Spectrex Xenon 700 SIL User Manual

Xenon 700 SIL
Open-Path Gas Detection System
User's and Maintenance Manual
ATEX, IECEx Approved
Ex II 2(1) G, Ex d e ia [ia Ga] IIC T5 Gb Ta
Document Ref: TM 799200, Rev. C January 2014
Phone: +1 (973) 239 8398 Fax: +1 (973) 239 7614
Web-Site: www.spectrex.net; Email: spectrex@spectrex.net
SafEyeTM Xenon 700S SIL Gas Detector
ii

Legal Notice

Warning: This manual should be carefully read by all individuals who have or will have responsibility for using, maintaining or servicing the product.
The Source and Detector are not field-repairable due to the meticulous alignment and calibration of the sensors and the respective circuits. Do not attempt to modify or repair the internal circuits or change their settings, as this will impair the system's performance and void the Spectrex, Inc. Product warranty.
The SafEye monitoring system described in this document is the property of Spectrex, Inc. No part of the hardware, software or documentation may be reproduced, transmitted,
transcribed, stored in a retrieval system, or translated into any language or computer language, in any form or by any means, without prior written permission of Spectrex, Inc.
While great efforts have been made to assure the accuracy and clarity of this document, Spectrex, Inc. assumes no liability resulting from any omissions in this document, or from misuse of the information obtained herein. The information in this document has been carefully checked and is believed to be entirely reliable with all of the necessary information included. Spectrex Inc. reserves the right to make changes to any products described herein to improve reliability, function, or design, and reserves the right to revise this document and make changes from time to time in content hereof with no obligation to notify any persons of revisions or changes. Spectrex, Inc. does not assume any liability arising out of the application or any use of any product or circuit described herein; neither does it convey license under its patent rights or the rights of others.
.
Legal Notice iii
SafEyeTM Xenon 700S SIL Gas Detector
Rev
Date
Revision History
Prepared by
Approved by
A
June 2011
First Release
Ian Buchanan
Eric Zinn
B
August 2013
Second Release
Ian Buchanan
Eric Zinn
C
January 2014
Third Release
Ian Buchanan
Eric Zinn

Release History

iv Release History

About this Guide

This guide describes the SafEyeTM Xenon 700S SIL Gas Detection System and its features and provides instructions on how to install, operate and maintain the detector.
This guide includes the following chapters and appendixes: Chapter 1, Scope, provides a general introduction and overview of the
product and the guide, with a brief description of its content.
Chapter 2, Technical Description, describes the Detector’s theory of
operation.
Chapter 3, Operation Mode, describes the Detector’s operation
modes, user interface and indications.
Chapter 4, Technical Specifications, describes the Detector’s
electrical, mechanical and environmental specifications.
Chapter 5, Installation Instructions, describes how to install the
Detector, including wiring and mode settings.
Chapter 6, Operating Instructions, describes the operating
instructions and power-up procedures.
Chapter 7, Maintenance Instructions, contains the maintenance and
support procedures.
Chapter 8, Troubleshooting, describes the solutions to problems that
may arise with the Detector.
Appendix A, Wiring Selection Tables, describes the electrical wire
selection according to the installation configuration.
Appendix B, Wiring Option Configurations, provides wiring diagrams
for installation.
Appendix C, Special Condition to Comply to SIL-2 Requirements,
describes the special conditions in order to comply with SIL-2 requirements.
About this Guide v
SafEyeTM Xenon 700S SIL Gas Detector
Abbreviation
Meaning
ATEX
Atmosphere Explosives
AWG
American Wire Gauge
BIT
Built In Test
EMC
Electromagnetic Compatibility
EOL
End of Line
FOV
Field of View
HART
Highway Addressable Remote Transducer­communication protocol
IAD
Immune at Any Distance
IECEx
International Electrotechnical Commission Explosion
IPA
Isopropyl Alcohol
IR
Infrared
JP5
Jet Fuel
Latching
Refers to relays remaining in the ON state even after the ON condition has been removed
LED
Light Emitting Diode
LPG
Liquefied Petroleum Gas
mA
MilliAmps (0.001 amps)
MODBUS
Master-slave messaging structure
N.C.
Normally Closed
N.O.
Normally Open
N/A
Not Applicable
NFPA
National Fire Protection Association
NPT
National Pipe Thread
SIL
Safety Integrity Level
UNC
Unified Coarse Thread
VAC
Volts Alternating Current

Abbreviations and Acronyms

vi Abbreviations and Acronyms
Table of Contents
Xenon 700 SIL Open-Path Gas Detection System
User's and Maintenance Manual ....................................................................... i
Legal Notice .................................................................................................... iii
Release History ............................................................................................... iv
About this Guide .............................................................................................. v
Abbreviations and Acronyms ............................................................................. vi
1 Scope ......................................................................................................... 1
1.1 Product Overview ................................................................................... 1
2 Technical Description ................................................................................ 3
2.1 Features ............................................................................................... 3
2.2 Applications ........................................................................................... 4
2.3 Principle of Operation ............................................................................. 4
2.3.1 Definitions of Terms ......................................................................... 4
2.3.2 Spectral Finger Print ......................................................................... 5
2.3.3 Optical Path .................................................................................... 5
2.3.4 Microprocessor Based ....................................................................... 5
2.3.5 Gas Sensitivity ................................................................................ 5
2.3.6 Gas and Mixture Selection and Setting ................................................ 6
2.3.7 Flash Source ................................................................................... 6
2.3.8 Heated Optics .................................................................................. 7
2.3.9 HART Protocol ................................................................................. 7
2.3.10 Handheld Unit ................................................................................. 8
2.3.11 Modbus RS-485 ............................................................................... 8
2.3.12 Tilt Mount ....................................................................................... 8
2.4 Product Marking ..................................................................................... 8
2.5 Models and Types .................................................................................. 9
2.6 Description ........................................................................................... 10
2.6.1 Flash Source Unit ............................................................................ 11
2.6.2 Detector Unit .................................................................................. 12
3 Operational Modes ................................................................................... 15
3.1 Operation Modes ................................................................................... 15
3.1.1 Normal Mode .................................................................................. 15
3.1.2 Maintenance Call Mode (3mA Output) ................................................ 15
3.1.3 Fault Mode ..................................................................................... 16
Table of Contents vii
SafEyeTM Xenon 700S SIL Gas Detector
3.1.4 Zero Calibration Mode (1mA Output) ................................................. 16
3.2 Output Signals ...................................................................................... 17
3.2.1 Standard 0-20mA Current Output ..................................................... 17
3.2.2 Dry Contact Relays ......................................................................... 18
3.2.3 RS-485 Interface ............................................................................ 18
3.3 System Set Up ..................................................................................... 18
3.3.1 Detection Function Programming ...................................................... 19
3.3.2 Detection Set-Up Function ............................................................... 19
3.3.3 Default Detector Set up ................................................................... 21
4 Technical Specifications ........................................................................... 23
4.1 General Specifications ........................................................................... 23
4.2 Electrical Specifications .......................................................................... 24
4.2.1 Power Consumption ........................................................................ 24
4.2.2 Electrical Input Protection ................................................................ 24
4.2.3 Electrical Interface .......................................................................... 24
4.2.4 Electrical outputs ............................................................................ 24
4.3 Mechanical Specifications ....................................................................... 26
4.4 Environmental Specifications .................................................................. 26
4.4.1 High Temperature ........................................................................... 26
4.4.2 Low Temperature ............................................................................ 26
4.4.3 Humidity ........................................................................................ 27
4.4.4 Salt and Fog ................................................................................... 27
4.4.5 Water and Dust .............................................................................. 27
4.4.6 Shock and Vibration ........................................................................ 27
4.4.7 Electromagnetic Compatibility (EMC): ................................................ 27
5 Installation Instructions .......................................................................... 29
5.1 Introduction ......................................................................................... 29
5.2 General Considerations .......................................................................... 29
5.2.1 Personnel ....................................................................................... 29
5.2.2 Tools Required................................................................................ 29
5.2.3 Site Requirements .......................................................................... 30
5.2.4 The Source and Detector ................................................................. 30
5.2.5 Guidance Tips for Gas Detector Locations........................................... 30
5.3 Preparations for Installation ................................................................... 30
5.3.1 Preparing for the Installation ............................................................ 31
viii Table of Contents
TM 799200 Rev (C), January 2014
5.4 Certification Instructions ........................................................................ 32
5.5 Conduit/Cable Installation ...................................................................... 33
5.6 Detector/Source Mounting ...................................................................... 33
5.6.1 Tilt Kit P/N 799640.......................................................................... 33
5.6.2 Detector/Source Installation ............................................................. 33
5.7 Detector Wiring .................................................................................... 34
5.8 Detector Terminal Wiring ....................................................................... 38
5.9 Flash Source Wiring............................................................................... 39
5.9.1 Wiring ........................................................................................... 39
5.9.2 Terminal Wiring .............................................................................. 39
6 Operating Instructions ............................................................................ 41
6.1 SafEye Operation .................................................................................. 41
6.2 Alignment of Unit .................................................................................. 41
6.3 Powering up the System ........................................................................ 42
6.4 Safety Precautions ................................................................................ 43
6.5 Signal Verification ................................................................................. 43
6.5.1 Signal Values Limitation ................................................................... 43
6.6 Zero Calibration .................................................................................... 44
6.7 Functional Check of Unit ........................................................................ 45
7 Maintenance Instructions ........................................................................ 47
7.1 General Maintenance ............................................................................. 47
7.2 Periodic Maintenance ............................................................................. 47
7.2.1 Routine Optical Surface Cleaning ...................................................... 48
7.2.2 Signal Verification ........................................................................... 48
7.2.3 Function Check of Unit ..................................................................... 48
8 Troubleshooting ...................................................................................... 49
8.1 Troubleshooting Problems ...................................................................... 49
Appendices .................................................................................................... 51
A Wire Selection Tables .............................................................................. 53
A.1 General Instructions For Electrical Wiring ................................................. 53
B Wiring Option Configurations .................................................................. 55
B.1 RS-485 Communication Network ............................................................. 60
Table of Contents ix
SafEyeTM Xenon 700S SIL Gas Detector
C Special Conditions to Comply to SIL-2 Requirements .............................. 61
C.1 Safety relevant parameters (700S) ......................................................... 61
C.2 Guidelines for Configuration, Installation, Operation and Maintenance ......... 61
C.2.1 Conditions for Safe Operation ........................................................... 61
C.2.2 Alarm Operation using the 0-20mA Signal Current .............................. 62
C.2.3 Relay Outputs ................................................................................ 62
C.3 Miscellaneous ....................................................................................... 62
Technical Support ......................................................................................... 63
x Table of Contents
TM 799200 Rev (C), January 2014
List of Figures
Figure 1: Wiring Options .................................................................................... 10
Figure 2: Flash Source Unit ................................................................................ 11
Figure 3: Detector Unit ...................................................................................... 13
Figure 4: Tilt Mount ........................................................................................... 35
Figure 5: Detector and Tilt Mount Assembly .......................................................... 36
Figure 6: Detector with Cover Removed ............................................................... 37
Figure 7: Source with Cover Removed ................................................................. 40
Figure 8: Magnetic Mode Selector ........................................................................ 44
Figure 9: Wiring Option 1 ................................................................................... 55
Figure 10: Wiring Option 2 ................................................................................. 56
Figure 11: Wiring Option 3 (Default) .................................................................... 57
Figure 12: 0-20mA Wiring (Sink) For Wiring Option 1 and 3 ................................... 58
Figure 13: 0-20mA Wiring (Source) For Wiring Option 1 and 3 ................................ 58
Figure 14: 0-20mA 3 Wire Connection for Option 1 and 3 ....................................... 59
Figure 15: RS-485 Networking for Wiring Option 3 ................................................ 60
List of Figures xi
SafEyeTM Xenon 700S SIL Gas Detector
List of Tables
Table 1: Gas Concentrations Measurement Terms .................................................. 4
Table 2: Gas Calibrations .................................................................................... 6
Table 3: Model Numbers and Installation Distances ................................................ 9
Table 4: Detector Types ..................................................................................... 12
Table 5: Standard (default) 0-20mA Current for the Gas Channel ............................ 17
Table 6: Optional - Discrete Reading of 0-20mA at Different Detector Modes ............ 18
Table 7: Sensitivity Levels Options ...................................................................... 19
Table 8: Other Functions .................................................................................... 20
Table 9: Detector Default Set Up ......................................................................... 21
Table 10: Source Default Set-Up ......................................................................... 21
Table 11: Detection Distance Range .................................................................... 23
Table 12: Detector and Source Maximum Power Consumption ................................ 24
Table 13: Dry Contact Relays.............................................................................. 25
Table 14: Tilt Mount Kit ...................................................................................... 33
Table 15: Tilt Mount Description .......................................................................... 35
Table 16: Detector and Tilt Mount Assembly Description ........................................ 36
Table 17: Detector with Cover Removed Description .............................................. 37
Table 18: Wiring Options .................................................................................... 38
Table 19: Source with Cover Removed Description ................................................ 40
Table 20: Maintenance Channels Limited Values .................................................... 43
Table 21: Troubleshooting Problems .................................................................... 49
Table 22: Maximum DC resistance at 68˚F for copper wire ..................................... 53
Table 23: Wiring length in feet (meter) ................................................................ 54
xii List of Tables

