Tektronix 2213A Instruction Manual

Tektronix-
2213A
OSCILLOSCOPE
OPERATORS
INSTRUCTION MANUAL
TABLE OF
Page
LIST OF ILLUSTRATIONS
......................................
M
LIST OF TABLES ..................................................... ii
OPERATORS SAFETY SUMMARY
........................
Mi
Section 1 GENERAL INFORMATION
INTRODUCTION
................................
1-1
SPECIFICATION
..................................
1-1
CALIBRATION
......................................
1-9
REPACKAGING FOR SHIPMENT .... 1-9
Section 2 PREPARATION FOR USE
FIRST-TIME START UP
....................
2-1
SAFETY
..........................................
2-1
LINE VOLTAGE
..............................
2-1
POWER CORD
..............................
2-1
LINE FUSE
....................................
2-1
INSTRUMENT COOLING
..............
2-2 CONTROLS, CONNECTORS, AND INDICATORS
......................................
2-3
DISPLAY, POWER, AND PROBE
ADJUST
..........................................
2-3
VERTICAL
......................................
2-4
HORIZONTAL
................................
2-5
TRIGGER
........................................
2-6
REAR PANEL
................................
2-7
Section 3 OPERATORS FAMILIARIZATION
GENERAL OPERATING INFORMATION
....................................
3-1
GRATICULE
....................................
3-1
GROUNDING
..................................
3-1
SIGNAL CONNECTIONS
..............
3-1
CONTENTS
Section 3 OPERATORS FAMILIARIZATION (cont)
INPUT COUPLING CAPACITOR PRECHARGING
............................
3-1
OPERATOR’S ADJUSTMENTS
........
3-3
INTRODUCTION
............................
3-3
BASELINE TRACE
........................
3-3
TRACE ROTATION
........................
3-3
PROBE COMPENSATION
............
3-3
Section 4 OPERATING PROCEDURES
BASIC APPLICATIONS
......................
4-1
INTRODUCTION
............................
4-1
INDEX TO BASIC APPLICATION
PROCEDURES
..............................
4-1
VOLTAGE MEASUREMENTS
........
4-1
TIME MEASUREMENTS
..............
4-4
TELEVISION DISPLAYS
................
4-8 DELAYED-SWEEP MAGNIFICATION
............................
4-9
Section 5 OPTIONS AND ACCESSORIES
INTRODUCTION
............................
5-1
OPTIONS
........................................
5-1 INTERNATIONAL POWER CORDS 5-1 STANDARD ACCESSORIES
........
5-1 OPTIONAL ACCESSORIES
..........
5-1
FRONT PANEL ILLUSTRATION
CHANGE INFORMATION
REV DEC 1983
I
2213A Operators
LIST OF ILLUSTRATIONS
Figure Page
The 2213A Oscilloscope ....................................................................................................................................... iv
1-1 Maximum input voltage vs frequency derating curve for CH 1 OR X, CH 2 OR Y, and EXT INPUT connectors .. 1-7
1- 2 Physical dimensions of the 2213A Oscilloscope.................................................................................................... 1-8
2- 1 Optional power cords............................................................................................................................................ 2-1
2-2 Fuse holder and power cord connector................................................................................................................ 2-2
2-3 Power and display controls and indicators and PROBE ADJUST output
...........................................................
2-3
2-4 Vertical controls and connectors .......................................................................................................................... 2-4
2-5 Horizontal controls................................................................................................................................................ 2-5
2-6 Trigger controls, connector, and indicator............................................................................................................ 2-6
2- 7 Rear panel connector............................................................................................................................................ 2-7
3- 1 Graticule measurement markings.......................................................................................................................... 3-1
3- 2 Probe compensation ............................................................................................................................................ 3-4
4- 1 Peak-to-peak waveform voltage .......................................................................................................................... 4-2
4-2 Instantaneous voltage measurement.................................................................................................................... 4-3
4-3 Common-mode rejection ...................................................................................................................................... 4-4
4-4 Voltage ratios........................................................................................................................................................ 4-4
4-5 Time Duration........................................................................................................................................................ 4-5
4-6 Rise Tim e.............................................................................................................................................................. 4'6
4-7 Time difference between pulses on time related signals....................................................................................... 4-6
4-8 Phase difference.................................................................................................................................................... 4~7
4-9 High-resolution phase difference .......................................................................................................................... 4-8
4-10 Delayed-sweep magnification................................................................................................................................ 4-9
LIST OF TABLES
Table Page
1-1 Electrical Characteristics........................................................................................................................................ 1-2
1-2 Environmental Characteristics...............................................................................................................................
1- 3 Physical Characteristics........................................................................................................................................ 1"7
2- 1 VERT MODE Trigger Source................................................................................................................................ 2-7
1
2213A Operators
#
1
J
J
1
1 J
h
i
OPERATORS SAFETY SUMMARY
The general safety information in this part o f the summary is fo r both operating and servicing personnel. Specific warnings
and cautions will be found throughout the manual where they apply and do not appear in this summary.
Terms in This Manual
CAUTION statements identify conditions or practices that could result in damage to the equipment or other property.
WARNING statements identify conditions or practices
that could result in personal injury or loss of life.
Terms as Marked on Equipment
CAUTION indicates a personal injury hazard not immedi ately accessible as one reads the markings, or a hazard to property, including the equipment itself.
DANGER indicates a personal injury hazard immediately
accessible as one reads the marking.
Symbols in This Manual
This symbol indicates where applicable
A
cautionary or other information is to be found.
For maximum input voltage see Table 1-1.
Symbols As Marked on Equipment
A
DANGER High voltage.
Protective ground (earth) terminal.
ATTENTION Refer to manual.
LvJ
Power Source
This product is intended to operate from a power source that does not apply more than 250 volts rms between the supply conductors or between either supply conductor and
ground. A protective ground connection by way of the grounding conductor in the power cord is essential for safe operation.
Grounding the Product
This product is grounded through the grounding conductor of the power cord. To avoid electrical shock, plug the power cord into a properly wired receptacle before connecting to the product input or output terminals. A protective ground connection by way of the grounding conductor in the power cord is essential for safe operation.
Danger Arising From Loss of Ground
Upon loss of the protective-ground connection, all accessi ble conductive parts (including knobs and controls that may appear to be insulating) can render an electric shock.
Use the Proper Power Cord
Use only the power cord and connector specified for your product.
Use only a power cord that is in good condition.
For detailed information on power cords and connectors
see Figure 2-1.
Use the Proper Fuse
To avoid fire hazard, use only a fuse of the correct type, voltage rating and current rating as specified in the parts list for your product.
Do Not Operate in Explosive Atmospheres
To avoid explosion, do not operate this product in an explo sive atmosphere unless it has been specifically certified for such operation.
Do Not Remove Covers or Panels
To avoid personal injury, do not remove the product covers or panels. Do not operate the product without the covers and panels properly installed.
m
2213A Operators
The 2213A Oscilloscope.
IV
Section 12213A Operators
1 1
1
" 1
1
GENERAL INFORMATION
INTRODUCTION
The TEKTRONIX 2213A Oscilloscope is a rugged, light weight, dual-channel 60 MHz instrument that features a bright, sharply defined trace on an 80- by 100 mm cathode- ray tube (crt). Its vertical system supplies calibrated deflec tion factors from 2 mV per division to 5 V per division. Trigger circuits enable stable triggering over the full band width of the vertical system. The horizontal system provides calibrated sweep speeds from 0.5 s per division to 50 ns per division, along with a delayed-sweep feature. A X I0 magni fier circuit extends the maximum sweep speed to 5 ns per division when the SEC/DIV switch is set to 0.05 fis per division.
The instrument is shipped with the following standard
accessories:
1 Operators manual 2 Probe packages 1 Power cord
For part numbers and information about instrument ac cessories, refer to the Options and Accessories section of this manual.
The service manual and all other optional accessories are orderable from Tektronix, Inc. A local Tektronix Field Office,
representative, or the Tektronix product catalog can provide
ordering and product information.
