Sonoscape A6T, A6, A5 User Manual

Page 1
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
I
1.1 Safety Overview .................................................................... 1-1
1.2 Warning Signs .......................................................................1-3
1.3 Biological safety .................................................................... 1-4
1.4 Ultrasound Safety.................................................................. 1-4
1.5 Operating Environment ......................................................... 1-4
1.6 Transport and storage ........................................................... 1-5
1.7 Electrical Requirements......................................................... 1-5
1.8 Electrical Safety Classification............................................... 1-6
2.1 General.................................................................................. 2-1
2.2 Application............................................................................. 2-1
2.3 Contraindication .................................................................... 2-2
2.4 System Configuration ............................................................ 2-2
2.4.1 Configuration.................................................................. 2-2
2.4.2 Standard Configuration................................................... 2-3
2.4.3 Optional Accessories...................................................... 2-3
2.4.4 Optional Peripherals....................................................... 2-3
2.5 Technical Specifications ........................................................ 2-5
2.6 System Principle.................................................................... 2-8
3.1 Operation Notice ................................................................... 3-1
3.2 Transducer Connection ......................................................... 3-1
3.3 System Power ON/OFF......................................................... 3-2
3.4 General Operation................................................................. 3-3
3.4.1 Keyboard Operation ....................................................... 3-3
3.4.2 User Interface................................................................. 3-6
3.4.3 System Setting ............................................................... 3-7
3.4.3.1 General Setting ................................................ 3-8
3.4.3.2 Dicom Setting................................................... 3-9
3.4.3.3 Formula Select............................................... 3-10
3.4.3.4 Formula Define .............................................. 3-12
3.4.3.5 Comment Setting ........................................... 3-14
3.4.4 Patient Information ....................................................... 3-17
3.4.3.6 Disk Management .......................................... 3-18
3.4.3.7 LCD Setting.................................................... 3-22
3.4.5 BodyMark ..................................................................... 3-23
TABLE OF CONTENTS
Page 2
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
II
3.4.6 How to use biopsy ........................................................3-24
3.4.7 Report........................................................................... 3-25
3.4.8 Use the peripherals ......................................................3-26
3.4.8.1 USB device ....................................................3-26
3.4.8.2 Footswitch ...................................................... 3-27
3.4.8.3 Video Out .......................................................3-27
3.4.8.4 Network Device .............................................. 3-27
3.5 Imaging................................................................................3-28
3.5.1 Basic operation............................................................. 3-28
3.5.2 Parameter Adjust .......................................................... 3-29
3.6 Measurement....................................................................... 3-32
3.6.1 B mode basic measurement.........................................3-33
3.6.1.1 Distance Measurement ..................................3-33
3.6.1.2 Angle measurement .......................................3-34
3.6.1.3 Ratio measurement ........................................3-35
3.6.1.4 Area/Circumference measurement ................ 3-36
3.6.1.5 Volume measurement (2-Axis, 3-Axis)...........3-38
3.6.2 B mode special measurement ......................................3-40
3.6.2.1
Basic OB measurement .................................3-41
3.6.2.1.1 GS (Gestational Sac) measurement .......3-41
3.6.2.1.2 CRL measurement .................................. 3-42
3.6.2.1.3 BPD (Biparietal Distance) measurement. 3-43
3.6.2.1.4 HC (Head Circumference) measurement 3-43
3.6.2.1.5 AC measurement ....................................3-44
3.6.2.1.6 FL (Femur Length) measurement ...........3-45
3.6.2.1.7 OFD measurement ............................... 3-46
3.6.2.1.8 CER measurement.................................. 3-47
3.6.2.1.9 AFI (Amniotic Fluid Index) measurement 3-47
3.6.2.1.10 Report .....................................................3-48
3.6.2.2 High-level OB measurement ..........................3-49
3.6.2.2.1 APAD measurement................................ 3-49
3.6.2.2.2 AA (Abdominal Area) measurement........3-50
3.6.2.2.3 HL (Humerus Length) measurement.......3-51
3.6.2.2.4 THD measurement..................................3-52
3.6.2.2.5 FTA (Fetal Trunk Area) measurement .....3-53
3.6.2.2.6 OOD measurement ................................. 3-53
3.6.2.2.7 TIBIA measurement ................................3-54
3.6.2.2.8 ULNA measurement................................ 3-54
3.6.2.2.9 Input Last Menstrual Period (Input LMP). 3-55
Page 3
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
III
3.6.2.2.10 Fetal Growth Curve (FGC)...................... 3-55
3.6.2.2.11 Fetal Biophysical Profile (FBP) ............... 3-56
3.6.2.3 GYN measurement ........................................ 3-60
3.6.2.3.1 UT measurement .................................... 3-61
3.6.2.3.2 Endo measurement ................................ 3-62
3.6.2.3.3 L.Ovary (Left Ovary) measurement......... 3-63
3.6.2.3.4 R.Ovary (Right Ovary) measurement...... 3-64
3.6.2.3.5 L.Follicle (Left Follicle) measurement ..... 3-65
3.6.2.3.6 R. Follicle (Right Follicle) measurement . 3-66
3.6.2.3.7 CX (Cervix) measurement ...................... 3-67
3.6.2.3.8 CX-L /UT-L.............................................. 3-68
3.6.2.4 Cardiac measurement.................................... 3-68
3.6.2.4.1 LV function .............................................. 3-69
3.6.2.4.2 Right Ventricle inner diameter (RV)......... 3-76
3.6.2.4.3 Pulmonary Artery inner diameter (PA)..... 3-77
3.6.2.4.4 Left Ventricular Mass (LV Mass) ............. 3-77
3.6.2.4.5 Left Ventricle Mass Weight Index............ 3-78
3.6.2.5 Small Part measurement ............................... 3-79
3.6.2.5.1 URO measurement (RUV) ...................... 3-79
3.6.2.5.2
Prostate measurement (PSV) ................. 3-80
3.6.2.5.3 Thyroid measurement (THY) .................. 3-82
3.6.3 M mode measurement ................................................. 3-83
3.6.3.1 Distance Measurement .................................. 3-83
3.6.3.2 Time measurement ........................................ 3-84
3.6.3.3 Slope measurement....................................... 3-84
3.6.3.4 HR measurement........................................... 3-85
3.6.3.5 LV function measurement (LV)....................... 3-86
3.6.2.5.1 TEICHHOLZ ........................................... 3-86
3.6.2.5.2 CUBE...................................................... 3-87
3.6.2.5.3 GIBSON.................................................. 3-88
3.6.3.6 Mitral Valve measurement (MV)..................... 3-88
3.6.3.7 Aortic valve measurement (AO) ..................... 3-90
3.6.3.8 Ejection Time (ET) ......................................... 3-92
3.6.3.9 TEI measurement .......................................... 3-92
3.7 Other Operation................................................................... 3-93
3.7.1 Save............................................................................. 3-93
3.7.2 Saving exam mode....................................................... 3-93
3.7.2.1
Saving image file............................................ 3-94
3.7.3 Cineloop ....................................................................... 3-95
Page 4
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
IV
3.7.3.1 Single image cineloop ....................................3-95
3.7.3.2 Continuous cineloop.......................................3-95
3.7.3.3 M mode cineloop............................................3-96
3.7.3.4 Review images...............................................3-96
3.7.4 Clipboard ...................................................................... 3-96
3.7.5 Software Upgrade......................................................... 3-97
4.1 Use transducer ......................................................................4-1
4.2 Use biopsy guide ................................................................... 4-1
4.3 Transducer Specifications...................................................... 4-2
5.1 Main Unit ...............................................................................5-1
5.1.1 Maintenance ...................................................................5-1
5.1.2 Installation ......................................................................5-2
5.1.3 System Transport ...........................................................5-2
5.2 Transducer.............................................................................5-3
5.2.1 Maintenance ...................................................................5-3
5.2.1.1 Use endocavity transducer...............................5-5
5.2.2 Cleaning .........................................................................5-6
5.2.3 Disinfection.....................................................................5-6
5.3 Service Method...................................................................... 5-7
5.3.1 Service Method...............................................................5-7
5.3.2 Service responsibility......................................................5-8
5.3.3 Service contact information ............................................5-8
Page 5
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
1
Caution
This manual contains necessary information to operate the system safely. Please read this manual before attempting to use the system.
This system has been designed with safety consideration of the operators and the patients. However, to ensure the system reliability and safety, please pay attention to the following:
z This system shall be operated only by or under the guidance
of a qualified person.
z This system complies with Type B general equipment, Class
I of EN606601-1:1990+A1:1993+A2:1995 standard.
z Do not modify this system in any way. Necessary
modifications can be made only by the manufacturer or its designated agents.
z This system is fully adjusted at the factory. Do not adjust
any fixed adjustable parts.
z In the event of a malfunction, turn off the power supply
immediately and inform the manufacturer or its designated agents.
z The power cable of the system must be connected to a
grounded power socket. Do not remove the ground cable for any reason.
z Only the devices comply with
EN606601-1:1990+A1:1993+A2:1995 standard can be connected with the system, either electronically or mechanically. Recheck the leakage current and other safety performance indices of the entire system to avoid potential system damage caused by leakage from a current superposition.
z The system has no any specialized design when it is
configured with high-frequency operation devices. The operator shall use caution in these types of applications.
Page 6
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
2
z The system is designed to provide medical information on
clinical diagnosis for doctors. The doctors shall be responsible for determining diagnostic results, while the manufacturer shall not assume any responsibility for such diagnostic process.
z Please back up the important data of the system to external
storage media, such as a U disk or a DVD.
z The manufacturer shall not be responsible for any loss of
stored data in the system due to any malfunction.
Due to functional update, this user’s manual may be slightly different from the system you are using.
Regulatory Requirement
This product complies with regulatory requirements of the following European directive 93/42/EEC concerning medical devices.
Page 7
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
1-1
1 System Safety
1.1 Safety Overview
Safety measures for the operator and the patient are discussed in this section. To ensure the safety of the operator and patient, please read the relevant details in this chapter carefully before operating this system. Meanwhile, in this manual the description on warning means that any violation of the relevant rules may result in personal injury or even risk of life to the operator or the patient.
The following precautions should be adhered to:
1. This system complies with Type B general equipment, Class I, EN606601-1:1990+A1:1993+A2:1995 standard. Please follow Section 1. “System Safety” in the user’s manual to operate this system properly.
2. Do not modify this system in any way. Necessary modifications can be made only by the manufacturer or its designated agents.
3. This system is fully adjusted before delivery from the factory. Do not adjust any fixed adjustable parts.
4. If any disorder occurs during operation, please turn off the power supply immediately and inform our designated contact persons or us.
5. The power cable of the system should be connected to a grounded power socket, and the ground cable should not be removed.
6. Only the devices comply with EN606601-1:1990+A1:1993+A2:1995 standard can be connected with the system, either electronically or mechanically. It is required to recheck the leakage current and other safety performance indexes of the whole system to avoid any potential hazard to the system resulted from leakage current superposition.
Page 8
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
1-2
7. When it is configured with high-frequency operation devices, no special protective measures are supplies for the system. The operator should be vigilant to such application.
