SOR RF Continuous Level Transmitter User Manual

SORtrax Level Transmitter
General Instructions
SORtrax is a 4-20 mA continuous level transmitter. It produces a 4-20mA current superimposed on the 12-55 VDC loop supply lines. The 4-20mA current is proportional to the level sensed by the instrument.
SORtrax detects level based on process admittance. Bench or field calibration is easily accomplished. Coarse and fine adjustments are provided to allow the user full control over the zero and span settings, and to insure the most accurate operation possible. Recalibration is needed each time the content of the process changes, due to the different dielectric constant of each material.
NOTE: If you suspect that a product is defective, contact the factory or the
®
Representative in your area for a return authorization number (RMA).
SOR This product should only be installed by trained and competent personnel.
670K9
Probe Installation
Probes are mounted vertically from the top of a vessel. The probe must be electrically isolated from the vessel; make no connection between the probe and the vessel other than the process connection and (if applicable) the threaded weight at the probe tip.
Do not weld any part of this instrument.
Make sure that the sensor can be fully inserted and tightened without interference from obstructions inside the tank or vessel. The probe should be mounted away from inlet fill paths. Spray from a fill path may cause false level indications.
Table of Contents
Probe Installation ...............................1
Design and specifications are subject to change
without notice.
For latest revision, go to
www.sorinc.com
Electrical Connection ...........................2
Empty and Fill Calibration .....................4
Bench Calibration ...............................5
Blind Calibration ................................6
Probe and Transmitter Performance Veri cation
..7
CE Mark Installation ............................7
RF Probe Grounding Scheme .................8
Control Drawings ...............................9
Troubleshooting ............................... 12
Dimensions .................................... 13
Form 837 (05.13) ©SOR Inc.
Registered Quality System to ISO 9001
1/16
For pressurized vessels, seal the flanged or threaded process connection to prevent leakage.
Do not use the sensor base as a handle to tighten the process connection.
Use suitable mounting bolts to mount a flanged probe on a flanged process connection.
Proper grounding is imperative for correct operation of the unit.
Sensor
Base Hub
Apply thread
sealant
Apply wrench to metal hex or wrench flats only!
Customer Ground Lead (optional)
Circuit Board Ground Lead
SORtrax Housing
Ground Screw
Open Sump or Basin
Do not suspend the unit by rigid conduit installed in the electrical hub. When installing the unit over an open sump or basin, use a suitable bracket to support the instrument.
Condensation build-up inside the electronics housing may damage the sensitive circuitry. To prevent the ingress of moisture, use drip loops or conduit runs which slope down from the enclosure. (See
)
The unit can be grounded two ways. If the tank is metallic, grounding is provided by the tank. If the tank is not metallic, or does not adequately make contact with the sensor, a separate ground wire must be provided by the customer. Run ground wiring through the conduit and into the electronics housing. Attach the ground wire per
Bracket
.
Electrical Connection
Ensure that wiring conforms to all applicable local and national electrical codes and install unit(s) according to relevant national and local safety codes.
Electrical power must be disconnected from explosion proof models before the cover is removed. Failure to do so could result in severe personal injury or substantial property damage.
2/16
Form 837 (05.13) ©SOR Inc.
This product must be installed with an explosion proof breather vent per agency requirements and the Nationsl Electric Code - Article 501, Section F, paragraph 3.
Use 18 - 22 AWG shielded twisted pair wire to make all signal and power connections. Ensure that wiring conforms to all applicable local and national electrical codes and install unit(s) according to relevant national and local safety codes.
Make sure the power source is
turned off.
Remove the housing cover.
Pull power and signal wire through
the conduit connection and into the control housing.
Locate TB1 on the control board.
(See
) Terminals are labeled
Current Meter or Resistive
Load (optional)
+
+
-
DC Power
Supply
TB1 Loop Power
Terminal Block
+ Terminal
-
Terminal
Control
Board
“+” and “-”. Connect power leads to the proper terminals.
Do not exceed the maximum loop resistance for the circuit. (See )
Use the following formula to determine the maximum loop resistance of your circuit:
R (ohms) = (input voltage - 12) ÷ 20.0 mA
Loop Resistance vs. Power Supply Voltage
2200
NOTE: Refer to the individual device speci cations of user-supplied equipment to determine the resistive load (in ohms) and the power requirements.
2100 2000 1900 1800 1700 1600 1500 1400 1300 1200 1100 1000
900 800 700 600 500 400 300
Maximum Loop Resistance (Ohms)
200 100
0
0 10 15 20 25 30 35 40 45 55
600 Ohms
at 24 VDC
1400 Ohms
at 40 VDC
1 Ohm
at 12 VDC
2150 Ohms
at 55 VDC
900 Ohms
at 30 VDC
-
Form 837 (05.13) ©SOR Inc.
Power Supply Voltage (Volts)
3/16
Empty and FIll Calibration
Electrical power must be disconnected from explosion proof models before the cover is removed. Failure to do so could result in severe personal injury or substantial property damage.
Electrical power must be disconnected from explosion proof models before the cover is removed. Failure to do so could result in severe personal injury or substantial property damage.

