Techne Liquid Calibration Bath User Manual

The Calibration Bath
Make sure you have read this manual carefully before you use the thermoregulator.
Techne supply a range of liquid calibration baths from 5 to 12 litres on which you can fit one of the thermoregulator range of units. The bath inner containers are manufactured from stainless steel, for maximum corrosion resistance. The 5 and 7 litre baths are deep drawn with large easy clean corner radii. The 12 litre bath is of welded construction.
The thermoregulator is designed to fit all standard laboratory baths, especially Techne baths. It will heat, circulate and safely control the temperature of the liquid in the bath within precise limits.
See the Operator's manual for full operating details for the thermoregulator.
Set Up
1 There are two modes of operation for the pump; circulation
internal to the bath; circulation external to the bath. For internal circulation the blanking caps on the top of the
outlet and return pipes should be securely in place. They screw on and, for safety, they may be tight.
For external circulation the blanking caps need to be removed. With the blanking cap removed, and an external circuit set up, the liquid will be split between internal flow and external flow.
2 CAUTION: DO NOT SWITCH THE THERMOREGULATOR
ON UNLESS EITHER THE BLANKING PLUGS ARE FITTED TO THE TOP OF THE PIPES or AN EXTERNAL CIRCUIT IS FULLY CONNECTED.
IF YOU ARE PUMPING TO AND FROM A SECOND OPEN BATH A RESTRICTION MUST BE PUT ON THE PRESSURE SIDE OF THE PIPES. THE PRESSURE VOLUME IS GREATER THAN THE SUCTION VOLUME AND YOU COULD OVER-FILL
THE SECOND BATH 3 Ensure that the bath is set up on a flat level surface. 4 Fit the unit securely to the bath. THE UNIT MUST ALWAYS
BE MOUNTED WITH THE BACK AND THE SWITCH END
OUTSIDE THE AREA OF THE BATH. This will reduce the
infiltration of hot vapours into the cooling system of the
thermoregulator. Ensure that at all times the air inlet and
outlet remain clear of obstructions. Free circulation of air
inside the unit is essential for proper cooling of the
electronics and pump motor. 5 The Thumb Screw secures the unit to the bath at the mains
Switch end. 6 If the pump has been set correctly for external circulation,
suitable hoses should be fitted to the outlet and return pipes.
A suitable hose must be capable of withstanding both the
temperature of operation and the liquid being used. Always
securely clip the hoses in place.
Material Allowable Temperature Range Comments
PVC 10°C to 60°C (50°F to 140°F) For water only
Silicone -40°C to 200°C (-40°F to 362°F)NOT for silicone oil
Viton -20°C to 250°C (-4°F to 482°F)
7 You will need to remove the cooling coil when you are using
the thermoregulator on one of the Techne Calibration Baths. Remove the screw which is used to fit the cooling coil to the base of the unit. Keep the cooling coil in a safe place in case you need to replace it.
A cooling coil is fitted to the bath. Connect a hose from a tap to one end of the coil and from the other end of the coil to drain. Adjust the water flow to give the required cooling.
8 For lower temperatures a Flow Cooler, such as the Techne
FC-200 or FC-500 is required.
See the instruction Manuals for these units for connections and setup.
9 Fill the bath to between the minimum and maximum levels
stated in the specification. If you are using oil at the highest temperatures, remember that
they will expand as the temperature rises. Even if you fill
the bath with oil to the minimum level, it may well reach the maximum level before the top temperature is reached. Take all precautions and monitor the level
through the full range. If water is used, demineralised water is preferred to reduce
the formation of scale. If scale should form, use only mild de­scaling agents to remove it. DO NOT attempt to hammer, chip or scrape the deposits from the surface of the bath
10 Recommended liquids:
Temperature Liquid
-40oC to 0oC 40% water 40% ethylene glycol 20% alcohol
-20oC to 30oC 50% water 50% ethylene glycol 5oC to 95oC water, preferably de-ionised with neutral pH 10oC to 150oC Dow Corning Silicone Oil 200 series * 10°C to 250°C Dow Corning Silicone Oil 210H/100cs series *
* Warning: check gel life at top end of range.
Extraction may be necessary at high temperatures; always check the manufacturer's data and safety sheets before using any of the liquids.
In all cases the OVERTEMPERATURE CUT-OUT must be set correctly for the liquid being used and the temperature at which it is to operate
11 If you are using a low viscosity oil with a TU-20C controller,
the pump on the controller may be too powerful and cause the oil to splash from the bath. There is a flow restrictor supplied with the Calibration Bath. Push the "top hat" restrictor into the outlet of the pump and fit the retaining plate over it. Remove the appropriate screw from the pump and fix the retaining plate with the same screw.