8. The installation shall be performed by personnel authorized by the manufacturer only. Do not attempt to install the system by yourself
9. Service shall be performed by service engineer authorized by the manufacturer only.
10. The system shall be used by a qualified operator or under the supervision of a qualified person.
11. The system is not designed to be operated in the presence of flammable substances. Such an environment could cause an explosion.
12. Power down the system before cleaning. Protect the system from water or other liquids that could drip into the system.
13. To ensure the safety of the patient, continuous scanning and imaging for a long period on the same part of the patient is prohibited.
14. When the system is used for ultrasound scanning, it shall be used with ultrasound gel without problem in biological compatibility. To ensure safety, only qualified ultrasound gel in compliance with related standard can be used.
15. At the end of the life time of this product, please contact your local agent or our company for disposal.
Page 9
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
1-3
1.2 Warning Signs
Signs Description
Type B Applied Part symbol classified as electric shock prevention
Dangerous voltage
Attention; To secure system safety, wherever this sign is found in the system or documents please refer to relevant instructions in the user’s manual.
OFF (The power disconnected from the unit)
ON (The power connected to the unit)
Equi-potentiality sign; The terminal shall be connected with the corresponding connection terminal of other devices, to ensure the equation of electric.
Alternating current
Operating the system in the presence of flammable gas is prohibited; Otherwise it may cause an explosion.
Direct current
Page 10
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
1-4
1.3 Biological safety
This product should be used only for valid reasons and should be used both for the shortest period of time and at the lowest power settings necessary (ALARA - As Low As Reasonably Achievable) to produce acceptable diagnostically image.
1.4 Ultrasound Safety
Keep the transducers clean. The transducers shall be cleaned before scanning another patient.
Patients should always be exposed to the lowest transmit power for the shortest time. Freeze the system or keep the probe away from the patient if no scanning is performed. Do not scan the patient with the probe at a fixed position of the body for a long time; Please refer to Appendix C for acoustic report of the system.
1.5 Operating Environment
¾ The unit should be operated within the parameter outline below:
z Relative humidity: 30%~75% non-condensing
z Ambient temperature: 10 ~ 40
z Atmosphere Pressure: 7001060hPa
¾ Strong radiation source or powerful electromagnetic waves, e.g.
electromagnetic waves from radio broadcasting, may result in image ghosting or noise. The system shall be isolated from such radiation source or electromagnetic waves.
¾ Do not put anything on top of the main unit.
¾ When the system is not in use, be sure to turn off the power
supply. Cover the main unit with a dust cover. The system shall be operated in the environment without dust.
Page 11
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
1-5
1.6 Transport and storage
¾ The following transport and storage environmental conditions are
within system tolerances:
z Relative humidity: 20%90 non-condensing
z Ambient temperature: -20 ~40
z Atmosphere Pressure: 7001060hPa
1.7 Electrical Requirements
¾ Power Requirements: 100-240V~, 50/60Hz
¾ Power Consumption: 130 VA
¾ Main unit voltage: With a fluctuation range of less than ±10%
otherwise the system may be damaged.
¾ Grounding: Before connecting the power cable, connect the
attached ground protection cable to a special grounding device.
NOTE:
Please follow the above requirements for power supply to the system.
The requirements of power supply may vary in different areas. Please refer to the symbols at the rear panel of the system for the detailed requirements.
Page 12
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
1-6
1.8 Electrical Safety Classification
¾ This safety standard of this system complies with Type B, Class I
EN606601-1:1990+A1:1993+A2:1995;
¾ It belongs to continuous operation (for more than 8 hours)
according to operation mode;
¾ According to the classification on harmful liquid
protection(Main unit:IPX0 Transducers:IPX4)
¾ The system should NOT be used under the condition with the
inflammable mixture of anesthesia gas and air, or oxygen and nitrous oxide;
¾ This equipment meets the requirements for EMC/EMI
IEC60601-1-2 Class A (CE);
¾ The system shall be added with a ground device protection.
During operation, please make sure that the system is connected to an extra grounding device. Otherwise it may be interfered by alternating current and affected in imaging quality.
Page 13
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
2-1
2 System Description
2.1 General
The system is a full digital portable B/W diagnostic ultrasound system with high performance. With the most advanced full digital ultrasound design in the industry, the system is configured with high performance super large scale integration, and human tissue ultrasound information can be acquired to its best capability; In combination with professional built-in Linux computer system, rich data ports, professional application packages and standard Windows style interface, the system is featured with high performance, as well as powerful software function and easy operation.
A high-performance industrial-grade LCD is used on the systems. Other than the traditional systems, clear image display and abundant gray scales can be found on this system, meanwhile the system weight is reduced significantly.
Moreover, the professional built-in operation system and the standard PC ports enable the user to perform system upgrade and maintenance without effort, while maintaining the advantages of its technologies.
2.2 Application
The system, with electronic scanning ultrasound diagnostic capabilities, is suitable for all kinds of clinical diagnosis. It can work with many multi-freq transducers, and can apply to general organs. Its main applications are as follows:
z Small parts (breast, thyroid, testes, etc.)
z Vascular (cerebrovascular, peripheral vascular)
z Abdominal (liver, spleen, gallbladder, kidney, etc.)
z Gynecological
z Cardiac
z Urological
z Musculo-skeletal
Page 14
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
2-2
2.3 Contraindication
No Contraindication. But the system is not intended for OPHTHALMIC use or any use that causes the acoustic beam to pass through the eye(s).
2.4 System Configuration
2.4.1 Configuration
The system consists of a main unit and transducer(s).
1. RS232 Port
2. LAN PortDICOM Port
3. VGA Port
4. S-Video Port
5. Double USB Port
Page 15
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
2-3
2.4.2 Standard Configuration
z Host 1 pc
z Power cable 1 pc
z Configurable transducers (models and quantity are subject to
customer requirements. See Section 4.3 for transducer models. )
z Protective ground cable 1 pc
z Fuse 1.6A/250V 2 pcs
z S-terminal cable 1 pc
z BNC video cable 1 pc
z USB2.0 cable 1 pc
z VGA video cable 1 pc
z Ultrasound gel 1 bottle
2.4.3 Optional Accessories
z DVD Combo
z Footswitch
z Biopsy guide
z Trolley
z Special backpack
z Configurable transducers (See Section 4.3 for transducer
models.)
2.4.4 Optional Peripherals
z USB2.0 memory stick
z USB2.0 mobile hard disk
z USB2.0 DVD-RW
Page 16
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
2-4
z USB2.0 video printer
z Composite video printer
z Network printer
z VCR
1. REMOTE Control Port
2. S-Video Port
3. Foot Switch
4. Ground Pole
5. Power Connector with Fuse-box
6. Power Switch
7. Probe Connectors
Page 17
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
2-5
2.5 Technical Specifications
z All digital beam former
z Dynamic receive focusing, dynamic aperture, dynamic
apodization;
z Transducer: multiple-frequency transducer, 5-frequency,
frequency range 2.0-12.0MHz;
z Probe connector: 2 slots
z Scanning range:
Convex array: Max depth300mm
Linear array: depth 123mm, width 46mm
Micro-convex array: Max angle 180°, during 2.0MHz, max
Depth 320mm
z Grayscale: 256
z Dynamic range: 108dB;
z Imaging mode: THI , B , B+B , B+M , M , 4B;
z Scanning lines: 256 lines in B mode max. 256 lines in M mode
max;
z Line density: low, high;
z Edge enhancement: 8 steps;
z Grayscale curve: 16 types;
z Persistence: 8 steps;
z Chroma: 8 types;
z Frequency: 5 steps;
z Focusing method: dynamic focusing, variable aperture;
z Focus number: Max 4, focus number and depth controllable
z Frame rate: max 60 frames;
z Gain control: gain 0255, encoder adjust; 8-TGC; independent
Page 18
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
2-6
gain control for B and M modes;
z Zoom factor: max. 8.0, local zoom function; z M mode: sweep speed 12345s; z M mode video invertion; z Biopsy guide line: on/off, position and angle adjustment; z User adjustable parameters: dynamic range, frequency, GSC,
persistence, gain, grayscale curve, left/right invert, up/down invert, image rotation angle, sweep angle/width, depth, line
density, zoom factor, focus number and position, biopsy guide
line, acoustic power, chroma pitcure;
z Diagnostic mode: applications such as cardiac, liver, kidney,
muscular-skeletal, gynecological, obstetric, thyroid, carotid, breast and pancreas; exam mode preset by the user;
z Cine and storage:
In B mode, 256 frames with high line density, or 512 frames
with low density; 128s in M mode;
A flash can store up to 2000 images in bmp format, 10000
images in JPG format after power off.
Cine speed adjustable: high, medium, low; Measurement when reviewing images;
z Clipboard function: the latest reviewed images (up to 32 images)
can be shown at the bottom right of the screen;
z Body mark: not less than 80 type; z Annotation: rich annotations, the user can add new annotations; z Measurement:
Basic measurement: distance, ratio, area (ellipse),
circumference, angle, volume, time, slope, heart rate
OB measurement: BPD, CRL, FL, AC, HC, GS, VOL, ANG
LE, UT, OFD, CI, AFI, FGC
Page 19
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
2-7
Cardiac measurement: MV, LV, AO, EDV, ESV, EF, CO, SV,
FS, ET
Other measurement: RUV, PV, and so on
z Report: cardiac report, vascular report, OB/GYN, growth curve,
urology;
z Display: 12-inch industrial-grade LCD, the user can adjust the
monitor through system menu;
z Built-in Linux system;
z Standard PS2-port keyboard, backlight display, high-brightness
control, keyboard with beeper prompt;
z Multiple language operation interface: Simplified Chinese/
Traditional Chinese/ English/ Russian/ Spanish/ French/ Italian;
z Input in Chinese/ English;
z Video out format: composite video, S-Video, VGA;
z Output ports: LAN, 2×USB2.0, composite video, VGA, S-Video,
print control;
z Input port: footswitch;
z DVD-RW (Option);
z Optional peripherals: footswitch, video printer, USB video printer,
USB disk, mobile hard disk, network device, VCR;
z DICOM function: (Option);
z File output format:
Image: BMP, JPG, AVI,DCM;
Report: .RPT;
z Software update: USB update by the user or update through
Flash replacement;
z Power in: 100-240V, 50/60Hz auto adaption;
z External dimension: 336×180×346mm(L×W×H
Page 20
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
2-8
z Net weight: 6.5Kg;
z Environmental requirement:
Operation Storage and Transport
Ambient temperature: 10~40 -2055
Relative humidity: 30%~75% 20%90
Atmospheric pressure: 7001060hPa 700~1060hPa
2.6 System Principle
The fundamental operating principle of diagnostic ultrasound system is: through the transducer, electric excitation signals are converted into acoustic signals, which are propogated, attenuated and reflected in the patient’s body. Then the echos from the body are converted into electric signals through the transducer. Going through complicated signal processing circuit, a scanned tomographic image is displayed on the monitor.
