Stryker 5212-309-001A Secure II MedSurg Bed Model 3002 Operations Manual

Page 1
iiBBeed® SSeerrvveerr vv22..55
IInnssttaallllaattiioonn//CCoonnffiigguurraattiioonn MMaannuuaall
5212
Connected Hospital®
5212-309-001 Rev A.1
EN
2019/11
Page 2
Page 3
SSyymmbboollss
~
Caution
Manufacturer
Direct current
Alternating current
5212-309-001 Rev A.1 EN
Page 4
Page 5
TTaabbllee ooff CCoonntteennttss
Warning/Caution/Note Definition ..............................................................................................................................2
Summary of safety precautions ................................................................................................................................3
Introduction ...............................................................................................................................................................4
Indications for use...................................................................................................................................................4
i
Bed Server software ...............................................................................................................................................4
System requirements and recommendations.............................................................................................................4
Device connection requirements ..............................................................................................................................6
Contact information.................................................................................................................................................6
Installation.................................................................................................................................................................7
Server configuration ................................................................................................................................................7
Windows Server 2012.........................................................................................................................................7
Windows Server 2016.......................................................................................................................................17
i
Bed Server application.......................................................................................................................................... 29
i
Bed Server Tools.............................................................................................................................................36
i
Bed Wireless Configuration Tool ...........................................................................................................................37
Editing Windows configuration ...............................................................................................................................38
Verify
i
Bed Server ................................................................................................................................................. 39
Setup....................................................................................................................................................................... 41
Adding devices (clients) to the Master Device List....................................................................................................41
Adding
Adding an additional Stryker interface..................................................................................................................... 44
Adding a third-party interface ................................................................................................................................. 50
Configuring the wireless router (Stryker device configuration) .............................................................................. 51
Configuring wireless network connection settings ................................................................................................. 53
Resetting the wireless module to factory default settings (Med-Surg bed)............................................................ 54
Resetting the wireless module to factory default settings (Model FL27 IInnTToouucchh) ................................................55
Troubleshooting ...................................................................................................................................................... 56
Basic.................................................................................................................................................................... 56
Advanced.............................................................................................................................................................59
Smart Equipment Management (option) troubleshooting ..........................................................................................65
Email alerts ..........................................................................................................................................................67
i
Bed Locator IDs and hospital locations ....................................................................................................... 43
Smart Equipment Management (SEM) ............................................................................................................... 44
Integrating RRaauullaanndd RReessppoonnddeerr®® 5 .................................................................................................................50
Connectivity issues (total device counts) ............................................................................................................ 59
Third party communication issues......................................................................................................................60
Device not connecting to server.........................................................................................................................61
Connectivity issues to one / multiple devices ......................................................................................................61
Cannot communicate with device (verify ports are open) ..................................................................................... 63
Cannot communicate with device (DNS only) ..................................................................................................... 64
5212-309-001 Rev A.1 1 EN
Page 6
WWaarrnniinngg//CCaauuttiioonn//NNoottee DDeeffiinniittiioonn
The words WWAARRNNIINNGG, CCAAUUTTIIOONN, and NNOOTTEE carry special meanings and should be carefully reviewed.
WWAARRNNIINNGG
Alerts the reader about a situation which, if not avoided, could result in death or serious injury. It may also describe potential serious adverse reactions and safety hazards.
CCAAUUTTIIOONN
Alerts the reader of a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient or damage to the product or other property. This includes special care necessary for the safe and effective use of the device and the care necessary to avoid damage to a device that may occur as a result of use or misuse.
NNoottee -- Provides special information to make maintenance easier or important instructions clearer.
EN 2 5212-309-001 Rev A.1
Page 7
SSuummmmaarryy ooff ssaaffeettyy pprreeccaauuttiioonnss
Always read and strictly follow the warnings and cautions listed on this page. Service only by qualified personnel.
CCAAUUTTIIOONN
• Before proceeding with this installation, make sure that a previous version of the installed on the target system. If a previous version was installed, uninstall the software. If you attempt to install the application on a system where a previous version was installed, the installer behavior may be unpredictable.
• Before proceeding with this installation, make sure that a previous version of the application is not currently installed on the target system. If a previous version was installed, uninstall the software. If you attempt to install the application on a system where a previous version was installed, the installer behavior may be unpredictable.
ii
BBeedd Server application is not currently
ii
BBeedd Wireless configuration tool
5212-309-001 Rev A.1 3 EN
Page 8
IInnttrroodduuccttiioonn
This manual assists you with the operation or maintenance of your Stryker product. Read this manual before operating or maintaining this product. Set methods and procedures to educate and train your staff on the safe operation or maintenance of this product.
NNoottee -- Stryker continually seeks advancements in product design and quality. This manual contains the most current product information available at the time of printing. There may be minor discrepancies between your product and this manual. If you have any questions, contact Stryker Customer Service or Technical Support at 1-800-327-0770.
IInnddiiccaattiioonnss ffoorr uussee
The intended use for theiiBBeedd® Wireless (withiiBBeedd Awareness) is to assist clinical staff to monitor bed parameters on specific Stryker beds. The desired bed parameters are set by operators at the bedside. The intended for use with specifically enabled Stryker beds that have been verified and validated with the software, and is not intended to provide bed status information for non-Stryker beds. The intended to communicate any patient status information, nor to permanently store any type of data. The
ii
BBeedd Awareness System is not intended to provide automated treatment decisions or as a substitute for professional healthcare judgment. The monitoring or alert equipment. All patient medical diagnosis and treatment are to be performed under direct supervision and oversight of an appropriate healthcare professional.
ii
BBeedd SSeerrvveerr ssooffttwwaarree
ii
BBeedd Wireless withiiBBeedd Awareness System is not a replacement or substitute for vital signs
ii
BBeedd Wireless software is only
ii
BBeedd Wireless
ii
BBeedd Wireless software is not
ii
BBeedd Wireless with
The includediiBBeedd Server installation CD (5212-501-001) contains the software required for theiiBBeedd Server installation.
NNoottee -- Before uninstalling the application, copy the following files to the PPuubblliicc DDooccuummeennttss folder. The files will import to the new application. Using the
Install directory\Stryker\iBedServerApplication\Data
BBIDList.xml DeviceURLs.xml
DeviceBBIDLocationAssociation.xml HospitalLocationList.xml
SSttrryykkeerr ddiissccllaaiimmss aallll rreessppoonnssiibbiilliittyy ffoorr iinnffoorrmmaattiioonn ttrraannssmmiitttteedd ooffff ooff iittss ddeevviicceess..
ii
BBeedd Server Tool, configure the Master Device List and the Locator Associations.
SSyysstteemm rreeqquuiirreemmeennttss aanndd rreeccoommmmeennddaattiioonnss
NNoottee
• If minimum system requirements are not met, system performance will be impacted.
• Apply relevant software patches annually.
HHaarrddwwaarree::
Minimum requirements for the
11 -- 330000 ccoonnnneecctteedd bbeeddss::
2.x GHz processor or higher with a total of 4 cores
Memory: 8 GB RAM
Hard Drive: 150 GB
ii
BBeedd Server hardware is dependent on the number of beds connected to the system.
330011 -- 660000 ccoonnnneecctteedd bbeeddss::
2.x GHz processor or higher with a total of 8 cores
Memory: 16 GB RAM
EN 4 5212-309-001 Rev A.1
Page 9
Hard Drive: 150 GB
660011 -- 880000 ccoonnnneecctteedd bbeeddss::
2.x GHz processor or higher with a total of 16 cores
Memory: 32 GB RAM
Hard Drive: 150 GB
880011 -- 11,,000000 ccoonnnneecctteedd bbeeddss::
2.x GHz processor or higher with a total of 24 cores
Memory: 32 GB RAM
Hard Drive: 150 GB
11,,000011 -- 11,,330000 ccoonnnneecctteedd bbeeddss::
2.x GHz processor or higher with a total of 32 cores
Memory: 64 GB RAM
Hard Drive: 150 GB
NNoottee
• For systems that have over 1,300 connected beds, add a core for every 50 additional beds.
• Two server environments are recommended for the
• The
ii
BBeedd Wireless System is supported in either physical or virtual environments.
ii
BBeedd Wireless System: TEST and PROD
SSooffttwwaarree::
WWiinnddoowwss SSeerrvveerr 22001122 RR22 // 22001166
Add Roles
Web Server (IIS) (Installed)
Roles Services
– Application development
ASP.NET 3.5 (Installed)
ASP.NET 4.5 (or higher) (Installed)
ASP (Installed)
– Management tools
IIS Management Console (Installed)
Features
.NET Framework 3.5 features (Installed)
.NET Framework 4.5 (or higher) features (Installed)
Telnet Client
– WCF Services
HTTP Activation
All current Microsoft High Priority Updates (Installed) and optional update for .NET Framework 4.5 (or higher)
Additional configuration or setup may be required depending on equipment and other variables. If you have difficulties during installation, setup, configuration, or while attempting to establish a connection between the wireless clients, contact Stryker Technical Support at 1-800-327-0770.
5212-309-001 Rev A.1 5 EN
ii
BBeedd Server and Stryker
Page 10
DDeevviiccee ccoonnnneeccttiioonn rreeqquuiirreemmeennttss
NNoottee -- You are required to use the DNS naming convention for each Stryker device if the devices can travel to multiple sub­nets.
DHCP connections using a reserved IP address for each device via its MAC address
Static connections using static IP address for each device via its MAC address
DNS naming convention using each device host name that is hard-coded to the device (Host name example = SYK-
82453f21f0c2 [SYK device MAC address])
CCoonnttaacctt iinnffoorrmmaattiioonn
Contact Stryker Customer Service or Technical Support at: 1-800-327-0770.
Stryker Medical 3800 E. Centre Avenue
Portage, MI 49002 USA
E-mail: medicaliBedWirelessSupport@stryker.com
EN 6 5212-309-001 Rev A.1
Page 11
IInnssttaallllaattiioonn
SSeerrvveerr ccoonnffiigguurraattiioonn
WWiinnddoowwss SSeerrvveerr 22001122
1. In the SSeerrvveerr MMaannaaggeerr navigate to the Dashboard.
2. Click on the AAdddd rroolleess aanndd ffeeaattuurreess link (Figure 1).
FFiigguurree 11 –– AAdddd rroolleess aanndd ffeeaattuurreess
3. Click the NNeexxtt button in the AAdddd RRoolleess aanndd FFeeaattuurreess WWiizzaarrdd (Figure 2).
FFiigguurree 22 –– AAdddd RRoolleess aanndd FFeeaattuurreess WWiizzaarrdd
5212-309-001 Rev A.1 7 EN
Page 12
4. In the IInnssttaallllaattiioonn TTyyppee step, select the RRoollee--bbaasseedd oorr ffeeaattuurree--bbaasseedd iinnssttaallllaattiioonn if not already selected and click
NNeexxtt (Figure 3).
