinet. It is
not necessary to ground the instrument to avoid electric shock.
Fuse Ċ The 222PS has no userĆreplaceable fuses.
Do Not Disassemble the Cabinet Ċ To avoid personal injury, do
not operate the instrument without a properly assembled cabinet. The
cabinet of the instrument should be disassembled only by qualified
service personnel.
Do Not Operate in Explosive Atmospheres Ċ The 222PS
provides no explosion protection from static discharges or arcing
components. Do not operate the 222PS in an atmosphere of explosive
gasses.
Electric Overload Ċ Never apply a voltage to a probe or connector
on the 222PS that is outside the range specified for that probe or
connector.
1Ć4
Overview
Page 27
Page 28
Page 29
At a Glance
This chapter describes the controls, connectors, and display readouts
of the 222PS PowerScout. It is intended to help orient you and to
provide basic information. For more detailed operating instructions for
various features, see the appropriate section in the chapter entitled In
Detail.
This section provides page references to the In Detail chapter for
further information.
222PS Operator Manual
2Ć1
Page 30
At a Glance
Front Panel Controls
The front panel for the 222PS PowerScout appears as shown in
Figure 2Ć1.
1
2
3
4
65
2Ć2
Figure 2Ć1:ă222PS Front Panel
1.The display area includes the screen and associated buttons. The
screen shows signal traces, readouts, and menu items. Buttons
along the side of the screen allow you to manipulate menus. See
page 2Ć6 for a more complete description of the display area.
2.The vertical controls allow you to manipulate the vertical aspects
of your signal. See page 3Ć77 for a more complete description of
the vertical controls.
3.The trigger controls allow you to manipulate the aspects of your
signal having to do with triggering. See page 3Ć63 for a more
complete description of the trigger controls.
4.The horizontal controls allow you to manipulate the horizontal
aspects of your signal. See page 3Ć25 for more information on the
horizontal controls.
At a Glance
Page 31
At a Glance
5.AUTO SETUP allows you to set up the instrument with the push
of a single button. Press this button for a quick, informative
display of any signal between 20 Hz and 1 MHz. See page 3Ć5 for
more information on the AUTO SETUP button.
6.The ON button toggles the instrument on or off. The instrument
beeps when it is turned on.
Vertical Controls
Figure 2Ć2 shows the vertical controls that are located on the front
panel.
321
Figure 2Ć2:ăVertical Controls
1.The outer knob sets the volts per division, which is the vertical
scale of your display. Turn the knob clockwise to decrease the
volts per division and counterclockwise to increase the volts per
division. See page 3Ć78 for more information on setting the volts
per division.
2.The inner knob sets the vertical position of the signal. Turn the
knob clockwise to move the signal towards the top of the screen
and counterclockwise to move the signal towards the bottom of
the screen. For more information on setting the vertical position
see page 3Ć77.
222PS Operator Manual
2Ć3
Page 32
At a Glance
3.The 222PS can display signals acquired through either or both of
Trigger Controls
The trigger controls are on the front panel of the 222PS PowerScout.
They appear as shown in Figure 2Ć3.
You can also use this knob to change the size of a signal to an
arbitrary number of divisions. To do so, see page 3Ć78.
its two channels. The channel selector buttons allow you to select
the channel that is affected by changes to the controls. With these
buttons you can also select channel coupling or turn a channel off
so that the signal it is measuring is not displayed. For more
information on channels, see page 3Ć17.
12
2Ć4
3456
Figure 2Ć3:ăTrigger Controls
1.When you push the button labeled SLOPE, you toggle between a
positiveĆ and negativeĆtrigger slope. For more details, see
page 3Ć70.
2.When you push the button labeled MODE, you invoke a menu
that allows you to specify the trigger mode. For a complete
explanation of trigger modes, see page 3Ć72.
At a Glance
Page 33
At a Glance
3.When you push the button labeled SOURCE, you invoke a menu
that allows you to specify the trigger source. For a complete
explanation of trigger sources, see page 3Ć64.
4.The light labeled TRIG'D turns on when the instrument is trigĆ
gered. See page 3Ć63 for more details.
5.The inner button, labeled AUTOLVL: PUSH, sets the trigger level
automatically. When you push it, it determines the peak values
and sets the trigger level to the midpoint of the signal. For more
information on this button, see page 3Ć71.
6.The outer knob sets the trigger level Ċ the threshold voltage the
signal must cross in order to trigger the instrument. Turn it clockĆ
wise to raise the trigger level; turn it counterclockwise to lower the
trigger level. For more information on the trigger level see
page 3Ć70.
Horizontal Controls
The horizontal controls are on the front panel of the 222PS PowerScĆ
out. They appear as shown in Figure 2Ć4.
Figure 2Ć4:ăHorizontal Controls
1.The outer knob sets the seconds per division. This is the horizonĆ
tal scale of your display. Turn the knob clockwise to decrease the
seconds per division; turn it counterclockwise to increase the
seconds per division. Setting the seconds per division is disĆ
cussed in more detail on page 3Ć26.
222PS Operator Manual
1
2
2Ć5
Page 34
At a Glance
2.The inner knob sets the horizontal position of the signal. Turn the
knob clockwise to move the signal to the right. Turn it counterĆ
clockwise to move the signal to the left. For more information on
setting the horizontal position see page 3Ć25.
You can also use this knob to magnify the signal by ten times. To
do so, see page 3Ć29.
The Display
The 222PS display shows waveforms that represent electrical signals.
However, it also shows two other kinds of informationĊreadouts and
menus.
Readouts
Readouts are numeric or symbolic information associated with a
signal.
The 222PS displays readouts at three places on the screen: along the
top, along the bottom, and slightly above the bottom.
Readouts along the top show information associated with the vertical
controls. Readoutsalong the bottom show informationassociated with
the triggerand horizontal controls. The readoutsjust above them show
information associated with saved waveforms. Figures 2Ć5, 2Ć6, and
2Ć7 show these readouts.
2Ć6
Vertical Readouts Ċ Figure 2Ć5 shows the vertical readouts along
the top of the display. The readouts on the left refer to channel 1. The
readouts on the right refer to channel 2.
At a Glance
Page 35
At a Glance
213
.... ....................................
6754
Figure 2Ć5:ăVertical Readouts
1.The > indicates that the waveform is uncalibrated. For more
information on uncalibrated waveforms, see page 3Ć78.
2.The ~ indicates AC coupling. For more information on coupling,
see page 3Ć19.
A
indicates ground coupling.
3.This is the channel 2 coupling. The = indicates DC coupling.
4.The box around the channel information indicates that this chanĆ
nel is selected.
5.This number is the volts per division for channel 2 Ċ its vertical
scaling. For more details on vertical scaling, see page 3Ć78.
6.This number is volts per division for channel 1.
If either channel is off, the voltsĆperĆdivision number is
replaced by an OFF.
7.The downwardĆpointing arrow indicates that the channel is
inverted. For more information on inverting a channel, see
page 3Ć22.
222PS Operator Manual
NOTE
2Ć7
Page 36
At a Glance
Saved Waveform Readouts Ċ Figure 2Ć6 shows the saved
waveform readouts above the bottomof the display. The readouts refer
to the last saved waveform displayed. For more information on saved
waveforms, see page 3Ć51.
21
.... ....................................
543
Figure 2Ć6:ăSaved Waveform Readouts
2Ć8
1.This indicates that the waveform is uncalibrated. For more inforĆ
mation on uncalibrated waveforms see page 3Ć78.
2.This is the channel coupling for the saved waveform. For more
information on coupling see page 3Ć19.
3.This is the seconds per division setting for the saved waveform.
4.This is the volts per division setting for the saved waveform.
5.This is the memory location to which the waveform was saved. In
this case, the waveform is saved in memory location 1. See page
3Ć51.
