Safe Working Load indicates the sum of the patient, mattress, and accessory weight
WARNING/CAUTION/NOTE DEFINITION
The words WARNING, CAUTION and NOTE carry special meanings and should be carefully reviewed.
WARNING
Alerts the reader about a situation which, if not avoided, could result in death or serious injury. It may also describe
potential serious adverse reactions and safety hazards.
CAUTION
Alerts the reader of a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the
user or patient or damage to the equipment or other property. This includes special care necessary for the safe and
effective use of the device and the care necessary to avoid damage to a device that may occur as a result of use or
misuse.
NOTE
Provides special information to make maintenance easier or important instructions clearer.
Return To Table of Contents
41080 -0 09- 001 REV Awww.stryker.com
Introduction
This manual is designed to assist you with the operation of Stryker Model 1080 Fluoroscopy Stretcher. Read this
manual thoroughly before using the equipment or beginning maintenance on it. To ensure safe operation of this
equipment, it is recommended that methods and procedures be established for educating and training staff on the safe
operation of this stretcher.
SPECIFICATIONS
Safe Working Load
Note: Safe Working Load
indicates the sum of the patient,
mattress, and accessory weight.
Overall Stretcher Length/Width87” x 30.5”221 cm/77.5 cm
Patient Surface Length/Width84” x 26”213.4 cm/66 cm
Minimum/Maximum Stretcher Height24”/34”61 cm/86.4 cm
Fowler Angle0° to 90°
Trendelenburg/Reverse Trendelenburg-8° to +8°
500 lb227 kg
Fluoroscopy Window19.5” x 66”49.5 cm/167.6 cm
C−Arm Access Width54”13 7. 2 c m
Stryker reserves the right to change specifications without notice.
www.stryker.com1080 -009- 001 REV A5
Return To Table of Contents
Summary of Safety Precautions
Carefully read and strictly follow the warnings and cautions listed on this page.
Service only by qualified personnel. See the maintenance manual for additional information.
WARNING
• Always apply the brakes when a patient is getting on or off the stretcher. Push on the stretcher to ensure that the
brakes are securely locked. Always engage the brakes unless the stretcher is being moved. Injury could result if
the stretcher moves while a patient is getting on or off the stretcher.
• After raising the siderails, pull firmly on the siderail to ensure that it is securely locked into the fully raised position.
Siderails are not intended to serve as a patient restraint device to keep patients from exiting the unit. Siderails are
intended to keep a patient from inadvertently rolling off the unit. It is the responsibility of the attending medical
personnel to determine the degree of restraint necessary to ensure that a patient will remain in place. Failure to
utilize the siderails properly could result in patient injury.
• When lowering the siderail to the collapsed position, keep extremities of patients and staff away from the siderail
spindles or injury could occur.
• During patient transfer, keep patient and operator extremities away from collapsed siderails or injury could occur.
• Operation of the fowler is a manual procedure. Use caution when raising the fowler while a patient is on the
stretcher. Use proper lifting techniques and get additional assistance, if necessary. Failure to use proper lifting
techniques could cause injury to the operator.
• Keep hands/fingers clear of the area around the fowler release handle and the fowler frame when lowering the
fowler. Injury could result if care is not taken when lowering the fowler.
• Physical restraints, even if properly secured, may result in serious harm to patients and caregivers. The use of
restraint straps may potentially cause entanglement, entrapment, physical injury, and/or death. Caution must be
used in affixing restraint straps to avoid potential injury to both patients and caregivers.
• Restraint straps and/or devices must be attached only at the identified attachment points of the unit. Failure to do
so may result in patient or caregiver injury. Do not attach restraints straps to the siderail.
• This unit accommodates the use of ankle, chest, wrist, and body restraints. The use of restraint straps is regulated
by state and federal restrictions. Users, caregivers, and/or practitioners should refer to the applicable state and
federal restrictions and the appropriate facility protocols before using any restraint strap and/or device.
CAUTION
• To avoid injury or damage to the equipment, do not allow the siderail to lower on its own.