SPECIFICATION
The following electrical characteristics (Table 1-1) are valid for the 2213A when it has been adjusted at an ambient temperature between +20°C and +30°C, has had a warm up period of at least 20 minutes, and is operating at an ambient temperature between 0°C and +50°C (unless oth
erwise noted).
Items listed in the Performance Requirements” column are verifiable qualitative or quantitative limits, while items listed in the Supplemental Information” column are either explanatory notes, calibration setup descriptions, perfor
mance characteristics for which no absolute limits are speci fied, or characteristics that are impractical to check.
Environmental characteristics are given in Table 1-2. The 2213A meets the requirements of MIL-T-28800C, para graphs 4.5.5.1.3, 4.5.5.1.4, and 4.5.5.1.2.2 for type III, Class 5 equipment, except where otherwise noted.
Physical characteristics of the instrument are listed in
Table 1-3.
REV DEC
1983
1-1
General Information 2213A Operators
Table 1-1
Electrical Characteristics
Characteristics Performance Requirements
Supplemental Information
VERTICAL DEFLECTION SYSTEM
Deflection Factor
Range
2 mV per division to 5 V per division in a 1-2-5 sequence.
5 mV per division to 5 V per division gain is adjusted with VOLTS/DIV switch set to 10 mV per division. 2 mV per division gain is adjusted with VOLTS/DIV switch set to 2 mV per division.
Accuracy
±3%
Range of VOLTS/DIV Variable Control Continuously variable between
settings. Increases deflection factor by at least 2.5 to 1.
Step Response Rise Time
Rise time is calculated from the formula:
Rise Time
0.35
0°C to + 35°C
Bandwidth (3 dB)
5 mV per Division to 5 V per Division
5.8 ns or less.
0°C to 50°C
2 mV per Division to 5 V per Division
7.0 ns or less.
Bandwidth (3 dB)
0°C to +35°C
2 mV per Division
Dc to at least 50 MHz.
Measured with a vertically centered 6-division reference signal from a 50 Q
source driving a 50 (1 coaxial cable that is terminated in 50 fl, both at the input connector and at the probe input, with the VOLTS/DIV Variable
control in the CAL detent.
5 mV per Division to 5 V per Division
Dc to at least 60 MHz.
0°C to +50°C
2 mV per Division to 5 V per Division
Dc to at least 50 MHz.
AC Coupled Lower Limit
10 Hz or less at - 3 dB.
Bandwidth Limiter
Upper limits ( - 3 dB) bandpass at 10 MHz ±15%.
Chop Mode Switching Rate
500 kHz ±30%.
Input Characteristics
Resistance
1 Mfl ±2%.
Capacitance
20 pF ±2 pF.
Maximum Safe Input Voltage 2 ^ ^
DC Coupled
400 V (dc + peak ac) or 800 V ac p-p to 10 kHz or less.
See Figure 1 -1 for derating curve.
AC Coupled
400 V (dc + peak ac) or 800 V-ac p-p
to 10 kHz or less.
1-2
REV MAR 1984
K
General Information2213A Operators
1
1
1 1 1
1 1
Table 1-1 (cont)
Characteristics
Performance Requirements
Supplemental Information
VERTICAL DEFLECTION SYSTEM (cont)
Common-Mode Rejection Ratio (CMRR)
At least 20 to 1 at 25 MHz.
Checked at 10 mV per division for common-mode signals of 6 divisions or less with VOLTS/DIV Variable control adjusted for best CMRR at 50 kHz.
Trace Shift with Attenuator Rotation 0.75 division or less.
VOLTS/DIV Variable control in CAL detent.
Trace Shift as VOLTS/DIV Variable Control is Rotated
1.0 division or less.
Trace Shift with Invert 1.5 division or less. Channel Isolation Greater than 100 to 1 at 25 MHz.
TRIGGER SYSTEM
TRIGGER Sensitivity
P-P AUTO/TV LINE and NORM
Modes
Internal
External trigger signal from a 50 Q source driving a 50 Q coaxial cable terminated in 50 Q at the input connector.
5 MHz 60 MHz
0.3 div
1.0 div
External
40 mV
150 mV
Lowest Useable Frequency in P-P AUTO Mode
20 Hz with 1.0 division internal or 100 mV external.
TV FIELD Mode
1.0 division of composite sync.
EXT INPUT
Maximum Input Voltage
400 V (dc + peak ac) or 800 V ac p-p at 10 kHz or less.
See Figure 1 -1 for derating curve.
Input Resistance
1 Mfl ±2%.
Input Capacitance
20 pF ±2.5 pF.
AC Coupled
10 Hz or less at lower - 3 dB point.
LEVEL Control Range (NORM)
INT
Can be set to any point of the trace that can be displayed.
EXT, DC
At least ± 1.6 V, 3.2 V p-p.
EXT, DC -=- 10
At least ± 16 V, 32 V p-p.
VAR HOLDOFF Control
Increases Sweep holdoff time by at least a factor of 10.
u
REV DEC 1983 1-3
General Information2213A Operators
Table 1-1 (cont)
Characteristics
Performance Requirements Supplemental Information
HORIZONTAL DEFLECTION SYSTEM
Sweep Rate 0.5 s per division to 0.05 ns per
division in a 1-2-5 sequence. XI0 magnifier extends maximum sweep speed to 5 ns per division.
Accuracy Unmagnified Magnified Sweep accuracy applies over the
center 8 divisions. Exclude the first 25 ns of the sweep for magnified sweep speeds and anything beyond the 100th magnified division.
+ 1C to +35°C
±3%
±4%
0°C to +50°C
±4%
±5%
POSITION Control Range Start of sweep to 10th division in XI or
100th divisions in XI0 will position past
the center vertical graticule line.
Sweep Linearity
±7%.
Linearity measured over any 2 of the center 8 divisions. With magnifier in XI0, exclude the first 25 ns and anything past the 100th division.
Variable Control Range
Continuously variable between calibrated settings. Extends the sweep speed by at least a factor of 2.5.
Delay Time
Delay Positions
Minimum less than 1.0 ns, 20 vs, and
0.4 ms.
MULTIPLIER
Increases delay time by at least a factor of 50.
Jitter
One part or less in 10,000 (0.01%) of
the maximum available delay time.
X-Y OPERATION (XI MAGNIFICATION)
Deflection Factory
Same as Vertical Deflection System (with VOLTS/DIV Variable controls in CAL detent).
Accuracy
X-Axis
±4%.
Measured with a dc-coupled, 5-division reference signal.
Y-Axis
Same as Vertical Deflection System.
Bandwidth (-3dB )
X-Axis
Dc to at least 2 MHz.
Measured with a 5-division reference signal.
Y-Axis
Same as Vertical Deflection System.
Phase Difference Between X- and Y-Axis Amplifiers
±3° from dc to 100 kHz.
With dc-coupled inputs.
1-4
REV DEC 1983
General Information2213A Operators
Table 1-1 (cont)
Characteristics Performance Requirements
Supplemental Information
PROBE ADJUST
Output Voltage of PROBE ADJUST Jack
0.5 V ±5%.
Repetition Rate
1 kHz ±20%.
Z-AXIS INPUT
Sensitivity
5 V causes noticeable modulation. Positive-going input decreases intensity.
Useable frequency range is dcto 10 MHz.
Maximum Safe Input Voltage
30 V (dc + peak ac) or 30 VC p-p ac at 1 kHz or less.
Input Resistance
lO kfl ±10%.
POWER SOURCE
Line Voltage Ranges
90 V to 250 V.
Line Frequency
48 Hz to 440 Hz. Maximum Power Consumption 40 W (70 VA). Line Fuse
1.0 A, 250 V, slow-blow. CATHODE-RAY TUBE
Display Area
80 by 100 mm.
Standard Phosphor
P31.
Nominal Accelerating Voltage
14 kV.
REV DEC 1983
1-5
General Information2213A Operators
Table 1-2
Environmental Characteristics
___________________________________________________________________
£
Characteristics
Description
Temperature
NOTE
The instrument meets the requirements of MIL-T-28800C, paragraphs 4.5.5.1.3,
4.5.5.1.4, and 4.5.5.1.2.2 for Type III, Class 5 equipment, except where other wise noted.