Trigger and Horizontal Readouts Ċ Figure 2Ć7 shows the
horizontal and trigger readouts along the bottom of the display.
At a Glance
Page 37
At a Glance
12
543
Figure 2Ć7:ăTrigger and Horizontal Readouts
1.The second from the left readout is the trigger slope. A + indiĆ
cates that triggering occurs on a rising edge. A - indicates that
triggering occurs on a falling edge. For more information on
trigger slope, see page 3Ć70.
2.The second from the right readout is the seconds per division Ċ
the horizontal scale factor. For more information on horizontal
scaling, see page 3Ć26.
3.At the right is the magnification indicator. For more details on the
magnification feature, see page 3Ć29.
4.The middle readout indicates that the instrument is in store mode.
For more information on store mode, see page 3Ć61.
5.At the left is the trigger level in volts. For more details on trigger
level see page 3Ć70.
Menus and Menu Buttons
Menus are lists of choices that you can select in order to perform some
action, such as placing the instrument in XY mode or turning off the
timeĆout feature.
A number of buttons on the front and top panels of the 222PS invoke
menus when pressed. When a menu is on the display, you can select
one of its items to perform an action. Figure 2Ć8 illustrates the parts of
a menu.
222PS Operator Manual
2Ć9
Page 38
At a Glance
1
4
CHOICE 1NAME:
3
CHOICE 2
CHOICE 3
CHOICE 4
2
Figure 2Ć8:ăParts of a Menu
1.The button labeled CLEAR erases the menu from the display.
2.The menu buttons are next to the screen, along its right edge.
Pressing the button next to a menu item performs the action
represented by that item.
3.The menu items appear along the right edge of the display. Up to
four items can appear on a menu. Each represents a possible
action you can perform.
4.The name of the menu appears at the top left of the display,
followed by a colon.
2Ć10
At a Glance
Page 39
Top Panel Controls
The top panel for the 222PS PowerScout appears as shown in FigĆ
ure 2Ć9.
At a Glance
1
8
7
23
56
4
Figure 2Ć9:ă222PS Top Panel
1.Pressing this button invokes a menu that allows you to recall
saved waveforms. See page 3Ć53.
2.Pressing this button toggles the instrument in or out of store
mode. For more information about store mode, see page 3Ć61.
3.Pressing this button invokes a menu that allows you to specify the
acquisition mode of the instrument. For more information about
acquisition modes, see page 3Ć1.
222PS Operator Manual
2Ć11
Page 40
At a Glance
4.Pressing this button invokes a menu that allows you to specify
5.Pressing this button invokes a menu that allows you to execute a
whether to invert a channel, display signals in XY mode, or display
signal readouts. For more information about inverting channels,
see page 3Ć22. For more information about XY mode, see
page 3Ć81. For more information about displaying or clearing
signal readouts, see page 3Ć21.
variety of special functions. In some cases, you may execute
items from two or three layers of menus.
HYou can check the display alignment and view the instrument
identification and firmware number. See page 3Ć9.
HYou can start selfĆcalibration routines for either channel or the
external trigger input. See page 3Ć9.
HYou can enable or disable the timeĆout feature, set the baud
rate, activate the modem, or select different probe types. For
more information about the timeĆout feature, see page 3Ć37.
For more information about setting the baud rate or activating
the modem, see page AĆ11. For more information about
configuring the 222PS for the correct probes, see page 3Ć45.
HYou can enable or disable MOTOR TRIG, which places a
special filter in the trigger path to allow triggering on motor
drive signals that are pulseĆwidth modulated and on 50/60 Hz
line signals. See page NO TAG.
2Ć12
6.Pressing this button invokes a menu that allows you to specify the
trigger position. See page 3Ć71.
7.Pressing this button invokes a menu that allows you to save or
recall frontĆpanel setups. See page 3Ć55.
8.Pressing this button invokes a menu that allows you to save
waveforms. See page 3Ć51.
At a Glance
Page 41
Side Connectors
The right side of the 222PS PowerScout appears as shown in FigĆ
ure 2Ć10.
Figure 2Ć10:ăSide of 222PS, Storage Pouch Removed
At a Glance
21
3
1.This is the probe connector for channel 1.
2.This is the probe connector for channel 2.
3.This is thebattery connection. See page 3Ć35 for more information
on connecting the battery.
The 222PScomes with an attached storage pouch. Store the probes in
the pouch when you are not using them. You do not need to disconĆ
nect the probes before you store them.
222PS Operator Manual
2Ć13
Page 42
At a Glance
Rear Panel Controls and Connectors
The rear panel for the 222PS PowerScout appears as shown in FigĆ
ure 2Ć11.
1
876
23
Figure 2Ć11:ă222PS Rear Panel
4
5
2Ć14
1.This panel points to the probe inputs on the instrument's side and
indicates the 222PS maximum input voltage rating.
2.This is the RSĆ232 connection port for remote communications.
For more information on remote communications, see page AĆ9.
3.This is the instrument serial number. You will need it if you must
ever arrange to ship the instrument back for maintenance. For
more maintenance information, see page 3Ć31.
4.This knob focuses the 222PS screen. Insert a small screwdriver
into the slot and turn it to adjust the focus.
At a Glance
Page 43
At a Glance
5.This knob varies the brightness of the 222PS screen. See
page 3Ć23 for more information on screen brightness.
6.This is the external trigger common reference connector. To use a
grounded reference with your external trigger source, connect the
reference signal here. See page 3Ć65.
WARNING
To avoid possible injury, do not connect the trigger comĆ
mon reference input to voltages greater than 42ĂV peak.
The trigger common reference input is not insulated.
7.This is the external trigger input connector. To use an external
signal as a trigger source, connect the external trigger signal here.
For more information on external triggering, see page 3Ć65.
WARNING
To avoid possible injury or damage to the 222PS or
equipment connected to it, do not float the external trigger
common connector, the RSĆ232 communications port, or
the external power input above 42ĂV peak. These inputs
are not electrically isolated from each other.
8.This is the external power input. Connect the External Power AC
Adapter to the input to run the instrument from line voltage. See
page 3Ć41 for more information on external power.
222PS Operator Manual
2Ć15
Page 44
At a Glance
The Tilt Stand
The 222PS PowerScout comes with a tilt stand so that you can view
the frontĆpanel and screen more easily. The tilt stand folds under the
instrument when not in use. To use it, lift the instrument and pull the tilt
stand forward until the instrument rests on it.
Figure 2Ć12:ăThe 222PS With Tilt Stand
2Ć16
At a Glance
Page 45
Page 46
Page 47
Acquisition Modes
Acquiring signals involves accepting an analog electrical signal,
sampling it, and producing a waveform. The 222PS allows you to
specify how the instrument acquires the signal and constructs a
waveform. This is the acquisition mode of the instrument.
Kinds of Acquisition Modes
The 222PS provides four acquisition modes: normal, average, enveĆ
lope, and continuous envelope. These modes help you to examine
and manage noisy signals.
Normal Acquisition Mode
Normal acquisition mode is the most common, and it is the instrument
default.
In normal acquisition mode, the instrument displays a waveform with
one sample point for each acquired point.
Average Acquisition Mode
Average acquisition mode displays a waveform that is the average of
the last four waveforms acquired. This mode is useful for reducing
random noise and displaying a cleaner signal.
Average acquisition mode works only when the instrument
is triggered. In auto level and autoĆbaseline trigger modes,
untriggered displays appear identical to those using
normal acquisition mode.
In auto baseline trigger mode, the instrument displays the
last waveform acquired; it is unaveraged.
