This manual is designed to assist you with the maintenance of the 2100 Childbearing Bed. Read it thoroughly
before using the equipment or beginning any maintenance on it.
SPECIFICATIONS
Maximum Weight Capacity300 pounds (135 kilograms)
Electrical3 motor function: Head–Bed–Foot
6 function patient comfort
120 VAC, 60 Hz, 4 Amp – 105 to 135 VAC operating range.
Current leakage less than 100 microamperes.
Hospital grade plug and 3–wire heavy duty cord.
Compatible with non–flammable anesthetic agents and oxygen by nasal catheter or mask.
WARNING
The 2100 is equipped with a hospital grade plug for protection against shock hazard. It must be plugged directly into a properly grounded three–prong receptacle. Grounding reliability can be achieved only when a
hospital grade receptacle is used.
WARNING / CAUTION / NOTE DEFINITION
The words WARNING, CAUTION and NOTE carry special meanings and should be carefully reviewed.
WARNING
The personal safety of the patient or user may be involved. Disregarding this information could result in injury
to the patient or user.
CAUTION
These instructions point out special procedures or precautions that must be followed to avoid damaging the
equipment.
NOTE
This provides special information to make maintenance easier or important instructions clearer.
To assure its proper use and the safety of patients and staff, the 2100 has been marked with the following
caution and warning labels:
DANGERExplosion Hazard – do not use in the presence of flammable anesthetics.
CAUTIONThis unit is equipped with a hospital grade attachment plug. Grounding reliability can be
achieved only when equipment is connected to equivalent receptacle.
CAUTIONElectrical shock hazard. Do not remove cover panels. Refer all servicing to qualified personnel.
CAUTIONDisconnect power cord before hand cranking. Remove hand crank before connecting power
cord.
2
Page 4
Unpacking and Initial Set–Up Procedures
UNPACKING INSTRUCTIONS
Refer to unpacking instructions located on outside of marked crate.
SET–UP PROCEDURES
It is important that the Stryker Adel 2100 Childbearing Bed is working properly before it is put into service.
The following list will help assure that each part of the bed is tested.
SPlug the bed into a properly grounded, hospital grade wall receptacle.
WARNING
The 2100 is equipped with a hospital grade plug for protection against shock hazard. It must be plugged directly into a properly grounded three–prong receptacle. Grounding reliability can be achieved only when a
hospital grade receptacle is used.
SAssure the siderails raise and lower smoothly and lock in the up position.
SAssure that the brakes hold when the brake pedal is engaged (see page 4 for adjustment).
SRun through each function on the hand pendant to assure that each function is working properly.
Beds equipped with siderail control option only:
SPlug the interface cable into the 37 pin connector in the litter frame at the head end of the bed, and
into the ”Patient Station”, ”Head Wall”, ”Docker Station”, or equivalent (whichever applies).
SRun through each function on the siderail control panels to assure that each function is working prop-
erly.
3
Page 5
Preventative Maintenance
PREVENTATIVE MAINTENANCE PROGRAM
Observation of the following scheduled maintenance and inspection program will help assure continuous
trouble–free operation of your 2100 Childbearing Bed.
Lubricate screw threads on
leg support clamping
knobs.
Clean and lubricate vertical
slide rods that support the
foot panels.
Inspect and adjust wheel
brake shoes on central
locking casters.*
Lubricate pin joints on lift
mechanism.
Inspect power cord and
plug for damage due to
abrasion or cutting.
Inspect exposed metal surfaces and apply touch–up
paint where required.
Check electrical safety by
testing for leakage current
and ground continuity.
Scratched wood components. Remove loose dust;
spray a light coat of Varathane over the area. Apply
a second coat in approximately six hours.
Remove stains from mattress covers.
Each occurrenceSee page 23.
6
6
24
6
12
12
12
N/A
N/A
N/A
Phillips screwdriver
N/ALubricant
N/A
N/A
Electrical testing equipment
N/A
Lubricant
Lubricant
N/A
N/A
Touch–up paint**
N/A
Satin Varathane #91
*T o adjust brakes, remove 6 mm screw from caster. Rotate housing to the right (counterclockwise) to reduce
braking force or to the left (clockwise) to increase braking force. Reinstall the 6 mm screw.
NOTE
Left front caster is not a brake caster
**Touch–up paint
Tan Brown (spray) p/n 0064–06–00
Tan Brown (touch–up) p/n 0064–08–00
Camel (spray) p/n 0064–07–00
Camel (touch–up) p/n 0064–09–00
4
Page 6
Service Information
TROUBLESHOOTING GUIDE
The following chart is intended to assist qualified service personnel with identifying and correcting problems
associated with the electrical/pneumatic systems. The schematic diagrams on the following pages should
be used for reference. A volt–ohm meter will be required.