Operating
0°C to +50°C (+32°F to +122°F).
Nonoperating
-55°C to +75°C (-67°F to +167°F). Tested to MIL-T-28800C paragraphs
4.5.5.1.3 and 4.5.5.1.4, except in 4.5.5.1.3 steps 4 and 5 (0°C operating test) are
performed ahead of step 2 (55 °C nonoperating test). Equipment shall remain off upon return to room ambient during step 6. Excessive condensation shall be removed before operating during step 7.
Altitude
Operating
To 4,500 m (15,000 ft). Maximum operating temperature decreased 1 °C per 1,000 ft above 5,000 ft.
Nonoperating To 15,000 m (50,000 ft).
Humidity (Operating and Nonoperating) 5 cycles (120 hours) referenced to MIL-T-28800C paragraph 4.5.5.1.2.2 for Type
III, Class 5 instruments. Operating and non-operating at 95% +0% to -5 % a relative humidity. Operating at +50°C and +30°C. Non-operating at +30°C to
+60°C.
Vibration (Operating)
15 minutes along each of 3 major axes at a total displacementof 0.015 inch p-p (2.4 gs at 55 Hz) with frequency varied from 10 Hz to 55 Hz to 10 Hz in 1-minute sweeps. Hold for 10 minutes at 55 Hz in each of the 3 major axes. All major resonances must be above 55 Hz.
Shock (Operating and Nonoperating)
30 gs, half-sine, 11-ms duration, 3 shocks per axis each direction, for a total of 18 shocks.
EMI
Meets radiated and conducted emission requirements per VDE 0871 Class B.
1-6
REV DEC 1983
Table 1-3
Physical Characteristics
General Information2213A Operators
Characteristics
Description
Weight With Power Cord
With Cover, Probes, and Pouch
6.0 kg (13.1 lb).
Without Cover, Probes, and Pouch
5.0 kg (10.9 lb).
Domestic Shipping Weight
7.0 kg (15.4 lb).
Height
With Feet and Handles
137 mm (5.4 in).
Width
With Handle
360 mm (14.2 in).
Without Handle
327 mm (12.9 in).
Depth
With Front Cover
445 mm (17.5 in).
Without Front Cover
440 mm (17.3 in).
With Handle Extended
511 mm (20.1 in).
VOL T S
10 KHz 50 KHz 100 KHz 500 KHz 1 MHz 100 MHz
FREQUENCY
C4207-28
Figure 1-1. Maximum input voltage vs. frequency derating curve for CH 1 OR X, CH 2 OR Y, and EXT INPUT connectors.
REV DEC 1983
1-7
General Information2213A Operators
1-8
ADD DEC 1983
Figure 1-2. Physical dimensions of the 2213A Oscilloscope.
General Information2213A Operators
CALIBRATION
Instrument performance should be checked after every
2000 hours of operation or once each year if used infre
quently. A more frequent interval may be necessary if your
instrument is subjected to harsh environments or severe usage.
REPACKAGING FOR SHIPMENT
If the instrument is to be shipped to a Tektronix Service Center for service or repair, attach a tag showing; owner (with address) and the name of an individual at your firm that can be contacted. Include complete instrument serial number and a description of the service required.
Save and reuse the package in which your instrument was shipped. If the original packaging is unfit for use or not available, repackage the instrument as follows:
Surround the instrument with polyethylene sheeting to protect its finish. Obtain a carton of corrugated cardboard having a carton test strength of 275 pounds and having inside dimensions of no less than six inches more than the instrument dimensions. Cushion the instrument by tightly packing three inches of dunnage or urethane foam between carton and instrument, on all sides. Seal carton with ship ping tape or industrial stapler.
ADD DEC 1983
1-9
Section 2—2213A Operators
PREPARATION FOR USE
FIRST-TIME START UP
SAFETY
Refer to the Operators Safety Summary at the front of this manual for power source, grounding, and other safety considerations pertaining to the use of the 2213A. Before connecting the instrument to a power source, carefully read the following about line voltages, power cords, and fuses.
LINE VOLTAGE
The instrument is capable of continuous operation using input voltages that range from 90 V to 250 V nominal at frequencies from 48 Hz to 440 Hz.
POWER CORD
A detachable three-wire power cord with a three-contact plug is provided with each instrument to permit connection to both the power source and protective ground. The plug protective-ground contact connects (through the protective- ground conductoi)to the accessible metal parts of the instru ment. For electrical-shock protection, insert this plug only into a power outlet that has a securely grounded protective- ground contact.
This instrument is shipped with the required power cord as ordered by the customer. Available power-cord informa tion is illustrated in Figure 2-1, and part numbers are listed in Section 5 of this manual Tkxitact your Tektronix represen tative or local Tektronix Field Office for additional power- cord information.
LINE FUSE
The instrument fuse holder is located on the rear panel (see Figure 2-2) and contains the line fuse. The following procedure can be used to verify that the proper fuse is in stalled or to install a replacement fuse.
1. Unplug the power cord from the power-input source
(if applicable).
Plug
Configuration
Usage
Line
Voltage
Reference Standards
North
Am erican
120V /
15 A
12 0V
ANSI C73 .11 NEM A 5-15 -P IEC 83
Universal
Euro
240V /
10-16 A
24 0V
CEE |7).11,1V,VII IEC 83
UK
240V /
13 A
24 0V
BS 1363 IEC 83
Au stralian
240V /
10 A
24 0V
AS C1 12
North
Am erican
240V /
15 A
24 0V
AN SI C 73.20 NEM A 6 -15-P IEC 83
Sw itzerland
220V /
10A
22 0V SEV
Abbreviations:
ANS I - American N ational Standards Institu te AS Standards Association o f Australia BS British Standards Institutio n CEE Internationa l Com mission on Rules for the
Ap proval of Ele ctric al Equipment IEC Internatio nal Electrotechn ical Com mission NEM A Na tional Electrical M anu fa cturer's Association SEV Schw eizevischer E le ktrotechischer Verein C4732-02
Figure 2-1. Optional power cords.
2. Press in and slightly rotate the fuse-holder cap coun
terclockwise to release it.
3. Pull the cap (with the attached fuse inside) out of the
fuse holder.
4. Verify proper fuse value (1.0 A, 250 V, slow blow).
5. Reinstall the fuse (or replacement fuse) and the fuse-
holder cap.
REV APR 1985
2-1
Preparation For Use2213A Operators
LINE
FUSE
A
CAUTION
fo r c o n t in u ed fi re pr ote ct io n
REPLACE ONLY W ITH SPECIFIED TYPE ANO RATEO FUSE DISCONNECT POWER INPUT BEFORE REPLACING FUSE
LINE VOLTAGE RANGE FUSE 250V
90-25 0VAC IA SLOW
CAUTION
TO AVOID ELECTRIC SHOCK. THE POWER CORO PROTECTIVE GROUNDING CONDUCTOR " MUST BE CONNECTED TO GROUND
POWER
CORD
C4207-03
INSTRUMENT COOLING
Always maintain adequate instrument cooling. The venti lation holes on both sides of the instrument cabinet and on the rear panel must remain free of obstruction.
Figure 2-2. Fuse holder and power cord connector.
Preparation For Use2213A Operators
CONTROLS, CONNECTORS, AND INDICATORS
The following descriptions are intended to familiarize the operator with the location, operation, and function of the instrument’s controls, connectors, and indicators.
DISPLAY, POWER, AND PROBE ADJUST
Refer to Figure 2-3 for location of items 1 through 8.
M ) Internal GraticuleEliminates parallax viewing error
between the trace and graticule lines. Rise-time ampli tude and measurement points are indicated at the left edge of the graticule.
© POWER SwitchTurns instrument power on and off.
Press in for ON; press again for OFF.
( 3) Power IndicatorAn LED that illuminates when the
instrument is operating.