222PS Operator Manual
NOTE
3Ć1
Page 48
Acquisition Modes
Envelope Acquisition Mode
In envelope mode, the instrument displays the positive and negative
peak signal values that occur during a display sample interval. This
mode is useful for detecting glitches such as unwanted peaks in a
signal.
Envelope acquisition mode functions only for time base
settings between 20Ăms and 20Ăs per division. If you set the
instrument at a faster setting, it functions as if it were in
normal acquisition mode.
Envelope mode samples the signal at 10ĂMHz, thereby acquiring
many samples for each point it displays.
Because of the 10ĂMHz sampling rate, the instrument
cannot detect glitches that last less than 100Ăns.
NOTE
NOTE
3Ć2
Continuous Envelope Acquisition Mode
Continuous envelope mode is similar to envelope mode. The differĆ
ence is that continuous mode accumulates and displays peak values
until you press the button labeled INIT.
NOTE
Continuous envelope acquisition mode functions only for
time base settings between 20Ăms and 20Ăs per division. If
you specify a faster setting, the instrument functions as if it
were in normal acquisition mode.
Changes to most frontĆpanel control settings also act like the INIT
button, discarding the old waveform data and starting the envelope
sequence anew. The only frontĆpanel controls that do not affect
continuous envelope mode are the horizontal and vertical position
knobs and the trigger level knob.
In Detail
Page 49
Figure 3Ć1 summarizes the effects of averaging and envelope modes
on a signal.
Figure 3Ć1:ăNormal, Average, and Envelope Signals
Selecting an Acquisition Mode
To choose an acquisition mode, follow these steps.
ĂĂStep 1:ăPress the ACQ button on the top panel to invoke the
acquisition menu.
The normal acquisition mode is boxed (as in Figure 3Ć2) unless
you have previously selected another acquisition mode.
Acquisition Modes
ĂĂStep 2:ăPress the button next to the acquisition mode you wish to
select. The acquisition mode takes effect, and the menu disapĆ
pears.
Figure 3Ć2:ăThe Acquisition Menu
222PS Operator Manual
NORMACQ:
ENV
AVG
CONT ENV
3Ć3
Page 50
Acquisition Modes
3Ć4
In Detail
Page 51
The 222PS allows you to obtain a readable display of a waveform by
pushing a single button. You can also set up the instrument in a
specific way and then save the setting in memory (see page 3Ć55).
This section discusses various automatic ways to set up the instruĆ
ment.
Procedure
If you wish to view a signal quickly, follow these steps.
ĂĂStep 1:ăConnect the probe tip to the signal you wish to see.
ĂĂStep 2:ăPress the button marked AUTO SETUP on the front
panel.
Pressing AUTO SETUP gets a quick, informative display of any signal
between 20 Hz and 1 MHz.
Parameter Effects
Auto Setup
The AUTO SETUP button has these effects on setup parameters.
Channels
When you press AUTO SETUP, the instrument first determines which
channels to display. It checks both probe tips for a signal on either
channel.
HIf a channel is receiving a signal, the 222PS turns the channel on.
HIf neither channel is receiving a signal, the 222PS turns on chanĆ
nel 1 and turns off channel 2.
222PS Operator Manual
3Ć5
Page 52
Auto Setup
Vertical Scaling
The 222PS then determines the characteristics of the signal, so it can
produce a useful display.
The 222PS sets the vertical position to display the signal in the center
of the screen. If both channels have a signal, the 222PS displays both
signals.
The instrument also sets the volts per division to display each signal
with several divisions of amplitude.
Horizontal Scaling
If only one channel is receiving a signal, the 222PS sets the horizontal
position to display 1-5 waveform cycles. It determines the peak
values and sets the trigger level at the midpoint.
If both channels have a signal, the instrument uses the channel 1
signal to set the seconds per division and trigger level. If the two
signals are synchronized, they both appear stable. Otherwise, the
channel 2 signal is untriggered.
Low and High Frequencies
3Ć6
AUTO SETUP avoids the timeĆbase modes used with the slower time
scales. It therefore does not set the seconds per division to 0.1 s or
slower and does not produce a readable display for signals slower
than 20ĂHz.
For signals with frequencies above 100ĂkHz, AUTO SETUP always
uses a secondsĆperĆdivision setting of 5Ă
The 222PS may display such highĆfrequency signals with too many or
too few cycles on the screen; therefore, they may appear confusing.
You may need to make small corrections to the seconds per division
knob to get a useful display.
Table 3Ć1 shows all AUTO SETUP actions.
ms to minimize search time.
In Detail
Page 53
TableĂ3Ć1:ă Auto Setup Settings
Auto Setup
Control
VOLTS/DIVAs determined by signal
VOLTS/DIV VARCalibrated
CouplingAC if in AC before, and if a signal exists; otherwise
Acquisition modeNormal
STORE/NONSTORENONSTORE
InvertOff (not inverted)
XY DisplayOff
SEC/DIV
Trigger sourceVertical
Trigger modeAuto level
Trigger positionPost
Trigger slopePlus
Trigger levelMidpoint of signal
X10 MAGOff
ReadoutsOn
Setting After Auto Setup
DC if a signal exists; otherwise OFF
As determined by signal; 5Ăms for signals above
100ĂkHz
Selected channelChannel 1 if signal exists, or if channel 2 has no
signal
NOTE
AUTO SETUP does not disable the motor trigger selection.
222PS Operator Manual
3Ć7
Page 54
Auto Setup
3Ć8
In Detail
Page 55
Calibration
The 222PS has a vertical channel selfĆcalibration routine to maintain
best balance and accuracy with temperature variations.
Running the SelfĆCalibration Routine
To achieve the specified performance, you should recalibrate the
222PS any time the ambient temperature has changed by more than
5° C. If the trace jumps when you rotate the volts per division knob
with no signal applied, the instrument probably needs recalibration.
To run the selfĆcalibration routine, follow these steps.
ĂĂStep 1:ăDisconnect both probes from the signal source.
NOTE
To ensure that the selfĆcalibration routine produces accuĆ
rate results, do not run the selfĆcalibration routine while
either probe is connected to a signal source.
ĂĂStep 2:ăPress the AUX FUNCT button on the top panel to display
the auxiliary functions menu. The display appears as shown in
Figure 3Ć3.
222PS Operator Manual
3Ć9
Page 56
Calibration
ALIGNAUX:
SELF CAL
CONFIG
MOTOR TRIG
Figure 3Ć3:ăThe Auxiliary Functions Menu
ĂĂStep 3:ăPress the menu button next to the SELF CAL menu item
to access the calibration menu. The display now appears as
shown in Figure 3Ć4.
3Ć10
CH1SELF CAL:
CH2
EXT TRIG
PROBE
Figure 3Ć4:ăThe Calibration Menu
NOTE
The PROBE selection on the SELF CAL menu is for use by
qualified service personnel only.
In Detail
Page 57
Calibration
ĂĂStep 4:ăTo begin the selfĆcalibration routine for channel 1, press
the menu button next to CH1.
This routine takes about a minute to perform. When it is finished,
the instrument displays PASS or FAIL.
If PASS appears on the screen, channel 1's recalibration was
successful.
If FAIL appears on the screen, run the calibration routine again. If
the failure persists, refer the instrument to a qualified service
person.
If the selfĆcalibration routine fails, the instrument sends a coded
error message. To capture this message, connect the instrument
to a terminal (or a PC emulating a terminal) through the RSĆ232
port on the rear panel and run the failed routine again. See
Appendix B: Remote Communications for a discussion of the error
codes.
ĂĂStep 5:ăNow calibrate channel 2. Invoke the menu again, and
press themenu button next to CH2. Follow the procedure outlined
in Step 4.
ĂĂStep 6:ăNowrecalibrate the external trigger input. Before you start
the calibration, connect the external trigger input connector to the
trigger common reference connector on the rear panel. Use a
jumper cable with a banana plug connector on each end.
