Display Board, Parts List. .............................
Model 1758 Battery Pack, Parts List ....................
Model 1758 Spare Pats List. ..........................
............................
...............................
...............
...............
.............
Title
.......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
Page
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
......
iv
SAFETY PRECAUTIONS
The following safety precautions should be observed before operating
the Model 175.
This instrument is intended for use by qualified personnel who
recognize shock hazards and are familiar with the safety precautions required to avoid possible injury. Read over the manual carefully before
operating this instrument.
Exercise extreme caution when a shock hazard is present at the instru-
ment’s input. The American National Standards Institute (ANSI) states
that a shock hazard exists when voltage levels greater than 30V rms or
42.4V peak are present. A good safety practice is to expect that a
hazardous voltage is present in any unknown circuit before measuring.
Inspect the test leads for possible wear, cracks or breaks before each
use. If any defects are found, replace with test leads that have the same
measure of safety as those supplied with the instrument.
For optimum safety do not touch the test leads or the instrument while
power is applied to the circuit under test. Turn the power off and
discharge all capacitors, before connecting or disconnecting the instru-
ment.
Do not touch any object which could provide a current path to the
common side of the circuit under test or power line (earth) ground.
Always make measurements with dry hands while standing on a dry,
insulated surface, capable of withstanding the voltage being measured.
Exercise extreme safety when testing high energy power circuits (AC
line or mains, etc.). Refer to the operation section.
Do not exceed the instrument’s maximum allowable input as defined in
This instruction manual contains the necessary information
for operating and maintaining the Model 175 Autoranging
Multimeter and the Model 1758 Rechargeable Battery Pack.
The information is divided into the following sections:
Section 1 contains general information and provides
guidelines for using this manual. Important safety information is also presented here.
Section 2 contains detailed bench operation information
for the Model 175.
Section 3 contains the information needed to verify the
accuracy of the Model 175. Performance verification can
be done upon receipt of the unit or whenever the basic
accuracy is in question.
For the more technically oriented, information on theory
of operation, and maintenance and servicing is contained
in Section 4 through 6.
NOTE
The Model 1753 IEEE-488 interface comes supplied with its own instruction manual.
1.2 GE-iTING STARTED
Perform the following steps in sequence to acquaint yourself
auicklv and safelv with the basic operation of the Model 175.
Verify that the Model 175 was not damaged in transit, es
explained in paragraph 1.3.
Carefully read the safety precautions and warnings found
preceding this section and the first two sections (General
Information and Bench Operation) of this manual.
Referring to paragraph 2.3.1 (Line Power) set the line
voltage switch and plug the power cord into a properly
grounded outlet. If the optional battery pack is installed
the charge circuitry will be activated.
Acquaint yourself with the controls and display of the
Model 175 as follows:
A. Turn on the Model 175 by pressing in the ON/OFF
pushbutton. All of the zeroee will be displayed briefly.
B. Connect the supplied test leads to the
VOLTS/OHMS/mA and COM input jacks, and short
them together.
C. Select AC volts and autoranging by pressing in the
AC/DC, V and AUTO pushbuttons. The AC, mV and
AUTO annunciators will be displayed. Pressing any of
the other range pushbuttons will put the Model 175 in
manual ranging es indicated by the absence of the
AUTO annunciator.
Select DC volts by releasing (outi the AC DC
D.
pushbutton (V still selected]. The AC annunciator will
turn off.
Select autoranging ohms by pressing in the !!
E.
pushbutton (DC still selected) and AUTO pushbutton.
The n annunciator will turn on. Press the AC:DC
pushbutton in (AC selected) and note the “Err”
message indicating that this is an invalid mode.
Select AC or DC current by setting the AC DC
F.
pushbutton accordingly and pressing I,? ihe A
pushbutton. The annunciator that reflects the selected
range will turn on. Note that current will not autorange
and that the 10 AMPS and COM input jacks must be
used on the 10A range.
Select
G.
H.
I.
5. When you are comfortable with the controls of the Model
175, go on and make the desired meaeurements using
Section 2, Bench Operation es a guide.
1.3 UNPACKING AND INSPECTION
The Model 175 Bench DMM was carefully inspected, both
mechanically and electrically, before shipment. Upon recelv
ing the Model 175, carefully unpack all items from the ship-
ping carton and check for any obvious signs of physical
damage that might have occurred during shipment. Report
the damage to the shipping agent immediately. Retain the
dB by placing the Model 175 in AC or DC volts
and pressing the dB pushbutton. The dB annunc~etor
will turn on. Press the dB button again fo take the
Model 175 out of the dB measurement mode.
REL (relative) can be used with any measurement furw
tion: volts, ohms, amps or dB. For example, place the
Model 175 in ohms and autorange. The display wll
read approximately 00.141!, which is the test lead
resistance. Press the REL pushbutton. The REL any
nunciator will turn on and the display will now reed
OO.OOI2. The relative level of 0.1411 will be subtracted
from all subsequent ohm measurements. Press the
REL pushbutton a second time to cancel the REL
level.
To activate the 100 point DATA LOGGER with
MIN/MAX, press and hold in the STOiCLR pushbut-
ton. When the reading rate A = 0 is displayed let go of
the button, The ST0 annunciator will turn on Press
the RCL pushbutton and the last data point will be
displayed briefly followed by the reading (data]. Other
data points can be displayed by holding in the RCL button. Turn off the DATA LOGGER by pressing the
STOlCLR pushbutton again.
1-l
original packing materials in case reshipment becomes
necessan/. The following items are included with every Model
‘Model 1751 Safety Test Lead
*Additional accessories as ordered.
1.4 SPECIFICATIONS
Detailed Model 175 specifications may be found immediately
preceding the table of contents of this manual.
1.5 WARRANTY INFORMATION
Warranty information may be found on the inside back cover
of this manual. Should it be necessary to exercise thewarranty, contact your Keithley representative or the factory to
determine the correct course of action. Keithley Instruments
maintains service facilities in the United States, West
Germany, Great Britain, France, the Netherlands.
Switzerland and Austria. Information concerning the application, operation or service of your instrument may be directed
to the applications engineer at any of theses locations. Check
the inside front cover of this manual for addresses.
1.6 MANUAL ADDENDA
Information concerning improvements or changes to the in-
strument which occur after the printing of this manual will be
found on an addendum sheet included with this manual. Be
sure to review these changes before attempting to operate or
service the instrument.
1.7 SAFETY SYMBOLS AND TERMS
The following safety symbols and terms are used in this
manual or found on the Model 175.
The symbol ! on the instrument indicates that the user
should refer to the operating instructions in this manual.
The symbol ,$
of 1OOOV or more may be present on the terminal(s). Standard safety practices should be observed when such
dangerous voltages are encountered.
The WARNING heading in this manual explains dangers that
could result in personal injury or death.
The CAUTION heading in this manual explains hazards that
could damage the instrument.
1.8 OPTIONAL ACCESSORIES
The following accessories can be used with the Model 175.
Model 1010 Single Rack Mounting Kit-Use to mount
one Model 175 in a standard 5 x X 19” rack.
on the w,trument indicates that a potential
Model 1017 Dual Rack Mounting Kit-Use to mount two
Model 175’s in a standard 5 x X 19” rack.
Model 1301 Temperature Probe-A rugged low cost
temperature probe designed to allow precision temperature
measurements from -55OC to 150°C.
Range: -55’C to 150°C
Output: lmV/OC; compatible with any DMM with at least
IOM12 input impedance.
Accuracy: f2OC from O” to 100°C; *YC from -5Y to
0°C and 100’ to 150°C
Power: 9V alkaline or C-Zn (NEDA 1604) battery.
Model 1600A High Voltage Probe-Extends the DMM to
40kV.
Maximum Input: 40kV DC or peak AC to 300Hr
Input Resistance:lOOOM12
Division Ratio: 1OOO:l
Ratio Accuracy: f2.5% from 1kV to 40kV DC, *3.5% if
200mV or 2V ranges of Model 175 are used; -3dB at 300Hz
AC
Operating Temperature: 0’ to 50°C
Model 1651 50.Ampere Current Shunt-The external
O.OOlI2 f I%,
measurements from O-50A DC or AC.
Model 1681 Clip-On Test Lead Set-contains two leads,
1.2m (48 inches) long terminated with banana plugs and
spring action clip on probes.
Model 16ElZA RF Probe-Permits voltage measurements
from 1OOkHz to 250MHr.
AC to DC transfer accuracy: *ldB from 1OOkHz to
250MHr at IV, peak responding, calibrated in rms of a sine
wave.
Maximum Allowable Input: 42V AC peak, 200V IDC + AC
peak)
Model 1684 Hard Shell Carrying Case-Hard vinyl case,
1OOmm x 300mm x 350mm 14 x I3 x I4 inches1 has a fit-
ted foam insert with room for the Model 175, instruction
manual and small accessories.
Model 1685 Clamp-On AC Probe-Measures AC current
by clamping onto a single conductor. Interruption of the cir-
cuit under test is unnecessary. The Model 1685 detects cur-
rent by sensing the changing magnetic field produced by the
current flow.
Range: 2, 20 and 200A rms
Accuracy: 54% of range at 60Hz; + 6% of range at 50Hz
Temperature Coefficient: +0.05%/°C on 20A and 200A
range; k0.3%/°C on 2A range
Maximum Allowable Current: 300A rms
Maximum Conductor Voltage: 6OOV rms
Conversion Ratio: O.lV/A rms
Model 1751 Safety Test Leads-This is the test lead set
supplied with each Model 175. Finger guards and shrouded
banana plugs help minimize the chance of making contact
with live circuitry.
Model 1753 IEEE-488 Interface-Field installable programmable option provides isolated data output. Switch-
selectable talk onlyoraddressable modes. Mounts within and
powered by the Model 175.
a-terminal shunt permits current
l-2
Model 1754 Universal Test Lead Kit-12 piece test lead kit,
with interchangeable plug-in accessories. Kit includes: one set
of test leads (i-red. l-black), two spade lugs, two standard banana plugs, two phone tips (.060DIA.), two hooks and two miniature alligator clips (with boots).
Model 1758 Rechargeable Battery Pack-Provides six hours
minimum operation from full charge, recharges within 10 hours
and is field installable.
Model 7008 IEEE-488 Digltal Cabl+Useful for connecting
the Model 1753 to the IEEE-488 bus. The Model 7008-3 is 0.9m
(3ft) in length and has a standard IEEE-488 connector at each
end. The Model 7008-6 is 1.8m (6ft) in length.
l-311-4
SECTION 2
BENCH OPERATION
2.1 INTRODUCTION
This section contains the information needed to prepare and
operate the Model 175 as a bench DMM. Bench operation
consists of using the Modal 175 to perform basic voltage,
currant, resistance and dB measurements. Also, the operation of the data logger is covered hare. The capabilities of the
Model 175 can be enhanced with the addition of the Model
1753 IEEE-488 interface. IEEE operation is covered in the
Modal 1753 Instruction Manual.
2.2 HIGH ENERGY CIRCUIT PRECAUTIONS
To optimize safety when measuring voltage in high energy
distribution circuits, read and use the directions in the follow-
ing warning.
WARNING
Dangerous arcs of an explosive nature in a high
energy circuit can cause sevare personal injury
or death. If the mater is connected to a high
energy circuit when set to a currant range, low
resistance range or any other low impedance
range, the circuit is virtually shorted. Dangerous
arcing can result awn when the mater is sat to a
voltage range if the minimum safety spacing is
reduced.
When making measurements in high energy circuits use test
leads that meat the following requirements:
1. Test leads should be fully insulated.
2. Only use test leads that can be connected to the circuit
leg. alligator or spade plugs) for hands-off measurement.
3. Do not use test leads that decrease voltage spacing. This
diminishes arc protection and creates a hazardous condition.
Use the following sequence when testing power circuits:
1. De-energize the circuit using the regular installed connactdisconnect device such as the circuit breaker, main switch,
etc.
2. Attach the test leads to the circuit under test. Use appropriate safety rated leads for this application.
3. Sat the DMM to the proper function and range.
4. Energize the circuit using the installed connect-disconnect
device and make measurements without disconnecting the
DMM.
