Stryker 240-030-960 User Manual

VISION ELECT HDTV
26” High-Definition
LCD Monitor
REF 240-030-960
User Guide
E 570-104-0001moc.rekyrts.www11/8002
English
616.4 mm (24.27”) mm
mm
mm
mm
mm
mm
mm
mm
mm
3mm
mm
mm
100mm
552.1 mm (21.74”)
110.2 mm (4.34”) 3 mm (0.12”) 194 mm (7.64”)
48 mm (1.89”)
327 mm (12.91”)
53 mm (2.04”)
228.8 mm (9.01”)
3 mm (0.12”)
121.2 mm (4.77”)
428.8 mm (16.88”)
345.8 mm (13.61”)
100 mm (3.94”)
100 mm (3.94”)
EN-1
English
VISION ELECT HDTV
26” High-Definition
LCD Monitor
Table of Contents
Warnings and Cautions ...................................................................EN-2
Warnings......................................................................................EN-2
Cautions.......................................................................................EN-3
Symbol Definitions............................................ ... .... ... ... ... ... .... ... ... ..EN-5
Product Description and Intended Use ..........................................EN-6
Product Contents .........................................................................EN-6
System Interconnection ...............................................................EN-8
Power Connection........................................................................EN-8
Operating the Monitor......................................................................EN-9
Front Panel Controls....................................................................EN-9
Rear Panel Panel.......................................................................EN-10
Input Port Layout.... .... ... ... ... ... .... ... ... ... .... ... ... ... ..........................EN-10
Fiber Optic Module (optional) Installation ..................................EN-11
Input Selection List. .... ... ... ... ... .... ... ... ... .... ... ... ... .... ... ... ... ... .... ... ...EN-11
Standard On-Screen Display (OSD) Operation .........................EN-12
Stryker Camera Preset Modes...................................................EN-12
OSD Function Description .........................................................EN-13
Cleaning the Monitor.................. .... ... ................................ ... .... ... ...EN-14
Cleaning display plastic area .....................................................EN-14
Cleaning display filter area.........................................................EN-14
Troubleshooting.............................................................................EN-15
Technical Specifications......... ... .... ... ................................ ... .... ... ...EN-16
Classification and Approvals......................................................EN-17
Electromagnetic Compatibility......................................................EN-18
Warranty..........................................................................................EN-23
Service and Claims.................. ... .... ... ................................ ... .... ... ...EN-23
Declaration of Conformity .............................................................EN-24
EN-2
Warn ings and Cautions
Please read this manual and follow its instructions carefully. The words Warning, Caution, and Note carry special meanings and should be carefully reviewed:
Warning The personal safety of the patient or physician may be
involved. Disregarding this information could result in injury to the patient or physician.
Warning A lightning bolt within a triangle is intended to warn of
the presence of hazardous voltage. Refer all service to authorized personnel.
Caution Special service procedures or precautions must be
followed to avoid damaging the instrument.
Note Special information to make maintenance easier or
important information more clear.
Warnings
Warning To avoid potential serious injury to the user and the
patient, and/or damage to this device, the user must:
1. Restrict this device to use by, or on the order of, a physician, because of Federal law (United States of America).
2. Read the operating manual thoroughly and be familiar with its contents prior to using this equipment.
3. Carefully unpack the unit and check if any damage occurred during shipment.
4. Test this equipment prior to a surgical procedure. This monitor was fully tested at the factory before shipment.
5. Do not place the monitor or any other heavy object on the power cord. Damage to the cable can cause fire or electric shock.
6. Should any solid object or liquid fall into the panel, unplug the unit and have it checked by qualified personnel before operating it any further.
7. Unplug the unit if it is not to be used for an extended period of time. To discon nect the cord, unscrew plug first, then pull it out by the plug. Never pull the cord itself.
8. To avoid electric shock, avoid removing control unit covers.
EN-3
English
9. Do not attempt internal repairs or adjustments not specifically detailed in this operating manual.
10. Pay close attention to the care and cleaning instructions in this manual. A deviation may cause damage (refer to the Cleaning section).
11. Do not sterilize the monitor.
12. Use appropriate caution to prevent cont act with fluids if the unit is being used with a power supply in patient environments.
