DB-3 and DB-3A
Dri-Block
®
Operator’s Manual
Issue 21
Date of issue:
3rd May 2005.
Techne is a trade mark
CONTENTS
DB-3 and DB-3A Operator’s Manual
SAFETY and INSTALATION
THE DRI-BLOCK SPECIFICATION
OPERATION 18
ADDITIONAL INFORMATION 22
D B - 3 a n d D B - 3 A O P E R A T O R ’ S M A N U A L
page
English 4 Français 6 Deutsch 8 Español 10 Italia 12
®
14
DB-3 15 DB-3A 16 Working Environment 17 Unpacking 17
Preparation 18 The overtemperature cutout 19 Setting the Operating Temperature 20 After Use 21
Insert Blocks 22 Operator maintenance 23 Replacement parts 24
P A G E 1
D B - 3 a n d D B - 3 A O P E R A T O R ’ S M A N U A L
P A G E 2
D B - 3 a n d D B - 3 A O P E R A T O R ’ S M A N U A L
NOTES
P A G E 3
D B - 3 a n d D B - 3 A O P E R A T O R ’ S M A N U A L
Introduction
Please read all the information in this booklet before using the unit.
Warning
HIGH TEMPERATURES ARE DANGEROUS: they can cause serious burns to operators and ignite combustible material. Techne have taken great care in the design of these units to protect operators from hazards, but Operators should pay attention to the following points:
USE CARE AND WEAR PROTECTIVE GLOVES TO PROTECT HANDS;
DO NOT put hot objects on or near combustible objects;
DO NOT operate the unit close to inflammable liquids or gases;
DO NOT place any liquid directly in your unit;
At all times USE COMMON SENSE.
Operator Safety
All Operators of Techne equipment must have available the relevant literature needed to ensure their safety. It is important that only suitably trained personnel operate this equipment, in accordance with the instructions contained in this manual and with general safety standards and procedures. If the equipment is used in a manner not specified by Techne the protection provided by the equipment to the Operator may be impaired. All Techne units have been designed to conform to international safety requirements and are fitted with an overtemperature cutout. On some models, the cutout is adjustable and should be set to suit the application. On all other models the cutout is preset to protect the unit. If a safety problem should be encountered, switch off at the mains socket and remove the plug from the supply.
Installation
1. All Techne units are supplied with a power cable. This may be integral or plug-in.
2. Before connecting the mains supply, check the voltage against the rating plate. The rating plate
is on the rear of the unit. Connect the mains cable to a suitable plug according to the table below.
Note that the unit must be earthed to ensure proper electrical safety.
The fused plug supplied with the mains lead for use in the UK is fitted with the following value fuse to protect the cable: 5 AMP. The fuse in the unit protects the unit and the operator Note that units marked 230V on the rating plate work at 220V; units marked 120V work at 110V. In both cases, however, the heating rate will degrade by approximately 8%.
3. Plug the mains cable into the socket on the rear of the unit.
4. Place the unit on a suitable bench or flat workspace, or in a fume cupboard if required, ensuring
that the air inlet vents on the underside are free from obstruction.
5. Note that the following symbols may be next to the indicator lamps on the front panel of the units and have the following meanings:
6. Symbols on or near the power switch of the unit have the following meanings:
Connections 220V-240V 110V-120V
Live Brown Black Neutral Blue White Earth Green/yellow Green
: the power indicator
: the heater indicator
: the overtemperature indicator
I : mains switch On O : mains switch Off
P A G E 4
D B - 3 a n d D B - 3 A O P E R A T O R ’ S M A N U A L
After use
When you have finished heating samples, remember that parts of the unit – the tubes, blocks and associated accessories – may be very hot. Take the precautions listed earlier.
Guarantee
The unit is guaranteed against any defect in material or workmanship for the period specified on the enclosed guarantee card. This period is from the date of purchase, and within this period all defective parts will be replaced free of charge provided that the defect is not the result of misuse, accident or negligence. Servicing under this guarantee should be obtained from the supplier. Notwithstanding the description and specification(s) of the units contained in the Operator’s Manual, Techne hereby reserves the right to make such changes as it sees fit to the units or to any component of the units. This Manual has been prepared solely for the convenience of Techne customers and nothing in this Instruction Book shall be taken as a warranty, condition or representation concerning the description, merchantability, fitness for purpose or otherwise of the units or components.
Operator maintenance
NOTE: THAT THIS EQUIPMENT SHOULD ONLY BE DISMANTLED BY PROPERLY TRAINED PERSONNEL. REMOVING THE SIDE, FRONT OR REAR PANELS EXPOSES POTENTIALLY LETHAL MAINS VOLTAGES. THERE ARE NO OPERATOR MAINTAINABLE PARTS WITHIN THE EQUIPMENT. In the unlikely event that you experience any problems with your unit which cannot easily be remedied, you should contact your supplier and return the unit if necessary. Please include any details of the fault observed and remember to return the unit in its original packing. Techne accept no responsibility for damage to units which are not properly packed for shipping: if in doubt, contact your supplier. See the Decontamination Certificate supplied with your unit.
