Stryker 4700-009-001B Sigma Delta Vaporizer User Instruction Manual

Adelr Maternity Bed
Model 4700 & 5012 OPERATIONS MANUAL
For Parts or Technical Assistance 1−800−327−0770
Table of Contents
Introduction 2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Specifications 2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Warning / Caution / Note Definition 2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Safety Tips And Guidelines 4. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Unpacking Instructions and Set−Up Procedures 5. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Siderail Operation Guide 7−9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Pendant Operation Guide 10−12. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Operating Bed Features
Using the Brake And Steer Pedals 13. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Using the Manual Motor Crank 13. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Activating Instant CPR 13. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Locking Out the Siderail Controls 13. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Removing the Head Board 13. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Using the Labor Grips 14. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Adjustable Seat Depth 14. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Nurse Call Battery 14. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Night Light Usage 15. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Operating the Foot Section 15. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Lumbar and Perineal Pillows 15. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Patient Positioning
Birthing Bed Positioning 16. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Birthing Chair Positioning With Labor Bar 16. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Delivery Table Positioning 16. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Trendelenburg Positioning 16. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Pelvic Tilt 16. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Using Accessories
Using Leg Supports 17. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Using the Labor Bar 17. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Using the Arm Board 17. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Using the Permanently Attached I.V. Pole 18. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Cleaning 19. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Preventative Maintenance Checklist 20. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Limited Warranty
Obtaining Parts and Service 21. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Supplemental Warranty Coverage 21. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Return Authorization 22. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Freight Damage Claims 22. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Introduction
INTRODUCTION
This manual is designed to assist you with the operation of the Stryker Adelr Maternity Bed. Read it thorough­ly before using the equipment or beginning any maintenance on it.
SPECIFICATIONS
Maximum Weight Capacity 500 pounds (227 kilograms) Bed Length/Width 90 x 41 (with siderails up) 37 (with siderails down)
229 cm x 104 cm (with siderails up) 94 cm (with siderails down) Bed Height (to top of seat) Low 19 (49 cm), High 37 (95 cm) (w/8 Casters) Mattress Size Head 49.5 (126 cm) x 33 (84 cm) x 5 (13 cm), Foot 31 (79 cm) x
30 (77 cm) x 3 (8 cm), Patient Sleep Surface 81 (206 cm) Critical Angles Maximum Elevation Head 70_, Trendelenburg 8_ BreakAway Point from Wall 60 (152 cm) Electrical Standard 4 motor function: HeadBedFootTrendelenburg
120 VAC, 60 Hz, 6 Amp 105 to 135 VAC operating range.
Optional: 230 VY, 50/60 Hz, 6 A 220 to 240 VAC operating range
Current leakage less than 100 microamperes.
Hospital grade plug and 3wire heavy duty cord.
Compatible with nonflammable anesthetic agents and oxygen by nasal
catheter or mask. Caster Size 8 Standard; 6 Optional
Stryker reserves the right to change specifications without notice.
WARNING / CAUTION / NOTE DEFINITION
The words WARNING, CAUTION and NOTE carry special meanings and should be carefully reviewed.
WARNING
The personal safety of the patient or user may be involved. Disregarding this information could result in injury to the patient or user.
CAUTION
These instructions point out special procedures or precautions that must be followed to avoid damaging the equipment.
NOTE
This provides special information to make maintenance easier or important instructions clearer.
To ensure its proper use and the safety of patients and staff, the Stryker Adelr Maternity Bed has been marked with the following caution and warning labels:
DANGER Explosion Hazard do not use in the presence of flammable anesthetics. CAUTION This unit is equipped with a hospital grade attachment plug. Grounding reliability can be
achieved only when equipment is connected to equivalent receptacle. CAUTION Electrical shock hazard. Do not remove cover panels. Refer all servicing to qualified personnel. CAUTION Disconnect power cord before using the back up hand crank. Remove hand crank from back
up outlet before connecting power cord.
2
Bed Symbols
Warning, Refer to Service/Maintenance Manual
~
Alternating Current
Type B Equipment: equipment providing a particular degree of protection against elec­tric shock, particularly regarding allowable leakage current and reliability of the protec­tive earth connection.
Class 1 Equipment: equipment in which protection against electric shock does not rely on BASIC INSULA TION only, but which includes an additional safety precaution in that means are provided for the connection of the EQUIPMENT to the protective earth con­ductor in the fixed wiring of the installation in such a way that ACCESSIBLE METAL PARTS cannot become live in the event of a failure of the BASIC INSULATION.
