AquaShock® pH Kit
850057K
Page 2
AquaShock® pH Kit 850057K
Copyright ©2014 by Sper Scientic
ALL RIGHTS RESERVED
Printed in the USA
The contents of this manual may not be reproduced or transmitted in any form or by any means electronic, mechanical, or other means that do not yet exist or may be developed, including photocopying, recording, or any information storage and retrieval
system without the express permission from Sper Scientic.
Sper Scientic 8281 E. Evans Rd., Suite #103, Scottsdale, AZ 85260 Tel: (480) 948-4448 Fax: (480) 967-8736 Web: www.sperscientic.com
2 3
Page 3
TABLE OF CONTENTS
INTRODUCTION .....................................4
FEATUR ES ..........................................5
MATERIALS SUPPLIED................................5
LCD DISPLAY .......................................6
POWER SUPPLY .....................................7
KEYPA D ............................................8
SETUP MODE .......................................9
CALIBRATION ......................................11
MEASUREMENT PROCEDURES.......................12
OPTIONAL ACCESSORIES ...........................14
SPECIFICATIONS ...................................15
WARRANTY .......................................16
3
Page 4
INTRODUCTION
The Sper Scientic AquaShock® pH Meter is the most waterproof, rugged and reliable pH meter on the market. It is fully waterproof, including probe connections, when used with the AquaShock® probe. As with all AquaShock® meters, the pH Meter is fully shockproof, oats, and features a double injection molded
case with an integral protective soft grip outer layer, high end
antishock, LED backlight display, and is powered by the latest
environmentally friendly, long lasting, rechargeable Lithium Ion
battery. The meter is highly accurate and stable, very intuitive and includes all of the functions required by most users. Sper Scientic guarantees you will not nd any meters more rugged and reliable than AquaShock®.
4 5
Page 5
FEATURES
• Floats
• Shockproof
• IP67 Waterproof
• Protective soft grip outer layer
• Rechargeable Lithium Ion battery
• Also accepts standard BNC probes
• Multi-line display LCD screen
• Reads pH and mV
• Automatic or manual temperature compensation
• Automatic buffer recognition
• Celsius or Fahrenheit selectable
• 99 data points for analysis
• Calibration point review
• Internal clock and calendar
• Maximum, minimum and average
• Hold function
• Low battery indicator
• Backlight
MATERIALS SUPPLIED
• Meter
• Dual pH/ATC Probe
• pH Buffers, Distilled Water
• 1 Lithium 1450 mAh 7.4 V Battery
• DC Power Adapter
• Instruction Manual
• Hard Carrying Case
5
Page 6
LCD DISPLAY
1
3
4
5
2
6
7
8
Name Description
1 Ready Displayed when measuring
results are stable
2 Electrode
Durability Icon 3 Scale Unit Current measurement scale unit 4 Measuring Results Current measurement value
5 ATC/MTC Automatic/Manual Temperature
6 Temperature Scale Current temperature
7 Temperature Value Current temperature
8 Battery Icon Current battery power level
Current electrode durability
Compensation indicator
measurement unit
measurement value
6 7
Page 7
LCD DISPLAY
Probe Life Icon:
More than 90% life remaining
More than 85– 90% life remaining
More than 80- 85% life remaining
Less than 80% life remaining
Battery Power Icon:
Battery voltage is more than 7.7V shows full.
Battery voltage between 7.2V and 7.7V shows 2 cells.
Battery voltage between 6.6V and 7.2V shows 1 cell.
Battery voltage less than 6.6V shows 0 cells and ashes.
Display showing battery charging from
less than 6.6V to fully charged.
POWER SUPPLY
This meter is powered by one rechargeable Lithium 7.4V (1450 mAh) battery. The DC power adapter can be used to charge the battery even when the meter is turned off. This meter includes a
charge protection function and will automatically stop charging
when the battery is full. A complete re-charge takes approximately 8 hours.
