Tanita MC-780MA Instruction Manual

Please read this Instruction Manual carefully and keep it for future reference.
Temperature : 5°C – 35°C Relative Humidity : 30% – 80%
(without condensation)
Maximum Altitude : 2,000m ASL Pressure : 86kPa - 106kPa
<Storage Conditions><Usage Conditions>
BODY COMPOSITION ANALYZER
Instruction Manual
To avoid malfunctions, avoid storing the equipment in a place with direct sunlight, significant temperature changes, a risk of dampness, a large amount of dust or a risk of vibration or impact, or in the vicinity of flames.
MC-780MA
2
GB
Efficacy
Intended Use
•This equipment can be used in the screening of certain adult diseases and conditions related to body weight and composition.
•The equipment can be used in the monitoring and prevention of conditions caused by excessive deposits of fatty tissue such as diabetes, hyperlipidemia, cholelithiasis (gallstones) and fatty liver.
•The equipment can be used in the monitoring of changes in individuals‘ body composition related to the ratio of fatty tissue to lean tissue.
•The equipment can be used to assess the effectiveness of individuals‘ nutrition and exercise programmes for health and phy
sical fitness.
1.This product is simple to use, and requires no specialized facilities or expertise is
required to take measurements.
2.Measurements can be taken quickly and easily, causing minimal inconvenience to the
patient during measurement.
3
GB
Contents
Before Use
How to Use
Troubleshooting 25
Specifications 26
If Necessary
Body Composition Mode 17
Measurement Results 19
Scale Mode 21
Weight Lock Mode 22
Output and Storage of Measurement Results 23
Taking a Measurement 17
For Your Safety 4
Part Names & Connection 8
Preparation 10
Power Supply 14
Settings 15
4
GB
Before Use
(cautions for safety)
For Your Safety
Prohibited
Warning
Caution
his mark indicates actions that are prohibited.
This mark indicates instructions that must always be followed.
Warning
Do not handle the plug with wet hands.
This may result in electric shock, fire, or leakage.
Keep this equipment away from flammable gas and oxygen-rich environments.
Do not modify this equipment in any way.
This can cause electric shock or injury, or affect the accuracy of analysis.
Do not use a multi-plug adapter.
This may cause fire.
This section explains precautionary measures to be taken to avoid injury to the users of this device and others, and to prevent damage to property. Please familiarise yourself with this information to ensure safe operation of this equipment.
Failure to follow instructions highlighted with this mark could result in death or severe injury.
This equipment must not be used on subjects with pacemakers or other mechanical implants.
This equipment passes a weak electrical current through the body which could interfere with and cause the malfunction of electrical medical implants, resulting in serious harm.
Failure to follow instructions highlighted with this mark could result in injury or damage to property.
5
GB
Before Use
(cautions for safety)
Always…
Keep away from water.
Do not jump on the equipment.
Avoid using on subjects with metal allergies.
Allergic reactions may be caused by the stainless steel used in the electrodes of this device.
Do not lean on the equipment.
Be sure to use the included AC adapter.TR30M120)
Do not lean against the equipment.
Unplug the AC cable from the equipment when moving it.
Tighten the adjustable feet when moving the equipment.
Do not insert fingers into gaps or holes.
Do not use this equipment near other products that emit electromagnetic waves.
Do not apply force to the display.
The screen panel may break and cause injury.
Assist persons with disabilities.
Be sure to clean the scale platform with appropriate disinfectant after each use.
Do not place items sensitive to magnetic forces near the equipment.
Caution
Prohibited
Stand clear of the subject during measurement to avoid electric shock and ensure accuracy.
The magnet of the impedance meter may cause corruption of data on devices such as floppy disks and USB memory sticks if these are placed near the equipment.
Another person must perform measurements for persons with disabilities that impair their ability to do so alone.
Interpretation of analysis results (e.g. evaluation of measurements and formulation of exercise programmes based on results) must be performed by a professional.
Weight loss measures and exercise based on self-analysis could be detrimental to you health. Always follow the advice of a qualified professional.
This equipment is designated a Class B IT device (mainly for systems intended to be used in domestic environments) and is CE (EMC) certified, but it may affect devices that are sensitive to electromagnetic waves.
If connecting a computer or peripheral devices to this equipment, please use devices complying with IEC60601-1 (EN60601-1). Power must be supplied from a medical isolation transformer for IEC60950(EN60950) devices. Keep a distance of 1.5m between units during operation. Failure to do so may cause electric shock to subjects or malfunction.
