Tektronix 2213 schematic

COMMITTED TO EXCELLENCE
PLEASE CHECK FOR CHANGE INFORMATION AT THE REAR OF THIS MANUAL.
2213
OSCILLOSCOPE
OPERATORS
INSTRUCTION MANUAL
500
97077
Serial Number __________
070-3397-00 Product Group 46
First Printing JUN 1981 Revised AUG 1982
2213 Operators
CONTROLS, CONNECTORS, AND INDICATORS
The following descriptions are intended to familiarize the operator with the location, operation, and function of the instrument’s controls, connectors, and indicators.
POWER, DISPLAY, AND
PROBE ADJUST
Refer to Figure 3 for location of items 1 through 7.
1
Internal Graticule-Eliminates parallax viewing error
0
between the trace and graticule lines. Rise-time amplitude and measurement points are indicated at the left edge of the graticule.
2
POWER Switch-Turns instrument power on and off.
0
Press in for ON; press again for OFF.
3
AUTO FOCUS Control-Adjusts display for optimum
0
definition. Once set, the focus of the crt display will
be maintained as changes occur in the intensity level of the trace.
PROBE ADJ
4
0
0.5
V,
negative going, square-wave voltage (at ap­proximately 1 pensate voltage probes and to check operation of the oscilloscope vertical system. It is not intended to ver­ify the accuracy of the vertical gain or time-base calibration.
BEAM FIND Switch-When held in, compresses the
5
0
display to within the graticule area and provides a visible viewing intensity to aid in locating off-screen displays.
TRACE ROTATION Control-Screwdriver control
6
0
used to align the crt trace with the horizontal graticule lines.
AUTO INTENSITY Control-Adjusts brightness of
7
0
the crt display. This control has no effect when the
BEAM FIND switch is pressed in. Once the control
is set, intensity is automatically maintained at approximately the same level between SEC/DIV switch settings from 0.5 ms per division to 0.05 per division.
Connector-Provides an approximately
kHz)
that permits the operator to com-
pus
Figure 3. Power, display, and probe adjust controls, connector, and
indicator.
4
VERTICAL
Refer to Figure 4 for location of items 8 through 16.
8
SERIAL and Mod Slots-The SERIAL slot is im-
0
printed with the instrument’s serial number. The Mod slot contains the option number that has been installed in the instrument.
9
CH 1 OR X and CH 2 OR Y Connectors-Provide
0
for application of external signals to the inputs of the vertical deflection system or for an X-Y display.
In the X-Y mode, the signal connected to the CH 1 OR X connector provides horizontal deflection, and the signal connected to the CH 2 OR Y con­nector provides vertical deflection.
10
GND Connector-Provides direct connection to
0
instrument chassis ground.
REV NOV 1981
MHz OSCILLOSCOPE
Figure 4. Vertical controls and connectors.
3397-05
2213 Operators
10X
PROBE-Indicates the deflection factor
selected when using a 10X probe.
VOLTS/DIV
13
0
ter clockwise out of their detent positions, these
controls provide continuously variable, uncalibrated deflection factors between the calibrated settings of the
VOLTS/DIV brated deflection factor to 25 volts per division with IX probe (a range of at least 2.51).
INVERT Switch-Inverts the Channel 2 display when
14
0
button is pressed in. Push button must be pressed in a second time to release it and regain a noninverted display.
VERTICAL MODE Switches-Two three-position
15
0
switches are used to select the
the vertical amplifier system.
CH I-Selects only the Channel 1 input signal for
display.
BOTH-Selects both Channel 1 and Channel 2 input signals be selected for either ADD, ALT, or CHOP
operation.
Variable Controls-When
switches.Extends maximum uncali-
for display.
The BOTH position must
rotated
mode of operation
coun-
for
u
Input Coupling (AC-GND-DC) Switches-Used to
select the method of coupling input signals to the vertical deflection system.
AC-Input signal is capacitively coupled to the vertical amplifier. The dc component of the input signal is blocked. Low-frequency limit (-3 point) is approximately 10 Hz.
GND-The input of the vertical amplifier is grounded to provide a zero (ground) voltage display (does not ground the input signal). This switch position allows coupling capacitor.