FFiigguurree 33 –– IInnssttaallllaattiioonn TTyyppee
5. In the SSeerrvveerr SSeelleeccttiioonn step, click SSeelleecctt aa sseerrvveerr ffrroomm tthhee sseerrvveerr ppooooll and make sure that the server is correct in
the SSeerrvveerr PPooooll box and click NNeexxtt (Figure 4).
FFiigguurree 44 –– SSeerrvveerr SSeelleeccttiioonn
EN 8 5212-309-001 Rev A.1
Page 13
6. In the SSeerrvveerr RRoolleess step, scroll through the options in the RRoolleess box and select WWeebb SSeerrvveerr ((IIIISS)) (Figure 5).
FFiigguurree 55 –– WWeebb SSeerrvveerr ((IIIISS))
7. In the pop-up, AAdddd ffeeaattuurreess tthhaatt aarree rreeqquuiirreedd ffoorr WWeebb SSeerrvveerr ((IIIISS)), click the AAdddd FFeeaattuurreess button (Figure 6).
FFiigguurree 66 –– AAdddd FFeeaattuurreess
5212-309-001 Rev A.1 9 EN
Page 14
8. In the SSeerrvveerr RRoolleess step, click the NNeexxtt button (Figure 7).
FFiigguurree 77 –– SSeerrvveerr ccoonnffiirrmmaattiioonn
9. In the FFeeaattuurreess step, select ..NNEETT FFrraammeewwoorrkk 33..55 FFeeaattuurreess, ..NNEETT FFrraammeewwoorrkk 44..55 FFeeaattuurreess, and TTeellnneett CClliieenntt in
the FFeeaattuurreess box (Figure 8).
FFiigguurree 88 –– FFeeaattuurreess sseelleeccttiioonn
EN 10 5212-309-001 Rev A.1
Page 15
10.In the WWeebb SSeerrvveerr RRoollee ((IIIISS)) step, click the NNeexxtt button (Figure 9).
FFiigguurree 99 –– WWeebb SSeerrvveerr RRoollee ((IIIISS))
11.In the RRoollee SSeerrvviicceess step, click the NNeexxtt button (Figure 10).
FFiigguurree 1100 –– RRoollee SSeerrvviicceess
5212-309-001 Rev A.1 11 EN
Page 16
12.In the CCoonnffiirrmmaattiioonn step, click the IInnssttaallll button to start the installation of the role and features (Figure 11).
FFiigguurree 1111 –– IInnssttaallll ccoonnffiirrmmaattiioonn
13.When the installation is finished, click the CClloossee button (Figure 12).
FFiigguurree 1122 –– CCoommpplleetteedd iinnssttaallllaattiioonn
EN 12 5212-309-001 Rev A.1
Page 17
14.Click on the AAdddd rroolleess aanndd ffeeaattuurreess link (Figure 13).
FFiigguurree 1133 –– AAdddd rroolleess aanndd ffeeaattuurreess
15.Click the NNeexxtt button in the AAdddd RRoolleess aanndd FFeeaattuurreess WWiizzaarrdd (Figure 14).
FFiigguurree 1144 –– AAdddd RRoolleess aanndd FFeeaattuurreess WWiizzaarrdd
5212-309-001 Rev A.1 13 EN
Page 18
16.In the IInnssttaallllaattiioonn TTyyppee step, select the RRoollee--bbaasseedd oorr ffeeaattuurree--bbaasseedd iinnssttaallllaattiioonn and click NNeexxtt (Figure 15).
FFiigguurree 1155 –– IInnssttaallllaattiioonn TTyyppee
17.In the SSeerrvveerr SSeelleeccttiioonn step, click SSeelleecctt aa sseerrvveerr ffrroomm tthhee sseerrvveerr ppooooll and verify that the server is correct in the
SSeerrvveerr PPooooll box and click NNeexxtt (Figure 16).
FFiigguurree 1166 –– SSeerrvveerr SSeelleeccttiioonn
EN 14 5212-309-001 Rev A.1
Page 19
18.In the SSeerrvveerr RRoolleess step in the RRoolleess box, expand the WWeebb SSeerrvveerr ((IIIISS)) heading, WWeebb SSeerrvveerr heading, and then
AApppplliiccaattiioonn DDeevveellooppmmeenntt. Select AASSPP..NNEETT 33..55 and AASSPP..NNEETT 44..55 and click NNeexxtt (Figure 17).
FFiigguurree 1177 –– SSeerrvveerr RRoolleess
19.In the pop-up window, click AAdddd FFeeaattuurreess (Figure 18).
FFiigguurree 1188 –– AAdddd FFeeaattuurreess
5212-309-001 Rev A.1 15 EN
Page 20
20.In the SSeerrvveerr RRoolleess step, select AASSPP and click NNeexxtt (Figure 19).
FFiigguurree 1199 –– AASSPP
21.In the FFeeaattuurreess step, click NNeexxtt (Figure 20).
FFiigguurree 2200 –– FFeeaattuurreess sseelleeccttiioonn
EN 16 5212-309-001 Rev A.1
Page 21
22.In the CCoonnffiirrmm iinnssttaallllaattiioonn sseelleeccttiioonnss step, click IInnssttaallll (Figure 21).
FFiigguurree 2211 –– CCoonnffiirrmm iinnssttaallllaattiioonn
23.When installation is complete, click the CClloossee button (Figure 22).
FFiigguurree 2222 –– IInnssttaallllaattiioonn ccoommpplleettee
24.Restart the server.
25.Run WWiinnddoowwss UUppddaattee to look for any important and optional updates and install them. Restart the server if required.
WWiinnddoowwss SSeerrvveerr 22001166
1. In the SSeerrvveerr MMaannaaggeerr navigate to the Dashboard.
5212-309-001 Rev A.1 17 EN
Page 22
2. Click on the AAdddd rroolleess aanndd ffeeaattuurreess link (Figure 23).
FFiigguurree 2233 –– AAdddd rroolleess aanndd ffeeaattuurreess
3. Click the NNeexxtt button in the AAdddd RRoolleess aanndd FFeeaattuurreess WWiizzaarrdd (Figure 24).
FFiigguurree 2244 –– AAdddd RRoolleess aanndd FFeeaattuurreess WWiizzaarrdd
EN 18 5212-309-001 Rev A.1
Page 23
4. In the IInnssttaallllaattiioonn TTyyppee step, select the RRoollee--bbaasseedd oorr ffeeaattuurree--bbaasseedd iinnssttaallllaattiioonn if not already selected and click
NNeexxtt (Figure 25).
FFiigguurree 2255 –– IInnssttaallllaattiioonn TTyyppee
5. In the SSeerrvveerr SSeelleeccttiioonn step, click SSeelleecctt aa sseerrvveerr ffrroomm tthhee sseerrvveerr ppooooll and make sure that the server is correct in
the SSeerrvveerr PPooooll box and click NNeexxtt (Figure 26).
FFiigguurree 2266 –– SSeerrvveerr SSeelleeccttiioonn
5212-309-001 Rev A.1 19 EN
Page 24
6. In the SSeerrvveerr RRoolleess step, scroll through the options in the RRoolleess box and select WWeebb SSeerrvveerr ((IIIISS)) (Figure 27).
FFiigguurree 2277 –– WWeebb SSeerrvveerr ((IIIISS))
7. In the pop-up, AAdddd ffeeaattuurreess tthhaatt aarree rreeqquuiirreedd ffoorr WWeebb SSeerrvveerr ((IIIISS)), click the AAdddd FFeeaattuurreess button (Figure 28).
FFiigguurree 2288 –– AAdddd FFeeaattuurreess
EN 20 5212-309-001 Rev A.1
Page 25
8. In the SSeerrvveerr RRoolleess step, click the NNeexxtt button (Figure 29).
FFiigguurree 2299 –– SSeerrvveerr ccoonnffiirrmmaattiioonn
9. In the FFeeaattuurreess step, select ..NNEETT FFrraammeewwoorrkk 33..55 FFeeaattuurreess, ..NNEETT FFrraammeewwoorrkk 44..55 FFeeaattuurreess, and TTeellnneett CClliieenntt in
the FFeeaattuurreess box (Figure 30).
FFiigguurree 3300 –– FFeeaattuurreess sseelleeccttiioonn
5212-309-001 Rev A.1 21 EN
Page 26
10.In the WWeebb SSeerrvveerr RRoollee ((IIIISS)) step, click the NNeexxtt button (Figure 31).
FFiigguurree 3311 –– WWeebb SSeerrvveerr RRoollee ((IIIISS))
11.In the RRoollee SSeerrvviicceess step, click the NNeexxtt button (Figure 32).
FFiigguurree 3322 –– RRoollee SSeerrvviicceess
EN 22 5212-309-001 Rev A.1
Page 27
12.In the CCoonnffiirrmmaattiioonn step, click the IInnssttaallll button to start the installation of the role and features (Figure 33).
FFiigguurree 3333 –– IInnssttaallll ccoonnffiirrmmaattiioonn
13.When the installation is finished, click the CClloossee button (Figure 34).
FFiigguurree 3344 –– CCoommpplleetteedd iinnssttaallllaattiioonn
5212-309-001 Rev A.1 23 EN
Page 28
14.Click on the AAdddd rroolleess aanndd ffeeaattuurreess link (Figure 35).
FFiigguurree 3355 –– AAdddd rroolleess aanndd ffeeaattuurreess
15.Click the NNeexxtt button in the AAdddd RRoolleess aanndd FFeeaattuurreess WWiizzaarrdd (Figure 36).
FFiigguurree 3366 –– AAdddd RRoolleess aanndd FFeeaattuurreess WWiizzaarrdd
EN 24 5212-309-001 Rev A.1
Page 29
16.In the IInnssttaallllaattiioonn TTyyppee step, select the RRoollee--bbaasseedd oorr ffeeaattuurree--bbaasseedd iinnssttaallllaattiioonn and click NNeexxtt (Figure 37).
FFiigguurree 3377 –– IInnssttaallllaattiioonn TTyyppee
17.In the SSeerrvveerr SSeelleeccttiioonn step, click SSeelleecctt aa sseerrvveerr ffrroomm tthhee sseerrvveerr ppooooll and verify that the server is correct in the
SSeerrvveerr PPooooll box and click NNeexxtt (Figure 38).