• To avoid damage, the weight of the I.V. bags should not exceed 40 lb.
• To avoid damage while transporting the stretcher, verify that the I.V. pole is at a low enough height to allow it to
safely pass through door openings and under light fixtures.
• Do not use the I.V. pole as a push/pull device because equipment damage could occur.
• Do not use the footboard/chartholder as a push/pull device because equipment damage could occur.
• To avoid damage, do not put items weighing more than 40 lb. in the upright oxygen bottle holder.
• Do not use the upright oxygen bottle holder as a push/pull device because equipment damage could occur.
• Before returning the unit to service after cleaning, ensure that the unit is functioning properly by verifying that all
labels are intact, raise/lower the stretcher height, brake/steer pedal locks properly in both positions, latch/unlatch
the siderails, raise/lower the fowler and gatch, and check all components for proper lubrication.
• Some cleaning products are corrosive in nature and may cause damage to the product if used improperly. If the
products suggested above are used to clean Stryker patient handling equipment, measures must be taken to
ensure that the stretcher is wiped with a damp cloth soaked in clean water and thoroughly dried following cleaning.
Failure to properly rinse and dry the stretcher will leave a corrosive residue on the surface of the stretcher, possibly
causing premature corrosion of critical components. Failure to follow the above directions when using these types
of cleaners may void this product’s warranty.
Return To Table of Contents
61080 -0 09- 001 REV Awww.stryker.com
OPERATING THE BASE CONTROLS
Operation Guide
D
D
A
B
C
TOP VIEW
A
Press to lower end of litter where you are standing.
B
Press to lower end of litter opposite where you are standing.
B
A
C
Lower both ends of the litter together by pressing down
on both pedal (A) and pedal (B) at the same time.
Pump pedal repeatedly to raise litter.
C
STEER
D
NEUTRAL
BRAKE
Lift fully up on pedal (D) to engage the steer caster.
Push fully down on pedal (D) to engage the stretcher brakes.
WARNING
Always apply the brakes when a patient is getting on or off the stretcher. Push on the stretcher to ensure that the brakes
are securely locked. Always engage the brakes unless the stretcher is being moved. Injury could result if the stretcher
moves while a patient is getting on or off the stretcher.
Return To Table of Contents
www.stryker.com1080 -009- 001 REV A7
Operation Guide
OPERATING THE SIDERAILS
Raising and lowering the siderails safely is a two-handed operation. Use one hand to hold and position the siderail and
the other hand to operate the siderail latch.
To raise the siderails, pull up on the siderail and raise it to the highest position until the latch engages.
WARNING
After raising the siderails, pull firmly on the siderail to ensure that it is securely locked into the fully raised position.
Siderails are not intended to serve as a patient restraint device to keep patients from exiting the unit. Siderails are
intended to keep a patient from inadvertently rolling off the unit. It is the responsibility of the attending medical personnel
to determine the degree of restraint necessary to ensure that a patient will remain in place. Failure to utilize the siderails
properly could result in patient injury.
To lower the siderails, pull up on the latch and guide the siderail to the lowest position.
WARNING
• When lowering the siderail to the collapsed position, keep extremities of patients and staff away from the siderail
spindles or injury could occur.
• During patient transfer, keep patient and operator extremities away from collapsed siderails or injury could occur.
CAUTION
To avoid injury or damage to the equipment, do not allow the siderail to lower on its own.
Make sure that the siderail latching mechanism is working properly at all times. If it is not, see the stretcher maintenance
manual for ”Siderail Latch Adjustment”.
Return To Table of Contents
81080 -0 09- 001 REV Awww.stryker.com
Operation Guide
OPERATING THE FOWLER
To raise the fowler, squeeze the fowler handle (A or B) for pneumatic assist until the fowler has reached the desired
angle.
To lower the fowler, squeeze the fowler handle (A or B) and push down until the fowler has reached the desired angle.
Note: You can operate the fowler from both ends of the stretcher litter for convenience when placing a patient on the
stretcher and to allow the patient to be reversed on the table when taking X-rays.