222PS Operator Manual
3Ć11
Page 58
Calibration
NOTE
To calibrate the external trigger input, the TRIG COM
connector must be connected to the EXT TRIG INPUT
connector.
ĂĂStep 7:ăInvoke the menu again, and press the menu button next
to EXT TRIG. A new menu appears on the display, as shown in
Figure 3Ć5.
GND EXT TRIG:
CAL
3Ć12
Figure 3Ć5:ăThe Ground External Trigger Calibration Menu
In Detail
Page 59
Calibration
ĂĂStep 8:ăPress the menu button next to CAL to begin the calibraĆ
tion. Follow the procedure outlined in Step 4.
ĂĂStep 9:ăWhen you are done, press the CLEAR button above the
menu buttons to return the instrument to normal operation.
222PS Operator Manual
3Ć13
Page 60
Calibration
3Ć14
In Detail
Page 61
Capturing Random
Events
Capturing random electric events in circuits is difficult and timeĆconĆ
suming. You can use the 222PS's single sequence mode to make the
task easier.
SingleĆsequence mode is one of four trigger modes you can use with
the 222PS. In singleĆsequence mode, the instrument acquires one
triggered signal. It then displays the signal and holds it until you press
the button labeled INIT to start the sequence all over again. Changing
a control that affects the signal acquisition also starts the sequence
again.
The following procedure will help you use singleĆsequence mode to
capture a random event.
ĂĂStep 1:ăPreset the instrument controls to display a baseline
signal.
ĂĂStep 2:ăApply a test signal to the channel 1 input to set the trigger
level. Make sure the test signal is the same amplitude and general
type (negative or positive pulse or sinusoidal) as the signal you
want to trigger on.
ĂĂStep 3:ăPress AUTO SETUP to obtain a quick frontĆpanel control
setup for the test signal. If the resulting vertical or horizontal
scaling result is not precisely what you want, adjust the voltsĆperĆ
division and secondsĆperĆdivision controls as you wish. If necesĆ
sary, reposition the trace vertically.
ĂĂStep 4:ăSet the trigger mode to normal.
ĂĂStep 5:ăSet the trigger source to channel 1.
ĂĂStep 6:ăAchieve a stable display by adjusting the trigger level
control.
ĂĂStep 7:ăTo set the trigger mode to singleĆsequence, Press the
button labeled MODE on the front panel. Press the menu button
next to the menu item SSEQ.
222PS Operator Manual
3Ć15
Page 62
Capturing Random Events
ĂĂStep 8:ă Check that the sweep triggers each time the INIT button
is pressed. If it does not, readjust the trigger level control slightly
until the sweep triggers each time you press INIT. The TRIG'D
indicator lights when the instrument triggers.
ĂĂStep 9:ăDisconnect the test signal from the oscilloscope and
apply the random signal to the input.
ĂĂStep 10:ăPress INIT to arm the trigger system. The instrument
then waits for the trigger event. The TRIG'D indicator lights when
the instrument triggers.
ĂĂStep 11:ăAfter the instrument triggers and completes the single
sequence, press INIT again to acquire another signal.
In scroll mode, the singleĆsequence trigger mode is useful for capturĆ
ing an event that occurs either randomly or infrequently. If the event is
also very narrow, use the envelope mode as well.
Initialize the singleĆsequence function and let the oscilloscope watch
for the event. The 222PS acquires data continuously up to the trigger
point. When the trigger event occurs, the instrument acquires the data
it needs to fill the rest of the display. It then halts the acquisition and
displays the waveform (with the captured trigger event) until you press
the INIT button again.
3Ć16
In Detail
Page 63
The 222PS has two fully isolated input channels with which you can
make floating measurements. With signals of up to 600 VAC
you can make measurements as you would with a volt meter.
This section explains how to select a channel, display the signal it
acquires, and choose the right channel coupling.
Selecting a Channel
You must select a channel before you can change its settings using
the frontĆpanel controls. You can select only one channel at a time. The
display indicates which channel is currently selected.
To select a channel, push the appropriate channel button (CH1 or
CH2) on the front panel.
Figure 3Ć6 shows the 222PS displaying signals for two channels.
Channel 1 is selected.
Channels
input,
RMS
-46.7mV+
Figure 3Ć6:ăChannel One Selected
222PS Operator Manual
5mV=.5V~
mS
2
3Ć17
Page 64
Channels
Displaying a Channel
When a channel is on, the 222PS displays any signal it acquires
through that channel. When a channel is off, the 222PS does not
display its signal. If either channel is off, the screen displays OFF"
instead of its voltsĆperĆdivision value.
Even when a channel is off, the instrument can still use it as a trigger
source.
By default, the 222PS displays both channels. To turn a channel off,
follow these steps.
ĂĂStep 1:ăIf the channel is not already selected, select it by pushing
the appropriate channel button (CH1 or CH2) on the front panel.
ĂĂStep 2:ă Press the channel button again to display the channel
menu. The screen appears as shown below.
DCCH 2 CPLG:
3Ć18
AC
GND
CH 2 OFF
Figure 3Ć7:ăThe Channel Menu
ĂĂStep 3:ăPress the button next to the bottom menu item,
CH 2 OFF. This turns the selected channel off and clears the
menu from the display. However, until you select channel 1 for
another purpose, channel 2 is still the selected channel.
In Detail
Page 65
To display a channel you have turned off, repeat the above procedure
until you see the menu on the screen. Then select the type of coupling
you wish to use for the channel signal. The following section describes
how to select the channel coupling.
When a channel is off, you can still change its vertical
settings with the POS, VOLTS/DIV,orVAR VOLTS/DIV
controls. First, select the channel. Then make the changes
you wish. The changes take effect when you turn the
channel back on.
Setting Channel Coupling
There are three possible couplings for each channel:
HDC coupling passes all frequencies of the input signal up to the
useful bandwidth of the instrument.
HAC coupling blocks any DC component of the signal and is the
most commonly used.
Channels
NOTE
HGround coupling disconnects the input signal and grounds the
input for the selected channel.
To select the coupling for a channel, follow these steps.
ĂĂStep 1:ăIf the channel is not already selected, select it by pushing
the appropriate channel button (CH1 or CH2) on the front panel.
ĂĂStep 2:ăPress the channel button again to invoke the channel
menu. The screen appears as in Figure 3Ć7.
ĂĂStep 3:ăPress the button next to the menu item representing the
type of coupling you wish to use. For example, press the second
button from the top to select AC coupling.
If the type of coupling you wish already appears boxed, it is
already the selected coupling. Press the CLEAR button to remove
the menu from the display.
222PS Operator Manual
3Ć19
Page 66
Channels
3Ć20
In Detail
Page 67
The 222PS display shows you menus, signals, and readouts. This
section explains how you can control the information displayed to you.
Readouts
Readouts include volts per division for each signal, seconds per
division, and trigger level. The instrument also displays additional
information, depending on the characteristics of the signal you display
and the mode of the instrument.
Displaying and Clearing the Readouts
Unless you explicitly turn off the readouts, the 222PS displays them. If
you wish to turn the readouts off and view only the signals, follow
these steps.
The Display
ĂĂStep 1:ăPress the DISPL button on the top panel to access the
display menu. Figure 3Ć8 shows the display menu.
Figure 3Ć8:ăThe Display Menu
222PS Operator Manual
INV1DISPLAY:
INV2
XY
RO OFF
3Ć21
Page 68
The Display
ĂĂStep 2:ăPress the menu button next to the bottom menu item RO
OFF. The menu disappears and the screen displays only the
signal.