DC-All frequency components of the input
signal are coupled to the vertical deflection system.
12
CH 1
0
VOLTS/DlV Used to select the vertical deflection factor in a l-2-5 sequence. To obtain a calibrated deflection factor, the
VOLTS/DIV
1X
PROBE-Indicates the deflection factor
selected when using either a 1X probe or a coaxial
cable.
and CH 2
variable control must be in detent.
precharging
VOLTS/DIV Switches-
dB
reference-
the input
CH
2-Selects
display.
ADD-Displays the algebraic sum of the Channel 1 and Channel 2 input signals.
ALT-Alternately displays Channel 1 and Channel 2 input signals. The alternation occurs during
retrace at the end of each sweep.
is useful for viewing both input signals at sweep speeds from 0.05 ps per division to 0.2 ms per division.
CHOP-The display switches between the Chan-
nel 1 and Channel 2 input signals during the sweep. The switching rate is approximately 250
kHz.
This mode is useful for viewing both Channel
1 and Channel 2 input signals at sweep speeds from 0.5 ms per division to 0.5 s per division.
POSITION Controls-Used to vertically position the
16
0
display on the crt. When the SEC/DIV switch is set to X-Y, the Channel 2 POSITION control moves the display vertically (Y-axis), and the Horizontal POSITION control moves the display horizontally (X-axis).
only the Channel 2 input signal for
This
mode
REV OCT 1981
5
2213 Operators
HORIZONTAL
Refer to Figure 5 for location of items 17 through 22.
DELAY
17
0
with select the amount of delay time between the start of the sweep and the beginning of the intensified zone.
SEC/DIV
10
0
for the sweep generator in a l-2-5 sequence.
calibrated sweep speeds, the control must
wise).
TIME-Two controls are used in conjunction
INTENS
Range Selector
selects 0.5 increase the sweep delay from the calibrated setting of the Range Selector switch, rotate the MULTIPLIER control clockwise.
MULTIPLIER Control-Provides variable sweep delay from less than 1 to greater than 20 times the setting of the Range Selector switch.
and DLY’D HORIZONTAL MODE to
Switch-This three-position switch
rc(s,
10
,us,
and 0.2 ms of delay time. To
Switch-Used to select the sweep speed
SEC/DIV
be in the calibrated detent (fully clock-
For
Variable
SEC/DIV
19
0
variable, uncalibrated sweep speeds to at least 2.5 times the calibrated setting. It extends the slowest sweep speed to at least 1.25 s per division.
X10
20
0
speed by a factor of 10, pull out the Variable knob. The fastest sweep speed can be extended to 5 ns per division. Push in the Variable control knob to regain the Xl sweep speed.
HORIZONTAL MODE Switch-This three-position
21
0
switch determines
horizontal deflection system.
Variable
Magnifier Switch-To
NO DLY-Horizontal deflection is provided by the sweep generator, sweep speed determined by the
INTENS-Horizontal deflection is provided by the sweep generator at a sweep speed determined by the
SEC/DIV provides an intensified zone on the display. The start of the intensified zone represents the sweep-
start point when DLY’D HORIZONTAL MODE
is selected.
Control-Provides continuously
increase displayed sweep
the mode of operation for the
without a delayed start, at a
SEC/DIV
switch. The sweep generator also
SEC/DIV
SEC/DIV
switch.
SEC/W
Figure 5. Horizontal controls.
3397-06
DLY’D-Horizontal deflection is provided by the
sweep generator at a sweep speed determined by the
SEC/DIV sweep is delayed from the initial sweep-trigger point by a time determined by the setting of the
DELAY TIME Range Selector switch and MULTI-
PLIER control.
POSITION Control-Positions the display hori-
22
0
zontally
in all modes.
switch setting. The start of the
TRIGGER
Refer to Figure 6 for locations of items 23 through 30.
23
EXT INPUT
0
ducing external signals into the trigger generator.
24
EXT COUPLING
0
used
AC-Signals above 60 Hz are capacitively coupled
to the input of the Trigger circuit.
ponents are blocked, and signals below 60 Hz are
attenuated.
Connector-Provides a means of intro-
Switch-Determines the method
to
couple external signals to the Trigger circuit.
