Spectrum SP-UV 200 Series, SP-VIS 200, SP-UV 200 Operating Manual

SP-UV 200
SP-UV 200 Series UV-Vis Spectrophotometer
Operating Manual
User’s Manual SP-UV 200 1 Spectrum Instruments
SPECTRUM INSTRUMENTS AUST PTY LTD.
Add: 1211/ 7, Riverside Quay, South Bank, Victoria 3006, Australia. TEL: +61 3 90415554, FAX: +61 3 90415554 Email: info@spectrum-instruments.com.au
General information about Spectrum Instruments Aust Pty Ltd. on the internet; http://www.spectrum-instruments.com.au
Copyrights and Trademarks
Microsoft, Windows 2000, Windows XP, Windows NT, MS Excel are registered trademarks of Microsoft Corp.
Edition – September 2013
Technical documentation made by: Spectrum Instruments
This documentation describes the state of this product at the time of publishing. It needs not necessarily agree with future versions of the product. Subject to change!
© Copyright 2013 Spectrum Instruments Aust Pty Ltd.
User’s Manual SP-UV 200 2 Spectrum Instruments
Content Page
1 Introduction……………………………………………... 4
1.1 Warning and safety labels on SP-UV 200………….…. 4
1.2 Notes on the User's Manual……………………………. 5
2 Safety notes…………………………………………….. 6
3 Disposal…………………………………………….……. 8
4 Performance Specifications…………………….……. 8
5
Design of the SP-UV 200………………………….…….. 9
6 Unpacking instructions……………………….………. 10
6.1 Installation and transport conditions………….……….. 10
6.2 Installation and start-up……………………….………… 10
6.3 Select light source…………………………………….…. 11
6.4 Turn the Lamps on/off……………………………….….. 11
7 Maintenance……………………………………………. 12
7.1 Replacing Halogen lamp………………………….…….. 12
7.2 Replacing Deuterium lamp……………………………... 13
8 Accessories……………………………………………... 15
9 Display and Buttons of SP-UV 200………….………. 17
10 Software function…………………………….………… 18
11 Connect to PC…………………………………………...
24
User’s Manual SP-UV 200 3 Spectrum Instruments
1 Introduction
The SP-UV 200 is a single beam spectrophotometer. The instrument is compact, easy-to-use, and
ruggedness. The large sample compartment can accept various accessories and cell from 10mm to 100mm
path length. The built-in RS-232 serial interface enables the instrument to be connected to a printer.
The SP-UV 200 is ideal to use in a clinical laboratory, biochemical laboratory, petrochemical laboratory,
environmental protection and various fields of quality control. This user’s manual provides the description of
the SP-UV 200, a list of main specifications and available accessories.
After carefully unpacking the content, check the materials against the packing list to ensure that you
have received everything in good condition. If some parts are missing, damaged or have any other defects,
please contact you dealer or the sales representative immediately.
1.1 Warning and safety labels on SP-UV 200
The following warning and safety labels are located on the SP-UV 200:
Caution! Hazardous area. Do not remove this cover under any circumstances. Trained
service peoples only. No serviceable components inside. Select the right voltage.
Disconnect the power cable before opening the device cover or replacing fuses!
User’s Manual SP-UV 200 4 Spectrum Instruments
1.2 Notes on the User's Manual
In this manual, the following symbols are used to refer you to warnings and notes:
Warning!
Warning messages alert you to a specific procedure or practice which, if not followed correctly, could cause
personal injury.
Caution!
Caution messages refer you to procedures which, if not observed, could result in damage to the equipment.
Caution! Dangerous electric voltage!
Caution! Emission of UV radiation!
Disconnect power cable!
Note
This note must be followed to obtain correct measurement results.
For easier navigation within the manual, the manual uses the following system:
Chapters and illustrations are numbered consecutively.
Every illustration has its own caption.
Steps of operation are numbered.
Cross-references are marked by an arrow (e.g. Notes on the User's Manual p.5)
User’s Manual SP-UV 200 5 Spectrum Instruments
2 Safety notes
For your personal safety and for a trouble-free operation, carefully read this chapter before starting up
the SP-UV 200.
Observe all safety notes given in this manual and all messages and notes displayed by the control
software on the screen.
In addition, observe the safety notes for system components of other manufacturers (e.g. PC, printer)
supplied together with the device.
In particular, you should also observe the safety notes given on the labels and the information on the
use, storage and disposal of reagents, reaction cells and the notes on the package inserts of test kits.
Intended use!
