Symmons 4-10A Service Manual

Page 1
Mechanical
4-10C
Same as 4-10A except with 1/2” IPS tempered outlet connection
Mixing Valve
4-10 Series
INSTALLATION, OPERATION AND SERVICE INSTRUC­TIONS
4-10A
The 4-10A Mechanical Mixing Valve is furnished with all ferrule fittings attached. To install, remove the three (3) nuts and ferrules from the unit and install 3/8” copper tubing into compression fitting nipple. Flexible braided stainless steel hose will also fasten to the 9/16 – 24 thread of the compression nipple once the nuts and ferrules are removed. Slide the nut and then the ferrule over the copper tubing in preparation for tighten­ing. Torque the nut onto the compression nipple so as to compress the ferrule around the tubing. The nut should be tightened so as to prevent any leaks by the ferrule. HOT water is supplied to the left side of the valve (when looking at the adjustment handle) and COLD water is supplied to the right side.
4-10B
The 4-10B is installed similar to the 4-10A valve, except a 3/8” compres­sion tee is supplied to deliver COLD water both to the mixing valve and to the cold side of a fixture. The ferrule fittings of the compression tee are secured exactly like those of the mixing valve.
Temperature Selection
The temperature of 4-10 Mechanical Mixing Valve is adjusted by inserting a 1/4” hex key (not supplied) into the temperature adjustment slot in mix­ing valve and turning the key either clockwise or counter-clockwise. Turn­ing the handle clockwise decreases the output temperature of the unit and turning the handle counterclockwise increases output temperature.
System Cleaning
The integral checks installed in the 4-10 mixing valve should be kept clean to insure proper operation of the unit. Unwanted debris may cause the checks to remain open and allow a cross connection of hot and cold water. To clean checks, shut off water supply to the unit. Remove mixing valve from piping assembly and prepare for check removal. Unthread the threaded nipple from the body using a 5/16” HEX key. Dislodge the rub­ber washer from the threaded hole using a small flat blade screwdriver. In­sert the screw driver or needle nose pliers into the small hole in the plastic insert and gently remove. The duckbill check will remove with insert. Clean checks or replace with MV-107 repair kit as needed. Reverse instructions to re-install checks into main body.
Compliance:
- ASME A112.18.1/CSA B125.1
- NSF 372 (CA 116875)
- NSF 61.9
4-10A Mechanical Mixing Valve with 3/8” compression connections and (2) integral checks (piping not supplied)
4-10B Mechanical Mixing Valve with 3/8” compression connec­tions, (2) integral checks and 3/8” compression tee (piping not supplied).
4-10C Mechanical Mixing Valve with (2) 3/8” compression connec­tions and (1) 1/2” IPS connection.
Note 1/4” Hex Key (not supplied) used to adjust temperature
Symmons Single Lever Faucet
4-10B
3
/8” compression
connections
Typical
(faucet not included)
4-10C
(2) 3/8” compression connections
3
/8” compression tee
Symmons Metering Faucet
Typical
(faucet not included)
1
/2
IPS connection
Symmons Industries, Inc. 31 Brooks Drive Braintree, MA 02184 (800) 796-6667, (781) 848-2250 Website: www.symmons.com
©2013 Symmons Industries, Inc. Printed in U.S.A.
Fax (800) 961-962 1, ( 7 8 1) 843-3849
Email: customerse rvice@sy m mo n s .com
ZV-353 111813
Loading...