Any dc com-
__
2213 Operators
DC-All
components of the signal are coupled to
the trigger circuitry. This position
displaying low-frequency or low-repetition-rate signals.
DC+lO-External
a factor of 10.
25
SOURCE Switch-Determines the source of the
0
trigger signal that is coupled to the input of the trigger circuit.
INT-Permits triggering on signals that are applied to the CH 1 OR X and CH 2 OR Y input con­nectors. selected by the INT switch.
LINE-Provides a triggering signal from a sample of the ac-power-source waveform. This trigger source is useful when channel-input signals are time related (multiple or submultiple) to the frequency on the power-source-input voltage.
EXT-Permits triggering on signals applied to the EXT INPUT connector.
trigger signals are attenuated by
The source of the internal signal is
is
useful
for
INT Switch-Selects the source of the triggering
26
0
signal when the
CH I-The signal applied to the CH 1 OR X input connector is the
VERT MODE-The internal trigger source is determined by the signals selected for display by the VERTICAL MODE switches.
CH
connector is the source of the trigger signal.
LEVEL Control-Selects the amplitude point on the
27
0
trigger signal at
TRIG’D Indicator-The light-emitting diode (LED)
28
0
illuminates to indicate that the sweep is triggered.
SLOPE Switch-Selects the slope of the signal that
29
triggers the sweep (also refer to TV Signal Displays at
0
the end of “Instrument Familiarization”).
I-Sweep is
portion of the trigger signal.
SOURCE switch is set to INT.
source of the trigger signal.
2-The signal applied to the CH 2 OR Y input
which
the sweep is triggered.
triggered on the positive-going
Figure 6. Trigger controls, connector, and indicator.
3397-07
I-Sweep
portion of the trigger signal.
MODE Switch-Determines the trigger mode for the
30
0
sweep.
AUTO-Permits triggering on waveforms having repetition rates of at least 20 Hz. Sweep free-runs
in the absence of an adquate trigger signal or when the repetition rate is below 20 Hz. The range of the TRIGGER LEVEL control will compensate for the amplitude variations of the trigger signals.
NORM-Sweep is initiated when an adequate trigger signal is applied.
signal, no baseline trace will be present. Triggering on television lines is accomplished in this mode.
TV FIELD-Permits triggering on television field­rate signals (refer to of “Instrument Familiarization”.
VAR HOLDOFF
31
0
trol of the
holdoff control improves the ability to trigger on aperiodic signals (such as complex digital waveforms).
is triggered on the negative-going
In the absence of a trigger
TV
Signal Displays at the end
Control-Provides continuous con-
holdoff
time between sweeps. Increases
time by at least a factor of four. This
REV NOV 1981
2213 Operators
REAR PANEL
Refer to Figure 7 for location of item 32.
32
EXT
Z
0
necting
AXIS Connector-Provides a means of
external signals to the Z-axis amplifier to
con-
REPLACE ONLY WITH SPECIFIED
TYPE
AND RATED FUSE DISCONNECT
POWER INPUT BEFORE REPLACING FUSE
intensity modulate the crt display. Applied signals do not affect display waveshape. Signals with fast rise times and fall times provide the most abrupt intensity change, and a 5-V p-p signal will produce noticeable modulation. The Z-axis signals must be time-related to the display to obtain a stable ation
on the crt.
GROUNDING CONDUCTOR
lOK0
POSITIVE GOING
present-
Figure 7. Rear-panel connector.
32
0
3397-08
OPERATING CONSIDERATIONS
2213 Operators
This section contains basic operating information and
techniques that should be considered before attempting
any measurements.
GRATICULE
The graticule is internally marked on the faceplate of the crt to enable accurate measurements without parallax error (see Figure 8). It is marked with eight vertical and ten horizontal major divisions. Each major division is divided
into five subdivisions. The vertical deflection factors and
horizontal timing are calibrated to the graticule so that accurate measurements can be made directly from the crt. Also, percentage markers for the measurement of rise and fall times are located on the left side of the graticule.