WARNING
• Operation of the fowler is a manual procedure. Use caution when raising the fowler while a patient is on the
stretcher. Use proper lifting techniques and get additional assistance, if necessary. Failure to use proper lifting
techniques could cause injury to the operator.
• Keep hands/fingers clear of the area around the fowler release handle and the fowler frame when lowering the
fowler. Injury could result if care is not taken when lowering the fowler.
If the pneumatic fowler is difficult to operate, see the stretcher maintenance manual for ”Pneumatic Fowler Adjustment”.
A
B
www.stryker.com1080 -009- 001 REV A9
Return To Table of Contents
Operation Guide
OPERATING THE TWO-STAGE PERMANENTLY ATTACHED I.V. POLE
Note: The two-stage permanently attached I.V. pole is an option and may have been installed at either the head, foot
or both ends of the stretcher. The choice was made at the time that the stretcher was purchased.
To use the two-stage permanently attached I.V. pole:
1. Lift and pivot the pole from the storage position and push down until it is locked into the receptacle.
2. To raise the height of the pole, pull up on the telescoping portion (A) until it locks into place at its fully raised
position.
3. Rotate the I.V. hangers (B) to desired position and hang the I.V. bags.
4. To lower the I.V. pole, turn the latch (C) until section (A) lowers.
A
C
B
A
C
I.V. Pole Latch Detail
I.V. Pole
CAUTION
• To avoid damage, the weight of the I.V. bags should not exceed 40 lb.
• To avoid damage while transporting the stretcher, verify that the I.V. pole is at a low enough height to allow it to
safely pass through door openings and under light fixtures.
• Do not use the I.V. pole as a push/pull device because equipment damage could occur.
Return To Table of Contents
101080 -009- 001 REV Awww.stryker.com
Optional Accessories
The accessories listed below can be purchased and installed on the Model 1080 Fluoroscopy Stretcher.
AccessoryPart NumberPage
Footboard/Chart Holder1080-134-000page 12
I.V. Pole, Removable0390-025-000page 13
Oxygen Bottle Holder, Upright1020-130-000page 13
Restraints, Ankle0946-043-000
Restraints, Body/Chest0390-019-000
page 14
Restraints, Wrist0946-044-000
Restraints, Full Strap Package1010-077-000
www.stryker.com1080 -009- 001 REV A11
Return To Table of Contents
Operation Guide
USING THE FOOTBOARD/CHART HOLDER
To use the footboard/chart holder, insert the footboard/
chart holder supports (A) into the corresponding holes
located at the foot end of the stretcher.
CAUTION
Do not use the footboard/chart holder as a push/pull device
because equipment damage could occur.
A
Footboard/Chart Holder
Return To Table of Contents
121080 -0 09-001 REV Awww.stryker.com
Operation Guide
OPERATING THE REMOVABLE I.V. POLE
To use the removable I.V. pole:
1. Remove the I.V. pole from the storage trough under the litter and insert
into the receptacle on the corner of the litter frame.
2. To raise the height of the pole, turn the knob (A) counterclockwise
and pull up on the telescoping portion (B) of the pole to raise it to the
desired height.
3. Turn the knob (A) clockwise to lock the telescoping portion in place.
B
CAUTION
• To avoid damage, the weight of the I.V. bags should not exceed 40 lb.
• To avoid damage while transporting the stretcher, verify that the I.V.
pole is at a low enough height to allow it to safely pass through door
openings and under light fixtures.
• Do not use the I.V. pole as a push/pull device because equipment
damage could occur.
INSTALLING THE UPRIGHT OXYGEN BOTTLE HOLDER
To install the upright oxygen bottle holder:
1. Insert the support bar (A) into the I.V. socket at any
of the four litter corners.
2. Insert the cotter pin (B) through the hole in the
support bar to hold the bottle holder in place.
A
Removable I.V. Pole
CAUTION
• To avoid damage, do not put items weighing more
than 40 lb. in the upright oxygen bottle holder.