1 Scope

Product Overview
page 1
In this chapter…

1.1 Product Overview

The SafEye ‘Xenon’ IR Open-Path Gas Detector 700S employs an advanced Xenon Flash Source and integrated electronics package, both housed in improved, low-profile stainless steel housings to provide high quality and performance, fast response, line of sight gas monitoring. This high quality is backed by 3 years warranty for the complete SafEye system and 10 years for the Xenon Flash source bulb.
SafEye ‘Xenon’ detects ambient combustible gases over a path length of up
to 459 ft. (140 m), even in harsh environments where dust, fog, rain, snow or vibration can cause a high reduction of signal. Due to its unique combination of triple optics, dual-spectrum reference sensor, and heated optics, the SafEye Xenon can maintain operation in up to 90% signal obscuration and ±1 degree of misalignment.
SafEye Xenon is manufactured only in stainless steel with heated optical window to improve performance in icing, snow, and condensation conditions. The programmable functions are available through a RS-485 port used with Host software supplied by Spectrex and a standard PC or an I.S handheld unit.
The SafEye Source and Detector unit enclosures are ATEX and IECEx certified Exd flameproof with an integral segregated rear, Exe terminal compartment (which avoids exposure of the sensors and electronics to surrounding environment), and an internal Exia intrinsically safe circuit power for the heating of the window. The Detector also has a plug interface for connection to handheld PC, which is intrinsically safe. Hence the combined approval - Ex d e ia [ia Ga] IIC T5 Gb Ta.
This manual consists of a full description of the Detector and its features. It contains instructions on the installation, operation and maintenance.
To use the Host software with the harness USB RS-485 converter change
the required functions and for maintenance, refer to Manual TM 899050 for instructions
To use the I.S Handheld Unit to change the required functions and for
maintenance, refer to Manual TM 799060 for instructions.
To use the non I.S Mini laptop Unit, refer to Manual TM777070 for
instructions.
To use HART Protocol to change the required functions and for
maintenance refer to Manual TM 899030.
Product Overview 1
SafEyeTM Xenon 700S SIL Gas Detector
2 Product Overview

2 Technical Description

Features
page 3
Applications
page 4
Principle of Operation
page 4
Product Marking
page 8
Models and Types
page 9
Description
page 10
In this chapter…

2.1 Features

Long Range Gas Detection up to 459ft (140m) Simultaneous detection of C1-C8 flammable gases High Sensitivity and fast response to Hydrocarbon gases Heated optics to improve performance in icing, condensation and snow
conditions
Continuous operation in extreme and harsh environmental Conditions Solar blind and immune to industrial environments Withstands extreme vibration conditions Interfaces with most commonly used control panels Standard 0-20 mA and Dry Contact Relays Outputs HART Protocol: Communication Protocol (see HART Protocol, p. 6) RS-485 Output Modbus compatible for PC Communication Network for a
maximum of 247 systems
Simple one person installation, alignment, and calibration ATEX & IECEx approved per Ex II 2(1) G, Ex d e ia [ia Ga] IIC T5 Gb Ta
SIL-2 Approved by TUV per SIL-2 Requirement Programmable configuration via the handheld unit Fast connection to I.S. approved handheld diagnostic/calibration unit New operation mode (3mA) “maintenance call” Warranty 3 years for the complete SafEye system Warranty 10 years for Xenon Flash bulb
Features 3
SafEyeTM Xenon 700S SIL Gas Detector
Term
Description
LEL
Lower Explosive Limit: The minimum concentration of a substance (gas/vapor) in air mixture that can be ignited. This mixture is different for every gas/vapor, measured in % of LEL.
LEL.m
Integral of Concentration in LEL units (1 LEL = 100% LEL) and the operation distance in meters (m).

2.2 Applications

The SafEye Xenon system may be used to monitor flammable gas concentration in various applications, such as:
Petrochemical, pharmaceutical and other chemical storage and
production areas
Flammable and toxic chemical storage sites and hazardous waste
disposal areas
Refineries, oil platforms, pipelines, refueling stations and fuel storage
facilities
Hazardous loading docks, transportation depots and shipping
warehouses
Engine rooms Compressor and pumping stations Test cells LNG-LPG Systems Offshore Floating Production Storage and Shipping vessels (FPSO), and
fixed oil rigs

2.3 Principle of Operation

The SafEye system detects gases through a dual spectral range monitoring, analyzing the absorption of radiation caused by gases in the atmosphere and comparing it (ratio) to background atmospheric absorption.

2.3.1 Definitions of Terms

Table 1 defines gas concentrations measurement terms that are used in this manual.
Table 1: Gas Concentrations Measurement Terms
4 Principle of Operation
TM 799200 Rev (C), January 2014

2.3.2 Spectral Finger Print

Each hazardous material is detected at a specific wavelength selected according to its specific spectral absorption or "finger print". There are three IR sensors: two signals and one reference. The detection process involves two separate filters, one transmitting radiation that is absorbed by a particular gas and one that is not sensitive to it.

2.3.3 Optical Path

The presence of hazardous airborne vapors, gases, or aerosols in a monitored area is detected when the defined substance crosses/enters the optical path between the radiation source unit and the detector.
Hazardous gases /vapors present in the atmosphere cause absorption of the radiation pulse in the optical path between the radiating source and the detector unit at some specific wavelengths. This causes a change in the signal intensity received by the detector, which is translated into an output related to the detector’s measuring scale.
The system analyzes the defined open path at the spectral bands specific to the materials being monitored. The Automatic Gain Control (AGC) unit compensates for environmental disturbances such as fog, rain, etc., through a constant comparison with its dual spectral beam.

2.3.4 Microprocessor Based

The incoming signals are analyzed by the built in microprocessor. A sophisticated mathematical algorithm calculates between the various functions of the detected signal thresholds. Statistics, ratio algorithms, data communications, diagnostics and other functions are performed.