FFiigguurree 3388 –– SSeerrvveerr SSeelleeccttiioonn
5212-309-001 Rev A.1 25 EN
Page 30
18.In the SSeerrvveerr RRoolleess step in the RRoolleess box, expand the WWeebb SSeerrvveerr ((IIIISS)) heading, WWeebb SSeerrvveerr heading, and then
AApppplliiccaattiioonn DDeevveellooppmmeenntt. Select AASSPP..NNEETT 33..55 and AASSPP..NNEETT 44..55 and click NNeexxtt (Figure 39).
FFiigguurree 3399 –– SSeerrvveerr RRoolleess
19.In the pop-up window, click AAdddd FFeeaattuurreess (Figure 40).
FFiigguurree 4400 –– AAdddd FFeeaattuurreess
EN 26 5212-309-001 Rev A.1
Page 31
20.In the SSeerrvveerr RRoolleess step, select AASSPP and click NNeexxtt (Figure 41).
FFiigguurree 4411 –– AASSPP
21.In the FFeeaattuurreess step, click NNeexxtt (Figure 42).
FFiigguurree 4422 –– FFeeaattuurreess sseelleeccttiioonn
5212-309-001 Rev A.1 27 EN
Page 32
22.In the CCoonnffiirrmm iinnssttaallllaattiioonn sseelleeccttiioonnss step, click IInnssttaallll (Figure 43).
FFiigguurree 4433 –– CCoonnffiirrmm iinnssttaallllaattiioonn
23.When installation is complete, click the CClloossee button (Figure 44).
FFiigguurree 4444 –– IInnssttaallllaattiioonn ccoommpplleettee
24.Restart the server.
25.Run WWiinnddoowwss UUppddaattee to look for any important and optional updates and install them. Restart the server if required.
EN 28 5212-309-001 Rev A.1
Page 33
ii
BBeedd SSeerrvveerr aapppplliiccaattiioonn
CCAAUUTTIIOONN -- Before proceeding with this installation, make sure that a previous version of theiiBBeedd Server application is not currently installed on the target system. If a previous version was installed, uninstall the software. If you attempt to install the application on a system where a previous version was installed, the installer behavior may be unpredictable.
1. Begin the
and select RRuunn aass AAddmmiinniissttrraattoorr.
2. If Microsoft SQL Server 2012 Express is not already installed, the IInnssttaallllSShhiieelldd WWiizzaarrdd rreeqquuiirreemmeennttss window will
open. Click the IInnssttaallll button (Figure 45).
NNoottee -- If Microsoft SQL Server 2012 Express is already installed, go to step 11.
ii
BBeedd Server Application installation by right clicking the 55221122--550022--000011 SSeerrvveerr AApppplliiccaattiioonn SSeettuupp..eexxee file
FFiigguurree 4455 –– MMiiccrroossoofftt SSQQLL sseerrvviiccee
3. Click the YYeess button in the confirmation pop-up to start the SQL Server installation (Figure 46).
FFiigguurree 4466 –– SSQQLL ccoonnffiirrmmaattiioonn
5212-309-001 Rev A.1 29 EN
Page 34
4. Select II aacccceepptt tthhee lliicceennssee tteerrmmss bbooxx and then click the NNeexxtt button (Figure 47).
FFiigguurree 4477 –– SSQQLL lliicceennssee
5. In the FFeeaattuurree SSeelleeccttiioonn step, leave the FFeeaattuurreess selection at the defaults. If the SShhaarreedd ffeeaattuurree ddiirreeccttoorryy does not
default to CC::\\PPrrooggrraamm FFiilleess, browse or create the new destination location for the installation and then click NNeexxtt (Figure 48).
FFiigguurree 4488 –– SSQQLL ffeeaattuurreess
EN 30 5212-309-001 Rev A.1
Page 35
6. In the IInnssttaannccee CCoonnffiigguurraattiioonn step, click NNeexxtt (Figure 49).
FFiigguurree 4499 –– IInnssttaannccee CCoonnffiigguurraattiioonn
7. In the SSeerrvveerr CCoonnffiigguurraattiioonn step, click NNeexxtt (Figure 50).
FFiigguurree 5500 –– SSeerrvveerr CCoonnffiigguurraattiioonn
5212-309-001 Rev A.1 31 EN
Page 36
8. In the DDaattaabbaassee EEnnggiinnee CCoonnffiigguurraattiioonn step, click NNeexxtt (Figure 51).
FFiigguurree 5511 –– DDaattaabbaassee EEnnggiinnee CCoonnffiigguurraattiioonn
9. In the EErrrroorr RReeppoorrttiinngg step, click NNeexxtt to start the SQL Server 2012 install process (Figure 52).
FFiigguurree 5522 –– EErrrroorr RReeppoorrttiinngg
EN 32 5212-309-001 Rev A.1
Page 37
10.When installation is complete, click CClloossee in the CCoommpplleettee window which will start theiiBBeedd Server Application
installation (Figure 53).
FFiigguurree 5533 –– SSQQLL sseerrvveerr iinnssttaallll ccoommpplleettee
11.In the IInnssttaallllSShhiieelldd WWiizzaarrdd window, click NNeexxtt (Figure 54).
FFiigguurree 5544 –– IInnssttaallllSShhiieelldd WWiizzaarrdd
5212-309-001 Rev A.1 33 EN
Page 38
12.In theiiBBeedd SSeerrvveerr AApppplliiccaattiioonn system configuration window, enter the sites information and click NNeexxtt (Figure 55).
FFiigguurree 5555 –– AApppplliiccaattiioonn ccoonnffiigguurraattiioonn
13.The CChhoooossee DDeessttiinnaattiioonn LLooccaattiioonn screen will appear (Figure 56).
a. If using the default location, click NNeexxtt.
b. If using a different location, click CChhaannggee and then NNeexxtt to confirm.
FFiigguurree 5566 –– CChhoooossee DDeessttiinnaattiioonn LLooccaattiioonn
EN 34 5212-309-001 Rev A.1
Page 39
14.The SSeelleecctt PPrrooggrraamm FFoollddeerr screen will appear (Figure 57).
a. If using the default folder, click NNeexxtt.
b. If using a different folder, create a different folder name and click NNeexxtt.
FFiigguurree 5577 –– SSeelleecctt PPrrooggrraamm FFoollddeerr
15.Click IInnssttaallll to begin the installation (Figure 58).
NNoottee -- To return to the SSeelleecctt PPrrooggrraamm FFoollddeerr, click BBaacckk.
FFiigguurree 5588 –– IInnssttaalllliiBBeedd SSeerrvveerr
5212-309-001 Rev A.1 35 EN
Page 40
16.Click FFiinniisshh to exit the IInnssttaallllSShhiieelldd WWiizzaarrdd (Figure 59).
FFiigguurree 5599 –– FFiinniisshh
ii
BBeedd SSeerrvveerr TToooollss
1. Begin theiiBBeedd Server Tools installation by right clicking the 55221122--550022--000011 SSeerrvveerr TToooollss SSeettuupp..eexxee file and select
RRuunn aass AAddmmiinniissttrraattoorr.
2. The CChhoooossee DDeessttiinnaattiioonn LLooccaattiioonn screen will appear (Figure 60).
a. If using the default location, click NNeexxtt.
b. If using a different location, click CChhaannggee and then NNeexxtt to confirm.
FFiigguurree 6600 –– CChhoooossee DDeessttiinnaattiioonn LLooccaattiioonn
3. Click FFiinniisshh to exit the IInnssttaallllSShhiieelldd WWiizzaarrdd (Figure 61).
EN 36 5212-309-001 Rev A.1
Page 41
FFiigguurree 6611 –– FFiinniisshh
ii
BBeedd WWiirreelleessss CCoonnffiigguurraattiioonn TTooooll
CCAAUUTTIIOONN -- Before proceeding with this installation, make sure that a previous version of theiiBBeedd Wireless configuration tool application is not currently installed on the target system. If a previous version was installed, uninstall the software. If you attempt to install the application on a system where a previous version was installed, the installer behavior may be unpredictable.
1. Run the
SSeettuupp..eexxee file on the source location.
2. In the IInnssttaallllSShhiieelldd WWiizzaarrdd screen, click NNeexxtt (Figure 62).
ii
BBeedd Wireless Configuration Tool by double-clicking the 55221122--550033--000011iiBBeedd WWiirreelleessss ccoonnffiigguurraattiioonn ttooooll
FFiigguurree 6622 –– IInnssttaallllSShhiieelldd WWiizzaarrdd
5212-309-001 Rev A.1 37 EN
Page 42
3. The CChhoooossee DDeessttiinnaattiioonn LLooccaattiioonn screen will appear (Figure 63).
a. If using the default location, click NNeexxtt.
b. If using a different location, click CChhaannggee and then NNeexxtt to confirm.
FFiigguurree 6633 –– CChhoooossee DDeessttiinnaattiioonn LLooccaattiioonn
4. Click FFiinniisshh to exit the IInnssttaallllSShhiieelldd WWiizzaarrdd (Figure 64).
FFiigguurree 6644 –– FFiinniisshh
EEddiittiinngg WWiinnddoowwss ccoonnffiigguurraattiioonn
1. Browse to the machine.config file.
C:\Windows\Microsoft.NET\Framework\v2.0.50727\CONFIG\ machine.config (32 bit)
C:\Windows\Microsoft.NET\Framework64\v2.0.50727\CONFIG\ machine.config (64 bit)
2. To edit the previous file to increase the thread count for the .NET Framework, replace <processModel autoConfig=
"false" with <processModel autoConfig="false" maxWorkerThreads="1000" maxIoThreads="1000" minWorkerThreads= "50" minIoThreads="50"/>.
EN 38 5212-309-001 Rev A.1
Page 43
3. Browse to: CC::\\PPrrooggrraamm FFiilleess ((xx8866))\\SSttrryykkeerr\\iiBBeedd SSeerrvveerr AApppplliiccaattiioonn\\HHBB\\SSEEWWSSII..HHeeaarrttBBeeaattWWiinnddoowwssSSeerrvviiccee..eexxee..
ccoonnffiigg.
4. To increase the available ports (TCP connections):
a. Execute using a command prompt: netsh int ipv4 set dynamicport tcp start=1025 num=64510
VVeerriiffyyiiBBeedd SSeerrvveerr
To verifyiiBBeedd Server:
1. Open the SSttrryykkeerr
your desktop, or by clicking SSttaarrtt >> AAllll PPrrooggrraammss >> SSttrryykkeerr >>
ii
BBeedd SSeerrvveerr TTooooll by either double-clicking theiiBBeedd Server Tools Suite shortcut located on
ii
BBeedd SSeerrvveerr TToooollss (Figure 65).