5. De-energize the circuit using the installed connectdisconnect device.
6. Disconnect the test leads from the circuit under test.
2.3 PREPARATION FOR USE
2.3.1 Line Power
The Model 175 is provided with a three-wire line cord which
mates with third-wire grounded receptacles. Connect the ins
strumant to AC line power as follows:
1. Sat the LINE VOLTAGE switch on the back of the instrument to correspond to line voltage available. Ranges are
105.12511 or 210.250V 50/60Hz AC.
CAUTION
Connect only to the line voltage selected.
Application of incorrect voltage can
damage the instrument.
2. Plug the power cord into a properly grounded outlet.
WARNING
Ground the instrument through a properly
grounded receptacle before operation.
Failure to ground the instrument can result
in savere injury or death in the avent of
short circuit or malfunction.
NOTE
Although the Model 175 is specified at 50 and
60Hz the instrument may be operated at 400Hz
and 440Hz. Add ona. count to instrument
specifications under this condition.
2.3.2 Battev Pack Power
The Modal 175 may also be operated from rechargeable
sealed nickel-cadmium batteries contained in the optional
Model 1758 Rechargeable Batwry Pack. The battery pack will
operate the Modal 175 for up to six hours The BAT annunciator will turn on when the battery charge is insufficient to
maintain accurate readings. Refer to Section 5, paragraph 5.3
for installation procedures.
2.3.3 Battery Charging
After the Model 1758 Battery Pack is installed in the Model
175 it can be charged and recharged as follows:
1. Connect the instrument to line power as described in
paragraph 2.3.1.
2. With the power switch off, the battery charge circuitry is
automatically energized to charge the battery at the maximum rata. When the battery pack is first installed, or if it is
completely discharged, allow it to charge for ten hours.
2-1
NOTE
For maximum battery efficiency only charge the
battery pack after it has become discharged and
only charge until it is fully charged (=,lOhours).
Continuous charging over long periods of time
will not damage the batteries but, usefu life will
gradually decrease. This loss is not permanent
and may be restored by cycling the battery pack
through several complete charge/discharge
cycles. The battery pack is capable of 500 to
1000 charge/discharge cycles before replacement is needed.
Do not make measurements with the BAT an-
nunciator on as the readings may be erroneous.
3. When the Model 175 is in use on line power, the battery
charger maintains a trickle charge on the battery pack.
2.4 FRONT PANEL FAMILIARIZATION
The following paragraphs and Figure 2-l provide a brief
description of the display, front panel controls and input ter-
minals.
2.4.1 Display
The Model 175 has a 4 % digit liquid crystal display (LCD).
The minus sign is displayed. The plus sign is implied by the
absence of the minus sign. The following annunciators are
displayed on the LCD.
BAT-Low battery indicator for the Model 1758.
AC-AC selected (DC implied by absence of AC annun-
ciator).
mV or V-Millivolts or volts selected.
n, kR or Mn-Ohms, kilohms or megohms selected.
p, mA or A-Microamps, milliamps or amps selected.
RMT (Remote)-Model 175 being controlled over the
IEEE-488 bus (Model 1753 installed).
C-Model 175 in calibration mode.
AUTO-Autorange selected.
REL--Relative selected.
dB-Decibel selected.
STO-Data being stored.
RCL-Data being recalled. RCL flashes when buffer is full
during logging cycle.
2.4.2 Front Panel Controls
ON/OFF-Pressing in this pushbutton turns the Model 175
on. Releasing (out) this pushbutton turns the instrument off.
REL (Relative)-This pushbutton allows readings to be
made with respect to any baseline value. Also allows zeroing
of on range readings. See paragraph 2.7.2 for more detailed
information on REL.
dB-This pushbutton selects the dB function and is used
along with the ACV or DCV function. Measurements are
made in dBm referenced to 600% REL can be used to make
any voltage level the OdB reference point for dB
measurements.
DATA LOGGER-Has min/max and 100 point reading
storage capacity; records data at one of six selectable rates
from every reading to 1 rdglhr. Records maximum and
minimum conversion during the period the data logger is active at the rate of 3lsec.
1. STO/CLR-Pressing this button initiates the logging sequence. Pressing the button a second time shuts off the
data logger.
2. RCL-Pressing and holding this button in scrolls the data
pointer. To read the data at a particular point, simply
release the button.
AC/DC-This switch is used along with the volts (V), current (A), and dB functions. Depressing (in) this pushbutton
selects AC and releasing (out) this pushbutton selects DC.
V-Depressing this pushbutton selects the volts function.
O-Depressing this pushbutton selects the ohms function.
The AC/DC pushbutton must be released (out).
A-Depressing this pushbutton selects the current function.
Range Push Buttons
1. AUTO-Depressing this pushbutton causes volts and
ohms to autorange. In current, it selects the 10A range (no
autoranging in current).
2-2
%iEiq
175 AUTORANGING MVLTIMETER
I
Figure 2-l. Model 175 Front Panel
2. Manual ranging is accomplished by depressing the ap
propriate range button.
2.4.3 Input Terminals
The input terminals are intended to be used with safety
shrouded test leads to help minimize the possibility of contact
with live circuits. Safety shrouded test leads are supplied with
the Model 175.
VOLTS/OHMS/mA and COM (Red and Black)-Use this
pair of terminals for all volt, ohm, milliamp and dB
measurements.
10 AMPS and COM (White and Black)-Use this pair exclusively for measuring currant up to 10A (up to 20A for 15
seconds,.
2.4.4 Current Fuse Replacement
The current fuse protects the 200pA through 2000mA ranges
from an input currant greater than 2A. To replace the current
fuse, perform the following steps:
1. Turn off the power and disconnect the power line and test
leads.
2. Place the end of a flat-blade screwdriver into the slot in the
fuse holder on the front panel. Press in slightly and rotate
the fuse carrier one-quarter turn counterclockwise.
Release pressure and remove the fuse carrier and the fuse.
Table 2-1. Error Messages
3. Remove the defective fuse and replace it with the following type: ?A, 250V. 3AG. normal-blow Keithley part
number FU-13, or equivalent.
CAUTION
Use only the recommended fuse type. If a fuse
with a higher current rating is installed, instrument damage may occur.
2.5 ERROR MESSAGES
Table 2-l lists the error messages associated with basic front
panel operation. Note that the instrument has a number of
other massages that are discussed in the appropriate sections
of this manual.
2.6 OPERATING CONDITIONS
2.6.1 Environmental Conditions
All measurements should be made at an ambient temperature
within the range of OOC to 5OOC. and with a relative humidity
of 0% to 80% up to 35OC. For instruments above 35OC
derate humidity 3% per OC up to 50°C. If the instrument has
been subjected to extremes of temperature, allow sufficient
time for internal temperatures to reach environmental conditions. Typically, it takes one hour to stabilize a unit that is
10°C (18OF, out of specified temperature range.
Comments
Model 175 locks up. See Section 5 for troubleshooting information.
Model 175 locks up, but operation can be restored by pressing
any one of the four momentary pushbuttons. If restored,
calibration is invalid as indicated by the flashing “C” annunciator.
See Section 5 for troubleshooting information.
Overrange input applied to the Model 175. Leading minus sign
indicates that input signal has a negative value.
“AC” and “R” annunciators flash. Correct problem by releasing
(out, AC/DC pushbutton.
Table 2-2. Model 175 Maximum Allowable Inputs
Ranges -
200mV. 2V
20-1ooov
200mV
2.750V
200w-2000mA
10A
All
Maximum Allowable Inputs
1OOOVDC or peak AC for less than 10s~ par minute.
300Vrms continuous.
1OOOVDC or peak AC.
750Vrms 1OOOV peak for lass than 1Osec per minute.
300Vrms continuous. lO’V*Hz maximum.
750Vrms. 1OOOV peak. 107V*Hz maximum.
2A, 250VDC or rms (fuse protected).
10A continuous; 20A for lbec (unfused,.
450VDC or peak AC
2-3
2.6.2 Maximum Allowable Inputs
2.7.2 Relative Mode
Table 2-2 lists the maximum allowable inputs for the Model
175.
2.7 BASIC BENCH MEASUREMENTS
Basic measurement techniques for using the Model 175 to
measure AC and DC volts, resistance, AC and DC current
and dB are covered in the following paragraphs. Also included is the operation of the MIN/MAX and 100 point data
logger.
WARNING
Before operating the Model 175, observe
the safety precautions found preceding
Section 1. When testing high energy power
circuits follow the procedure found in
paragraph 2.2 High Energy Circuit Precautions. Failure to observe these and other
safety precautions found in this manual
could result in severe injury or death.
The COM terminal on the Model 175 is designed to float
above earth ground to avoid ground loop problems.
WARNING
Hazardous voltages may be applied to the
COM terminal. The maximum allowable
voltage between the COM terminal and
chassis ground is 500V. Destruction of insulation, which could present a shock
hazard, may occur if the 500V maximum is
exceeded.
CAUTION
Do not exceed the maximum input limits
shown in Table 2-2.
2.7.1 Power-Up
NOTE
The software revision level of the Model 175 can
be displayed upon power-up by running the
diagnostic program. See Section 5,
Maintenance, for more information.
Turn on the Model 175 by pressing in the ON/OFF switch.
The following will occur:
1. Reset-All zeros will be briefly displayed before going into
the measurement mode.
2. RAM Test-If this test fails the Model 175 will lock up with
zeros displayed.
3. NVRAM Test-If this test fails the display will show the error
message “cErr”
Refer to Table 2-l for mc~re information pertaining to error
messages.
When the relative mode is selected with an on-scale reading
on the display the following occurs:
1. The REL annunciator is displayed.
2. The next reading is stored.
3. The stored reading is then algebraically subtracted from all
subsequent readings and displayed.
A REL level can be established for any measurement function
(Volts, Ohms, Amps and dB1 and is effective only on that
function. Changing functions will not affect a REL level
already established. However, if another REL level is set (on
any function) the previous REL level will be cancelled. For example, place the 175 in the R function and select the 2OOn
range. Short the test leads and press the REL button. Note
that the REL annunciator is on. Select DCV and note that the
REL annunciator is off, indicating that there is not a REL level
established for DCV. Switch back to 0 and note that the REL
level is still there. Again, go to DCV and set a REL level of
+ IV. The REL annunicator will go on. Switch back to n and
note that the REL annunciator is off, indicating that the REL
level for 0 is cancelled.
Once a REL level is established for a measurement function,
that stored level will be the same regardless of what range the
Model 175 is on. For example, if + IV is established as the
REL level on the 20VDC range, + IV will also be the REL level
on the 1OOOVDC range.
It is important to note that the use of REL reduces the
dynamic range of measurements by that level. For instance,
assume that the REL level is +lV and the Model 175 is
manually set to the 2V range. The maximum positive
displayed reading, before overranging, would be +0.9999\1.
This is because the A/D converter would be seeing 1.9999V
(maximum) from the input. Thus, the dynamic range of
measurement is -1.9999V to + 0.9999V (2.999BV) as compared to the normal -1.9999V to + 1.9999V (3.9998V). The
dynamic range of measurement has been reduced by IV. The
effects on dynamic range can be reduced by selecting a
higher range or using autorange.
2.7.3 DC Voltage Measurements
The Model 175 can make DC voltage measurements between
1OhV and IOOOV. The basic procedure is as follows:
I. Connect the test leads to the VOLTS and COM terminals
of the Model 175.
2. Select the DCV function.
3. Select a range consistent with the expected voltage. For
automatic range selection, press in the AUTO pushbutton.
NOTE
Manual ranging is recommended for routine
measurements above 200V.
2-4
4. Connect the test leads to the source as shown in Figure
2-2. If the positive source terminal is connected to the
COM terminal of the instrument, the display will show a
negative value. If the negative source terminal is connected to the COM terminal, the display will show a
positive value.
5. Observe the display; if the “OL” message is shown, select
a higher range until a normal reading is shown. Always use
the lowest possible range for the best resolution.
6. Take the reading from the display.
NOTE
High input impedance j VlOOOMRl DC volts
measurements can be made on the 200mV and
2V ranges by releasing (out) all the function
pushbunons (AC/DC, V, R, A).