ĂĂStep 3:ăIf you wish to turn the readouts back on again, repeat the
above procedure. The next time you access the display menu, the
RO OFF item appears boxed, indicating that it is selected and that
readouts have therefore been turned off. Press the button again to
deselect the menu item. The menu disappears and readouts
again appear.
Inverting the Display
You can invert the waveforms displayed for either channel. Figure 3Ć9
shows an example of a normal and an inverted waveform.
Normal
Waveform
Inverted
Waveform
3Ć22
Figure 3Ć9:ăNormal and Inverted Waveforms
Some signal processingcircuits,such as circuits thatprocesscomposĆ
ite video signals, automatically invert waveforms. While working on
these circuits you may want to invert these signals to view them
normally.
To invert a signal, follow these steps.
ĂĂStep 1:ăPress the DISPL button on the top panel to invoke the
display menu. See Figure 3Ć8.
In Detail
Page 69
ĂĂStep 2:ăPress the button next to the menu item corresponding to
the channel you wish to invert (INV1 OR INV2).
After you press the menu button, the menu disappears and the
inverted signal (if any) reappears on the screen. A small downĆ
arrow appears in front of the volts per division readout for an
inverted channel.
ĂĂStep 3:ăIf you wish to display the channel normally, repeat the
above procedure. The next time you access the display menu, the
inverted channel item appears boxed, indicating that it is selected.
Press the button again to deselect the menu item. The menu
disappears, and the channel is no longer inverted.
Varying the Brightness
Use the intensity control on the back panel (labeled INTEN) to adjust
the display's brightness.
To changethe brightness, insert a small screwdriver into thecenter slot
of the knob and rotate the knob until the display appears as you wish.
The Display
222PS Operator Manual
3Ć23
Page 70
The Display
3Ć24
In Detail
Page 71
Horizontal Operation
This section discussescontrollingthe horizontal aspects ofyoursignal.
The knobs used to do this are at the bottom left of the front panel, as
shown in Figure 3Ć10.
Figure 3Ć10:ăThe Horizontal Controls
The horizontal controls affect signals acquired through both channels.
Horizontal Positioning
To position the signals horizontally, rotatethe inner knob, labeled POS.
You can position the trace to the right or the left up to oneĆhalf the
length of the screen. Readouts are not affected by horizontal positionĆ
ing.
For information on the positioning of XY traces, see page 3Ć81.
When X10 magnification is on, the horizontal position control scrolls
through the magnified waveform. For more information about magnifyĆ
ing waveforms, see page 3Ć29.
The horizontal position control affects the position of a saved waveĆ
form. The 222PS displays a saved waveform at the horizontal position
presently in effect, not the horizontal position at which it was saved.
222PS Operator Manual
3Ć25
Page 72
Horizontal Operation
Seconds per Division
The 222PS can display a signal with a range of 20Ăs to 50Ăns. If magniĆ
fication is on, the instrument's range is 2Ăs to 5Ăns. This means that the
222PS can display a waveform that represents as long a period as
200Ăseconds or as short a period as 50Ăns (in the latter case, with
magnification on).
To change the secondsĆperĆdivision setting, turn the outer knob
(labeled SEC/DIV). The instrument displays the resulting seconds per
division at the bottom of the screen. The 222PS uses a 1-2-5 switchĆ
ing sequence: this means that each click of the knob changes the time
scale from, for example, 1Ăms to 2Ăms and then to 5Ăms, before going
to 10Ă ms.
If you are using battery power at slow time bases, turn off
the timeĆout feature described on page 3Ć37. Otherwise,
the instrument may time out and turn itself off before it can
completely acquire the signal.
NOTE
3Ć26
Aliasing
Aliased waveforms are waveforms that appear to have a frequency
much lower than is accurate.
Aliasing can occur when the seconds per division setting, and thereĆ
fore the sample rate, is too low to display a high frequency waveform
accurately. When this occurs, the instrument does not sample the
signal often enough. The resulting waveform it displays is misleading.
Figure 3Ć11 illustrates an aliased waveform.
A common symptom of aliasing is an unstable display even when the
TRIG'D light is on.
In Detail
Page 73
Horizontal Operation
HighĆFrequency
Apparent LowĆFrequency
Waveform Due to Aliasing
Sampled Points
To avoid aliasing, make sure that your sample rate is fast enough for
the frequency of the signal you wish to view Ċ usually, at least twice
as fast as the highest frequency component of the signal.
Actual
Waveform
Figure 3Ć11:ăAn Aliased Waveform
TimeĆBase Mode
When you choose a seconds per division setting, the instrument
automatically selects the timeĆbase mode required to display the
resulting signal. The timeĆbase mode can also depend on the trigger
and acquisition modes. The 222PS uses four timeĆbase modes:
record, equivalentĆtime, scroll, and scrollĆscan.
Record Ċ For most seconds per division settings, the instrument
uses the timeĆbase mode you are most accustomed to: record. When
a trigger occurs, the 222PS acquires and displays a full screen of the
waveform in one pass.
EquivalentĆtime Ċ However, when the seconds per division setting
is too fast, the instrument cannot sample fast enough to capture all 512
samples and display them in one pass. Therefore, the instrument
depends on successive repetitions of the same waveform to fill the
display with samples.
Scroll and ScrollĆscan Ċ When the seconds per division setting is
very slow, the display takes too long to fill using record timeĆbase
mode. Therefore, the instrument uses one of scrolling timeĆbase
modes.
222PS Operator Manual
3Ć27
Page 74
Horizontal Operation
The instrument uses scroll mode for slow time bases when either
autoĆlevel or autoĆbaseline trigger mode is in effect.
The instrument uses scrollĆscan mode for slow time bases when
normal or single sequence trigger mode is in effect.
Averaging and continuous envelope acquisitionmodesacquireseveral
records of data before displaying a waveform. Therefore, if either of
these acquisition modes is in effect, the instrument uses record
timeĆbase mode at slow time bases instead of a scrolling mode. Under
these circumstances, the display updates slowly.
Table 3Ć2 shows these dependencies and the actual secondsĆperĆdiviĆ
sion settings associated with each timeĆbase mode. The boundary
between the medium and the slow ranges differs according to whether
the instrument is in store or nonstore mode.
TableĂ3Ć2:ă TimeĆBase Modes
Store Mode
On or off
On
Off
On
Off
Seconds per
Division
Fast:
2Ăms to 50Ăns
Normal:
50Ăms to 5Ăms
20Ăms to 5Ăms
Slow:
20Ăs to 0.1Ăs
20Ăs to 50Ăms
Acquisition
Mode
AnyEquivalentĆtime
AnyRecord
Normal or
Envelope
Averaging or
Continuous
Envelope
TimeĆbase
Mode
Scroll or
ScrollĆscan
RecordĊ
slow update
3Ć28
In Detail
Page 75
Magnifying the Signal
You can magnify waveforms by ten times. To do so, push the inner
horizontal control knob, labeled POS.
When magnification is on, each division contains five data points
horizontally instead of the normal 50. The secondsĆperĆdivision readĆ
outs show corresponding values. Magnification also affects saved
waveforms on the display and their secondsĆperĆdivision readouts.
To view the rest of the magnified waveform, turn the horizontal position
knob. This action allows you to pan through the magnified waveform
from side to side.
When you pan through a magnified waveform, you may also be
moving the trigger position. The trigger position indicator, shown as a
+, cannot move off the screen. Therefore, if you pan the trigger
position off the screen to either side, the trigger position indicator
remains at the edge of the screen to show the direction of the trigger
point.
When magnification is on, a 10X indicator appears at the bottom of the
screen, to the right of the seconds per division readout. Figure 2Ć7
shows the bottom readouts.
Horizontal Operation
Displays in XY mode cannot be magnified.
When you magnify displays in scroll and scrollĆscan timeĆbase modes,
they update only after the instrument completely acquires the waveĆ
form.