2.3.5 Gas Sensitivity

The SafEye IR Model uses wavelengths around 3.4µ spectral band to measure air flammability potential between the source and detector. At this wavelength, all hydrocarbon materials have a strong absorption peak. This peak enables the detector to achieve both regular sensitivity of 0-5 LEL.m or high sensitivity of 0-2 LEL.m according to functions set up. However, since the desired detection information is air flammability and the actual measurement is radiation absorption around the 3.4μ spectral band, the detector has a different sensitivity to different gases or gas combinations.
This difference in sensitivity is irrelevant if in the protected area the gas composition is known. However, if in a protected area the gas composition can vary substantially, then the difference in sensitivity should be considered in determining the Detector calibration.
Principle of Operation 5
SafEyeTM Xenon 700S SIL Gas Detector
Gas Type
Description
Gas 1
Pure methane and is for use in methane storage and piping applications
Gas 2
92% methane, 4% propane and 4% ethane (default) Universal oil and gas production mixture to be used in all cases where methane concentration in the mixture does not exceed 98%. It can also be used for pure Ethane applications.
Gas 3
LPG - 60% Propane and 40% Butane.
Gas 4
99% methane and 1% propane and is for use in detecting methane mixture with heavier gases where the methane component can vary between 100% to 95%. It is also good for protecting areas where a leak can be either pure methane or pure propane.

2.3.6 Gas and Mixture Selection and Setting

At the 3.4µ spectral band of the SafEye, the least sensitive gas is pure (100% vol) methane and the most sensitive gases are various mixtures of Methane with heavier alkanes where the Methane percentage is less than 90%. For pure Ethane the sensitivity is close to the high sensitivity gases and for pure Propane it lies somewhere between the two extremes. Every SafEye has 4 built-in gas calibration settings that can be changed by function set up The setting for each SafEye is detailed on the calibration datasheet provided with each unit.
The calibration settings are designed for use in various applications. Models 701, 702, 703 are calibrated to 4 different gases as follows:
Table 2: Gas Calibrations
Model 721, 722, 723 are calibrated to one gas only Gas 1 – Ethylene 100%. The four internal gas calibrations cover most of the flammable gas detection
applications. Actual selections should be made by the user in consultation with experts to provide for safety requirements. However, for special cases where none of the four calibrations are appropriate, Spectrex, through local agents can advise how to calibrate a SafEye Detector to any specific gas.

2.3.7 Flash Source

The Xenon Flash Source was originally introduced in the first SafEye development and was designed to overcome false alarms, which were experienced by early generations of the open path system. The new SafEye Xenon 700S employs the latest generation of flash bulbs to provide even more power and extended operation life (10 years).
6 Principle of Operation
TM 799200 Rev (C), January 2014

2.3.8 Heated Optics

SafEye Xenon includes heated optics for the Detector and source. The heater increases the temperature of the optical surface by 5-8°F (3-5°C) above the ambient temperature to improve performance in icing, condensation and snow conditions. The heated optics is configured to automatically operate when the change in temperature requires the heating (default).
However, the heated optics can be defined as one of the following modes:
1 Not operated (not an option on source unit) 2 On continuously 3 Automatic, per temperature change (default)
See System Set Up, on page 18. When operated ‘per temperature change’, the user can define the start
temperature below which the window will be heated (default 41°F (5°C)). This temperature can be defined between 32°F (0°C) to 122°F (50°C). The heating will stop when the temperature is 27°F (15°C) above the start temperature.

2.3.9 HART Protocol

The Xenon 700S uses the HART Protocol. HART Communication is a bi-directional industrial field communication
protocol used to communicate between intelligent field instruments and host systems. HART is the global standard for smart instrumentation and the majority of smart field devices installed in plants worldwide are HART­enabled.
HART technology is easy to use and very reliable. Through the HART connection, the SafEye is able to perform:
Detector set-up Detector troubleshooting Detector health and status
For more details, refer to the HART Manual TM 899030.
Principle of Operation 7
SafEyeTM Xenon 700S SIL Gas Detector

2.3.10 Handheld Unit

The new I.S.-approved handheld diagnostics unit (Part no. 799810) is available to make installation and maintenance easier. This is an all-on-one diagnostic / calibration / interrogation plug-in unit that allows for one­person installation and maintenance.
The handheld unit can be used:
For on-site function programming and set up changes to the Detector. During installation, the handheld unit will display all the detector’s
parameters and confirm that the installation has been completed successfully. It is also required to perform the necessary Zero calibration function.
For Maintenance and Trouble Shooting – The handheld unit will provide
recommendation of maintenance action to overcome and optimize the Detector’s performance.
For more details, refer to Manual TM 799060.

2.3.11 Modbus RS-485

For more advanced communications, the SafEye Xenon has a RS-485 Modbus-compatible output that provides data communication from a network (up to 247 Detectors) to a host computer or universal controller for central monitoring. This feature enables easy maintenance, local and remote diagnostic tools.

2.3.12 Tilt Mount

The new design stainless steel tilt mount provides a smaller installation footprint that can conform to limited space constraints, while the sturdy construction maintains alignment even in constant vibration. The improved ‘X’ and ‘Y’ axis worm-gear adjustments provide quick and easy alignment for installation and maintenance procedures.

2.4 Product Marking

The Xenon 700S open path gas Detector and source unit are certified to: ATEX Ex II 2(1) G per SIRA 03ATEX1406X
Ex d e ia [ia Ga] IIC T5 Gb Ta = -40°C to +55°C
IECEx per IECEx SIR 10.0003X
Ex d e ia [ia Ga] IIC T5 Gb Ta = -40°C to +55°C
This product is suitable to use in hazardous zones 1 and 2 with IIC gas group vapors present.
8 Product Marking
TM 799200 Rev (C), January 2014
Model
No.
Detector
Source
Min.
Installation
Distance
Max.
Installation
Distance
701S
XDSS-C-111XX
XSS-C-11X
13 ft. (4 m)
65 ft. (20 m)
702S
XDSS-C-112XX
XSS-C-11X
50 ft. (15 m)
230 ft. (70 m)
703S
XDSS-C-113XX
XSS-C-12X
165 ft. (50 m)
459 ft. (140m)
721S
XDSS-C-121XX
XSS-C-11X
13 ft. (4 m)
65 ft. (20 m)
722S
XDSS-C-122XX
XSS-C-11X
50 ft. (15 m)
230 ft. (70 m)
723S
XDSS-C-123XX
XSS-C-12X
165 ft. (50 m)
459 ft. (140m)

2.5 Models and Types

Series 700S offers 2 families of detected gases: Models 701S, 702S and 703S used for methane, LPG and gases mixture
for most offshore and onshore applications. Models 721S, 722S and 723S used for ethylene only. Table 3 describes the installation distances for each of the Series 700S
models.
Table 3: Model Numbers and Installation Distances
The 700S Series has 3 wiring options:
Option 1: Power, Alarm Relay, Fault Relays, 0-20mA Option 2: Power, Alarm Relays, Fault Relays, Accessory Relay Option 3: Power, Alarm Relay, RS-485, 0-20mA, RS-485 RTN (default)
Note: SIL2 covers only 0-20mA outputs. The wiring option must be set at the factory; cannot be changed at
customer facility.
Models and Types 9
SafEyeTM Xenon 700S SIL Gas Detector
Figure 1: Wiring Options

2.6 Description

The SafEye is comprised of two main units:
Flash Source Unit, page 11 Detector Unit, page 12
Xenon 700S detects gases over an open path transmitted from the Flash Source to the Detector.
10 Description
TM 799200 Rev (C), January 2014
1
Main Housing
6
Telescope Site
2
Junction Box
7
N/A
3
Back Cover
8
Earth Terminal
4
Cable Inlet
9
Front Window/Lens Section
5
Holding Plate

2.6.1 Flash Source Unit

The Flash Source unit emits IR radiation pulses at the rate of two pulses per second. The pulse width (5-10msec) is very powerful. The front of the source SafEye has a coated lens that collimates the IR beam for maximum intensity. The coated lens blocks all the UV and visible lights from passing through the lens and prevents the flash pulse being visible to the eye. The lens is heated to improve performance in icing, condensation and snow conditions.
There are two source types: Short and Medium Range: For use in models 701S, 702S, 721S, and
722S Long Range: For use in models 703S and 723S
Description 11
Figure 2: Flash Source Unit
SafEyeTM Xenon 700S SIL Gas Detector
Detector Type
Range
(in ft.)
Range
(in meters)
Short range
13 - 65 ft.
4 – 20 m
Mid-range
50 - 230 ft.
15 – 70 m
Long range
165 - 459 ft.
50 – 140 m

2.6.2 Detector Unit

The Detector receives the transmitted pulsed radiation signals from the Flash Source. The signals are then amplified and fed into an analog to digital signal converter to be processed by the internal microprocessor. When the signals drop below a prescribed level, the internal microprocessor will compensate for them. This will allow the signal to be maintained even in severe weather conditions. The data is sent to the output interface section. The signals can be sent to a standard control panel or to a central computer.
The front window of the Detector is heated to improve performance in icing, condensation and snow conditions.
Table 4 describes the detector types and their ranges.
Table 4: Detector Types
12 Description
TM 799200 Rev (C), January 2014
1
Main Housing
6
Telescope Site
2
Junction Box
7
Handheld Fast Connection
3
Back Cover
8
Earth Terminal
4
Cable Inlet
9
Front Window/Window Section
5
Holding Plate
Figure 3: Detector Unit
Description 13
SafEyeTM Xenon 700S SIL Gas Detector
14 Description

3 Operational Modes

Operation Modes
page 15
Output Signals
page 17
System Set Up
page 18
In this chapter…

3.1 Operation Modes

The SafEye 700S has four operational modes:
Normal Mode, page 15 Maintenance Call Mode (3mA Output), page 15 Fault Mode, page 16 Zero Calibration Mode (1mA Output), page 16

3.1.1 Normal Mode

This mode is used for gas detection. In normal mode, the following statuses are possible:
Normal: Signal received from gas detection is at safe levels Warning: Gases have been detected at warning levels Alarm: Gases have been detected at alarm levels
Note: For the standard 0-20mA output, the Warning and Alarm levels are not relevant. The user chooses these alarm levels at the controller. The output Detector will be 4mA at zero reading and 20mA for full-scale reading.
Warning and Alarm states are relevant where the stepped 0-20mA output is used and will energize the warning and alarm relays.
If the RS485 output is used, the Detector changes its status from ‘N’ to ‘W’ at Warning level and to ‘A’ at Alarm.