7
Page 8
KEYPAD
1
3
5
7
Name 1st Function 2nd Function
1 Power/Backlight/
Low Battery
2 Setup Setting parameters Conrm setting
3 MI↑ Store data
4 MR↓ Recall data
5 Max/Min/Ave Max/Min/Ave
6 Range/Temp Switch measuring
7 Hold/Exit Hold Exit
8 CAL Calibration
Power on/off Backlight on/off
Set temperature value
scale
2
4
6
8
8 9
Page 9
SETUP MODE
The Setup Mode allows you to customize the following meter preferences and defaults:
• Temperature Units
• Clock Setting
• Restore Factory Defaults
• View Calibration Data
• View Probe Parameters
1. Press POWER for 1 second to turn the meter on before
performing any setup function.
Temperature Units
1. Press SETUP to enter the Setup Mode.
2. Press SETUP to enter the Set Temperature screen.
3. The unit °C or °F will ash on the LCD.
4. Press or to select the °C or °F temperature unit.
5. Press SETUP to save and return to Normal Mode.
6. Press EXIT to return to Normal Mode without saving the selection.
Real Time Clock
1. Press SETUP to enter the Setup Mode.
2. Press or to select the Set Clock screen.
3. Press SETUP to set the year. The value will ash on the LCD.
4. Press to increase the value by 1. Hold down
to continually increase the value.
Press ↓ to decrease the value by 1. Hold down to continually decrease the value.
Press SETUP to save and the meter automatically
moves to the month setting.
5. Repeat step 4 to adjust the month, day, hour, and minute.
6. Press EXIT to return to Normal Mode without saving.
Note...
This procedure adjusts the meter’s internal clock. The internal clock will function when the meter is turned off but only when the battery is installed.
9
Page 10
Restore Factory Settings
1. Press SETUP to enter the Setup Mode.
2. Press or to select the Restore Factory Settings screen.
3. Press SETUP to conrm selection. “NO” will ash on the LCD.
4. Press or to select “YES” or “NO.”
5. When “YES” is displayed, press SETUP to conrm and all
parameters will restore to factory settings (except the clock) and the meter returns to Normal Mode.
6. Press EXIT to return to Normal Mode without saving.
View Calibration Data
1. Press SETUP to enter the Setup Mode.
2. Press or to select the View CAL Data screen.
3. Press SETUP to enter the mode.
4. Press ↓ to view the previous calibration point. Press to
view the next calibration point. The date and time is on cyclical display.
5. Press EXIT to return to Normal Mode.
View Probe Parameters
1. Press SETUP to enter
the Setup Mode.
2. Press or to select the View
Probe Parameters screen.
Note...
If the meter is not calibrated, the screen will show:
Press SETUP or EXIT to
return to Normal Mode.
10 11
Page 11
CALIBRATION
Calibration should be performed whenever the probe has been
out of use for 48 hours or longer. To prepare for calibration:
1. Immerse the pH probe in standard buffer solution for
30 minutes or longer.
Calibrate With Temperature Sensor/ATC pH Probe
1. Press CAL for 2 seconds to enter Calibration Mode.
2. The meter will automatically recognize the calibration
buffer. The value and “Finished” will display on the LCD when the procedure is complete.
3. Press EXIT to return to Normal Mode.
4. Repeat Steps 1-2 to calibrate other points.
Note...
All other keys will be inoperable while using the calibration function.
Calibrate Without Temperature Sensor/MTC pH Probe
1. Press CAL for 2 seconds to enter Temperature Setup Mode.
The temperature value will ash on the LCD.
2. Press to increase the value. Hold down
to continually increase the value.
Press ↓ to decrease the value. Hold down ↓ to continually decrease the value.
For °C, the range is 0 ~ 60.0°C; for °F, the range is 32 ~ 140.0°F.
Press SETUP and the meter will automatically recognize the
calibration buffer. The value and “Finished” will display on the LCD when the procedure is complete.
Repeat Step 2 to calibrate other points.
3. Press EXIT to return to Normal Mode.
11
Page 12
MEASUREMENT PROCEDURES
Turning the Unit On/Off
1. Press POWER to turn the meter on.
2. Press and hold POWER for 2 seconds to turn the meter off.
Backlight
1. With the meter on press POWER to turn the backlight on/off.
Scale Selection
1. Press RANGE to switch between the pH and mV scales.
Setting the MTC Temperature
Do NOT connect the temperature sensor or non-ATC probe
while setting the temperature.
