Solartron 7151 Service Manual

5 (1)

7151

COMPUTING MULTIMETER

MAINTENANCE MANUAL

Issue 1: June 1984

SOLARTRON

Schlumberger

PartNo. 71510011

Solartron Instruments, Victoria Road, Farnborough

Hampshire, England GU14 7PW Telephone: Farnborough (0252) 544433

Telex: 858245 Solfar G Cables: Solartron Famborough

A division of Schlumberger Electronics (UK) Ltd

Solartron pursues a policy of continuous development and product improvement

19©84

The specification in this document may therefore be changed without notice

n

7151 COMPUTING MULTIMETER

MAINTENANCE MANUAL

1642g/0072g

CONTENTS

Chapter

1

General

Chapter

2

Calibration Procedures

Chapter

3

Dismantling & Reassembly

Chapter

4

Circuit Descriptions & Diagrams

Chapter

5

Fault

Diagnosis Guide

Chapter

6

Parts

Lists & Component Layout

1642g/0072g

r

Chapter 1 General

1642g/0072g

f

CHAPTER 1

1.GENERAL

The Solartron 7151 Computing Multimeter performs all common measurement functions, and offers: a library of programs; clock controlled measurements; and a programmable power-on status.

The instrument is suitable for general purpose bench work, or for use within a system where 7151 would be operated via one of its remote control interfaces. The interfaces provided are the IEEE 488 (1978) STD system and the RS232C V24 serial system.

2.SAFETY

The 7151 multimeter has been designed in accordance with the recommendations of IEC 348. To ensure the user's safety, and the continued safe operation of the instrument, it is advisable to fully observe the procedures and specifications given in the Operating Manual (Part No. 71510010).

An Earth wire is provided to ensure the user's safety. Therefore, if an extension mains cable is used, check that the Earth connection is maintained throughout the length of the extension.

When using 7151 on equipment which is capable of delivering high voltages (e.g. inductive circuitry giving high back emf's such as the secondary of a large mains transformer), it is most important that 7151's test leads are disconnected from the equipment before switched it off. This ensures that harmful back-emf's do not reach 7151. Care should always be exercised when handling the input leads, especially where high voltages are known to be present, or where high transients could occur.

Whenever it is likely that the safety of

the

instrument

has been

impaired

- e.g. if

it

shows visible signs of

damage,

if it fails

to

perform

correctly,

or

if the specifications

have

been exceeded

in

any way

- it

should be made inoperative and referred to a suitable repair depot. Any maintenance, adjustment or repair of the multimeter must be carried out by skilled personnel only, in accordance with the procedures and precautions detailed in this Maintenance Manual (part no. 71510011).

A Wherever this symbol appears on the front or rear panel it is advisable to consult the appropriate section of the Operating Manual for further information.

3.SUMMARY OF OPERATION

A schematic block diagram of the 7151 is shown in Fig. 1.1. 7151 is essentially a voltage measuring instrument which uses the pulse width technique of analog to digital conversion.

All inputs to the instrument are first converted to dc voltages before being passed to the input amplifier. This is simple enough for current (dc) and resistance, but ac inputs also undergo rms conversion to dc.

All inputs are suitably scaled by the input amplifier and passed to the

0072g/1634g

1.1

analog to digital converter (ADC). With no input, the ADC produces two balanced pulse trains of mark space ratio 1:1. When an input is received, the mark-space ratios of the trains respond in an equal and opposite manner, proportional to the size of the input. These trains are then converted to a single end and gated into a reversible counter. The nett result is a pulse count proportional to the measure of the input.

The measuring circuits are controlled by what is termed the 'floating' logic and consists essentially of a 8-bit microprocessor with 'Piggyback' ROM. The other circuits of 7151 are organised in a bus arrangement which is controlled by the 'earthy' logic and consists essentially of another 8-bit microprocessor. Isolated communication between the floating and earthy logic is acheived by opto coupled serial links. It is the earthy logic which is responsible for effective control of measurements, processing, remote control, the real time clock, the displays, and so on.

