System Sensor CDM100M User Manual

CD100M
Hand Held CO2 and Temperature Monitor User Manual
Introduction
The unit features a large, easy-to-read dis­play with a push-button interface which al­lows for easy calibration, quick adjustments for altitude correction, and simple toggling between °C and °F temperature readings.
1
14
13
12
9
10
8
11
7
2
4
5
6
Features
1. Lightweight- Made of ABS Plastic. This durable, lightweight plastic adds protection while weighing less than one pound.
2. Display - The large display allows for CO and Temperature readings at a glance. The display also allows for adjustments to mode settings using the push button interface.
3. Soft Touch- Attractive, gives comfort and extra durability.
4. Up/Down - Used to increase or de­crease values while in the edit mode.
5. Mode - Toggles between the menu options.
6. Power - Turns the power on and off.
7. Enter - When toggling between menu op­tions, the E nter button will select desired menu option for editing (e. g. Altitude settings). After changes have been made, the Enter key will also confirm s ettings and return to the main menu.
8. Calibration Port - Used for gas connec­tion during calibration.
9. Calibration Activation Switch ­Located in the battery compartment, this switch is used for activating the calibration process for CO
10. Voltage Output- The sensor is equipped with an output for datalogging.
11. Built-in Kicks tand- For continuous monitoring. The built-in kickstand simply swings out from the base to support the moni­tor in an upright position.
12. Power Connection - For connecting the supplied 6V power adapter.
13. Battery Compartment - 4 AA Batteries are required for portable operation.
14. Battery Cover
or Temperature.
2
Display
Listed below are the modes of operation visible on the display and accessible through the on-board buttons.
Warm-Up Mode - When the power button is pressed, the monitor enters a one minute warm-up indicated by the word WARM-UP in the upper left corner.
Normal Operating Mode - After warm-up the sensor will stabilize and display current conditions.
Elevation Settings Mode - This feature al­lows the user to correct the monitor for el­evation changes.
Calibration Settings Mode - Allows user to set the CO when performing a calibration.
Calibration in Prog Mode - Displays when calibration is in progress.
Low Battery - Displays when the power source is low.
3
Start-Up
Battery Operation - For portable use, the monitor operates on 4 AA batteries. E x­pected battery life is as follows:
Normal (Non-Alkaline)..Up to 50 hours
Rechargeable............Up to 60 Hours
Industrial Alkaline ...Up to 70 Hours
If the monitor will be used frequently, we recommend rechargeable batteries. The ini-
2
tial investment may seem costly but may quickly pay for itself.
Battery Ins tallation - Remove the battery cover (located on the back) by pressing the pressure clip (located on the bottom near the kickstand) and pull upward.
Follow the battery diagram imprinted on the plastic on the back cover for proper installa­tion. R eplace battery cover when battery installation is complete.
Low Battery - Depending on the battery, a warning signal will flash (indicated by the blinking Low Battery LCD) when there is less than 30 minutes of battery life. At this point the batteries should be replaced or the AC adapter should be used as a substitute.
If operation continues, the unit will become inoperable and only the Low Battery will be blink on the LCD display.
AC Power - The sensor is shipped with a 6V DC 500mA AC/DC adapter.
When possible, we suggest using the power adapter to guarantee continuous power during long term monitoring or datalogging operation.
To use the AC adapter, connect the plug into the back of the unit and plug the trans­former into any standard wall outlet.
Use the supplied adapter. Using
the wrong adapter may cause
concentration or Temperature
2
NOTE
damage to the unit.
Operation
Power-Up
1. Press the Power Button.
2. A 2 second delay will occur before the display becomes visible.
3. 10 seconds will elapse before current CO2 readings are displayed.
4. “Warm-up” will display for approxi­mately one minute. During this one minute warm-up, adjustments can not be made to the sensor.
Elevation Correction - The sensor is shipped with the elevation setting set at “zero” or sea level. The sensor, like any other gas measuring device is directly af­fected by altitude changes. If you are at an altitude greater than 500 feet, an adjustment should be made to assure sensor accuracy. To change the default setting using Elevation Correction follow the steps below.
1. While in Normal Mode press the “Mode” button once. The Elevation LCD will be­gin blinking.
2. Press Enter.
3. Pres s mode to toggle the elevation reading between feet (ft) and meters (m)
4. Use the Up/Down button to adjus t the alti­tude. The altitude can be adjusted in incre­ments of 500 (feet) or 100 (meters). Once the correct altitude is set, pres s Enter to lock the setting and return to normal mode.
Display in Fahrenheit or Celsius - To change the temperature to read in Celsius or Fahrenheit, or to turn the temperature dis­play off, use the Up or the Down button.
Featuring
TE C H N O L O G Y
Stand-alone Monitoring - Once the batter­ies have been installed and the Elevation correction has been made (as described in the steps above), the sensor will begin to accurately display current room conditions.
Kickstand - The sensor is equipped with a kickstand located on the base of the unit. For desktop monitoring, pull the kickstand out to prop the unit on an angle.
Monitoring using an External Datalogger - Voltage outputs for both CO and temperature are available via an RJ-45 jack on the rear of the unit.
2
If power is lost during operation, battery op­eration will not function as a back up.
Calibration (CO2)
1. The calibration procedure will last ap­proximately 5 minutes. Before performing the calibration procedure, remove the bat­tery cover (see “Battery Installation”) to provide access to the CO2 calibration ac­tivation switch. Connect the supplied AC adapter to the back of the sensor. If you do not have the AC adapter, new batter­ies should be used.