Cautions
1. The AC Adapter must be plugged in to a grounded power outlet.
2. Use only the proprietary VISION ELECT HDTV power supply (P/N 240-030-950, Manufacturer: JEC Korea, Model No: JMW1150KA2400F07) for the VISION ELECT HDTV monitor (model 240-030-960). Completely secure the connection between the DC power cord and the extension cord.
3. Grounding reliability can only be achieved when the equipment is connected to an equipment receptacle labeled “Hospital Only” or “Hospital Grade.”
4. To connect to an international po wer supply, use an attachment plug appropriate for the power outlet.
5. Turn power off when the unit is not in use.
6. Remove the power module and connection when transpo rting the unit.
7. Save the original carton and associated packing material. They will be useful should you have to transport or ship the unit.
8. Handle the monitor with care. Do not strike or scratch the screen.
9. Never operate the unit right after having been transported from a cold location directly to a warm location.
10. Do not expose the monitor to moisture or apply liquid cleaner s directly to the screen. Spray the cleaning solution into a sof t cloth and clean gently.
11. Allow adequate air circulation to prevent internal heat buildup.
12. Do not place the unit on surfaces (rugs, blankets, etc.) or near materials (curtains, draperies) that may block the vent ilation slot s. The monitor is cooled by natural convection and has no fan.
13. Do not install the unit near sunlight, excessive dust, mechanical vibration, or shock.
14. The unit is designed for operation in a horizontal position. Do not operate the unit in a vertical position.
EN-4
15. Keep the unit away from equipment that uses strong magnets (i.e., large loudspeakers).
16. Do not touch the patient with signal input or output connectors. Equipment with SIP/SOP connectors should either comply with IEC 60601-1 and/or IEC 60601-1-1 harmonized national standards or the combination should be evaluated.
17. CAUTION: Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.
Note This equipment has been tested and found to comply with
the limit for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses, and can radiate radio frequency energy an d, if not inst a lled and used in accordance with the instructions, may cause harmful interference to radio communications. There is no guarantee that interference will not occur in a particular installation, which can be determined by turning the equipment off and on. The user is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving device .
Increase the separation distance between the equipment.
Connect the equipment to an outlet on a circuit different from that to which the other device(s) are connected.
Consult the manufacturer or field service technician for help.
18. To ensure electromagnetic compatibility, refer to the “Electromagnetic Compatibility” section of this manual. The VISION ELECT HDTV (model 240-030-960) monitor must be installed and operated according to the EMC information provided in this manual.
The VISION ELECT HDTV (model 240-030-960) moni tor has been tested under the UL 60601-1 standard and is UL listed for medical application.
The warranty is void if any of these warnings or cautions are disregarded.
EN-5
English
Symbol Definitions
The following symbols appear on the product, its labeling, or the product packaging. Each symbol carries a special definition, as defined below:
Dangerous: High Voltage
Consult accompanying documents
Direct Current
Indicates protective earth ground
For indoor use only
DC power control switch
Serial Number
Top - Bottom
Fragile
Do not get wet
Maximum Stacking
Manufacturer
European Authorized Representative
Indicates proof of conformity to applicable European Economic Community Council directives and to harmonized standards published in the official journal of the European Communities
51LJ
Medical Equipment
E215822
Medical Equipment is in accordance with UL 60601-1 an d CAN/CSA C22.2 No.
601.1 in regards to electric shock, fire hazards, and mechanical hazards
Tested to comply with FCC Class B standards
IPX1
Degrees of protection against the ingress of water
This symbol indicates that the waste of electrical and electronic equipment must not be disposed as unsorted municipal waste and must be collected separately. Please contact the manufacturer or other authorized disposal company to decommission your equipment.
DC power on at locker switch
SN
3
EN-6
Product Description and Intended Use
The VISION ELECT HDTV (model 240-030-960) High Definition LCD Monitor is an intelligent, microprocessor-based TFT-LCD monitor intended for use in endoscopic sur gic al ap plic at ion s . It ha s an ergonomically designed display and is com patible with most an alo g RGB (Red, Green, Blue) display standards.
The VISION ELECT HDTV (model 240-030-960) monitor has the following features:
• Advanced Viewing Solution (AVS): A sophisticated filter
extends the viewing angle of the screen image, without sacrificing contrast ratio and brightness.