WARNING
Because these problem solving methods may involve interchanging motor plugs or pneumatic air lines to isolate the fault or probable cause, the patient should always be removed from the bed before corrective action
is taken.
PROBLEM/SYMPTOMPOSSIBLE CAUSERECOMMENDED ACTION
No motors functioning.No power supply.
Defective power cord or plug.
Defective electrical controller.
One motor not functioning – head,
bed, or foot.
Incorrect travel limits – head, bed or
foot.
One pneumatic solenoid not functioning – lumbar, seat left or seat
right.
No inflation of lumbar, seat left, or
seat right.
No inflation or deflation of lumbar,
seat left, or seat right.
No air to operate lumbar, seat left, or
seat right.
Slow filling or emptying of air from
lumbar, seat left, or seat right.
Squeal or noise in lift mechanism Or,
bed motor labors when bed is fully
loaded with 300 pounds.
Loose wire at motor plug or receptacle.
Defective motor.
Defective pendant controller.
Defective membrane switch.
Defective electrical controller (circuit
board).
Physical blockage of movement.
Defective motor or drive screw.
Motor drive tube out of adjustment.
Loose wire at solenoid plug.
Defective solenoid.
Defective pendant controller.
Defective membrane switch.
Defective electrical controller (circuit
board).
Leaking bladder.Replace head panel–mattress or
Kinked lines.
Defective solenoid.
Defective compressor.
Loose wire at motor connectors.
Kinked lines.
Contamination in valves.
Dry joints.
Defective counterbalance spring.
Test voltage at wall receptacle
Examine and test for continuity.
Test input and output voltages (see
schematic page 7). Listen for relay
actuation.
Inspect connectors; check continuity.
Interchange motor plugs to isolate.
Interchange motor plugs to isolate.
Check continuity.
Test input and output voltages (see
schematic page 7). Listen for relay
actuation.
Visually inspect moving parts.
Replace motor (see pages 9–11).
Reset (see motor replacement–
pages 9–11)
Inspect connectors; check continuity.
Interchange solenoid plugs to isolate. Replace coil.
Interchange solenoid plugs to isolate.
Check continuity.
Test input and output voltages (see
schematic page 7).
seat panel–mattress (see page 15 or
16).
Inspect lines for kinking: re–route, if
necessary.
Interchange solenoid plugs to isolate.
Replace compressor (see page 14).
Inspect connectors: check continuity;
test voltage.
Inspect lines for kinking: re–route, if
necessary.
Isolate by switching lines. Replace
manifold (see page 17).
See prev. maintenance (page 4).
Check for obstruction, bent rod, or
bent housing. Remove obstruction.
Replace spring (see page 20).
5
Page 7
Service Information
TROUBLESHOOTING GUIDE (CONTINUED)
PROBLEM/SYMPTOMPOSSIBLE CAUSERECOMMENDED ACTION
Bed does not tilt when lever is depressed.
Bed tilts when weight is applied to
opposite ends.
Brakes do not lock wheels.Tire wear.Readjust casters
Brakes hard to apply.Brakes over–adjusted.Readjust casters
NURSE COMMUNICATION SIDERAIL OPTION TROUBLESHOOTING GUIDE
Before attempting to diagnose an apparent problem, be sure the power cord is plugged into a properly
grounded hospital grade wall receptacle and the bed has power. Be sure the nurse call cable has a good
connection at both the wall and the bed. If both these things are done, proceed with diagnosis.
CAUTION
Membrane switches are fragile. Never depress a membrane dome unless the switch is lying on a flat, firm
surface. Doing otherwise will cause the switch to fail eventually. When applying a new membrane switch,
lay it down evenly; do not roll it down. Once a membrane switch has been peeled back, it cannot be reused.
Tilt spring out of adjustment.
Defective tilt spring.
Tilt spring out of adjustment.
Defective tilt spring.
Readjust spring length.
Replace spring (see page 21).
Readjust spring length.
Replace spring (see page 21).
PROBLEM/SYMPTOMPOSSIBLE CAUSERECOMMENDED ACTION
Some or all of the pillow speaker
functions aren’t working properly.
A control button on one side of the
siderail doesn’t work or the T.V.
sounds different on one side.*
Defective cable (from bed to wall).Switch the cable with one that is
known to work. If this solves the
problem, replace the cable. If not, try
a bed from another room that is
known to be O.K. If the problem persists on the bed that is known be
O.K., the problem is in the wiring in
the wall.
Replace switch.**
Replace siderail PC board.**
Switch the siderail cables at the splitter box. If the problem switches
sides, replace splitter box. If the
problem stays in the same siderail,
replace cable.
* For audio problems, check resistance across pins 4 and 35 on the 37 pin connector at the rear of the bed.