FFiigguurree 6655 –– SSttrryykkeerriiBBeedd SSeerrvveerr TTooooll
2. Click VVeerriiffiiccaattiioonn>>SSeerrvveerr in the task bar.
3. In the
5212-309-001 Rev A.1 39 EN
ii
BBeedd SSeerrvveerr VVeerriiffiiccaattiioonn window, click VVeerriiffyy SSeerrvveerr (Figure 66).
Page 44
FFiigguurree 6666 –– VVeerriiffyyiiBBeedd SSeerrvveerr
a. If the server interface verification is successful, the system returns a green check (Figure 66).
b. If the server interface verification is unsuccessful, the system returns a red X.
NNoottee -- Before you continue installing beginning of the installation process and make sure that all steps were executed properly. For further troubleshooting details, see
4. Proceed to
Adding devices (clients) to the Master Device List
Troubleshooting
ii
BBeedd Server, you must resolve this error. To resolve the error, return to the
(page 56).
(page 41).
EN 40 5212-309-001 Rev A.1
Page 45
SSeettuupp
AAddddiinngg ddeevviicceess ((cclliieennttss)) ttoo tthhee MMaasstteerr DDeevviiccee LLiisstt
1. Open theiiBBeedd Server Tool by either double-clicking theiiBBeedd Server Tools Suite shortcut located on your
desktop, or by clicking SSttaarrtt >> AAllll PPrrooggrraammss >> SSttrryykkeerr >> iiBBeedd SSeerrvveerr TToooollss >>
2. In the
ii
BBeedd SSeerrvveerr TTooooll window, click EEddiitt>>MMaasstteerr DDeevviiccee LLiisstt (Figure 67).
ii
BBeedd SSeerrvveerr TToooollss.
FFiigguurree 6677 –– EEddiitt MMaasstteerr DDeevviiccee LLiisstt
3. In the AAdddd NNeeww DDeevviiccee UURRLL box, type in the URL of the device and then click AAdddd DDeevviiccee UURRLL (example: http//
10.32.56.101:1639 or http://syk-84253f2356a.stryker.com:1639) (Figure 68).
NNoottee -- Make sure to enter the web extension http:// and the Stryker communication port :1639 to the IP or DNS name for each device (client).
5212-309-001 Rev A.1 41 EN
Page 46
FFiigguurree 6688 –– AAdddd DDeevviiccee UURRLL
4. Repeat step 3 until all new devices have been added.
5. Once all new devices are in the list, click the SSeelleecctt AAllll box of the MMaasstteerr DDeevviiccee UURRLL LLiisstt and then click the UUppllooaadd
MMaasstteerr DDeevviiccee UURRLL LLiisstt ttoo tthhee SSeerrvveerr button (Figure 69).
FFiigguurree 6699 –– MMaasstteerr DDeevviiccee UURRLL LLiisstt
6. Click the OOKK button in the AAlleerrtt window confirming that the MMaasstteerr DDeevviiccee UURRLL LLiisstt was uploaded to the server.
7. To make sure that the MMaasstteerr DDeevviiccee UURRLL LLiisstt uploaded, go to the
SSeerrvveerr.
NNoottee -- Allow time for synchronization before you make sure that the MMaasstteerr DDeevviiccee UURRLL LLiisstt was uploaded.
EN 42 5212-309-001 Rev A.1
ii
BBeedd SSeerrvveerr TTooooll window and click VViieeww>>iiBBeedd
Page 47
AAddddiinnggiiBBeedd LLooccaattoorr IIDDss aanndd hhoossppiittaall llooccaattiioonnss
1. Open theiiBBeedd Server Tool by either double-clicking theiiBBeedd Server Tools Suite shortcut located on your
desktop, or by clicking SSttaarrtt >> AAllll PPrrooggrraammss >> SSttrryykkeerr >> iiBBeedd SSeerrvveerr TToooollss >>
2. In the
ii
BBeedd SSeerrvveerr TTooooll window, click EEddiitt>>LLooccaattiioonn AAssssoocciiaattiioonn (Figure 70).
FFiigguurree 7700 –– LLooccaattiioonn AAssssoocciiaattiioonn
ii
BBeedd SSeerrvveerr TToooollss.
3. In the LLooccaattiioonn AAssssoocciiaattiioonn LLiisstt box, click in the LLooccaattoorr IIDD box and type the ID from the Locator (Figure 71).
FFiigguurree 7711 –– AAdddd LLooccaattiioonn AAssssoocciiaattiioonn
4. In the LLooccaattiioonn AAssssoocciiaattiioonn LLiisstt box, click in the HHoossppiittaall LLooccaattiioonn box and type the Hospital Location (Figure 71).
NNoottee -- The Hospital Location is normally formatted using the location HL7 alias name.
5. To add another association, click the AAdddd RRooww button (Figure 71).
6. Repeat steps 3 and 4 until all new Locator IDs and Hospital Locations have been entered.
7. Once all new location associations have been made, click the upper left SSeelleecctt AAllll box of the Location Association List,
and then click UUppllooaadd LLooccaattiioonn AAssssoocciiaattiioonn((ss)) ttoo SSeerrvveerr (Figure 72).
5212-309-001 Rev A.1 43 EN
Page 48
FFiigguurree 7722 –– UUppllooaadd LLooccaattiioonn AAssssoocciiaattiioonn
8. Click the OOKK button in the AAlleerrtt window to confirm that the Locator ID List, Hospital Location List, and Location
Association List have been successfully uploaded to the server.
9. To verify in the
NNoottee -- Allow time for synchronization before you make sure that the Master Device URL List was uploaded.
ii
BBeedd SSeerrvveerr TTooooll window, click VViieeww>>iiBBeedd SSeerrvveerr.
AAddddiinngg aann aaddddiittiioonnaall SSttrryykkeerr iinntteerrffaaccee
SSmmaarrtt EEqquuiippmmeenntt MMaannaaggeemmeenntt ((SSEEMM))
NNoottee -- If the facility has an existing LLIIFFEENNEETT account, navigate to AAccccoouunntt DDeeffiinniittiioonn and check the SSmmaarrtt EEqquuiippmmeenntt MMaannaaggeemmeenntt box under FFeeaattuurreess.
CCrreeaattiinngg aa LLIIFFEENNEETT®® aaccccoouunntt
NNoottee -- Make sure that the facility does not have a pre-existing LLIIFFEENNEETT account. If they do not, continue to step 1.
1. In an internet browser window, go to hhttttppss::////wwwwww..lliiffeenneettssyysstteemmss..ccoomm// to reach the LLIIFFEENNEETT website.
2. Click on the AAccccoouunntt tab and click AAccccoouunnttss in the drop down list (Figure 73).
FFiigguurree 7733 –– LLIIFFEENNEETT ccrreeaattee aaccccoouunntt
3. To create a new account, click AAdddd AAccoouunntt (Figure 74).
EN 44 5212-309-001 Rev A.1
Page 49
4. Submit information for all of the required fields:
a. Add Account (Figure 75)
b. Account Address (Figure 76)
FFiigguurree 7744 –– AAdddd AAccccoouunntt
FFiigguurree 7755 –– AAdddd AAccccoouunntt
FFiigguurree 7766 –– AAccccoouunntt AAddddrreessss
c. Account Definition (Figure 77)
NNoottee -- Make sure that you select the following: LIFENET, Live, Hospital, Expiration Date, Smart Equipment Management, and Other Devices
5212-309-001 Rev A.1 45 EN
Page 50
d. Account Configuration
NNoottee -- This step can be skipped.
e. Account Administrator (Figure 78)
FFiigguurree 7777 –– AAccccoouunntt DDeeffiinniittiioonn
FFiigguurree 7788 –– AAccccoouunntt AAddmmiinniissttrraattoorr
f. Notes
NNoottee -- This step can be skipped.
5. Click DDoonnee which will prompt an email to be sent to the newly created Account Administrator. This email will contain a
link to the SEM home page and a temporary password that needs to be changed before installing the SEM Device Data Manager (Figure 79).
EN 46 5212-309-001 Rev A.1
Page 51
FFiigguurree 7799 –– SSEEMM hhoommee ppaaggee
IInnssttaalllliinngg tthhee SSttrryykkeerr SSEEMM DDeevviiccee DDaattaa MMaannaaggeerr
1. Upload the Stryker SEM Device Data Manager install file (5212-504-001) onto the Stryker server where the
Application is installed.
2. Begin the Stryker SEM Device Data Manager installation by right clicking the 55221122--550044--000011 DDeevviiccee DDaattaa MMaannaaggeerr file
and select RRuunn aass AAddmmiinniissttrraattoorr.
3. In the SSttrryykkeerr SSEEMM DDeevviiccee DDaattaa MMaannaaggeerr -- IInnssttaallllSShhiieelldd WWiizzaarrdd window, click NNeexxtt (Figure 80).
FFiigguurree 8800 –– IInnssttaallllSShhiieelldd WWiizzaarrdd
ii
BBeedd Server
4. In the SSttrryykkeerr SSEEMM DDeevviiccee DDaattaa MMaannaaggeerr -- IInnssttaallllSShhiieelldd WWiizzaarrdd window, enter the following values and click NNeexxtt
(Figure 81).
i
Server IP address: IP address of the machine where theiServer is installed
Hospital name: Name of the hospital
Agent Serial Number: Enter the MAC address of the server machine where the SEM Device Data Manager is installed
LLIIFFEENNEETT UserName - Username that is used to register Device Data Manager with LLIIFFEENNEETT
LLIIFFEENNEETT Password - Password that is used to register Device Data Manager with LLIIFFEENNEETT
LLIIFFEENNEETT URL - The URL for registering the Device Data Manager installed at the hospital with LLIIFFEENNEETT
NNoottee -- The LLIIFFEENNEETT user name and password are located in the email sent to the Account Administrator in
LIFENET® account
5212-309-001 Rev A.1 47 EN
(page 44).
Creating a
Page 52
FFiigguurree 8811 –– FFiieelldd vvaalluueess
5. In the CChhoooossee DDeessttiinnaattiioonn LLooccaattiioonn step, click NNeexxtt to install the SEM Device Data Manager files to the default location or click CChhaannggee to select a different destination folder (Figure 82).
FFiigguurree 8822 –– CChhoooossee DDeessttiinnaattiioonn LLooccaattiioonn
6. Click FFiinniisshh to close the SSttrryykkeerr SSEEMM DDeevviiccee DDaattaa MMaannaaggeerr -- IInnssttaallllSShhiieelldd WWiizzaarrdd window (Figure 83).