• Upgraded video cards or software will not require buying a new
monitor because of the wide auto-scanning compatibility range.
• The internal microprocessor digitally controls auto-scanning, for
horizontal scan frequenc ies be tween 31.47 KHz and 78.88 KHz, and vertical scan frequencies between 50.0 Hz and 85.1 Hz. In each frequency mode, the microproces sor-based circuitry allows the monitor to function at the precision of a fixed frequency.
• The resident memory allows for storing factory default settings
and also additional user adjustment parameters.
• The maximum resolution achievable is WUXGA (1920 × 1200,
60Hz).
• The monitor is compliant with the VESA-DPMS power-
management standard. In order to sa ve energy, the monitor must be connected to a system compliant with the standard.
• The monitor is designed to mount an accessory (such as a
camera or speaker) with a 2 lbs (max) load per hole.
• The monitor is certified by UL International in accordance with
medical standard UL 60601-1. It is also CE marked for sale in the European Community for integration or use with medical products.
Product Contents
1. Carefully unpack the VISION ELECT HDTV (model 240-030-9 60) monitor. Save the original carton and associated packing material. They will be useful should you have to transport or ship the unit.
2. Check the monitor and its accessories for any damage that may have occurred during shipping. Immediately report any damag e to the shipping company.
EN-7
English
3. Check the contents of the accessories package with the following list to make sure all components have been included. In addition to this manual, you should find:
Included Accessories
(a) 1 AC adapter (Stryker P/N 240-030-950,
Manufacturer: JEC, Model No: JMW1150KA2400F07) (b) 4 M4 × 10mm VESA screws (c) 2 Cable-managerment clamps
Optional Accessories (not shown)
1. Fiber optic module (Stryker P/N 240-030-962)
2. 15-ft. (5 pin) DC extension cable (Stryker P/N 240-030-951)
3. 75-ft. (5 pin) DC extension cable (Stryker P/N 240-030-952)
4. 2-to-5 pin DC convertor adapter (Stryker P/N 240-030-953)
5. DVI cable
6. VGA HDDB15 cable
7. BNC cable
8. S-Video cable
9. Hospital-grade AC power cord
EN-8
System Interconnection
Power Connection
Video Input Signals
eg:
SDC
SIDNE
VIDEO PRINTER
DVD/VCR
etc...
Video Sources: Camera (1188 Series, 1088 Series, 988 Series)
Video Peripheral
Signal Output DVI RGBHV S-Video C-Video
(Model 240-030-930)
VISIONELECT
VISION ELECT HDTV
(Model 240-030-960)
Video Input Signals
Video Peripheral
Signal Output:
DVI RGBHV S-Video C-Video
SDI
Fiber (Optional)
eg:SDC SIDNE VIDEO PRINTER DVD/VCR
etc....
Video Sources: Camera (xx88 Series)
SD/HD-SDI output
Additional Inputs: RGBS YPbPr SD/HD-SDI SOG
Hospital-grade power cord
Power Supply 240-030-950 Manufacturer: JEC Korea Model No: JMW1150KA2400F07
Extension cord (Optional) Manufacturer: JEC Korea Stryker P/N: 240-030-951 (15ft) or 240-030-952 (75ft)
VISION ELECT HDTV (Model 240-030-960)
EN-9
English
Operating the Monitor
Front Panel Controls
Operate the monitor using the rotary control located on the front panel. A list of the monitor controls and their functions is provided below.
Figure 1: The VISION ELECT HDTV LCD Monitor front panel controls.
1. Power LED: Indicates menu current status. Displays green if monitor is powered on; blinks red if monitor is in Standby mode.
2. Power Switch (Soft): Turns the power ON or OFF.
3. Rotary Control (Turn Right / Left): With the on-screen display menu activated, increases/decreases the value of the selected parameter. With the on-screen display deactivated, activates the video source selection menu.
4. Rotary Control Switch (Push): Accesses/selects on-screen display menu.
5. Rotary Control Switch (Push and Hold): Exits on-screen display menu.
Rotary Control
Power LED Power Switch
(Soft)
EN-10
Rear Panel Panel
Figure 2: The VISION ELECT HDTV LCD Monitor rear panel.
