Power up screen.................................................................................................................................... 22
The front panel controls ......................................................................................................................... 22
Setting the temperature units ................................................................................................................ 23
Setting the operating temperature ......................................................................................................... 23
Setting the timer .................................................................................................................................... 23
After use ................................................................................................................................................ 24
User maintenance ................................................................................................................................. 24
Support contacts.................................................................................................................................... 27
REPLACEMENT PARTS .......................................................................................... 29
DECLARATION OF CONFORMITY ......................................................................... 30
4
INTRODUCTION
Voltage,
Please read all the information in this manual before using the unit.
Techne Dri-Block® heaters provide a safe, dry, constant temperature source in the laboratory. The
units can be used for incubation, boiling, inactivation, wet ashing, sample concentration, enzyme
analysis and many other clinical and industrial purposes.
The units cover the temperature range from ambient +5°C to 100°C or 200°C (depending on model)
using machined aluminium alloy blocks as the heat transfer medium. Independent indicator lights
indicate mains power, heater action and over-temperature cut-out. An ON/OFF switch is mounted on
the rear panel, together with the mains connector and a double fuse holder.
Temperature setting is via a simple push button display. Temperatures can be displayed in °C or °F.
Each unit also has a built-in count-down timer which can be set in 1 minute intervals to 99h 59m.
The units are constructed of painted aluminium alloy and polyphenylene sulphide (PPS) plastic,
which is strong yet lightweight and can resist sustained temperatures as high as 260°C. Even though
the units heat up rapidly, venting and highly efficient insulation ensure that the cases remain cool
enough to handle even at maximum operating temperatures.
Each unit is supplied with an extraction tool for easy removal of the insert blocks.
The operating instructions in this Operator’s Manual cover the following range of digital Dri-Block®
heaters:
The features of individual models are outlined in the Specifications section of this manual.
Note: This manual covers the digital Dri-Block® heater range. The analogue Dri-Block® heaters DB-2A,
DB-3 and DB-3A are covered in a separate user manual.
Before using the Dri-Block® heater please make sure you have read this manual carefully. If there is
any doubt relating to the proper use of this equipment, the staff at Bibby Scientific Ltd. or your
supplier will be happy to assist you.
UNPACKING
When unpacking the unit please ensure that the following have been removed from the packaging:
• Dri-Block® heater
• 6101308 Extraction tool
• Mains cables (UK and EU or US)
• Operator’s manual
• Guarantee card
The user is advised to keep the original packaging in case the instrument ever needs to be returned
for service or repair. Bibby Scientific Ltd. accepts no responsibility for damage incurred unless the
unit is correctly packed and transported in its original packaging.
The Dri-Block® heater range
6
SAFETY AND INSTALLATION
Connection
s 230V
, 50/60Hz
115V
, 50/60Hz
Live Brown
Black
Neutral
Blue White
Earth
Green/yellow
Green
Please read all the information in this manual before using the Dri-Block® heater.
WARNING
HIGH TEMPERATURES ARE DANGEROUS: they can cause serious burns to operators and ignite
combustible material.
Techne have taken great care in the design of these units to protect operators from hazards, but
operators should pay attention to the following points:
• USE CARE AND WEAR PROTECTIVE GLOVES TO PROTECT HANDS
• DO NOT use combustible substances near hot objects
• DO NOT operate the instrument in the vicinity of inflammable liquids or gases.
• DO NOT place any liquid directly into the instrument.
OPERATOR SAFETY
All operators of Techne equipment must have available the relevant literature needed to ensure their
safety. It is important that only suitably trained personnel operate this equipment, in accordance with
the instructions contained in this manual and with general safety standards and procedures. If the
equipment is used in a manner not specified by Bibby Scientific Ltd. the protection provided by the
equipment to the operator may be impaired.
All Techne instruments are designed to conform to international safety requirements and are fitted
with an over-temperature cut-out.
If a safety problem should be encountered, switch off the unit at the mains socket and remove the
plug from the electricity supply.
INSTALLATION
The instrument should be carried using both hands. Never move or carry the instrument when in use
or connected to the mains electricity supply.
1. All Techne instruments are supplied with a power cable; this may be integral or plug-in.
2. Before connecting the instrument to the mains electricity supply, check the voltage
against the rating plate (located on the back of the unit). Please note that the unit must be earthed to ensure proper electrical safety. Connect the mains cable to a
suitable plug according to the table below.
UK ONLY: The fused plug supplied with the mains cable is fitted with a 5 amp fuse to protect the
instrument and the operator.
Note that units marked 230V on the rating plate work at 207V; units marked 115V work at 104V. In
both cases, however, the heating rate will degrade by approximately 8%.
3. Place the unit on a suitable flat bench or in a fume cupboard if required, ensuring that the air inlet
vents on the underside are free from obstruction.