To turn off magnification, push the horizontal POS knob again.
222PS Operator Manual
NOTE
3Ć29
Page 76
Horizontal Operation
3Ć30
In Detail
Page 77
Maintenance
The 222PS is covered by a standard Tektronix threeĆyear warranty. If it
fails during the warranty period, return it to Tektronix for free servicing
(subject to the conditions of the warranty statement).
To arrange for warranty service or get an estimate for outĆofĆwarranty
repairs, call 1Ć800ĆTEKĆWIDE (1Ć800Ć835Ć9433).
To help diagnose the problem, have the instrument serial number and
firmware version number available. The serial number is located at the
top right of the rear panel. To get the firmware identification number,
follow the steps below.
If your instrument must be returned for servicing, package it as deĆ
scribed on page 3Ć33.
Identifying the Firmware Version
To identify the 222PS firmware version, follow these steps.
ĂĂStep 1:ăPress the AUX FUNCT button on the top panel to access
the auxiliary functions menu. The display now appears as shown
in Figure 3Ć12.
222PS Operator Manual
3Ć31
Page 78
Maintenance
ĂĂStep 2:ăPress the menu button next to ALIGN. This calls up the
ALIGNAUX:
SELF CAL
CONFIG
MOTOR TRIG
Figure 3Ć12:ăThe Auxiliary Functions Menu
XY Alignment Menu as shown in figure 3Ć13.
3Ć32
XYALIGN:
IDENT
Figure 3Ć13:ăXY Alignment Menu
ĂĂStep 3:ăPress the menu button next to IDENT. A message apĆ
pears in the middle of the display, identifying the instrument, as
shown in Figure 3Ć14. The version number you see may differ.
In Detail
Page 79
TEKĆ222PS VER: x.xx
Figure 3Ć14:ăThe Firmware Version
Repackaging for Shipment
Remember to put the instrument in its carry case before repacking. If
the original packing materials are unfit or unavailable, then repackage
the instrument in the following manner:
Maintenance
1.Use a corrugated cardboard shipping carton with a test strength
of at least 125 kg (275 lb) and an interior size at least 15 cm (6 in)
greater than the instrument size in all dimensions. See AppenĆdix D: Specifications for instrument dimensions.
2.Enclose the following information:
Hthe owner's name and address
Hthe name and phone number of a contact person
Hthe serial number of the instrument
Hthe reason for returning the instrument
Ha complete description of the service required
3.Disconnect the battery before packing the instrument.
4.Completely wrap the instrument with polyethylene sheeting or its
equivalent to protect the outside finish and keep harmful subĆ
stances out of the instrument.
222PS Operator Manual
3Ć33
Page 80
Maintenance
5.Cushion the instrument on all sides with three inches of padding
6.Seal the shipping carton with an industrial stapler or strapping
7.Call 1Ć800ĆTEKĆWIDE (1Ć800Ć835Ć9433) for shipping instructions.
material or urethane foam, tightly packed between the carton and
the instrument.
tape.
3Ć34
In Detail
Page 81
You can operate the 222PS by using the internal battery or by plugĆ
ging it into external power. This section explains how to do both. It also
explains how to charge and change the battery when necessary.
Battery Operation
The 222PS comes supplied with a battery for use when portable
operation is convenient.
Completely recharge the battery as soon as possible after each use of
the instrument under battery power.
The instrument is shipped from the factory with the battery
charged. However, the battery may not retain its charge
while in transit to you. Therefore, we recommend that you
charge the battery for three hours before operating the
222PS for the first time.
Power
NOTE
Even when the instrument is off, current trickles slowly
from the battery. If the current drawn off in this way deĆ
pletes the battery below 7.32ĂV, the instrument cannot start
on battery power. If this condition occurs, recharge the
battery immediately. Instructions for charging the battery
are on pageĂ3Ć37.
Connecting the Battery
The battery is charged at the factory. It is shipped disconnected to
prolong its shelf life. To connect the battery, follow these steps.
ĂĂStep 1:ăPlace the oscilloscope on its left side as viewed from the
front panel.
222PS Operator Manual
NOTE
3Ć35
Page 82
Power
ĂĂStep 2:ăOpen the probe pouch and disconnect the probes.
ĂĂStep 3:ăSlide the battery compartment cover and probe pouch
toward the rear of the instrument to unlock the locking tabs.
ĂĂStep 4:ăLift the battery compartment cover and remove it. The
instrument now appears as shown in Figure 3Ć15.
Figure 3Ć15:ăSide View Without Battery Cover
ĂĂStep 5:ăConnect the threeĆwire battery connector to the pins at
the rear of the instrument. The orientation of the connector does
not matter.
3Ć36
ĂĂStep 6:ăReplace the battery compartment cover. Position the
cover locking tabs into the matching slots in the battery compartĆ
ment. Make sure the locking tabs are all the way in the slots on
both the top and the bottom. If the tabs do not seat easily, first
seat the top tabs and then press on the bottom of the cover to
seat the bottom tabs.
ĂĂStep 7:ăPush forward on the rear of the battery compartment
cover to lock the tabs.
ĂĂStep 8:ăAs soon as possible, charge the battery for three hours.
See the following instructions for charging the battery.
In Detail
Page 83
Power
Charging the Battery
During periods of heavy use in a harsh environment, you will need to
recharge the battery after three hours of operation. Under better
circumstances, it may need recharging less often. The battery will last
longer if you recharge the instrument after each use.
When the battery charge is low, a lowĆbattery indicator
the upper right corner of the display. If the battery voltage drops below
7.32ĂV, the instrument automatically turns itself off.
In order to recharge the battery, plug in the External Power AC Adapter
and leave the instrument turned off for three hours.
appears in
NOTE
The 222PS battery recharges whenever you plug it into
external power. However, it recharges faster if the instruĆ
ment is off.
Charging the Battery Externally
You can charge the battery outside the instrument using the external
battery charger accessory. See the OptionalAccessories information in
Appendix E. You can also use any other charger that supplies 9.8ĂVDC
at 20
_C with the supply current limited to 1ĂA. For best results in
various temperatures, thermally compensate the charging voltage by
-10 mV per degree C.
For example, at 50
9.80 V + [(50 - 20)
To charge the battery, follow the steps on the data sheet that comes
with the battery charger unit.
_C, the charging voltage is
-10 mV] = 9.50 V
Time Out
An automatic timeĆout feature prevents the battery from losing power
when the instrument is unattended for a long period. When you enable
the timeĆout feature, the 222PS turns itself off after two minutes of
operating under battery power with no changes to the controls.
222PS Operator Manual
3Ć37
Page 84
Power
You can disable the timeĆout feature so that the instrument stays on for
the life of the battery's charger. You can also enable the timeĆout
feature again when you wish. To do so, follow these steps.
ĂĂStep 1:ăPress the AUX FUNCT button on the top panel to display
the auxiliary functions menu. Figure 3Ć16 shows the display.
ALIGNAUX:
SELF CAL
CONFIG
MOTOR TRIG
Figure 3Ć16:ăThe Auxiliary Functions Menu
3Ć38
ĂĂStep 2:ăPress the menu button next to CONFIG to access the
configuration menu. Figure 3Ć17 shows the configuration menu.
TIME OUTCONFIG:
BAUD
MODEM ON
PROBE
Figure 3Ć17:ăThe Configuration Menu
In Detail
Page 85
Power
ĂĂStep 3:ăThe TIME OUT menu item appears boxed when the
timeĆout feature is enabled. To disable the timeĆout feature press
the top menu button.
This menu item is a toggle. Repeating this procedure reĆenables
the timeĆout feature and boxes the menu item again.
ĂĂStep 4:ăTo clear the menu from the display, press the button
labeled CLEAR.