DC
VOLTAGE
SOURCE
Figure 2-2. DC Voltage Measurements
2.7.4 TRMS AC Voltage Measurements
The Model 175 can make TRMS AC voltage measurements
between IOwV and 75OV. Proceed as follows:
1. Connect the test leads to the VOLTS and COM terminals
of the Model 175.
2. Select the ACV function.
3. Select a range consistent with the expected voltage. For
automatic range selection, press in the AUTO pushbutton.
NOTE
Manual ranging is recommended for routine
measurements above 200V.
4. Connect the test leads to the source as shown in Figure
2.3.
5. Observe the display; if the “OL” message is shown, select
a higher range until a normal reading is shown. Always use
the lowest possible range for the best resolution.
6. Take the reading from the display.
AC
VOLTAGE
Figure 2.3. TRMS AC Voltage Measurements
2.7.5 Resistance Measurements
The Model 175 can make resistance measurements between
1OmfI and 200MQ The 2MIf. 20MR and 200M12 ranges will
autorange when the MI1 pushbutton is pressed in. Proceed as
follows to make resistance measurements:
1. Connect the test leads to the OHMS and COM terminals of
the Model 175.
2. Select the II function.
NOTE
The message “Err” and flashing II and AC annunciators will be displayed if the AC/DC pushbutton is pressed in. This is an invalid mode. To
correct, simply release (out) the AC/DC pushy
button.
3. Select a range consistent with the expected resistance. For
automatic range selection. use the autorange mode.
NOTE
Zeroing may be necessary to compensate for
test lead resistance on the 20011 and 2kR ranqos.
Zero the display as follows:
A. Short the to?., leads together.
B, Press the REL pushbutton. The display will zero.
C. Proceed to step 4.
4. Connect the test leads to the resistance to be measured as
shown in Figure 2-4.
5. Observe the display: if the “OL” message reading is
shown, select a higher range until a normal reading is
shown. Always use the lowest possible range for the best
resolution.
6. Take the reading from the display
NOTE
See paragraph 2.8 for TRMS considerations.
NOTE
It is helpful to shield resistances greater than
106R (1MR) if a stable reading is expected. Place
the resistance in a shielded enclosure and elect
trically connect the shield to COM of the Model
175
2-5
Figure 2-4. Resistance Measurements
2.7.6 Current Measurements (DC or TRMS AC)
The Model 175 can make DC or TRMS AC current
measurements between lO@A and IOA (20A for 15 seconds).
If the expected current level is in question, make the initial
measurement with the 10A range. This will help prevent the
inadvertent blowing of the 2A current fuse.
NOTE
For routine measurements above 10A it is
recommended that the Model 1651, 50.Ampere
current shunt be used.
1. For current measurements between 2000mA and 20A.
A. Connect the test loads to the 10 AMPS and COM ter-
minals of the Model 175.
B. Select the ACA or DCA function.
C. Select an appropriate range for the expected current.
Current measurements cannot autorange.
D. Connect the test leads to the current source as shown
in Figure 2-6. If an overrange indication is displayed
select a higher range until a normal reading is shown.
Use the lowest possible range for the best accuracy.
E. Make the reading from the display.
CURRENT
TWISTED
SOURCE
Figure 2-5. Current Measurements Between 2000mA
and 20A
CURRENT
SOURCE
NOTE
The test leads used must be capable of handling
20A and it is recommended that thev be twisted
(see Figure 2-5) to minimize external iields which
could affect the Model 175 or other equipment.
Also, keep the test leads es short as possible to
minimize voltage drop.
B. Select the ACA or DCA function.
C. Select the 10A range. Current does not autorange.
D. Connect the test leads to the current source as shown
in Figure 2-5 and make the reading from the display.
NOTE
Up to 5A may be applied continuously without
degradation of the measurement due to selfhosting effects. Above 5A derate 0.15% rdg per
amp for self-heating. For currents between IOA
and 20A. specified accuracy can only be obtained when measurements are limited to a maximum of 15 seconds.
2. For current measurements up to 2000mA:
A. Connect the test leads to the mA and COM terminals of
the Model 175.
Figure 2-6. Current Measurements up to 2000mA
2.7.7 AC Plus DC Measurements
Use the Model 175 to measure TRMS on a signal which has
both AC and DC components as follows:
1. Measure and record the TRMS AC component as described in paragraph 2.7.4.
2. Measure and record the DC component as described in
paragraph 2.7.3.
3. Compute the rms value from the following equation:
E
RMS - DC
-\/E
2 + E,,?
2.7.8 dB Measurements
The dB function makes it possible to compress a large range
of readings into a much smaller scope. The relationship between dB and voltage can be expressed by the following
equation.
dB = 20 log __
VW,
VFW
2-6
Tables 2-3 and 2-4 list the dB specifications for DC volts and
AC volts.
The Model 175 can make dBm measurements referenced to
the standard 6OOn impedance or to other impedances. The
relative feature allows measurements in dB independent of
impedance.
The basic procedure for placing the instrument in the dB
mode is to first select AC or DC volts and then press the dB
button. Note that once dB is selected (dB annunciator on),
pressing in the 0 or A function pushbuttons will turn dB off.
1. dBm Measurements with 600R Reference Impedance
dBm is defined as decibels above or below a 1mW
reference. The standard reference impedance of the
Model 175 is 600% What that means is that the Model 175
is designed to read OdBm when the calculated voltage
needed to dissipate 1mW through a 600R impedance is
applied to the Model 175. That calculated voltage level is
0.7746V as derived from the basic power equation.
E=\lP.R
E = IlO-3W.6000
E = 0.7746V
Thus with a 600R reference impedance the Model 175 will
read OdBm whenever 0.7746V is applied.
NOTE
Do not confuse reference impedance with input
impedance. The input impedance of the instrument is still 1OMR (see specifications) in the dB
mode.
To make dBm measurements referenced to 60012, pro-
ceed es follows:
A. Connect the test leads to the VOLTS and COM tern
minals of the Model 175.
8. Select the ACV or DCV function.
C. Select autorange for optimum resolution.
D. Press the dB button.
E. Connect the test leads to the voltage source.
F. Make the dBm reading from the display.
2. dBm Measurements with Other Reference Impedances
dBm measurements can be made with other reference imp
pedances. The most convenient method for using other
reference impedances is to algebraically subtract the
calculated dB offset for the desired reference impedance
from the reading on the display of the Model 175. Table
2-5 lists common reference impedances and the cork
responding offset values. The following equation can be
used to calculate the offset for impedances not listed in
Table 2-5:
Offset (foi dBml : 10 log -$,o;j
To make dBm measutements referenced to another em-
pedance, proceed as follows:
A. Choose the desired reference impedance.
B. Calculate or look up the offset value in Table 2-5 for the
desired reference impedance.
C. Determine dBm at the desired reference impedance as
follows:
dBm (at ref 2) = 175 reading offset
Examole: Make dBm meawrements references to a
lOOI deference impedance
New ref I
Table 2-3. dB Specifications for DC Volts 16OOI1 Refl.
Table 2-4. dB Specifications for AC Volts l600R Refl
dS Mode lref: 60001
!
Range / Input
200mV I
1mV to 2mV
I i-58 to -52dBm)
~ (Ei, 2v.750v
*Up to 1 ktiz
-
OHz- 1
i?
IkHz
2*
2
0.2
0.2
Accuracy (* dSm)
IOkHz- / ZOkHz2OkHz 1 50kHz
! -
3
0.3 1
~
0.26
i
I
0.56
50kHz1OOkHz
-
-
1.2
2-7
i. lOOn is not listed in Table 2-5 so the offset must be
2.7.9 dB Measurement Considerations and Applica-
calculated as follows: tions
offset = 10 log (j$$)
offset = -7.78dB
ii. Subtract -7.78dB from all subsequent displayed
readings on the Model 175.
dBm measurements. referenced to another impedance,
can be read directly from the display of the Model 175 by
utilizing the REL feature, and an accurate voltage source.
The basic procedure is as follows:
A. Calculate or look up the equivalent voltage level (Table
2-5) for OdBm at the desired reference impedance.
8. Input that voltage level to the Model 175.
C. With the Model 175 in the dB mode, press the REL
button.
D. dBm measurements referenced to the desired im-
pedance can now be read directly from the display of
the Model 175.
3. dBW Measurements
dBW is defined as decibels above or below a one watt
reference. The procedure is the same as that found in
paragraph 2.7.8 step 2. The only difference is that the
reference point is OdBW (1W) rather than OdBm (1mW).
4. dElV Measurements
dBV is defined as decibels above or below 1V (OdBV
point). This is a voltage relationship independent of impedance. The basic procedure is to simply subtract 2.22
dl3 ITable 2-5) from all subsequent displayed readings on
the Model 175..
5. Relative dB Measurements
Just about any voltage level within the measurement limits
of the Model 175 can be established as the OdB point. The
basic procedure is to establish that level as the OdB point
by using REL and make the desired dB measurements.
Table 2-5. Levels for Other Reference Impedances
Reference
Impedance
(cl)
8 0.0894 2.828
50
75 0.2739
150
300 0.5477
600 0.7746
1000 1 .oooo
Equiv. Voltage
Level for:
(600!7 Refl
OdBm OdBW OdBm
-18.75 11.25
0.2236
-10.79
9.03
0.3873
6.02
3.01
0.00
1 2.22
Offset
OdBW
V, for OdBm = U10~3W-Z,,,
V;z;;:: for OdBW =\6,,,
Offset (for dBm) = 10 log[6:& )
1. Typical Instrument Performance
Typically, the Model 175 will perform better than its
published dB specification. The following example will illustrate this point:
A. Using the Model 175 in the dB mode 6000 ret)
measure a 1mV RMS. 1kHz source (common application in the communications field). Typically, the Model
175 will read -57.7dBm.
B. The calculated dBm level for that source is -57.8dBm.
C. The O.ldBm error is considerably better than the
k2dBm specification. The specifications are intended
to covet worst measurement conditions.
2. Measuring Circuit Gain/Loss
Any point in a circuit can be established as the OdB point.
Measurements in that circuit are then referenced to that
point expressed in terms of gain I + dB) or loss (-dBl. To
set the OdB point:
A. Place the Model 175 in volts, autorange and dB.
B. Connect the Model 175 to the desired location in the
circuit.
C. Press the REL button. The display will read OdB.
D. Gain/Loss measurements can now be made referenc-
ed to the OdB point.
3. Measuring Bandwidth
The Model 175 can be used to determine the bandwidth of
an amplifier as follows:
A. Connect a signal generator to the input of the
amplifier.
B. Set the Model 175 to ACV and autorange.
C. Connect the DMM and a frequency counter to the load
of the amplifier.
D. Adjust the frequency of the signal generator until a
peak AC voltage reading is measured on the Model
175.
E. Press the dB button and then press the REL button.
The OdB point is now established.
F. Increase the frequency input until the Model 175 reads
-3.OOdB. The frequency measured on the frequency
counter is the high end limit of the bandwidth.
G. Decrease the frequency input until the dB reading
again falls to -3dB. The frequency measured on the
signal generator is the low end limit of the bandwidth.
4. Determining Q
The (1 of a tuned circuit can be determined as follows:
A. Determine the center frequency and bandwidth as ex-
plained in paragraph 2.7.9 step 3.
B. Calculate Q by using the following formula:
Q = Center Frequency/Bandwidth
Offset (for ,,B,,,,) = ,O
log
2.7.10 MINIMAX and 100 Point Data Logger Operation
2.7.11 Diode Test
The data logger can store up to 100 readings and store the
minimum and maximum readings recorded during the period
that the data logger is active. The Data Logger remains active
even after 100 points of data are stored, which means the
MINIMAX readings continue to update. The only way to
deactivate the Data Logger is to press the STOiCLR button
(ST0 annunciator off1 or cycle power. The 100 points of data
are stored at one of six selectable rates from three per second
to one reading per hour. Readings for minimum and maxim
mum are sampled at the rate of three per second regardless
of the selected rate. The procedure for operating the data
logger is as follows:
Connect the desired measurement configuration to the
Model 175. Make sure that the controls of the Model 175
are set appropriately.