1. VESA mounting holes (100mm): Use to mount the monitior.
2. Power connector: DC power connector.
3. Power switch (Hard): Turns the input DC power ON or OFF.
4. Connectors tag: Indicate types of connectors.
5. Cable-managerment clamps: Cables organizer.
Input Port Layout
Cable Managerment Clamps
VESA mounting holes (100mm)
Power Switch (Hard)
Connector tags
Power connector
Power 24V
Optical (optional)
S-Video HD/SD
SDI IN
G/Y H-sync R/Pr B/Pb V-sync
Power Switch (hard)
DVI VGA C-Video/
SOG
HD/SD SDI OUT
RS232
EN-11
English
Fiber Optic Module (optional) Installation
1. Power OFF the monitor using th e ha rd pow er switch .
2. Unscrew the two screws from the cover plate, then remove the cover plate.
3. Insert the Fiber Optic Module (Stryker P/N 240-030-962) into the Fiber Optic Module slot.
Caution Excessive force or misalignment during insertion
might damage the module connector.
4. Power ON the monitor using the hard power switch.
Input Selection List
Digital RGB Analog RGB HD/SD-SDI Component (Y/Pb/Pr) RGBS S-Video C-Video SOG Exit
Unscrew those 2 screws
EN-12
Standard On-Screen Display (OSD) Operation
The monitor provides an on-screen display to help navigate through the various monitor-adjustment menus.
1. Press the Rotary Control to activate the on-screen display (OSD) menu.
2. Rotate the Rotar y Control to move up or down throug h the menu. The parameter will be highlighted when selected.
3. Press the Rotary Control to enter the next level OSD.
4. Rotate the Rotary Control to increase or decrease the value of the selected parameter, or to make a selection on different options.
5. To exit the OSD menu screen from the second- or third-level OSD menu, select the Exit option. To completely exit the OSD, press and hold the Rotary Control. If no keys are presse d for a time period, the OSD automatically times out.
6. While the OSD menu is deactivated, rotate the Rotary Control to activate the input signal selection menu. The current input signal will be highlighted with a dot. Rotate the Rotary Control to select the preferred input signal.
Stryker Camera Preset Modes
Camera Resolution
(H x V)
Horizontal
Frequency
(KHz)
Vertical
Frequency
(Hz)
988 1024 x 768 49.09 59.90
988i 1024 x 768 41.25 50.00
1088/SDC Pro2 1024 x 768 50.03 60.00
1088i/SDC Pro2 1024 x 768 41.10 50.00
1088/1188/SDC HD 1280 x 1024 64.02 60.10
1088i/1188i/SDC HD 1280 x 1024 59.99 50.00
1188w720 1280 x 720 45.00 60.00
1188iw720 1280 x 720 37.50 50.00
EN-13
English
OSD Function Description
Actual on-screen display values may vary with updated versio ns of the firmware and user setting. * Color Temperature RGB adjustment is avilable only for Standard, Arth and Lap settings. * PACS and Norm selection only available under SOG input. ** Only available under SDI, S, or C video input. *** Only available under VGA input.
Item Function Description Range
Specialty
Color Temperature *
Chooses between color temperatures for Standard, Arth, Lap, PACS, or
Norm Red Red balance -128 to 128 Green Green balance -128 to 128 Blue Blue balance -128 to 128
Gamma Gamma value
0.1 to 2.5, S0, S1, S2
Setting
Brightness Increases or decreases the brightness 0-100 Contrast Increases or decreases the contrast 0-100 Phase** Increases or decreases the Phase level 0-100 Chroma** Increases or decreases the Chroma level 0-100 Image Sharpness Sets image sharpness 1-10 Video Sharpness** Increases or decreases the video sharpness 0-100
Image Effect
Scale Mode
Chooses scale mode between Fill All, V -Fill , H-Fill, O ne To One or Fill To
Aspect Freeze Frame Enables or Disables freeze frame Zoom/Pan Enables zoom-in and pan function PIP Enables PIP (Picture In Picture) function POP Enables POP (Picture On Picture) function PBP Enables PBP (Picture By Picture) functio n
Advanced
OSD Control Controls OSD Menu Position, Background, and Timeout Screen Control*** Controls and adjust Horizontal, Vertical, Frequency, Phase
DPMS
Chooses DPMS (Display Power Management
Signaling)
ON, OFF, 30min, 60min, 90min,
120min Auto Source Select Adjusts Auto Source Select between on and off Restore Factory
Settings
Sets to factory default
Key lock Sets to Key lock mode
Information
User Name Entry Enters custom username display for boot-up display Serial Number Displays monitor serial number Runtime Displays current run time of monitor Input Format Displays current input format
EN-14
Cleaning the Monitor
Caution Do not expose the monitor to moisture or apply liquid
cleaners directly to the screen. Spray the cleaning solution into a soft cloth and clean the screen gently.