Continually monitor both the subject and the equipment for anomalies.
If an anomaly in the subject or equipment is discovered,take appropriate action, such as stopping the equipment, while ensuring the safety of the subject.
6
GB
For Accurate Measurements
Before Use
(cautions for safety)
Avoid measuring after strenuous exercise.
Avoid measuring in multiple locations with greatly differing temperatures.
This may cause inaccurate measurements. Please take measurements after sufficient rest.
Use in a stable location.
Errors in measurements may occur if the device is used in an unstable location.
Avoid measuring after excessive food or fluid intake or when dehydrated.
Ensure that your arms are not touching your sides and that your inner thighs are not touching each other during measurement. If necessary, place a dry towel between your arm and side and/or between your thighs.
Do not take measurements while using transmitting devices, such as mobile phones, which may affect readings.
Make sure the soles of your feet are free of excess dirt, as this may block the mild electric current.
Always hold both arms straight down when taking measurements to prevent measurement errors such as undermeasurement of body fat.
Bare feet should be placed correctly on the electrode platform. Place arms straight down during measurement.
This may cause inaccurate measurements. For greater accuracy, avoid using directly after waking up. Use at the same time on each occasion, at least three hours after a meal.
Use the equipment under the same conditions and in the same position as much as possible for accurate tracking of changes.
Readings are greatly affected by the level of hydration and position of the body. Please use at the same time of day each time, under the same conditions and in the same body position.
This may cause inaccurate measurements. Allow the equipment to stand for at least 2 hours before using if it is moved to a new location with a temperature difference of 20°C or more.
7
GB
Scheduled Maintenance
Before Use
(cautions for safety)
Update settings, replace items, or call for service as necessary according to the results of the visual inspections. Do not use the unit if you see any signs of damage. Equipment that has been damaged must be checked for proper operation by qualified personnel before using again.
TANITA recommends that each facility conduct periodic checks of each unit.
1. Check the following at least daily:
• That the unit is on a stable and level surface
• Date and time settings
2. Visually inspect the following at least weekly:
• The display for any damage or contamination
• All cables, cords, and connector ends for damage or contamination
• All safety-related labeling for legibility
• All accessories (sensors, electrodes, etc.) for wear or damage
3. Visually inspect the following at least monthly:
 •Mounting screws on stand
8
GB
Before Use
(cautions for safety)
* For accurate measurement, place the machine as
level as possible.
* Rotat e the a djustab le feet in 4 po sitions for
adjustment so that the bubbles of the level gauge reach the centre.
Status when the level gauge is viewed from above
Checking the lev
Control unit Connection of Plugs Platform Connection of Plugs
el
black ring
air bubble
Part names & Connection procedure
Adjustable feet x 4
Level Gauge
Electrode x 8
Casters (x 2)
Platform
Control Unit
Grip
RS-232C
Communication cable
USB A USB B
AC adapter
SD card slot
WEEE - Waste Electrical and Electronic Equipment Directives
Date of manufacture
For indoor use only
See the instructions
Symbols and their Meanings
Positive polarity
Conformity with Medical Device Directive 93/42/EEC
Type BF applied part
- Grips and platform
Serial interface
Direct current
Class II Equipment
Input, Output
Alternating current
Caution Refer to the attached notes.
SD card
*The SD logo is a registered trademark
of the SD Association.
0123
9
GB
Before Use
(cautions for safety)
Turn ON / OFF the power
Reset zero point
Display Body Fat (percentage and mass)
*Not measured value but calculated value
*Not measured value but calculated value
*Not measured value but calculated value
Display Visceral Fat Rating
Display Body Muscle (percentage and mass)
Display Basal Metabolic Rate
Display Body Water (percentage and mass)
Printout result
Set preset value (Clothes weight)
Set various functions
Confirms the entered numerical value.
Select measurement mode
Meanings of the LED Indicators and Keys
Indicates whether "Standard mode or Athletic mode" is selected as the body type.
Indicates whether "Male or Female" is selected as the gender
Enter the age between "5 - 99 years" Enter the height between "90.0 - 249.9 cm"
Display and Keys
MC-780 III Max 270kg Min 2kg e=0.1kg T = -10kg
90.0 - 249.9cm
5 - 99
Male
Age
Height
Female
AthleticStandard
Male
Female
AthleticStandard
90.0 - 249.9cm
Height
5 - 99
Age
1
2 3
9
4 5
6
0
7 8
CE
Enter
Enter
PT
PT
Setting
Setting
Loading...
+ 19 hidden pages