Tektronix 191 schematic

TB 9-6625-2176-35
DEPARTMENT OF THE ARMY TECHNICAL BULLETIN
CALIBRATION PROCEDURE FOR
CONSTANT AMPLITUDE SIGNAL GENERATOR
AN/USM-272
AND
TEKTRONIX, TYPE 191
16 May 1988
TB 9-6625-2176-35, 15 April 1987, is changed as follows:
1. Remove old pages and insert new pages as indicated below. New or changed material is indicated by a vertical
bar in the margin of the page.
Change 2
Remove pages Insert pages
1 and 2 1 and 2 7 and 8 7 and 8
2. File this change sheet in front of the publication for reference purposes.
By Order of the Secretary of the Army:
CARL E. VUONO
General, United States Army
Chief of Staff
Official:
R. L. DILWORTH
Brigadier General, United States Army
The Adjutant General
Distribution: To be distributed in accordance with DA Form 12-34C, Block No. 319, requirements for calibration procedures publications.
PIN NO: 061693-002
DEPARTMENT OF THE ARMY TECHNICAL BULLETIN
CALIBRATION PROCEDURE FOR
CONSTANT AMPLITUDE SIGNAL GENERATOR
AN/USM-272
TEKTRONIX, TYPE 191
Headquarters, Department of the Army, Washington, DC
31 March 1988
TB 9-6625-2176-35, 15 April 1987, is changed as follows:
TB 9-6625-2176-35
Change 1
AND
1. Remove old pages and insert new pages as indicated below. New or changed material is indicated by a vertical
bar in the margin of the page.
Remove pages Insert pages
7 and 8 7 and 8
2. File this change sheet in front of the publication for reference purposes.
By Order of the Secretary of the Army:
CARL E. VUONO
General, United States Army
Chief of Staff
Official:
R. L. DILWORTH
Brigadier General, United State Army
The Adjutant General
Distribution: To be distributed in accordance with DA Form 12-34C, Block No. 319, requirements for calibration procedures publications.
THIS PUBLICATION IS A COURTESY QUICK COPY FRCM THE UNITED STATES ARMY PUBLICATICNS DISTRIBUTION CENTER, ST. LOUIS, MISSOURI TO MEET YOUR NEEDS WHILE WE REPLENISH OUR RULAR STOCK.
*TB 9-6625-2176-35
SUPERSEDED COPY DATED 2 AUGUST 1978
DEPARTMENT OF THE ARMY TECHNICAL BULLETIN
CALIBRATION PROCEDURE FOR
CONSTANT AMPLITUDE SIGNAL GENERATOR
AN/USM-272 AND TEKTRONIX, TYPE 191
Headquarters, Department of the Army, Washington, DC
15 April 1987
Approved for public release; distribution is unlimited.
REPORTING OF ERRORS
You can help improve this publication by calling attention to errors and by recommending improvements and stating your reasons for the recommendations. Your letter or DA Form 2028, Recommended Changes to Publications, should be mailed directly to Commander, U.S. Army TMDE Support Group, ATTN: AMXTM-LPP, Redstone Arsenal, AL 35898-5400. A reply will be furnished directly to you.
SECTION I. IDENTIFICATION AND DESCRIPTION
Test instrument identification ................................................... 1 2
Forms, records, and reports-.................................................... 2 2
Calibration description.............................................................. 3 2
II. EQUIPMENT REQUIREMENTS
Equipment required.................................................................. 4 3
Accessories required................................................................ 5 3
III. CALIBRATION PROCESS
Preliminary instructions............................................................ 6 4
Equipment setup ..................................................................... 7 4
Amplitude accuracy.................................................................. 8 5
Frequency accuracy................................................................. 9 7
Frequency response................................................................. 10 7
Power supply ........................................................................... 11 9
Final procedure........................................................................ 12 9
Paragraph Page
*This bulletin supersedes TB 11-6625-2384-35, 2 August 1978, including all changes.
1

SECTION I

IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification. This bulletin
provides instructions for the calibration of Constant Amplitude Signal Generator, AN/USM-272 and Tektronix, Type 191. The manufacturer's manual and TM 116625-2384-15 were used as the prime data sources in compiling these instructions. The equipment being calibrated will be referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations. None. b. Time and Technique. The time required for
this calibration is approximately 4 hours, using the dc and low frequency technique.
2. Forms, Records, and Reports Table 1. Calibration Description
Test instrument
parameters
Amplitude (at 50 kHz into Range: 0.5 to 5 V 50W load) Accuracy: ±3% of indicated amplitude
Frequency (into 50W load) Range: 50 kHz and 0.35 to 100 MHz Frequency response Range: 0.35 to 100 MHz at .5 to 5.5 V
TB 9-6625-2176-35
a. Forms, records, and reports required for
calibration personnel at all levels are prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which they appear. When adjustments are in tables, the (R) follows the designated adjustment. Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance specifications which pertain to this calibration are listed in table 1.
Performance
specifications
Range: 50 to 500 mV Accuracy: ±4% of indicated amplitude Range: 5 to 50 mV Accuracy: ±5% of indicated amplitude
Accuracy: ±2%
Accuracy: £ ±3% from 50 kHz value
Change 2 2

SECTION II

EQUIPMENT REQUIREMENTS
TB 9-6625-2176-35
4. Equipment Required. Table 2 identifies the specific
equipment to be used in this calibration procedure. This equipment is issued with Secondary Transfer Calibration Standards Set AN/GSM-286. Alternate items may be used by the calibrating activity when the equipment listed in table 2 is not available. The items selected must be verified to perform satisfactorily prior to use and must bear evidence of current calibration. The equipment must meet or exceed the minimum use specifications listed in table 2. The accuracies listed in
Table 2. Minimum Specifications of Equipment Required
Item Common name Minimum use
A1 AUTOTRANSFORMER Range: 105 to 125 V ac
Accuracy: ±1%
A2 DC VOLTAGE STANDARD Range: 0 to 5 V dc
Accuracy: ±0.01%
A3 DIGITAL VOLTMETER Range: 0 to -12 V dc
0.0167 to 1.820 V rms Accuracy: ±0.75%
A4 DIFFERENTIAL VOLTMETER Range: 0.9 to 1.1 mV
Accuracy: ±0.75%
A5 FREQUENCY COUNTER Range: 50 kHz to 100 MHz
Accuracy: 0.5%
A6 THERMAL CONVERTER Range: 50 kHz to 100 MHz
at 5 V Accuracy: ±0.75%
table 2 provide a four-to-one ratio between the standard and TI.
5. Accessories Required. The accessories listed in table 3 are issued as indicated in paragraph 4 above and are used in this calibration procedure. When necessary, these items may be substituted by equivalent items, unless specifically prohibited.
specifications
Manufacturer and model
(part number)
General Radio, Model W10MT3AS3 or Ridge, Model 9020F (7910809) John Fluke, Model 332B/AF (332B/AF) Hewlett-Packard, Model 3490AOPT060 (3490AOPT060) Dana, Model 5000, or Dana, Model 5000 w/641 John Fluke, Model 887AB/AN (887AB/AN) Hewlett-Packard, Model 5345A (MIS-28754/1 Type 1) w/5355A and K87-59992A Ballantine, Model 1394A (MIS-10221)
3
Loading...
+ 9 hidden pages