0072g/1634g

1.2

 

 

 

A DRIVE

OHMS

 

 

 

 

 

 

 

 

 

 

 

CONTROL

 

A DRIVE

 

 

 

 

 

 

 

CONTROL

 

 

I/P

o-

INPUT SWITCHING

INPUT

 

 

 

+/-REF

TERMINALS »

 

AMP

 

 

CONVERTER

REFERENCE

 

 

 

 

 

 

 

 

GAIN

 

COUNT

 

 

 

 

 

CONTROL

 

 

 

 

 

 

 

 

 

 

 

 

 

FLOATING

LOGIC

RESET

AUTO-CAL

 

 

 

 

MEMORY

 

 

 

 

 

 

 

A.C.

HOLD OFF

WATCH-DOG

CONDITIONING

ATTEN. &

 

 

GAIN CONTROL

 

OPTO COUPLERS

 

 

2 WIRE

SERIAL

 

 

 

 

 

 

 

 

LINK

 

 

 

 

 

 

 

 

 

 

 

KEYBOARD

8.

4

ROWS

KEYBOARD

&

 

 

 

 

FRONT/REAR

SW.

FRONT/REAR

SW.

 

 

 

 

 

 

 

 

 

 

DECODERS

 

 

 

DECODERS

 

 

HOLD -OFF

4 COLUMNS

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

WATCH-DPG

RESET

EARTHY LOGIC

 

 

 

 

 

 

 

 

 

 

1 COLUMN

 

 

 

 

 

 

 

 

 

 

INTERFACE

8

ROWS

INTERFACE

SW.

 

A8 TO A15

AO/DO TO

A7/D7

DECODERS

 

 

DECODERS

 

DATA BUS

ADDRESS BUS

ENABLE

ADDRESS

CLOCK

SERIAL

LATCHES

 

 

At) TO A7

7151 MULTIMETER FUNCTIONAL BLOCK DIAGRAM

TO FLOATING

FLOATING

POWER

POWER RAILS

SUPPLY

 

TO EARTHY

EARTHY

POWER

POWER RAILS

SUPPLY

 

DISPLAY

LATCHES DRIVER

DRIVER

INTERFACE BUS

LEVEL

CHANGING

TO

GPIA

TRANSCEIVERS

■ REAR

PANEL

 

GPIB

SOCKET

 

ANALOG OUTPUT

 

 

 

 

 

TO

 

 

 

REAR

PANEL

 

LEVEL

CHANGING

AUXILIARY SOCKET

TRANSCEIVERS

7151 MULTIMETER FUNCTIONAL BLOCK DIAGRAM * FIG 1.1

Calibration Procedures

Chapter

1642g/00'72g

{

I

(

{

1

CHAPTER 2

Setting-Up And Calibration Procedures

General

These procedures enable the instrument to be set-up and calibrated to the factory despatch standards.

The procedures are categorised into the following sections:

1.Setting-up procedures

2.Initial calibration procedures

3.Final calibration procedures

Safety

The instrument must be disconnected from the mains supply when dismantling it to gain access to the preset controls and also when it is being reassembled (see Chapter 4 for dlssembly instructions).

When adjusting preset controls beware of high test voltages, the guard potential on the guard plate and also the mains input supply.

Calibration Method

Owing to the automatic calibration circuits incorporated in 7151, it can only be calibrated by connecting it to a remote controller and then using the appropriate calibration commands. Alternatively, a calibration program can be used which is a much faster method of calibrating 7151. Solartron can supply, on tape cassettes, a calibration program for the more common types of controllers.

The user is advised to re-calibrate 7151 annually.

If the instrument's existing state of calibration is judged to be satisfactory, the user can simply re-write the existing calibration constants by sending the REFRESH command to 7151 once it is in the calibration mode.

Calibration Source

It is recommended that the calibration source has an accuracy of at least two times better than the accuracy specified for the various 7151 functions. The 7151 specification is given in the Operating Manual and the important percentage accuracies are as follows:-

DC Volts

0.002%

DC Current

0.02%

AC Volts

0.05%

AC Current

0.05%

Resistance

0.002%

0072g/1624g

2.1

ENTERING CALIBRATION MODE

Insert a shorted 2.5mm jack plug Into the rear panel CAL socket, causing the front panel CAL Indicator to repeatedly flash. The short may be within the plug Itself, or externally via a switch. The plug must remain fitted throughout the calibration, and can be removed after calibration is complete.