3.1.2 Maintenance Call Mode (3mA Output)

This mode indicates low signal or low signal ratio that may be caused by a dirty window; misalignment; poor source or that one of the Detectors parameters is at the ‘limit’ value.
The Detector continues to function and reads any gas present, but provides a (3mA) pre-warning signal that a maintenance procedure is required.
Operation Modes 15
SafEyeTM Xenon 700S SIL Gas Detector

3.1.3 Fault Mode

In the Fault status, there are three fault types: Fault 1 (2mA Output) – Non critical: If this occurs, it is due to
blockage, poor alignment, very low signal or in the case of partial
obscuration or full beam block and detection is no longer possible.
The Detector’s proper operation can be restored (Auto reset) during
operation if the condition causing the problem is removed or resolved.
This mode will occur after a delay of 60 sec. from the moment of the
fault. This delay is important to eliminate momentary obscuration due to
passing through the beam. Low Voltage Fault (0mA Output):In this case, detection is disabled
due to low voltage being supplied to the Detector. The Detector returns
to proper operation only when proper voltage level restored.
Fault 2 (0mA Output) – Critical: In this case, the detection is
disabled due to electrical / software operational failure or central device
(memory, processor) fault. Such fault causes the Detector to cease
operation.

3.1.4 Zero Calibration Mode (1mA Output)

This mode zeros the base level from which the Detector will read gas. It should only be performed when there are:
No combustible gases present Clear path between the Flash Source and Detector Clear weather conditions
Zero calibration must be done after installation, re-alignment and window cleaning, using the handheld unit (or Host software on PC)
16 Operation Modes
TM 799200 Rev (C), January 2014
Current Reading
Status and Description
0mA +0.3mA
Fault 2 or Low Voltage (Critical)
1mA ±0.3mA
Zero Calibration (In Progress)
2mA ±0.3mA
Fault 1 (non-Critical)
3mA ±0.3mA
“Maintenance Call”
4mA±0.5mA
No gas present
0-20mA
Continuous measuring of gas concentration at a range between 0 and full scale
21mA
Concentration is over the range limit (more than full-scale concentration).

3.2 Output Signals

The SafEye system provides the following outputs depending upon the wiring option selected (see Models and Types, page 9):
Standard 0-20mA Current Output, page 18 Dry Contact Relays, page 18 RS-485 Interface, page 18

3.2.1 Standard 0-20mA Current Output

The 0-20mA output can provide the Detector status measurement in one of the two following methods:
Default - It can be measured proportionally showing a continuous
reading of the exact gas concentration (see Table 1). Option - It can be a discrete indication according to the Detector mode
or the warning or alarm signal at a defined gas concentration (see Table
2). Setting of this method can be done by maintenance host or Handheld
unit. The 0-20mA functions as current sink but can be configured as Source – see
Appendix B. The maximum permitted load resistance for the 0-20mA output is 600Ω
Table 5: Standard (default) 0-20mA Current for the Gas Channel
Output Signals 17
SafEyeTM Xenon 700S SIL Gas Detector
Current Reading
Status and Description
LEL.m Setting
0+0.3mA
Fault 2 or Low Voltage
-
1±0.3mA
Zero Calibration
-
2±0.3mA
Fault 1 - 3±0.3mA
“Maintenance Call”
-
4±0.5mA
Standby
-
14±0.5mA
Warning
1
19±0.5mA
Alarm
3
21mA
Concentration is over the range limit (more than full-scale concentration).
Table 6: Optional - Discrete Reading of 0-20mA at Different Detector Modes

3.2.2 Dry Contact Relays

The detector may include up to three of the following relays depending on the wiring configuration:
Fault relay Alarm relay Accessory relay
Alarm and Accessory relays are normally non-energized open. When in the alarm or warning status the appropriate relay is closed.
Fault relay is normally energized closed, and when it is in the fault condition the relay is opened.

3.2.3 RS-485 Interface

The RS-485 input/output sends complete data information to a PC and receives data or control commands from the PC. The protocol is Modbus compatible. The communication with the PC that is operated through the interface is executed only when used with appropriate host software.

3.3 System Set Up

This section includes the following options:
Detection Function Programming, page 19 Detection Set-Up Function, page 19
18 System Set Up
TM 799200 Rev (C), January 2014
Sensitivity
Full Scale
Warning Level
Alarm Level
Normal
5 LEL.m
1 LEL.m
3 LEL.m
High
2 LEL.m
0.4 LEL.m
1 LEL.m

3.3.1 Detection Function Programming

The SafEye 700S series incorporates several functions that can be set by the customer using:
Handheld unit (P/N 799810). Refer to manual TM 799060 for
programming instructions. The connection of the handheld to the
detector is fast and intrinsically safe and allows function change with no
need to open the detector. To use the non I.S mini laptop Unit, refer to Manual TM 777070 for
instructions. Host software with harness USB RS-485 converter, refer to manual
TM 899050 for programming instructions.

3.3.2 Detection Set-Up Function

For the default settings see Default Detector Set up, page 21
3.3.2.1 Gas Calibration
Four gas types can be selected for maximum compatibility to the required measured gas(es).
Gas Types for Models 701S, 702S, 703S:
Methane Mixture 92% methane 4% ethane 4% propane (default) LPG – 60% Propane, 40% Butane Mixture 99% methane 1% propane
Gas Type for Models 721S, 722S, 723S: Ethylene 100% These four calibrations are standard calibrations. The Detector can be
calibrated up to eight (8) different types of gases or mixtures upon special request.
Each Detector will be supplied with a calibration sheet that will define the calibrated gas set up, full-scale, alarm and warning level.
3.3.2.2 Full Scale
Two full scales are available:
Table 7: Sensitivity Levels Options
When choosing a full scale, the warning and alarm level will change automatically per the table.
System Set Up 19
SafEyeTM Xenon 700S SIL Gas Detector
Function
Accessory Relay
Accessory Relay is activated at warning level
Accessory Relay is activated at alarm level
Alarm Latching
No latching function at alarm relay
Alarm relay is latched. Latching reset can be performed by momentary power disconnection or when Detectors are set to alignment mode
4-20 mA mode
Continue reading of the 0-20mA per the gas concentration level (see Table 1)
Discrete reading of the 0-20mA output according to the Detector status (see Table 2)
Beam Block During Alarm
Non latching of alarm indication during blocking mode
Alarm outputs are latched when the Detector turns to blocking mode from alarm position. Latching reset can be provided only if the Detector turn to normal mode
3.3.2.3 Zero Calibration
Zero Calibration can be enabled or disabled as follows:
Enable – Zero calibration is performed according to background Disable – The Detectors are not updated due to change of background
3.3.2.4 Other Functions
Table 8: Other Functions
3.3.2.5 Address Setup
The Detector provides up to 247 addresses that can be used in RS-485 communication link.
3.3.2.6 Heated Optic Operation
The heated optics for the Detector unit can be defined as one of the following modes:
OFF – Not operated ON – Operated continuously AUTO – On, per temperature change (default)
When operated per temperature change, the user can define the start temperature below which the window will be heated between 0 to 50°C. The Detector will stop heating the window when the temperature will be 15°C above the defined temperature.
This feature relates to the Detector only.
20 System Set Up
TM 799200 Rev (C), January 2014
Function
Setting
Gas Type
2
Full scale sensitivity
5 LEL.m
BG Zero Calibration
Enabled
Accessory Relay
Warning*
Alarm Latching
No
0-20mA
Continuous**
Beam block during alarm
No latch
Heat mode
Auto
Heat on
5
Function
Setting
Heat mode
Auto
Heat on
5
The Source heated optic must be defined with the order in two options:
Heated continuously Start heating below 41°F (5°C). (default)

3.3.3 Default Detector Set up

The Detector has 8 functions that can be programmed according to the customer requirement at factory or at customer facility using a software Host or a handheld unit. The standard set up is as follows:
Table 9: Detector Default Set Up
* Only refers to wiring option with Accessory relay output ** Only refer to the wiring options with 0-20mA
Table 10: Source Default Set-Up
The Source default can be changed only at the factory.
System Set Up 21
SafEyeTM Xenon 700S SIL Gas Detector
22 System Set Up

4 Technical Specifications

General Specifications
page 23
Electrical Specifications
page 24
Mechanical Specifications
page 26
Environmental Specifications
page 26
Detected Gases:
Simultaneous detection of C1-C8 flammable gases
Detection Distance range:
Table 11
Model
No.
Detector
Source
Min.
Installation
Distance
Max.
Installation
Distance
701S
XDSS-C-111XX
XSS-C-11X
13 ft. (4 m)
65 ft. (20 m)
702S
XDSS-C-112XX
XSS-C-11X
50 ft. (15 m)
230 ft. (70 m)
703S
XDSS-C-113XX
XSS-C-12X
165 ft. (50 m)
459 ft. (140m)
721S
XDSS-C-121XX
XSS-C-11X
13 ft. (4 m)
65 ft. (20 m)
722S
XDSS-C-122XX
XSS-C-11X
50 ft. (15 m)
230 ft. (70 m)
723S
XDSS-C-123XX
XSS-C-12X
165 ft. (50 m)
459 ft. (140m)
Response Time:
23 sec to T 90
Spectral Response:
3.0 - 4.0 micron
Sensitivity Range:
0-5 LEL.m
0-2 LEL.m
Field of View:
Line of Sight
Alignment Tolerance:
± 1°
Drift:
Long term ± 5% of full scale
Temperature Range:
-40°F (-40°C) to +131°F (+55°C)
Immunity to false alarm:
Does not produce false alarm and is not influenced by Solar Radiation, Hydrocarbon Flames and other external IR Radiation Sources.
In this chapter…

4.1 General Specifications

Table 11: Detection Distance Range
General Specifications 23
SafEyeTM Xenon 700S SIL Gas Detector
Operating Voltage:
18-32 VDC
Without Heated Optic (Max.)
With Heated Optic (Max.)
Detector
220mA
300mA
Source
220mA
300mA

4.2 Electrical Specifications

4.2.1 Power Consumption

Table 12: Detector and Source Maximum Power Consumption

4.2.2 Electrical Input Protection

The input circuit is protected against voltage-reversed polarity, voltage transients, surges and spikes according to MIL-STD-1275B.