Mode of Operation: Continuous
IPX4: Protection from liquid splash
Dangerous Voltage Symbol
Protective Earth Terminal
Potential Equalization Symbol
Medical Equipment Classified by Underwriters Laboratories Inc. with Respect to Elec­tric Shock, Fire, Mechanical and Other Specified Hazards Only in Accordance with UL 26011 and CAN/CSA C22.2 No. 601.1
3
Introduction
SAFETY TIPS AND GUIDELINES
Before operating the Stryker Adel Maternity Bed, it is important to read and understand all information in this manual. Carefully read and strictly follow the safety guidelines listed on this page. It is important that all users have been trained and educated on the inherent hazards associated with the usage of electric beds.
WARNING
S The Stryker Adel Maternity Bed is equipped with a hospital grade plug for protection against shock haz-
ard. It must be plugged directly into a properly grounded threeprong receptacle. Grounding reliability can be achieved only when a hospital grade receptacle is used.
S Powered bed mechanisms can cause serious injury. Operate bed only when all persons are clear of the
mechanisms.
S To help reduce the number and severity of falls by patients, always leave the bed in the lowest position
when the patient is unattended.
S Leave the siderails fully up and locked when the patient is unattended. When raising the siderails, listen
for the click that indicates the siderail has locked into the up position. Pull firmly on the siderail to ensure it is locked into position. Siderails are not intended to be a patient restraint device. It is the responsibility of the attending medical personnel to determine the degree of restraint necessary to ensure a patient will remain safely in bed.
S Always apply the caster brakes when a patient is getting on or off the bed. Always keep the caster brakes
applied when a patient is on the bed (except during transport). Serious injury could result if the bed moves while a patient is getting in or out of bed. After the brake pedal is applied, push on the bed to ensure the brakes are locked. When moving the bed, toggle the steer pedal to put the bed in the steer mode. This locks the swivel motion of the right foot end caster and makes the bed easier to move.
S The instant CPR release is for emergency use only. When activating the instant CPR, all persons and
equipment must be removed from the area below and around the Fowler (back rest) section of the bed or serious personal injury or damage to the equipment could occur.
S When large spills occur in the area of the circuit boards, 110 volt cables and motors, immediately unplug
the bed power cord from the wall socket. Remove the patient from the bed and clean up the fluid. Have maintenance completely check the bed. Fluids can have an affect on operational capabilities of any elec­trical product. DO NOT put the bed back into service until it is completely dry and has been thoroughly tested for safe operation.
S Do not steam clean or hose off the bed. Do not immerse any part of the bed. The internal electric parts
may be damaged by exposure to water. Hand wash all surfaces of the bed with warm water and mild detergent. Dry thoroughly. Inspect the mattress cover after each use. Discontinue use if any cracks or rips are found in the cover which may allow fluids to enter the mattress. Exposure to fluids may cause injury to patient and/or user.
S Preventative maintenance should be performed at a minimum of annually to ensure all bed features are
functioning as designed. Close attention should be given to safety features including, but not limited to:
S Safety side latching mechanisms, S Caster braking system, S Frayed electrical cords and components, S Leakage current 100 MA max. S No controls or cabling entangled in bed mechanisms, S All electrical controls return to off or neutral position when released, S For additional maintenance information, refer to your maintenance manual.
S Always unplug bed during service or cleaning. When working under the bed with the bed in the high posi-
tion, always place blocks under the litter frame and set the brakes to prevent injury in case the Bed Down switch is accidently pressed.
S Unplug the bed power cord from the wall socket before using the hand crank. Remove the hand crank
before plugging the power cord into the wall socket or personal injury could result.
4
Unpacking and Initial SetUp Procedures
UNPACKING INSTRUCTIONS
Refer to unpacking instructions attached to the bed inside the crate.
SETUP PROCEDURES
It is important that the Stryker Adelr Maternity Bed is working properly before it is put into service. The follow­ing list will help ensure that each part of the bed is tested.
S Plug the bed into a properly grounded, hospital grade wall receptacle.
WARNING
The Stryker Adelr Maternity Bed is equipped with a hospital grade plug for protection against shock hazard. It must be plugged directly into a properly grounded threeprong receptacle. Grounding reliability can be achieved only when a hospital grade receptacle is used.