2. Power up the sensor and wait for the Warm-up to end.
3. Next verify the Elevation correction has been set. R efer to the steps in Elevation Correction for procedure.
4. Once the Elevation is verified, press the Mode button twice. The Calibration mode will begin blinking.
5. Press Enter.
6. Use the Up/Down Rocker button to adjust the lower reading on the display to the current ambient conditions. Pressing the button once will change the readings in increments of 10 ppm. To increase the speed, press and hold the button.
For best accuracy, a reference or
known concentration of CO
should be used when adjusting
7. Next, on the backside of the unit locate the push button switch (under the battery cover, in the small round hole to the right of the connector jack), use a small pointed object to depress and hold the switch for 5 seconds. The Calibration light will begin to blink.
8. Press Enter.
9. Calibration In Progress will begin to blink. At this point the unit will program it­self based on the CO2 value that was in­put in Step 6. The calibration process will take approximately 5 minutes.
When Calibration is complete, the display will return to the steady Calibration mode. Press Enter to return to the nor­mal operation mode.
NOTE
2
the reading.
T emperature Adjustment
Use this procedure to adjust the temperature output when, for example, you wish to have the temperature output match a reference sen­sor. The accuracy of a field adjustment is de­pendent upon the stability of the environment in which the procedure is performed, and upon the accuracy of the reference sensor.
1. Before performing the temperature adjust­ment, connect the supplied AC adapter to the back of the sensor. If you do not have the AC adapter, new batteries should be used.
2. Power up the sensor and, using the kickstand, place it on a flat surface in a stable environment relatively free of drafts or temperature changes. Wait 30 minutes for the unit to fully equilibrate with the environment. Do not hold the unit in your hand during this period.
3. Press the Mode button three times. The word “TEMPERATURE” at the bottom of the display will begin blinking.
4. Press Enter. Both the word “TEM­PER ATURE” and the numeric tempera­ture display will begin blinking in unison.
5. Use the Up/Down Rocker button to ad­just the temperature reading to match the reference.
6. Press Enter. The temperature offset is immediately adjusted, the blinking stops, and the unit is now in normal operating mode.
Specifications
Method
Dual Beam Absorption Infrared™
Sample Method
Diffusion or flow through (50 - 100 ml/min)
Warranty
18 months parts and labor
Performance
CO2 Channel Measurement Range
0-4,000 ppm voltage output 0-10,000 ppm display
Sensitivity
± 1 ppm
Accuracy
±50 ppm or ±5% of reading, whichever is greater
Repeatability
±20 ppm
Temperature Dependence
±0.1% of reading per °C or ±2 ppm per °C, whichever is greater, referenced to 25°C
Pressure Dependence:
0.13% of reading per mm Hg (Corrected via user input for elevation)
Annual Drift
± 20 ppm typical
Response Time
<60 seconds for 0..90% of step change
Warm-Up Time
<60 seconds at 22°C
Operating Conditions
32-122°F (0-50°C) 0-95% RH, non-condensing
Storage Temperatures
–40 to 140°F (-40 to 60°C)
Calibration Interval
12 months, offset adjustment using single gas at 0-1000 ppm CO2. Full factory calibration available
Temperature Channel
Temperature Range
Voltage output 32 to 104°F (0 to 40°C) Display 32 to 122°F (0 to 50°C)
Calibration Interval
12 months, offset adjustment using temperature standard at 50 to 86°F (10 to 30°C). Full factory calibration available
Display Resolution
0.1°F (0.1°C)
Display Options
°F, °C, or Off. Set with panel button.
Accuracy
±2°F (±1°C)
Response Time
20-30 minutes (case must equilibrate with environment)
Output - Analog
CO2
0-4 VDC, 1mV/ppm (4,000 ppm max)
Temperature
0-4 VDC linear, 32-104°F (0-40°C)
Output Impedance
100 Ohms
Wiring Connection
Via RJ-45 to DB9 serial port cable
Display
LCD with independent CO2/ temperature readings (panel buttons set elevation, °F/°C, calibration functions)
Power Supply
Battery Type
Four AA batteries, not included
Battery Operation
80 hours (alkaline)
External
6 VDC from external AC/DC adapter, included
Power Requirements
100 mA Peak, 20 mA average from 6V
Certification
FCC Class 15 Part B
Note
Specifications subject to change without notice. Covered by United States Patents: 5,060,508 and 5,163,332
Warranty
System Sensor warrants its enclosed sensor to be free from defects in materials and workmanship under normal use and service for a period of 18 months from date of purchase by the original owner. System Sensor makes no other express warranty for this sensor. No agent, representative, dealer, or employee of the Company has the authority to increase or alter the obligations or limitations of this Warranty. The Companyís obligation of this Warranty shall be limited to the repair or replacement of any part of the sensor which is found to be defective in materials or workmanship under normal use and service during the 18 month period commencing with the date of purchase by the original owner. After phoning System Sensorís toll free number 800-SENSOR2 (736-7672) for a Return Authorization number, send defective units postage prepaid to: System Sensor, Warranty Service, RA#_______, 3825 Ohio Avenue, St. Charles, IL 60174. Please include a note describing the malfunction and suspected cause of failure. The company shall not be obligated to repair or replace units which are found to be defective because of damage, unreasonable use, modifications, or alterations occurring after the date of manufacture. In no case shall the Company be liable for any consequential or incidental damages for breach of this or any other Warranty, expressed or implied whatsoever, even if the loss or damage is caused by the limitation of incidental or consequential damages, so the above limitation or exclusion may not apply to you. This Warranty gives you specific legal rights, and you may also have other rights, and you may also have other rights which may vary from state to state.
P/N 62507B 10599
System Sensor
3825 Ohio Avenue St. Charles, IL 60174 1-800-SENSOR2 (736-7672) Fax: 630-377-6495
Featuring
Absorption Infrared™ is a trademark of Engelhard Sensor Technologies
Telaire®is a registered trademark of Engelhard Sensor Technologies
Loading...