• Do not use the upright oxygen bottle holder as a
push/pull device because equipment damage could
occur.
www.stryker.com1080 -009- 001 REV A13
A
B
Upright Oxygen Bottle Holder
Return To Table of Contents
Operation Guide
USING THE RESTRAINT STRAPS
This unit allows the use of ankle, chest, wrist, and body restraints. See the figure below for restraint strap attachment
points. Do not attach restraints straps to the siderail. Stryker makes no recommendation for the use of restraints.
Restraint Strap Locations
WARNING
• Physical restraints, even if properly secured, may result in serious harm to patients and caregivers. The use of
restraint straps may potentially cause entanglement, entrapment, physical injury, and/or death. Caution must be
used in affixing restraint straps to avoid potential injury to both patients and caregivers.
• Restraint straps and/or devices must be attached only at the identified attachment points of the unit. Failure to do
so may result in patient or caregiver injury. Do not attach restraints straps to the siderail.
• This unit accommodates the use of ankle, chest, wrist, and body restraints. The use of restraint straps is regulated
by state and federal restrictions. Users, caregivers, and/or practitioners should refer to the applicable state and
federal restrictions and the appropriate facility protocols before using any restraint strap and/or device.
Return To Table of Contents
14108 0- 009 -0 01 REV Aww w.stryker.com
Cleaning
Hand wash all surfaces of the stretcher with warm water and mild detergent. DRY THOROUGHLY. Do not steam clean,
pressure wash, hose off or ultrasonically clean. Using these methods of cleaning is not recommended and may void
this product’s warranty.
• Chlorinated Bleach Solution (5.25% − less than 1 part bleach to 100 parts water)
Avoid oversaturation and ensure the product does not stay wet longer than the chemical manufacturer’s guidelines for
proper disinfecting.
CAUTION
Some cleaning products are corrosive in nature and may cause damage to the product if used improperly. If the products
suggested above are used to clean Stryker patient handling equipment, measures must be taken to ensure that the
stretcher is wiped with a damp cloth soaked in clean water and thoroughly dried following cleaning. Failure to properly
rinse and dry the stretcher will leave a corrosive residue on the surface of the stretcher, possibly causing premature
corrosion of critical components. Failure to follow the above directions when using these types of cleaners may void this
product’s warranty.
For mattress cleaning instructions, please see the tag on the mattress, or contact the mattress manufacturer.
Clean Velcro® AFTER EACH USE. Saturate Velcro® with disinfectant and allow disinfectant to evaporate. Appropriate
disinfectant for nylon Velcro® should be determined by the hospital.
www.stryker.com1080 -009- 001 REV A15
Return To Table of Contents
Preventative Maintenance
At a minimum, preventative maintenance should be performed annually. A preventative maintenance program should
be established for all Stryker Medical equipment. Preventative maintenance may need to be performed more frequently
based on the usage level of the product.
All fasteners secure
Siderails move and latch properly
Engage brake pedal and push on the stretcher to ensure all casters lock securely
All casters secure and swivel properly
Body restraints working properly
I.V. pole intact and operating properly
Oxygen bottle holder intact and operating properly (Optional Equipment)
Stryker Medical Division, a division of Stryker Corporation, warrants to the original purchaser of the Stryker Model 1080
Fluoroscopy Stretcher to be free from defects in material and workmanship for a period of one (1) year after date of
delivery. Stryker’s obligation under this warranty is expressly limited to supplying replacement parts and labor for, or
replacing, at its option, any product which is, in the sole discretion of Stryker, found to be defective. If requested by
Stryker, products or parts for which a warranty claim is made shall be returned prepaid to the factory. Any improper
use or any alteration or repair by others in such manner as in Stryker’s judgment affects the product materially and
adversely shall void this warranty. Any repair of Stryker products using parts not provided or authorized by Stryker shall
void this warranty. No employee or representative of Stryker is authorized to change this warranty in any way.