4.2.3 Electrical Interface

There are three output-wiring options. These options must be defined at factory per the customer order. See Appendix B for wiring/terminal diagram for each option.
Note: Option 3 is the default unless otherwise specified at time of order.
Option 1:Power, Alarm Fault Relay, 0-20mA (Figure 9). Option 2:Power, Alarm, Fault & Accessory Relay(Figure 10) Option 3: Power, Alarm Relay, RS 485, 0-20mA, RS-485 RTN (Figure 11)
Note: SIL2 covers only 0-20mA outputs.

4.2.4 Electrical outputs

4.2.4.1 0-20mA Current Output
The 0-20mA is isolated sink option. This output can be configured as Source – see Appendix B.
The maximum permitted load resistance is 600 ohm. Two 0-20mA options:
Continuous reading – default (see Table 2) Discrete reading (see Table 3)
4.2.4.2 Communication Network
The Detector is equipped with an RS-485 communication link that can be used in installations with computerized controllers.
24 Electrical Specifications
Communication is compatible with the Modbus protocol: This protocol is a standard and is widely used.
TM 799200 Rev (C), January 2014
Relay
Type
Normal
Maximum Ratings
Alarm
SPST
NO
5A at 30VDC
Accessory
SPST
NO
5A at 30VDC
Fault
SPST
NO
5A at 30VDC
It enables continuous communication between a single standard Modbus
controller (Master device) and a serial Network of up to 247 Detectors. It enables connection between different types of Spectrex Detectors or
other Modbus devices to the same Network.
4.2.4.3 Relays Output
The Detector may include up to three of the following relays depending on the wiring configuration selected.
Table 13: Dry Contact Relays
Alarm and accessory relays are normally de-energized open. When in the alarm or warning status the appropriate relay is closed.
Fault Relay is normally energized closed, the contact will be closed, and when it is in Fault situation, the relay is opened.
4.2.4.4 HART Protocol
The HART Protocol is a digital communication signal at low level on top of the 0-20mA.
This is a bi-directional field communications protocol used to communicate between intelligent field instruments and the Host System.
Through the HART Protocol the detector can:
Display set-up Reconfigure set-up Display detector status and definition Perform detector diagnostic Troubleshooting
Electrical Specifications 25
SafEyeTM Xenon 700S SIL Gas Detector
Enclosure:
The Detector, Source and tilt mount are St. St. 316 Electro chemical and passivized coating.
Explosion Proof:
ATEX
Ex II 2(1) G, Ex d e ia [ia Ga] IIC T5 Gb Ta
Temp -40°F (-40°C) to 131°F (55°C)
IECEx
Ex II 2(1) G, Ex d e ia [ia Ga] IIC T5 Gb Ta
Temp -40°F (-40°C) to 131°F (55°C)
Water and Dust Tight:
IP 66 and IP 67
NEMA 250 type 6p
Electrical Modules:
Conformal coated
Electrical Connection:
(Two options - specified at time of order)
2 X M25 (ISO)
2 X 3/4” - 14NPT conduits
Dimensions:
Detector
8.2 x 5.7 x 6 in
(210 x 145 x 154mm)
Source
10 x 5.3 x 6.9 in
(255 x 135 x 175mm)
Tilt Mount
4.7 x 4.7 x 5.5 in
(120 x 120 x 140mm)
Weight:
Detector
9.2 lb.
(4.2 kg)
Source
10.1 lb.
(4.6 kg)
Tilt Mount
4.2 lb.
(1.9 kg)
Operating Temperature:
+131 ºF (+55 ºC)
Storage Temperature:
+149 ºF (+65 ºC)
Operating Temperature:
-4 ºF (-20 ºC)
Storage Temperature:
-40 ºF (-40 ºC)

4.3 Mechanical Specifications

4.4 Environmental Specifications

The SafEye system is designed to withstand harsh environmental conditions. The Source and Detector units compensate for adverse conditions while maintaining accuracy.

4.4.1 High Temperature

Designed to meet MIL-STD-810C, method 501.1 procedure II

4.4.2 Low Temperature

Designed to meet MIL-STD-810C, method 502.1, procedure I
26 Environmental Specifications
TM 799200 Rev (C), January 2014
Dust:
Totally protected against dust
Liquids:
Protected against immersion between 15 cm and 1m in depth. Protected against all water jets from all directions.
Vibration:
Designed to meet MIL-STD-810C, method 514.2, procedure VIII.
Mechanical Shock:
Designed to meet MIL-STD-810C, method 516.1, procedure I.
Radiated Emission:
EN61000-6-3
Conducted Emission:
EN61000-6-3
Radiated Immunity:
EN61000-4-3
Conducted Immunity:
EN61000-4-6
ESD:
EN61000-4-2
Burst:
EN61000-4-4
Surge:
EN61000-4-5

4.4.3 Humidity

Designed to meet MIL-STD-810C, method 507.1, procedure IV relative humidity of up to 95% for the operational temperature range.

4.4.4 Salt and Fog

Designed to meet MIL-STD-810C, method 509.1 procedure I. Exposure to a 5% salt solution for 48 hours.

4.4.5 Water and Dust

IP67 per EN60529 IP66 per EN60529

4.4.6 Shock and Vibration

4.4.7 Electromagnetic Compatibility (EMC):

This product is in conformance with EMC directive 89/336/EC.
Environmental Specifications 27
SafEyeTM Xenon 700S SIL Gas Detector
28 Environmental Specifications

5 Installation Instructions

Introduction
page 29
General Considerations
page 29
Preparations for Installation
page 30
Certification Instructions
page 32
Conduit/Cable Installation
page 33
Detector/Source Mounting
page 33
Detector Wiring
page 34
Detector Terminal Wiring
page 38
Flash Source Wiring
page 39
In this chapter…

5.1 Introduction

The Detector and Flash Source units can be installed and maintained with the use of general-purpose common tools and equipment. The installation procedure has to be performed by suitably qualified personnel.
This section does not attempt to cover all of the standard practices and codes of installation. Rather, it emphasizes specific points of consideration and provides some general rules for suitably qualified personnel. Special safety precautions are stressed wherever applicable.

5.2 General Considerations

5.2.1 Personnel

Only suitably qualified personnel, familiar with the local codes and practices, trained for gas detection maintenance should be employed. Wiring should only be performed or supervised by someone with knowledge of electronics and in particular wiring installation.

5.2.2 Tools Required

The SafEye system requires the following tools:
Set of screwdrivers Set of hex keys/Allan wrenches (supply with commissioning kit) Voltage Multimeter
General Considerations 29
SafEyeTM Xenon 700S SIL Gas Detector

5.2.3 Site Requirements

The installation position of the SafEye system must take into account if the gas being monitored is heavier or lighter than air, and the individual site requirements. The site selected must give the Detector a direct view to the Source. The mounting point for each item should be secure and stable with minimal vibrations. Equipment should be either mounted in a position where it cannot be knocked out of alignment, or guarded from physical impact.

5.2.4 The Source and Detector

The appropriate Detector should be selected for the length of open path to be monitored. To allow for ageing of the Source and a reduction of the IR signal due to adverse weather we recommend using a Detector that is not at the limit of its operating range. As a guide, select a Detector that is installed at a distance from the Source of not more than 75% of the specified operating distance. In severe weather conditions such as offshore oil production and exploration, this should be reduced to 50%.
The open path between the Source and Detector and the immediate surroundings should be kept clear of obscuration that might hinder the free movement of air in the protected area or block the infrared beam.

5.2.5 Guidance Tips for Gas Detector Locations

Guidance Tips for Gas Detector locations in order to provide the best detection coverage:
Below potential leak source for heavier than air gases Above potential leak sources for lighter than air gases Near to leak sources along the expected leak trajectory, taking into
account prevailing wind directions
Between leak source and potential ignition sources In area with expected heavy fog, rain or snow, consider the effect of
long range installation and install the Detector at shorter range with the
maximum intensity model available.

5.3 Preparations for Installation

Installation should comply with local, national and international regulations and norms as applicable to gas Detectors and approved electrical devices installed in hazardous areas. The Detectors can be installed with general­purpose common tools and equipment.
The System should include the following (in addition to this manual):
Detector unit: XDSS-C-11XXX (See Models and Types on page 9) Source unit: XSS-C-XX (See Models and Types on page 9)
30 Preparations for Installation
TM 799200 Rev (C), January 2014
Two Tilt Mount Bases P/N 799640
1 base is used for the Detector
1 base is used for the Flash Source. The Commissioning kit includes Function Check Filter and the Telescope
Kit, which is used during each SafEye installation and then removed.
They can be reused for all other SafEye installations on the site.
Therefore, only one set is provided for several detectors.
Telescope Kit P/N 799210 Function Check Filter P/N 792260 (1-5)* Handheld Unit P/N 799810
* There are 5 options depending on full scale and type of calibrated gas.