The reading should be 22.5 "5 ohms for 45 ohm systems or 4 "2 ohms for 8 ohm systems. A reading
double these values indicates no connection to the speaker in one of the siderails. If the sound is inadequate, turn the volume up to its highest level by holding the button for 30 seconds. Adjust the remote speaker output volume on the TV to its highest level. Turn the TV back down until the volume at the remote speakers just begins to be affected to reduce distortion caused by overdriving the speakers. Cover the speaker
grills separately and compare the sound loss to be sure the sound is balanced between the two speakers.
** For problems inside the siderails, peel back the outside membrane switch to expose the seven screws hold-
ing the plastic housing to the siderail assembly. Remove the screws and carefully disconnect the membrane switch from the siderail PC board.
*** For problems in the splitter box, remove the box and examine the LED’s next to the relays. If the LED comes
on when the function button is pressed, a faint click should be audible. If no click is heard, replace the relay .
If replacing the relay doesn’t solve the problem or if the LED doesn’t light, replace the splitter box PC board.
If the problem still persists, replace the splitter cable. (Refer to Splitter Box Assembly (page 49) for part
reference.)
6
Page 8
Schematic Diagram
7
Page 9
Nurse Communication Schematic Diagram
LEFT
SIDERAIL
CABLE
8
Page 10
Nurse Communication Schematic Diagram
THIS WIRE
COLORED
STRIPE
TO BED
CIRCUIT
BOARD
8a
Page 11
Notes
Page 12
Service Information
HEAD MOTOR REPLACEMENT
HEAD MOTOR KIT #88–3979–18–00
Required Tools:
7/16” Hex Socket & Ratchet1/4” Allen Wrench
Phillips ScrewdriverLoctite 242
Replacement Procedure:
1. Raise bed to full up position.
CAUTION
DIsconnect power plug from wall receptacle.
2. Use hand crank to lower head panel.
3. Remove headboard, foot mattress, foot extension, foot pans, I.V. pole and leg supports.
4. Set brake with casters turned as though the bed is being pushed backwards. Assure the head end of the
bed is at least five feet away from the wall when it is upright.
5. With the assistance of another person, lift the foot end of the bed chassis and rotate the locked rear casters
until the upper frame–headboard mounts touch the floor.
6. Disconnect the head motor plug and unscrew the cord strain relief clamp.
7. Remove the four hex screws from the motor trunnion mounts.
8. Remove the two hex screws and washers from the drive tube/pull bar.
9. Motor, drive tube and mounting hardware can now be removed as a unit.
NOTE
The head motor is supplied to you with the drive tube retracted (ref. 1/8” from motor case). Do not allow drive
tube to rotate.
10. Remove the two white trunnion bearings from the defective motor and place them onto the new motor trunnion mounts. Place the motor assembly back into the litter frame and attach with the four hex screws.
Use Loctite 242 to secure the screws.
11. To hook up the head panel to the drive tube, plug the motor lead into the control box, replace the strain relief
clamp, and plug the power cord into the wall receptacle. Hold the drive tube to prevent rotation and press
the ”HEAD DOWN” button on the hand pendant until the drive tube aligns with the head panel pull bar
holes. Install the two hex screws and washers.
12. Be sure all screws, cables and electrical connections are secure.
13. With the assistance of another person, tilt the bed back down to the floor.
14. Operate the head motor from full down position to full up to full down again. Make sure the motor stops
electrically before the drive tube bottoms out on the mechanical stops. Check motor cable for proper routing and attachment to prevent interference.
15. Check electrical safety in conformance with specified hospital requirements (leakage current, ground continuity, etc.)
16. Remount all items removed in step three and return the bed to service.
Disconnect power plug from wall receptacle.
The 2100 bed is equipped with gas springs to assist the lift mechanism. To prevent the seat section from
expanding away from the chassis when the bed motor is removed, use the hand crank to raise the bed to
full up before placing the bed in servicing position. If the motor is jammed or cannot be hand cranked, tie the
seat section to the chassis with straps or strong rope.
2. Remove headboard, foot mattress, foot extension, foot pans, I.V. pole and leg supports.
3. Set brake with casters turned as though the bed is being pushed backwards. The casters at the head end
of the bed must be five feet away from the wall.
4. With the assistance of another person, lift the foot end of the bed chassis and rotate the locked rear casters
until the upper frame–headboard mounts touch the floor.
5. Disconnect the bed motor plug and unscrew the cord strain relief clamp.
6. Remove the four hex screws from the motor trunnion mounts.
7. Remove the two Allen screws and sleeves from the drive tube/crank.
8. Motor, drive tube, and mounting hardware can now be removed as a unit.
9. Remove the two white trunnion bearings from the defective motor and place them onto the new motor trunnion mounts. Place the motor assembly back into the litter frame and attach with the four hex screws.
Use Loctite 242 to secure the screws.
NOTE
The bed motor is supplied with the drive tube retracted (ref. 1/8” from motor case). Do not allow drive tube
to rotate.