Loaainn Data:
Press and hold the STOiCLR pushbutton. The following reading rates will scroll on the display:
r = 0 (every reading)
r = 1 (1 rdglsecl
I = 2 (1 rdgll0 set)
r = 3 (1 rdglmin)
r = 4 (1 rdg/lO min)
r = 5 (1 rdgihr)
NOTE
There is no need to select a rate if just
minimum/maximum readings are desired.
Momentarily press the STO/CLR button to start
the logger.
Release the STO/CLR pushbutton when the desired
reading rata is displayed. The ST0 annunciator will turn
on and data will be logged at the selected rate.
NOTE
The logging cycle can be terminated at any time
bv pressina the STOiCLR button. This shuts off
tie’ data logger. However, data is retained and
can be recalled at any time as long as the instrument remains on.
Data Retrieval
Data can be retrieved at any time, but a flashing RCL annunciator indicates that the maximum number of readings
(100) have been stored.
Press and hold in the RCL pushbutton. The display will
A
scroll through the data points and MINIMAX (LO/HI).
The first data point displayed will be the last stored
reading. The next two data points will be the HI and LO
readings made during that logging cycle. Notice that
the longer the RCL pushbutton is held in the faster the
data points will scroll on the display.
6,
Release the RCL pushbutton at the desired data point
and note the reading (data) on the display. The data
pointer can be incremented by steps of one by momantarily holding in the RCL pushbutton.
Shut off the data logger by pressing the STO/CLR
pushbutton. All stored data will be retained until a new
store cycle has commenced.
The 2kU and 200kfI ranges can be used for testing semiconductor junctions as follows:
1. Select 12 function.
2. Press 2k and 200k pushbuttons [diode symbols1 in
simultaneously.
3. Display reads forward V drop of diode at 0.7mA lup to 2VI.
Red terminal is positive.
2.8 TRMS CONSIDERATIONS
Most DMMs actually measure the average value of an input
waveform but are calibrated to read its RMS equivalent. This
poses no problems as long as the waveform being measured
is a pure, low-distortion sine wave. For complex. none
sinusodial waveforms, however. measurements made with
an averagrng type meter can be grossly inaccurate. Because
of its TRMS (True Root Mean Square1 measuring
capabilities,
the Model 175 provides accurate AC
measurements for a wide variety of AC input waveforms
2.8.1 AC Voltage Offset
Typically the Model 175 will display 25 counts or less of offset
on AC volts with the input shorted. This offset is caused by
amplifier noise and offset of the TRMS converter. This offset
will not affect reading accuracy and should not be zeroed out
using the REL feature. The following equation expresses how
this offset IV
OFFsETJ is added to the signal input IV,,):
Displayed reading -\,&I2 ~+ iVoiFstri2
As long as V,, is at least 10 times larger than V,,,,,,, negligi~
ble error will occur.
Example: Range = WAC, range
Offset = 25 counts
input = 200mV RMS
Displayed Reading :: “!Q@!2 + (.Q025j2~
~$:~~o’,,“,;“oooo”3
= .2OOOV RMS
If REL is used to zero the display. the 25 counts of offset
would be subtracted from V,, resulting in an error of 25
counts in the displayed reading.
2.8.2 TRMS Measurement Comparison
The RMS value of a pure sine wave is equal to 0.707 times its
peak value. The average value of such a waveform is 0.637
times the peak value. Thus. for an average-responding meter,
a correction factor must be designed in. This correction fact
tar, K, can be found by dividing the RMS value by the
average value as follows:
K = 0.707 = 1.1,
0.637
2-9
By applying this correction factor to an averaged reading, a typical meter can be designed to give the RMS equivalent. This
works fine as long as the waveform is a pure sine. but the ratios
between the RMS and average values of different waveforms is
far from constant, and can vary considerably.
Table 2-6 shows a comparison of common types of waveforms.
For reference, the first waveform is an ordinary sine wave with a
peak amplitude of 1OV. The average value of the voltage is
6.37V, while its RMS value is 7.07V. If we apply the 1 .I 1 correction factor to the average reading, it can be seen that both meters will give the same reading, resulting in no error in the aver-
age-type meter reading.
Table 2-6. Comparison of Average and TRMS Meter Readings
Wavefoml
Sine
+10--
0
n
The situation changes with the half-wave rectified sine wave. As
before, the peak value of the waveform is 1 OV. but the average
value drops to 3.18V. The RMS value of this waveform is 5.OOV,
but an average responding meter will give a reading of 3.53V
(3.18 x 1 .l l), creating an error of 29.4%.
A similar situation exists for the rectified square wave, which
has an average value of 5V and a” RMS value of 5V. Here the
average responding meter gives a reading of 5.55V (5 x 1 .i 1).
while the Model 175 gives a TRMS reading of 5V. Other
waveform comwrisons can be found in Table 2-6.
AVerage
Responding
Meter Readine
7.07v
Half-Wave Rectified Sine
*b”n-
Full-Wave Rectified Sine
“,“m
SqlJX?
+10- -
0
Fr
Rectified Square Wave
3.53v
7.07-9
ll.lOV
1ov
1ov
5.cw
1ov ‘v?j
5.55v
ll.lV.ll
2-10
1ov 5.77v
1
5.55v
2.8.3 Crest Factor
The crest factor of a waveform is the ratio of its peak value to
its RMS value. Thus, the crest factor specifies the dynamic
range of a TRMS instrument. For sinusodial waveforms, the
crest factor is 1.414. For a symmetrical square wave, the crest
factor is unity
tang&r pulse is related to its duty cycle; as the duty cycle
decreases, the crest factor increases. The Model 175 has a
crest factor of 3. which means the instrument will give act
curate TRMS measurements of rectangular waveforms with
duty cycles as low as 10%
2.8.4 Extended Frequency Response
The crest factor of other waveforms will. of course. depend
on the waveform in question because the ratio of peak to
RMS value will vary. For example, the crest factor of a rec-
50%
Figure 2-7 illustrates the extended frequency response of the
AC volts ranges up to 1MHr.
1OHz
1OOHz
1kHz
1OkHz
0.1 FULLSCALE 2.750V RANGES
Figure 2-7. Model 175 Typical ACV Frequency Response
1OOkHz ;
1MHz
2.11/2-12
SECTION 3
PERFORMANCE VERIFICATION
3.1 INTRODUCTION
This section contains information necessary to verify that the
Model 175’s performance is within specified accuracy. Model
175 specifications may be found at the front of this manual.
Ideally, performance verification should be performed when
the instrument is first received to enwre that no damage or
change in calibration has occurred during shipment. The
verification procedure may also be performed whenever instrument accuracy is suspect or following calibration. If performance on any of the ranges or functions is substandard.
calibration can be performed as described in Section 5.
NOTE
If the instrument does not meet specifications
and it is still under warranty (less than 12 months
since date of shipmentl, contact your Keithley
representative or the factory to determine the
action to be taken.
3.2 ENVIRONMENTAL CONDITIONS
All measurements should be made at an ambient temperature
between 1W and 28X (65O to 82°F) with a relative humidity
less than 80%.
3.3 RECOMMENDED TEST EQUIPMENT
Equipment for veri’fying the performance of the Model 175 is
listed in Table 3-l. Alternate equipment may be used as long
as the equipment accuracy is at least as good as the
specifications listed in Table 3-l.
3.4 INITIAL CONDITIONS
Before performing the verification procedures, make sure the
Model 175 meets the following conditions:
1. If the instrument has been subject to temperatures below
l@‘C 165°F) or above 2E°C 18Z°F1. allow sufficient time for
the instrument to reach temperatures within the range.
Generally, it takes one hour fo stabilize an instrument that
is 10°C (18°F) outside of this range.
2. Turn on the Model 175 DMM and allow it to warm up for
one hour. The instrument may be operated from either
line power or battery pack power, as long as the battery
pack has been fully charged as described in paragraph
2.3.3.
3.5 VERIFICATION PROCEDURE
The following paragraphs give the basic verification pro-
cedure for the following functions: DC volts, AC volts,
resistance, and current.
WARNING
The following procedures require that high
voltages may be applied to the input ter-
minals of the Model 175. Use normal
precautions to avoid possible electrical
shock which could result in personal injury
or death.
3.5.1 DC Voltage Accuracy Check
CAUTION
Do not exceed 1OOOV between the
VOLTSlOHMSlmA and COM terminals or
damage to the instrument may occur.
Mfg
Fluke
Fluke
Fluke
Table 3-l. Equipment Specifications
Resistance Calibrator
DC Current Calibrator
Specifications
200mV. 2v, zov, zoov, 1ooov ranges f 0.005%
100mV. 1V. 1OV. 1OOV ranaes:
20Hz-‘50H; +ti.l% I
20kHz-100kHz *0.05%
1ooov range:
lOHz-30Hz f0.12%
50kHz-100kHz -tO.lO%
3-l
1. Select the DC V function and 200mV range.
2. Connect the calibrator to the instrument.
3. Apply positive lOO.OOOmVDC to the Model 175. The
reading must be within the limits specified in Table 3.2.
4. For each remaining range, apply the required voltage as
specified in Table 3-2, and verify that the reading is within
specifications.
5. Repeat all checks with negative voltage.
Table 3-2. Limits for dc Voltage Verification
dcV Applied
RaIIlX dc Voltaee
200 mv
I I
2 v 1.lxQoo v
20 v lO.oiml v
200 v
loo0 v
3.52 AC Voltage Accuracy Check
Do not exceed 750V RMS, IOOOV peak
lOW*Hz, between the VOLTSlOHMSlmA
and COM terminals or instrument damage
may occur.
1. Select the AC V function and the 200mV range.
2. Connect the calibrator to the DMM.
3. Set the calibrator output to lOO.OOOmV AC at a frequency
of 2QHz. Verify that the reading is within the limits
specified in Table 3-3.
4. Repeat the 1OOmV AC measurement at the other frequencies specified in Table 3-3.
5. Check the 2V. 2OV. ZOOV, and 750V ranges by applying
the required voltages and frequencies specified in Table
3-3 end verifying that the readings are within the specified
limits.
3.5.3 Resistance Accuracy Check
100.000 mV
100.000 v
1000.00 v
CAUTION
Allowable Readings
(18” c to 280 C)
99.94 to loo.06
0.9995 to l.oco5
9.995 to 10.005
99.95 to 100.05
999.5 to 1000.5
1. Select the R function (AC/DC pushbutton must be out)
and the ZOOQ range.
2. Connect the test leads to the Model 175 end short the
other ends together.
3. Press the REL pushbutton to compensate for the test lead
resistance.
4. Disconnect the short end connect the test leads to the
calibrator.
5. Set the calibrator to lOO.OOOfl and verify that the reading is
within the limits specified in Table 3-4.
6. Check the 2kR, 20kR, 200k0, and ML2 ranges by applying
the required resistances specified in Table 3-4 end verifying
that the readings are within the specified limits.
Table 3-4. Limits for Resistance Verification
c
L
3.5.4 DC Current Accuracy Check
CAUTION
Do not exceed 2A to the VOLTS/OHMS/
mA and COM terminals or the amp fuse will
blow.
1. Select the DC A function end initially, the 2000mA range.
2. Connect the calibrator to the VOLTSiOHMSimA end
COM terminals of the Model 175.
3. Apply lOO.OOO~A end switch the Model 175 to the 200pA
range. The reading must be within the limits specified in
Table 3-5.
4. Check the 2mA through 2000mA ranges by applying the
required current specified in Table 3-5 and veribing that
the readings are within the specified limits.
IWC to 28W
99.93 to 100.07
0.9994 to 1.0006
9.993 to 10.007
99.94 to 100.06
0.9993 to 1.0007
CAUTION
Do not exceed 450VDC or peak AC between the VOLTSlOHMSlmA and COM
terminals or instrument damage may occur.