The block diagram of the system is displayed below:
Fig.2-6a Block Diagram
Page 21
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
2-9
According to the design, the Digital Beam Former sends out transmit pulse signals at a certain frequency. Going through signal driving and logic switch, the signals are sent to the elements of the transducer. With the oscillation of element, ultrasonic waves are generated. The combinations of many element signals enable the transducer to send out a dynamic directional compounded beam. Then the beam is propogated and attenuated in the human body. When the echo returned to the element crystal, it will be changed to electric signals again. Being amplified by the pre-amplifier, the signals are sent for A/D conversion, through which analog signals are converted to digital signals, compounded in the beam former, go through digital signal processing such as filtering, logarithm compression, detection and form digital beam data stream. The beam data steam is sent to the CPU board goes through scan transform and video processing. Finally, an ultrasonic image is displayed on the monitor.
The core of the CPU board is a built-in Linux system. In addition to completing display of ultrasonic images, it accepts keyboard operation commands from the user, and changes them to internal executable commands. Moreover, it will achieve all kinds of measurement and calculation, data input and output of peripheral ports, and so on. The system can also work with peripheral devices such as a printer, a network device or an external memory, so as to extend system functions.
Page 22
Page 23
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-1
3 System Operation
3.1 Operation Notice
Warning:
1 The system shall be operated by a trained medical operator;
2 Before using this system, please read the relevant sections in the
manual carefully.
3.2 Transducer Connection
The transducer ports are at the back of the system.
With the cable side of transducer socket downward, plug in the transducer and turn the transducer knob clockwise to lock it. To remove the transducer, turn the knob counter clockwise and plug out the transducer. The transducer can by plugged in either of the two ports.
See the following figures:
Fig. 3.2a How to connect a transducer
Page 24
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-2
NOTE:
To ensure normal system functioning, please plug in or out the transducer only when the system is powered off or in frozen state.
3.3 System Power ON/OFF
Warning:
Before the system is powered on, read the relevant description in Section
1.8 Electrical Safety carefully.
Before the system is powered on, please make sure that:
1 The specifications of the power complies with the labeled power
spec at the rear panel of the system;
2 The protective grounding terminal of the rear panel is connected
well with the ground via the ground cable provided with the system;
3 The power switch on the rear panel is in OFF state;
¾ System power connection
Connect the system to the power supply with the power cable provided with the system.
¾ Power on the system
When the system power is connected, press the power switch on the rear panel to turn on the system. When the system is turned on, the backlight on keyboard will on, and the screen will show logo of splash screen. The system will complete the initialization in about 1 minute and go to exam mode select screen.
Page 25
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-3
¾ Turning off the system
Switch the power switch on the rear panel to off state, and the system will be turned ff immediately.
¾ Reset
If any abnormal situation, such as software failure occurs on the system, press the reset key on the side board to reset the system; The reset process will restart the main board work state, and start the software system, but does not restart the power supply.
NOTE:
The reset key can be used for reset function with the tip of a non-metal object such as a pencil. Generally the user is not recommended to perform such operation directly. It shall be carried out by an engineering staff only.
3.4 General Operation
3.4.1 Keyboard Operation
The system is designed with backlit keyboard. Based on user operation habit, the keys are arranged in a optimized method, so that the user can immediately achieve all the complicated operations.
There are 65 keys (10 numerical keys[09], 26 character keys [AZ], 6 control keys, and 23 function keys), 3 digital encoders with buttons (FUNC, MENU, GAIN) and 8-step TGC on the keyboard.
By pressing 【Shift】key, the upper case of that key can be actived.
Page 26
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-4
The keyboard control panel is shown as below.
Fig. 3-4-1a Keyboard Control Panel
Basic functions of each key are shown in the following table.
Item English Description of Function
(1) Exam Select an exam mode for the transducer
(2) Patient Input patient information
(3) Comment Input comments
(4) File Go to disk management system
(5) Clipboard Go to clipboard function
(6) THI Tissue Harmonic Imaging
(7) Freq. Change transducer frequency
(8) Character,
number key,
Control key
Input characters or numbers; Caps Lock:
capital lock; Shift: to activate upper case
definition; Backspace: to delete characters
one by one; Clear: clear all information
Page 27
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-5
Item English Description of Function
(9) TGC Adjust ultrasound echo gain in depth direction
(10) B B mode
(11) B/B
Dual B mode, or switch the activation between
the left and right images
(12) Up/Down Invert the image vertically
(13) B/M B/M and full M mode
(14) 4B 4B mode
(15) L/R Invert the image horizontally
(16) SAVE
Save exam mode; save single images or
cine-loop
(17) REPORT Go to exam report screen
(18) PRINT Print the current screen
(19) FUNC
Press the key to switch between depth, focus
or zoom rate functions, and adjust parameter
of the current state
(20) MENU
Press the key to go to system setup menu and
image adjust menu; Use the knob to adjust
menu parameter of the current image
(21) Set Input confirm key, increase, left key function;
(22) Distance Distance measurement mode
(23) Area Go to ellipse, trace area measurement mode
(24) Calc B mode or M mode measurement mode
(25) Bodymark Go to body mark mode
(26) Esc Cancel input key, decrease, right key function
(27) GAIN Adjust B or M image gain
(28) Freeze Freeze or unfreeze the image
Table 3.4.1 Description of Key Function
Page 28
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-6
3.4.2 User Interface
The main user interface is shown below:
The functions of the main sections on the interface are shown below:
1 Logo of manufacturer
2 Image parameter area
3 Measurement result area
4 Image/measurement menu area
5 System information area
6 Imaging area
7 User prompt
8 Grayscale bar/TGC curve
Fig. 3.4.2a Main User Interface
Page 29
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-7
3.4.3 System Setting
In exam mode select page, press the【MENU】 key, and it will show the System Setting page. Press the【MENU】key again, or click [Exit], it will exit the System Setting page.
The System Setting screen is shown below:
Fig. 3.4.3a System Setting
System Setting includes: General Setting, Dicom Setting, Formula
Setting, Formula Define, Comment Setting.Click the item selected to enter the
relative interface.
Page 30
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-8
3.4.3.1 General Setting
In the System Setting page, click [General Setting] to go to General Setting page.
General Setting page is shown below:
Fig. 3.4.3.1-a General Setting page
The setup options of General Setting are shown below:
Hospital Name
Date To input the system date as YYYY-MM-DD.
Time To input system time as per the format HH:MM: SS.
Language
Screen Save combo-box select, OFF, 5 Min, 30 Min.
B/M Type combo-box select, 1:1 LB/RM, 1:1 TB/BM, 1:2 TB/BM.
Image Format combo-box select, JPG, BMP, DCM.
Key in by the keyboard, up to 20 Chinese characters or 40 English characters can be inputted.
Select one language from the combo-box: Simplified Chinese, Traditional Chinese, English, Spanish, Russian, Italian or French.
Page 31
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-9
Video Format combo-box select, NTSC, PAL.
System Color combo-box select, Gray, Color.
System Ver. Un-modifiable, to review system release information.
Confirmation exit and save the curent setting
Cancel exit without saving
Load default load default the setting imformation
3.4.3.2 Dicom Setting
In General Setting page, click [DICOM Setting]. If the system is configured with DICOM, it will go to DICOM Setting screen; otherwise the screen will show a prompt message.
DICOM Setting page is shown below:
Fig. 3.4.3.1-b Dicom Setting page
Page 32
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-10
The setting items for Dicom Setting are shown below:
Local Setting:
z Active
z
DHCP Setting The system IP address configuration.
z
Fix IP Configure Static IP, Sub IP, Gateway, Printer
IP address.
DICOM Server:
z Transfer to USB User can transfer DICOM files through
USB;
z Transfer to Server
In General Setting, click [Ok], it will exit and save the current setting info; click [Cancel], it will exit without saving.
3.4.3.3 Formula Select
In System Setting page, click [Formula Select], go to Formula Select page. Click [Save] to exit and save the current setting info; click [Exit] to exit without saving the current setting info.
To turn on/off network; to check Active select box means to enable the function; only when the function is active, the system can perform network connection; when the function is active, select DHCP Setting or Fix IP.
User can configure remote IP, port, Access number. When it is setted, DICOM files can be transferred through network;
Page 33
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-11
The formula select page is shown below:
Fig. 3.4.3.3 Formula select label
The measurement items and the setting method of formula select are shown below:
GS Select GS measurement formula from the
combo-box.
FL Select FL measurement formula from the
combo-box.
THD Select THD measurement formula.
Save Exit and save the formula selected
Exit Exit without saving
Load Default Load the default formula for all measurement.
Page 34
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-12
3.4.3.4 Formula Define
In System Setting page, click [Formula Define], go to Formula Define page. Click [Save] to exit and save the current setting info; click [Exit] to exit without saving the current setting info.
Fig. 3.4.3.4 User-defined formula label
Page 35
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-13
In the User Define formula, the user can create a user define table for each measurement.
The measurement items and the setting method of User Define formula are shown below:
Measurement Select the measurement item to be edited from
the com-box.
Formula Select the formula of measurement item from
the com-box.
Dist Key in the distance value of the measurement
table, value is divided into 32 steps.
GA Set the GA result value corresponds to each
distance.
Save Click it, and in the pop-up input dialog box, input
the name of the new formula. Click Save to save the new formula and exit; The system will add one “U” before the new formula name automatically, which means that it is user defined;
Exit Click to exit Measure Setting.
Del Formula Delete the selected formula;
NOTE:
Only the user-defined formula can be deleted.
Page 36
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-14
3.4.3.5 Comment Setting
In System Setting page, click [Comment Setting] to go to Comment Setting page. Click [Exit] to exit Comment Setting. Comment Setting page is shown below:
Fig. 3.4.3.5 Comment Setting page
The Comment Setting page can be divided into the following areas:
(1) Application select area;
(2) System predefined comment area;
(3) User Define comment area, each application have 6 additional
user-defined annotations;
(4) User defines comment operation area.
In Comment Setting, different measurement parts are displayed in screens, but the operation methods are consistent.
Page 37
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-15
Here is an instruction on how to use the User Define comment:
Select Click any of the buttons in User Define comment display
area, and the user defined comment for that button will be displayed in the input box.
Input Use the keyboard to input new user defined comments.
Add
Del
z How to use Comment
In an exam mode, press the【Comment】 key to active the comment screen. Click [Exit], or press the【Comment】key again to exit adding comments.
When adding a comment, the user may follow the steps:
Start to comment
Confirm position
Manual input
Click add, and the user defined comment is added to the user defined comment display area.
Click Del, the selected user-defined comment is deleted from the user defined comment dis
play
area.
Press the【Comment】key key, and the cursor will change to “|”shape to mean that it is now in the comment state;
Move the cursor to the position to add the comment. Press the【SET】key to confirm the position. The cursor will changes to “|”, which means that it is ready to input comment. Manual input and Auto input
Input the comment from the keyboard, numbers and letters can be inputted; With theCaps Lockkey, the user can switch between upper and lower cases of letters; By press the【Shift】+【Space】key, user can switch to the Chinese characters input state. After the input, press the【Enter】key to confirm.
Page 38
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-16
Auto input
Fig. 3.4.3.5-b Comment Setting page
Move the comment Move the cursor to comment, the cursor
changes to “ “ , which means that the
comment is moveable. Press the【SET】key
to select the comment, now a box is enclosed around the comment, which means that the comment can be moved with trackball movement. Move the trackball to the target position. Press the SETkey, and the comment is moved to a new position.
Delete a comment Select a comment, press the【Backspace
key to delete the selected comment. Press
theClearto clear all the inputted comments.