10. Using Loctite 271, install the two 1/4” Allen screws and sleeves.
11. Plug the motor lead into the control box and replace the strain relief clamp.
12. Be sure all screws, cables, and electrical connections are secure.
13. With the assistance of another person, tilt the bed back down to the floor.
14. Plug the power cord into the wall receptacle. Operate the bed motor from full up position to full down to
full up again. Be sure the motor stops electrically before the drive tube bottoms out on the mechanical
stops. Check motor cable for proper routing and attachment to prevent interference.
15. Check electrical safety in conformance with specified hospital requirements (leakage current, ground continuity, etc.)
16. Remount all items removed in step two and return bed to service.
2. Use hand crank to lower foot pans to full down position.
3. Remove headboard, foot mattress, foot extension, foot pans, I.V. pole and leg supports.
4. Set brake with casters turned as though the bed is being pushed backwards. The casters at the head end
of the bed must be five feet away from the wall.
5. With the assistance of another person, lift the foot end of the bed chassis and rotate the locked rear casters
until the upper frame–headboard mounts touch the floor.
6. Disconnect the foot motor plug and unscrew the cord strain relief clamp.
7. Remove the four hex screws from the motor trunnion mounts.
8. Remove the two Allen screws and sleeves from the drive tube/crank.
9. Motor, drive tube and mounting hardware can now be removed as a unit.
10. Remove the two white trunnion bearings from the defective motor and place them onto the new motor trunnion mounts. Place the motor assembly back into the litter frame and attach with the four hex screws.
Use Loctite 242 to secure the screws.
NOTE
The bed motor is supplied with the drive tube retracted (ref. 1/8” from motor case). Do not allow drive tube
to rotate.
11. Install the two 1/4” Allen screws and sleeves.
12. Plug the motor lead into the control box and replace the strain relief clamp.
13. Be sure all screws, cables, and electrical connections are secure.
14. Plug the power cord into the wall receptacle. Operate the foot motor from full down position to full up to
full down again. Be sure the motor stops electrically before the drive tube bottoms out on the mechanical
stops. Check for proper cable routing and attachment to prevent interference.
15. With the assistance of another person, tilt the bed back down to the floor.
16. Check electrical safety in conformance with specified hospital requirements (leakage current, ground continuity, etc.)
17. Remount all items removed in step three and return the bed to service.
11
Page 15
Service Information
CONTROLLER (CIRCUIT BOARD) REPLACEMENT
CONTROLLER KIT #88–0316–02–00 (for beds with serial numbers 880596 and below)
CONTROLLER KIT #88–0316–21–00 (for beds with serial numbers 880597 and above)
2. Remove headboard, foot mattress, foot extension, foot pans, I.V. pole and leg supports.
3. Set brake with casters turned as though the bed is being pushed backwards. The casters at the head end
of the bed must be five feet away from the wall.
4. With the assistance of another person, lift the foot end of the bed chassis and rotate the locked rear casters
until the upper frame–headboard mounts touch the floor.
5. Unplug all connectors on controller box.
6. Remove the two screws holding on the controller box cover, remove the cover, and disconnect the power
cord.
7. Remove the three mounting nuts and 2 stand–offs and loosen the cable clamps. Remove the defective
circuit board.
8. Attach replacement circuit board to control box base with the four star washers and screws. (On later model
beds without a control box base, attach the circuit board to the litter.)
9. Remount the box to the litter frame with four star washers and nuts.
10. Connect the power cord and attach the cover with the two screws.
11. Plug all connectors into the controller box and re–tighten the cable clamps.
12. Be sure all screws, cables, and electrical connections are secure.
13. With the assistance of another person, tilt the bed back down to the floor.
14. Plug the power cord into the wall receptacle. Run through all functions on the hand pendant at least three
times to ensure proper operation. Check for proper cable routing and attachment to prevent interference.
15. Check electrical safety in conformance with specified hospital requirements (leakage current, ground continuity, etc.)
16. Remount all items removed in step two and return bed to service.
12
Page 16
Service Information
HAND PENDANT REPLACEMENT
HAND PENDANT KIT #88–0316–03–00
Required Tools:
Phillips Screwdriver3/8” Nut Driver
Replacement Procedure:
1. Raise bed to full up position.
2. Remove screw at head end of motor cover and remove cord strain relief clamps.
3. Unplug pendant cable from controller box and remove.
4. Route replacement pendant cable to control box and plug into controller.
5. Attach strain relief clamps and install screw at head end of motor cover.
6. Be sure all screws and connections are secure.
7. Run through all the functions on the hand pendant at least three times to ensure proper operation.
8. Return bed to service.
MEMBRANE SWITCH REPLACEMENT
MEMBRANE SWITCH KIT #88–0316–04–00
Required Tools:
5/64” Allen Wrench
Replacement Procedure:
CAUTION
Disconnect power plug from wall receptacle.