2 v 1.00000 v 0.9880 to1.0120 0.993U to1.0070 0.9860 to 1.0140
20 v 1o.m v 9.880 to10.120 9.930 to10.070 9.860 to 10.140
200 v 1oo.m v 98.80 to101.20 99.30 to100.70 98.60 to 101.40
750 v 750.w v 740.5 to 759.5 744.2 to 755.8
Applied ac
Do not exceed 10A continuously or 20A for
15 seconds to the 10 AMPS and COM ter-
minals or instrument damage may occur.
Allowable Readings (18°C to 28°C)
5okHz
!
91.25 to 108.75
0.9675 to 1.0325 0.9300 to 1.07W
9.675 to 10.325 9.300 to 10.700
96.75 to 103.25 93.w to 107.w
CAUTION
1MlkHz
-
5. Set the Model 175 to the 10A range and connect the DC
current source to the 10 AMPS and COM terminals.
6. Apply a current of 1.90000A to the Model 175. The
reading must be within the limits specified in Table 3-5.
3.5.5 AC Current Accuracv Check
Since AC current uses the same circuitry as AC volts and DC
current already checked in paragraphs 3.52 and 3.5.4, no additional accuracy checks are necessary.
i for DC Current Verification Table 3-5. Limit!
i-Dt;#;tnt ““;,os,“‘;,,~;$d;gs
ZOOpA ~ lOO.OOOfiA 99.83 to 100.17
2mA 1 .OOOOOmA 0.9983 to 1~0017
20mA : lO.OOOOmA 9.983 to 10.017
lOO.OOOmA
99.78 to 100.22
2000mA lOOO.OOmA 997.8 to 1002.2
J 1 .goF?oa
1.885 10 1.914
3-313-4
SECTION 4
THEORY OF OPERATION
4.1 INTRODUCTION
This section contains an overall functional description of the
Model 175. Information pertaining to the Model 1758 Batten/
Pack option is also included. Detailed schematics and component layout drawings are located at the end of this instruction manual.
4.2 OVERALL FUNCTIONAL DESCRIPTION
The Model 175 is a 4 % digit _+20,000 count DMM with five
AC and DC voltage ranges. 7 resistance ranges and 5 AC and
DC current ranges. A simplified block diagram of the Model
175 is shown in Figure 4-l. The heart of the Model 175 is the
AID converter that translates the conditioned analog input
signals into a form usable by the microcomputer.
4.3 ANALOG CIRCUITRY
The following paragraphs conlair a description of the input
multiplexer, buffer amplifier, -2V reference and AID cons
wrier circuits. These circuits may be found on schematic
diagram number 175-106 located at the end of this manual.
10 AMPS
RESISTORS
Figure 4-1. Similplifisd Block Diagram
RANGE CONTROL
n OFlIVE
FUNCTION
MlCRO COMPUTER
LCD DISPLAY
4-1
4.3.1 Multiplexer
The multiplexer connects one of four signals to the buffer
amplifier: Signal, zero, reference, ohms reference. The
multiplexer, shown in Figure 4-2, is made up of 4 JFETS
which are controlled by the microprocessor through U114.
The FETs are driven by U109 and part of U106. The drivers
convert the digital signals of the microprocessor to signals
usable by the FETs.
a110
l
+
OUTPUT TO
OHMS
REFERNECE
a112
BUFFER
AMPLIFIER
-4)
INPUT FROM
MULTIPLEXER
CONVERTER
Figure 4-3. Simplified Schematic of the Input Buffer
Amplifier
4.3.3 -2V Reference Source
0113
REFERENCE
._. _
FROl
Figure 4-2. Simplified Schematic of the Multiplexer
Ordinarily; FET switching creates transients which could be
seen in the final measurement. These effects are minimized in
the Model 175 through the use of software generated delays
and by shorting the multiplexer bus to signal common before
each signal measurement through Cllll.
4.3.2 Input Buffer Amplifier
The input buffer amplifier provides the necessan/ isolation
between the input signal and the AID converter. The
amplifier is a noninverting, low noise, high impedance circuit
with xl or x10 gain. The amplifier gain is controlled by the
microprocessor and is range and function dependent. Figure
4-3 shows the simplified schematic of the input buffer
amplifier.
The Model 175 voltage and current measurements are based
on comparing the unknown signal with an internal -2V
reference voltage. During each measurement cycle the
microprocessor samples the unknown and uses it along with
a zero measurement and -2V signal measurement to compute
the unknown voltage.
The -2V reference is made up of a highly stable zener diode
WRlOl). an op-amp and a resistive voltage divider. U103 and
R120 A, B, C act as a constant current source to minimize the
zener voltage variations. Fill7 C, D is then used to divide
down the -6.35V zener voltage to -2V.
The output of U103 (-7V) is used as a reference voltage for
the A/D converter and as a negative supply for various components.
4.3.4 AID Converter
The Model 175 uses a combination constant frequency
charge balance, single slope analog-to-digital converter. A
simplified schematic of the A/D used in the Model 175 is
shown in Figure 4-4 with an associated output waveform.
4-2
SINGLE SLOPE
” CHARGE BALANCE
RlZlA
INTEGRATOR
OUTPUT WAVEFCJRM
I I
ELAY
NEXT -
MEASUREMENT
PHASE
The charge balance phase begins when the input
enable/disable line is set high. This occurs at the end of a
software-generated delay period that allows the signal to settle after the appropriate multiplexer FET is turned on. The actual delay period depends on the selected range and function.
Once this occurs the signal from the buffer amplifier is added
to the offset from RlZOH. This converts the bipolar signal
from the buffer (* 2V) to an unipolar input to the integrator.
The integrator ramps up until it just passes the chargebalance comparator threshold voltage. When thejsing edge
of Q3 occurs from U122 or when Ut 19 goes low. Q goes high
forcing IcB into the integrator input. Since IcB is much
greater than the current through R120G and R120H the integrator output voltage will ramp in the negative direction.
The integrator will continue ramping downward until U1198
goes low. Each time the output U121A goes high it is gated
(inside the microprocessor) with the microprocessor’s internal clock and these pulses are counted. Once U121A goes
IOM, the process repeats itself.
The charge balance phase continues for 100msec. At the end
of the charge balance phase, the output of the integrator is
resting at sc~me positive voltage. Since the integrator output
is connected to the noninverting input of the UlOBA, its out-
put will stay high until the integrator ramps negative. During
single slope Q114 is turned off and R121H is connected to
+5V. The single slope comparator is then gated with the
microprocessor’s internal clock and counted. Once the comparator output goes low the microprocessor stops counting
and can compute the reading.
PHASE
I:-
PHASE
0
‘CB
A
INPUT
ENABLE/DISABLE
1.w
Figure 4-4. AID Converter
4.3.5 input Signal Conditioning
For DCV and ACV the signal conditioning is performed by
R106, its shunt capacitors, KlOl. K102. K103 b Q106.
The following attenuation is provided:
ln the DCV mode: + 1 is used on the 200mV Et 2V ranges.
+ 10 is used on the 20V range.
+ 100 is used on the 200V range.
+ 1000 is used on the 1OOOV range.
In the ACV mode: f 1 is used on the 200mV range.
+ 10 is used on the 2V range.
i 100 is used on the 20V range.
~1000 is used on the 200V Et 1OOOV
range.
Protection for the AC Et DC voltage ranges is provided by
R103, R108, 0107 and 0108. RI03 and RlOE are used exclusively on the lower ranges of ACV Et DCV to limit current
to 0107 and Q108 during overload. During the overload Q107
and QlO8 clamp the maximum voltage on the signal FET line
to within 0.7V of the supplies.
Signal conditioning for current is performed by R109. RllO
and I?119 current shunts. For DC current measurements the
shunt voltage drop (200mV full scalel is applied directly to the
input signal FET for conversion. In AC current, the shunt
voltage drop is treated as a 200mV AC signal and is switched
to the AC converter section. Overload clamping occurs at 3
diode voltage drops which is a level high enough to permit
high crest factor current waveforms.
4-3
In DCV the properly scaled signal is applied directly to Ql IO
through R107 and CllO. In the AC V mode the scaled analog
signal is applied to the AC converter for transformation to a
DC signal that is applied to QllO.
Resistance measurements are made using the ratiometric
technique (see Figure 4-5). When the resistance function is
selected a series circuit is formed between the ohms source,
reference resistor and the external unknown resistance.
I
RX = VZ.RREF
“1
Figure 4.5. Resistance Measurements-Ratiometric
Technique
Three reference resistors are used on the ohms ranges RlOl,
R102 and R106A. RlOl is used for the 200R Et 2kR ranges,
R102 for the 20kR and 200kcl ranges and R106A for 2MR.
20Mn & 200Mn. Drive for the ohms ranges is ultimately controlled by the microprocessor through Ulll and U112.
Switching for the ohms ranges is done using low leakage
base to collector diodes of 0102, 0104 and Q105. The appropriate transistor is turned on by driving the base high
(+ 5V). The simplified schematics for the ohms circuitry is
shown is Figure 4-6.
RX
I3101 RRs~
lkn
FIT101
4.15kn
a1 04
TO 0112 OF
MULTIPLEXER
Ri06ARREF
hM
1 OMR
T
>TOQ113OF
MULTIPLEXER
RX
Figure 4-6. Simplified Schematics of Ohms Circuitry
By measuring the four inputs to the AID converter the
unknown resistance can be computed by the microprocessor
using this equation:
Osense HI - &~nse LO
RX =
For the 200% 20k0, 2Ml7 ranges fI sense HI is actually
multiplied by a factor of 10 in the buffer circuit:
Protection on the ohms ranges is accomplished by Q103,
RTIOI, 0101, R103, Q107and Q108. Foran inputvoltageapplied to the R input terminals, QlOl clamps the voltage across
RIO1 to a safe level. RTIOI limits current to Ql03 which
clamps the voltage at 0104 to a safe limit (<12V).
nref HI--nref LO
A. 2OOn
4-4
-:‘
and 2kn Ranges
For the 20k0 and 200kfl ranges protection is provided by
R102, R104 and R105. Rl06A provides protection for the
2MiI. 20Mn and 200MR ranges by limiting current.
4.3.6 AC Converter
All AC voltage inputs pass through UIOI for a x2.5 voltage
amplification. The gain stage is used to permit accurate
voltage measurement at higher frequencies and lower input
lW&.
In order to drive the display correctly four voltages are
obtained from R126. The clock required by U201 is obtained
from U122.
The output of UlOl is applied to the TRMS converter chip
which converts the AC input signal to the corresponding DC
level. The DC output is then -2.5 and applied to the signal
FET.
4.4 DIGITAL CIRCUITRY
Model 175 operation is controlled by the internal microcornputer. This section briefly describes the operation of the
Various Sections of the microcomputer and associated
digital
circuitry. A simplified block diagram is included for user
reference; for more complete circuit details refer to schematic
diagram number 175.106 at the end of this manual,
4.4.1 Microcomputer
The microcomputer centers around the 146805E2 CMOS
microprocessor. It is an 8 bit microprocessor with direct ad-
dressing of alp to 8k bytes on a shared address and data bus.
Timing of me microprocessor is accomplished by the use of
Y101; a 3.2768MHz crystal. Internally this frequency is divided down by 5 to obtain a bus operating frequency of
655.36kHz. This is present on the address strobe of U123 (pin
6) and supplies timing to all other parts of the instrument
through the binary divider U122.
The software for the MPU is stored in U115 (PROM). Temporary storage is provided by Ul13. U113 is used to share the
calibration constants on power up and as RAM for the
microprocessor’s in-house functions. It also stores readings
for the data logger. UllO is the NVRAM and is used to store
the calibration constants.
4.4.2 Address Decoding
U120 is used to latch in the address that is on the bus when
the address strobe of U123 goes high and presents it to the
PROM (U115) during data strobe.
The display board also houses the special function keys; d8,
REL, STO/CLR and RCL.
4.5 DIGITAL CALIBRATION
The Model 175 uses digital calibration to eliminate all poten-
tiometers in the instrument to facilitate calibration. The constants that the Model 175 uses are stored in a nonvolatile
electrically alterable read only memory (UllOl, and are read
on power-up of the instrument. There is one constant for
each range on DCV, ACV and I?. On the DCA and ACA func-
tions the 200mV DC and 200mV AC constants are used
respectively.