Cleaning display plastic area
No specific liquid or chemical is necessary for cleaning the VISION ELECT HDTV (model 240-030-960) LCD monitor. Use only non-abrasive cloths and cleaning solutions to clean similar equipment used in hospitals.
1. Clean with a dry soft cloth, or a soft cloth lightly moistened with mild detergent solution. Do not use any type of solvent, such as alcohol or benzine, which might damage the finish.
2. Apply alcohol to glass surfaces with soft cotton applicator to aid in cleaning and drying without leaving spots or streaks.
3. Dry thoroughly with soft towel or gauze surgical sponge.
Acceptable cleaning agents for bezel cleaning include:
• Cidex (2.4% glutaraldehyde solution)
• 0.5% Chlorhexidine in 70% isopropyl alcohol
Cleaning display filter area
Clean with a dry soft cloth, or soft cl oth lightly moistened with warm water . Other acceptable cleaning agents are listed below:
• 70% isopropyl alcohol
• Cidex (2.4% glutaraldehyde solution)
• 0.5% Chlorhexidine in 70% isopropyl alcohol
EN-15
English
Troubleshooting
Before returning your LCD monitor for service, consult the troubleshooting list below:
Problem Current Status Remedy
No Picture
LED ON
Using the OSD, adjust the brightness and cont rast to maximum, or reset them to their default settings.
LED OFF
Check the power switch at the front and back of the monitor.
Check if the AC power cord is properly connected to the AC adapter.
LED Blinking
Check if the video signal cable is properly connected at the back of the monitor.
Check if power of the video signal source system is ON.
Abnormal
Picture
Oversized, undersized display, missing display, or center shift
Using the Screen Menu, adjust the PHASE, FREQUENCY, HORIZONTAL, and VERTICAL settings with non-standard video signal timing.
Wait a few seconds after initial sync of video signals, or power cycle the monitor.
EN-16
Technical Specifications
Display
LCD Monitor Panel 25.54 inches
(a-Si TFT Active matrix LCD) Synchronization 2.5 - 5.0 Vpp separated sync Pixel Pitch 0.2865(W) × 0.2865(H) Response Time <25ms Typ View Angle +/-89° (L/R) × +/-89° (U/D) Display Colors 16 million colors Native Resolution 1920 dots × 1200 dots Input Signal 1 × DVI
1 × VGA
1 × HD/SD-SDI
1 × C-Video/SOG
1 × S-Video
1 × Component (Y/G, Pb/B, Pr/
R, H/CS, VS)
1 × Optical (optional) Maximum Pixel Clock 170MHz
Electrical
Power Adapter AC 100-240V; DC 24V Power Consumption 150W (max) Current Direct
Dimensions
Dimensions (W × H × D) 616.4 × 428.8 × 121.2mm
(24.27 × 16.88 × 4.77 inches) Weight 19.62 lbs VESA Mounting Interface VESA 100 × 100mm
Operating Conditions
Operating Temperature 50 to 104ºF (10 to 40°C) Relative Humidity 30 to 75% Atmospheric Pressure Range 500 to 1060 hPa Electrical Input Rating 24V DC 6.25A
Transport & Storage Conditions
Stor age -4 to 140°F (-20 to 60°C) Relative Humidity Range 10 to 85% Atmospheric Pressure Range 500 to 1060 hPa
EN-17
English
Classification and Approvals
Class I Equipment Medical equipment with respect to electric shock, fire and mechanical
hazards only in accordance with UL 60601-1 and CAN/CSA C22.2 No. 601.1.