1. Remove the four screws from the back of the hand pendant.
2. Unplug the membrane switch assembly at the ribbon cable connector.
3. Plug the replacement membrane switch ribbon into the connector.
4. Place the membrane switch into the pendant housing, being sure to properly locate the strain relief bar.
Insert and tighten the four socket screws.
5. Run through all the functions of the membrane switch at least three times to ensure proper operation.
6. Return bed to service.
13
Page 17
Service Information
COMPRESSOR REPLACEMENT
COMPRESSOR KIT #88–0316–24–00
Required Tools:
11/32” Nut DriverBlade Screwdriver
Side (Wire) Cutter
Replacement Procedure:
1. Raise the bed to the full up position.
CAUTION
Disconnect power plug from wall receptacle.
2. Remove headboard, foot mattress, foot extension, foot pans, I.V. pole and leg supports.
3. Set brake with casters turned as though the bed is being pushed backwards. The casters at the head end
of the bed must be five feet away from the wall.
4. With the assistance of another person, lift the foot end of the bed and rotate the locked rear casters until
the upper frame–headboard mounts touch the floor.
5. Remove ground wire screw. Pull lead wire connectors apart and cut tubing line next to compressor fitting.
6. Remove the three mounting nuts and washers and the defective compressor.
7. Install the replacement compressor and shock mounts in the litter frame with the three star washers and
nuts.
8. Connect the ground wire with the star washer and screw. Push the electrical connectors together and push
the tubing end over the compressor outlet fitting.
9. Be sure all screws, cables and electrical connections are secure.
10. With the assistance of another person, tilt the bed back down to the floor.
1 1. Plug the power cord into the wall receptacle. Run through all functions on the hand pendant at least three
times to ensure proper operation. Check for proper cable routing and attachment to prevent interference.
12. Check electrical safety in conformance with specified hospital requirements (leakage current, ground continuity, etc.)
13. Remount all items removed in step two and return bed to service.
14
Page 18
Service Information
HEAD PANEL–MATTRESS REPLACEMENT
HEAD PANEL–MATTRESS KIT #88–0316–10–00
Required Tools:
7/32” Allen WrenchPhillips Screwdriver
Side (Wire) Cutter
Replacement Procedure:
1. Raise head panel to approximately 30_.
2. Raise bed to the full up position.
CAUTION
Disconnect power plug from wall receptacle.
3. Remove headboard, foot mattress, foot extension, foot pans, I.V. pole and leg supports.
4. Set brake with casters turned as though the bed is being pushed backwards. The casters at the head end
of the bed must be five feet away from the wall.
5. Unfasten Velcro strap and remove over–mattress cover.
6. Cut tie–wraps to free tubing (note routing of tubing to ensure it will be reinstalled properly).
7. Behind the head panel, separate the head mattress vinyl tubing at the middle of the two inch ”J” clip.
8. Remove the eight cap screws from the head panel support arms to release the head panel assembly with
mattress.
9. Feed lumbar tubing line through litter access hole and place head panel base ribs in support arm tracks.
Avoid kinking or pinching tubing.
10. Install the eight cap screws.
11. With barb supplied in kit, match and cut tubing and splice together.
12. Replace over–cover and pull it taut with the Velcro strap.
13. Plug the power cord into the wall receptacle. Run through all functions on the hand pendant at least three
times to ensure proper operation. Check for proper cable routing and attachment to prevent interference.
14. Remount all items removed in step three and return bed to service.
15
Page 19
Service Information
SEAT PANEL–MATTRESS REPLACEMENT
SEAT PANEL–MATTRESS KIT #88–0316–11–00
Required Tools:
3/8” Nut DriverPhillips Screwdriver
SIde (Wire) Cutter
Replacement Procedure:
1. Raise head panel to approximately 30_.
2. Raise bed to the full up position.
CAUTION
Disconnect power plug from wall receptacle.
3. Remove headboard, foot mattress, foot extension, foot pans, I.V. pole and leg supports.
4. Remove over–mattress cover, loosening Velcro straps and screws on underside of head panel.
5. Set brake with casters turned as though the bed is being pushed backwards. The casters at the head end
of the bed must be five feet away from the wall.
6. With the assistance of another person, lift the foot end of the bed chassis and rotate the locked rear casters
until the upper frame–headboard mounts touch the floor.
7. Cut old air lines on foot end of barbs located about 8” from the manifold.
8. Cut tie–wraps to free tubing (note routing of tubing to ensure it will be reinstalled properly).
9. Remove the four screws to release the defective seat panel assembly from the litter.
10. Install the four screws to hold the replacement seat panel assembly to the litter.
1 1. Route tubing as it was before removal and push ends of tubing onto manifold fitting. Install the tie–wraps.