4.6 POWER SUPPLY
Fuse F102 is the LINE FUSE which is internally accessible.
SlOl is the power on/off switch and 5102 selects 115V or
230V operations by placing the transformer primary windings
in parallel or series.
TlOl, the power rransformer has two secondary windings:
one for the Model 175 and the other for the IEEE option
(Model 1753). The bridge rectifier lCR104l functions as a
fullwave rectifier for both the plus and minus supplies. R123
limits current 10 the 12V zener (VR1021 and to the batteries (if
installed1 for charging. The zener acts as a pre-regulator to
the + 5V regulator.
4.7 MODEL 1768 BATTERY OPTION
Maximum battery charging rate is achieved when the instrument is connected to line power and the on/off switch is off,
Fullwave rectified voltage from CR104 is applied to R102 and
BTlOl to charge the batteries. Cl101 acts as a current sink if
the charging current rises above 150mA. The batteries are of
the quick recharge type and will charge in 8 to 10 hours. With
the instrument turned on the batteries will trickle charge at
approximately 40mA.
4.4.3 PIA
U114 provides for most of the control of the instrument. It
controls all ranging hardware, AID converter, and data output and input for the IEEE option.
4.4.4 Display Board
The LCD display is driven by a flat pack LCD controller chip
U201 and it communicates to the microprocessor through 4
control lines. During power-up the microprocessor configures
U201 to drive the triplexed display.
With the batten/ pack installed, the negative supply is
generated using a CMOS voltage inverter (UlOll. The output
of the inverter is applied to CR101 and Cl01 for filtering.
Low battery detection is accomplished by fhe comparator
IU102) and the microcprocessor. A voltage level of 8.8V
across BTlOl signals the end of useful batten, life. The trip
level for the comparator is set by R103 and R104.
4-514-0
SECTION 5
MAINTENANCE
5.1 INTRODUCTION
This section contains installation, service and calibration in-
formation for the Model 175 and 1758 Battery Pack. In addi-
tion to front panel calibration, a program for calibrating the
Model 175 over the IEEE bus is included.
WARNING
The procedures described in this section
are for use only by qualified service personnel. Do not perform these procedures
unless qualified to do so. Many of the steps
covered in this section may expose the individual to potentially lethal voltages that
could result in personal injury or death if
normal safety precautions ere not ob-
served.
5.2 TOP COVER REMOVAL/INSTALLATION
The top cover of the Model 175 must be removed in order to
service the unit or to install the Model 1758 battery pack
and/or the Model 1753 IEEE-488 interface. Proceed es
follows:
WARNING
Disconnect the line cord and 811 other
sources and cables before removing the top
cover.
1. Turn off the power, disconnect the line cord and remove all
teet leads from the terminals of the Model 175.
2. Turn the unit over and remove the four screws from the
bottom of the case.
3. Turn the unit over again and separate the top cover from
the rest of the unit.
4. To reinstall the top cover, position the tilt bail properly into
the bottom cover and reverse the above procedure.
5.3 BAlTERY PACK (Model 1758) INSTALLATION
Refer to Figure 6-l and perform the following procedure to in-
stall the battery pack:
WARNlNG
Installation of the battery peck should only
be performed by qualified personnel.
Disconnect line cord and remove all test
leads from the terminals of the Model 175.
1. Remove the top cover as explained in paragraph 5.2.
2. Remove the shield by pulling up on either side until the
back lip disengages from the retaining clip. Ease the shield
out of the unit.
Position the battery board as shown in Figure 6~1 and
secure it to the shield using two supplied screws. The
screws are fed through the shield into the battery board
fasteners.
Place the battery pack in the bracket and position if on lhe
shield as shown. Feed the two screws through the shield
into the bracket and tighten.
CAUTlON
Do not allow the battery leads to short
together or damage to the batteries may
occur.
5. Carefully place the shield iwith battery pack1 back into the
Model 175 so that it seats properly on the two spacers.
Press down firmly on the back of the shield to engage it ins
to the retaining (and ground1 clip.
6. Connect the ribbon cable from the battery board to the
male connector (marked BATTERY) on the mother board.
CAUTION
Make a close visual inspection to ensure
that the connectore are properly mated or
damage to the instrument may result.
7. Connect the red battery lead to the + RED terminal pin on
the battery board. Connect the black batten, lead to the
-ELK terminal pin on the batten, board.
8. Reinstall the top cover as explained in paragraph 5.2.
9. Charge the batten/ pack per instructions in paragraph
2.3.3.
5.4 TROUBLESHOOTING
The troubleshooting instructions contained in this section are
intended for qualified personnel having a basic understanding
of analog and digital circuitry. The individual should also be
experienced et using test equipment es well es ordinary
troubleshooting procedures. The information presented here
has been wrinen to assist in isolating a defective circuit or
circuit section; isolation of the specific componenf is left to
the technician.
NOTE
Avoid touching the PC Board or its component
parts. Handle the PC Board by its edges.
5.4.1 Recommended Test Equipment
The success or failure in troubleshooting the Model 175
depends not only on the skill of the technician, but also relies
heavily on accurate, reliable test equipment. Table 5-1 lists
the minimum equipment and specifications recommended for
troubleshooting the Model 175. Other equipment such as
logic analyzers, capacitance meters, etc, could also be helpful
in difficult situations.
5-l
Dual-trace. triqqered-sweeD oscillo-
1 scope, Dd to%MHz bandwidth.
5.42 Self Diagnostic Program
To use the self diagnostic program, hold in the dB button and
turn on the Model 175. The following will occur:
1. All LCD digits and annunciators will turn on.
2. The software revision level will be displayed (i.e. All.
3. The sequential display test will run.
4. The Model 175 will go into the troubleshooting test mode.
If the dB button is released the instrument will flag either
RAM or NVRAM self test failures, should they occur. If
neither RAM nor NVRAM fails, the instrument will default to
the troubleshooting test mode.
RAM Test--If the RAM test fails, the Model 175 will lock up
with all zeroes displayed. Replacing U113 may correct
problem.
Non-Volatile RAM Test-If the NVRAM test fails the following message will be displayed:
GE I- r_j
This is a sufficient message indicating that the instrument is
probably not properly calibrated since calibration constants
are stored in the non-volatile RAM. The Model 175 will lock
up at this point if the test fails, but operation may be restored
for troubleshooting by pressing any front panel control button. The flashing “C” annunciator will indicate that the unit
failed the NVRAM test.
At this point try calibrating the instrument with the constants
already entered by simultaneously pressing in REL and dB un-
til CAL is displayed, and then again until Stor is displayed. If
the error gets corrected, indicating that the NVRAM is probably good, a full calibration will be needed. If the error persists try replacing the NVRAM chip (UllO). Again, the Model
175 must undergo a complete calibration after the problem is
corrected.
-
f
I I
Figure 5-l. Segment Identification
1. The “a” segments of the digits, and the dB and V annunciators are displayed.
2. The “‘b” segments of the digits are displayed.
3. The “c” segments of the digits are displayed.
4. The “d” segments of the digits, and the m (mAI RCL and
RMT annunciators are displayed.
5. The “e” segments of the digits, minus sign and the REL.
M and R annunciators are displayed.
6. The “f” segments of the digits, and the AUTO, BAT, m
(mV) and c annunciators are displayed.
7. The “g” segments of the digits, and the AC and k annunciators are displayed.
8. The decimal points, most significant digit and the STO, F
and A annunciators are displayed.
Troubleshooting Test Modes-The troubleshooting mode
is designed to switch on various switching FETs, transistors
and logic levels to allow signal tracing through the instrument. The first displayed mode will reflect the selected function and range. For example, assume that “0~1” is displayed.
The ‘Ii corresponds to the volts function, the “1” cor-
responds to the 200mV range and the “0” is the test
number. The test number can be changed by pressing in the
dB button. Table 5-2 lists the test modes for all functions and
ranges.
To update the test mode, select the new function and range
and hold in the dB button until the function symbol changes.
b
d
Sequential Display Test-Segments and annunciators are
sequentially displayed in eight steps. Use Figure 5-1 for seg-
ment identification. The steps are as follows:
5-2
Troubleshooting consists of selecting the desired test mode
and using the data found in Table 5-2 to signal trace the cirwit.
NOTES:
1. When a different function or range is selected the d6 button must be pressed and held in to update the display with
the corresponding test mode
Table 5-2. Troubleshooting Modes
2. Do not use AUTO when in AC or DC volts.
3. Use AUTO when checking circuitry on the 20MI1 and
200MR ranges (X&l.
Function
Et Flange
200MVDC
2VDC
20VDC
200VDC
1OOOVDC
200mVAC
2VAC
20VAC
200VAC
750VAC
ACAbDCA
ALL Ranges
Test
Mode
Oul
lul
2ul
3ul
ou2
lu2
2u2
3~2
ou3
1
lu3
2~3
3u3
ou4 -
lu4
2~14
t
3u4
t
ou5
lu5
2~15
3u5
Oul
lul
2ul
3ul
ou2
lu2
2u2
3~2
ou3
lu3
2~3
3u3
ou4
lu4
2~14
3u4
ou5
lu5
2~5
3u5
OAl-OA6
lAl-lA6
2A-2A6
3Al-3A6
AMP
Gain
IU105*I
x10
x10
Xl
Xl
Xl
Xl
Xl
Xl
Xl
Xl
Xl
Xl
Xl
Xl
Xl
Xl
Xl
Xl
Xl
Xl
x10
x10
Xl
Xl
x10
x10
Xl
Xl
x10
x10
Xl
Xl
x10
x10
Xl
Xl
x10
x10
Xl
Xl
x10
x10
Xl
Xl
Multiplexer
FET On
QllO
Qlll
Q113
Qlll
(1110
Qlll
Q113
Qlll
QllO
a111
Q113
Qlll
QllO
Qlll
Q113
Qlll
QllO
Qlll
(1113
Qlll
QllO
Qlll
Q113
Qlll
QllO
Qlll
(1113
Qlll
QllO
Qlll
Q113
Qlll
~0110
Qlll
Q113
a111
QllO
Qlll
Q113
Qlll
QllO
Qlll
Q113
Qlll
Ohms ;
I
; 1 0
Range Control
Logic levels on U114
1
1 0
1 0
1 0
1 0
1 0
1
0 0
0 0
0 0
0; 0
01 0
0’ 0
0 0
01 0
0
0
0'1
0; 1
0:l
0 1
-~~~~ +-1 1
1 1
1 1
1: 1
1; 1
0: 1
0: 1
0: 1
0 1
5-3
Table 5-2. Troubleshooting Mode (Cont.)
Function Et
Range
200R
2kR
20kIl
200kfl
2MII
lML2)
20 b 200M0
(Auto)
* f4 to 5V (Logic 1) at pin 9 of U106El selects X10 gain, OV (Logic 0) at pin 9 of U106B selects the Xl
5.4.3 Power Supply and Battery Pack (Model 1758)
Checks
Table 5-3 shows the various checks that can be made to the
power supplies within the Models 175 and 1758. In addition
to the normal voltage checks, it is a good idea to check the
various supplies with an oscilloscope to make sure no noise
or ringing is present.
5.4.4 A/D Converter and Display
Make sure the A/D converter and display are operating pro-
Test
Mode
001
101
201
301
401
501
601
002
102
202
302
402
502
602
003
103
203
303
403
503
603
004
104
204
304
404
504
604
005
105
205
305
405
505
605
006
106
206
306
406
506
606
Amp
Gain
(U105’
x10
x10
Xl
Xl
Xl
Xl
Xl
Xl
Xl
Xl
Xl
Xl
x10
x10
Xl
Xl
Xl
Xl
Xl
Xl
Xl
Xl
Xl
Xl
x10
x10
Xl
Xl
Xl
Xl
Xl
Xl
Xl
Xl
Xl
’ i
1
Multiplexer
FET On
QllO
Qlll
Q113
a1 10
0111
Q113
Qlll
Q113
0110
Qlll
Q112
QllO
Qlll
Q113
QllO
Qlll
Q112
(3110
Qlll
Q113
QllO
Qlll
Q112
QllO
Qlll
0113
QllO
0111
Q112
QllO
Qlll
0113
0110
Qlll
Q113
Ohms
Range
rransistors On
Q104~
Q104
Q104
(1104
Q104
-
perly before attempting to troubleshoor the signal conditioning circuits. Check these circuits using the information in
Tables 5-4 and 5-5.