1. Cleaning
Before cleaning your unit ALWAYS disconnect it from the power supply and allow it to cool below 50° C. Your unit can be cleaned by wiping with a damp soapy cloth. Care should be exercised to prevent water from running inside the unit. Do not use abrasive cleaners.
2. Overtemperature cutout
In the event of no heater power, check the mains plug and lead. Repeated operation of the cutout indicates a serious fault: you may need to return the unit to your supplier for repair.
3. Fuses
Your unit is protected by one or two fuses. These should only be changed by suitably qualified personnel. If the fuses blow persistently, a serious fault is indicated and you may need to return the unit to your supplier for repair.
Contact Information
For technical, sales or servicing information, contact your local Techne dealer or,
Barloworld Scientific Ltd Beacon Road, Stone Staffordshire ST15 0SA, United Kingdom
Telephone: +44(0)1785 812121 Fax: +44(0)1785 813748 e-mail: equipment@barloworld-scientific.com
or,
Techne Inc, 3 Terri Lane, Suite 10, Burlington, New Jersey 08016, USA.
Telephone: 609-589-2560 Toll free: 800-225-9243 ext 306 Fax: 609-589-2571 e-mail: labproducts@techneusa.com Web site: www.techneusa.com
P A G E 5
D B - 3 a n d D B - 3 A O P E R A T O R ’ S M A N U A L
Introduction
Veuillez lire attentivement toutes les instructions de ce document avant d’utiliser l’appareil.
Avertissement
DANGER DE TEMPERATURES ELEVEES : les opérateurs peuvent subir de graves brûlures et les matériaux combustibles risquent de prendre feu. Techne a apporté un soin tout particulier à la conception de ces appareils de façon à assurer une protection maximale des opérateurs, mais il est recommandé aux utilisateurs de porter une attention spéciale aux points suivants :
PROCEDER AVEC SOIN ET PORTER DES GANTS POUR SE PROTEGER LES MAINS.
NE PAS poser d’objets chauds sur ou près de matériaux combustibles.
NE PAS utiliser l’appareil à proximité de liquides ou de gaz inflammables.
NE PAS verser de liquide directement dans l’appareil.
FAIRE TOUJOURS PREUVE DE BON SENS.
Sécurité de l’opérateur
Tous les utilisateurs de produits Techne doivent avoir pris connaissance des manuels et instructions nécessaires à la garantie de leur sécurité. Important : cet appareil doit impérativement être manipulé par un personnel qualifié et utilisé selon les instructions données dans ce document, en accord avec les normes et procédures de sécurité générales. Dans le cas où cet appareil ne serait pas utilisé selon les consignes précisées par Techne, la protection pour l’utilisateur ne serait alors plus garantie. Tous les appareils Techne sont conçus pour répondre aux normes de sécurité internationales et sont dotés d’un coupe-circuit en cas d’excès de température. Sur certains modèles, ce coupe-circuit est réglable pour s’adapter à l’application désirée. Sur d’autres modèles, il est pré-réglé en usine pour assurer la protection de l’appareil. Dans le cas d’un problème de sécurité, coupez l’alimentation électrique au niveau de la prise murale et enlevez la prise connectée à l’appareil.
Installation
1. Tous les appareils Techne sont livrés avec un câble d’alimentation qui peut être intégré à l’appareil ou à raccorder.
2. Avant de brancher l’appareil, vérifiez la tension requise indiquée sur la plaque d’identification. Raccordez le câble électrique à la prise appropriée en vous reportant au tableau ci-dessous. Il est
important que l’appareil soit relié à la terre pour assurer la protection électrique requise.
Le fusible à l’intérieur de l’appareil est destiné à assurer la protection de l’appareil et de l’opérateur. Remarque : les appareils dont la plaque indique 230 V peuvent fonctionner sur 220 V, et ceux dont la plaque indique 120 V peuvent fonctionner sur 110 V. Dans les deux cas cependant, la capacité de chauffage diminuera d’environ 8 %. La plaque d’identification se trouve à l’arrière de l’appareil.
3. Raccordez le câble d’alimentation à la prise située à l’arrière de l’appareil.
4. Placez l’appareil sur un plan de travail ou surface plane, ou le cas échéant, dans une hotte d’aspiration, en s’assurant que les trous d’aération situés sous l’appareil ne soient pas obstrués.
5. Les symboles ci-dessous situés à côté des témoins lumineux sur la face avant de l’appareil ont la signification suivante :
6. Les symboles situés sur ou à côté de l’interrupteur de l’appareil ont la signification suivante :
Connexions 220V-240 V 110V-120 V
Phase marron noir Neutre bleu blanc Terre vert/jaune vert
: témoin d’alimentation : témoin de chauffage : témoin d’excès de température
O : arrêt l : marche
P A G E 6