12. Reconnect air lines at the barbs, noting that the center of the manifold is the right seat bladder.
13. Replace over–cover by pulling it taut with the Velcro straps.
14. Be sure all the screws, cable and electrical connections are secure.
15. With the assistance of another person, tilt the bed back down to the floor.
16. Plug the power cord into the wall receptacle. Run through all functions on the hand pendant at least three
times to ensure proper operation. Check for proper cable routing and attachment to prevent interference.
17. Remount all items removed in step three and return bed to service.
16
Page 20
Service Information
MANIFOLD REPLACEMENT
MANIFOLD KIT #88–0316–12–00
Required Tools:
Phillips Screwdriver3/8” Nut Driver
SIde (Wire) Cutter
Replacement Procedure:
1. Raise bed to the full up position then lower about 2” to gain access to lower manifold screw.
CAUTION
Disconnect power plug from wall receptacle.
2. Remove headboard, foot mattress, foot extension, foot pans, I.V. pole and leg supports.
3. Set brake with casters turned as though the bed is being pushed backwards. The casters at the head end
of the bed must be five feet away from the wall.
4. With the assistance of another person, lift the foot end of the bed chassis and rotate the locked rear casters
until the upper frame–headboard mounts touch the floor.
5. Disconnect the manifold electrical connectors at the control box.
6. Remove the two manifold mounting screws, washers and cable clamps.
7. Cut and remove air lines to the seat–left, seat–right and lumbar at the manifold fittings and cut air lines at
compressor.
8. Connect the air lines to the replacement manifold fittings.
9. Mount the manifold to the litter with the two star lock washers and screws.
10. Reconnect the electrical plugs to the control box and replace the cable clamps.
11. Be sure all screws, cables and electrical connections are secure.
12. With the assistance of another person, tilt the bed back down to the floor.
13. Plug the power cord into the wall receptacle. Run through all functions on the hand pendant at least three
times to ensure proper operation. Check for proper cable routing and attachment to prevent interference.
14. Check electrical safety in conformance with specified hospital requirements (leakage current, ground continuity, etc.)
15. Remount all items removed in step two and return bed to service.
17
Page 21
Service Information
LIMIT SWITCH REPLACEMENT
LIMIT SWITCH KIT #88–0316–14–00 (for beds with serial numbers 880596 and below)
LIMIT SWITCH KIT #88–0318–62–00 (for beds with serial numbers 880597 and above)
2. Remove headboard, foot mattress, foot extension, foot pans, I.V. pole and leg supports.
3. Set brake with casters turned as though the bed is being pushed backwards. The casters at the head end
of the bed must be five feet away from the wall.
4. With the assistance of another person, lift the foot end of the bed chassis and rotate the locked rear casters
until the upper frame–headboard mounts touch the floor.
5. Remove the two Phillips screws and lock washers and pull the switch assembly away from the bracket.
6. Remove the electrical connectors from the switches.
7. Remove the two screws, washers, spacers and nuts to release the limit switches.
8. Install the replacement switches using the fasteners removed in step 7.
9. Reconnect the electrical connectors to the switches.
10. Be sure all screws, cables and electrical connections are secure.
11. With the assistance of another person, tilt the bed back down to the floor.
12. Plug the power cord into the wall receptacle. Run through all the functions on the hand pendant at least
three times to ensure proper operation.
13. Raise bed to full up position. Lower foot pans to full down position, then lower bed. Note the automatic
raising of the foot pans. Raise foot pans to the full up position. Lower the bed to the full down position,
then lower foot pans. Note the automatic stop of the foot pans. Check for proper cable routing and attachment to prevent interference.
14. Check electrical safety in conformance with specified hospital requirements (leakage current, ground continuity, etc.)
15. Remount all items removed in step two and return bed to service.
18
Page 22
Service Information
COIL REPLACEMENT
COIL KIT #88–0316–16–00
Required Tools:
Phillips Screwdriver
Replacement Procedure:
1. Raise bed to the full up position.
CAUTION
Disconnect power plug from wall receptacle.
2. Remove headboard, foot mattress, foot extension, foot pans, I.V. pole and leg supports.
3. Set brake with casters turned as though the bed is being pushed backwards. The casters at the head end
of the bed must be five feet away from the wall.
4. With the assistance of another person, lift the foot end of the bed chassis and rotate the rear casters until
the upper frame–headboard mounts touch the floor.
5. Remove retaining clip on top of coil (slide to one side) and remove defective coil.
6. Loosen screw on side of coil connector and pull connector away from solenoid.
7. Install the replacement coil on the valve post and fasten with the slide clip.
8. Plug connector onto solenoid and tighten screw.
9. Be sure all screws, cables and electrical connections are secure.
10. With the assistance of another person, lower the bed back down to the floor.
1 1. Plug the power cord into the wall receptacle. Run through all functions on the hand pendant at least three
times to ensure proper operation. Check for proper cable routing and attachment to prevent interference.