Note: Do not switch mains power on or off when the shorting plug Is fitted, otherwise the Internal calibration constants may be altered.

Using the controller, send the command CALIBRATE ON to 7151, putting It Into the calibration mode. The CAL Indicator should then be steady. Also displayed Is the word, 'CAL'. Once the calibration mode has been selected, the following conditions apply:

(a) Three commands cannot be used: TRIG

'OFF'

status

Is

adopted.

TRACK

NULL

'OFF'

status

Is

adopted,

 

all

 

nulls being deleted.

(b) Four commands become available: HI

 

 

 

 

 

LO

 

 

 

 

 

WRITE

for

refreshing existing

REFRESH

 

cal.

constants.

 

0072g/1624g

2.2

CALIBRATING MEASUREMENT RANGE

Using the controller, select the function and range to be calibrated by sending the appropriate MODE and RANGE commands.

7151 must then be supplied with two precisely known reference inputs (non-negative) one at approximately nominal full scale (referred to as the Hi point), and one at approximately zero (referred to as the Lo point), in the case of ac ranges the Lo point should not be less than approximately 5% of nominal full scale rather than zero. This ensures that all inputs are within the optimum part of 7151's linear range.

After a reference input is applied, 7151 must be informed of the precise value of the input. This is achieved by using the HI command for a Hi point, and the LO command for a Lo point. These commands must be accompanied by an integer argument number, of up to six digits in length, which expresses the applied input in terms of 5 x 9's count.

An integer value of 200000 corresponds to nominal full scale for any range.

For example, applying 2V on the 2V range, enter 200000 applying 20V on the 20V range, enter 200000 applying 5V on the 200V range, enter 005000

Apply the Hi point input to 7151 for the requisite function/range.

For example, 2.00843V on 2V dc range.

Using the controller, send the HI command to 7151.

For

example,

HI200843.

 

 

 

 

 

 

7151

responds

by displaying

'Hi Pt' for

about

1.5 seconds,

during which

time

it measures the

applied

reference

input.

When finished, the

It is

instrument displays

(and

outputs) its measured count, e.g.

214576.

of no consequence if

the

displayed count differs from the applied

input.

Repeat the above procedure for the Lo point. For example, reference = OV (short circuit), and send the LO command. For example LOO (leading zeroes need not be specified).

Having specified the Hi point and Lo point (in any order), send the command WRITE to 7151 (no argument required). This causes the calibration constants for the selected range/function to be calculated and stored in memory. If successful, the message 'Good' is displayed. If unsuccessful, an error message will be displayed and output to the controller.

Repeat the above instructions for each function/range to be calibrated.

0072g/1624g

2.3

RESTORING THE MEASUREMENT FUNCTIONS

Using the controller, send 7151 the command CALIBRATE OFF. The CAL indicator will then flash indicating that the CAL shorting plug is still fitted.

Withdraw the CAL shorting plug. The CAL indicator should then be invisible, the instrument being ready for normal use.

SUMMARY

(a)Insert CAL shorting plug (2.5mm) in rear panel socket.

(b)Select the calibration mode by sending the CALIBRATE ON command.

(c)Select the requisite function and range to be calibrated and perform the calibration sequence. Repeat for each range/function to be calibrated.

(d)De-select the calibration mode by sending the CALIBRATE OFF command.

(e)Remove CAL plug.

0072g/1624g

2.4

Setting-Up Procedures

DC Power Supply Checks

Measure the dc supplies on PCB1 and PCB2 at the output pins of the appropriate regulator IC's. Tolerances of the most important supplies, mains voltage 240V, follow:

floating 15V unregulated floating 15V regulated floating 5V unregulated floating 5V regulated earthy 5V unregulated earthy 5v regulated

Display Checks

between 20.7V and 21.6V 15 ± 0.75V

between 8.8V and 9.1V 5 + 0.25V

between 9.5V and 9.8V 5 ± 0.25V

The contrast of the display can be adjusted by means of RV301. Make the digits appear as black as possible but without introducing slurring when a reading changes.

Keyboard Checks

The following sequence exercises all 16 keys.