The SP-UV 200, including original accessories, may only be used for the applications described in this
manual. The manufacturer cannot assume any liability for any other application, including that of individual
modules or single parts. This also applies to all service or repair work that is not carried out by authorized
service personnel. All warranty claims shall be forfeited.
Local safety regulations!
Observe the local safety regulations relevant to the operation of the device (e.g. labor safety regulations,
regulations for the prevention of and protection against accidents).
References to possible risks given in this manual do not replace the relevant labor safety regulations that
are to be followed.
Personnel!
The SP-UV 200 may only be operated by appropriately qualified and trained personnel.
For that, the knowledge of the information provided by this manual is indispensable and taken for granted.
Emergency stop!
In the case of emergency, disconnect the SP-UV 200and its components from line power by removing the
power plug from the power outlet.
Caution! This procedure involves the risk of data loss and damage to the operating system!
Electric shocks!
The SP-UV 200 is electrically powered. At several parts inside the system, extremely hazardous electrical
voltages are accessible!
User’s Manual SP-UV 200 6 Spectrum Instruments
To ensure Protection Class I (protective earth connectio
n) of the device, connect the power plug of the
device only to a power outlet with protective earth conductor. The protective effect must not be undone by
the use of an extension cable without protective earth conductor.
Before opening the device or removing any device covers, disconnect the power cable!
Before connecting the SP-UV 200 to the power outlet, make sure the line voltage supplied agrees with the
operating voltage specified on the rating plate at the rear panel of the device. Operation of the device with
any other operating voltage than that specified may result in the destruction of the device.
Only use fuses of the specified type.
Do not operate the device in explosion-risk rooms!
UV radiation!
Never look directly or indirectly, e.g. via a mirror, into the radiation emitted by the lamp
to avoid the risk of getting conjunctivitis!
Therefore, switch off the lamp for lamp replacement!
Lamp damage!
Do not grasp functioning halogen and deuterium lamps on their glass bulbs.
Contamination on the bulb affects the radiation properties of the lamps.
Especially protect the beam-exit window of the deuterium lamp.
Accumulation of heat
Heat accumulation might result in overheating and faults on the device.
Make sure not to cover the ventilation slots of the SP-UV 200!
Before you switch on the SP-UV 200, remove its dust cover!
Handling liquids
Take care to ensure that no liquids can enter the SP-UV 200 to avoid any damage to the device.
Risk of corrosion
Do not set up the device in the vicinity of aggressive vapors, e.g. strongly corrosive acid or caustic vapors!
The vapors might corrode the connections, mechanical and optical components of the device.
Avoid exposing the sample compartment to strongly corrosive substances! For the analysis of such samples,
User’s Manual SP-UV 200 7 Spectrum Instruments
use vapor-tight stoppered cells.
After the measureme
nt, do not leave the samples unnecessarily long in the sample compartment of the
device!
3 Disposal
The owner/operator of the SP-UV 200 must ensure to dispose of the waste material (samples) cropping up
in the analyses in accordance with the relevant legal and local regulations.
When the service life of the device is over, dispose of the device with its electronic components as electronic
scrap observing the relevant regulations.
4 Performance Specifications
Model SP-UV 200 SP-VIS 200
Optic system 1200 grooves/mm diffraction
grating monochromator.
1200 grooves/mm diffraction
grating monochromator.
Optic type Single beam Single beam
Light path 10mm-100mm 10mm-100mm
Spectrum band width 2nm 4nm
Wavelength range 190-1100nm(0.1nm increment) 320-1100nm(0.1nm increment)
Wavelength accuracy ±0.5nm ±0.5nm
Wavelength repeatability ±0.3nm ±0.3nm
Stray light 0.05%T (at 220nm NaI& 340nm,
NaNO
2
)
0.3%T (at 360nm, NaNO
2
)
Photometric mode T, A and Conc. measurement T, A and Conc. measurement
Photometric range 0-125.0%T, -1.0-1.000A,
0.000-9,999 Conc.
0-125.0%T, -1.0-4.000A,
0.000-9,999 Conc.
Photometric accuracy ±0.3%T ±0.3%T
Photometric repeatability ±0.2%T ±0.2%T
Baseline stability ±0.002A/hr(@500nm, after lamp
turn on for 1 hour)
±0.002A/hr(@500nm, after lamp
turn on for 1 hour)
Baseline flatness ±0.002A ±0.003A
Noise level 0.001A(at 0A, 500nm) 0.001A(at 0A, 500nm)
Light source Tungsten-Halogen & Deuterium Tungsten-Halogen
Detector Silicon Photodiode Silicon Photodiode
User’s Manual SP-UV 200 8 Spectrum Instruments
Loading...