( T ) FOCUS ControlAdjusts for optimum display
definition.
H j j PROBE ADJUST ConnectorProvides an approxi-
mately 0.5 V, negative-going, square-wave voltage (at approximately 1 kHz) that permits an operator to compensate voltage probes and to check operation of the oscilloscope vertical system. It is not intended for verifying the accuracy of the vertical gain or time-base circuitry.
BEAM FIND SwitchWhen held in, compresses the display to within the graticule area and provides a visi ble viewing intensity to aid in locating off-screen displays.
C l j TRACE ROTATION ControlScrewdriver adjust-
ment used to align the crt trace with the horizontal graticule lines.
f 8J INTENSITY ControlDetermines the brightness of
y the sweep trace.
2-3
Figure 2-3. Power and display controls and indicators and PROBE ADJUST output.
Preparation For Use2213A Operators
VERTICAL
Refer to Figure 2-4 for location of items 9 through 16.
(jT ) CH 1 VOLTS/DIV and CH 2 VOLTS/DIV Switches—
Used to select the vertical deflection factor in a 1-2-5 sequence. To obtain a calibrated deflection factor, the VOLTS/DIV variable control must be in the calibrated (CAL) detent (fully clockwise).
IXIndicates the deflection factor selected when
using either a 1X probe or a coaxial cable.
10X PROBEIndicates the deflection factor se
lected when using a 10X probe.
® VOLTS/DIV Variable ControlsWhen rotated coun
terclockwise out of their calibrated detent positions, these controls provide continuously variable, uncalibrated deflection factors between the calibrated settings of the VOLTS/DIV switches.
POSITION ControlsUsed to vertically position the display on the crt. When the SEC/DIV switch is set to X-Y, the Channel 2 POSITION control moves the dis play vertically (Y-axis), and the Horizontal POSITION control moves the display horizontally (X-axis).
Figure 2-4. Vertical controls and connectors.
^ 2 ) Input Coupling (AC-GND-DC) SwitchesThree-po
sition switches that select the method of coupling the input signals to the instrument deflection system.
ACInput signal is capacitively coupled to the ver tical amplifier. The dc component of the input sig nal is blocked. Low-frequency limit (-3 dB point) is approximately 10 Hz.
GNDThe input of the vertical amplifier is grounded to provide a zero (ground) reference-volt age display (does not ground the input signal). This switch position allows precharging the input cou pling capacitor.
DCAll frequency components of the input signal are coupled to the vertical deflection systems.
(l3 ) CH 1 OR X and CH 2 OR Y Input Connectors
Provide for application of external signals to the verti cal deflection system or for an X-Y display. In the X-Y mode (SEC/DIV switch set to X-Y), the signal con nected to the CH 1 OR X input connector provides horizontal deflection (X-axis) and the signal connected to the CH 2 OR Y input connector provides vertical deflection (y-axis).
M4J VERTICAL MODE SwitchesTwo three-position
switches and two button switches are used to select the mode of operation for the vertical amplifier system.
CH 1Selects only the Channel 1 input signal for display.
BOTHSelects both Channel 1 and Channel 2 in put signals for display. The CH 1-BOTH-CH 2 switch must be in the BOTH position for either
ADD, ALT, or CHOP operation.
CH 2Selects only the Channel 2 input signal for display.
ADDDisplays the algebraic sum of the Channel 1 and Channel 2 input signals.
ALTAlternately displays Channel 1 and Channel 2 input signals. The alternation occurs during re trace at the end of each sweep. This mode is useful for viewing both input signals at sweep speeds from 0.05 iis per division to 0.2 ms per division.
CHOPThe display switches between the Chan nel 1 and Channel 2 input signals during the sweep. The switching rate is approximately 500 kHz. This mode is useful for viewing both Channel 1 and Channel 2 input signals at sweep
2-4
Preparation For Use—2213A Operators
speeds from 0.5 ms per division to 0.5 s per division.
BW LIMIT—When pressed in, this button switch limits the bandwidth of the vertical amplifier and the Trigger system to approximately 10 MHz. Button must be pressed a second time to release it and regain full 60 MHz bandwidth operation. Provides a method for reducing interference from high-fre quency signals when viewing low-frequency signals.
CH 2 INVERT SwitchInverts the Channel 2 dis
play when button is pressed in. Button must be pressed in a second time to release it and regain a
noninverted display.
M5J GND ConnectorProvides direct connection to the
instrument chassis ground.
M 6 ) SERIAL and Mod SlotsThe SERIAL slot is im-
V-" / printed with the instrument’s serial number. The Mod
slot contains the option number that is installed in the instrument.
HORIZONTAL
Refer to Figure 2-5 for location of items 17 through 22.
(]] J
SEC/DIV SwitchesUsed to select the sweep
speeds for the sweep generator in a 1 -2-5 sequence. To obtain calibrated sweep speeds, the SEC/DIV Variable control must be in the calibrated detent (fully clockwise).
M 8J SEC/DIV Variable ControlProvides continuously
variable, uncalibrated sweep speeds to at least 2.5 times slower than the calibrated setting. It extends the slowest sweep speed to at least 1.25 s per division.
(jtj) X10 Magnifier SwitchTo increase displayed sweep
speed by a factor of 10, pull out the SEC/DIV Variable knob. The fastest sweep speed can be extended to 5 ns per division. Push in the SEC/DIV Variable knob to regain the XI sweep speed.
(2 0 ) POSITION ControlPositions the display horizontally
in all modes.
(
2 1
) HORIZONTAL MODE Switch Three-position switch
' ^ determines the mode of operation for the horizontal
deflection system.
NO DLYHorizontal deflection is provided by the
sweep generator, without a delayed start, at a
sweep speed determined by the SEC/DIV switch setting.
INTENSHorizontal deflection is provided by the sweep generator at a sweep speed determined by the SEC/DIV switch. The sweep generator also provides an intensified zone on the display. The start of the intensified zone represents the sweep- start point when DLYD HORIZONTAL MODE is selected.
DLYDHorizontal deflection is provided by the sweep generator at a sweep speed determined by the SEC/DIV switch setting. The start of the sweep is delayed from the initial sweep-trigger point by a time determined by the setting of the DELAY TIME Range Selector switch and MULTIPLIER control.
(22 )
DELAY TIME—Two controls are used in conjunction with INTENS and DLYD HORIZONTAL MODE to se lect the amount of delay time between the start of the sweep and the beginning of the intensified zone.
Range Selector SwitchThis three-position switch selects 0.4 ms, 20 ns, and 1.0 ^s of delay time. To increase the sweep delay from the cali brated setting of the Range Selector Switch, rotate the MULTIPLIER control clockwise.
MULTIPLIER ControlProvides variable sweep delay from less than 1 to greater than 50 times the setting of the Range Selector switch.
2-5
Preparation For Use2213A Operators
TRIGGER
Refer to 2-6 for location of items 23 through 31.
(23) TRIGGER Mode SwitchesThree section switch
that determines the trigger mode for the sweep.
NORMSweep is initiated when an adequate trig ger signal is applied. In the absence of a trigger signal, no baseline trace will be present.
P-P AUTO-TV LINE Permits triggering on wave forms and television lines having repetition rates of at least 20 Hz. Sweep free-runs in the absence of an adequate trigger signal or when the repetition
rate is below 20 Hz. The range of the TRIGGER
LEVEL control is restricted to the peak-to-peak
range of the trigger signal. TV FIELDPress in both P-P AUTO and NORM
buttons. Permits triggering on television field sig nals. TRIGGER LEVEL control should be rotated fully counterclockwise when triggering on TV sig nals with negative going sync and clockwise for
positive going sync. SGL SWP RESETPress in the spring-return but
ton momentarily to arm the trigger circuit for a sin gle-sweep display. In this mode, the trigger system operates the same as NORM, except only one sweep is displayed for each trigger signal. Another sweep cannot be displayed until the SGL SWP
RESET button is momentarily pressed in again to reset the trigger circuit. This mode is useful for dis playing and photographing either nonrepetitive sig nals or signals that cause unstable conventional
~ displays (e.g., signals that vary in amplitude,
shape, or time).