1ST OR LEFT
VERTICAL
GRATICULE
LINE
CENTER
VERTICAL
GRATICULE
LINE
11TH
OR RIGHT
VERTICAL
GRATICULE
LINE
SIGNAL CONNECTIONS
Generally, probes offer the most convenient means of connecting an input signal to the instrument. They are shielded to prevent pickup of electromagnetic interference, and the supplied 10X probe offers a high input impedance that minimizes circuit loading. This allows the circuit under test to operate with a minimum of change from its normal condition as measurements are being made.
Coaxial cables may also be used to connect signals to the
input connectors, but they may have considerable effect on
the accuracy of a displayed waveform. To maintain the
original frequency characteristics of an applied signal,
only high-quality, low-loss coaxial cables should be used.
Coaxial cables should be terminated at both ends in their
characteristic impedance. If this is not possible, use suitable
impedance-matching devices.
INPUT COUPLING CAPACITOR
PRECHARGING
RISE AND CENTER
FALL TIME HORIZONTAL
MEASUREMENT GRATICULE
PERCENTAGE LINE
MARKERS
Figure 8. Graticule measurement markings.
GROUNDING
The most reliable signal measurements are made when
the 2213 and the unit under test are connected by a com-
mon reference (ground lead), in addition to the signal lead or probe. The probe’s ground lead provides the best grounding method for signal interconnection and ensures the maximum amount of signal-lead shielding in the probe cable. A separate ground lead can also be connected from the unit under test to the oscilloscope GND connector
located on the front panel.
4115-16
When the input coupling switch is set to GND, the input
signal is connected to ground through the input coupling
capacitor in series with a
l-MS2
resistor to form a
pre­charging network. This network allows the input coupling capacitor to charge to the average dc-voltage level of the signal applied to the probe. Thus, any large voltage transients that may accidentally be generated will not be applied to the amplifier input when the input coupling switch is moved from GND to AC. The
precharging
net­work also provides a measure of protection to the external circuitry by reducing the current levels that can be drawn from the external circuitry during capacitor charging.
The following procedure should be used whenever the probe tip is connected to a signal source having a different dc level than that previously applied, especially if the level difference is more than 10 times the
VOLTS/DIV
dc-
switch setting:
1. Set the AC-GND-DC switch to GND before con-
necting the probe tip to a signal source.
2. Insert the probe tip into the oscilloscope GND
connector.
@
9
2213 Operators
3. Wait several seconds for the input coupling capacitor
to discharge.
4. Connect the probe tip to the signal source.
5. Wait several seconds for the input coupling capacitor
to charge.
6. Set the AC-GND-DC switch to AC. The display will
remain on the screen, and the ac component of the signal
can be measured in the normal manner.
INSTRUMENT COOLING
To maintain adequate instrument cooling, the ventila­tion holes on both sides and rear panel of the equipment cabinet must remain free of obstructions.
INSTRUMENT FAMILIARIZATION
INTRODUCTION
The procedures in this section are designed to assist you in quickly becoming familiar with the 2213. They provide information which demonstrates the use of all the controls,
connectors, and indicators and will enable you to effi­ciently operate the instrument.
Before proceeding with these instructions, verify that
the POWER switch is OFF (push button out), then plug the power cord into the ac-power-input-source outlet.
If during the performance of these procedures an im­proper indication or instrument malfunction is noted, first verify correct operation of associated equipment. Should the malfunction persist, refer the instrument to qualified service personnel for repair or adjustment.
The equipment listed in Table 1, or equivalent equip­ment, procedures.
is
required to complete these familiarization
Equipment Required for Instrument
Familiarization Procedure
Description
Calibration Generator
Dual-Input Coupler
Cable (2 required)
Adapter
Termination
Standard-ampl Signal amplitude: 2 mV to 50 V. Output signal: 1 -kHz square wave.
Fast-rise repetition rate: 1 to
Rise time: 1 ns or less. Signal amplitude: 100 Aberrations:
Connectors: bnc-female-to-dual-bnc-male.
Impedance: 50 Connectors: bnc.
Connectors: bnc-female-to-bnc female.
Impedance: 50
Table 1
Minimum Specification
itude accuracy:
100
mV
to 1 V.
+2%.
s2.
Length: 42 in.
a.
Connectors: bnc.
50.25%.
kHz.
10
2213 Operators
BASELINE TRACE
First obtain a baseline trace, using the following
procedure.
