SX-B
Epilator Manual
•Introduction
•Technical Data Sheet
•Accessories & Controls
•Theory
•Treatment Techniques
•Hints & Tips
•Diathermy/Thermolysis only operation
•Testing Procedure
Introduction
Thank you for buying the Sterex SX-B Blend epilator.
It is essential that you read the following pages of this manual and work through the practical CD, which is included with your SX-B Blend epilator, before you operate
this unit. (Video available on request)
|
Technical Data Sheet |
• ADDRESS: |
Sterex Electrolysis International Ltd |
|
174 Kings Rd |
|
Tyseley |
|
Birmingham |
|
B11 2AP |
|
U.K. |
|
Tel : 0121 708 2404 Fax: 0121 707 0028 |
•EQUIPMENT CLASSIFICATION:
Class 1 Type B Applied Parts.
The equipment is designed for continuous operation.
The equipment is not protected against ingress.
Do not use the equipment near flammable anaesthetic gases.
Type B Equipment
No other electrical equipment may be connected to the patient simultaneously when using this epilator equipment.
•SYMBOLS
Applied parts Type B
Potentially Hazardous Output
The manual must be read before using this equipment.
The EUT (Equipment Under Test) complies with European Electromagnetic Compatibility requirements. However, the EUT may emit electromagnetic radiation, which may interfere with other equipment in the vicinity. Likewise other equipment may interfere with the EUT. If such effects are to be minimised, increasing the distance between equipment and reducing cable lengths where relevant should help.
It is advisable not to use such equipment on people fitted with pace makers and to take medical advice if fitted with any electrical devices.
It is not recommended to use epilators within 3 metres of sensitive electrical appliances such as computers, medical equipment etc.
•MEDICAL DEVICES DIRECTIVE
The SX-B conforms to the Medical Devices EN60601 – 1:1990: Notified Body number = 0120
•FUNCTION
The function of the epilator is to remove body and facial hair using Thermolysis/Diathermy and Galvanic or a combination of the two techniques called “Blend”.
•MAINTENANCE
The exterior of the epilator should be cleaned regularly with a dusting brush/cloth. Dirt, which is difficult to remove, can be cleaned off with a moistened cloth (99% water 1% mild detergent). The equipment must be wiped dry with a lint free cloth. Under no circumstances must the cleaning fluid get into the equipment. The use of other cleaning agents may attack the plastic and paint surface.
Safety/Preventative Maintenance
The equipment operates according to safety Class 1 (three-conductor power cord with protective earth contact). The mains/line plug shall only be inserted into a socket outlet protected with a protective earth contact. The protective action must not be negated by the use of an extension cord without a protective conductor.
The ground accessible metal parts (case, sockets, jacks) and the mains/line supply contacts (line/live, neutral) of the epilator have been tested against insulation breakdown.
Whenever it is likely that protection has been impaired, the equipment shall be made inoperative and be secured against any unintended operation.
The protection is likely to be impaired if, for example, the equipment/mains cable:
1.Shows visible signs of damage
2.Fails to perform the intended user operation
3.Has been subjected to prolonged storage under unfavourable conditions e.g. in an open or moist environment
4.Has been subjected to severe transport stress
•SERVICING OF EQUIPMENT
Servicing of the equipment may only be carried out by Sterex Electrolysis International Ltd or by persons authorised by them.
• SPECIFICATIONS |
|
Operating Temperature |
10 degrees C to 50 degrees C |
Lines/Mains Voltage |
230/240V ac |
Mains Frequency |
50/60 Hz |
Power Consumption |
35 Watts approx. |
Maximum Ambient Temperature |
-10 degrees C to + 40 degrees C |
Protective System |
Class 1 Type B |
Weight |
2.75 kilos approx. |
Maximum available Thermolysis/RF |
0-2 Watts measured using a 5 K Ω non- |
Output Power |
inductive Load Resistor |
Thermolysis/RF Output Frequency |
13.56 MHz |
DC/Galvanic Output |
0-15 Volts at a maximum current of 1 mA |
•RANGE & ACCURACY OF DISPLAYED VALUES
|
Meter |
|
Accuracy |
Available Output Power |
Thermolysis/RF |
0-100% |
In |
1% steps |
0-2 Watts |
Galvanic/DC |
0-100% |
In |
1% steps |
0-15 mW |
Timer |
0-9.95 secs |
In |
0.5 second steps |
|
•ENVIRONMENTAL & STORAGE CONDITIONS
The permissible ambient temperature range for storage or transportation is -40 degrees C to +70 degrees C. The maximum operating altitude is up to 15000M.
The maximum relative humidity is up to 80%.
•OPERATING CONDITIONS
The equipment has been designed for indoor use. The permissible ambient temperature range during operation is -10 degrees C to +40 degrees C. If condensed water exists in the instrument it should be acclimatised before switching on. The epilator should be kept in a clean dry room, and must not be operated in explosive, corrosive, dirty or moist environments.