Techne PS139 User Manual

Description
Part code
Quantity
Description
Part code
Quantity
PS139
Fitting stacking plates and wall plates on a Hybrigene
A trained technician should carry out this work only after reading the full instructions.
FHB4DSTK: Stacking and wall mounting kit for two units
Wall Bracket 6105214 1 M4 x 8 Pozi Pan Screw 6008652 9 M4 Crinkle Washers 6008639 9 Thumbscrew 6009552 1 Wall Plate 6105213 1 Stacking Plate 6105212 1
FHB4DKIT: Stacking and wall mounting kit for three units
Wall Bracket 6105214 1 M4 x 8 Pozi Pan Screw 6008652 15 M4 Crinkle Washers 6008639 15 Thumbscrew 6009552 1 Wall Plate 6105213 1 Stacking Plate 6105212 2
Fixing Procedure for two Hybrigenes
1. Ensure the units are turned off and remove the power to them.
2. Place one unit on the top of the other.
3. The stacking plate has three holes and three slots. It has three slots to allow for the adjustment of the feet for levelling the units.
4. Use three of the screws to attach the plate to the bottom of the top unit.
5. Use another three screws to attach the plate to the bottom hybridiser.
6. Ensure that the units are level and the screws are tight.
Page 1 of 3
PS139 Issue 4 07/2014
Fixing the stacking plate
7. For safety it is recommended that the top unit is fixed to a wall to prevent toppling. Use the final three screws to attach the wall bracket to the top of the top unit. Ensure that the screws are tight.
8. Put the hybridisers in the position you want to use them.
9. Position the wall plate and mark the position of the holes.
10. Use the appropriate screws and fixings to attach the wall plate to the wall; ensure that the screws are tight.
Wall plate
Wall bracket
Hybrigene with wall bracket attached
11. Re-position the hybridisers placing the wall bracket over the wall plate.
12. Use the thumb-screw to attach the wall bracket to the wall plate.
Wall fixings
Fixing Procedure for three Hybrigenes
If you have three units, repeat the above process attaching a stacking plate between each pair of units.
Page 2 of 3
PS139 Issue 4 07/2014
UK
North and South America
France
Asia
Middle East
Contact Information
For technical, sales or servicing information, contact your local Techne dealer or:
Bibby Scientific Ltd. Beacon Road Stone Staffordshire ST15 0SA UK Tel: +44 (0)1785 812121 Fax: +44 (0)1785 810405 E-mail: sales@bibby-scientific.com
www.techne.com
Bibby Scientific Limited Bâtiment Le Deltaparc Parc Silic PN2 7 rue du Canal BP 55437 VILLEPINTE 95944 ROISSY Charles de Gaulle France Tel: +33(0)148 63 78 03 Fax: +33(0)148 63 78 01 E-mail: ventes@bibby-scientific.com
www.bibby-scientific.com
Bibby Scientific Middle East Ltd. PO Box 27887, Engomi 2433 Nicosia Cyprus Tel: +357 22 660 423 Fax: +357 22 660 424 e-mail: sales@bibby-scientificme.com
Bibby Scientific US Inc. t/a Techne Inc. 3 Terri Lane, Suite 10 Burlington, N.J. 08016 USA Toll Free (in NA): 800-225-9243 Tel: +1 609 589 2560 Fax: +1 609 589 2571 E-mail: labproducts@techneusa.com
www.techneusa.com
Bibby Scientific - Singapore Prudential Tower, Level 26 30 Cecil Street Singapore 049712 Tel: +65 6631 2976 Fax: +44 (0) 1785 810405 e-mail: info@bibby-scientific.com
www.bibby-scientific.com
Page 3 of 3
PS139 Issue 4 07/2014
Loading...