5.3.1 Preparing for the Installation

The following procedure is used for preparing the installation of the 700S Gas Detector:
1 Verify the appropriate Purchase Order. Record the Part No. and the
Serial No. of the detectors and source units and the installation date in
the appropriate log book.
2 Open the container package immediately prior to detector installation
and visually inspect the detectors; sources and accessories.
3 Verify that all components required for the detector installation are
readily available before commencing the installation. In the event that
the installation is not completed in a single session, secure and seal
detectors and conduits.
4 For wiring, use color-coded conductors or suitable wire markings or
labels. 12 to 20 AWG (0.5 mm² to 3.5 mm²) wires may be used for site
wiring. The selection of wire gauge should be based on the number of
detectors used on the same loop and the distance from the control unit,
in compliance with specifications (See Appendix A, page 53)
Preparations for Installation 31
SafEyeTM Xenon 700S SIL Gas Detector
Warning: Do not open the detector, even when isolated, when flammable atmosphere is present.

5.4 Certification Instructions

Use the following certification instructions: The cable entry point may exceed 158°F (70°C) suitable precautions
should be taken when selecting the cable. The equipment may be used with flammable gases and vapors with
apparatus groups IIA, IIB and IIC T5 in the ambient temperature
range -40°F (-40°C) to +131°F (+55°C). Installation shall be carried out by suitably trained personnel in
accordance with the applicable code of practice e.g. EN 60079-14:1997. Inspection and maintenance of this equipment shall be carried out by
suitably trained personnel in accordance with the applicable code of
practice e.g. EN 60079-17. Repair of this equipment shall be carried out by suitably trained
personnel in accordance with the applicable code of practice
e.g. EN 60079-19. The certification of this equipment relies upon the following materials
used in its construction:
Enclosure: 316L Stainless Steel
Window: Sapphire Glass
Window Cement: Ciba Geigy epoxy putty AV 138M & HV 998
Seals: Buna-n If the equipment is likely to come into contact with aggressive
substances, then it is the responsibility of the user to take suitable
precautions that prevent it from being adversely affected, thus ensuring
that the type of protection provided by the equipment is not
compromised.
Aggressive substances: For example. acidic liquids or gases that may
attack metals, or solvents that may affect polymeric materials.
Suitable precautions: For example, regular checks as part of routine
inspections or establishing from the material’s data sheets that it is resistant to specific chemicals.
Special conditions for safe use: The series 700S Open Path Gas
Detectors and Flash Source units shall not be used as safety related
devices in accordance with directive 94/9/EC.
32 Certification Instructions
TM 799200 Rev (C), January 2014
Item
Qty
Type/Model
Tilt Mount
1
799220
Screw
1
5/16” – 18 UNC x 3/4”
Spring Washer
1
No. 5/16”

5.5 Conduit/Cable Installation

The conduit and cable installation must comply with the following guidelines:
1 To avoid water condensation in the detector, install the Detector with the
conduits/cable entries facing downwards
2 Use flexible conduits/cables for the last portion connecting to the
detector
3 When pulling the cables through the conduits, ensure that they are not
tangled or stressed. Extend the cables about 12 inches (30cm) beyond
the Detector location to accommodate wiring after installation.
4 After the conductor cables have been pulled through the conduits,
perform a continuity test

5.6 Detector/Source Mounting

The Detector Source should be mounted with the Tilt Mount Kit, Model
799640. The Tilt enables the Detector to be rotated up to 45º in all directions and fine alignment of up to 3º.

5.6.1 Tilt Kit P/N 799640

The following contents are included with the Tilt Mount Kit:
Table 14: Tilt Mount Kit

5.6.2 Detector/Source Installation

(Figure 4 and Figure 5)
To install the Detector/Source:
1 Place the Tilt Mount holding plate (item 1) in its designated location and
secure it with (4) fasteners through four (4) holes dia. 8.5mm.
Notes:
1. Skip this step if the Tilt Mount is already installed.
2. Detector removal for maintenance purpose does not require Tilt Mount
removal.
2 Place the Detector, with its conduit/cable inlets pointing downwards on
the Detector holding plate of the Tilt Mount (item 2). Secure the
Detector with 5/16”-18UNC x 3/4" screw with No. 5/16” spring washers
(9,10) to the Tilt Mount. Use ¼” Hex Key for 5/16” screw (item 9).
Conduit/Cable Installation 33
3 Repeat 1 – 2 to install the Source.
SafEyeTM Xenon 700S SIL Gas Detector

5.7 Detector Wiring

(Figure 5 and Figure 6)
To install the Detector wiring:
1 Remove the four (4) socket-head-screws (item 14, Figure 5) that secure
the Detector back cover (item 15, fig. 4). The chamber is now revealed.
2 Remove the protective plug mounted on the Detector conduit/cable
entry inlet; pull the wires through the Detector inlet (item 4, Figure 6).
Use a 3/4” – 14NPT or M25x1.5 explosion-proof conduit connection /
cable gland to assemble the cable / explosion-proof conduit to the
Detector.
3 Connect the wires to the required terminals (item 2, Figure 6) according
to the wiring diagram. See Detector Terminal Wiring, page 38 and
Figure 9, Figure 10,Figure 11 Appendix B.
4 Connect the grounding wire to the ground screw outside Detector
(item 3. Figure 6). The Detector must be well grounded to Earth Ground.
5 Check the wires for secure mechanical connection and press them neatly
against the terminal to prevent them from interfering while closing the
cover (item 15, Figure 5)
6 Place and secure the Detector back cover using four (4) socket-head-
screws (item 14, Figure 5)
34 Detector Wiring
TM 799200 Rev (C), January 2014
1
Tilt Mount Holding Plate
5
Vertical Fine Alignment Tightening Screw
2
Detector/Source Holding Plate
6
Vertical Crude Alignment Tightening Screw
3
Horizontal Crude Alignment Tightening Screw
7
Vertical Fine Alignment Screw
4
Horizontal Fine Alignment Tightening Screw
8
Horizontal Fine Alignment Screw
Figure 4: Tilt Mount
Table 15: Tilt Mount Description
Detector Wiring 35
SafEyeTM Xenon 700S SIL Gas Detector
1
Tilt Mount Holding Plate
9
Detector Tightening Screw
2
Detector/Source Holding Plate
10
Detector Tightening Washer
3
Horizontal Crude Alignment Tightening Screw
11
Detector
4
Horizontal Fine Alignment Tightening Screw
12
Telescope
5
Vertical Fine Alignment Tightening Screw
13
Telescope Tightening Bolt
6
Vertical Crude Alignment Tightening Screw
14
Back Cover Tightening Bolt
7
Vertical Fine Alignment Screw
15
Detector Back Cover
8
Horizontal Fine Alignment Screw
Figure 5: Detector and Tilt Mount Assembly
Table 16: Detector and Tilt Mount Assembly Description
36 Detector Wiring
TM 799200 Rev (C), January 2014
1
Housing
5
Internal Earth Connection
2
Terminal Board
6
Connection to Handheld Unit
3
Earth Terminal
7
Detector Holding Plate
4
Inlet Conduit
8
Detector Telescope Site
Detector Wiring 37
Figure 6: Detector with Cover Removed
Table 17: Detector with Cover Removed Description
SafEyeTM Xenon 700S SIL Gas Detector
Wiring
Option
Terminal Number
1 2 3 4 5 6 7 8 9
3
Figure
11
(Default)
24Vdc
0Vdc
Alarm Relay
(1)
RS-485+
(3)
RS-485-
(3)
0-20mA +
(4)
0-20mA -
(4)
RS-485
RTN
(6)
2
Figure
10
24Vdc
0Vdc
Alarm Relay
(1)
Fault Relay
(2)
Accessory Relay
(5)
not used
1
Figure 9
24Vdc
0Vdc
Alarm Relay
(1)
Fault Relay
(2)
0-20mA +
(4)
0-20mA -
(4)
not used

5.8 Detector Terminal Wiring

The Detector has three output wiring options in the rear, segregated, “Exde“
terminal section, with 9 Terminals, labeled 1-9.
Option 3 is the default unless otherwise specified at time of order
The following describes the function of each electrical terminal of the Detectors for all wiring options:
Table 18: Wiring Options
Notes:
1 The Alarm output is a N.O. contact relay (SPST).
The contacts are closed in Gas Alarm state.
2 The Fault output is N.C. SPST relay. The contacts are closed when the
Detector is in its normal operational condition.
3 Used for communication network as specified in Appendix B. 4 See Appendix B for more details 5 The Accessory output is N.O. (SPST) relay. The Accessory Relay may act
in parallel with the ALARM relay to activate another external device or it
may provide a warning signal, depending on Function Set Up.
6 Used as RTN to the RS-485 communication port.
38 Detector Terminal Wiring
TM 799200 Rev (C), January 2014

5.9 Flash Source Wiring

5.9.1 Wiring

To install the wiring:
1 Remove the four (4) socket-head-screws (item 14, Figure 5) that secure
the Source back cover (item 15, Figure 5). The chamber is now open.
2 Remove the protective plug mounted on the source conduit/cable entry
inlet; pull the wires through the source inlet (item 4, Figure 7). Use a
3/4” – 14NPT or M25x1.5 explosion-proof conduit connection / cable
gland to assemble the cable / explosion-proof conduit to the source.
3 Connect the wires to the required terminals (item 2, Figure 7) (See
Terminal Wiring, page 39)
4 Connect the ground wire to the ground screw outside the source (item 3,
Figure 7). The source must be well grounded to earth ground.
5 Check wires for secure mechanical connection and press them neatly
against the terminal to prevent them from interfering while closing the
cover (item 15, Figure 5)
6 Place and secure the source back cover using four (4) socket-head-
screws (item 14, Figure 5)

5.9.2 Terminal Wiring

The Flash Source contains a three-wire strip terminal:
Terminal 1 – positive (+) power supply Terminal 2 – common return Terminal 3 – GND (ground)
Flash Source Wiring 39
SafEyeTM Xenon 700S SIL Gas Detector
1
Housing
5
Internal Earth Connection
2
Terminal Board
6
N/A
3
Earth Terminal
7
Detector Holding Plate
4
Inlet Conduit
8
Detector Telescope Site
40 Flash Source Wiring
Figure 7: Source with Cover Removed
Table 19: Source with Cover Removed Description

6 Operating Instructions

SafEye Operation
page 41
Alignment of Unit
page 41
Powering up the System
page 42
Safety Precautions
page 43
Signal Verification
page 43
Zero Calibration
page 44
Functional Check of Unit
page 45
In this chapter…

6.1 SafEye Operation

After the system is positioned, it will monitor for possible specified gases automatically sending signals to a standard control panel or to a PC. This section describes the alignment, calibration and operation.
Important! Accurate Alignment is essential for proper operation of the SafEye system.