EN 48 5212-309-001 Rev A.1
Page 53
FFiigguurree 8833 –– IInnssttaallllSShhiieelldd WWiizzaarrdd CCoommpplleettee
VVeerriiffyyiinngg tthhee SSttrryykkeerr SSEEMM DDeevviiccee DDaattaa MMaannaaggeerr iiss iinnssttaalllleedd
1. In Windows, click SSttaarrtt.
2. Enter rruunn in the search box.
3. In the run window, enter iinneettmmggrr.
4. Expand CCoonnnneeccttiioonnss in the left panel.
5. Under SSiitteess navigate to and click DDeevviicceeDDaattaaMMaannaaggeerr (Figure 84).
5212-309-001 Rev A.1 49 EN
Page 54
FFiigguurree 8844 –– DDeevviicceeDDaattaaMMaannaaggeerr
6. Right click on DDaattaaMMaannaaggeerr..ssvvcc and click BBrroowwssee.
7. A browser window should open to the Data Manager Service web page.
AAddddiinngg aa tthhiirrdd--ppaarrttyy iinntteerrffaaccee
IInntteeggrraattiinngg RRaauullaanndd RReessppoonnddeerr®® 55
NNoottee -- This installation to be performed by Stryker personnel only.
To integrate RRaauullaanndd RReessppoonnddeerr 5, follow the installation steps found in SD-180 available from the Stryker quality system.
EN 50 5212-309-001 Rev A.1
Page 55
CCoonnffiigguurriinngg tthhee wwiirreelleessss rroouutteerr ((SSttrryykkeerr ddeevviiccee ccoonnffiigguurraattiioonn))
NNoottee --iiBBeedd Server 1.0 wireless clients only authenticate with TKIP encryption andiiBBeedd Server 2.0 wireless clients only authenticate with AES encryption. Below is an example of configuring a LINKSYS N600 E2500 router.
To configure the wireless router for Stryker devices:
1. Enter the router’s admin menu to configure the router for use.
2. In the SSeettuupp//BBaassiicc SSeettuupp tab, verify the router is set for DHCP (Figure 85).
FFiigguurree 8855 –– BBaassiicc SSeettuupp
5212-309-001 Rev A.1 51 EN
Page 56
3. In the WWiirreelleessss//BBaassiicc WWiirreelleessss SSeettttiinnggss tab, configure the 2.4 GHz and 5 GHz Wireless Settings so they match Figure 86.
FFiigguurree 8866 –– BBaassiicc WWiirreelleessss SSeettttiinnggss
4. In the WWiirreelleessss//WWiirreelleessss SSeeccuurriittyy tab, configure the 2.4 GHz and 5 GHz Wireless Security Settings so they match Figure 87.
FFiigguurree 8877 –– WWiirreelleessss SSeeccuurriittyy
5. Click the SSaavvee SSeettttiinnggss button.
EN 52 5212-309-001 Rev A.1
Page 57
CCoonnffiigguurriinngg wwiirreelleessss nneettwwoorrkk ccoonnnneeccttiioonn sseettttiinnggss
Applies to Gateway 1 (3002 SSeeccuurree®® II, 3002 / 3005 SS33®®), Gateway 2 (3002 / 3005 SS33), and Gateway 3 (2131 / 2141 IInnTToouucchh®®).
NNoottee -- You must have a laptop and a router configured for Stryker defaults.
1. Plug in the router configured for Stryker defaults (
51)).
2. Plug the power cord of the bed into the wall outlet and make sure that the wireless option is turned on.
3. Connect the laptop to the syk_med_install SSID which the configured router is broadcasting.
4. Open the
5. On the bed, enter the CCoonnnneeccttiivviittyy IInnffoo menu by accessing the service menu on the footboard (see the appropriate product maintenance manual).
6. Scroll down to the IP address which the router provided for the bed.
7. Enter the bed IP address in the WWiirreelleessss DDeevviiccee UURRLL//IIPP box (Figure 88).
ii
BBeedd Wireless Configuration Tool.
Configuring the wireless router (Stryker device configuration)
(page
FFiigguurree 8888 –– WWiirreelleessss DDeevviiccee UURRLL
8. Click the GGeett WWiirreelleessss DDeevviiccee CCoonnffiigguurraattiioonn button. This will retrieve the bed wireless default settings and connect to the bed, which is shown in the left column of the tool.
9. Click the OOKK button of the retrieval confirmation.
10.Enter the facility network information and make sure that you fill in all of the appropriate blanks in the right column of the tool.
11.Select the RRaaddiioo MMooddee for the facility requirements.
12.Click the UUppllooaadd CCoonnffiigguurraattiioonn ttoo DDeevviiccee button.
NNoottee -- If the radio does not connect, make sure that you check the settings entered into the column on the right. If they are incorrect, the radio will need to be reset to defaults and then repeat the configuration process.
5212-309-001 Rev A.1 53 EN
Page 58
RReesseettttiinngg tthhee wwiirreelleessss mmoodduullee ttoo ffaaccttoorryy ddeeffaauulltt sseettttiinnggss ((MMeedd--SSuurrgg
B
A
bbeedd))
TToooollss rreeqquuiirreedd::
Large paper clip
PPrroocceedduurree::
1. Raise the product to its highest position.
2. Using a large paper clip, insert it into the reset hole (A) on the bottom side of the wireless module (B) (Figure 89).
3. Hold the large paper clip inside the reset hole for five seconds.
FFiigguurree 8899 –– WWiirreelleessss mmoodduullee
EN 54 5212-309-001 Rev A.1
Page 59
RReesseettttiinngg tthhee wwiirreelleessss mmoodduullee ttoo ffaaccttoorryy ddeeffaauulltt sseettttiinnggss ((MMooddeell FFLL2277
A
B
C
IInnTToouucchh))
NNoottee -- For IInnTToouucchh 2.1-4.0, follow all of the steps below. For IInnTToouucchh 5.0 or higher, see step 1 to reset Wi-Fi.
TToooollss rreeqquuiirreedd::
Appropriate FL27 IInnTToouucchh maintenance manual
PPrroocceedduurree::
1. Use the appropriate FL27 IInnTToouucchh maintenance manual to access the CCoonnffiigguurraattiioonn menu (For IInnTToouucchh 2.1-4.0 see Figure 90. For IInnTToouucchh 5.0 or higher, see Figure 92).
2. Tap WWii--FFii CCoonnffiigguurraattiioonn (A) (Figure 90).
3. Tap the AAddvvaanncceedd tab (B) (Figure 91).
4. Tap RReesseett (C) (Figure 91).
5. Tap SSaavvee.
IInnTToouucchh 22..11 -- 44..00
FFiigguurree 9900 –– IInnTToouucchh CCoonnffiigguurraattiioonn MMeennuu
IInnTToouucchh 55..00 oorr hhiigghheerr
FFiigguurree 9911 –– IInnTToouucchh WWii--FFii CCoonnffiigguurraattiioonn mmeennuu
FFiigguurree 9922 –– IInnTToouucchh WWii--FFii RReesseett
5212-309-001 Rev A.1 55 EN
Page 60
TTrroouubblleesshhoooottiinngg
BBaassiicc
PPrroobblleemm PPoossssiibbllee ccaauussee SSoolluuttiioonn
Make sure the URL in the Web Service URL box is correct.
1. If the URL is incorrect, enter the following URL into the Web Service
The Web Service URL is incorrect.
URL box (
39)):hhttttpp::////llooccaallhhoosstt//SSEEWWSSII// SSEEWWSSIIWWeebbSSeerrvviicceeCCoonnttrraacctt.. aassmmxx.
2. Click VVeerriiffyy SSeerrvveerr.
1. Make sure that the Heartbeat service is running.
Verify iBed Server
(page
Cannot verify the
ii
BBeedd Server Tool
The Heartbeat service is not active.
The IIS Manager does not have all of the required Default Web Sites.
1.1. If the Heartbeat service has
not started, start the service, restart the server, and then reverify the application.
1. Make sure that the IIS Manager has all of the required Default Web Sites.
1.1. Go to IInntteerrnneett IInnffoorrmmaattiioonn
SSeerrvviicceess ((IIIISS)) MMaannaaggeerr.
1.2. Navigate to WWeebb SSiitteess >>
DDeeffaauulltt WWeebb SSiittee to verify that the three web sites are listed: CHAP / SERVERSCHEMA / SEWSI.
1.3. If the Default Web Site is not
listed, reinstall the Server application.
i
Bed Server
ii
BBeedd
EN 56 5212-309-001 Rev A.1
Page 61
PPrroobblleemm PPoossssiibbllee ccaauussee SSoolluuttiioonn
1. If the troubleshooting steps in the appropriate bed maintenance manual were followed:
1.1. Use the MAC address from
1.2. Ping the wireless module
Cannot see the device in the Server application
ii
BBeedd
The MAC address does not have the bed’s IP address or DNS name.
1.3. Press EEnntteerr.
the bed and verify with IT what wireless module’s IP address or DNS name is.
from the server machine using the Command Prompt (ping IIPP aaddddrreessss oorr DDNNSS nnaammee :1639).
1.3.1. If you receive four (4) responses, then the bed is on the wireless network.
1.3.2. If you receive no response, check the wireless module on the bed.
PPrroobblleemm PPoossssiibbllee ccaauussee SSoolluuttiioonn
1. Using the server tool, select VVeerriiffyy DDeevviiccee DDiirreeccttllyy.
2. Click GGeett EExxiissttiinngg DDeevviiccee((ss)), and then select the device in question.
3. Add it to the DDeevviiccee LLiisstt.
Cannot see a device in the Server Tools under the Device Location List, but the device shows under the connected device URL List
The device was not added to the device list.
4. In the DDeevviiccee LLiisstt, select the device, and then click VVeerriiffyy DDeevviiccee((ss)).
5. Check the CCoonnnneeccttiioonn SSttaattuuss once it responds.
5.1. If there is a response, the
device is communicating properly on the wireless network.
5212-309-001 Rev A.1 57 EN
Page 62
PPrroobblleemm PPoossssiibbllee ccaauussee SSoolluuttiioonn
Identify the bed serial number that is stored in the CPU. It should match the serial number on the serial label on the bed.
NNoottee -- If there are two matching serial numbers in the device list, the server will only recognize the
The serial number in the CPU and the serial number on the bed specification label do not match.
PPrroobblleemm PPoossssiibbllee ccaauussee SSoolluuttiioonn
first device that logs on to the server.
To find the bed’s serial number, put the bed into the Bed Configuration mode and navigate to SSeerriiaall NNuummbbeerr >> CCuurrrreenntt SSNN to verify the serial number.
If the serial number does not match, go to the EEddiitt SSNN menu to edit the serial number, and then save.
Cannot see a location
The Locator ID was entered incorrectly.