Replacing the Battery
If you use the instrument on battery power often, you may wish to
obtain and charge extra batteries to take with you. Then, when the
battery charge inside the instrument gets low, you can switch to one of
the fully charged spare batteries.
NOTE
Because the 222PS loses saved data after 30 s without
power, have the spare battery handy before beginning this
procedure.
To replace the battery, follow these steps:
ĂĂStep 1:ăOpen the battery compartment. Follow the procedure on
pageĂ3Ć35.
ĂĂStep 2:ăDisconnect the battery from the threeĆwire battery conĆ
nector.
ĂĂStep 3:ăLift the battery pack out of the battery compartment.
ĂĂStep 4:ăPlace the charged replacement battery into the battery
compartment with the battery leads on the bottom facing toward
the threeĆwire battery connector.
ĂĂStep 5:ăConnect the battery to the threeĆwire battery connector.
ĂĂStep 6:ăClose the battery compartment. Follow the procedure on
pageĂ3Ć35.
ĂĂStep 7:ăRecharge the low battery as soon as possible. See the
procedure on page 3Ć37.
222PS Operator Manual
3Ć39
Page 86
Power
Storing the Instrument
When storing the instrument for a period shorter than two months,
leave the battery connected. When the instrument is off, the current
drawn from the battery is less than 1ĂmA. With the battery in place,
waveform settings and front panel setups remain in memory; they are
available when you turn the oscilloscope on again.
When storing the instrument for a period longer than two months,
follow these steps to extend the life of your battery and instrument.
ĂĂStep 1:ăCharge the battery fully. Follow the instructions on
pageĂ3Ć37.
ĂĂStep 2:ăRemove the battery from the instrument. Use the proceĆ
dure on pageĂ3Ć35 to remove the battery cover.
ĂĂStep 3:ăStore the fully charged battery in a cool place.
Deep Discharge
Under certain circumstances, the battery can become deeply disĆ
charged. When in this state, the battery accepts a charge very slowly.
In some cases, it may not accept a charge at all.
3Ć40
A deep discharge condition is caused by three situations:
Husing the instrument until the battery charge is low and then
storing it without recharging it
Hstoring the battery in a discharged state
Hstoring the instrument for over two months without removing the
battery
If the battery becomes deeply discharged, you may be able to recover
it with the following procedure.
ĂĂStep 1:ăCharge the battery for 24Ăhours. Follow the instructions
on pageĂ3Ć37.
ĂĂStep 2:ăIf the battery does not accept the charge, remove it from
the instrument and try again to charge it using a 20ĂV power
supply that is currentĆlimited to 100ĂmA.
In Detail
Page 87
ĂĂStep 3:ăDuring this operation, check the power supply frequently
for a currentĆlimited state. If the battery recovers from its state of
deep discharge, it will cause the power supply to currentĆlimit. Do
not leave the battery connected to the external power supply
without checking it frequently.
ĂĂStep 4:ăIf the power supply shows that it is currentĆlimited,
reinstall the battery in the instrument.
ĂĂStep 5:ăContinue to recharge the battery.
ĂĂStep 6:ăIf the battery does not recover, return it to Tektronix for
safe disposal or dispose it in accordance with local environmental
regulations.
External Power Operation
The 222PS has an external power input connector so that it need not
use the battery power. You can connect the 222PS to a wall socket
using the External Power AC Adapter or you can use your own exterĆ
nal power source.
Power
You can also operate the instrument on external power without the
battery present. For instructions on removing the battery, see
pageĂ3Ć39.
AC Line Operation
The 222PS comes with an External Power AC Adapter. This adapter
converts AC line voltage to the 16-20ĂVAC input voltage that the
instrument requires. The adapter also recharges the 222PS's battery.
In order to maintain the battery charge for times when you
require portable operation, we recommend that you use
the External Power AC Adapter whenever practical.
In order to operate the instrument from line power, follow these steps.
ĂĂStep 1:ăPlug the jack end of the External Power AC Adapter into
the external power input on the rear panel of the instrument.
222PS Operator Manual
NOTE
3Ć41
Page 88
Power
ĂĂStep 2:ăPlug the prong end of the External Power AC Adapter into
an AC power source.
ĂĂStep 3:ăPress the ON button.
When the oscilloscope is operating under external power, an
external power indicator
the display.
appears in the upper right corner of
Other Sources of External Power
You can operate the 222PS from your own external power source. The
power source must supply at least 15ĂW or 16ĂvoltĆamperes.
HAn AC power source must provide 16-20ĂVAC at 47-400 ĂHz.
HA DC power source must provide 12-28ĂVDC.
The external power input connector has two contacts. DC power of
either polarity can be between contacts.
NOTE
To prevent blowing internal fuses, do not force either pin
lower than .5 volts more negative than the instrument
chassis. The instrument chassis is connected to the
ground pin of the RSĆ232 communications port and to the
external trigger connector common.
3Ć42
WARNING
To avoid possible injury or damage to the 222PS or
equipment connected to it, do not float the external trigger
common connector, the RSĆ232 communications port, or
the external power input above 42ĂV peak. These inputs
are not electrically isolated from each other.
In Detail
Page 89
Probes
The 222PS comes with two P850 10X probes. An additional 1X probe,
the P400 probe, is available as an optional accessory.
The P850 probes provide high attenuation so that you can scale
signals approaching 600 VAC
useful for measuring sensitive highĆimpedance electronic circuits or
highĆvoltage divider circuits. The optional P400 probes measure
lowĆlevel signals requiring high sensitivity.
To prevent improper operation and the risk of electric
shock, use only Tektronix P400 or P850 probes with this
instrument.
Actual probe attenuation factors are 3X for the P400 probe
and 30X for the P850 probe. The instrument is calibrated to
compensate for these attenuation factors. Other probes or
input devices will therefore give incorrect amplitude
displays.
for better display. They are also
RMS
CAUTION
The optional P400 1X probe limits the maximum deflection factor of the
222PS to 50Ăvolts per division. It decreases the probe tip input impedĆ
ance to 1ĂMW and is sufficient for minimal loading of sensitive circuits.
The P400 probe also limits input frequency to 20 MHz.
To avoid personal injury or damage to the 222PS or the
probes, do not apply more than 850ĂV peak between
probe tip and earth ground, between probe tip and probe
common, or between probe common and earth ground.
222PS Operator Manual
WARNING
3Ć43
Page 90
Probes
Connecting the Probes
The input connectors for the probes are inside the pouch over the
battery compartment, on the right side of the oscilloscope as you face
the screen. You must unzip or remove the pouch to access the conĆ
nectors.
Figure 3Ć18:ăSide View Showing Probe Connectors
(Pouch Removed)
Probe Connectors
3Ć44
To connect a probe, place its jack end into a channelĂinput connector.
Press until you feel the probe is firmly seated.
You do not need to disconnect the probes before storing them in the
pouch.
CAUTION
The exposed probe tips are sharp for probing through
solderĆresin and oxide layers. When placing the probes in
the side pouch, store them with the retractable hook tip
attached to prevent unnecessary damage to the pouch.
In Detail
Page 91
Configuring the Probes
To ensure that the 222PS is operating with the correct settings, confiĆ
gure the instrument to match the probes you are using.
To set the probe configuration, follow these steps.
ĂĂStep 1:ăPress the AUX FUNCT button on the top panel to display
the auxiliary functions menu, as shown in Figure 3Ć19.
Probes
ALIGNAUX:
SELF CAL
CONFIG
MOTOR TRIG
Figure 3Ć19:ăThe Auxiliary Functions Menu
ĂĂStep 2:ăPress the menu button next to the CONFIG menu item to
access the configuration menu. The display now appears as
shown in Figure 3Ć20.