In comment state, press the【Esc】key, an auto input adding page will be shown. Select the desired comment, press the SETkey, the comment will be added. Then the user may continue to input another comment.
Page 39
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-17
3.4.4 Patient Information
Press the【Patient】key, active the Patient Information page.
Click Save, save the current Patient Information; click [Exit], or press the Patientkey again, it will exit without saving the current information.
The Patient Information page is shown below:
Fig. 3.4.4 Patient Information page
The setting items and setting method of Patient Information page are shown below:
Name Input from the keyboard. 5 Chinese characters or 10
english letters.
Sex Male or female can be selected by com-box.
Height Input from the keyboard. 3 numbers, in the unit of cm.
ID Input from the keyboard. Patient ID, 10 letters or
numbers.
Age Input from the keyboard. 3 numbers.
Weight Input from the keyboard. 3 numbers, in the unit of kg.
ClinicNo Input from the keyboard. Clinical number, 10 letters or
numbers.
Page 40
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-18
BedNo Input from the keyboard. Bed number, 10 letters or
numbers.
LMP Input the last menstrual period of the patient by the
keyboard.
Comment Input from the keyboard. 40 Chinese characters or
80 English letters.
New Patient
More
Tel Input by the keyboard.
Add Input by the keyboard.
Memo Input by the keyboard.
3.4.3.6 Disk Management
Disk Management function is to manage files and directories on the system disk or external USB disk.
Press the 【File】key, or in [System Setting] page, click [Disk Management], active Disk Management page.
Press the【File】key again, or click [Exit] to exit Disk Management page.
The Patient Information page will be extended, which includes: Tel, Address, Memo. Click it again to go back to the initial Patient
All the information of the current patient will be cleared when new patient data input, including the report for the performed measurement.
Page 41
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-19
Disk Management page is shown below:
Fig. 3.4.3.2-a Disk Management page
Disk Management screen is divided into:
(1) Driver and current path display area;
(2) Directory and file operation select area;
(3) Directory display area for the current path;
(4) File display area for the current path;
(5) Other operations.
The operation of Disk Management is described below:
Drive drop down frame selectionA is the system disk;B、C is the external U disk.
z Dir (Directory)
The Dir is only valid for an external USB disk. The user shall select the USB disk (Driver B, C) from the driver com-box, and on the left of the file manager it displays the USB disk directory.
.. Click ‘.’ to go to the upper level directory.
Page 42
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-20
Subdirectory Click to select the target directory; Click it again to
select the directory, and it goes into that directory.
Select the target directory, and then click the operation item:
New
Copy To copy a directory and get ready for Paste
Paste To paste a directory, the Paste operation is invalid for the
system disk
Del To delete a directory. There will be a confirm dialog box.
Rename To rename a directory. In the pop-up input confirm dialog
box,input the name of a new directory. Click Ok to confirm.
NOTE:
If a repeated directory name is displayed when performing New or Rename, it will prompt that such operation fails.
z File
Part of the File operation, the Copy, Del and Rename operations are all valid for system disk and the USB disk. Paste works for the USB disk only. The user shall select the USB disk (Driver, C) from the driver drop-down box, and on the right of the file manager it displays all the files under the current directory. Select the target file, and then click any of the items under File:
Copy To copy a file, copy the target file to the buffer area and
get ready for Paste.
Paste To paste a file (before to paste, change the current
directory to paste the file), paste the file in the buffer area to the current directory, and clear the buffer area. In the system disk, the Paste operation is invalid.
Del To delete a file, In the pop-up confirm dialog box, click
Ok to confirm and complete the deletion of a target file;
Rename To rename a file, in the pop-up input confirm dialog box,
input the name of a new directory. Click Ok to confirm.
To create a new directory. In the pop-up input dialog box, input the name of a new directory. Click Ok to confirm;
Page 43
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-21
If the inputted file name already exists, it will prompt operation fails. If the inputted file name already exists when performing Paste, it will prompt whether to overwrite or not.
z Others
DICOM Trans If the system has DICOM function, it will select to
transfer files in DICOM format.
Load To load the selected file (image or report) to the
system.
Burning Files
Burning Files page is shown below:
Fig. 3.4.3.2-b Burning Files screen
To go to the burning file page.
Click [Burning Files] in [Disk Management] page, and it goes to Burning Files page. Click [Exit] to exit Burning Files page.
Page 44
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-22
The instruction for performing Burning Files is described below:
Test Insert a CD, click Test, and it displays the CD
information and the capacity available.
>/>> To add a file; Select the target file, click’>’, to add one
file to Burning Files buffer area; click’>>’, to add all the files in the current directory to the Burning Files buffer area. The percentage of the Burning Files buffer area capacity to the whole CD capacity is prompted in Used and Available display area. The total size of the selected files shall not exceed the capacity of the CD.
</<< To delete a file. Select one file from the Burning Files
buffer area, click’<’, and the selected file is deleted from the Burning Files buffer area; click’<<’, and all the content in the Burning Files buffer area sis deleted.
Burning Files Click it to start Burning Files operation. After
Burning Files is finished, the system will prompt “Success”. For the user, Burning Files enables auto continuing burning without inserting a new CD every time.
3.4.3.7 LCD Setting
In System Setting page, click [LCD Setting] to LCD Setting page. Click [Ok],it will exit and save the current setting; click [Cancel], it will exit without saving the setting info. LCD Setting page is shown below:
Fig. 3.4.3.3-a LCD Setting page
Page 45
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-23
The setting items of LCD Setting are described below:
Brightness Press the【Trackball+SETkey to select
brightness slider position.
Default To export the factory default LCD Setting.
3.4.5 BodyMark
Press the【BodyMark】key to active the BodyMark page.
Select the desired bodymark and exit BodyMark page;
Click Exit, or press the【BodyMark】 key again to exit BodyMark screen.
The BodyMark page is shown below:
The operation of BodyMark is follows:
Add
Go to BodyMark page, click a bodymark, and the selected bodymark appears at the left-bottom of the image area; Move the 【Trackballto adjust the probe position in the bodymark; Rotate the【MENU】key to adjust the probe orientation in the bodymark; Press the SETkey to fix the probe orientation. Note: To insert a new bodymark will clear the original bodymark.
Page 46
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-24
Move Move the trackball cursor to the bodymark, and the cursor
changes to “ “, which means that bodymark is moveable.
Press the【SET】key to select the bodymark. Now a box is
enclosed around the bodymark, which means that the
bodymark can be moved with trackball movement. Move
the trackball to the target position, press the【SET】key to
move the bodymark to the new position.
Del Go to BodyMark page, click Exit, or press the【BodyMark
key to exit BodyMark page and clear the added bodymark.
3.4.6 How to use biopsy
In [image menu], click [Biospy] item; When biospy is in ’on’ state, the parameters of Position and Angle will display. Meanwhile the biopsy guide line is displayed in the imaging area.
The control menu of Biospy is shown below:
Fig. 3.4.6 Biospy control menu
The setting method for Biospy menu is shown below:
Position
Angle
Translate the biopsy, -10mm ~ +10mm selectable. Press the【SET】key to increment, or press the【SET】key to decrement; or rotate the【MENU】knob to increment or decrement, 1mm each step.
Biospy angle, -10° ~ +10° selectable. Press the【SET】key to increment, or press the【SET】key to decrement; Or rotate the MENUknob to increment or decrement, 1° each step.
Page 47
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-25
3.4.7 Report
In frozen mode, press the【Report】key to active Report page. Click [Exit], or press the【Report】key again to exit Report page.
The Report page is shown below:
Fig. 3.4.8 Report management page
The Report screen is divided into:
(1) Patient basic information area
(2) Measurement information area
(3) Calculation result information area
(4) Exam and other operation area
The operation of Report is shown below:
Change The user may change measurement value in the
measurement information area.
Page 48
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-26
NOTE:
Many relevant numbers will be changed in the report automatically.
Print Click it and the report content will be printed out .
Save
OK Click it to save the change of current report.
3.4.8 Use the peripherals
3.4.8.1 USB device
The system adopts general USB ports to connect to peripheral devices. The selectable peripheral devices include: USB disk, mobile hard drive, USB printer, USB CD writer, USB mouse, etc.
NOTE:
Due to the fact that there are many brands and models of such devices, to ensure the normal functioning of the peripheral devices, please try to use the recommended models by the manufacturer.
The operation method for USB peripherals is follows:
USB disk
Click it. In the pop-up input dialog box, input a new report name, and the system will add a default extend file name “.rpt”. Click OK to save the new report and exit the Report page; When the report is saved, the user may manage, review, recall, change or print in the Disk Management.
When a USB disk is plugged in, the status prompt area will displays a
icon. Before plug out the disk,
click the
icon. In the uninstall USB disk dialog, click OK to uninstall the USB disk, and the
icon
disappears;
Page 49
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-27
NOTE:
The USB disk can only be removed when it is uninstalledTo remove the USB disk by force during operation will damage the system files, destroy the user data, or even lead to system failure to initialize
USB printer
USB CDRW
3.4.8.2 Footswitch
Connect the special footswitch accessory to the real panel.
3.4.8.3 Video Out
The system video output format includes: VGA, S-video, composite video (NTSC/PAL);
Composite video is TV signals. It can be connected to all kinds of TV monitors. There are 2 major TV formats used worldwide: NTSC and PAL, the user may select the local format in System Setting; when it is connected to a composite video printer, connect the video signal cable and the remote control cable to the rear panel of the main unit, use the Print key on the main unit keyboard to achieve printing through a remote control video printer;
3.4.8.4 Network Device
Through the standard net port, the main unit can be connected to a DICOM system. See Section 3.4.3.1 for the detailed operation;
When a system supported USB printer is plugged in, the status prompt area will displays a
icon;
Press the 【Print】key, and a print dialog is displayed. Adjust the report format, click OK to print out the report;
When a system supported USB CD writer is plugged in, and when it is properly connected to the system, active Burning Files page to perform burning. See Section 3.4.3.2 Disk Management on how to perform burning;
Page 50
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-28
3.5 Imaging
3.5.1 Basic operation
The system supports many image modes such as B, B/B, 4B, B/M, M, and THI. Here is the operation:
B mode press the【Bkey to switch to B mode.
2B mode press the 【Dual Bkey to switch to dual B mode,
the left image is alive. Press the Dual B key again, and it alternate between left and right image. The current live image is indicated by a triangle below.
4B mode press the【Quad B key to switch to 4B mode, press
the【Quad B】key again, and it activate the upper left, upper right, bottom left, bottom right image by turn, The current live image is indicated by a triangle below.
B/M mode press theB/M modekey to switch to B/M mode,
press theB/M modekey again to switch to full M mode. The display method for M mode, B/M mode, B/M mode can be set up in System Setting.
THI mode press the【THIkey to switch to THI (Tissue
Harmonic Imaging) mode, press THI again to exit.
Vertical invert Press the 【Up/Down key to invert the
current active image vertically, and the small dot marking the image orientation is inverted as well.
Horizontal invert Press the【L/Rkey to invert the current
active image horizontally, and the small dot marking the image orientation is inverted as well.
Page 51
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-29
3.5.2 Parameter Adjust
For the system, the adjustable imaging parameters include:
Focus The transmit focus numbers can be selected: 1~4.