1. Using the Tool, select VViieeww CCuurrrreenntt LLooccaattiioonn AAssssoocciiaattiioonnss.
2. Click GGeett LLooccaattiioonn AAssssoocciiaattiioonnss from server.
3. Look through the HHoossppiittaall LLooccaattiioonn and verify the Locator ID.
NNoottee -- The Locator ID is alphanumeric and case sensitive. You must use lower case to enter the Locator ID letters, or the server will not recognize the locator.
3.1. If the Locator ID is incorrect,
3.2. If the Locator ID is missing,
ii
BBeedd Locator Association
use the UUppddaattee function to update the error.
see
Adding iBed Locator IDs
and hospital locations
43).
(page
EN 58 5212-309-001 Rev A.1
Page 63
PPrroobblleemm PPoossssiibbllee ccaauussee SSoolluuttiioonn
1. Make sure the SEWSI wweebb..ccoonnffiigg file has the correct Web URL..
1.1. Navigate to the directory
Third party does not receive bed status information
2. Make sure PPoorrtt 8800 is not blocked.
where the is installed.
1.1.1. Navigate to the
1.1.2. Find the line which
ii
BBeedd Server Tool
SEWSI folder and open the wweebb..ccoonnffiigg file.
contains SSeerrvveerrUURRLL==
1) Make sure the local host was replaced with the IP of the server
NNoottee -- This cannot be a DNS name.
2.1. PPoorrtt 8800 needs to be open between the Stryker server and products (bi-directional), and the Stryker server and any third-party servers (bi­directional) ((LLIINNKK HHEERREE)).
2.2. If PPoorrtt 8800 is blocked, work with IT to open the port.
3. Search SEWSI logs to make sure there are no communication failures. Find messages generated for the third-party ((LLIINNKK HHEERREE))
AAddvvaanncceedd
CCoonnnneeccttiivviittyy iissssuueess ((ttoottaall ddeevviiccee ccoouunnttss))
View total device counts to identify drops.
Navigate to the directory where iBed Server Tools is installed and open the LOGS folder.
1. Find SSttrryykkeerrMMaaiinneennaanncceeSSeerrvviiccee__llooggffiillee..ttxxtt.
2. Highlight several of the SSttrryykkeerrMMaaiinneennaanncceeSSeerrvviiccee__llooggffiillee..ttxxtt files during the time when there may have been an outage.
a. Right click and select EEddiitt wwiitthh NNootteeppaadd++++..
3. In NNootteeppaadd++++ press CCttrrll++FF to open a find window.
a. Search for the below string and select FFiinndd iinn AAllll OOppeenneedd DDooccuummeennttss.
NNoottee -- [[TToottaall CCoonnnneecctteedd DDeevviicceess CCoouunntt eexxcclluuddiinngg sseerrvveerr ==
5212-309-001 Rev A.1 59 EN
Page 64
b. The FFiinndd RReessuulltt pane in NNootteeppaadd++++ will populate with the lines of the selected log files (Figure 93).
FFiigguurree 9933 –– TToottaall ddeevviiccee ccoouunnttss
c. Review the list until you locate the drop in the number of connected devices and the duration of time the devices
were offline.
Work with the local IT department to investigate.
TThhiirrdd ppaarrttyy ccoommmmuunniiccaattiioonn iissssuueess
Search SEWSI log files for third party messages.
1. Open the
2. Navigate to the subscription list tab.
NNoottee -- When you search for messages from a product, make sure that you use the correct Device ID.
3. Copy the subscription ID for a device which is going to the correct third party.
ii
BBeedd Server Tool.
a. Verify subscriber URL to find messages to a specific third party.
4. Navigate to the directory where
a. Open the LLOOGGSS folder.
b. Find SSEEWWSSII__llooggffiillee..ttxxtt.
5. To search for messages in a specific timeframe, highlight multiple SSEEWWSSII__llooggffiillee..ttxxtt in that range based on the date modified in Windows Explorer.
a. Right click the highlighted files and select EEddiitt wwiitthh NNootteeppaadd++++.
6. Press CCttrrll++FF to open a find window.
a. Paste the SSuubbssccrriippttiioonn IIDD copied in step 3.
b. Select FFiinndd iinn aallll ooppeenneedd ddooccuummeennttss.
c. The Find result pane will open with all messages containing the SSuubbssccrriibbeerr IIDD.
7. Click the XML message from the Find pane to open the file to that line.
a. XML messages will start in format <<??xxmmll vveerrssiioonn==””11..00”” eennccooddiinngg==uuttff--88””??>>.
NNoottee -- Example message with the payload XML data highlighted: <?xml version="1.0" encoding="utf-8"?><soap: Envelope xmlns:soap="http://schemas.xmlsoap.org/soap/envelope/" xmlns:xsi="http://www.w3.org/2001/XMLSchema­instance" xmlns:xsd="http://www.w3.org/2001/XMLSchema"><soap:Body><ReceiveSubscriptionRequest xmlns= "http://SEWSI.ServiceContracts/2008/09"><SubscriberID>C3000-000-000_180415201155204742</ SubscriberID><<PPaayyllooaaddXXMMLL>>&&lltt;;ddeevviiccee iidd==""33000000--000000--000000__118800441155220011""&&ggtt;;&&lltt;;BBeeddHHeeiigghhttIInn&&ggtt;;2277&&lltt;;// BBeeddHHeeiigghhttIInn&&ggtt;;&&lltt;;LLaassttLLooggggeeddWWeeiigghhttLLbb&&ggtt;;110022..990000000099&&lltt;;//LLaassttLLooggggeeddWWeeiigghhttLLbb&&ggtt;;&&lltt;; BBeeddEExxiittAAllaarrmmiinngg&&ggtt;;ffaallssee&&lltt;;//BBeeddEExxiittAAllaarrmmiinngg&&ggtt;;&&lltt;;//ddeevviiccee&&ggtt;;<<//PPaayyllooaaddXXMMLL>><<TTiimmeeSSttaammpp>>22001188--0077-­1166TT1111::5599::1144..33441100664477--0044::0000<<//TTiimmeeSSttaammpp>></ReceiveSubscriptionRequest></soap:Body></soap:Envelope>] []
ii
BBeedd Server Tool is installed.
EN 60 5212-309-001 Rev A.1
Page 65
DDeevviiccee nnoott ccoonnnneeccttiinngg ttoo sseerrvveerr
Search for duplicate serial numbers (devices newer than GW1).
1. Open the
2. Navigate to the CClliieenntt DDiiaaggnnoossttiicc IInnffoo tab.
3. Click the far left blank cell of the header to highlight the entire table.
a. Copy and paste the data in MMiiccrroossoofftt EExxcceell.
4. Highlight the CClliieenntt IIDD column.
5. In the HHoommee tab of the ribbon, click CCoonnddiittiioonnaall FFoorrmmaattttiinngg.
a. Highlight CCeellllss RRuulleess -- DDuupplliiccaattee VVaalluueess.
b. Click OOKK to highlight duplicates in red.
c. Highlight the row above where the data was pasted, click the DDaattaa tab, and select FFiilltteerr.
Any cells highlighted (except NNeettwwoorrkk EExxcceeppttiioonn // TTiimmeeoouuttss) are duplicate serial numbers which need to be resolved.
ii
BBeedd Server Tool.
CCoonnnneeccttiivviittyy iissssuueess ttoo oonnee // mmuullttiippllee ddeevviicceess
Not receiving specific bed status over port 80.
Verify communication from devices to the server with SOAP UI.
NNoottee -- Requires Stryker supplied project file.
1. Open SSOOAAPP UUII.
2. In the PPrroojjeeccttss pane, expand the SSEEWWSSII folder.
3. Click CCoonnnneeccttiivviittyy MMoocckk to open CCoonnnneeccttiivviittyy MMoocckk window (Figure 94).
FFiigguurree 9944 –– CCoonnnneeccttiivviittyy MMoocckk
5212-309-001 Rev A.1 61 EN
Page 66
4. Click the gear icon to edit settings for Mock Third Party.
a. Modify host URL to the IP address of the server where SOAP UI is running (Figure 95).
FFiigguurree 9955 –– SSOOAAPP UUII
b. Click OOKK to save.
5. Click the green PPllaayy icon to start the Mock Third Party Service.
6. Click the green arrows to open the service webpage to verify.
a. Copy the URL of the webpage to the clipboard
7. In the Projects Pane, expand SSEEWWSSIIWWeebbSSeerrvviicceeCCoonnttrraacctt.
a. Expand SSuubbssccrriibbeeOOnnCChhaannggee.
b. Double click VVTTFF -- SSuubb 33 ((7722 hhoouurr)).
c. Modify the Subscriber URL with the URL from step 3.3.1 (Figure 96).
FFiigguurree 9966 –– VVTTFF -- SSuubb 33 ((7722 hhoouurr))
d. Change the Device ID to the Device ID being searched for (Figure 96).
e. Click the Play icon in VVTTFF -- SSuubb 33 ((7722 hhoouurr)) window.
EN 62 5212-309-001 Rev A.1
Page 67
8. In the CCoonnnneeccttiivviittyyMMoocckk window, you will begin to see messages appear in the message log if the ports are open and the server is able to communicate with the device (Figure 97).
NNoottee -- If messages do not populate, port 80 may be closed or the WWeebbSSeerrvviicceeUURRLL in the SSEEWWSSII wweebb..ccoonnffiigg file may be incorrect.
FFiigguurree 9977 –– MMeessssaaggee LLoogg
9. Double click a message to view the XML generated (Figure 98).
FFiigguurree 9988 –– MMeessssaaggee VViieewweerr
CCaannnnoott ccoommmmuunniiccaattee wwiitthh ddeevviiccee ((vveerriiffyy ppoorrttss aarree ooppeenn))
Verify communication with device over port using TTEELLNNEETT.
1. Open a ccoommmmaanndd prompt.
5212-309-001 Rev A.1 63 EN
Page 68
2. Enter the command tteellnneett <<IIPPAAddddrreessss>> <<PPoorrtt>>, where IIPP AAddddrreessss is the address of the device and PPoorrtt is the port number for verification (Figure 99).
FFiigguurree 9999 –– CCoommmmaanndd pprroommpptt
A blank box with a blinking cursor indicates that the port is open.
CCoonnnneeccttiinngg ttoo indicates that the port is closed or the device is offline (Figure 100).
FFiigguurree 110000 –– CCoonnnneeccttiinngg
NNoottee -- Only use TTEELLNNEETT for a device that is online and pingable.
