The Sper Scientic AquaShock® pH Meter is the most waterproof,
rugged and reliable pH meter on the market. It is fully waterproof,
including probe connections, when used with the AquaShock®
probe. As with all AquaShock® meters, the pH Meter is fully
shockproof, oats, and features a double injection molded
case with an integral protective soft grip outer layer, high end
antishock, LED backlight display, and is powered by the latest
environmentally friendly, long lasting, rechargeable Lithium Ion
battery. The meter is highly accurate and stable, very intuitive and
includes all of the functions required by most users. Sper Scientic
guarantees you will not nd any meters more rugged and reliable
than AquaShock®.
45
Page 5
FEATURES
• Floats
• Shockproof
• IP67 Waterproof
• Protective soft grip outer layer
• Rechargeable Lithium Ion battery
• Also accepts standard BNC probes
• Multi-line display LCD screen
• Reads pH and mV
• Automatic or manual temperature compensation
• Automatic buffer recognition
• Celsius or Fahrenheit selectable
• 99 data points for analysis
• Calibration point review
• Internal clock and calendar
• Maximum, minimum and average
• Hold function
• Low battery indicator
• Backlight
MATERIALS SUPPLIED
• Meter
• Dual pH/ATC Probe
• pH Buffers, Distilled Water
• 1 Lithium 1450 mAh 7.4 V Battery
• DC Power Adapter
• Instruction Manual
• Hard Carrying Case
5
Page 6
LCD DISPLAY
1
3
4
5
2
6
7
8
NameDescription
1ReadyDisplayed when measuring
results are stable
2Electrode
Durability Icon
3Scale UnitCurrent measurement scale unit
4Measuring ResultsCurrent measurement value
5ATC/MTCAutomatic/Manual Temperature
6Temperature ScaleCurrent temperature
7Temperature ValueCurrent temperature
8Battery IconCurrent battery power level
Current electrode durability
Compensation indicator
measurement unit
measurement value
67
Page 7
LCD DISPLAY
Probe Life Icon:
More than 90% life remaining
More than 85– 90% life remaining
More than 80- 85% life remaining
Less than 80% life remaining
Battery Power Icon:
Battery voltage is more than 7.7V shows full.
Battery voltage between 7.2V and 7.7V shows 2 cells.
Battery voltage between 6.6V and 7.2V shows 1 cell.
Battery voltage less than 6.6V shows 0 cells and ashes.
Display showing battery charging from
less than 6.6V to fully charged.
POWER SUPPLY
This meter is powered by one rechargeable Lithium 7.4V (1450
mAh) battery. The DC power adapter can be used to charge the
battery even when the meter is turned off. This meter includes a
charge protection function and will automatically stop charging
when the battery is full. A complete re-charge takes approximately
8 hours.
7
Page 8
KEYPAD
1
3
5
7
Name1st Function2nd Function
1Power/Backlight/
Low Battery
2SetupSetting parametersConrm setting
3MI↑Store data↑
4MR↓Recall data↓
5Max/Min/AveMax/Min/Ave
6Range/TempSwitch measuring
7Hold/ExitHoldExit
8CALCalibration
Power on/offBacklight on/off
Set temperature value
scale
2
4
6
8
89
Page 9
SETUP MODE
The Setup Mode allows you to customize the following
meter preferences and defaults:
• Temperature Units
• Clock Setting
• Restore Factory Defaults
• View Calibration Data
• View Probe Parameters
1. Press POWER for 1 second to turn the meter on before
performing any setup function.
Temperature Units
1. Press SETUP to enter the Setup Mode.
2. Press SETUP to enter the Set Temperature screen.
3. The unit °C or °F will ash on the LCD.
4. Press ↑ or ↓ to select the °C or °F temperature unit.
5. Press SETUP to save and return to Normal Mode.
6. Press EXIT to return to Normal Mode without saving the selection.
Real Time Clock
1. Press SETUP to enter the Setup Mode.
2. Press ↑ or ↓ to select the Set Clock screen.
3. Press SETUP to set the year. The value will ash on the LCD.
4. Press ↑ to increase the value by 1. Hold down ↑
to continually increase the value.
Press ↓ to decrease the value by 1. Hold down ↓
to continually decrease the value.
Press SETUP to save and the meter automatically
moves to the month setting.
5. Repeat step 4 to adjust the month, day, hour, and minute.
6. Press EXIT to return to Normal Mode without saving.
Note...
This procedure adjusts the meter’s internal clock. The internal
clock will function when the meter is turned off but only when the
battery is installed.
9
Page 10
Restore Factory Settings
1. Press SETUP to enter the Setup Mode.
2. Press ↑ or ↓ to select the Restore Factory Settings screen.
3. Press SETUP to conrm selection. “NO” will ash on the LCD.
4. Press ↑ or ↓ to select “YES” or “NO.”
5. When “YES” is displayed, press SETUP to conrm and all
parameters will restore to factory settings (except the clock)
and the meter returns to Normal Mode.
6. Press EXIT to return to Normal Mode without saving.
View Calibration Data
1. Press SETUP to enter the Setup Mode.
2. Press ↑ or ↓ to select the View CAL Data screen.
3. Press SETUP to enter the mode.
4. Press ↓ to view the previous calibration point. Press ↑ to
view the next calibration point. The date and time is on
cyclical display.