Stryker MedicalStretcherproducts are designed for a ten (10)year expected service life under normal use, conditions,
and with appropriate periodic maintenance as described in the maintenance manual for each device. Stryker warrants
to the original purchaser that the welds on its Stretcherproducts will be free from structural defects for the expected
ten (10) year life of the Stretcher product as long as the original purchaser owns the product.
This statement constitutes Stryker’s entire warranty with respect to the aforesaid equipment. Stryker makes no
other warranty or representation, either expressed or implied, except as set forth herein. There is no warranty
of merchantability and there are no warranties of fitness for any particular purpose. In no event shall Stryker be
liable here under for incidental or consequential damages arising from or in any manner related to sales or use of
any such equipment.
Warranty does not include any disposable items, I.V. poles (except for Stryker permanently attached poles), mattresses,
batteries, or damage resulting from abuse.
TO OBTAIN PARTS AND SERVICE
Stryker products are supported by a nationwide network of dedicated Stryker Field Service Representatives. These
representatives are factory trained, available locally, and carry a substantial spare parts inventory to minimize repair
time. Simply call your local representative, or call Stryker Customer Service USA at 1-800-327-0770.
SERVICE CONTRACT COVERAGE
Stryker has developed a comprehensive program of service contract options designed to keep your equipment operating
at peak performance at the same time it eliminates unexpected costs. We recommend that these programs be activated
before the expiration of the new product warranty to eliminate the potential of additional equipment upgrade charges.
A Service Contract helps to:
• Ensure equipment reliability
• Stabilize maintenance budgets
• Diminish downtime
• Establish documentation for JCAHO
• Increase product life
• Enhance trade-in value
• Address risk management and safety
www.stryker.com1080 -009- 001 REV A17
Return To Table of Contents
Warranty
SERVICE CONTRACT PROGRAMS
Stryker offers the following service contract programs:
Service Agreement OptionsPremiumCompleteStandard *
Annually scheduled preventative maintenanceXX
All parts**, labor, and travelXX
Unlimited emergency service callsXX
Priority one contact: two hour phone responseXX
Most repairs will be completed within 3 business daysXX
JCAHO documentationXXX
On-site record of PM & emergency serviceXX
Factory-trained Stryker service technicianXXX
Stryker authorized parts usedXXX
Service during regular business hours (8-5)XXX
* Replacement parts and labor for products under PM contract will be discounted.
** Does not include any disposable items, I.V. poles (except for Stryker permanently at tached poles), mattresses, batteries, or damage resulting from
abuse.
Stryker Medical also offers personalized service contracts.
Pricing is determined by age, location, model and condition of product.
For more information on our service contracts,
please call your local representative.
RETURN AUTHORIZATION
Merchandise cannot be returned without approval from the Stryker Customer Service Department. An authorization
number will be provided which must be printed on the returned merchandise. Stryker reserves the right to charge
shipping and restocking fees on returned items. Special, modified, or discontinued, items not subject to return.
DAMAGED MERCHANDISE
ICC Regulations require that claims for damaged merchandise must be made with the carrier within fifteen (15) days of
receipt of merchandise. Do not accept damaged shipments unless such damage is noted on the delivery receipt at the time of receipt. Upon prompt notification, Stryker will file a freight claim with the appropriate carrier for damages
incurred. Claim will be limited in amount to the actual replacement cost. In the event that this information is not received
by Stryker within the fifteen (15) day period following the delivery of the merchandise, or the damage was not noted on
the delivery receipt at the time of receipt, the customer will be responsible for payment of the original invoice in full.
Claims for any short shipment must be made within thirty (30) days of invoice.
INTERNATIONAL WARRANTY CLAUSE
This warranty reflects U.S. domestic policy. Warranty outside the U.S. may vary by country. Please contact your local
Stryker Medical representative for additional information.
Return To Table of Contents
181080 -0 09- 001 REV Awww.stryker.com
United States
Stryker Medical
3800 E. Centre Ave.,
Portage, Michigan USA
49002
ECREP
European Representative
Stryker France S.A.S.
ZAC - avenue Satolas Green
69881 MEYZIEU Cedex
France
2011/121080-009-001 REV Awww.stryker.com
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.