CCaannnnoott ccoommmmuunniiccaattee wwiitthh ddeevviiccee ((DDNNSS oonnllyy))
Verify Hostname with NNSSLLOOOOKKUUPP
1. Open a ccoommmmaanndd prompt.
2. Enter the command nnssllooookkuupp <<IIPPAAddddrreessss>> where IIPP AAddddrreessss is the address of the device for DDNNSS eennttrryy verification.
A reply with the hhoossttnnaammee and IIPP aaddddrreessss indicates success (Figure 101).
NNoottee -- The first line is the DDNNSS server name and IP address. The second line is the device hostname/IP.
FFiigguurree 110011 –– DDNNSS ssuucccceessss
EN 64 5212-309-001 Rev A.1
Page 69
A reply that states <<DDNNSSSSeerrvveerrNNaammee>> ccaann’’tt ffiinndd <<IIPPAAddddrreessss>> indicates failure (Figure 102).
FFiigguurree 110022 –– DDNNSS ffaaiilluurree
3. To verify a reverse lookup, enter the command nnssllooookkuupp <<hhoossttnnaammee>> where the hhoossttnnaammee is the device you are verifying DNS entry for.
NNoottee -- The server reply should be the same as seen in step 2.
SSmmaarrtt EEqquuiippmmeenntt MMaannaaggeemmeenntt ((ooppttiioonn)) ttrroouubblleesshhoooottiinngg
1. To verify DDeevviiccee DDaattaa MMaannaaggeerr UUttiilliittyy:
In Windows, click SSttaarrtt.
Enter SSttrryykkeerrDDeevviicceeDDaattaaMMaannaaggeerrRReeggiissttrraattiioonn in the search box.
Make sure that the SSttrryykkeerrDDeevviicceeDDaattaaMMaannaaggeerrRReeggiissttrraattiioonn utility is displayed in the search results.
2. To verify in
ii
BBeedd SSeerrvveerr TTooooll:
Locate the machine where the Stryker
Click the SSuubbssccrriippttiioonn LLiisstt tab and make sure that the DDeevviiccee IIDD and SSuubbssccrriippttiioonn IIDD columns are populated.
Make sure that the server has a general subscription and each device has both a general and alarm subscription
(Figure 103).
ii
BBeedd Server application is installed and open theiiBBeedd SSeerrvveerr TTooooll.
FFiigguurree 110033 –– GGeenneerraall ssuubbssccrriippttiioonn
Make sure that the SSuubbssccrriibbeerr UURRLL column displays the same URL shown in the WWeebb..ccoonnffiigg file of the Device
Data Manager (Figure 104 and Figure 105).
5212-309-001 Rev A.1 65 EN
Page 70
FFiigguurree 110044 –– SSuubbssccrriippttiioonn LLiisstt
FFiigguurree 110055 –– WWeebb..ccoonnffiigg ffiillee
3. To verify server data in the Device Data Manager log file:
Search for string ““DDeevviicceeIIDD==<<SSeerriiaallNNuummbbeerr>> JJssoonnCCoonntteenntt==”” in the log file (Figure 106)
NNoottee -- Replace <<SSeerriiaallNNuummbbeerr>> with serial number in the WWeebb..ccoonnffiigg file.
FFiigguurree 110066 –– EExxaammppllee:: DDeevviicceeIIDD == 11223344556677889900 JJssoonnCCoonntteenntt==
Search for string ““DDaattaa UUppllooaaddeedd ffoorr SSeerrvveerr ssuucccceessssffuullllyy!!”” in the log file (Figure 107)
FFiigguurree 110077 –– DDaattaa uuppllooaaddeedd iinn lloogg ffiillee
EN 66 5212-309-001 Rev A.1
Page 71
4. To verify connected devices in the log file:
Search for string ““DDeevviicceeIIDD == <<ddeevviicceeIIDD>> JJssoonnCCoonntteenntt==”” in the log file (Figure 108)
NNoottee -- Replace <<ddeevviicceeIIDD>> with device ID of the connected device.
FFiigguurree 110088 –– EExxaammppllee:: DDeevviicceeIIDD == 33000000--000000--000000__DDeevviicceeSS11 JJssoonnCCoonntteenntt==
Search for string ““DDaattaa UUppllooaaddeedd ssuucccceessssffuullllyy ffoorr DDeevviicceeIIDD == <<ddeevviicceeIIDD>>”” in the log file (Figure 109)
NNoottee -- Replace <<ddeevviicceeIIDD>> with device ID of the connected device.
FFiigguurree 110099 –– EExxaammppllee:: DDaattaa UUppllooaaddeedd ssuucccceessssffuullllyy ffoorr 33000000--000000--000000__DDeevviicceeSS11 iinn LLIIFFEENNEETT
EEmmaaiill aalleerrttss
The following email alerts may be generated by theiiBBeedd server application.
EEmmaaiill iinnffoo SScceennaarriioo
EEmmaaiill SSuubbjjeecctt:: Stryker iBed Server Alert
EEmmaaiill BBooddyy:: 2015-04-20 00:24:59,298 [2964] ERROR Stryker.IServer. BusinessLogic.SEWSI.RuleManager. EMailToAdminForLowBatteryStatus - SEND EMAIL: Stryker room locator f5d5b2130000 in room 205 has a low battery. Currently connected to device ID 3000-000-000_
130316141.
EEmmaaiill SSuubbjjeecctt:: Stryker iBed Server Alert
EEmmaaiill BBooddyy:: 2015-04-20 00:35:51,784 [2328] ERROR Stryker.IServer. BusinessLogic.SEWSI.RuleManager. EMailToAdminForLowBatteryStatus - SEND EMAIL: Stryker room locator f5d5b2130000 is not in the location list and has a low battery, update the location and association lists as required. Currently connected to device ID 3000­000-000_130316141.
Low Battery Status is true for the BBID mapped with a location
Low Battery Status is true for a BBID which is not mapped to any room or the BBID is missing
5212-309-001 Rev A.1 67 EN
Page 72
EEmmaaiill iinnffoo SScceennaarriioo
EEmmaaiill SSuubbjjeecctt:: Stryker iBed Server Alert
EEmmaaiill BBooddyy:: 2015-04-20 00:45:22,254 [2328] ERROR Stryker.IServer. BusinessLogic.SEWSI.RuleManager. EmailToAdminForMissingBBID - SEND EMAIL: Stryker
BBID is sent in the subscription payload but it is not present in the “BBIDList.xml” file
room locator f5d5b2130000 is not in the location list and does not have an association to a room, update or create the location association as required. Currently connected to device ID 3000-000-000_130316141.
EEmmaaiill SSuubbjjeecctt:: Stryker iBed Server Alert
EEmmaaiill BBooddyy:: 2015-04-20 00:48:02,395 [2328] ERROR Stryker.IServer. BusinessLogic.SEWSI.RuleManager. EmailToAdminForMissingBBID - SEND EMAIL: Stryker
BBID is sent in payload and it is present in “BBIDList.xml” file but not present in “DeviceBBIdLocationAssciation.xml”
file room locator f5d5b2130000 is in the location list but is not associated to a room, update the location association as required. Currently connected to device ID 3000-000-000_
130316141.
EEmmaaiill SSuubbjjeecctt:: Stryker iBed Server Alert
EEmmaaiill BBooddyy:: 2015-04-20 00:50:52,536 [2328] ERROR Stryker.IServer. BusinessLogic.SEWSI.RuleManager. EmailToAdminForMissingBBID - SEND EMAIL: Stryker room locator f5d5b2130000 is in the location list but is associated to an undefined room, update the location association as required. Currently connected to device ID 3000-000-000_130316141.
EEmmaaiill SSuubbjjeecctt :: SSttrryykkeerr iiBBeedd SSeerrvveerr UUrrggeenntt AAlleerrtt
EEmmaaiill BBooddyy:: AApppplliiccaattiioonn HHeeaalltthh CChheecckk FFaaiilleedd aatt SStteepp 11 -->> iiBBeeddSSeerrvveerr iiss uunnaabbllee ttoo aacccceessss tthhee ddaattaabbaassee.. PPlleeaassee rreessttaarrtt tthhee MMSS--SSQQLL ddaattaabbaassee.. AAfftteerr rreessttaarrtt,, iiff tthhee mmeessssaaggee ccoonnttiinnuueess,, pplleeaassee ccoonnttaacctt SSttrryykkeerr ssuuppppoorrtt..
EEmmaaiill SSuubbjjeecctt:: SSttrryykkeerr iiBBeedd SSeerrvveerr UUrrggeenntt AAlleerrtt
EEmmaaiill BBooddyy:: AApppplliiccaattiioonn HHeeaalltthh CChheecckk FFaaiilleedd aatt SStteepp 22 -->> TThhee MMSS-­SSQQLL ddaattaabbaassee iiss ccoorrrruupptt.. FFoolllloowwiinngg TTaabblleess aarree mmiissssiinngg iinn DDaattaabbaassee -->>DDeevviicceeCCoonnnneeccttiioonnIInnffooLLiisstt.. PPlleeaassee ccoonnttaacctt SSttrryykkeerr ssuuppppoorrtt ttoo rreebbuuiilldd..AApppplliiccaattiioonn HHeeaalltthh CChheecckk FFaaiilleedd aatt SStteepp 33 -->> iiBBeeddSSeerrvveerr iiss uunnaabbllee ttoo aacccceessss SSEEWWSSII.. PPlleeaassee rreessttaarrtt tthhee WWoorrlldd WWiiddee WWeebb PPuubblliisshhiinngg SSeerrvviiccee aanndd SSttrryykkeerrSSEEWWSSIIHHeeaarrttbbeeaattsseerrvviiccee.. AAfftteerr rreessttaarrtt,, iiff tthhee mmeessssaaggee ccoonnttiinnuueess,, pplleeaassee ccoonnttaacctt SSttrryykkeerr ssuuppppoorrtt..
BBID is sent in payload and it is present in “BBIDList.xml”
file and in “DeviceBBIdLocationAssciation.xml” file but
location for that BBID is missing in
“DeviceBBIdLocationAssciation.xml” file
When DB is down
DeviceConnectionInfoList table doesn't exist
EN 68 5212-309-001 Rev A.1
Page 73
EEmmaaiill iinnffoo SScceennaarriioo
EEmmaaiill SSuubbjjeecctt:: SSttrryykkeerr iiBBeedd SSeerrvveerr UUrrggeenntt AAlleerrtt
EEmmaaiill BBooddyy:: AApppplliiccaattiioonn HHeeaalltthh CChheecckk FFaaiilleedd aatt SStteepp 22 -->> TThhee MMSS-­SSQQLL ddaattaabbaassee iiss ccoorrrruupptt..FFoolllloowwiinngg TTaabblleess aarree mmiissssiinngg iinn DDaattaabbaassee -->>MMaasstteerrSSuubbssccrriippttiioonnIInnffoo.. PPlleeaassee ccoonnttaacctt SSttrryykkeerr ssuuppppoorrtt ttoo rreebbuuiilldd..