1. From Normal Mode, press RANGE for 2 seconds to enter
Set Temperature Mode. The temperature value will ash on the LCD.
2. Press to increase the value. Press and hold
to increase the value continuously.
Press ↓ to decrease the value. Press and hold ↓ to decrease the value continuously.
For °C, the range is 0 ~ 60.0°C; for °F, the range is 32 ~ 140.0°F.
3. Press SETUP to save and the meter returns to Normal Mode.
4. Press EXIT to return to Normal Mode without saving.
Saving to Memory
1. Press MI to save the current value and view the data point
number. There are 99 data points available.
12 13
Page 13
Recall Memory
1. Press RANGE to select the scale (pH or mV) for recall.
2. Press MR to enter Recall Mode and view the last measured
value. The date and time are on cyclical display.
3. Press or to scroll saved data points.
4. Press EXIT to return to Normal Mode.
Clear Memory
1. Press MI and MR simultaneously for 2 seconds to clear all
memory. The meter will return to Normal Mode.
Maximum, Minimum, and Average
1. Press RANGE to select the scale.
2. Press MAX/MIN/AVE, the maximum recorded value will
appear on the LCD. The date and time is on cyclical display.
3. Press MAX/MIN/AVE, the minimum recorded value will
appear on the LCD. The date and time is on cyclical display.
4. Press MAX/MIN/AVE, the average recorded value will appear
on the LCD. The date and time is on cyclical display.
5. Press MAX/MIN/AVE to return to Normal Mode.
Hold
1. Press HOLD to hold the value.
2. Press HOLD to release the hold value.
Note...
When in Hold Mode, all other functions will be disabled
except turning the meter off, backlight function, saving, and exiting Hold Mode.
13
Page 14
OPTIONAL ACCESSORIES
840016 Standard pH Probe (non-ATC)
840049 Spear Tip pH Probe (non-ATC) 840092 Bench-Top Tripod 840093 Field Tripod
850057P AquaShock® ATC pH Probe 860008 pH 4 Buffer Solution - 3 bottles of 40ml each 860009 pH 7 Buffer Solution - 3 bottles of 40ml each 860010 pH 10 Buffer Solution - 3 bottles of 40ml each 860011 Deionized Water - 3 bottles of 40ml each 860012 pH Buffer Set - 1 each pH 4, 7, 10
14 15
Page 15
SPECIFICATIONS
Mode pH Temperature mV Range 0 ~ 14 pH 32 ~ 140°F
(0 ~ 60°C)
-1999 ~
1999 mV Resolution 0.01 pH 0.1°F (0.1°C) 0.1 mV Accuracy ±0.01 pH ±0.9°F (±1.0°C) ±0.3 mV Input BNC
Connector Calibration
Memory
pH Slope Range
Temp Compen­sation
Operating Temp
Battery
Battery Life
Up to 5 points
Up to 99 data points
75% ~ 110%
Automatic (ATC) or manual (MTC) from
32 ~ 140°F or 0 ~ 60°C
32 ~ 122°F (0 ~ 50°C)
Lithium battery 7.4 V (1450 mA/h)
> 200 hours
Temperature Probe Connector
BNC Connector
Power Button Indicator Light
Power Off Light off Power On Green light for 5 seconds, then light off Low Battery Red light ashes every 5 seconds Charging Battery
Blue light on
15
Page 16
WARRANTY
Sper Scientic warrants this product against defects in materials and workmanship for a period of ve (5) years from the date
of purchase, and agrees to repair or replace any defective unit
without charge. If your model has since been discontinued, an equivalent Sper Scientic product will be substituted if available. This warranty does not cover probes, batteries, battery leakage,
or damage resulting from accident, tampering, misuse, or abuse
of the product. Opening the meter to expose its electronics will break the waterproof seal and void the warranty.
To obtain warranty service, ship the unit postage prepaid to:
SPER SCIENTIFIC LTD.
8281 East Evans Road, Suite #103
Scottsdale, AZ 85260
The defective unit must be accompanied by a description of the
problem and your return address. Register your product online at www.sperscientic.com, or return your warranty card within 10 days of purchase.
Revised 01/10/14