S Ensure the siderails raise and lower smoothly and lock in the up position. S Ensure that the brakes hold when the brake pedal is engaged. S Test each function on the (optional) hand pendant to ensure that each function is working properly.
Beds equipped with nurse communication siderail control option only: S Plug the interface cable into the 37 pin connector in the litter frame at the head end of the bed, and into
the ”Patient Station”, ”Head Wall”, ”Docker Station”, or equivalent (whichever applies).
S Run through each function on the siderail control panels to ensure that each function is working properly.
NOTE
To prevent a low battery condition when the bed is not plugged in, position the cord out switch at the head end of the bed to the off position. The switch is identified by the label shown below. If the switch is not posi­tioned as shown below and the bed power cord and pendant cord are unplugged, the life of the backup bat­tery will be significantly reduced. If the siderail lights LED (located on the outside of both siderails) is flashing, the Nurse Call battery needs to be replaced. The battery is located on the patients right side at the head end of the bed. No tools are required to replace the battery. Unplug the bed power cord from the wall socket and replace the battery. After replacing the battery, verify the siderail lights LED is no longer flashing and operates normally when the dif fer­ent light settings are selected.
5
Bed Illustration
WARNING
Potential pinch points
6
Siderail Operation Guide
OPERATING SIDERAILS
To engage the head end siderail, grasp the rail and swing it upward to full height. When the siderail is being raised, it does not lock in the intermediate position. To lower the siderail, push in the release handle and rotate the siderail until it locks in the intermediate position. To lower the siderail fully, push in the release handle again and rotate the siderail until it is completely lowered.
NOTE
To activate the siderail bypass mechanism, the rail must be fully lowered. If the rail is not completely lowered, the siderail will lock in the intermediate position when it is raised.
USING THE SIDERAIL CONTROL PANEL LIGHTS
The bed is equipped with lights to illuminate the head end siderail control panels and the red nurse call switches. Four settings are available for the control panel lights. Push once to turn on the lights at low intensi­ty, again to change to high intensity, a third time to leave just the nurse call light on, and a fourth time to turn all lights off.
INSIDE SIDERAIL FUNCTION GUIDE
1.
1.
1. Push to raise Fowler (back rest).
2. Push to raise foot section.
3. Push to lower Fowler (back rest).
4. Push to lower foot section.
1. Push to activate Nurse Call
NOTE
Yellow LED will light when button is pushed. Red LED will light with Nurse Station acknowledgement.
1. Push to turn TV or radio on and to select a channel.
2.
3.
2. Push to increase volume.
3. Push to decrease volume.
7
Siderail Operation Guide
INSIDE SIDERAIL FUNCTION GUIDE (CONTINUED)
1.
2.
3.
1.
2.
1. Push to turn the television on and to select a channel.
2. Push to increase the television volume.
3. Push to decrease the television volume.
1. Push to turn the room light on. Push again to turn off.
2. Push to turn the bed overhead light on. Push again to turn off.
1. Push for more support to the patient’s lower back.
2. Push for less support to the patients lower back.
3. Push for more support to the patient’s seat section.
4. Push for less support to the patients seat section.
5. Push to change the channel on the television.
6. Push to change the channel on the television.
8
Siderail Operation Guide
OUTSIDE SIDERAIL FUNCTION GUIDE
1. Push to raise Fowler (back rest).
2. Push to raise bed height.
3. Push to raise foot section.
4. Push to lower Fowler (back rest).
5. Push to lower bed height.
6. Push to lower foot section.
1. Push to lower head end of bed (Trendelenburg position). (This function is also used for pelvic tilt).
2. Push to toggle siderail lights to different settings: A. LOW B. HIGH C. NURSE CALL ONLY D. OFF
3. Push to raise head end of bed and/or return bed to level.
4. Push to lock out all bed motion. Push again to unlock.
(LED will light when bed motion is locked.)
5. Push to activate Nurse Call.
1. Push to decrease seat depth.
2. Push to increase seat depth.
NOTE: The Fowler (back rest) must be raised until the “EN- ABLED LED lights (a minimum of 35_) before the depth can be changed.
9
Communication Pendant Operation Guide
10
Motion Pendant Operation Guide
11
Motion Pendant with Nurse Call Operation Guide
12
Operating Bed Features
USING THE BRAKE AND STEER PEDALS
The brake pedals are located at the center of the base frame on both sides of the bed. T o engage the brakes, depress the pedal to the brake position. To disengage the brakes, depress the pedal to the neutral position. The steer pedal is located at the head end of the bed. To engage the steering function, depress the pedal to the ON position. To disengage the steering function, depress the pedal to the OFF position.
