Tektronix 2200-20-5, 2200-30-5, 2200-32-3, 2200-60-2, 2200-72-1 Specifications

Series 2200 Programmable DC Power Supplies
2200-20-5
2200-30-5
2200-32-3
2200-60-2
2200-72-1
DC output rating
Voltage
0 to 20 V
0 to 30 V
0 to 32 V
0 to 60 V
0 to 72 V
Current
0 to 5 A
0 to 5 A
0 to 3 A
0 to 2.5 A
0 to 1.2 A
Maximum power
100 W
150 W
96 W
150 W
86 W
Load regulation
Voltage
< 0.01% + 2 mV
< 0.01% + 2 mV
< 0.01% + 2 mV
< 0.01% + 2 mV
< 0.01% + 2 mV
Current
< 0.05% + 0.1 mA
< 0.05% + 0.1.5 mA
< 0.05% + 0.1 mA
< 0.05% + 0.5 mA
< 0.05% + 0.5 mA
Line regulation
Voltage
< 0.01% + 1 mV
< 0.0% + 1 mV
< 0.01% + 1 mV
< 0.01% + 2 mV
< 0.01% + 1 mV
Current
< 0.05% + 1 mA
< 0.05 % + 1 mA
< 0.05 % + 1 mA
< 0.05 % + 1 mA
< 0.05% + 1 mA
Ripple and noise (20 Hz to 7 MHz)
< 3 mVPP
< 4 mVPP
< 4 mVPP
< 5 mVPP
< 4 mVPP
Current
< 3 mA
RMS
< 4 mA
RMS
< 3 mA
RMS
< 3 mA
RMS
< 3 mA
RMS
Setting resolution
Voltage
1 mV
1 mV
1 mV
1 mV
1 mV
Current
0.1 mA
0.1 mA
0.1 mA
0.1 mA
0.1 mA
Setting accuracy (using remote sense, 25°C ±5°C)
Voltage
± 0.03% + 3 mV
± 0.03% + 3 mV
± 0.03% + 3 mV
± 0.03 % + 6 mV
± 0.03 % + 6 mV
Current
± 0.05% + 2 mA
± 0.05% + 2.5 mA
± 0.05% + 2 mA
± 0.05 % + 1.5 mA
± 0.05 % + 1 mA
Keithley Instruments 28775 Aurora Road Cleveland, Ohio 44139
Specifications
1-800-935-5595
tek.com/keithley

Specification conditions

This document contains specifications for the Models 2200-20-5, 2200-30-5, 2200-32-3, 2200-60-2, and 2200-72-1 Programmable dc Power Supplies. Specifications are the standards against which the Models 2200-20-5, 2200-30-5, 2200-32-3, 2200-60-2, and 2200-72-1 are tested. Upon leaving the factory, the Models 2200-20-5, 2200-30-5, 2200-32-3, 2200-60-2, and 2200-72-1 meet these specifications. Typical values are not warranted and are provided solely as useful information.

Model-specific specifications

Voltage
< 1 mV
RMS
SPEC-2200 Rev. B May 2020
< 1 mV
RMS
< 1 mV
RMS
*PSPEC-2200B*
< 1 mV
RMS
< 1 mV
RMS
1
Series 2200 Programmable DC Power Supplies Specifications
2200-20-5
2200-30-5
2200-32-3
2200-60-2
2200-72-1
Readback resolution
Voltage
1 mV
1 mV
1 mV
1 mV
1 mV
Current
0.1 mA
0.1 mA
0.1 mA
0.1 mA
0.1 mA
Readback accuracy (25 °C ± 5 °C)
Voltage
0.02% + 3 mV
0.02% + 3 mV
0.02% + 3 mV
0.02% + 6 mV
0.02% + 5 mV
Current
0.05% + 2 mA
0.05% + 2.5 mA
0.05% + 2 mA
0.05% + 1.5 mA
0.05% + 1 mA
Voltage transient response – settling time
Load change
<400 μs to within 75 mV following a change from 0.1 A to 1 A
Setting change
Rising
<Setting voltage from 0% to 100%; voltage change from 10% to 90% < 35 ms Falling
< Setting voltage from 100% to 0%; voltage change from 90% to 10% < 200 ms
NOTE:
Overvoltage protection
Range (typical)
1 V to 19 V
1 V to 29 V
1 V to 31 V
1 V to 59 V
1 V to 71 V
Accuracy
± 0.5% + 0.5 V
± 0.5% + 0.5 V
± 0.5% + 0.5 V
± 0.5% + 0.5 V
± 0.5% + 0.5 V
Response time
<10 ms
<10 ms
<10 ms
<10 ms
<10 ms
Communications
USB: Type B connector, USB-TMC compatible GPIB:
Display
Vacuum fluorescent display
Memory
40 setup memories
List mode
Up to seven lists can be defined, each with up to 80 steps. Each step includes a voltage limit and
Output, sense, status,
Removable screw terminal block carries the following signals:
Floating voltage rating
Up to 100 V (dc + peak ac) between protective earth (safety ground) and any output terminal
Model-specific specifications
NOTE: Specification does not include command decode time.
Specification does not include command decode time.
(typical)

General specifications

a current limit. For continuous sequences, each step also includes a duration.
and control:
IEEE-488.2 compliant
Output channel: Duplicates the front panel outputs
Remote sense lines: Connection for remote sense
Control input: Multifunction TTL input, which can function as a trigger input, output control
line, or digital input
Status Output: Multifunction TTL output, which can function as a fault indication or digital output
2