6.2 Alignment of Unit

Using the Telescope performs full alignment. Alignment procedure should be performed in two stages: Crude alignment
and fine adjustment. The Telescope includes a Periscope that consists of a prism and an ocular
located vertical to the Telescope assembly. This allows the user to observe into the opposite unit perpendicularly to the alignment when access from the rear of the unit is impossible. For installations where rear access is possible the Periscope is not necessary. In this case it can be removed by releasing the Periscope Fastening Screw.
Important Notes
1 Prior to Telescope Installation, verify that the Telescope and Its Sight
Mounting are free from any dirt to ensure proper alignment according to
factory calibration.
2 Do not attempt to change any factory calibration at the Telescope or its
mounting. This may prevent optimal alignment.
To align the unit (see Figure 10):
1 Make sure that the Detector and the Flash Source are installed properly.
Installation instructions are described in section 5 of this document.
Alignment of Unit 41
SafEyeTM Xenon 700S SIL Gas Detector
2 Install the Telescope Assembly (12) on the Telescope Site Mounting of
the Source according to the drawing. Fasten the Telescope with
Fastening Screw (13).
3 Crude Alignment:
a. Use ¼” Allen screw driver for all alignment screws.
b. Loosen screws 5, 6.
c. Approximately aim the source horizontally toward the detector.
d. Tighten screw 6.
e. Loosen screws 3 and 4.
f. Approximately aim the source vertically toward the detector.
g. Tighten screw 3.
4 Repeat step 3 for the Detector. 5 Fine Alignment:
a. Aim the Source to the Detector within horizontal axis using screws 7.
Aim the cross to the upper level of the telescope site or the Detector.
b. Tighten screw 5.
c. Aim within the vertical axis using screw 8.
d. Tighten screw 4.
e. Make sure the telescope cross is pointing at the top of the telescope
site of the Detector.
6 Repeat step 5 for the Detector alignment.

6.3 Powering up the System

Important! Prior to any operation or maintenance, follow the Safety Precautions in Safety Precautions, page 43 .
To power up the system:
1 Make sure that the Source and Detector are connected to power. 2 Make sure that the 0-20mA meter is connected to the Detector. 3 Power up the system 18-32VDC. 4 After 60 seconds, the current meter will indicate 4mA.
Note: Zero calibration should be performed after powering up the system
(see Zero Calibration, page 44).
42 Powering up the System
TM 799200 Rev (C), January 2014
Channel
Installation Distance
Maintenance
Min
Med
Max
Reference
1V Gain1
1V Gain2
1V Gain3
The minimum signal allowed is 2.5V at Gain4
Signal 1
1V Gain1
1V Gain2
1V Gain3
The minimum signal allowed is 2.5V at Gain4
Signal 2
1V Gain1
1V Gain2
1V Gain3
The minimum signal allowed is 2.5V at Gain4
Ratio 1
0.6-2
0.6-2
0.6-2
0.5-3
Ratio 2
0.6-2
0.6-2
0.6-2
0.5-3
NQRat 1
0.95 – 1.05
Must be 0.95 – 1.05
NQRat 2
0.95 – 1.05
Must be 0.95 – 1.05
LEL
0 LEL x m
0 LEL x m
Temp.
Up to 25° C beyond room temp.
Up to 25° C beyond room temp.
Voltage
32VDC > V > 17VDC
32VDC > V > 17VDC
Ref/Noise
More than 80
More than 30
Sig 1/Noise
More than 80
More than 30
Sig 2/Noise
More than 80
More than 30

6.4 Safety Precautions

After powering up the system, the Detector requires minimal attention in order to function properly, but the following should be noted:
1 Follow the instructions in the Manual; refer to the drawings and
specifications issued by the manufacturer.
2 Do not open the detector/source housing while power is supplied. 3 Disconnect external devices, such as automatic extinguishing systems
before carrying out any maintenance task warranty.

6.5 Signal Verification

The signal verification can be done through Host software supplied by Spectrex (see Detection Function Programming, page 19).

6.5.1 Signal Values Limitation

Table 20 describes the maintenance data channels limitation values.
Table 20: Maintenance Channels Limited Values
Note: The installation information refers to the installation distance.
Safety Precautions 43
Min: The minimum distance as defined on the model no. Med: Half of the maximum distance as defined on the model no. Max: The maximum distance as defined on the model no.
SafEyeTM Xenon 700S SIL Gas Detector

6.6 Zero Calibration

Zero calibration must be done after:
Installation Realignment Window Cleaning Any change in Detector or Source position
Precise alignment must be performed prior to the zero calibration procedure. Zero calibration should be performed in good weather conditions with insignificant gas concentrations in the surrounding environment or indoors.
To perform the zero calibration procedure:
1 Switch from Normal to Alignment mode indication. 2 Switch from Alignment to Standby mode. 3 Switch from Standby to Zero Calibration mode.
The 0-20mA output should now be at 1mA.
4 Wait up to 60 seconds until it switches to Normal mode. The Detector
reading is now set to normal. The 0-20mA output should now indicate
4mA.
Switching from each position (above first three instructions) can be done
by HOST HART (refer to TM888030) or RS485 (refer to TM888050 or
TM799060), or by moving the magnetic mode selector above the
magnetic switch (see Figure 8).
Figure 8: Magnetic Mode Selector
44 Zero Calibration
TM 799200 Rev (C), January 2014

6.7 Functional Check of Unit

The SafEye system has been calibrated at the factory for the User's specific gas or vapor detection requirements. The following procedure validates the functional operation of the system. The Functional Check Filter is a convenient operational check used to confirm that response has not changed from previous readings. The filter is not used for calibration, which is unnecessary, nor does it equate to a particular quantity of gas.
Caution: Automatic activation or any external device that should not be activated during the calibration check should be disconnected.
Notes:
1 This functional verification procedure is for a standard 0-20mA output. 2 Prior to starting the functional check, verify that the power to the units is
on and that the current of the 0-20mA channel is stable. Record the
reading.
To perform the functional check:
1 Position the Functional Check filter in front of the SafEye Detector. 2 The functional check filter's window must be centered over the viewing
window of the Detector.
3 Wait 20 seconds. 4 Read the 0-20mA current. Determine the difference between the reading
taken with and without the Functional Check Filter. This difference is the
0-20mA current variance.
5 Record the 0-20mA current variance in the Maintenance Logbook. If the
variance is more than a 30% change when compared to the previous
check (see delivery form), repeat the alignment.
Functional Check of Unit 45
SafEyeTM Xenon 700S SIL Gas Detector
46 Functional Check of Unit

7 Maintenance Instructions

General Maintenance
page 47
Periodic Maintenance
page 47
In this chapter…

7.1 General Maintenance

The SafEye Xenon system requires only simple periodic maintenance to provide satisfactory service and achieve maximum performance. The Detector and Source units can be maintained with the use of common tools and equipment. Record the periodic test results in Maintenance Logbook with a copy of the delivery form inside.

7.2 Periodic Maintenance

Recommended cleaning of optical surfaces to be performed periodically. Note: The frequency of cleaning operations is ultimately dependent upon
the existing environmental conditions and the applications used.
To perform periodic maintenance:
1 Proper maintenance will allow the SafEye system to retain maximum
performance and reliability.
2 The optical surfaces of the Source and Detector viewing windows should
be kept as clean as possible as it is an active device.
3 Alignment procedures must be performed each time that the Source or
the Detector unit has been opened or moved for any reason.
4 The Signal Verification Check corroborates the current signals from the
Flash Source compared to that of previous alignments. It is
recommended to perform this check every 6-12 months. The signal
should be checked according to threshold levels (see. Signal Verification,
page 43).
5 The functional check should be performed every 6 months (see
Functional Check of Unit, page 45).
6 Alignment procedure should be performed only if the signals are below
Threshold value (see Signal Verification, page 43).
7 Zero calibration (see Zero Calibration, page 44) must be done every
time the Detector or Source is realigned or windows cleaned.
Periodic Maintenance 47
SafEyeTM Xenon 700S SIL Gas Detector

7.2.1 Routine Optical Surface Cleaning

The SafEye system, being an optical device, must be kept as clean as possible. The optical surfaces concerned are the Source and Detector viewing windows.
To clean the optical window, proceed as follows:
1 Turn off the power to the SafEye detector and source. 2 In places where dust or dirt has accumulated on the optical surface,
clean the surface with a small, soft-bristle brush.
3 The surfaces must then be washed thoroughly with water and a mild
non-abrasive detergent.
4 Thoroughly rinse the glass surface with clean water; ensuring no residue
is left behind.
5 Dry the glass with a clean dry soft cloth. 6 Enter in the Maintenance Logbook: Date, name of person and company
who performed the maintenance service.
7 Turn on power to the SafEye detector and source. 8 Perform signal verification (see Signal Verification, page 43). 9 Perform zero calibration (see Zero Calibration, page 44). 10 Perform functional check (see Functional Check of Unit, page 45)

7.2.2 Signal Verification

The signal verification check determines the proper operation of the open path. It checks the alignment and cleanliness of the window or any problem in Source or Detector. Use the PC Host software to measure the signal verification.
Refer to Manual TM 799060 or use I.S handheld unit. Refer to Manual TM 777070 for non I.S Mini laptop Unit.