BRAKE PEDAL
STEER PEDAL
USING THE MANUAL MOTOR CRANK
WARNING
To avoid serious personal injury or damage to the equipment, unplug the bed power cord from the wall recep­tacle before using the hand crank and remove the hand crank from the motor before plugging in the power cord.
If there is a loss of electrical power to the bed, there is a manual motor crank stored in clips on the head end of the bed. Center the notch in the end of the crank on the roll pin on the bed motor assembly or the Fowler motor assembly and crank the bed to the desired position.
ACTIVATING INSTANT CPR
The CPR release lever is located at the head section on both sides of the bed. To activate the CPR release, grasp the lever and squeeze tightly. The Fowler (back rest) will lower to the lowest position instantly.
WARNING
The instant CPR release is for emergency use only. Before activating the instant CPR, verify all persons and equipment are away from the area below and around the Fowler (back rest) section of the bed or serious personal injury or damage to the equipment could occur.
LOCKING OUT THE SIDERAIL CONTROLS
The lock out feature disables the siderail bed motion controls. Lock out buttons are located on the outside of both siderails. Lock out the bed mo­tion controls by depressing the button once. The bed motion lockout LED will light. Reactivate the controls by pressing the button again.
13
Operating Bed Features
REMOVING THE HEAD BOARD
T o remove the head board, lift it straight up and of f the bed. To replace the head board, align the plastic inserts on the bottom of the head board with the pins on the bed and lower the head board until it completely engages the pins.
USING THE LABOR GRIPS
To position the labor grips, grasp the handle and pull the grip out and up above the mattress until the mecha­nism locks into position. To lower the grips, pull the lever to release the grip and, as it begins to rotate, release the lever and lower the grip to the stored position.
ADJUSTABLE SEAT DEPTH (OPTIONAL)
To accommodate patients of different sizes, the depth of the seat can be adjusted up to 4 inches (from 12.5 to 8.5 inches).
S The Fowler (back rest) must be raised to a minimum of 35_ before the seat depth can be changed. Once
the Fowler (back rest) is elevated to 35_, the “ENABLED” LED on the outside of the siderail will light and the seat depth can be changed. When the Fowler (back rest) is lowered (either electrically or using the emergency CPR release), the bed will automatically adjust the seat depth to 12.5”.
S To reduce seat depth, depress the button on the siderail or the hand pendant to move the Fowler
(back rest) toward the foot end of the bed. Release the button when the desired seat depth is attained.
S To increase seat depth, press the button to move the Fowler (back rest) toward the head end
of the bed.
NURSE CALL BATTERY
To prevent a low battery condition when the bed is not plugged in, position the cord out switch at the head end of the bed to the off position. The switch is identified by the label shown below. If the switch is not posi­tioned as shown below and the bed power cord and pendant cord are unplugged, the life of the backup bat­tery will be significantly reduced. If the siderail lights LED (located on the outside of both siderails) is flashing, the Nurse Call battery needs to be replaced. The battery is located on the patients right side at the head end of the bed. No tools are required to replace the battery. Unplug the bed power cord from the wall socket and replace the battery. After replacing the battery, verify the siderail lights LED is no longer flashing and operates normally when the dif fer­ent light settings are selected.
14
Operating Bed Features
NIGHT LIGHT USAGE
The bed is equipped with a night light to illuminate the floor area around the bed. There is a switch on the light to turn it on and off.
OPERATING THE GLIDEAWAY FOOT SECTION/ATTACHABLE CALF SUPPORTS
S
Rotate the foot rest into position by pulling it out and up over the foot end mattress until it clicks into place. Ensure the foot rest is securely locked in the upright position by grasping and attempting to move it.
S After they are in position, the foot rests are adjustable six ways:
1. Abduct/Adduct (pivoting) Pivoting in and out allows the clinician to position the patient’s feet and legs at the desired width. Pull down on the knob located below the foot mattress. Pivot the foot rest into the desired position and release the knob to lock the foot rest into place.
2. In/Out The in/out motion allows patients of different heights to be accommodated comfortably. Pull the lever marked with the PULL label to increase the distance of the foot rest from the Perineal edge. Release the lever when the foot rest is in the desired position.