4. Plug the mains cable into the socket on the back of the instrument.
5. Switch on the instrument. Symbols on or near the power switch of the unit have the following
meanings:
7
I Mains Switch On
O Mains Switch Off
Replacement cable
Should the mains lead need replacement, a cable of 1mm2 of harmonized code H05VV-F connected
to an IEC320 plug should be used. IF IN DOUBT CONSULT A QUALIFIED ELECTRICIAN.
ENVIRONMENTAL CONDITIONS
The Dri-Block® heateris designed operate under the following conditions:
• Indoor use
• Ambient temperature range +5ºC to +40ºC
• Altitude to 2000m
• Relative humidity up to 80%
• Mains supply fluctuations not exceeding 10%
• Over voltage category II IEC 60364-4-443
• Pollution degree 2
• Use with a minimum distance all-around of 200mm from walls or other items
Note: The control specifications are quoted at an ambient temperature of 20ºC. The specification
may deteriorate outside an ambient temperature of between 10ºC and 30ºC.
The instrument has been tested for radio frequency interference and is certified under EN61326.
GUARANTEE
The instrument is guaranteed against any defect in material or workmanship for the period specified
on the enclosed guarantee card. This period is effective from the date of purchase; within this period
all defective parts will be replaced free of charge provided that the defect is not the result of an
accident, misuse or negligence. Servicing under this guarantee should be obtained from the supplier
of the instrument.
This manual has been prepared for the convenience of Techne’s customers and nothing in this
manual shall be taken as a warranty, condition or representation concerning the description,
merchantability, fitness for purpose or otherwise of the unit or components.
Notwithstanding the description and specification(s) of the instruments contained in the operator’s
manual, Techne reserves the right to make such changes as it sees fit to the instruments or to any of
the components.
8
SÉCURITÉ ET CONSIGNES D’INSTALLATION
Connexion
s 230V
, 50/60
Hz 115V
,
50/60
Hz
Phase
Marron
Noir
Neutre
Bleu Blanc
Terre
Vert/jaune
Vert
Veuillez lire attentivement toutes les instructions de ce document avant d’utiliser le Dri-Block® heater.
AVERTISSEMENT
Les TEMPÉRATURES ÉLEVÉES SONT DANGEREUSES car elles peuvent provoquer de graves
brûlures chez l’opérateur et enflammer les matériaux combustibles.
Techne a apporté un soin tout particulier à la conception de ces appareils de façon à assurer une
protection maximale des opérateurs, mais il est recommandé aux utilisateurs de porter une attention
spéciale aux points suivants :
•PROCEDER AVEC PRUDENCE ET PORTER DES GANTS POUR SE
PROTEGER LES MAINS
• NE PAS utiliser de matériaux combustibles auprès d’objets chauds.
• NE PAS utiliser l’appareil à proximité de liquides ou de gaz inflammables
• NE PAS verser de liquides directement dans l’appareil.
SECURITE DE L’OPÉRATEUR
Tous les utilisateurs de produits Techne doivent avoir pris connaissance des consignes et instructions
nécessaires à la garantie de leur sécurité. Important: cet appareil doit impérativement être manipulé
par un personnel qualifié et être utilisé selon les instructions données dans ce document, en accord
avec les normes et procédures de sécurité générales. Dans le cas où cet appareil ne serait pas utilisé
selon les consignes précisées par Bibby Scientific Ltd., la protection pour l’utilisateur ne serait alors
plus garantie.
Tous les appareils Techne sont conçus pour répondre aux normes de sécurité internationales et sont
dotés d’un coupe-circuit en cas de surchauffe.
En cas de problème de sécurité, couper l’alimentation électrique au niveau de la prise murale et
enlevez la prise connectée à l’appareil.
INSTALLATION
Porter l'appareil à deux mains. Ne jamais déplacer ou transporter l’appareil lorsqu’il est en
fonctionnement ou branché à l’alimentation électrique.
1. Tous les appareils Techne sont livrés avec un câble d'alimentation, qui peut être intégré à l'appareil
ou à raccorder.
2. Avant de raccorder l’appareil à l’alimentation électrique sur secteur, vérifier la tension
requise indiquée sur la plaque d’identification (située au dos de l’appareil). Il est
important que l’appareil soit relié à la terre pour assurer la protection électrique
requise. Brancher le câble secteur sur une prise appropriée, voir tableau ci-après.
ROYAUME-UNI SEULEMENT: La prise avec fusible intégré fournie avec le câble secteur est
munie d’un fusible 5 A destiné à protéger l’appareil et l’utilisateur.
Remarque : les appareils dont la plaque indique 230 V peuvent fonctionner sur 207 V, et ceux dont
la plaque indique 115 V peuvent fonctionner sur 104 V. Dans les deux cas cependant, la capacité
de chauffage diminuera d’environ 8 %.
9