( S ) TRIGD-READY IndicatorLED illuminates when ei-
v> ther P-P AUTO or NORM Trigger Mode is selected
and the Sweep has been triggered (TRIG’D). In single sweep display, the LED illuminates to indicate that the Trigger circuit is armed (READY).
(25) TRIGGER LEVEL ControlSelects the amplitude
point on the trigger signal at which the sweep is triggered.
(26) SLOPE SwitchSelects the slope of the signal that
triggers the sweep.
OUTWhen button is released out, sweep is trig gered from the positive-going slope of the trigger signal.
INWhen button is pressed in, sweep is triggered from the negative-going slope of the trigger signal.
(27) SOURCE SwitchDetermines the source of the trig-
ger signal that is coupled to the input of the trigger circuit.
INTPermits triggering on signals that are applied to the CH 1 OR X and CH 2 OR Y input connec tors. The source of the internal signal is selected by the INT switch.
LINEThe power-source waveform is the source of the trigger signal. This trigger source is useful when vertical input signals are time related (multi ple or submultiple) to the frequency of the power- input source voltage.
EXTPermits triggering on signals applied to the
EXT INPUT connector.
(2 8 ) INT SwitchSelects the source of the internal trig-
gering signal when the SOURCE switch is set to INT.
CH 1The signal applied to the CH 1 OR X input connector is the source of the trigger signal.
VERT MODEThe internal trigger source is deter
mined by the signals selected for display by the VERTICAL MODE switches. See Table 2-1 for VERT MODE trigger source.
"N
--------------------------------------------------------------_
_________________
4734-os
Figure 2-6. Trigger controls, connector, and indicator.
2-6
Preparation For Use—2213A Operators
speeds from 0.5 ms per division to 0.5 s per division.
BW LIMIT—When pressed in, this button switch limits the bandwidth of the vertical amplifier and the Trigger system to approximately 10 MHz. Button must be pressed a second time to release it and regain full 60 MHz bandwidth operation. Provides a method for reducing interference from high-fre quency signals when viewing low-frequency signals.
CH 2 INVERT SwitchInverts the Channel 2 dis
play when button is pressed in. Button must be pressed in a second time to release it and regain a
noninverted display.
M5J GND ConnectorProvides direct connection to the
instrument chassis ground.
M 6 ) SERIAL and Mod SlotsThe SERIAL slot is im-
V-" / printed with the instrument’s serial number. The Mod
slot contains the option number that is installed in the instrument.
HORIZONTAL
Refer to Figure 2-5 for location of items 17 through 22.
(]] J
SEC/DIV SwitchesUsed to select the sweep
speeds for the sweep generator in a 1 -2-5 sequence. To obtain calibrated sweep speeds, the SEC/DIV Variable control must be in the calibrated detent (fully clockwise).
M 8J SEC/DIV Variable ControlProvides continuously
variable, uncalibrated sweep speeds to at least 2.5 times slower than the calibrated setting. It extends the slowest sweep speed to at least 1.25 s per division.
(jtj) X10 Magnifier SwitchTo increase displayed sweep
speed by a factor of 10, pull out the SEC/DIV Variable knob. The fastest sweep speed can be extended to 5 ns per division. Push in the SEC/DIV Variable knob to regain the XI sweep speed.
(2 0 ) POSITION ControlPositions the display horizontally
in all modes.
(
2 1
) HORIZONTAL MODE Switch Three-position switch
' ^ determines the mode of operation for the horizontal
deflection system.
NO DLYHorizontal deflection is provided by the
sweep generator, without a delayed start, at a
sweep speed determined by the SEC/DIV switch setting.
INTENSHorizontal deflection is provided by the sweep generator at a sweep speed determined by the SEC/DIV switch. The sweep generator also provides an intensified zone on the display. The start of the intensified zone represents the sweep- start point when DLYD HORIZONTAL MODE is selected.
DLYDHorizontal deflection is provided by the sweep generator at a sweep speed determined by the SEC/DIV switch setting. The start of the sweep is delayed from the initial sweep-trigger point by a time determined by the setting of the DELAY TIME Range Selector switch and MULTIPLIER control.
(22 )
DELAY TIME—Two controls are used in conjunction with INTENS and DLYD HORIZONTAL MODE to se lect the amount of delay time between the start of the sweep and the beginning of the intensified zone.
Range Selector SwitchThis three-position switch selects 0.4 ms, 20 ns, and 1.0 ^s of delay time. To increase the sweep delay from the cali brated setting of the Range Selector Switch, rotate the MULTIPLIER control clockwise.
MULTIPLIER ControlProvides variable sweep delay from less than 1 to greater than 50 times the setting of the Range Selector switch.
2-5
Section 32213A Operators
OPERATORS FAMILIARIZATION
GENERAL OPERATING INFORMATION
GRATICULE
The graticule is internally marked on the faceplate of the crt to enable accurate measurements without parallax error (see Figure 3-1). It is marked with eight vertical and ten hori zontal major divisions. Each major division is divided into five subdivisions. The vertical deflection factors and horizon tal timing are calibrated to the graticule so that accurate measurements can be made directly from the crt. Also, per centage markers for the measurement of rise and fall times are located on the left side of the graticule.
GROUNDING
The most reliable signal measurements are made when the 2213A and the unit under test are connected by a com mon reference (ground lead), in addition to the signal lead or
1ST OR LEFT 11TH OR RIGHT
VERTICAL VERTICAL
GRATICULE GRATICULE
LINE L IN E /
)
°
,
RISE AND
FALL TIME _P1.
MEASUREMENT
PERCENTAGE v '
MARKERS
CENTER
HORIZONTAL
LTjF". GRATICULE riCJAL line
ICULE
NE
C4207-09
probe. The probe’s ground lead provides the best grounding method for signal interconnection and ensures the maxi mum amount of signal-lead shielding in the probe cable. A separate ground lead can also be connected from the unit under test to the oscilloscope GND connector located on the front panel.
SIGNAL CONNECTIONS
Generally, probes offer the most convenient means of connecting an input signal to the instrument. They are shielded to prevent pickup of electromagnetic interference, and the supplied 10X probe offers a high input impedance that minimizes circuit loading. This allows the circuit under test to operate with a minimum of change from its normal condition as measurements are being made.
Coaxial cables may also be used to connect signals to the input connectors, but they may have considerable effect on the accuracy of a displayed waveform. To maintain the original frequency characteristics of an applied signal, only high-quality, low-loss coaxial cables should be used. Coaxial cables should be terminated at both ends in their characteristic impedance. If this is not possible, use suitable impedance-matching devices.
INPUT COUPLING
CAPACITOR PRECHARGING
When the Input Coupling switch is set to GND, the input signal is connected to ground through the input coupling capacitor in series with a 1 Mfi resistor to form a precharging network. This network allows the input coupling capacitor to charge to the average dc-voltage level of the signal applied to the probe. Thus any large voltage transients that may accidentally be generated will not be applied to the amplifier input when the Input Coupling switch is moved from GND to AC. The precharging network also provides a measure of protection to the external circuitry by reducing the current levels that can be drawn from the ex ternal circuitry during capacitor charging.
The following procedure should be used whenever the probe tip is connected to a signal source having a different
dc level than that previously applied, especially if the dc-
3-1
Figure 3-1. Graticule measurement markings.
Operators Familiarization—2213A Operators
level difference is more than 10 times the VOLTS/DIV switch setting:
1. Set the Input Coupling switch to GND.
2. Insert the probe tip into the oscilloscope GND connec tor and wait several seconds for the input coupling capacitor to discharge.
3. Connect the probe tip to the signal source and wait
several seconds for the input coupling capacitor to charge.
4. Set the Input Coupling switch to AC. The display will remain on the screen, and the ac component of the signal can be measured in the normal manner.
Operators Familiarization—2213A Operators
OPERATORS ADJUSTMENTS
INTRODUCTION trig g e r
To verify the operation and accuracy of your instrument before making measurements, perform the following adjust ment procedures. If adjustments are required beyond the scope of the operators’s adjustments, refer the instrument
to a qualified service technician.