EEmmaaiill SSuubbjjeecctt:: SSttrryykkeerr iiBBeedd SSeerrvveerr UUrrggeenntt AAlleerrtt
EEmmaaiill BBooddyy:: AApppplliiccaattiioonn HHeeaalltthh CChheecckk FFaaiilleedd aatt SStteepp 22 -->> TThhee MMSS-­SSQQLL ddaattaabbaassee iiss ccoorrrruupptt..FFoolllloowwiinngg LLooggiinnss aarree mmiissssiinngg iinn DDaattaabbaassee -->>NNTT AAUUTTHHOORRIITTYY\\LLOOCCAALL SSEERRVVIICCEE.. PPlleeaassee ccoonnttaacctt SSttrryykkeerr ssuuppppoorrtt ttoo rreebbuuiilldd..
EEmmaaiill SSuubbjjeecctt:: SSttrryykkeerr iiBBeedd SSeerrvveerr UUrrggeenntt AAlleerrtt
EEmmaaiill BBooddyy:: AApppplliiccaattiioonn HHeeaalltthh CChheecckk FFaaiilleedd aatt SStteepp 22 -->> TThhee MMSS-­SSQQLL ddaattaabbaassee iiss ccoorrrruupptt..FFoolllloowwiinngg LLooggiinnss aarree mmiissssiinngg iinn DDaattaabbaassee -->>NNTT AAUUTTHHOORRIITTYY\\NNEETTWWOORRKK SSEERRVVIICCEE.. PPlleeaassee ccoonnttaacctt SSttrryykkeerr ssuuppppoorrtt ttoo rreebbuuiilldd..
AApppplliiccaattiioonn HHeeaalltthh CChheecckk FFaaiilleedd aatt SStteepp 33 -->> iiBBeeddSSeerrvveerr iiss uunnaabbllee ttoo aacccceessss SSEEWWSSII.. PPlleeaassee rreessttaarrtt tthhee WWoorrlldd WWiiddee WWeebb PPuubblliisshhiinngg SSeerrvviiccee aanndd SSttrryykkeerrSSEEWWSSIIHHeeaarrttbbeeaattsseerrvviiccee.. AAfftteerr rreessttaarrtt,, iiff tthhee mmeessssaaggee ccoonnttiinnuueess,, pplleeaassee ccoonnttaacctt SSttrryykkeerr ssuuppppoorrtt..
MasterSubscriptionInfo table doesn't exist
LOCAL SERVICE doesn't exist
NETWORK SERVICE doesn't exist
EEmmaaiill SSuubbjjeecctt:: SSttrryykkeerr iiBBeedd SSeerrvveerr UUrrggeenntt AAlleerrtt
EEmmaaiill BBooddyy:: AApppplliiccaattiioonn HHeeaalltthh CChheecckk FFaaiilleedd aatt SStteepp 33 -->> iiBBeeddSSeerrvveerr iiss uunnaabbllee ttoo aacccceessss SSEEWWSSII.. PPlleeaassee rreessttaarrtt tthhee WWoorrlldd WWiiddee WWeebb PPuubblliisshhiinngg SSeerrvviiccee aanndd SSttrryykkeerrSSEEWWSSIIHHeeaarrttbbeeaattsseerrvviiccee.. AAfftteerr rreessttaarrtt,, iiff tthhee mmeessssaaggee ccoonnttiinnuueess,, pplleeaassee ccoonnttaacctt SSttrryykkeerr ssuuppppoorrtt..
AApppplliiccaattiioonn HHeeaalltthh CChheecckk FFaaiilleedd aatt SStteepp 44 -->> SSttrryykkeerrSSEEWWSSIIHHeeaarrttbbeeaattsseerrvviiccee iiss nnoott rruunnnniinngg.. PPlleeaassee rreessttaarrtt SSttrryykkeerrSSEEWWSSIIHHeeaarrttbbeeaattsseerrvviiccee.. AAfftteerr rreessttaarrtt,, iiff tthhee mmeessssaaggee ccoonnttiinnuueess,, pplleeaassee ccoonnttaacctt SSttrryykkeerr ssuuppppoorrtt..
Both SEWSI and HB services are down
5212-309-001 Rev A.1 69 EN
Page 74
EEmmaaiill iinnffoo SScceennaarriioo
EEmmaaiill SSuubbjjeecctt:: SSttrryykkeerr iiBBeedd SSeerrvveerr UUrrggeenntt AAlleerrtt
EEmmaaiill BBooddyy:: AApppplliiccaattiioonn HHeeaalltthh CChheecckk FFaaiilleedd aatt SStteepp 22 -->> TThhee MMSS-­SSQQLL ddaattaabbaassee iiss ccoorrrruupptt..FFoolllloowwiinngg TTaabblleess aarree mmiissssiinngg iinn DDaattaabbaassee -->>MMaasstteerrSSuubbssccrriippttiioonnIInnffoo .. PPlleeaassee ccoonnttaacctt SSttrryykkeerr ssuuppppoorrtt ttoo rreebbuuiilldd..
AApppplliiccaattiioonn HHeeaalltthh CChheecckk FFaaiilleedd aatt SStteepp 33 -->> iiBBeeddSSeerrvveerr iiss uunnaabbllee ttoo aacccceessss SSEEWWSSII.. PPlleeaassee rreessttaarrtt tthhee WWoorrlldd WWiiddee WWeebb PPuubblliisshhiinngg SSeerrvviiccee aanndd SSttrryykkeerrSSEEWWSSIIHHeeaarrttbbeeaattsseerrvviiccee.. AAfftteerr rreessttaarrtt,, iiff tthhee mmeessssaaggee ccoonnttiinnuueess,, pplleeaassee ccoonnttaacctt SSttrryykkeerr ssuuppppoorrtt..
AApppplliiccaattiioonn HHeeaalltthh CChheecckk FFaaiilleedd aatt SStteepp 44 -->> SSttrryykkeerrSSEEWWSSIIHHeeaarrttbbeeaattsseerrvviiccee iiss nnoott rruunnnniinngg.. PPlleeaassee rreessttaarrtt SSttrryykkeerrSSEEWWSSIIHHeeaarrttbbeeaattsseerrvviiccee.. AAfftteerr rreessttaarrtt,, iiff tthhee mmeessssaaggee ccoonnttiinnuueess,, pplleeaassee ccoonnttaacctt SSttrryykkeerr ssuuppppoorrtt..
EEmmaaiill SSuubbjjeecctt:: SSttrryykkeerr iiBBeedd SSeerrvveerr UUrrggeenntt AAlleerrtt
EEmmaaiill BBooddyy:: AApppplliiccaattiioonn HHeeaalltthh CChheecckk FFaaiilleedd aatt SStteepp 33 -->> iiBBeeddSSeerrvveerr iiss uunnaabbllee ttoo aacccceessss SSEEWWSSII.. PPlleeaassee rreessttaarrtt tthhee WWoorrlldd WWiiddee WWeebb PPuubblliisshhiinngg SSeerrvviiccee aanndd SSttrryykkeerrSSEEWWSSIIHHeeaarrttbbeeaattsseerrvviiccee.. AAfftteerr rreessttaarrtt,, iiff tthhee mmeessssaaggee ccoonnttiinnuueess,, pplleeaassee ccoonnttaacctt SSttrryykkeerr ssuuppppoorrtt..
AApppplliiccaattiioonn HHeeaalltthh CChheecckk FFaaiilleedd aatt SStteepp 44 -->> SSttrryykkeerrSSEEWWSSIIHHeeaarrttbbeeaattsseerrvviiccee iiss nnoott rruunnnniinngg.. PPlleeaassee rreessttaarrtt SSttrryykkeerrSSEEWWSSIIHHeeaarrttbbeeaattsseerrvviiccee.. AAfftteerr rreessttaarrtt,, iiff tthhee mmeessssaaggee ccoonnttiinnuueess,, pplleeaassee ccoonnttaacctt SSttrryykkeerr ssuuppppoorrtt..
Both SEWSI and HB services are down and
MasterSubscriptionInfo table doesn't exist
HB is down
EEmmaaiill SSuubbjjeecctt:: SSttrryykkeerr iiBBeedd SSeerrvveerr UUrrggeenntt AAlleerrtt
EEmmaaiill BBooddyy:: iiBBeedd SSeerrvveerr iiss uunnaabbllee ttoo ccoommmmuunniiccaattee wwiitthh SSttrryykkeerr
No devices are detected
cclliieennttss.. PPlleeaassee cchheecckk tthhee sseerrvveerr//nneettwwoorrkk ccoonnffiigguurraattiioonn.. IIff tthhee mmeessssaaggee ccoonnttiinnuueess,, pplleeaassee ccoonnttaacctt SSttrryykkeerr TTeecchhnniiccaall SSuuppppoorrtt..
EN 70 5212-309-001 Rev A.1
Page 75
EEmmaaiill iinnffoo SScceennaarriioo
EEmmaaiill SSuubbjjeecctt:: SSttrryykkeerr iiBBeedd SSeerrvveerr UUrrggeenntt AAlleerrtt
EEmmaaiill BBooddyy:: TToottaall nnuummbbeerr ooff cclliieennttss ccoonnnneecctteedd ttoo iiBBeedd SSeerrvveerr hhaass ggoonnee bbeellooww <<ccoonnffiigguurreedd tthhrreesshhoolldd>>.. PPlleeaassee cchheecckk tthhee sseerrvveerr//nneettwwoorrkk ccoonnffiigguurraattiioonn.. IIff tthhee mmeessssaaggee ccoonnttiinnuueess,, pplleeaassee ccoonnttaacctt SSttrryykkeerr TTeecchhnniiccaall SSuuppppoorrtt..
EEmmaaiill SSuubbjjeecctt:: SSttrryykkeerr iiBBeedd SSeerrvveerr UUrrggeenntt AAlleerrtt
EEmmaaiill BBooddyy:: TThhee ffoolllloowwiinngg eerrrroorrss wwiitthh TThhiirrdd PPaarrttyy CCoommmmuunniiccaattiioonn hhaass bbeeeenn rreeccoorrddeedd iinn llaasstt 11 hhoouurr.. <<EErrrroorr>>
Number of connected devices drops below the configured
threshold value
Error with third party
5212-309-001 Rev A.1 71 EN
Page 76
Stryker Medical 3800 E. Centre Avenue Portage, MI 49002 USA
5212-309-001 Rev A.1 WCR: AA.9
2019/11
Loading...