222PS Operator Manual
3Ć45
Page 92
Probes
TIME OUTCONFIG:
BAUD
MODEM ON
PROBE
Figure 3Ć20:ăThe Configuration Menu
ĂĂStep 3:ăPress the menu button next to the PROBE menu item to
access the probe menu, as shown in Figure 3Ć21.
3Ć46
CH 1 1XPROBE:
CH 1 10X
CH 2 1X
CH 2 10X
Figure 3Ć21:ăThe Probe Menu
In Detail
Page 93
Probes
ĂĂStep 4:ăBoxes appear around the items that represent the current
probe configuration. The default configuration for the 222PS
assumes 10X probes on both channels. Therefore, those menu
items appear boxed unless you have already changed the probe
configuration.
Press the menu button or buttons next to the menu items correĆ
sponding to the configuration you need. If the current probe
settings are appropriate, you need not press any buttons.
ĂĂStep 5:ăAfter you have configured the 222PS for the correct
probes, press the CLEAR button to remove the menu from the
display.
NOTE
The probe configurations are in the 222PS memory. They
remain there until you change them again or until the
memory loses power. If the probe configuration is lost, it
returns to the default value of 10X probes for both chanĆ
nels.
Probe Accessories
The P850 probes come with four accessories (shown in Figure 3Ć22):
Ha retractable hook tip
Han IC lead protection shroud
Ha detachable probe common lead
Htwo cableĆmarker rings
When removing the hookĆtip assembly from the probe, you
can accidentally disconnect the probe body from the
probe cable. If this occurs, no signal can pass from the
probe to the oscilloscope. To reconnect the probe body to
the cable, insert the connector at the end of the cable into
the probe body until it seats firmly.
222PS Operator Manual
NOTE
3Ć47
Page 94
Probes
1
2
3Ć48
Probe
34
Figure 3Ć22:ăProbe and Accessories
In Detail
Page 95
Probes
1.You can use the detachable probe common lead shown in FigĆ
ure 3Ć22 to connect the oscilloscope input common to the referĆ
ence point of the circuit being tested. The probe common lead is
not chassis ground and you can connect it to an active circuit
component. You can therefore make a floating measurement
across a component, with neither point connected to ground
potential.
WARNING
To avoid personal injury or damage to the 222PS or the
probes, do not apply more than 850 V peak between
probe tip and earth ground, between probe tip and probe
common, or between probe common and earth ground.
2.You can snap the colored cable marker rings shown in FigĆ
ure 3Ć22 into the grooves on the probe cable to distinguish
between the probes connected to channel 1 and channel 2.
3.When testing integrated circuit devices, remove the hook tip and
use the IC lead protection shroud on the probe tip. The shroud
shown in Figure 3Ć22 exposes the sharp probe tip, but it prevents
the probe from creating a short circuit across adjacent IC leads.
4.The hook tip shown in Figure 3Ć22 can connect to accessible test
points such as a component lead or test point connector. This
accessory frees your hands for other tasks.
222PS Operator Manual
3Ć49
Page 96
Probes
3Ć50
In Detail
Page 97
Saving and
Recalling Data
The 222PS has memory to store saved waveforms and frontĆpanel
setups. Data in this memory remains as long as the instrument has
access to one of these power sources:
Ha battery with a charge greater than 7.32ĂV (see page 3Ć35 for a
discussion of battery operation)
Hline voltage coming through a power cord plugged into a wall
socket through the external AC adapter (see page 3Ć41 for a
discussion of line operation)
Hpower coming in from another external power source (see
pageĂ3Ć42 for a description of acceptable external power sources)
The instrument does not need to be on for it to access power for the
memory. However, if you turn off the instrument, disconnect it from any
external power source, and remove its battery, it will lose any data in
memory after 30 s.
Waveforms
A saved waveform is a record of a single acquisition cycle. Think of it
as a snapshot of a waveform. The 222PS lets you save up to four
waveforms in memory. It can recall these at any time.
Saving a Waveform
Use these steps to save a waveform.
ĂĂStep 1:ăSelect the channel whose waveform you wish to save or,
if you wish, put the instrument in XY mode.
ĂĂStep 2:ăUsing the selected channel, acquire and display the
waveform you wish to save.
ĂĂStep 3:ăPosition the waveform where you wish it to be saved.
222PS Operator Manual
3Ć51
Page 98
Saving and Recalling Data
ĂĂStep 4:ăPress the SAVE button on the top panel. Acquisition
stops, the screen freezes, and a menu appears as shown in
Figure 3Ć23.
Figure 3Ć23:ăSaved Waveforms Menu
If the waveform you are saving uses channel 1, the menu name is
SAVE CH1 WV:. If the waveform you are saving uses channel 2,
the menu name is SAVE CH2 WV:. If the waveform you are saving
uses XY mode, the menu name is SAVE XY WV:.
1SAVE CH1 WV:
2
3
4
3Ć52
The example in Figure 3Ć23 assumes a waveform acquired using
channel 1.
ĂĂStep 5:ăPress the button next to the memory location in which
you wish to save the waveform.
NOTE
You can save only one waveform to each memory location;
the 222PS can store only four waveforms. If you choose a
memory location that already holds a waveform, the
instrument replaces it with the one you are presently
saving.
The 222PS displays readouts of the parameters of the saved
waveform at the bottom of the screen, as shown in Figure 3Ć24.
In Detail
Page 99
Saving and Recalling Data
In Figure 3Ć24, the waveform has been saved to memory locaĆ
tion 1 and therefore named W1. Its vertical setting is 10 mV per
division, its coupling is DC (as indicated by the =), and its horiĆ
zontal setting is 10 ms per division.
1SAVE CH1 WV:
2
3
W1 10 mV = 10 ms
4
Figure 3Ć24:ăSaved Waveform Parameters
A waveform saved in XY mode shows the scale factor for the x
axis (channel 1)in the position of thevolts per division setting. The
scale factor for the y axis (channel 2) appears in the position of the
seconds per division setting.
ĂĂStep 6:ăTo clear the menu, press the button labeled CLEAR or
invoke another menu. Changing one of the following frontĆpanel
controls also clears the menu: volts per division, seconds per
division, X10 magnification, or autolevel.
ĂĂStep 7:ăAfter clearing the menu, the 222PS continues to display
the saved trace and its readout. To clear them, press the button
labeled CLEAR again.
Recalling a Saved Waveform
To display a saved waveform, follow these steps.
ĂĂStep 1:ăPress the RCL button on the top panel. A menu appears
as shown in Figure 3Ć25.
222PS Operator Manual
3Ć53
Page 100
Saving and Recalling Data
Figure 3Ć25:ăThe Recall Waveforms Menu
ĂĂStep 2:ăPress the button next to the memory location that holds
the waveform you wish to view. The menu disappears and the
222PS displays the waveform at the same position in which it was
saved, with the same parameters in effect. The instrument also
displays readouts of these parameters at the bottom of the
screen.
1RECALL WVFRM:
2
3
4
3Ć54
If the instrument displays two saved waveforms, the last one you
recall is the one whose parameters appear on the screen.
If the 222PS is already displaying a saved waveform, the memory
location that holds it appears boxed in the menu. If you press a
button next to a memory location holding an already displayed
waveform, the waveform disappears from the screen.
If you press a button corresponding to an empty memory locaĆ
tion, the instrument beeps and the menu remains displayed.
ĂĂStep 3:ăTo clear the menu press the CLEAR button or invoke
another menu.
ĂĂStep 4:ăAfter you clear the menu, the instrument continues to
display the recalled waveform. To clear the recalled waveform,
press the CLEAR button again. This action clears all recalled
waveforms.
ĂĂStep 5:ăTo erase a waveform from a memory location, save
another waveform to the same location.
In Detail
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.