5.4.5 Signal Conditioning
These circuits can be checked by using the diagnostic pro-
gram (troubleshooting modes). See paragraph 5.4.2.
Q104
-
Q104
Q104
Q104
Q104
Q104
Q104
-
Q105
Q105
Q105
Q105
Q105
Q105
Q105
Q105
Q105
Q105
Q105
Q105
Q102
Q102
0102
Q102
Q102
Q102
-
Q102
0102
Q102
0102
Q102
Range ControlI
igic levels on U114 LO
PA0 PA
1
1
1
1 1
1
1
1
1
1
1
1
1
0
0
0
0
0
0
0
0
0
0
0
0
1
1
1
1
1
1
1
1
1
1
1
1
PA2
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
PA3
1
0
1
0
1
0
1
0
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
0
0
0
0
0
0
E
0
0
0
1
1
gain.
5-4
Table 5-3. Power Supply Checks end Battery Pack (Model 17581 Checks
Step Item/Component
1 5102 Line Switch
F102 Line Fuse
2
3 Line Cord
Required Condition
Set to 115 on 230V as required
Continuity.
Plugged into line receptacle;
Remarks
S102 externally accessible from rear panel.
**If U116 is replaced, be sure that the device is properly seated on the mother board so that it will not touch
the IEEE board (if installed’.
Table 5.4. A/D Converter Checks
step
Item/Component
1
Required Condition
Turn on power; select 2V DC
Remarks
range. Short input.
U123, pin 38
2
U122, pin 10
3
U122, pin 6
4
U122, pin 5
5
U122, pin 7
6
U122, pin 12
7
3.2768MHz Clock
655.36kHz Clock
81.92kHz Clock
40. 96kHz Clock
163.84kHz Clock
1.28kHz Clock
Crystal (YlOl’
Address strobe
Synchronous clock for A/D
Synchronous clock for A/D
Synchronous clock for AID
Integrator timebase and real time
interrupt.
U119B. pin 6
8
+5V to =OV pulse train,
Charge balance synchronization signal.
3pec duration every 22~~
9 U104, pin 6
10 U108, pin 6
11 UlOEB, pin 7
Integrator Ramp
= 1.5v
Variable pulse train OV to
Comparator Reference
Comparator Output
+5v.
12 U121A, pin 6
Variable pulse train, OV to
Reference current generator.
+5v.
U107A. pin 15
13
Variable pulse train, OV to
+5v.
U107A. pin 10
14
15 U121B. pin 9
I 5msec positive going
pUlSEIS.
1OOmsec positive going pulse.
Control line for charge balance/single
slooe.
Control line for integrator
5-5
r-m-~-
step
Item/Component
‘i
!
2
P1006. pin 5
3
P1006, pin 6
4
P1006. pin 7
5
P1006. pin 2
6
P1006. pin 1
7
P1006, pin 12
8
P1006. pin 13
9
P1006. pin 14
10 LCD
11 Connector (P1006)
Table 5-5. Display Board Checks
Required Conditions
Turn on power: select the ZVDC range.
+ 3.33v
+ 1.66V
+5v i5%
81.92kHz
ov to + 5v Pulses
ov to + 5v Pulses
ov to + 5v Pulses
+ 5v to ov Pulses
Check that LCD is positioned properly.
Check that connector is not reversed.
If reversed, display test will run, then
display will blank.
Check that they are positioned properly.
Remarks
Vlcdl
Vlcd2
Power to display
clock.
Data From up
Data From up
Data From lp
Data From pp
5.5 LINE FUSE REPLACEMENT
The line fuse is located internally in the Model 175. For exact
fuse location, refer to Figure 6.1. To replace the fuse proceed
as follows:
WARNING
Disconnect the line cord and all other
sources before removing the top cover.
1. Remove the top cover as explained in paragraph 5.2.
2. If the Model 1753 IEEE-488 interface is installed it must be
removed to gain access to the fuse. The IEEE board is
secured to the mother board by a support post at the rear.
and a connector on the left side. To remove. lift board up
until it disenages from the connector and support post.
3. Replace the blown fuse with the following type: 1BA.
25OV. 3AG. Sk-Blc IKeithley P/N-F&20).
CAUTION
Do not use a fuse with a rating higher than
specified or instrument damage may occur.
If the instrument persistently blows fuses, a
problem may exist within the instrument. If
so. the problem must be rectified before
continuing operation.
the Model 175. when handling these devices, use the following precautions to avoid damaging them.
1. The ICs listed in Table 5-6 should be transported and
handled only in containers specially designed to prevent
static build-up. Typically. these parts will be received in
static-protected containers until ready for installation.
2. Remove the devices from their protective containers only
at a properly grounded work station. Also ground yourself
with a suitable wriststrap.
3. Handle the devices only by the body; do not touch the
pins.
4. Any printed circuit board into which the device is to be in-
serted must also be grounded to the bench or table.
5. Use only antistatic type solder suckers.
6. Use only grounded soldering irons.
Table 5-6. Static Sensitive Devices
4. If the IEEE interface was installed, reinstall by reversing
the procedure in step 2.
5. Reinstall the top cover as explained in paragraph 5.2.
5.6 SPECIAL HANDLING OF STATIC SENSITIVE
DEVICES
CMOS devices are designed to operate at very high impedance levels for low power consumption. As a result, any
normal static charge that builds up on your person or clothing
may be sufficient to destroy these devices if they are not
handled properly. Table 5-6 lists the static sensitive devices in
5-6
5.7 FRONT PANEL CALIBRATION
Calibration should be performed every 12 months, or if the per-
formance verification procedures in Section 3 show that the
Model 175 is out of specification. If any of the calibration procedures in this section cannot be performed properly, refer to the
troubleshooting information in this section. If the problem persists, contact your Keithley representative or the factory for further information.
The entire calibration procedure may be performed without having to make any internal adjustments if frequency compsnsatiw (see paragraph 5.7.6) has been verified. It is suggested that
the 2V. ZOV, and ZOOV ranges at 1OkHz be checked (Section 3
Performance Verification) before proceeding.
TheModel 175 Multimeter has been improved to make calibration of the instrument more convenient for the user. Later war-
sions of the Model 175 (with Revision B or C software) can now
be calibrated without having to open up the unit.
To determine what level your instrument has, hold in tha dB button and turn on the Model 175. First, all LCD digits and annunciators turn on, then the software revision level will be displayed
(for example, Al). The instrument will then run a display test
and go into troubleshooting diagnostics. Cycle power to return
unit to normal operation.
5.7.1 Recommended Calibration Equipment
Calibration can be performed using the Fluke Model 51018
Calibrator. Alternate test equipment may be used as long as the
equipmenfs accuracy is at least as good as the specifications
listed in Table 5-7.
Table 5-7. Recommended Callbratlon Equipment
WARNING
Disconnect the line cord and test leads from the
terminals of the Model 175 before removing the
top cover.
1. Remove the top cover as explained in paragraph 5.2.
2. The calibration jumper is located in the middle of the
mother board right behind the shield. Move the calibration
jumper from position A to position B as shown in Figure
5-2.
3. Replace the top cover; plug the line cord back in and turn
the instrument on.
Revision El Software
If you have Revision B level software in your Model 175, then
you no longer have to remove the top cover to calibrate the instrument if frequency compensation has been verified. The ins
strument may be calibrated from front panel buttons.
The Model 175 must be in calibration storage enable in order to
realize permanent storage of the calibration constants. If the 1”~
strument is not placed in enable. subsequent calibration will be
lost when the instrument is turned oft. Perform the following
steps to place the Model 175 in calibration storage enable:
1. If the Model 175 is presently on, turn it off using the ON.
OFF power switch.
2. While holding in ths STOiCLR button, turn the instrument
on using the ON/OFF power switch.
3. When the message ‘CAY is displayed, release the STO;
CLR button. ‘CAL” will disappear. the instrument will return
to the normal mode, and storage of calibration constants IS
enabled.
4. Press in the REL and dB pushbuttons simultaneously until
the message ‘CAL” is displayed again. Release the bull
tons. The unit is now ready to be calibrated as indicated by
the ‘c” annunciator an the display.
Revision C Software
Description
DC Voltage
Calibrator
Specifications
200mV, ZV, 2OV, 2OOV,
1 ioov ranges 10.005%
Fluke 51018
accuracy.
AC Voltage
Calibrator
200mV. 2V. 2OV, 2OOV.
11 oov ranges M.05% ac-
Fluke 51018
C”,TXY.
Resistance
Calibrator
lOOR, IkR. lOkR, 1ookR
ranaes M.005%.
Fluke 51018
1 Mii range so.01 %
IOMn range ~1~0.05%
5.72 Calibration Enable
If you have Revision A level software in your Model 175, then
there is an internal calibration jumper that enables permanent
storage of calibration constants.
The Model 175 is shipped with its internal calibrationjumperin a
disabling position. In this position, calibration cannot be done.
The jumper must be placed in its enabling position to allow calibration. Proceed as follows:
Lastly, if you have Revision C level sofhvare in your Model 175.
a rear panel external switch has been added to your instrument.
When this switch is in ENABLED, it allows you to permanently
store calibration constants.
The Model 175 is shipped from the factory with its external
CALIBRATION switch in the DISABLED position. In this pow
tion. calibration constants cannot be stored when entered from
the front panel or over ths IEEE-486 bus. The switch must be
moved to ENABLED to allow calibration constant storage.
The CALIBRATION switch is located in the middle of the rear
panel. Slide the switch to ENABLED. Plug in the line cord and
turn on the instrument. Press in the REL and dB pushbuttons
simultaneously until the message ‘CAL” is displayed Release
the buttons. The unit is now in calibration as indicated by the ‘C”
annunciator.
5.7.3 Environmental Condltlons
Calibration should be performed under laboratory conditions
having an ambient temperature of 23 k3”C and a relative hu-
midity of less than 70% With the instrument on, allow it to warm
up for one hour. If the instrument has been subjected to tem-
5-7
peratures outside this range, or to higher humidity, allow at least
one additional hour for the instrument to stabilize before beginning the calibration procedure.
7. Repeat only steps 1 and 2 for the remaining ranges using
Table 5-9 as a guide.
Table 5-9. AC Voltage Calibration
WARMNG
Some procedures require the use of high voltage. Take care to prevent contact with live circuits which could cause electrical shock result-
ing in injury or death.
NOTE
Calibration can be stopped at any time and only selected ranges can be calibrated if needed.
5.7.4 DC Voltage Calibration
Connect the calibration source to the VOLTS and COM termi-
nals of the Model 175.
1. Press in the REL and dB pushbuttons simultaneously until
the message ‘CAL” is displayed. Replace the buttons. The
unit is now in the calibration mode as indicated by the “G”
annunciator.
2. Set the voltage calibrator to output OV.
3. Press the REL button (REL on).
4. Set the Calibrator Voltage output according to the range as
shown in Table 5.6.
5. Adjust the display of the 175 to read the same a?. the voltage of the calibrator with the use of the STO/CLR and RCL
buttons. The STOKLR button increments the displayed
reading and the RCL button decrements the displayed
reading.
6. Select the next range.
7. Repeat steps 2 through 6 for the remaining DC voltage
ranges in Table 5-6.
Table 5-8. DC Voltage Calibration
175
Range
200MV
1 1000v
5.7.5 AC Voltage Calibration
Calibrator
Voltage
190.000mV 190.00mV
1 1ooo.oov 1 1ooo.ov
175
Reading
1
175
Range
200mV
‘dB annunciator must be on (indicated two-point calibration).
5.7.6 Resistance Calibration
With the Model 175 still in the calibration mode (“C annunciator
on), select the n function and connect the test leads to the
OHMS and COM terminals of the Model 175. Table 5-10 summarizes the procedure.
1. With the test leads connected to the Model 175, short the
other ends together.
