Tasco TA430D Instruction Manual

TA430D
Refrigerant Leak Detector
Instruction Manual
CONTENT Page
1. GENERAL INFORMATION ------------------1
2. FEATURES
3. SPECIFICATIONS
-------------------------------2
4. OPERATION GUIDE
---------------------------3
5. PART & CONTROL
------------------------------5
6. GETTING STARTED
----------------------------8
6-1 Installing Battery-------------------------------- 8
6-2 Ch
arging -----------------------------------------
8
6-3 Li
thium-ion Battery Care ---------------------8
6-4 Automatic Ambient Reset Feature-
--------10
6-5 Button Feature---------------------------------10
6-6 Sensitivity Adjustment Feature--------------13
7. OPERA
TING PROCEDURE
----------------15
8. CLEANING
----------------------------------------19
Infrared Refrigerant Leak Detector
1
1. GENERAL INFORMATION
Thank you for purchasing Infrared Refrigerant Leak Detector. Read through the instruction manual before operation for correct and safe usage. Please store and retain this instruction manual for future reference.
2. FEATURES
Infrared Refrigerant Leakage Detector uses a newly developed infrared gas sensor which is extremely sensitive to variety of general used refrigerant. It provides features of fast-response, high-reliability, and long-durability. It is an ideal tool for maintaining the air-condition or a cooling system with compressor and refrigerant.
Microprocessor Control with advanced digita
l
signal processing.
Multi-color visual display.
Mute function to turn off audible alarm.
Turbo-High-Medium-Low leak sensitivity selector
.
Low battery
indication.
Infrared gas sensor
.
Detection of Freon gases, including
R-134a,
R-410A, R-407C, R404A, R22, R32, HFO-1234yf,
etc.
Carryi
ng case included.
15.5” (40 CM) flexible st
ainless probe.
Ambient concentration reset at de
fault.
Disable
Auto-reset (Ambient concentration reset).
Infrared Refrigerant Leak Detector
2
Peak Hold function – find leaks in noisy
environment
s.
Automatic zero and background co
mpensation at
default.
AC USB Adapter: @5V 1A
3. SPECIFICATIONS
Detectable Gases:
R-134a, R-404A, R-407C, R-410A, R-404A, R-
22,
R32, HFO-1234yf etc.
Sensiti
vity:
T H M L
R-22, R-32, R-134a, 404A, 407C, 410A, HFO-1234
yf
3g/ year
10g/ year
50g/ year
100g/ year
Alarm Method:
Buzzer, Tricolor LED bar Indicator
Power Usage:
3.7Vdc rechargeable Lithium-ion battery
Snake Tube length: 40cm ( 15.5” )
Dimension / Weight:
198 x 70 x 42.5 mm ( approximately 365g )
Accessories:
User manual, carry case, Li-ion battery, Micro USB
Infrared Refrigerant Leak Detector
3
Cable, AC-USB adapter
Battery Life: Approximately 5 hours normal use
Battery Charge Time: approximately 6 hours
Auto power OFF: 10 minutes
Warm-Up Time: Approximately 45 seconds
Operating Temperature & Humidity:
0 ~ 40°C, < 80% RH
Storage Temperature & Humidity:
-10 ~ 60°C, < 70% RH
Altitude: < 2000M (6500’)
4. OPERATION GUIDE
(1) The Infrared Refrigerant Leak Detector unit is not
equipped with anti-explosive designs and
measures. Do not use this unit in the environment
with the burnable gases.
(2) There are some environmental conditions that
might cause error reading:
Pollutant places.
Large temperature variation.
Places with high wind velocity.
Organic solvent, adhesive vapor, fuel gas and
vesicant will cause abnormal response from the
sensor. Try to avoid the environment involved
with this substance.
Infrared Refrigerant Leak Detector
4
Places fill with too much to Freon gas.
(3
) Please note that a white filter must be installed into
t
he probe tip. If a white filter is missed and the
operator directly detects the leakage, the Infrared
Chamber system will be damaged. If a white filt
er
is dir
ty, please replace a new one.
Infrared Refrigerant Leak Detector
5
5. PARTS & CONTROL
5-1 Panel Description
Infrared Refrigerant Leak Detector
6
○1 USB Power ○2 Sensor Probe
3
Sensor Protector 4 LED Leak Indicators
5
Low Battery Indicator 6 Auto-Balance Indicator
7
Peak Hold Indicator 8 Mute Indicator
9
Sensitivity T/H/M/L Indicator
10
Mute Button 11 Auto Reset Button
12
Turbo Button 13 Peak Hold Button
14
Sensitivity H/M/L Button
15
Power On/Off ○16 Probe Clamp
17
Battery Cover ○18 Holster
19
White Filter
5-2 LED Leak Indicator Definition:
Low Concentration Indication