Depth To select transmit focus depth position.
DYN To select system dynamic range, 10 steps
selectable.
GSC To select grayscale curve, 16 curves selectable.
Enhance To enhance image edge, 8 steps selectable.
Persist Persistence processing, 8 steps selectable.
Line Density To select scan line density, High/Low, and the
default is High.
SPAN To select scan angle/width.
Biospy To set up biopsy setting, and the default is OFF; if it
is set ON, two lines of indented menu are displayed:
Position Biospy translate -10+10mm selectable, the increment is 1mm.
Angle Biopsy angle-10~+10° selectable, the increment is 1°.
Freq To select multiple frequencies, 5 steps selectable.
Rotate To select B mode image orientation, which
corresponds to 0°, 90°, 180°, 270°.
Chroma To select the chroma code for the ultrasound image;
Power To select acoustic power, 8 steps.
ZOOM To activate image zoom function.
Gain To adjust image gain in B or M mode.
TGC 8-step TGC, to adjust image depth compensation
gain. The adjusted parameter curve is displayed in the TGC curve display area, which disappears automatically in 3 seconds.
Page 52
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-30
Fig. 3-5-2 Image Menu
The operation of special knobs is follows:
Image display depth, focus position, image zoom function. Press the【FUNC】key, in turn it activates image display depth, focus position, image zoom function, the current function state is displayed in the system prompt area. Rotate the【FUNC】knob to adjust focus position and image depth. image For Zoom adjust, use the trackball and the SETkey, and the specific steps are shown below;
Press the【FUNC】key, select image zoom function; now in the image area it displays box, which means that it is a region that can be zoomed in. Rotate the【FUNC】knob to select the required zoom factor, now the size of the box changes; Move the trackball to to region to be zoomed in, press the【SET】key to go to zoom mode. Then rotate the FUNCknob to change the current image zoom factor; If the zoom factor is less than 1.0, the system is displayed full ; in zoom mode, press the【FUNC】key again or ESC, it will exit the zoom mode, and the image returns to normal mode.
Page 53
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-31
B mode, M mode image gain adjust. Rotate the【GAIN】 knob to perform image gain adjust; in B/M mode, press the FUNCkey to switch between B-gain and M-gain. In freeze mode, adjust the the【FUNC】knob will change the LCD brightness.
To adjust the selected parameter in the image menu. In live imaging mode, press the 【MENU】key to activate or exit mage adjust menu. In the image menu, use the cursor to select the parameter item, and that item is highlighted. Rotate the【MENU】knob to adjust that parameter. In exam mode screen, press the【MENU】key to activate System Setting menu.
Page 54
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-32
3.6 Measurement
The frame of measurement menu in B mode is as shown in Fig. 3.6.1.
Fig. 3.6.1 B mode measurement menu
Page 55
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-33
The system has included many measurement formulas and forms. Due to differences in race characteristics worldwide, the user may select an appropriate formula to perform measurement based on local race data, or to input local race data in the form of user-defined data tabulation. See Section
3.4.3.4 Measure Setting for details;
3.6.1 B mode basic measurement
Basic measurement includes:
Distance measurement
Angle measurement
Ratio measurement (distance ratio, area ratio)
Area/Cir measurement (Ellipse, Trace)
Volume measurement (2-Axis, 3-Axis)
3.6.1.1 Distance Measurement
Function: Distance Measurement will measure the distance between 2
points.
Measurement method:
1. In B mode frozen state, press the【Distance】key to go to distance measure mode, or press the 【Calc】key to pop up the Meas & Calc menu, move the trackball, use the cursor to select “Distance “ , press the 【SET】key to go to distance measure mode.
2. The cursor in the image area is displayed as a “+”, Use the trackball to shift the cursor to the measurement start point , press the 【SET】key, and it will display a fix mark “ X”, and it displays the measurement ruler number as the current one.
3. Use the trackball to move the cursor, the cursor “+” and the the start point are connected by a dotted line between the marks “X”. In the result window it displays
Page 56
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-34
3.6.1.2 Angle measurement
Function: To measure the inclination angle between 2 straight lines
(0~180°).
Measurement method:
4. Use the trackball to shift the cursor “+” to the measured
end point, press the 【SET】key, on the measurement end point it displays the fix mark “X”, the measurement result is fixed, and it ends the measurement.
5. Repeat steps 2~4 to go to the next Distance Measurement.
1. In B mode frozen state, press the 【Calc】key
to pop up the Meas & Calc menu, move the trackball, use the cursor to select “Angle”, press the 【SET】key to go to angle measurement mode.
2. Draw a section along one side of the measurement angle. See “Distance Measurement” for the measurement method.
3. Draw a section along the other side of the measurement angle. See “Distance Measurement” for the measurement method. The inclination between the two sections and the length of the sections are displayed in the result window.
4. Repeat steps 2 to 3 to perform the next measurement.
Page 57
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-35
3.6.1.3 Ratio measurement
Ratio measurement includes distance ratio and area ratio.
A Distance ratio
Function: To measure and calculate the ratio of two distance measurement
values. The first measured value is the numerator, and the next measured value is the denominator.
Measurement method:
B Area ratio
Function: To measure and calculate the ratio of two area measurement
values. The first measured value is the numerator, and the next measured value is the denominator.Use the Ellipse method to perform the area measurement, for the Ellipse method,see the section 3.6.1.4 for detail.
1. In B mode frozen state, press the 【Calc】key to
pop up the Meas & Calc menu, move the trackball, use the cursor to select “Ratio”, press the 【SET】 key to go to Ratio measurement mode.
2. Measure the first distance. The measurement method is the same as “Distance Measurement”.
3. Measure the second distance. Move the cursor to the image window area, and the cursor is displayed as a “+”. Use the trackball to shift the cursor to measurement the start point, press the SETkey, and on the measure start point it displays the fix mark “x”. Use the trackball to move the cursor, between “+” and “x” there is a connecting dotted line. In the result window it displays measurement value and calculation result in real-time.
4. Press the 【SET】key, the measurement ends, and the section length and “Ratio” calculation result is displayed in the result window.
5. Repeat steps 2 to 4 to perform the next measurement.
Page 58
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-36
3.6.1.4 Area/Circumference measurement (Ellipse, Trace)
Area/Circumference measurement includes Ellipse and Trace.
AEllipse
Function: Use an approximate ellipse to measure circumference and area
of an enclosed region.
Measurement method:
1. In B mode frozen state, press the【Calc】key to
pop up the Meas & Calc menu. Move the trackball, use the cursor to select “Area/Cir”, press the 【SET】key to go to the submenu. Select “Ellipse”, press the 【SET】key to go to “Ellipse” measurement mode.
2. Move the cursor to the start point of the fixing axis of the ellipse measurement area. Press the SETkey, and on the start point of the fixing axis it displays the mark “+”.
3. Move the cursor to the end point of the fixing axis of the ellipse measurement area. Between the cursor “+” and “x” there is a connecting dotted line. Press the 【SET】key, on the end point of the fixing axis, it displays the mark “x”, and on the screen it displays an ellipse.
4. Move the trackball to adjust the variable axis length of the ellipse to match the ellipse with the measured region. Move the trackball leftward, and the variable ellipse axis is getting small, Move the trackball rightward, and the variable ellipse axis is getting bigger.
5. Press the【SET】key to confirm the ellipse measurement area, and the result is displayed in the result window, and it ends the measurement.
6. Repeat steps 2 to 5 to perform the next measurement.
Page 59
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-37
a: Press Set to confirm the start point of the ellipse axis
b: Move theTrackballto shift the cursor to the end point of the ellipse
axis
c: Press theSETkey to confirm the end point of the ellipse axis
d: Move theTrackball to select the ellipse area
e: Press theSETkey to confirm the ellipse area
BTrace
Function: Use Trace measurement to measure circumference and area of
an enclosed region.
Measurement method:
1 In B mode frozen state, press the 【Calc】key to
pop up the Meas & Calc menu, move the trackball, use the cursor to select “Area/Cir”, press the 【SET】key to go to the submenu to select “Trace”. Press the【SET】key to go to “Trace” measurement mode.
2 Move the cursor to the measurement start point ,
press the【SET】key, on the measurement start point it displays the fix mark “ x”.
3 Move the trackball to enable the cursor to move
along the edge of the target area, and a trace is drawn.
4 Press the【SET】key, and the trackball start point
and the end point are enclosed by a straight line; or when the cursor and the trace start point distance are very close, the trace is enclosed automatically. The measurement result is displayed in the result window, and it ends the measurement.
5 Repeat steps 2 to 4 to perform the next
measurement.
Page 60
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-38
NOTE:
Always avoid any intersection of trace when performing Ttrace method, so as to prevent calculation error.
3.6.1.5 Volume measurement (2-Axis, 3-Axis)
Volume measurement consists of 2-Axis and 3-Axis.
A2-Axis
Function: Based on the perpendicular section of the measured object, to
calculate the volume of the measured object.
Principle: Formula for 2-Axis:
V = (π/6) × A × B
2
Where A is the long axis of the ellipse, and B is the short axis of the ellipse.
Measurement method:
1 In B mode frozen state, press the 【Calc】key
to pop up the Meas & Calc menu, move the trackball, use the cursor to select “volume”, press the【SET】key to go to the submenu, move the trackball to select “2-Axis”, press theSETkey to go to “2-Axis volume” measurement mode.
2 Move the cursor to the start point of the fixing
axis of the ellipse measurement area. Press theSETkey, and on the fixing axis start point , it displays a mark “+”.
3 Move the cursor to the end point of the fixing
axis of the ellipse measurement area. Between the cursor “+” and “x” there is a connecting dotted line. Press the【SET】key, on the fixing axis end point it displays the mark “x”, and on the screen it displays an ellipse.
Page 61
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-39
B3-Axis
Function: Based on the vertical section and the horizontal section of the
measured object, the object volume is calculated.
Principle: The volume can be estimated from the orthogonal D1, D2 and D3 values of the object.
The formula:
V = (π/6) × D1 × D2 × D3
4. Move the trackball to adjust the variable axis length of the ellipse to match the ellipse with the measured region. Move the trackball leftward, and the variable ellipse axis is getting small, Move the trackball rightward, and the variable ellipse axis is getting bigger.
5. Press the【SET】key to confirm the ellipse measurement area, and the result is displayed in the result window, and it ends the measurement.
6. Repeat steps 2 to 5 to perform the next measurement.
Page 62
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-40
Measurement method:
3.6.2 B mode special measurement
B mode special measurement includes:
Basic OB measurement
Applicable OB measurement
GYN measurement
Cardiac measurement
Urologic measurement
Small Part measurement
1. In B mode, scan and freeze the image, press
the Calckey to pop up the Meas & Calc menu, move the trackball, use the cursor to select “volume”, press the【SET】key to go to the submenu, move the trackball to select “3-Axis”, go to “3-Axis volume” measurement mode.
2. On the screen based on the size of the measured region an ellipse is drawn. See “Area/Cir” with ellipse for the ellipse drawing method.
3. Unfreeze the image; rescan to acquire a section orthogonal to the previous image. Freeze it, and measure the length of the 3rd axis. The measurement method is the same as that for “Distance Measurement”.