VAR HOLDOFF Mode SLOPE LEVEL INT SOURCE EXT COUPLING
NORM P-P AUTO OUT Midrange VERT MODE INT
AC
Before proceeding with these instructions, refer to
Preparation for Use” (Section 2).
Verify that the POWER switch is OFF (button out), then
plug the power cord into the power-source outlet.
2. Press in the POWER switch button (ON) and allow the instrument to warm up (20 minutes is recommended for maximum accuracy).
3. Adjust the INTENSITY control for desired display
If indications specified in these procedures cannot be ob- brightness,
tained, refer the instrument to a qualified service technician.
4. Adjust the Vertical and Horizontal POSITION controls as needed to center the trace on the screen.
BASELINE TRACE
First obtain a baseline trace, using the following
procedure.
1. Preset the instrument front-panel controls as follows:
TRACE ROTATION
Normally, the resulting trace will be parallel to the center horizontal graticule line, and the Trace Rotation adjustment should not be required. If adjustment is needed, perform the following procedure:
Display
INTENSITY FOCUS
Fully counterclockwise 1 Preset instrument controls and obtain a baseline Midrange trace.
Vertical (Both Channels)
POSITION VERTICAL MODE
BW LIMIT VOLTS/DIV VOLTS/DIV Variable CH 2 INVERT
Input Coupling
Midrange CH 1 Off (button out) 50 mV CAL detent Off (button out) AC
2. Use the Channel 1 POSITION control to move the
baseline trace to the center horizontal graticule line.
3. If the resulting trace is not parallel to the center hori zontal graticule line, use small flat-bit screwdriver to adjust the TRACE ROTATION control and align the trace with the center horizontal graticule line.
Horizontal
POSITION HORIZONTAL MODE SEC/DIV SEC/DIV Variable XI0 Magnifier Range Selector MULTIPLIER
Midrange NO DLY
0.5 ms CAL detent Off (knob in)
0.4 ms <X1
PROBE COMPENSATION
Misadjustment of probe compensation is a common
source of measurement error. Most attenuator probes are equipped with a compensation adjustment. To ensure opti mum measurement accuracy, always compensate the oscil loscope probes before making measurements. Probe compensation is accomplished as follows:
3-3
Operators Familiarization—2213A Operators
1. Preset instrument controls and obtain a baseline
trace.
2. Connect the two 10X probes (supplied with the in
strument) to the CH 1 and CH 2 input connectors.
3. Set both VOLTS/DIV switches to 10 mV and set
both Input Coupling switches to DC.
4. Select CH 1 VERTICAL MODE and insert the tip of
the Channel 1 probe into the PROBE ADJUST output jack.
5. Using the approximately 1 kHz PROBE ADJUST square-wave signal as the input, obtain a 5-division display of the signal.
6. Set the SEC/DIV switch to display several cycles of the PROBE ADJUST signal. Use the Channel 1 POSITION control to vertically center the display.
7. Check the waveform presentation for overshoot and rolloff (see Figure 3-2). If necessary, adjust the probe com pensation for flat tops on the waveforms. Refer to the in structions supplied with the probe for details of compensation adjustment.
Figure 3-2. Probe compensation.
8. Select CH 2 VERTICAL MODE and connect the
Channel 2 probe tip to the PROBE ADJUST output jack.
9. Use the Channel 2 POSITION to vertically center the
display and repeat step 7 for the Channel 2 probe.
10. Disconnect the probes from the instrument.
L
(
d
i: L
L L
3-4
Section 4—2213A Operators
OPERATING PROCEDURES
BASIC APPLICATIONS
INTRODUCTION
After becoming familiar with the capabilities of the 2213A Oscilloscope an operator can then easily develop conve nient methods for making particular measurements. The in formation in this section is designed to enhance operator understanding and to assist in developing efficient tech niques for making specific measurements. Recommended methods for making basic measurements with your instru ment are described in the procedures contained in this section.
When a procedure first calls for presetting instrument controls and obtaining a baseline trace, refer to the Oper ators Adjustments" part in Section 3 and perform steps 1 through 4 under Baseline Trace”.
INDEX TO BASIC
APPLICATION PROCEDURES
VOLTAGE MEASUREMENTS
................................
4-1
AC Peak-to-Peak Voltage
..................................
4-1
Instantaneous Voltage
........................................
4-2
Algebraic Addition................................................. 4-3
Common-Mode Rejection
....................................
4-3
Amplitude Comparison
........................................
4-4
TIME MEASUREMENTS
........................................
4-4
Time Duration....................................................... 4-4
Frequency............................................................. 4-5
Rise Time ............................................................. 4-5
Time Difference Between Pulses On Time
Related Signals..................................................... 4-6
Phase Difference
................................................
4-7
TELEVISION DISPLAYS
..........................................
4-8
TV Line Signal ........ 4-8
TV Field Signal ..................................................... 4-8
DELAYED-SWEEP MAGNIFICATION
......................
4-9
VOLTAGE MEASUREMENTS
AC Peak-to-Peak Voltage
To make a peak-to-peak voltage measurement, use the
following procedure:
NOTE
This procedure may also be used to make voltage measurements between any two points on the waveform.
1. Preset instrument controls and obtain a baseline
trace.
2. Apply the ac signal to either vertical-channel input connector and set the VERTICAL MODE switch to display the channel used.
3. Set the appropriate VOLTS/DIV switch to display about five divisions of the waveform, ensuring that the VOLTS/DIV Variable control is in the CAL detent.
4. Adjust the TRIGGER LEVEL control to obtain a stable
display.
5. Set the SEC/DIV switch to a position that displays
several cycles of the waveform.
6. Vertically position the display so that the negative
peak of the waveform coincides with one of the horizontal graticule lines (see Figure 4-1, Point A).
7. Horizontally position the display so that one of the
positive peaks coincides with the center vertical graticule line (see Figure 4-1, Point B).
8. Measure the vertical deflection from peak-to-peak
(see Figure 4-1, Point A to Point B).
4-1
Operating Procedures2213A Operators
NOTE
If the amplitude measurement is critical or if the trace is thick (as a result o f hum or noise on the signal), a more accurate value can be obtained by measuring
from the top of a peak to the top of a valley. This will eliminate trace thickness from the measurement.
9. Calculate the peak-to-peak voltage, using the follow
ing formula:
VOLTS/DIV
vertical switch setting
Volts (p-p) = deflection x indicated by 1X
(divisions) (or 10X PROBE when
10X probe is used)
2. Apply the signal to either vertical-channel input con nector and set the VERTICAL MODE switch to display the channel used.
3. Verify that the VOLTS/DIV Variable control is in the CAL detent and set the Input Coupling switch to GND.
4. Vertically position the baseline trace to the center horizontal graticule line. This establishes the ground refer ence location.
EXAMPLE: The measured peak-to-peak vertical deflec tion is 4.6 divisions (see Figure 4-1) using a 10X attenua tor probe with the VOLTS/DIV switch set to 5 V (at 10X
PROBE setting).
Substituting the given values:
NOTE
If measurements are to be made relative to a voltage level other than ground, set the Input Coupling switch to DC instead, and apply the reference voltage to the input connector. Then position the trace to the refer ence (horizontal graticule) line.
Volts (p-p) = 4.6 div x 5 V/div = 23 V.
Instantaneous Voltage
To measure instantaneous level at a given point on a
waveform, referred to ground, use the following procedure:
5. Set the COUPLING switch to DC. Points on the waveform above the ground reference location are positive. Those points below are negative.
1. Preset instrument controls and obtain a baseline
trace.
NOTE
If using Channel 2, ensure that the Channel 2 INVERT switch is in its noninverting mode (button out).
6. If necessary, repeat Step 4 using a different refer ence line which allows the waveform in Step 5 to be dis played on screen.
7. Adjust the TRIGGER LEVEL control to obtain a sta
ble display.