3. Height The foot end high/low adjustment allows the foot end of the bed to be positioned 6” below the top of the seat mattress to accommodate different sized patients when the foot rest or calf supports are being used.
S To store the foot rests, loosen the ball screws on the selfstoring calf supports (if used). Ensure the foot
rests are pointing toward the end of the bed. Pull the lever marked “PULL” located at the bottom of the foot rest support and the foot rest will glide under the foot mattress.
REMOVING THE FOOT SECTION MATTRESS
Before removing the foot mattress, put the foot rests into position above the foot mattress. Place the patients feet in the contoured foot rests. To remove the foot mattress, pull up on the red release handle on either side of the mattress and slide the mattress straight off the bed.
NOTE
If the foot section mattress is placed with the perineal edge toward the floor, it will stand independently. To reinstall the foot section of the bed, set the foot mattress back on the bed and slide it straight in until it locks
into place. Pull on the handles to ensure it is securely installed.
OPTIONAL LUMBAR PILLOW AND PERINEAL WEDGE
Use the onetouch lumbar and seat buttons on the siderail to adjust the amount of support given to the
S
patients lumbar and seat areas if the bed is equipped with this option.
NOTE
Be sure the locating pins on the underside of the mattress are inserted in the holes in the steel seat support.
15
Operating Bed Features
BIRTHING BED POSITIONING
Position the patient’s feet in the foot rests.
S
S Raise or lower the foot rests to a position comfortable for the patient. S Remove the foot section and tuck the drape into the drainage pan. S Raise the bed to a comfortable height by pressing the BedUp control on the siderail and position the
patients perineum out and over the edge of the seat section.
DELIVERY TABLE POSITIONING
Slide the patient down to the perineal edge.
S
S Lower the foot section to its lowest position. S Position the attached calf supports and place the patients legs in the supports.
TRENDELENBURG POSITIONING
Lower the Fowler (back rest) by pressing the control on the siderail or by squeezing the CPR lever. Press the Trendelenburg button on the siderail or the control pendant to lower the head end of the bed to the desired angle.
PELVIC TILT
Use the Trendelenburg button on the siderail or the control pendant to provide more pelvic tilt and comfort for the patient when the Fowler (back rest) is raised.
16
Using Optional Accessories
USING LEG SUPPORTS
To use fulllength leg supports on the bed, insert the leg support bar into the receptacle near the labor grips at the perineal edge.
USING THE LABOR BAR
To use the labor bar, insert it into the holes on each side of the seat section, under the mattress at the perineal edge. The labor bar can be used by the patient to aid with various birthing positions such as squatting or kneeling.
USING THE ARM BOARD
The arm board is held in place between the mattress and the frame. The arm board is provided with a remov­able mattress. The weight of the patient retains the arm board in the desired position.
17
Using Optional Accessories
OPERATING OPTIONAL 3STAGE PERMANENTLY ATTACHED I.V. POLE
A
E
C
C
B
D
A
E
DET AIL OF I.V. POLE LATCH
DET AIL OF I.V. POLE GRIP
NOTE
The 3stage permanently attached I.V. pole is an option and can only be installed at the head end of the bed.
To use the 3stage permanently attached I.V. pole:
1. Lift and pivot the pole from the storage position and push down until it is locked into the receptacle.
B
D
2. To raise the height of the pole, pull up on the telescoping portion (A) until it locks into place at its fully raised
position.
3. For a higher I.V. pole, pull up on section (B). Release section (B) at any desired height and it will lock
into place.
4. Rotate the I.V. hangers (C) to the desired position and hang the I.V. bags.
5. To lower the I.V. pole, push up on the red portion of grip (D) while holding onto section (B) until it lowers.
Turn latch (E) clockwise until section (A) lowers.
CAUTION
To avoid damage, the weight of the I.V. bags should not exceed 40 pounds. To avoid damage while transporting the bed, verify the I.V. pole is at a low enough height to allow it to pass
safely through door openings.
18
Cleaning
WARNING
Hand wash all surfaces of the bed with warm water and mild detergent. Dry thoroughly. DO NOT STEAM CLEAN, PRESSURE WASH, HOSE OFF OR ULTRASONICALLY CLEAN. Using these methods of cleaning is not recommended and may void this product’s warranty. Inspect the mattress cover after each use. Discontinue use if any cracks or rips are found in the cover which may allow fluids to enter the mattress. Exposure to fluids may cause injury to patient and/or user.