4. After ending the measurement, the measurement result for volume is displayed in the result window.
5. Repeat steps 2 to 4, to go to the next 3-Axis volume measurement.
Page 63
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-41
3.6.2.1 Basic OB measurement
3.6.2.1.1 GS (Gestational Sac) measurement
Function: Calculate GA (gestational age) based on GS.
Measurement method:
1. In B mode frozen state, press the【Calc】key to
pop up the Meas & Calc menu. Move the trackball; use the cursor to select “OB Basic”. Press the【SET】key to go to the submenu. Move the trackball to select “GS”. Press the SETkey to go to “GS” measurement mode.
2. Measure GS. See “Distance Measurement” for the detailed method.
3. The GS measurement result is displayed in the result window, meanwhile based on the preset formula GA is displayed.
4. Repeat steps 1 to 3 to go to the next GS measurement.
Page 64
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-42
3.6.2.1.2 CRL (Crown Rump Length) measurement
Function: Calculate GA based on CRL.
Measurement method:
1. In B mode, scan and freeze the image, press
theCalckey to pop up the Meas & Calc menu, move the trackball, use the cursor to select “OB Basic”, press the【SET】key to go to the submenu, move the trackball to select “CRL”, press the【SET】key to go to “CRL” measurement mode.
2. Measure CRL. See “Distance Measurement” for the detailed method.
3. CRL measurement result is displayed in the result window, meanwhile based on the preset formula GA is displayed.
4. Repeat steps 1 to 3 to go to the next CRL measurement.
Page 65
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-43
3.6.2.1.3 BPD (Biparietal Distance) measurement
Function: When the fetus is between 12 to 20 weeks, calculate GA based
on BPD.
Measurement method:
3.6.2.1.4 HC (Head Circumference) measurement
Function: Calculate GA based on HC.
Measurement method:
1. In B mode, scan and freeze the image, press
theCalckey to pop up the Meas & Calc menu, move the trackball, use the cursor to select “OB Basic”, press the【SET】key to go to the submenu, move the trackball to select “BPD”, press the【SET】key to go to “BPD” measurement mode.
2. Measure BPD. See “Distance Measurement” for the detailed method.
3. BPD measurement result is displayed in the result window, meanwhile based on the preset formula GA is displayed.
4. Repeat steps 1 to 3 to go to the next BPD measurement.
1. In B mode, scan and freeze the image, press theCalc】key to pop up the Meas & Calc menu, move the trackball, use the cursor to select “OB Basic”, press the【SET】key to go to the submenu, move the trackball to select “HC”, press the SETkey to go to “HC” measurement mode.
2. Measure HC. See “Distance Measurement” for the detailed method.
3. HC measurement result is displayed in the result window, meanwhile based on the preset formula GA is displayed.
4. Repeat steps 1 to 3 to go to the next HC measurement.
Page 66
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-44
3.6.2.1.5 AC (Abdominal Circumference) measurement
Function: Calculate GA based on AC.
Measurement method:
1. In B mode, scan and freeze the image, press
theCalckey to pop up the Meas & Calc menu, move the trackball, use the cursor to select “OB Basic”, press the【SET】key to go to the submenu, move the trackball to select “AC”, press the【SET】key to go to “AC” measurement mode.
2. Measure AC. See “Distance Measurement” for the detailed method.
3. AC measurement result is displayed in the result window, meanwhile based on the preset formula GA is displayed.
4. Repeat steps 1to 3 to go to the next AC measurement.
Page 67
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-45
3.6.2.1.6 FL (Femur Length) measurement
Function: Calculate GA based on FL.
Measurement method:
1 In B mode, scan and freeze the image, press
the Calckey to pop up the Meas & Calc menu, move the trackball, use the cursor to select “Basic OB”, press the【SET】key to go to the submenu, move the trackball to select “FL”, press the【SET】key to go to “AC” measurement mode.
2 Measure FL. See “Distance Measurement” for
the detailed method.
3 FL measurement result is displayed in the
result window, meanwhile based on the preset formula GA is displayed.
4 Repeat steps 1 to 3 to go to the next FL
measurement.
Page 68
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-46
3.6.2.1.7 OFD (Occipitofrontal Diameter) measurement
Function: Only when it is difficult to measure BPD correctly, calculate GA
based on OFD.
Measurement method:
1. In B mode, scan and freeze the image, press theCalckey to pop up the Meas & Calc menu, move the trackball, use the cursor to select “OB Basic”, press the【SET】key to go to the submenu, move the trackball to select “OFD”, press the【SET】key to go to “OFD” measurement mode.
2. Measure OFD. See “Distance Measurement” for the detailed method.
3. OFD measurement result is displayed in the result window, meanwhile based on the preset formula GA is displayed.
4. Repeat steps 1 to 3 to go to the next OFD
measurement.
Page 69
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-47
3.6.2.1.8 CER measurement
Function: Calculate GA based on FT
Measurement method:
3.6.2.1.9 AFI (Amniotic Fluid Index) measurement
Function:
Evaluate the fetal status based on AFI.
Measurement the formula:
AFI=AFD1 (depth1) + AFD2 (depth2) + AFD3 (depth3) +
AFD4 (depth4)
AFD1, AFD2, AFD3, AFD4 refer to amniotic fluid depths in 4 parts.
1. In B mode, scan and freeze the image, press
theCalckey to pop up the Meas & Calc menu, move the trackball, use the cursor to select “OB Basic”, press the【SET】key to go to the submenu, move the trackball to select “CER”, press the【SET】key y to go to “CER” measurement mode.
2. Measure CER. See “Distance Measurement” for the detailed method.
3. CER measurement result is displayed in the result window, meanwhile based on the preset formula GA is displayed.
4. Repeat steps 1 to 3 to go to the next CER measurement.
Page 70
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-48
Measurement method:
3.6.2.1.10 Report
See “3.4.8 Report”.
1. In B mode, scan and freeze the image, press
theCalckey to pop up the Meas & Calc menu, move the trackball, use the cursor to select “OB Basic”, press the【SET】key to go to the submenu, move the trackball to select “AFI”, press the【SET】key to go to “AFI” measurement mode.
2. Measure AFD1. See “Distance Measurement” for the detailed method.
3. Unfreeze the image, rescan the image, freeze, measure AFD2, the measurement method is the same as “Distance Measurement”.
4. Unfreeze the image, rescan the image, freeze, measure AFD3, the measurement method is the same as “Distance Measurement”.
5. Unfreeze the image, rescan the image, freeze, measure AFD4, the measurement method is the same as “Distance Measurement”.
6. The amniotic fluid depths in 4 parts are calculated, and the AFI result is displayed in the result window.
7. Repeat steps (1) to (6) to go to the next AFI measurement.
Page 71
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-49
3.6.2.2 High-level OB measurement
3.6.2.2.1 APAD measurement
Function: Calculate GA based on APAD.
Measurement method:
1. In B mode, scan and freeze the image, press
theCalckey to pop up the Meas & Calc menu, move the trackball, use the cursor to select “OB Prof”, press the【SET】key to go to the submenu, move the trackball to select “APAD”, press the【SET】key to go to “APAD” measurement mode.
2. Measure APAD. See “Distance Measurement” for the detailed method.
3. APAD measurement result is displayed in the result window, meanwhile based on the preset formula GA is displayed.
4. Repeat steps (1) to (3) to go to the next APAD measurement.
Page 72
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-50
3.6.2.2.2 AA (Abdominal Area) measurement
Function: Calculate GA based on AA.
Measurement method:
1. In B mode, scan and freeze the image, press
theCalckey to pop up the Meas & Calc menu, move the trackball, use the cursor to select “OB Prof”, press the【SET】key to go to the submenu, move the trackball to select “AA”, press the【SET】key to go to “AA” measurement mode.
2. Measure AA; see “Area/Cir measurement (Ellipse) “for the detailed method.
3. AA measurement result is displayed in the result window, meanwhile based on the preset formula GA is displayed.
4. Repeat steps 1 to 3 to go to the next AA measurement.
Page 73
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-51
3.6.2.2.3 HL (Humerus Length) measurement
Function: Calculate GA based on HL.
Measurement method:
1. In B mode, scan and freeze the image, press
theCalckey to pop up the Meas & Calc menu, move the trackball, use the cursor to select “OB Prof”, press the【SET】key to go to the submenu, move the trackball to select “HL”, press the【SET】key to go to “HL” measurement mode.
2. Measure HL. See “Distance Measurement” for the detailed method.
3. HL measurement result is displayed in the result window, meanwhile based on the preset formula GA is displayed.
4. Repeat steps 1 to 3 to go to the next HL measurement.
Page 74
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-52
3.6.2.2.4 THDThoracic Diametermeasurement
Function: Calculate GA based on THD.
Measurement method:
1. In B mode, scan and freeze the image, press
theCalckey to pop up the Meas & Calc menu, move the trackball, use the cursor to select “OB Prof”, press the【SET】key to go to the submenu, move the trackball to select “THD”, press the【SET】key to go to “THD” measurement mode.
2. Measure THD. See “Distance Measurement” for the detailed method.
3. THD measurement result is displayed in the result window, meanwhile based on the preset formula GA is displayed.
4. Repeat steps 1 to 3 to go to the next THD measurement.
Page 75
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-53
3.6.2.2.5 FTA (Fetal Trunk Area) measurement
Function: Calculate GA based on FTA.
Measurement method:
3.6.2.2.6 OOD measurement
Function: Calculate GA based on OOD..
Measurement method:
1. In B mode, scan and freeze the image, press
theCalckey to pop up the Meas & Calc menu, move the trackball, use the cursor to select “OB Prof”, press the【SET】key to go to the submenu, move the trackball to select “FTA”, press the【SET】key to go to “FTA” measurement mode.
2. Measure FTA. See “Distance Measurement” for the detailed method.
3. FTA measurement result is displayed in the result window, meanwhile based on the preset formula GA is displayed.
4. Repeat steps 1 to 3 to go to the next THD measurement.
1. In B mode, scan and freeze the image, press theCalckey to pop up the Meas & Calc menu, move the trackball, use the cursor to select “OB Prof”, press the【SET】key to go to the submenu, move the trackball to select “OOD”, press the【SET】key to go to “OOD” measurement mode.
2. Based on the pre-selected formula, if the corresponding measurement item is measured, then OOD is calculated automatically, and the result is displayed in the result window.
Page 76
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-54
3.6.2.2.7 TIBIA measurement
Function: Calculate GA based on TIBIA.
Measurement method:
3.6.2.2.8 ULNA measurement
Function: Calculate GA based on ULNA.
Measurement method:
1 In B mode, scan and freeze the image, press
the【Calc】key to pop up the Meas & Calc menu, move the trackball, use the cursor to select “OB Prof”, press the【SET】key to go to the submenu, move the trackball to select “TIBIA.”, press the【SET】key to go to “TIBIA.” measurement mode.
2 Measure TIBIA. See “Distance Measurement”
for the detailed method.
3 TIBIA measurement result is displayed in the
result window, meanwhile based on the preset formula GA is displayed.
4 Repeat steps 1 to 3 to go to the next TIBIA
measurement.