12. Check electrical safety in conformance with specified hospital requirements (leakage current, ground continuity, etc.)
13. Remount all items removed in step two and return the bed to service.
19
Page 23
Service Information
COUNTERBALANCE SPRING REPLACEMENT
COUNTERBALANCE SPRING KIT #88–0316–17–00
Required Tools:
Blade Screwdriver
Replacement Procedure:
1. Raise bed to full up position.
CAUTION
Disconnect power plug from wall receptacle.
2. Remove headboard, foot mattress, foot extension, foot pans, I.V. pole and leg supports.
3. Set brake with casters turned as though the bed is being pushed backwards. The casters at the head end
of the bed must be five feet away from the wall.
4. With the assistance of another person, lift the foot end of the bed chassis and rotate the rear casters until
the upper frame–headboard mounts touch the floor.
5. Remove the retaining clip wire off each of the two ball joint sockets of the defective spring.
6. Use the blade screwdriver to pry the socket off the ball joints and remove the faulty spring. To verify spring
is defective, hold large end firmly in your hands, place the small end on the floor, and push down. If the
spring collapses, the spring is defective. If it is not defective, remove the other springs and test.
7. Install the replacement spring by snapping the sockets on the ball joints and installing the retaining clip wire
through the ball joint socket.
NOTE
Be sure to install the spring in the same direction it was prior to removal.
8. Be sure all screws, cables and electrical connections are secure.
9. With the assistance of another person, tilt the bed back down to the floor.
10. Plug the power cord into the wall receptacle. Run through all functions on the hand pendant at least three
times to ensure proper operation. Check for proper cable routing and attachment to prevent interference.
11. Check electrical safety in conformance with specified hospital requirements (leakage current, ground continuity, etc.)
12. Remount all items removed in step two and return bed to service.
20
Page 24
Service Information
TILT/TRENDELENBERG SPRING REPLACEMENT AND ADJUSTMENT
TILT/TRENDELENBERG SPRING KIT #88–0316–18–00
Required Tools:
Snap Ring Pliers (External)Blade Screwdriver
11/16” Open End Wrench1/4” Drift Punch
5mm Allen WrenchLoctite 242
Replacement Procedure:
1. Raise bed to the full up position and tilt head down.
CAUTION
Disconnect power plug from wall receptacle.
2. Remove headboard, foot mattress, foot extension, foot pans, I.V. pole and leg supports.
3. Set brake with casters turned as though the bed is being pushed backwards. The casters at the head end
of the bed must be five feet away from the wall.
4. With the assistance of another person, lift the foot end of the bed chassis and rotate the locked rear casters
until the upper frame–headboard mounts touch the floor.
CAUTION
Replace gas springs one side at a time.
5. Remove shoulder bolt holding gas spring to tilt arm.
6. Remove pin connecting rear lift bar and tilt arm. Disconnect rear lift bar from drive crank.
7. Fully compress gas spring and remove from tilt arm.
8. Remove pin connecting front lift bar and tilt arm.
9. Loosen jam nut on defective gas spring. Unscrew gas spring from spring bracket and remove.
10. Turn replacement gas spring into spring bracket 7 1/2 turns. This will ensure both gas springs engage at
the same time when the Trendelenberg pedal is depressed.
11. Replace and tighten jam nut on gas spring.
12. Apply Loctite 242 to threads of shoulder bolt and attach gas spring to tilt arm.
13. Activate gas springs to align tilt arm.
14. Reconnect front lift bar to tilt arm.
15. Reconnect rear lift bar to drive crank and tilt arm.
16. Repeat steps 5 through 15 for other spring, if necessary.
17. With the assistance of another person, tilt the bed back down to the floor.
18. Remount all items removed previously and return the bed to service.
The 2100 is designed and constructed to minimize and simplify your housekeeping requirements. Large, flat
surfaces with a minimum of cracks and crevices make the bed easy to clean using standard hospital practices
and procedures. All parts of the bed should be washed using a mild, detergent–type, cleaning solution. General–duty sanitizers and disinfectants may be used on all bed components, but avoid excessive moisture or
abrasive cleaners on hardwood areas. Also, avoid leaving excessive moisture between the mattress and foot
panel by drying surface thoroughly before re–making the bed.
GENERAL CLEANING
In general, when used in those concentrations recommended by the manufacturer, either phenolic type or
quaternary type disinfectants can be used with Staph–Chek fabrics. Iodophor type disinfectants are not recommended for use on Staph–Chek fabrics because staining may result. The following products have been
tested by the Herculite Laboratory and have been found not to have a harmful effect on Staph–Chek fabrics
WHEN USED IN ACCORDANCE WITH MANUFACTURERS RECOMMENDED DILUTION.*
TRADE NAME DISINFECTANT
TYPE
A33QuaternaryAirwick (Professional Products Division)2 ounces/gallon
A33 (dry)QuaternaryAirwick (Professional Products Division)1/2 ounce/gallon
BeaucoupPhenolicHuntington Laboratories1 ounce/gallon
Blue ChipQuaternaryS.C. Johnson2 ounces/gallon
ElimstaphQuaternaryWalter G. Legge1 ounce/gallon
Franklin
This solution may be used to remove iodine stains from mattress cover and foam footrest pad surfaces.