1. Preset the instrument front-panel controls as follows:
Display
AUTO INTENSITY
AUTO FOCUS
Vertical (Both Channels)
AC-GND-DC VOLTS/DlV VOLTS/D IV Variable
VERTICAL MODE
INVERT
POSITION
Horizontal
SEC/DIV SEC/DlV
HORIZONTAL MODE X 10 Magnifier
Variable
POSITION DELAY TIME
Range Selector
MULTIPLIER
Trigger
SLOPE LEVEL MODE EXT COUPLING SOURCE VAR
HOLDOFF
INT
Fully counterclockwise (minimum) Midrange
AC
50m (IX) Calibrated detent (fully clockwise)
CH 1 Off (button out) Midrange
0.5
ms Calibrated detent (fully clockwise) NO DLY
Off (variable knob in)
Midrange
0.2 ms Fully counterclockwise
J(lever
Midrange AUTO AC
INT Fully counterclockwise
VERT MODE
up)
NOTE
Normally, the resulting trace will be parallel with the center horizontal graticule line and should not require adjustment. If trace alignment is required, see the under “Opera tor’s Adjustments.
“Trace Rotation
adjustment procedure
DISPLAYING A SIGNAL
ne
1
obtaining a baseli
After
connectan input signal and
1. Connect the calibration generator standard-amplitude
output to both the CH 1 and CH 2 inputs as shown in
Figure 9.
2.
Set the calibration generator for a standard-amplitude
1-kHz
square-wave signal and adjust its output to obtain
a vertical display of 4 divisions.
3. Adjust the Channel 1 POSITION control to center
the display vertically on the screen.
4. Adjust the TRIGGER LEVEL control, if necessary,
to obtain a stable triggered display.
The
TRIG’D
that the sweep is triggered.
5. Rotate the AUTO FOCUS control between its maximum The display should become blurred on either side of the optimum control setting.
indicator should illuminate to indicate
clockwise and counterclockwise positions.
trace,you are now ready to
di
splay it
NOTE
on the crt screen.
2. Press in the POWER switch button (ON) and allow
the instrument to warm up for 20 minutes.
3. Adjust the AUTO INTENSITY control for desired
display brightness.
4. Adjust the Vertical
controls to center the trace on the screen.
and Horizontal POSITION
@
6. Set the AUTO FOCUS control for a sharp,
defined display over the entire trace length.
7. Move the display off the screen using the Channel 1
POSITION control.
8. Press in and hold the BEAM FIND push button;
the display should reappear on the screen. Adjust the
Channel 1 and Horizontal POSITION controls to center
the trace both vertically and horizontally. Release the
well-
11
2213 Operators
BEAM FIND
button; the display should remain within
5.
Set the Channel 1 AC-GND-DC switch to AC.
the viewing area.
6. Observe that the display is centered approximately
at the center horizontal line.9. Adjust the AUTO INTENSITY control counterclock-
wise until the display disappears.
7.
Set the CH 1
10. Press in and hold the BEAM FIND push button;
observe that a 2-division vertical display appears. the display should reappear. Release the BEAM FIND button and adjust the AUTO INTENSITY control to desired display brightness. 8. Rotate the CH
fully counterclockwise.
Using the Vertical Section
1.
Set the Channel 1 AC-GND-DC switch to GND.
2. Adjust the trace to the center horizontal graticule
line.
3. Set the Channel 1 AC-GND-DC switch to DC.
4. Observe that the bottom of the display remains at
the center horizontal graticule line (ground reference).
9. Observe that minimum vertical deflection occurs
when the
VOLTS/DlV
clockwise.
IO. Rotate the CH 1 VOLTS/DIV
clockwise
11.
to the CAL detent.
Select
CH 2 VERTICAL MODE and again perform preceding steps 1 through 10 using Channel Performance should be similar to Channel 1.
VOLTS/DlV
1
switch to 0.1 (IX) and
VOLTS/DIV
Variable control
Variable contol is fully counter-
Variable control
2
controls.
fully
12
CALIBRATION
GENERATOR
DUAL-INP
COUPLE
50
!2
CABLE
/
Figure 9. Initial setup for instrument familiarization procedure.
AMPL
OUTPUT
3397-10
Loading...
+ 21 hidden pages