1 In B mode, scan and freeze the image, press
theCalc key to pop up the Meas & Calc menu, move the trackball, use the cursor to select “OB Prof”, press the【SET】key to go to the submenu, move the trackball to select “ULNA.”, press the【SET】key to go to “ULNA” measurement mode.
2 Measure ULNA. See “Distance Measurement”
for the detailed method.
3 ULNA measurement result is displayed in the
result window, meanwhile based on the preset formula GA is displayed.
4 Repeat steps 1 to 3 to go to the next ULNA
measurement.
Page 77
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-55
3.6.2.2.9 Input Last Menstrual Period (Input LMP)
Input the patient’s LMP, so as to calculate GA and the fetal delivery date.
3.6.2.2.10 Fetal Growth Curve (FGC)
Function:
FGC function is to compare the fetal measurement data with a normal FGC, so as to evaluate if the fetus grows properly. Generally it is determined by the inputted LMP, the measured data and the fetal growth indices.
Operation instruction:
1. Measure one or more fetal growth indices (GS, CRL, BPD, FL, AC, HC, AA, APAD, TAD, CER, FT, FTA, HA, HUM, OFD, THD).
2. Input LMP. In B mode, scan and freeze the image, press the 【Calc】key to pop up the Meas & Calc menu, move the trackball, use the cursor to select“OB Prof”, press the【SET】 key to go to the submenu, move the trackball to select “FGC”, press the【SET】key to go to FGC mode.
3. The default display for FGC is “BPD” screen; on its right the formula displayed is the basic formula of the current growth curve. Move the cursor to other formula item, press the【SET】 key to display the normal growth curve based on the algorithm of the formula, so as to evaluate the fetal growth.
4. The FGC displays the growth status of a certain index of the fetus, which is displayed in the image as a measurement value corresponding to the position on a growth curve (marked with “X” in the image).
5. The FGC can be displayed in one chart one screen, or four charts one screen. Press the matched button to setup. Press”Close” to exit FGC.
Page 78
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-56
Fig. 3.6.2.2.9 Fetal Growth Curve (one graph/ four graphs)
3.6.2.2.11 Fetal Biophysical Profile (FBP)
When conducting FBP, acquire indices of fetal growth through testing or measurement, and use the grading method to evaluate if the fetus is at risk, which serves as a guiding index in clinical processing.
Page 79
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-57
The FBP testing and measurement shall cover the following 6
indices:
Fetal Heart Rate (FHR)
Fetal Motion (FM)
Fetal Breathing Motion (FBM)
Fetal Tension of muscle (FT)
Amniotic Fluid (AF)
Placenta grading (PL)
Where in AF is acquired by measuring amniotic fluid depth; PL is based on placenta ultrasonic image, for which the placenta maturity is divided into 3 levels: Level 0,1,2.,3.
While the other indices are obtained through fetal reaction test, the fetal reaction test takes about 20-30 minutes.
Grading standard:
1) Fetal Heart Rate (FHR)
Observation duration: 20 minutes
Grading standard:
2: When fetal moves, FHR accelerates>=15 times/min,
duration>=15S, >=5 times;
1: When fetal moves, FHR accelerates >=15 times/min, duration
>=15S, 14 times;
0: When fetal moves, FHR accelerates <=1 time.
2) Fetal Motion (FM)
Observation duration: 30 minutes
Grading standard:
2: FM>=3 times
1: FM 12 times
0: No fetal movement
Page 80
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-58
3) Fetal Breathing Motion (FBM)
Observation duration: 30 minutes
Grading standard:
2: FBM>=1 time, duration >=60S
1: FBM>=1 time, duration 3060S
0: No FBM, or duration <=30S
4) Fetal Tension of muscle (FT)
Observation duration: 30 minutes
Grading standard:
2: >=1 time, extension and bending of extremities and the spine
1: >=1 time, extension and bending of extremities or the spine
0: extension of extremities, no bending, with hands open
5) Amniotic Fluid (AF)
Grading standard:
2: AF max. Depth>2cm
1: AF max. depth1~2cm
0: AF max. Depth<1 cm
6) Placenta grading (FL)
Grading standard:
2: PL<=2
1: back wall placenta, difficult to determine the PL
0: PL=3
Based on single indices the fetal growth status is acquired as below:
2: Normal
1: Slightly abnormal
0: Obviously abnormal
Page 81
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-59
Total mark and general fetal growth status:
7~12 Normal fetus, with low risk of chronic suffocation
3~6 Suspicious of fetal chronic suffocation
0-2 Highly suspicious of fetal chronic suffocation.
Measurement method:
1. In B mode scan the image and freeze the image, press the 【Calc】key to pop up the Meas & Calc menu, move the trackball, use the cursor to select “GYN” soft key, press the SETkey to go to the submenu, move the trackball to select “FBP”, press the【SET】key to go to “FBP” mode.
2. go to the fetal grading FBP menu; there are six items: FHR, FM, FBM, FT, PL, AF , select any of them from the drop-down box, next to each item you can select 0, 1, 2; When it is finished, select Cancel to quit the selected grade and exit; select Report and go to FBP to report the screen. The reports displays grade for each item, total grade, a brief evaluation. Select Print to print it out ; select Exit to exit;
Page 82
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-60
Fig. 3.6.2.2.10 FBP Grade Report
3.6.2.3 GYN measurement
GYN measurement includes:
Uterus measurement (UT)
Endometrium thickness measurement (ENDO)
Left ovary measurement (L.OV)
Right ovary measurement (R.OV)
Left follicle measurement (L.FO)
Right follicle measurement (R.FO)
Cervix long diameter measurement (CX)
Cervix length / Uterus length (CX-L /UT-L)
Page 83
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-61
3.6.2.3.1 UT measurement
Function: Measure length, height and width of uterus in sequence
UT-L Uterus length (mm)
UT-H Uterus height (mm)
UT-W Uterus width (mm)
UT = UT-L + UT-H + UT-W
Measurement method:
1. In B mode, scan and freeze the image, press theCalckey to pop up the Meas & Calc menu, move the trackball, use the cursor to select “GYN”, press the【SET】key to go to the submenu, move the trackball to select “UT”, press the【SET】key to go to “UT” measurement mode.
2. Measure UT-L, see “Distance Measurement” for the specific method.
3. Measure UT-H, see “Distance Measurement” for the specific method.
4. Measure UT-W, see “Distance Measurement” for the specific method
5. UT = UT-L + UT-H + UT-W, the measurement & calculation result is displayed in the result window.
6. Repeat steps 1 to 5 to go to the next UT measurement.
Page 84
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-62
3.6.2.3.2 Endo (endometrium thickness) measurement
Function: Endo measurement is to measure endometrium thickness.
Measurement method:
1. Press the 【Calc】key to pop up the Meas &
Calc menu, move the trackball, use the cursor to select “GYN”, press the【SET】key to go to the submenu, move the trackball to select “Endo”, press the【SET】key to go to “Endo” measurement mode.
2. Measure Endo; see “Distance Measurement” for the specific method.
3. The measurement result is displayed in the result window.
4. Repeat steps 1 to 3 to go to the next Endo measurement.
Page 85
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-63
3.6.2.3.3 L.Ovary (Left Ovary) measurement
Function: Based on left ovary length, height and width: Lt L, Lt H, Lt W,
to calculate left ovary volume: L.OV-V = 0.523 * Lt Lmm * Lt Hmm * LtWmm
Measurement method:
1. Press the 【Calc】key to pop up the Meas &
Calc menu, move the trackball, use the cursor to select “GYN”, press the【SET】key to go to the submenu, move the trackball to select “L.Ovary”, press the【SET】key to go to “L.Ovary” measurement mode.
2. Measure Lt L of left ovary, sees “Distance Measurement” for the specific method.
3. Measure Lt H of left ovary, sees “Distance Measurement” for the specific method.
4. Measure Lt W of left ovary, sees “Distance Measurement” for the specific method.
5. L.Ovary volume L.OV-V = 0.523 * Lt Lmm * Lt Hmm * Lt Wmm, and the measurement & calculation result is displayed in the result window.
6. Repeat steps 1 to 5 to go to the next L.Ovary volume measurement.
Page 86
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-64
3.6.2.3.4 R.Ovary (Right Ovary) measurement
Function: Based on right ovary length, height and width: Rt L, Rt H, Rt
W, to calculate right ovary volume: R.OV-V = 0.523 * Rt Lmm * Rt Hmm * Rt Wmm
Measurement method:
1. Press the 【Calc】key to pop up the Meas &
Calc menu, move the trackball, use the cursor to select “GYN”, press the【SET】key to go to the submenu, move the trackball to select “R.Ovary”, press the【SET】key to go to “R.Ovary” measurement mode.
2. Measure Rt L of right ovary; see “Distance Measurement” for the specific method.
3. Measure Rt H of right ovary; see “Distance Measurement” for the specific method.
4. Measure Rt W of right ovary; see “Distance Measurement” for the specific method.
5. R.Ovary volume R.OV-V = 0.523 * Rt Lmm * Rt Hmm * Rt Wmm, and the measurement & calculation result is displayed in the result window.
6. Repeat steps 1 to 5 to go to the next R.Ovary volume measurement.
Page 87
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-65
3.6.2.3.5 L.Follicle (Left Follicle) measurement
Function: To measure length and width of left follicle: L.FO L, L.FO W
Measurement method:
1. press the 【Calc】key to pop up the Meas &
Calc menu, move the trackball, use the cursor to select “GYN”, press the【SET】key to go to the submenu, move the trackball to select “L.Follice”, press the【SET】key to go to “L.Follice” measurement mode
2. Measure L.FO L of left follicle; see “Distance Measurement” for the specific method.
3. Measure L.FO W of left follicle; see “Distance Measurement” for the specific method.
4. The measurement result is displayed in the result area.
5. Repeat steps 1 to 4 to go to the next L.Follice measurement.
Page 88
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-66
3.6.2.3.6 R. Follicle (Right Follicle) measurement
Function: To measure length and width of right follicle: R.FOL, R.FOW.
Measurement method:
1. Press the 【Calc】key to pop up the Meas &
Calc menu, move the trackball, use the cursor to select “GYN”, press the【SET】key to go to the submenu, move the trackball to select “R. Follicle, press theSETkey to go to R. Follicle measurement mode.
2. Measure R.FO L of right follicle; see “Distance Measurement” for the specific method.
3. Measure R.FO L of right follicle; see “Distance Measurement” for the specific method.
4. The measurement result is displayed in the result area.
5. Repeat steps (1) to (4) to go to the next R. Follicle measurement.
Page 89
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-67
3.6.2.3.7 CX (Cervix) measurement
Function: Measure cervix length CX-L, height CX-H and width CX-W in
sequence.
Measurement method:
1. Press the 【Calc】key to pop up the Meas &
Calc menu, move the trackball, use the cursor to select “GYN”, press the【SET】key to go to the submenu, move the trackball to select “Cervix”, press the【SET】key to go to “Cervix” measurement mode.
2. Measure cervix length CX-L; see “Distance Measurement” for the specific method.
3. Measure cervix height CX-H; see “Distance Measurement” for the specific method.
4. Measure cervix width CX-W; see “Distance Measurement” for the specific method.
5. The measurement result is displayed in the result window.
6. Repeat steps 1 to 5 to go to the next Cervix measurement.
Page 90
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-68
3.6.2.3.8 CX-L /UT-L
Function: To measure uterus length UT-L and cervix length CX-L, then
to calculate their ratio UT-L/CX-L.