In general, when used in those concentrations recommended by the manufacturer, either phenolic type or quaternary type disinfectants can be used. Iodophor type disinfectants are not recommended for use be­cause staining may result. The following products have been tested and have been found not to have a harm­ful effect WHEN USED IN ACCORDANCE WITH MANUFACTURERS RECOMMENDED DILUTION.*
TRADE NAME
A33 Quaternary Airwick (Professional Products Division) 2 ounces/gallon A33 (dry) Quaternary Airwick (Professional Products Division) 1/2 ounce/gallon Beaucoup Phenolic Huntington Laboratories 1 ounce/gallon Blue Chip Quaternary S.C. Johnson 2 ounces/gallon Elimstaph Quaternary Walter G. Legge 1 ounce/gallon Franklin Phenomysan F2500 Phenolic Purex Corporation 1 1/4 ounce/gallon Franklin Sentinel Quaternary Purex Corporation 2 ounces/gallon Galahad Phenolic Puritan Churchill Chemical Company 1 ounce/gallon HiTor Quaternary Huntington Laboratories 1/2 ounce/gallon LPH Phenolic Vestal Laboratories 1/2 ounce/gallon Matar Phenolic Huntington Laboratories 1/2 ounce/gallon Omega Quaternary Airwick (Professional Products Division) 1/2 ounce/gallon Quanto Quaternary Huntington Laboratories 1 ounce/gallon Sanikleen Quaternary West Chemical Products 2 ounces/ gallon Sanimaster II Quaternary Service Master 1 ounce/gallon Vesphene Phenolic Vestal Laboratories 1 1/4 ounce/gallon Wexcide Phenolic Wexford Labs 1/2 ounce/gallon
DISINFECTANT
TYPE
MANUFACTURER
*MANUFACTURERS
RECOMMENDED
DILUTION
Quaternary Germicidal Disinfectants, used as directed, and/or Chlorine Bleach products, typically 5.25% So ­dium Hypochlorite in dilutions ranging between 1 part bleach to 100 parts water, and 2 parts bleach
to 100 parts water are not considered mild detergents. These products are corrosive in nature and may cause damage to your stretcher if used improperly. If these types of products are used to clean
Stryker patient handling equipment, measures must be taken to insure the stretchers are rinsed with clean water and thoroughly dried following cleaning. Failure to properly rinse and dry the stretchers will leave a cor­rosive residue on the surface of the stretcher, possibly causing premature corrosion of critical components.
NOTE
Failure to follow the above directions when using these types of cleaners may void this products warranty.
REMOVAL OF IODINE COMPOUNDS
This solution may be used to remove iodine stains from mattress cover and foam footrest pad surfaces.
1. Use a solution of 12 tablespoons Sodium Thiosulfate in a pint of warm water to clean the stained area.
Clean as soon as possible after staining occurs. If stains are not immediately removed, allow solution to soak or stand on the surface.
2. Rinse surfaces which have been exposed to the solution in clear water before returning bed to service.
19
Preventative Maintenance Checklist
All fasteners secure All welds intact, not cracked or broken No bent or broken tubing or sheet metal No debris in casters All casters secure and swivel properly Engage brake pedal and push on the bed to ensure all casters lock securely Steer caster latches properly Siderails move and latch properly Fowler operates properly Fowler (back rest) Slide operates properly Bed Up/Down operates properly Foot operates properly Trendelenburg operates properly I.V. pole intact and operating properly No rips or cracks in mattress cover Lubricate where required Replace Nurse Call 9V battery (annually) Power cord not frayed No cables worn or pinched All electrical connections tight All grounds secure to the frame Ground impedance not more than 100 milliohms Current leakage not more than 100 microamps
Serial No.
NOTE
Preventative maintenance should be performed at a minimum of annually. A preventative maintenance pro­gram should be established for all Stryker Medical equipment. Preventative maintenance may need to be performed more frequently based on the usage level of the product.
20
Warranty
Limited Warranty:
Stryker Medical Division, a division of Stryker Corporation, warrants to the original purchaser that its products should be free from defects in material and workmanship for a period of one (1) year after date of delivery. Strykers obligation under this warranty is expressly limited to supplying replacement parts and labor for, or replacing, at its option, any product which is, in the sole discretion of Stryker, found to be defective. Stryker warrants to the original purchaser that the frame and welds on its beds will be free from structural defects for as long as the original purchaser owns the bed. If requested by Stryker, products or parts for which a warranty claim is made shall be returned prepaid to Strykers factory. Any improper use or any alteration or repair by others in such manner as in Strykers judgement affects the product materially and adversely shall void this warranty. Any repair of Stryker products using parts not provided or authorized by Stryker shall void this warranty. No employee or representative of Stryker is authorized to change this warranty in any way.