IPX1 Water Ingress Protection Continuous operation
Warning This equipment is not suitable for use in the presence
of a FLAMMABLE ANESTHETIC MIXTURE WITH AIR, or WITH OXYGEN OR NITROUS OXIDE.
This monitor is intended for use on Health Care Facilities model 240-030-960.
No user serviceable parts inside. Ask qualified personnel before accessing internal componen ts.
Caution For disposal of waste product, follow the requirement
of the local code.
This product is considered electronic equipment. It must not be disposed of as unsorted municipal waste, and must be collected separately. Please contact the manufacturer or other authorized disposal company to decommission your equipment.
51 LJ
Medical Equipment
E215822
EN-18
Electromagnetic Compatibility
Like other electrical medical equipment, the VISION ELECT HDTV (model 240-030-960) monitor requires special precautions to ensure electromagnetic compatibility with other electrical medical devices. To ensure electromagnetic compatibility (EMC), the VISION ELECT HDTV (model 240-030-960) monitor m ust be installed and operated according to the EMC information provided in this manual.
Note The VISION ELECT HDTV (model 240-030-960) monitor
has been designed and tested to comply with IEC 60601-1-2:2003 requirements for EMC with other devices.
Caution The VISION ELECT HDTV (model 240-030-960) monitor
may be interfered with by other equipment, including portable and mobile RF communication equipment, even if such equipment meets the applicable emissions requirements.
Warning Using cables or accessories other than those provided
with the VISION ELECT HDTV (model 240-030-960) monitor, may result in increased electromagnetic emissions or decreased immunity to such emissions.
Warning If the VISION ELECT HDTV (model 240-030-960)
monitor is used adjacent to or stacked with other equipment, observe and verify normal operation o f th e VISION ELECT HDTV (model 240-030-960) monitor in the configuration in which it will be used pr ior to us ing it in a surgical procedure. Consult the tables below for guidance in placing the VISION ELECT HDTV (model 240-030-960) monitor.
Warning When this device is connected with other electrical
equipment, leakage currents may be additive. To minimize total leakage current per patient, ensure that all systems are inst alled according to the requirement s of IEC 60601-1-1.
EN-19
English
Manufacturer's declaration - electromagnetic emissions
VISION ELECT HDTV (model 240-030-960) monitor is intended for use in the electromagnetic environment specified below. The customer or the user of VISION ELECT HDTV (model 240-030-
960) monitor should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic Environment - guidance
RF emissions CISPR11
Group 1
The VISION ELECT HDTV (model 240-030-
960) monitor must emit electromagnetic energy in order to perform its intended function. Nearby electronic equipment may be affected.
RF emissions CISPR11
Class B
VISION ELECT HDTV (model 240-030-960) monitor is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Harmonic emissions IEC61000-3-2
Class D
Voltage Fluctuations/ flicker emissions IEC61000-3-3
Complies
EN-20
Manufacturer's declaration - electromagnetic immunity
VISION ELECT HDTV (model 240-030-960) monitor is intended for use in the elect r omagnetic environment specified below . The cust omer or the user of VISI ON ELECT HDTV (model 240-030-960) monitor should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic
environment -
guidance
Electrostatic Discharge (ESD) IEC61000-4-2
6kV contact 8kV air
6kV contact 8kV air
Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Electrical fast transient/ burst IEC61000-4-4
2kV for power supply lines 1kV for input/output lines
2kV for power supply lines 1kV for input/output lines
Mains power quality should be that of a typical commercial or hospital environment.
Surge IEC61000-4-5
1kV differential mode 2kV common mode
1kV differential mode 2kV common mode
Mains power quality should be that of a typical commercial or hospital environment.
Power frequency (50/60Hz) magnetic field IEC 61000-4-8
3.0 A/m 3.0 A/m
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
Voltage dips, short interruptions and voltage variations on power supply input lines IEC61000-4-11
<5% Ut (>95% dip in Ut) for 0.5 cycle 40% Ut (60% dip in Ut) for 5 cycles 70% Ut (30% dip in Ut) for 25 cycles <5% Ut (>95% dip in Ut) for 5 sec.
<5% Ut (>95% dip in Ut) for 0.5 cycle 40% Ut (60% dip in Ut) for 5 cycles 70% Ut (30% dip in Ut) for 25 cycles <5% Ut (>95% dip in Ut) for 5 sec.