Install and connect power to the unit per previous pages. Install a current meter to the unit per
Turn both potentiometers (fine adjustments for ZERO and SPAN) fully counterclockwise. (See
Zero Adjustment (4 mA)
.
ZERO AND SPAN
ADJUSTMENTS
Fine
)
Coarse
ZERO
Coarse
Fine
SPAN

Close switch #1 on coarse adjustment switches for ZERO and SPAN. (See )

Open switches #2, #3, and #4 on both coarse adjustment switches per .
With the tank at 0%, view the reading on the current meter. If the output is less than 4 mA, turn the fine ZERO adjustment clockwise until 4 mA is reached.
NOTE: Only one ZERO switch may be in the closed position at any time.

If the output is greater than 4 mA, close the other switches one at a time (opening
Coarse
Adjustment
switch #1 in
the closed
position, other
switches open
the previous switch) until a reading less than 4 mA is attained. Turn the fine ZERO adjustment clockwise until 4 mA is reached.
NOTE: If 4 mA cannot be obtained, see the troubleshooting chart on page 14.
Span Adjustment (20 mA)

Set the process to desired high level. (See )
Ensure SPAN coarse adjust switch #1 is in the closed position,
100%
with all other switches open.
Rotate the SPAN fine adjustment pot clockwise until the meter indicates 20 mA. If a 20 mA reading cannot be obtained, proceed to step 10.
Open SPAN switch #1 (coarse) and close SPAN switch #2. Repeat step 9. If a 20 mA reading cannot be obtained, repeat the procedure closing switch #3 or #4 until 20 mA is indicated.
4/16
Form 837 (05.13) ©SOR Inc.
0%
Bench Calibration
Electrical power must be disconnected from explosion proof models before the cover is removed. Failure to do so could result in severe personal injury or substantial property damage.
Remove the sensor. (Unscrew the sensor from the housing. Disconnect the ring terminal or banana plug from the sensor end.)
Connect power and a current meter to the unit per . Connect a capacitor substitution box (available from SOR) to the circuit to simulate the probe. Connect the
probe lead to one terminal of the capacitor substitution box, and connect the other terminal of the box to the control housing. (See eliminate any stray capacitance which may affect the calibration.
Set the capacitance box to desired zero level
capacitance.
Rotate ZERO and SPAN fine adjustment potentiometers (pots) at least 22 turns counter-clockwise.
) Use short connection wires to
Capacitance
SORtrax
Probe Wire
Substitution
Connection
Ground Wire
Zero Adjustment (4 mA)
Close ZERO switch #1 and SPAN switch #1. (See
View the reading on the current meter. If the
output is less than 4 mA, turn the fine ZERO
) Open all other switches per .
Capacitor
Substitution Box
adjustment clockwise until 4 mA is reached.
NOTE: Only one ZERO switch may be in the closed position at any time.
If the output is greater than 4 mA, (opening the previous switch) close each coarse
adjustment switch, one at a time, until a reading less than 4 mA is reached. Turn the fine ZERO adjustment clockwise until 4 mA is reached.
NOTE: If 4 mA cannot be obtained, please see the trouble shooting chart on page 14.
Span Adjustment (20 mA)
Set the capacitance box to the desired maximum capacitance value.
Ensure SPAN coarse adjust switch #1 is in the closed position, with all other switches open.
Rotate the SPAN fine adjustment pot clockwise until the meter indicates 20 mA. If a 20
mA reading cannot be obtained, proceed to step 11.
Open SPAN switch #1 (coarse) and close SPAN switch #2. Repeat step 10. If a 20 mA
reading cannot be obtained, repeat the procedure closing switch #3 or #4 until 20 mA is indicated.
NOTE: Only one SPAN switch may be in the closed position at any time. NOTE: If 20 mA cannot be obtained, please see the troubleshooting chart on page 14.
If the zero setting is changed, the span will change, and the unit must be recalibrated.
Form 837 (05.13) ©SOR Inc.
5/16
Blind Calibration
Electrical power must be disconnected from explosion proof models before the cover is removed. Failure to do so could result in severe personal injury or substantial property damage.
Use this calibration method when it is not possible to raise or lower the level of the process in the vessel or tank. Do not permanently install the unit until after setting the ZERO (4 mA) point.
Current Meter
- +
M- and M+ optional
wiring diagram
Use the wiring configuration in when the loop cannot be conveniently opened for current monitoring. Attach a current meter as shown.
Turn both potentiometers (fine adjustments
for ZERO and SPAN) fully counterclockwise. (See
)
TB1
Loop Power
Terminal Block
M- M+
Control Board
Zero Adjustment (4 mA)
Close switch #1 on coarse adjustment switches for ZERO and SPAN. (See )
Open switches #2, #3, and #4 on both coarse adjustment switches per .
With the end of the probe just touching the process, view the output. If the output is
less than 4 mA, turn the fine ZERO adjustment clockwise until 4 mA is reached.
NOTE: Only one ZERO switch may be in the closed position at any time.
If the output is more than 4 mA, (opening the previous switch) close each coarse
adjustment switch individually until a reading less than 4 mA is attained. Turn the fine ZERO adjustment clock-wise until 4 mA is reached.
NOTE: If 4 mA cannot be obtained, see the troubleshooting chart on page 14.
Span Adjustment (20 mA)
Calibrate for span by determining the present level in the tank, and setting the span
proportionally. For best results, the level should be at 50% or greater.