Stryker Medical beds are designed for a 15 year expected life under normal use conditions and appropriate periodic maintenance as described in the maintenance manual for each device.
This statement constitutes Strykers entire warranty with respect to the aforesaid equipment. STRYKER MAKES NO OTHER WARRANTY OR REPRESENTATION, EITHER EXPRESSED OR IMPLIED, EXCEPT AS SET FORTH HEREIN. THERE IS NO WARRANTY OF MERCHANTABILITY AND THERE ARE NO WARRANTIES OF FITNESS FOR ANY PARTICULAR PURPOSE. IN NO EVENT SHALL STRYKER BE LIABLE HEREUNDER FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING FROM OR IN ANY MANNER RELATED TO SALES OR USE OF ANY SUCH EQUIPMENT.
To Obtain Parts and Service:
Stryker products are supported by a nationwide network of dedicated Stryker Field Service Representatives. These representatives are factory trained, available locally, and carry a substantial spare parts inventory to minimize repair time. Simply call your local representative, or call Stryker Customer Service at (800) 3270770.
Service Contract Coverage:
Stryker has developed a comprehensive program of service contract options designed to keep your equip­ment operating at peak performance at the same time it eliminates unexpected costs. We recommend that these programs be activated before the expiration of the new product warranty to eliminate the potential of additional equipment upgrade charges.
A SERVICE CONTRACT HELPS TO:
S Ensure equipment reliability S Stabilize maintenance budgets S Diminish downtime S Establish documentation for JCAHO S Increase product life S Enhance tradein value S Address risk management and safety
21
Warranty
Stryker offers the following service contract programs:
SPECIFICATIONS GOLD SILVER PM* ONLY
Annually scheduled preventative maintenance X X All parts,** labor, and travel X X Unlimited emergency service calls X X Priority one contact; two hour phone response X X X Most repairs will be completed within 3 business days X X JCAHO documentation X X X Onsite log book w/ preventative maintenance & emergency service records X Factorytrained Stryker Service Technicians X X X Stryker authorized parts X X X End of year summary X Stryker will perform all service during regular business hours (95) X X X
* Replacement parts and labor for products under PM contract will be discounted.
** Does not include any disposable items, I.V. poles (except for Stryker HD permanent poles), mattresses, or damage re-
sulting from abuse.
Stryker Medical also offers personalized service contracts.
Pricing is determined by age, location, model and condition of product.
For more information on our service contracts,
please call your local representative or call (800) 327−0770 (option #2).
Return Authorization:
Merchandise cannot be returned without approval from the Stryker Customer Service Department. An autho­rization number will be provided which must be printed on the returned merchandise. Stryker reserves the right to charge shipping and restocking fees on returned items.
SPECIAL, MODIFIED, OR DISCONTINUED ITEMS NOT SUBJECT TO RETURN.
Damaged Merchandise:
ICC Regulations require that claims for damaged merchandise must be made with the carrier within fifteen (15) days of receipt of merchandise. DO NOT ACCEPT DAMAGED SHIPMENTS UNLESS SUCH DAMAGE IS NOTED ON THE DELIVERY RECEIPT AT THE TIME OF RECEIPT. Upon prompt notification, Stryker will file a freight claim with the appropriate carrier for damages incurred. Claim will be limited in amount to the actual replacement cost. In the event that this information is not received by Stryker within the fifteen (15) day period following the delivery of the merchandise, or the damage was not noted on the delivery receipt at the time of receipt, the customer will be responsible for payment of the original invoice in full.
Claims for any short shipment must be made within thirty (30) days of invoice.
International Warranty Clause:
This warranty reflects U.S. domestic policy. Warranty outside the U.S. may vary by country. Please contact your local Stryker Medical representative for additional information.
22
European Representative
Stryker EMEA RA/QA Director Stryker France ZAC Satolas Green Pusignan Av. De Satolas Green 69881 MEYZIEU Cedex France
6300 S. Sprinkle Road, Kalamazoo, MI 490019799
(800) 3270770
www.strykermedical.com
DH 10/03 4700009001 REV B
Loading...