5. Press EXIT to return to Normal Mode.
View Probe Parameters
1. Press SETUP to enter
the Setup Mode.
2. Press ↑ or ↓ to select the View
Probe Parameters screen.
Note...
If the meter is not calibrated, the screen will show:
Press SETUP or EXIT to
return to Normal Mode.
1011
Page 11
CALIBRATION
Calibration should be performed whenever the probe has been
out of use for 48 hours or longer. To prepare for calibration:
1. Immerse the pH probe in standard buffer solution for
30 minutes or longer.
Calibrate With Temperature Sensor/ATC pH Probe
1. Press CAL for 2 seconds to enter Calibration Mode.
2. The meter will automatically recognize the calibration
buffer. The value and “Finished” will display on
the LCD when the procedure is complete.
3. Press EXIT to return to Normal Mode.
4. Repeat Steps 1-2 to calibrate other points.
Note...
All other keys will be inoperable while using the calibration function.
Calibrate Without Temperature Sensor/MTC pH Probe
1. Press CAL for 2 seconds to enter Temperature Setup Mode.
The temperature value will ash on the LCD.
2. Press ↑ to increase the value. Hold down ↑
to continually increase the value.
Press ↓ to decrease the value. Hold down ↓
to continually decrease the value.
For °C, the range is 0 ~ 60.0°C; for °F,
the range is 32 ~ 140.0°F.
Press SETUP and the meter will automatically recognize the
calibration buffer. The value and “Finished” will display on the
LCD when the procedure is complete.
Repeat Step 2 to calibrate other points.
3. Press EXIT to return to Normal Mode.
11
Page 12
MEASUREMENT PROCEDURES
Turning the Unit On/Off
1. Press POWER to turn the meter on.
2. Press and hold POWER for 2 seconds to turn the meter off.
Backlight
1. With the meter on press POWER to turn the backlight on/off.
Scale Selection
1. Press RANGE to switch between the pH and mV scales.
Setting the MTC Temperature
Do NOT connect the temperature sensor or non-ATC probe
while setting the temperature.
1. From Normal Mode, press RANGE for 2 seconds to enter
Set Temperature Mode. The temperature value will ash
on the LCD.
2. Press ↑ to increase the value. Press and hold ↑
to increase the value continuously.
Press ↓ to decrease the value. Press and hold ↓
to decrease the value continuously.
For °C, the range is 0 ~ 60.0°C; for °F, the range is 32 ~ 140.0°F.
3. Press SETUP to save and the meter returns to Normal Mode.
4. Press EXIT to return to Normal Mode without saving.
Saving to Memory
1. Press MI to save the current value and view the data point
number. There are 99 data points available.
1213
Page 13
Recall Memory
1. Press RANGE to select the scale (pH or mV) for recall.
2. Press MR to enter Recall Mode and view the last measured
value. The date and time are on cyclical display.
3. Press ↑ or ↓ to scroll saved data points.
4. Press EXIT to return to Normal Mode.
Clear Memory
1. Press MI and MR simultaneously for 2 seconds to clear all
memory. The meter will return to Normal Mode.
Maximum, Minimum, and Average
1. Press RANGE to select the scale.
2. Press MAX/MIN/AVE, the maximum recorded value will
appear on the LCD. The date and time is on cyclical display.
3. Press MAX/MIN/AVE, the minimum recorded value will
appear on the LCD. The date and time is on cyclical display.
4. Press MAX/MIN/AVE, the average recorded value will appear
on the LCD. The date and time is on cyclical display.
5. Press MAX/MIN/AVE to return to Normal Mode.
Hold
1. Press HOLD to hold the value.
2. Press HOLD to release the hold value.
Note...
When in Hold Mode, all other functions will be disabled
except turning the meter off, backlight function, saving, and
exiting Hold Mode.
13
Page 14
OPTIONAL ACCESSORIES
840016 Standard pH Probe (non-ATC)
840049 Spear Tip pH Probe (non-ATC)
840092 Bench-Top Tripod
840093 Field Tripod
850057P AquaShock® ATC pH Probe
860008 pH 4 Buffer Solution - 3 bottles of 40ml each
860009 pH 7 Buffer Solution - 3 bottles of 40ml each
860010 pH 10 Buffer Solution - 3 bottles of 40ml each
860011 Deionized Water - 3 bottles of 40ml each
860012 pH Buffer Set - 1 each pH 4, 7, 10
Power OffLight off
Power OnGreen light for 5 seconds, then light off
Low BatteryRed light ashes every 5 seconds
Charging Battery
Blue light on
15
Page 16
WARRANTY
Sper Scientic warrants this product against defects in materials
and workmanship for a period of ve (5) years from the date
of purchase, and agrees to repair or replace any defective unit
without charge. If your model has since been discontinued, an
equivalent Sper Scientic product will be substituted if available.
This warranty does not cover probes, batteries, battery leakage,
or damage resulting from accident, tampering, misuse, or abuse
of the product. Opening the meter to expose its electronics will
break the waterproof seal and void the warranty.
To obtain warranty service, ship the unit postage prepaid to:
SPER SCIENTIFIC LTD.
8281 East Evans Road, Suite #103
Scottsdale, AZ 85260
The defective unit must be accompanied by a description of the
problem and your return address. Register your product online at
www.sperscientic.com, or return your warranty card within 10
days of purchase.