Determine the level present in the tank. Complete the unit installation, except for the housing cover.
Calculate the expected current output using the following formula: 20mA x (% of tank
fill) 20mA = output.
For example: Tank level is at 50%, set SPAN coarse and fine for 20mA reading (16mA x
50% + 4mA, If the material level is at 90%, set the span adjustment for 18.4 mA (16mA x 90% + 4mA, The 100% fill level should be verified at the earliest opportunity.
If the zero setting is changed, the span will change, and the unit must be recalibrated.
6/16
Form 837 (05.13) ©SOR Inc.
Probe and Transmitter Performance Veri cation
Probe Check Remove the transmitter from the probe. With no process on the probe, check the resistance from the probe to ground. Resistance less than 1MΩ indicates leakage in the probe. With process touching the probe, check the resistance from the probe to ground. Resistance of 100KΩ or less in the case of a bare probe with conductive material indicates that an insulated probe is required. In any other case, a defect in the probe is indicated.
Transmitter Drift Check If the output of the transmitter is drifting, it is important to verify if the drift is due to the probe or the transmitter. If the probe is connected and installed properly, it will not drift.
Remove the probe from the transmitter. Without touching the calibration settings, connect a capacitance across the probe to ground input. See Figure 9 on page 5. Change the capacitance until the transmitter output is 4.00 mA. Observe the zero point reading for twenty four hours. If the reading is stable, then the probe or the application is the source of the drift.
CE Mark Installation
When subjected to an RF interference, the 670K9 will maintain the +/-1% accuracy in all frequency ranges with the following exception:
The unit is susceptible to a conducted RF interference in the range of 31-40MHz, reducing the accuracy to +/-1.3%. The +/-1% accuracy is maintained above and below this range.
In order to achieve the stated accuracy for 670K9, a shielded cable, cable gland, shield beads, and the probe should be mounted in a metallic vessel. SOR recommends using a shielded cable made of PVC insulation around a tinned copper braid shield (Olflex CY cable or equivalent). Refer to illustrations for installation of shield beads.
Install end of bead as close to cable gland as possible
2 Loops
POSITIVE
NEGATIVE
1 Loop
POWER WIRE
SHIELD BEAD
2 Loops
Long/Slender
bead
PROBE WIRE
SHIELD BEAD
Short/Fat bead
Form 837 (05.13) ©SOR Inc.
7/16
SOR RF Probe Grounding Scheme
Power Supply
Do not provide separate earth grounding for the process connection. This can create a parallel grounding circuit that will impair operation and calibration.
IMPORTANT! Do not provide separate earth grounding for the process connection. This can create a parallel grounding circuit that will impair operation and calibration.
SOR RF Probe Grounding Scheme
Critical Grounding Path =
Circuit
Board
Line
Neutral
Ground
Line Neutral Ground
Electronics
Housing
Connection
Process
SOR Supplied Stilling Well (optional)
Probe Center Conductive Element
8/16
Form 837 (05.13) ©SOR Inc.
Control Drawing
Form 837 (05.13) ©SOR Inc.
9/16
Control Drawing
10/16