Measurement method:
3.6.2.4 Cardiac measurement
In B mode Cardiac measurement includes:
Left Ventricle (LV)
Right Ventricle (RV)
Pulmonary Artery(PA)
Left Ventricle Mass (LV Mass)
Left Ventricle Mass Weight Index (LVMWI)
1. In B mode freeze the image, press the 【Calc】
key to pop up the Meas & Calc menu, move the trackball, use the cursor to select “GYN”, press the【SET】key to go to the submenu, move the trackball to select “CX-L /UT-L”, press the【SET】key to go to “CX-L /UT-L” measurement mode.
2. Based on ”Distance Measurement” method, measure cervix length CX-L
3. Based on ”Distance Measurement” method, measure cervix height CX-H
4. The measurement result is displayed in the result area. Repeat steps (1) to (2) to perform the next measurement.
Page 91
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-69
3.6.2.4.1 LV function
The measurement items for LV function are:
Left Ventricular Area of Long axis diastolic LVALd
Left Ventricular Length of long axis diastolic LVLd
Left Ventricular Area of Long axis systolic LVALs
Left Ventricular Length of long axis systolic LVLs
Left Ventricular Area of short axis at Mitral valve
diastolic
LVAMd
Left Ventricular Diameter of short axis diastolic LVIDd
Left Ventricular Area of short axis at Mitral valve
systolic
LVAMs
Left Ventricular Diameter of short axis systolic LVIDs
Left Ventricular Area of short axis at Papilla
diastolic
LVAPd
Left Ventricular Area of short axis at Papilla
systolic
LVAPs
Calculation items and formulas involved are:
Calculation item Formula Unit
End-Diastolic Right
Ventricular Volume (EDV)
EDV ML
End-Systolic Right
Ventricular Volume (ESV)
ESV ML
Stroke Volume (SV) SV=EDV-ESV ML
Cardiac =Output(CO) CO=SV×HR L/Min
Ejection Fraction (EF) EF=SV/EDV No unit
Stroke Index (SI) SI=SV/BSA No unit
Cardiac Index (CI) CI=CO/BSA L/(Min·m2)
Fractional Shortening of RV
short axis (FS)
FS=(LVIDd-LVIDs) /LVIDd
Mean Ventricular
Circumference Fractiional
Shortening (MVCF)
MVCF=
(LVIDd-LVIDs)/(LVIDd×ET)
Page 92
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-70
Where
BSA=WT0.425×HT0.725×73.58/1000 (Estern, WT KG, HT CM)
BSA=WT0.425×HT0.725×71.84/1000 (Western, WT KG, HT CM)
In B mode right ventricular measurement & calculation methods are:
Single-Plane (SPEllipse)
Bi-Plane (BPEllipse)
Bullet (Bullet)
Simpson (Simpson)
z Single-Plane (SPEllipse)
Measurement items: LVALd , LVL d, LVAL s, LVL s
Calculation items: EDV, ESV, SV, CO, EF, SI, CI
Use the formula:
EDV=
LVLd
LVALd
×
×
π
2
3
8
ESV=
LVLs
LVALs
×
×
π
2
3
8
Page 93
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-71
Measurement method:
1. In B mode freeze the image, press the 【Calc】
key to pop up the Meas & Calc menu, move the trackball, use the cursor to select “Cardiac”, press the【SETkey to go to the submenu, move the trackball to select “LV”, press the【SET】key to go to the submenu, move the trackball to select “Single-Plane”, press the【SET】key to go to measurement state.
2. Measure LVALd, see “Area/Cir measurement (Ellipse) “for the specific measurement method.
3. Measure LVLd, see “Distance Measurement” for the specific method.
4. Unfreeze the image, rescan the image, and freeze the image.
5. Measure LVALs, see “Area/Cir measurement (Ellipse) “for the specific measurement method.
6. Measure LVLs, see “Distance Measurement” for the specific method.
7. Based on the formula you can get EDV, ESV, SV, CO, EF, SI, and CI. The measurement and calculation result is displayed in the result window.
8. Repeat steps 1 to 7 to perform the next measurement.
Page 94
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-72
z Bi-Plane (BPEllipse)
Measurement items: LVALd , LVAMd, LVI Dd, LVA Ls, LVAMs, LVID s;
Calculation items: EDV, ESV, SV, CO, EF, SI, CI
Use the formula:
EDV=
LVIDd
LVAMdLVALd
×
×
×
π
3
8
ESV=
LVIDs
LVAMsLVALs
×
×
×
π
3
8
Measurement method:
1 In B mode freeze the image, press the 【Calc】
key to pop up the Meas & Calc menu, move the trackball, use the cursor to select “Cardiac”, press the【SET】key to go to the submenu, move the trackball to select “LV”, press the SETkey to go to the submenu, move the trackball to select “Bi-Plane”, press the【SET】 key to go to measurement state.
2 Measure LVALd, see “Area/Cir measurement
(Trace)” for the specific measurement method.
3 Measure LVAMd, see “Area/Cir measurement
(Ellipse)” for the specific measurement method.
4 Measure LVIDd, see “Distance Measurement”
for the specific method.
5 Unfreeze the image, rescan the image, and
freeze the image.
6 Measure LVALs, see “Area/Cir measurement
(Trace)” for the specific measurement method.
7 Measure LVAMs, see “Area/Cir measurement
(Trace)” for the specific measurement method.
Page 95
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-73
z Bullet (Bullet)
Measurement items: LVAMd , LVLd, LVAM s, LVL s;
Calculation items: EDV, ESV, SV, CO, EF, SI, CI
Use the formula:
EDV=
LVAMdLVLd ××
6
5
ESV=
LVAMsLVLs××
6
5
8. Measure LVIDs, see “Distance Measurement” for the specific method.
9. Based on the formula you can get EDV, ESV, SV, CO, EF, SI, CI. The measurement and calculation result is displayed in the result window.
10. Repeat steps (1) to (9) to perform the next measurement.
Page 96
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-74
Measurement method:
1. In B mode freeze the image, press the 【Calc】
key to pop up the Meas & Calc menu, move the trackball, use the cursor to select “Cardiac”, press the【SET】key to go to the submenu, move the trackball to select “LV” , press the【SET】key to go to the submenu, move the trackball to select “Bullet”, press the SETkey to go to measurement state.
2. Measure LVAMd, see “Area/Cir measurement (Ellipse)” for the specific measurement method.
3. Measure LVLd, see “Distance Measurement” for the specific method.
4. Unfreeze the image, rescan the image, and freeze the image.
5. Measure LVAMs, see “Area/Cir measurement (Trace)” for the specific measurement method.
6. Measure LVLs, see “Distance Measurement” for the specific method.
7. Based on the formula you can get EDV, ESV, SV, CO, EF, SI, CI. The measurement and calculation result is displayed in the result window.
8. Repeat steps (1) to (7) to perform the next measurement.
Page 97
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-75
z Simpson (Simpson)
Measurement items: LVAMd , LVLd , LVAP D, LVAM s, LVLs, LVA Ps;
Calculation items: EDV, ESV, SV, CO, EF, SI, CI
Use the formula:
EDV=
(
)
LVAPdLVAMdLVAPdLVAMd
LVLd
×+×+×× 24
9
ESV=
(
)
LVAPsLVAMsLVAPsLVAMs
LVLs
×+×+×× 24
9
Measurement method:
1 In B mode freeze the image, press the 【Calc】
key to pop up the Meas & Calc menu, move the trackball, use the cursor to select “Cardiac”, press the【SET】key to go to the submenu, move the trackball to select “LV” , press the SETkey to go to the submenu, move the trackball to select “Simpson”, press the【SET】 key to go to measurement state.
2 Measure LVAMd, see “Area/Cir measurement
(Ellipse)” for the specific measurement method.
3 Measure LVLd, see “Distance Measurement”
for the specific method.
4 Measure LVAPD; see “Area/Cir measurement
(Ellipse)” for the specific measurement method.
5 Unfreeze the image, rescan the image, and
freeze the image.
6 Measure LVAMs, see “Area/Cir measurement
(Ellipse)” for the specific measurement method.
7 Measure LVLs, see “Distance Measurement”
for the specific method.
Page 98
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-76
3.6.2.4.2 Right Ventricle inner diameter (RV)
Function: To measure Right Ventricle inner diameter.
Measurement method:
8. Measure LVAPs, see “Area/Cir measurement (Ellipse)” for the specific measurement method.
9. Based on the formula you can get EDV, ESV, SV, CO, EF, SI, and CI. The measurement and calculation result is displayed in the result window.
10. Repeat steps (1) to (9) to perform the next measurement.
1. In B mode freeze the image, press the Calckey to pop up the Meas & Calc menu, move the trackball, use the cursor to select “Cardiac”, press the【SET】key to go to the submenu, move the trackball to select “RV”, press the【SET】key to go to measurement state.
2. Measure RV; see “Distance Measurement” for the specific method.
3. The measurement result is displayed in the result window. Repeat steps 1 to 2 to perform the next measurement.
Page 99
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-77
3.6.2.4.3 Pulmonary Artery inner diameter (PA)
Function: To measure Pulmonary Artery inner diameter.
Measurement method:
3.6.2.4.4 Left Ventricular Mass (LV Mass)
Function: Based on this measurement item, to calculate Left
Ventricular Mass.
The items to be measured are:
LVPWd Left Ventricular Posterior Wall thickness diastolic cm
IVSTd Left Ventricular Septal thickness diastolic cm
LVIDd Left Ventricular Short Axis Diameter diastolic cm
Calculation formula:
LV Mass (g) 1.04*(LVPWd+IVSTd+LVIDd) 3 -LVIDd 3
–13.6
1. In B mode freeze the image, press the
Calckey to pop up the Meas & Calc menu, move the trackball, use the cursor to select “Cardiac”, press the【SET】key to go to the submenu, move the trackball to select “PA”, press the【SET】key to go to measurement state.
2. Measure PA; see “Distance Measurement” for the specific method.
3. The measurement result is displayed in the result window. Repeat steps 1 to 2 to perform the next measurement.
Page 100
A6T/A6/A5 Portable Ultrasonic Diagnostic System
4701-0040-01A
3-78
Measurement method:
1. In B mode freeze the image, press the 【Calc】key to pop up the
Meas & Calc menu, move the trackball, use the cursor to select “Cardiac”, press the【SET】key to go to the submenu, move the trackball to select “LVM”, press the【SET】key to go to measurement state.
2. Measure LVPWd, see “Distance Measurement” for the specific method.
3. Measure IVSTd, see “Distance Measurement” for the specific method.
4. Measure LVIDd, see “Distance Measurement” for the specific method.
5. Based on the formula to calculate LV Mass, the measurement and calculation result is displayed in the result window.
6. Repeat steps 1 to 5 to go to the next LV Mass measurement.
3.6.2.4.5 Left Ventricle Mass Weight Index (LVMWI)
Function: Based on patient’s height, weight and LV Mass, to calculate
LVMWI.
The formula:
LVMWI=LV Mass (g)/BSA (m2)
Loading...