7.2.3 Function Check of Unit

The SafEye Xenon system has been calibrated at the factory per the user’s
specific gas or vapor detection requirements. This procedure validates the functional operation. The functional check must be done periodically. Refer to Functional Check of Unit, page 45 for instructions.
Caution: Automatic activation or any external device that should not be activated during the calibration check should be disconnected.
48 Periodic Maintenance

8 Troubleshooting

Troubleshooting Problems
page 49
Problem
Cause
Solution
“Maintenance call”
status or R, S1 and S2 are below 2.5V DC at Gain 4
Poor alignment
Perform alignment
Dirt on the window
Clean the window Poor light source
Replace the light source
Detector fault
Replace/repair Detector
NQRat1 and NQRat2 below the permitted limit
Gas in the path
Make sure that the path is clean and the weather conditions are good
NQRat1 and NQRat2 above the permitted limit
Poor alignment
Perform alignment
R/N, S1/N, S2/N below 50
Poor alignment
Perform alignment
Dirt on the window
Clean the window Poor light source
Replace the light source
Detector fault
Replace/repair Detector
Temperature higher than 25° C beyond the room temperature
Electronic problem
Replace/repair Detector
Ratio1 and Ratio2 out of the limit
Poor alignment
Perform alignment
Dirt on the window
Clean the window Detector fault
Replace/repair Detector
Voltage less than 16V DC. The Detector at “V” fault
Low input voltage
Check the power supply and installation
In this chapter…

8.1 Troubleshooting Problems

Table 21: Troubleshooting Problems
Troubleshooting Problems 49
SafEyeTM Xenon 700S SIL Gas Detector
50 Troubleshooting Problems

Appendices

Appendices 51
SafEyeTM Xenon 700S SIL Gas Detector
52 Appendices

A Wire Selection Tables

General Instructions For Electrical Wiring
page 53
AWG #
mm2
Ohm per 100 ft.
Ohm/100 meter
26
0.12 - 0.15
4.32
14.15
24
0.16 - 0.24
3.42
11.22
22
0.30 - 0.38
1.71
5.60
20
0.51 - 0.61
1.07
3.50
18
0.81 - 0.96
0.67
2.20
16
1.22 - 1.43
0.43
1.40
14
1.94 - 2.28
0.27
0.88
12
3.09 - 3.40
0.17
0.55
10
4.56 - 6.64
0.11
0.35
In this appendix
A.1 General Instructions For Electrical Wiring
Refer to Table 22 to determine the required wire gauge for general wiring, such as relay wiring. Calculate the permitted voltage fall with respect to loads current, wire gauge and length of wires.
Refer to Table 23 to select wire gauge for power supply wires. DO NOT connect any circuit or load to Detectors’ supply inputs.
Table 22: Maximum DC resistance at 68˚F for copper wire
1 Select "Number of Detectors" connected in one circuit. 2 Select "wiring length" per your installation requirements. 3 Refer to "power supply range" for voltage extreme applied.
General Instructions For Electrical Wiring 53
SafEyeTM Xenon 700S SIL Gas Detector
No. of Detectors
Recommended Wire Diameter
Power Supply Range (VDC)
24
18
16
14 - -
22-32
20
18
16
14 - -
22-32
16
20
18
16
14 - 22-32
12
20
18
16
14 - 20-32
8
20
18
16
14 - 20-32
4 and less
20
18
16
16
14
20-32
Feet
164
328
492
656
820
(Meters)
(50)
(100)
(150)
(200)
(250)
Max. Length from Power Supply to Last Detector
Table 23: Wiring length in feet (meter)
54 General Instructions For Electrical Wiring

B Wiring Option Configurations

RS-485 Communication Network
page 60
In this appendix
Figure 9: Wiring Option 1
Note: Only 0-20mA is SIL2 approved.
Wiring Option Configurations 55
SafEyeTM Xenon 700S SIL Gas Detector
Figure 10: Wiring Option 2
Note: This option does not comply with SIL2 requirements.
56 General Instructions For Electrical Wiring
TM 799200 Rev (C), January 2014
Figure 11: Wiring Option 3 (Default)
Note: Only 0-20mA is SIL2 approved.
Wiring Option Configurations 57
SafEyeTM Xenon 700S SIL Gas Detector
Figure 12: 0-20mA Wiring (Sink) For Wiring Option 1 and 3
Figure 13: 0-20mA Wiring (Source) For Wiring Option 1 and 3 Notes:
The Detectors are factory set to isolated 0-20mA-sink version. For non-isolated 0-20mA version (Source), connect Terminal 8 to
Terminal 1. The 0-20mA meter is connected between Terminal 7 and
Terminal 2.
58 General Instructions For Electrical Wiring
TM 799200 Rev (C), January 2014
Figure 14: 0-20mA 3 Wire Connection for Option 1 and 3
Wiring Option Configurations 59
SafEyeTM Xenon 700S SIL Gas Detector
B.1 RS-485 Communication Network
By using the RS-485 network capability of the SafEye Xenon 700S Detector and additional software, it is possible to connect up to 32 Detectors in an addressable system with four (4) wires only (2 for power and 2 for communication). Using repeaters, the number of Detectors can be much larger (32 Detectors for each repeater) up to 247 on the same four (4) wires. When using the RS-485 network, it is possible to read each Detector status (FAULT, WARNING, and ALARM).
For more details, consult the factory.
Figure 15: RS-485 Networking for Wiring Option 3
60 RS-485 Communication Network
C Special Conditions to Comply to SIL-2
Safety relevant parameters (700S)
page 61
Guidelines for Configuration, Installation, Operation and Maintenance
page 61 Miscellaneous
page 62
Requirements
In this appendix
Special conditions to comply with the requirements of EN 61508 for SIL-2 This appendix details the special conditions to comply with the requirements
of EN 61508 for SIL-2. The SafEye 700S can be used in low or high demand mode applications, see IEC 61508.4, Chapter 3.5.12.
C.1 Safety relevant parameters (700S)
Perform the following functional check of the detector half yearly: (see chapters Periodic Maintenance and Function Check of Unit)
HFT: 0 PFD: 3.3 x 10 -4 (3.3% of SIL-2)  PFH: 1.5 x 10 SFF: Fulfills the conditions of EN 61508 for SIL-2.
-4
(15% of SIL-2) - For 0-20mA signal current application.
C.2 Guidelines for Configuration, Installation,
Operation and Maintenance
Alarm conditions complying with SIL-2 can be implemented only with 0-20mA outputs.
C.2.1 Conditions for Safe Operation
The SafEye 700S shall consist only of the approved hardware and
software modules
The 24C power supply must comply with the conditions for “safe low
voltage” according to EN 60950 (PELV / SELV).
After installation and configuration the set-up parameters must be
verified and the function of the SafEye 700S (gas detection, function of
the 0-20mA interface, relay functions) must be checked completely.
Safety relevant parameters (700S) 61
SafEyeTM Xenon 700S SIL Gas Detector
Mode
Normal
Warning
Alarm
Continuous current with low sensitivity (Full Scale: 5 LEL.m)
4 mA (0 LEL.m)
7.2 mA (1 LEL.m)
13.6 mA (3 LEL.m)
Continuous current with high sensitivity (Full Scale: 2 LEL.m)
4 mA (0 LEL.m)
7.2 mA (0.4 LEL.m)
12 mA (1 LEL.m) Discrete current
4 mA
14 mA
19 mA
C.2.2 Alarm Operation using the 0-20mA Signal Current
The connected controller has to monitor the 0-20mA signal current for valid values: (see cap. Output Signals)
The connected controller has to monitor the 0-20mA signal current for values below 4mA and above 20mA.
C.2.3 Relay Outputs
Relay outputs don’t comply with SIL2 requirements.
C.3 Miscellaneous
The complete functional performance of the gas detector has to be verified half-yearly (gas detection, fault detection, performance of the 0-20mA signal current and the relays), gas alarms ON and OFF.
The HART® and RS 485 interface are not allowed to be used for the transmission of safety related data.
62 Miscellaneous

Technical Support

For all technical assistance or support, contact:
218 Little Falls Road Cedar Grove, NJ 07009, USA Tel: +1 (973) 239 8398 Fax: +1 (973) 239 7614
Email: spectrex@spectrex.net Web-site: www.spectrex.net
Your Local Office:
SPECTREX INC.
Texas (USA)
Mr. Jay Cooley, Regional Sales Manager: 16203 Park Row, Suite 150 Houston, Texas 77084 USA
Phone: +1 (832) 321 5229 Email: jay@spectrex.net
Europe
Mr. Ian Buchanan, Regional Manager: 6 Applecross Road Glasgow G66 3TJ United Kingdom
Phone: +44 (0) 141 578 0693 Email: ian@spectrex.net
Far East
Mr. Deryk Walker, Regional Sales Manager 59 Fen Ji Hu, Danshui Taipei County 25163 Taiwan (ROC)
Phone: +886 2 8626 2893 Mobile: +886 926 664 232 Email: deryk@spectrex.net
Technical Support 63
SafEyeTM Xenon 700S SIL Gas Detector
64 Technical Support
Loading...