Mains power quality should be that of a typical commercial or hospital environment. If the user of VISIO N ELECT HDTV (model 240-030-960) image intensifier requires continued operation during power mains interruptions, it is recommended that VISION ELECT HDTV (model 240­030-960) image intensifier be powered from an uninterruptible power supply or a battery.
NOTE: Ut is the A.C. mains voltage prior to application of the test level.
EN-21
English
Manufacturer's declaration - electromagnetic immunity
VISION ELECT HDTV (model 240-030-960) monitor is intended for use in the electromagnetic environment specified below. The customer or the user of VISION ELECT HDTV (model 240-030-960) monitor should ensure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance
Level
Electromagnetic Environment: Guidance
Portable and mobile RF communications equipment should be used no closer to any part of the VISION ELECT HDTV (model 240-030-960) monitor system, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Conducted RF IEC 61000-4-6
3 Vrms 150 kHz to 80 MHz
3 V
Recommended Separation Distance
Radiated RF IEC 61000-4-3
3 V/m 80MHz to 2.5 GHz
3 V/m
80 MHz to 800 MHz 800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacture r and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site
survey
(a)
, should be less than the
compliance level in each frequency range
(b)
. Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people. (a) Field strengths from fixed tra nsmitters, such as base stat ions for radio ( cellular/cordless) telep hones
and land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast, cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the VISION ELECT HDTV (model 240-030-960) monitor system is used exceeds the applicable RF compliance level above, the VISION ELECT HDTV (model 240-030-960) monitor system should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the VISION ELECT HDTV (model 240­030-960) monitor unit. (b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
d1.17P=
d1.17P=
d2.33P=
EN-22
Recommended Separation Distances Between Portable and Mobile RF Communications
Equipment and the VISION ELECT HDTV (model 240-030-960) monitor System
The VISION ELECT HDTV (model 240-030-960) monitor system is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The use r of t he VISION ELECT HDTV (model 240-030-960) monitor system can help prevent electromagne tic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the VISION ELECT HDTV (model 240-030-960) monitor system as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power (W) of transmitter
Separation distance (m) according to frequency of transmitter
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.74
11.171.172.33 10 3.70 3.70 7.37 100 11.70 11.70 23.30 For transmitters rated at a maximum output power not listed above, the recommended separation
distance (d) in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structu res, objects, and people.
d1.17P=
d1.17P=
d2.33P=
EN-23
English
Warranty
This S tryker Endoscopy p roduct is wa rranted to the original purchase r for a period of 90 days from the date of purchase to be free from defects in material and workmanship.
This warranty extends to all purchases and is limited to the repair or replacement of the product without charge when returned in the original shipping case to:
Stryker Endoscopy 5900 Optical Court San Jose, CA 95138 USA
Stryker Endoscopy cannot accept responsibility for returns or replacements which have not been authorized. This warranty does not cover damages caused by misuse or by failure to follow the procedures outlined in this manual or demonstrated by Stryker Endoscopy representatives.
There are no other expressed warranties.
Service and Claims
Do not attempt to service this product yourself. If service is needed either during or after the warranty period, contact Stryker Endoscopy at 1-800-624-4422, or phone your local Stryker Endoscopy sales representative.
EN-24
Declaration of Conformity
Manufacturer:
Stryker Endoscopy 5900 Optical Court San Jose, CA 95138 USA 1-800-624-4422 1-408-754-2000
Product Designation: VISION ELECT HDTV (model 240-030-960) 26” High Definition LCD Monitor
The above product complies with the requirements of EC Directive 93/42/EEC. This declaration is based on compliance testing to:
EN 60601-1/08-90 EN 60601-1 A1/05-93 EN 60601-1 A2/06-95
This document is on file. The CE mark will be affixed based on directive 93/42/EEC.
Stryker European Representative Regulatory Manager, S tryker France ZAC Satolas Green Pusignan Ave. de Satolas Green 69881 MEYZIEU Cedex, France
® Stryker and Stryker Endoscopy are registered trademarks of Stryker Corporation.
Stryker Endoscopy 5900 Optical Court San Jose, CA 95138 USA 1-408-754-2000, 1-800-624-4422 www.stryker.com
2008/11 www.stryker.com 1000-401-075 E