8. Set the SEC/DIV switch to a position that displays
several cycles of the signal.
9. Measure the divisions of vertical deflection between the ground reference line and the point on the waveform at which the level is to be determined (see Figure 4-2).
4-2
Figure 4-1. Peak-to-Peak waveform voltage.
Operating Procedures2213A Operators
10. Calculate the instantaneous voltage, using the fol
lowing formula:
VOLTS/DIV
Instanta- vertical switch setting
neous = deflection x polarity x indicated by IX
Voltage (divisions) (+ or ) (or 10X when 10X
probe is used)
EXAMPLE: The measured vertical deflection from the reference line is 4.6 divisions (see Figure 4-2), the wave form point is above the reference line, a 10X attenuator probe is being used, and the VOLTS/DIV switch is set to 2 V (at 10X probe setting).
Substituting the given values:
Instantaneous Voltage = 4.6 div x (+1) x 2V/div =
9.2 V.
Algebraic Addition
With the VERTICAL MODE switches set to BOTH and ADD, the waveform displayed is the algebraic sum of the signals applied to the Channel 1 and Channel 2 inputs (CH 1 + CH 2). If the Channel 2 INVERT button is pressed in, the waveform displayed is the difference between the signals applied to the Channel 1 and Channel 2 inputs (CH 1
- CH 2). When both VOLTS/DIV switches are set to the same deflection factor, the deflection factor in the ADD mode is equal to the deflection factor indicated by either VOLTS/DIV switch.
The following general precautions should be observed
when using the ADD mode.
a. Do not exceed the input voltage rating of the
oscilloscope.
b. Do not apply signals that exceed the equivalent of about eight times the VOLTS/DIV switch settings, since large voltages may distort the display. For example, with a VOLTS/DIV switch setting of 0.5 V, the voltage applied to that channel should not exceed approximately 4 volts.
c. Use Channel 1 and Channel 2 POSITION control settings which most nearly position the signal on each chan nel to midscreen, when viewed in either CH 1 or CH 2
VERTICAL MODE. This ensures the greatest dynamic range for ADD mode operation.
t
- ( r .\
Mtl-tMtNUt
LINE
-
J
t
VERTICAL
-
L
DEFLECTION
1
1
MEASURE POSITIVE
AMPLITUDE
® TO (D
OR NEGATIVE AMPLITUDE
© TO (D
C4732-08
Figure 4-2. Instantaneous voltage measurement.
d. To attain similar response from each channel, set both the Channel 1 and Channel 2 Input Coupling switches to the same position.
Common-Mode Rejection
The following procedure shows how to eliminate un wanted ac input-power frequency components. Similar methods could be used either to eliminate other unwanted
frequency components or to provide a dc offset.
1. Preset instrument controls and obtain a baseline
trace.
2. Apply the signal containing the unwanted line-fre
quency components to the CH 1 input connector.
3. Apply a line-frequency signal to the CH 2 input con nector. To maximize cancellation, the signal applied to Channel 2 must be in phase with the unwanted line-fre quency component on the Channel 1 input.
4. Select BOTH and ALT VERTICAL MODE and set both VOLTS/DIV switches to produce displays of approxi mately 4- or 5-divisions in amplitude.
5. Adjust the CH 2 VOLTS/DIV switch and CH 2 VOLTS/DIV Variable control so that the Channel 2 display is approximately the same amplitude as the undesired portion of the Channel 1 display (see Figure 4-3A).
4-3
Operating Procedures2213A Operators
CHI SIGNAL
V
^
m N
>
WITH UNWANTED
\
y
\
LINE FREQUENCY
COMPONENT
CH2 SIGNAL
FROM LINE
FREQUENCY
SOURCE
-s
: *
~
*
* ^
%
x
X
-
\
(A) CH I AND CH2 SIGNALS.
SIGNAL WITH
LINE FREQUENCY
COMPONENT
CANCELED
OUT
'
(B) RESULTANT SIGNAL.
C4732-09
Figure 4-3. Common-mode rejection.
6. Select ADD VERTICAL MODE and press in the INVERT button, and slightly readjust the CH 2 VOLTS/DIV Variable control for maximum cancellation of the undesired signal component (see Figure 4-3B).
Amplitude Comparison (Ratio)
In some applications it may be necessary to establish a set of deflection factors other than those indicated by the VOLTS/DIV switch settings. This is useful for comparing unknown signals to a reference signal of known amplitude.
To accomplish this, a reference signal of known amplitude is first set to an exact number of vertical divisions by adjusting the VOLTS/DIV switch and Variable control. Unknown sig
nals can then be quickly and accurately compared with the reference signal without disturbing the setting of the
VOLTS/DIV Variable control. This procedure is as follows:
3. Set the amplitude of the reference signal to five verti cal divisions by adjusting the VOLTS/DIV switch and VOLTS/DIV Variable control.
4. Disconnect the reference signal and apply the un known signal to be measured to the same channel input. Adjust the vertical position of the waveform so that its bot tom edge just touches the 0% line on the crt.
5. Horizontally position the waveform so that its top most features cross the center vertical graticule line (see
Figure 4-4).
6. Read the percent ratio directly from the graduations of the center line, referring to the 0% and 100% percentage marks on the left edge of the graticule (1 minor division equals 4% for a 5-division display).
REFERENCE
1
SIGNAL
UNKNOWN
SIGNAL
....
1 ;
....
-
fl
4
1
X
_L _L X
:
READ
PRECENT
RATIO
C4732-10
Figure 4-4. Voltage ratios.
TIME MEASUREMENTS
Time Duration
To measure time between two points on a waveform,
use the following procedure.
1. Preset instrument controls and obtain a baseline / 1. Preset instrument controls and obtain a baseline
trace. trace.
2. Apply the reference signal to either vertical channel input and set the VERTICAL MODE switch to display the channel used.
2. Apply the signal to either vertical-channel input con nector and set the VERTICAL MODE switch to display the channel used.
4-4
Operating Procedures2213A Operators
3. Adjust the TRIGGER LEVEL control to obtain a stable
display.
4. Set the SEC/DIV switch to display one complete pe
riod of the waveform. Ensure that the SEC/DIV Variable
control is in the CAL detent.
5. Position the display to place the time-measurement
points on the center horizontal graticule line (see Figure 4-5).
6. Measure the Horizontal distance between the time-
measurement points.
7. Calculate time duration using the following formula:
horizontal SEC/DIV
distance x switch
Time (division) setting
Duration =
_________________________
__
magnification factor
EXAMPLE: The distance between the time measurement
points is 8.3 divisions (see Figure 4-5), and the SEC/DIV switch is set to 2 ms per division. The X I0 Magnifier is off (knob in).
Substituting the given values:
Time Duration = 8.3 div x 2 ms/div = 16.6 ms
Frequency
The frequency of a recurrent signal can be determined
from its time-duration measurement as follows:
1. Measure the time duration of one waveform cycle us
ing the preceding Time Duration” measurement procedure.
2. Calculate the reciprocal of the time-duration value to
determine the frequency of the waveform.
EXAMPLE: The signal in Figure 4-5 has a time duration of 16.6 ms.
Calculating the reciprocal of time duration:
Frequency =
-----
- = -
--------
= 60 Hz
time duration 16.6 ms
Rise Time
Rise-time measurements use the same methods as time duration, except that the measurements are made between the 10% and 90% points of the low to high transition of the
selected waveform (see Figure 4-6) Fall time is measured between the 90% and 10% points of the high to low transi tion of the waveform.
1. Preset instrument controls and obtain a baseline
trace.
2. Apply an exact 5-division signal to either vertical- channel input connector and set the VERTICAL MODE switch to display the channel used.
4-5
Figure 4-5. Time duration.
Figure 4-6. Rise time.
Operating Procedures2213A Operators
3. Set the appropriate VOLTS/DIV switch and variable
control for an exact 5-division display.
5. Select BOTH VERTICAL MODE; then select either
ALT or CHOP, depending on the frequency of input signals.
4. Vertically position the trace so that the zero reference of the waveform touches the 0% graticule line and the top of the waveform touches the 100% graticule line.