12 A bath that is fitted with a lid or insulating ball blanket gives
the best operating conditions. A lid or ball blanket will prevent vapour loss, heat loss and give better temperature control. If an open bath is used above 80oC (ie where steam or other readily condensing vapours are present) the operation of the unit, particularly the digital display, may be affected.
Below about 80oC a cover becomes less important but will still give better temperature control.
Bath Sizes
Bath LCB-5 LCB-7 LCB-12 Overall Length A 351 mm 351 mm 351 mm
Overall Width B 260 mm 260 mm 260 mm Overall Height C 183 mm 233 mm 358 mm Opening Length D 140 mm 140 mm 140 mm Opening Width E 140 mm 180 mm 140 mm Max. Working Depth F 130 mm 140 mm 305 mm Min. Working Depth G 90 mm 90 mm 265 mm Maximum Capacity 5.0 l 6.5 l 12.0 l Minimum Capacity 4.0 l 5.5 l 11.0 l
D
E Lid
C G F
B Drain Point
A
Cooling Coil Standard Probe
Entry
Thermoregulator Fixing Point
Working Environment
The units are designed to work safely under the following conditions: Ambient temp. range 5°C to 40°C Humidity Up to 95% relative humidity, non-condensing Note: The control specifications quoted are for an ambient
temperature range of 10°C to 30°C. The specification may deteriorate outside this range but the unit will still work safely.
Radio frequency interference tested and passed to
EN50081-1.
Immunity Tested and passed to
EN50082-1
Additional Information
NOTE THAT THIS EQUIPMENT SHOULD ONLY BE DISMANTLED BY PROPERLY TRAINED PERSONNEL. REMOVING THE TOP
CASE EXPOSES POTENTIALLY LETHAL MAINS VOLTAGE.
THERE ARE NO OPERATOR MAINTAINABLE PARTS WITHIN THE EQUIPMENT.
In the unlikely event that you experience any problems with your Thermoregulator which cannot easily be remedied, you should contact your supplier and return the unit if necessary. Please include any details of the fault observed and remember to return the unit in its original packing. Techne accept no responsibility for damage to units which are not properly packed for shipping: if in doubt, contact your supplier.
Operator maintenance
1. Cleaning
Before cleaning your unit ALWAYS disconnect from the
power supply and allow to cool below 50° C.
Your Thermoregulator can be cleaned by wiping with a
damp soapy cloth. Care should be exercised to prevent
water from running inside the unit. Do not use abrasive
cleaners.
2. Overtemperature cut-out
The overtemperature cut-out is a sensitive mechanical
device and mechanical shock can cause it to trip.
In the event of no heater power, check the mains plug and
lead, then reset the cut-out control.
Repeated operation of the cut-out indicates a serious fault:
you may need to return the unit to your supplier for repair.
3. Fuses
Your unit is protected by two fuses.
For the approriate fuse see the User's Manaul for the unit.
These should only be changed by suitably qualified
personnel.
If the fuses blow persistently, a serious fault is indicated
and you may need to return the unit to your supplier for
repair.
Accessories
There are many accessories both for the thermoregulators and the baths. These are constantly being updated and details can be found in Techne Sales literature. However they do include:
Flow Coolers
FFC2D FC-200 230V UK FFC2D FC-200 230V Europe FFC2P FC-200 120V FFC5D FC-500 230V UK FFC5D FC-500 230V Europe FFC5P FC-500 120V
FTU232 "Thermsoft" kit inc: disc, cable and
connector for TU-20C
Contact Information
For technical, sales or servicing information, contact your local Techne dealer or,
Barloworld Scientific Ltd Beacon Road, Stone Staffordshire ST15 0SA, United Kingdom Telephone: +44(0)1785 812121 Fax: +44(0)1785 813748 e-mail: equipment@barloworld-scientific.com
or,
Techne Inc, 3 Terri Lane, Suite 10, Burlington, New Jersey 08016, USA. Telephone: 609-589-2560 Toll free: 800-225-9243 ext 306 Fax: 609-589-2571 e-mail: labproducts@techneusa.com Web site: www.techneusa.com
Liquid Calibation
Baths (LCB)
Reference Card
WARNING
Warning: HIGH TEMPERATURES ARE DANGEROUS Avertissement: DANGER DE TEMPERATURES ELEVEES Warnung: HOHE TEMPERATUREN SIND GEFÄHRLICH Aviso: LAS TEMPERATURAS ELEVADAS SON PELI
Issue 5
Date of issue:
3rd May 2005
Techne is a trade mark
© Techne 2002