1. Use a solution of 1–2 tablespoons Sodium Thiosulfate in a pint of warm water to clean the stained area.
Clean as soon as possible after staining occurs. If stains are not immediately removed, allow solution to
soak or stand on the surface.
2. Rinse surfaces which have been exposed to the solution in clear water before returning bed to service.
Stryker Adel warrants to the original purchaser that its products should be free from defects in material and
workmanship for a period of one (1) year after date of delivery . Stryker Adel ’s obligation under this warranty
is expressly limited to supplying replacement parts and labor for, or replacing, at its option, any product which
is, in the sole discretion of Stryker Adel, found to be defective. Stryker Adel warrants to the original purchaser
that the frame and welds on its beds will be free from structural defects for as long as the original purchaser
owns the bed. If requested by Stryker Adel, products or parts for which a warranty claim is made shall be
returned prepaid to Stryker Adel’s factory. Any improper use or any alteration or repair by others in such manner as in Stryker Adel’s judgement affects the product materially and adversely shall void this warranty. No
employee or representative of Adel is authorized to change this warranty in any way.
This statement constitutes Stryker Adel’s entire warranty with respect to the aforesaid equipment. STRYKER
ADEL MAKES NO OTHER WARRANTY OR REPRESENTATION, EITHER EXPRESSED OR IMPLIED,
EXCEPT AS SET FORTH HEREIN. THERE IS NO WARRANTY OF MERCHANTABILITY AND THERE
ARE NO WARRANTIES OF FITNESS FOR ANY PARTICULAR PURPOSE. IN NO EVENT SHALL STRYKER ADEL BE LIABLE HEREUNDER FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING
FROM OR IN ANY MANNER RELATED TO SALES OR USE OF ANY SUCH EQUIPMENT.
To Obtain Parts and Service:
Stryker Adel products are supported by a nationwide network of dedicated Stryker Adel Field Service Representatives. These representatives are factory trained, available locally, and carry a substantial spare parts
inventory to minimize repair time. Simply call your local representative, or call Stryker Adel Customer Service
at (800) 327–0770.
Supplemental Warranty Coverage:
Stryker Adel has developed a comprehensive program of extended warranty options designed to keep your
equipment operating at peak performance at the same time it eliminates unexpected costs. We recommend
that these programs be activated before the expiration of the new product warranty to eliminate the potential
of additional equipment upgrade charges. Stryker Adel offers the following Supplemental Warranties:
Extended (Parts and Labor)
SAll replacement parts (excluding mattresses and consumable items)
SLabor and travel for all scheduled and unscheduled calls
SBiannual Preventive Maintenance Inspections and repairs
SJCAHO paperwork for preventive maintenance
SPriority Emergency Service
Standard (Labor Only):
SLabor and travel for all scheduled and unscheduled calls
SBiannual Preventive Maintenance Inspections and repairs
SJCAHO paperwork for preventive maintenance
SPriority Emergency Service
Basic (Parts Only):
SAll replacement parts (excluding mattresses and consumable items)
SPriority Emergency Service
Please call your local representative, or call (800) 327–0770 for further information
58
Page 62
Warranty
Return Authorization:
Merchandise cannot be returned without approval from the Stryker Adel Customer Service Department. An
authorization number will be provided which must be printed on the returned merchandise. Stryker Adel reserves the right to charge shipping and restocking fees on returned items.
SPECIAL, MODIFIED, OR DISCONTINUED ITEMS NOT SUBJECT TO RETURN.
Damaged Merchandise:
ICC Regulations require that claims for damaged merchandise must be made with the carrier within fifteen
(15) days of receipt of merchandise. DO NOT ACCEPT DAMAGED SHIPMENTS UNLESS SUCH DAMAGE
IS NOTED ON THE DELIVERY RECEIPT AT THE TIME OF RECEIPT. Upon prompt notification, Stryker
Adel will file a freight claim with the appropriate carrier for damages incurred. Claim will be limited in amount
to the actual replacement cost. In the event that this information is not received by Stryker Adel within the
fifteen (15) day period following the delivery of the merchandise, or the damage was not noted on the delivery
receipt at the time of receipt, the customer will be responsible for payment of the original invoice in full.
Claims for any short shipment must be made within thirty (30) days of invoice.
59
Page 63
6300 Sprinkle Road, Kalamazoo, MI 49001–9799 (800) 327–0770
DH 2/94 0309–89–00 REV 1 ECN #4619
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.