Form 837 (05.13) ©SOR Inc.
Control Drawing
Form 837 (05.13) ©SOR Inc.
11/16
Troubleshooting
Symptom/Problem Possible Cause Corrective Action
Not current in the loop 1. Power supply turned off
2. Improper wiring at TB1
3. Excessive loop resistance
Zero point cannot be set to 4.00mA at low level
Span point cannot be increased to 20mA at high level
Span point cannot be decreased to 20mA at high level
1. Fine zero incorrect
2. Probe capacitance greater than 500pF when coarse ZERO switch #4 is closed (all other switches open)
1. Insufficient probe capacitance
2. Excessive loop resistance
3. Incorrect calibration
1. Low probe resistance to ground
2. Probe capacitance greater than 2,000pF
1. Check power supply
2. Check electrical connections, figure 4
3. Reduce loop resistance or adjust power supply
1. Check calibration
2. Use smaller diameter probe or replace bare probe w/insulated probe or locate probe farther from the vessel wall or consult the factory
1. Increase probe diameter or use probe w/higher dielectric insulation or locate probe closer to vessel wall or install a stilling well or consult the factory
2. Reduce loop resistance or adjust power supply
3. Check calibration
1. Consult the factory
2. Use a smaller diameter probe or replace bare probe w/insulated probe or locate the probe farther from the vessel wall
Transmitter output is 20mA or greater when the vessel is not full
Erratic transmitter output
Drifting transmitter output
Non-linear output 1. Extreme material build-up
1. Incorrect calibration
2. Probe shorted to ground
3. Material build-up on probe
1. Turbulent process
2. Radio frequency interference
3. Probe moving within the vessel
1. Process material properties are changing
2. Probe insulation is eroded
3. Transmitter malfunction
2. Non-parallel surfaces near the probe
3. Conducting liquid in an ungrounded vessel, eg: fiberglass tank
1. Recalibrate
2. Replace the probe
3. Clean probe and replace or possible misapplication, consult the factory
1. Install a stilling well
2. Install RFI/EMI filters
3. Improve probe anchoring
1. Consult the factory
2. Verify probe integrity
3. Consult the factory
1. Consult the factory
2. Mount the probe in a better location or install a stilling well, or install a probe with a ground rod
3. Connect earth ground to instrument ground
12/16
Form 837 (05.13) ©SOR Inc.
Dimensions
R Housing Con guration (Explosion Proof Remote)
Dimensions are for
reference only.
Contact the factory for certified drawings for a particular model
number.
Linear = mm/inches
Drawing 0390653
Form 837 (05.13) ©SOR Inc.
13/16
Dimensions
K Housing Con guration (Explosion Proof Remote)
Dimensions are for reference only.
Contact the factory
for certified drawings
for a particular model number.
Linear = mm/inches
Drawing 0390654
14/16
Form 837 (05.13) ©SOR Inc.
Dimensions
Other Sensors
Form 837 (05.13) ©SOR Inc.
Dimensions are for reference only.
Contact the factory for certified drawings
for a particular model number.
Linear = mm/inches
Drawing 0390654
15/16
Printed in USA www.sorinc.com
14685 West 105th Street, Lenexa, KS 66215  913-888-2630  800-676-6794 USA  Fax 913-888-0767
16/16
Registered Quality System to ISO 9001
Form 837 (05.13) ©SOR Inc.
Loading...