2. Select the 200R range and press the REL button on the
Model 175. The REL annunciator will turn on and the display will zero (test lead compensation).
3. Disconnect the short and connect the test leads to thecalibrator.
4. Set the calibrator to output lOOn and adjust the display,
using the STO/CLR and RCL buttons for a reading of
100.00f~.
5. Press the REL button and note that the REL annunciator
Calibrator
Voltage
190.000mV
Calibrator 175
Frequency
200Hz 190.00mV
Reading
turns off.
6. Select the 2kR range and again short the test leads together.
7. Press the REL button. The REL annunciator will turn on
and the display will zero.
6. Reconnect the test leads. set the calibrator to output IkQ
and adjust the Model 175 for a reading of 1 .OOOOkR.
9. Press the REL button to turn off REL.
10. Select the 20k range and set thwxfibrator tc Jatput 1 OWL.
Adjust the Model 175 to read 1 O.OOOka.
11. Repeat step 6 for the 2OOW1,2Mnand 20MR ranges using
Table 5-l 0 as a guide.
Table 5-10. Resistance Calibration
With the Model 175 still in the calibration mode (‘C annunciator
on), select the AC volts function and connect the calibration
source to the VOLTS and COM terminals of the Model 175.
Steps 1 through 6 must be performed in the exact sequence
listed.
1. Select the 200mV range and set the calibrator to output
190.000mV @ 200Hz.
2. Adjust the display using the STO/CLR and RCL buttons to
read 190.00mV AC.
3. Press the d6 button and verify that the dB annunciator is
on.
4. Set the calibrator to output 19.0000mV @ 200Hz.
5. Adjust the display to read 19.00mV.
6. Press the REL button and verify that the dB annunciator is
off
5-a
175
Range REL’
200R on
2ka on
2OkR
2ookQ off
2MR off
20Mn
*REL is used to compensate for test lead resistance an
the 200R and 2m ranges.
5.7.7 Frequency Compensation
Check high frequency (10kHz) AC volts accuracy as explained
in section 3, Performance Verification. In the event that fre-
off 1 OMn 1 O.OOOMR
Calibration 175
Resistance
1000 1 oo.oon
lks1
off I oka 1 o.oookR
lOOka 1 oo.oowz
1MR 1 .OOOOMn
Reading
1.00001&l
quency compensation must be performed. three internal trimmer capacitors will have to be adjusted. The trimmer capacitors
are accessible through the shield (see Figure 6-l). The shield
and PC board must be secured to the bottom cover to prevent
movement. This assembly can be secured with two screws and
nuts in place of the top cover. When making adjustments use a
flat bladed, insulated calibration tool. Proceed as follows:
1. Set the Model 175 for 200VAC and set the calibrator to outs
put lOO.OOOV @ IOkHz.
2. Adjust Cl01 for a reading of 100.00 *3 counts.
3. Select the 2V range and set the calibrator to output
1 .OOOOOV @ 10kHz.
4. Adjust Cl04 for a reading of 1.000 +8 counts.
5. Repeat steps 1 and 2.
6. Select the 20V range and set the calibrator to output
1 O.OOOOV @ 1 OkHz.-
7. Adjust Cl07 for a reading of 10.000 it3 counts
5.7.9 Calibration Storage
stead the message “our is displayed, the” calibration storage
was not enabled and the calibration constants will only be valid
until the Model 175 is turned off.
To save calibration constants after “out” appears on the display.
slide the calibration switch to ENABLED. Press REL and dE 51~
multaneously until “CAL’appears on the display. Then sinwita~
“eousiy press REL and dB until ‘Star” is displayed. The callbra
tion constants are now stored and the Model 175 is ready lor
normal operation.
Slide the CALIBRATION switch back to the DISABLED position.
5.9 CALIBRATION OVER THE IEEE-488 BUS
The Model 175, with the Model 1753 installed, can be calibrated
over the IEEE-486 bus with the use of a programmable calbra
tar and a controller.
Revision A Software
To store the calibration constants, simultaneously press the
REL and dB button until the message STOR is displayed. If the
new calibration constants are not stored, they will be lost when
the unit is turned off. If it is desired to disable front panel calibration place the calibration jumper in the appropriate position (see
Figure 5-2)
-
T
1
i-E
CALIBRATION DISABLE CALIBRATIOI
Figure 5-2. Calibration Jumper
Revision B Sofhvare
To Store the calibration constants and then exit calibration, si-
multaneously press the REL and dB buttons until the message
‘Star” is displayed. If instead the message “our” is displayed,
the” calibration storage was not enabled as explained in para-
graph 5.7.2 and the calibration constants will only be valid until
the Model 175 is turned off.
Now, cycle power to the Model 175 and begin normal operation.
Revision C Sottwarc
To store the calibration constants, simultaneously press the
REL and dB buttons until the message ‘Star” is displayed. If in-
POSITION
POSITION
VIEW FROM REAR
NABLE
The following program can be used to calibrate lhe Model 175
over the IEEE-466 bus. Use the equipment listed below:
1. Model 1753 IEEE-468 Interface.
2. Fluke Model 5100 Series Calibrator with Model 5100A~05
IEEE Interface.
3. Hewlett Packard Model HP-65 Computer w!th the foliow~
mg:
A. Model HP 62937A HP-If3 Interface
B. Model HP 92936A ROM Drawers
C. I/O ROM (0065~15003)
Detailed operating instructions for the Model 1753 can be found
in the Model 1753 Instruction Manuals
Programming Example-Use the following procedure. along
with the equipment listed previously, to calibrate the Model 175
over the IEEE-466 bus.
1. Place the calibration jumper or switch in the calibration ens
able position (see paragraph 5.7.3).
2. Configure the Model 175/1753, HP-65 and the Fluke
Model 5100 Series Calibrator as a system by connecting
the instruments together with IEEE cables.
3. Set the primary address of the Model 175/l 753 to (11000)
24. Set the Model 5100 Series Calibrator to 17 (10001).
4. Turn the instruments on and allow a one hour warm up.
5. Type the following program into the HP-95.
6. After step 5 is complete check the program to make sure
there are no mistakes. If the program has eve” a small mistake it will not operate as intended.
7. Connect the output of the Model 5100 Series Calibrator lo
the input terminals of the Model 175.
9. Press the RUN key on the HP-65 to initiate the program.
9. The program will stop at ceriain predetermined points to
prompt the user to change functions. When the prompt iw
structions have bee” completed press CONT lo resume
the program.
10. When the ‘calibration is now complete” message is disk
played, place the calibration jumper or switch b&k to the
calibration disable position.
5-9
PROQRAM
REd,“TE 72’4 I ; 1:
OUTPUT 724; "VO.lX"
1: L E H P
COMMENTS
Ensures that the 175 is properly in the calibration mode.
724 : “F!.T’+”
OIUTPUT 717 i “l;i;,N”
WFIIT 5000
OCTPLIT 724 i “Vi9X”
WAIT 3000
0 IQ T P Ll T
7 1 7 I ” S ”
OIJTPIJT 724 .z “R4’*”
C:LEF(R
BEEP 50,1000
[IISP ” ‘1
DISP ” blflRNING! THE FOLLOWIN
G STEPS USE HIGH VO
LTHGE ’ ”
[lJ$,P ” ”
@lSP 2)
PROGRPM”
rJ1:;p 81 11
PRESS CONT TO RESUflE
PFilJSE
OISP 8, 80
OUTP’JT 717 i “190’J,N’i
sets 175 to ZOOmV range.
outputs ov to 176.
Turns REL on.
outputs 19Omv to 175.
Calibrate8 2OCmV range.
sets 175 to 2u range.
outputs OV to 175.
Turns REL on.
outputs 1.9v to 176.
Calibrates 2V range.
sets 175 to 2ov range.
outputs 19V to 175.
Calibrates 20V range.
Sets 6101 B to standby.
sets 175 to 2oov range.
outputs 190V to 176.
COMMENTS
Calibrates 200V range.
Sets 175 to 1OOOV range.
Outputs 1ooov to 175.
Calibrates 1OOOV range.
Sets 51018 to standby.
AC Volts Calibration
Sets 175 to 200mV range.
Outputs 190mV at 200Hz to 175.
Calibrates high end of 2OOmV range.
Turns d9 annunciator on.
Outputs 19mV at 200Hz to 175.
Calibrates low end of 200mV range.
Turns d9 annunciator off.
Sets 175 to 2V range.
Outputs 1.9V at 200Hz to 175.
Calibrates 2V range.
Sets
175
to
20V
range.
Outputs 19V at 200Hz to 175.
Calibrates 20V range.
Sets 51019 to standby.
Sets 175 to 200V range.
5.11
PROQRAM
COMMENTS
Sets 51018 to standby.
sets 175 to 750v range.
outputs 750v at 400Hr to 175.
Calibrates 750V range.
Sets 51018 to standby.
sets 175 to 2oOn range.
Set3 program for one loop.
LIfiIT 2000
NEXT fi
ljlJTPlJT 724 ; “P3X”
Gl!TFlJT 71 7
WHIT S000
0 I.1 T F i-1 T 7 2 4
WHIT 2000
rJlJTPl.iT 724
outputs 1n to 175.
Enter displayed reading (D).
A = Test lead, resistance and cai error.
output loon to 175.
B = 1000 from 51018.
F = Cal point plus lead resistance and cal error.
C$ = Command that calibrates 175 using F.
Calibrate 2OOO range.
Loops to line 1250 once.
Sets 175 to Zkfl range.
Sets program for one loop.
outputs 10 to 175.
Enter displayed reading 16).
A = Test lead resistance and WI error.
Output lkfl to 175.
B = lOOOn from 51018.
F = Cal point plus lesd resistance and cal error.
C8 = Command that cslibrates 175 using F.
Calibratea 2k0 range.
Loops to line 1400 once.
Sets 175 to 20k0 range.
outputs lOk0 to 175.
Calibrates 2OkD range.
Sets 175 to 200kD range.
COMMENTS
Outputs 1OOkfl to 175.
Calibrates 200kll range.
Sets 175 to Mfl ranges.
Outputs lM0 to 175.
Calibrates 2Mfl range.
outputs tom to 175.
Calibrates 20MO range.
Stores calibration points in NVRAM.
S-13/614
SECTION 6
REPLACEABLE PARTS
6.1 INTRODUCTION
This section contains replacement parts information, compo-
nent location drawings and schematic diagrams for the
Model 175 and Model 1758.
6.2 REPLACEABLE PARTS
Parts are listed alpha-numerically in order of their circuit
designation. Table 6-l contains a parts list information for the
Model 175. Table 6-2 contains a parts list for the displaY
board. Table 6-3 contains a parts list for the Model 1758 Battery Pack. Miscellaneous replaceable parts are not listed in a
table can be identified in Figures 6-l and 62. Table 6-4 con-
tains a complement of spare parts that can be ordered to
maintain UP to 10 Model 175’s for appr&imatelY one year.
6.3 ORDERING INFORMATION
To place an order, or to obtain information concerning
replacement parts, contact Your Keithley representative or
the factory. See the inside front cover for addresses. When
ordering include the following information:
1. Instrument Model Number
2 .Instrument Serial Number
3. Part Description
4. Circuit Description (if applicable)
5. Keithley Part Number
6.4 FACTORY SERVICE
If the instrument is to be returned to the factory for service,
please complete the service form which follows this section
List all control settings. describe problem and check boxes that apply to problem.
q
q
Intermittent
q
IEEE failure
OFroot panel operational OAll ranges or functions are bad q Checked all cables
Display or output (circle one)
aDrifts
q
IJnstable
q
OverIoad
q
Calibration only
q
Data required
(attach any additional sheets as necessary.)
Show a block diagram of your measurement system including all instruments connected (whether power is turned on or not).
Also, describe signal source.
Analog output follows display q Particular range or function bad; specify
q
Obvious problem on power-up q Batteries and fuses are OK
q
lJnable to zero
q
Will not read applied input
UC of C required
Where is the measurement being performed? (factory. controlled laboratory, out-of-doors, etc.)
What power line voltage is used?
Relative humidity?
Any additional information. (If special modifications have been made by the user. please describe.)
Other?
Ambient Temperature?
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.