5. Horizontally position the display so the 10% point on
the waveform intersects the second vertical graticule line.
6. If the two signals are of opposite polarity, press in the Channel 2 INVERT button to invert the Channel 2 display (signals may be of opposite polarity due to 180° phase difference).
7. Adjust the TRIGGER LEVEL control for a stable
display.
6. Measure the horizontal distance between the 10% and 90% points (between Points A and B of Figure 4-6) and calculate the time duration using the following formula:
horizontal SEC/DIV
distance x switch
Rise (divisions) setting
Time = ________________________
magnification factor
Example: The horizontal distance between the 10% and 90% points is 5 divisions, and the SEC/DIV switch is set to 1 fis per division. The XI0 magnifier knob is off (knob in).
Substituting the given values in the formula:
Rise Time
5 div x 1 nsldiv
i
5 ns
8. Set the SEC/DIV switch to a sweep speed which pro vides three or more divisions of horizontal separation be tween the reference points on the two displays. Center each of the displays vertically (see Figure 4-7)
9. Measure the horizontal difference between the two signal reference points and calculate the time difference us ing the following formula:
SEC/DIV horizontal
switch x difference
Time setting (divisions)
Difference = ________________________
magnification factor
EXAMPLE: The SEC/DIV switch is set to 50 ^s per divi sion, the XI0 Magnifier is on (button out) and the hori
zontal difference between waveform measurement
points is 4.5 divisions.
Time Difference Between Pulses On Time-Related Signals
The calibrated sweep speed and dual-trace features of the 2213A allow measurement of the time difference be tween two separate events. To measure time difference,
use the following procedure:
1. Preset instrument controls and obtain a baseline
trace, then set the TRIGGER SOURCE switch to CH 1.
2. Set both Input Coupling switches to the same posi
tion, depending on the type of input coupling desired.
3. Using either probes or cables with equal tyne delays, connect a known reference signal to the Channel 1 input and the comparison signal to the Channel 2 input.
4. Set both VOLTS/DIV switches for 4- or 5-division
displays.
Substituting the given values in the formula:
Time
Difference
50 Ats/div x 4.5 div
10
22.5 ns
Figure 4-7. Time difference between pulses on time-related signals.
4-6
Operating Procedures2213A Operators
Phase Difference
In a similar manner to Time Difference Between Two Time-Related Pulses” phase comparison between two sig nals of the same frequency can be made using the dual trace feature of the 2213A. This method of phase difference measurement can be used up to the frequency limit of the vertical deflection system. To make a phase comparison, use the following procedure:
1. Preset instrument controls and obtain a baseline
trace, then set the INT switch to CH 1.
2. Set both Input Coupling switches to the same posi
tion, depending on the type of input coupling desired.
CHANNEL 1 CHANNEL 2
MEASURE
TIME FROM
® TO ( |)
HORIZONTAL
DIFFERENCE
C4207-19
3. Using either probes or cables with equal time delays, connect a known reference signal to the Channel 1 input and the unknown signal to the Channel 2 input.
4. Select BOTH VERTICAL MODE; then select either ALT or CHOP, depending on the frequency of input signals. The reference signal should precede the comparison signal in time.
5. If the two signals are of opposite polarity, press in the Channel 2 INVERT button to invert the Channel 2
display.
Figure 4-8. Phase difference.
horizontal
Phase horizontal graticule
Difference = difference x
(divisions) calibration
(deg/div)
Example: The horizontal difference is 0.6 division with a graticule calibration of 45° per division as shown in Fig ure 4-8.
6. Set both VOLTS/DIV switches and both Variable
controls so the displays are equal in amplitude.
7. Adjust the TRIGGER LEVEL control for a stable
display.
8. Set the SEC/DIV switch to a sweep speed which
displays about one full cycle on the waveforms.
9. Position the displays and adjust the SEC/DIV Vari able control so that one reference-signal cycle occupies ex actly 8 horizontal graticule divisions at the 50% rise-time points (see Figure 4-8). Each division of the graticule now represents 45° of the cycle (36
h- 8 divisions), and Jthe
horizontal graticule calibration can be stated as 45o/per division.
10. Measure the horizontal difference between cor responding points on the waveforms at a common horizon tal graticule line (50% of rise time) and calculate the phase difference using the following formula:
Substituting the given values into the phase difference
formula:
Phase difference = 0.6 div x 45°/div = 27°.
More accurate phase measurements can be made by us ing the X I0 Magnifier function to increase the sweep speed without changing the SEC/DIV Variable control setting.
EXAMPLE: If the sweep speed were increased 10 times with the magnifier (XI0 Magnifier button out), the magni fied horizontal graticule calibration would be 45°/division divided by 10 (or 4./division). Figure 4-9 shows the same signals illustrated in Figure 4-8, but magnifying the displays results in a horizontal difference of 6 divisions between the two signals.
Substituting the given values in the phase difference
formula:
Phase difference = 6 div x 4.5°/div = 27°.
Operating Procedures2213A Operators
6. Select the DLYD HORIZONTAL MODE and increase the sweep speed to magnify the intensified portion of the sweep (see Figure 4-10).
7. The apparent sweep magnification can be calculated
from the following formula:
Apparent
Delayed Sweep
Magnification
initial SEC/DIV setting
second SEC DIV setting
EXAMPLE: Determine the apparent magnification of a display with an initial SEC/DIV switch setting of 0.1 ms per division and the second SEC/DIV switch setting of
1 ixs per division.
Substituting the given values:
Apparent l x 1 0 4s
Magnification 1 x 106s
102 =100
4-10
Section 5—2213A Operators
OPTIONS AND
INTRODUCTION
This section contains a general description of instrument options available at the time of publication of this manual. Also included is a complete list (with Tektronix part number) of standard accessories included with each instrument and a partial list of optional accessories. Additional information about instrument options, option availability, and other ac cessories can be obtained either by consulting the current Tektronix Product Catalog or by contacting your Tektronix Field Office or representative.
OPTIONS
There are currently no options available for the 2213A.
INTERNATIONAL POWER CORDS
Instruments are shipped with the detachable power-cord configuration ordered by the customer. Descriptive informa tion about the international power-cords is provided in Sec tion 2, Preparation for Use”. The following list identifies the Tektronix part numbers for the available power cords.
ACCESSORIES
STANDARD ACCESSORIES
The following standard accessories are provided with
each instrument.
Qty Description Order
2 Probes, 10X 1,5-meter length with
accessories ........................................ 010-6122-01
1 Power Cord
.........................................
161-0104-00
1 Operators Manual
...............................
070-4734-00
OPTIONAL ACCESSORIES
The following optional accessories are recommended for
use with the 2213A Oscilloscope.
Description Order
Service Manual
.........................................
070-4733-00
Protective Front Panel Cover
...................
200-2520-00
Cord Wrap and Storage Pouch
............
016-0677-00
Protective Front Panel Cover, Cord
Wrap, and Storage Pouch
.........................
020-0672-00
Carrying Strap......................................... 346-0199-00
Carrying Case
...........................................
016-0694-00 Low-Cost, General-Purpose Camera .. . C-5C Option 04
Description Order
Universal Euro 10-16 A, 50 Hz
..................
020-0859-00
UK 240 V/13A, 50 Hz
................................
020-0860-00
Australian 240 V/10A, 50 H z
......................
020-0861-00
North American 240 V/15A, 60 H z
............
020-0862-00
Switzerland 220 V/10A, 50 Hz
..................
020-0863-00
Rackmount Adapter Kit
..................
016-0466-0C
Model 200C
SCOPE-MOBILE CARTOccupies less than 18 inches of aisle space, with stor age area in base
....................................
REV MAR 1984
5-1
Limited license: Tektronix, Inc. grants permission to reproduce this drawing as a visual aid for use in instrument operator training.
2213A Operators
Ife ktronix
COM M IT TED TO EX CE LLENCE
b -
I
Copyright © 1983, Tektronix, Inc.
All rights reserved.
Loading...