Key Press

FILT 2 press minimum

K£2

V~

V===

AUTO 2 presses minimum

V

A

LOCAL

NULL 2 presses minimuim 6x92 presses minimum TRACK 2 presses minimum SAMPLE with "HOLD" asserted

COMPUTE

MENU

Display Response

"FILT" on/off

Finish with "FILT" off ma-

mA===

KQ

V~

V===

"AUTO" on/off

Check for downranging Check for upranging

"GPIB nm" where nm is address value.

"NULL" on/off "6x9" on/off "HOLD" on/off

“HOLD" goes out briefly and returns.

"NO PROG" “PROBES?"

0072g/1624g

2.5

Initial Calibration Procedures

Test Equipment

1.General purpose DMM.

2.General purpose oscilloscope

3.Controller, e.g. Commodore PET fitted with BASIC III or BASIC IV firmware.

4.Calibrator, e.g. Fluke 5101 fitted with GPIB interface.

5.ACV Calibrator, e.g. Hewlett-Packard 745.

6.ACV High Voltage Amplifier e.g. Hewlett-Packard 746.

7.Capacitor O.lyF polypropylene attached to a twin 4mm banana plugs (3/4" centres).

Switch on 7151 and allow to warm up for at least one hour before calibration.

The initial calibration procedures are detailed in the following tables:

Table No.

Initial

Procedure

2.1

calibration,

DC Volts

2.2

Initial

calibration,

Resistance

2.3

Initial

calibration.

Current

2.4

Initial

calibration,

AC Volts

Please Note: The limits of error expressed in the following tables are those adhered to by the factory for a new instrument. As an instrument 'ages', "components become more noisy or their tolerances increase.

Therefore, when calibrating a used instrument, it may be necessary to accept limits of error that are marginally higher than those listed in these pages. However, the instrument should always conform to the commercial specification (see Operating Manual) after calibration.

0072g/1624g

2.6

Table

2.1

Initial Calibration, DC Volts

 

 

TEST

RANGE

&

INPUT

ACTION

LIMITS

COMMENTS

 

MODE

 

 

 

 

 

1Configure the rear panel interface switches and connect the controller to

7151.

Do not insert the calibration Key Jack yet. FRONT/REAR switch to 'FRONT'.

2

3

 

 

Power on

 

 

4

 

 

Insert Calibration

 

 

 

 

 

Key Jack

 

 

5

2VDC

s/c link

Adjust RV3.

±

100 yV

 

 

VHI-VLO

DVM between link

 

 

 

 

front

& ROME

 

 

6

2VDC

s/c

Check display ^

3

adjacent

 

 

VHI-VLO

'for2scatter

values

 

 

Front

 

 

 

7

0.2VDC

o/c

Check reading

0

± lOOyV

I/P amp gross offset null.

2V range noise test. The reading may jump every 10 secs at drift-correct.

Input current measurement. Value may be exceeded at drift-correct.

8

2VDC

4V< plus

Measure C4 with

 

 

overload

DMM referred

to

 

 

< -100V

ROME

 

9

2VDC

-4V< minus

As above

 

 

 

overload

 

 

 

 

< -100V

 

 

10

2VDC

±1,99999V

Adjust RV1

 

 

 

alternatively

CAL BAL

 

11

2VDC

+1.99999V &

Do calibration

 

 

0.00000V

routine over

the

 

 

 

interface

 

12

0.2VDC

0.199999V &

Do calibration

 

 

0.00000V

routine over

the

 

 

 

interface

 

13

20VDC

+19.9999V &

Do calibration

 

 

0.00000V

routine over

the

 

 

 

interface

 

+3.90

Positive

input-clamp

+3.05

test

(D6)

 

-3.05

Negative

input-clamp

-3.90

test

(D26)

+ and -

Cal. Bal

Adjustment

equal within Use continuous

1 bit

drift-correct (Yl).

+2V set-up

+2V set-up. Use the calibrator to deliver 0 volt.

+20V set-up.

0072g/1624g

2.7

Table 2.1 Cent.

 

 

TEST

RANGE &

INPUT

ACTION

 

 

MODE

 

 

 

14

200VDC +199.999V &

Do calibration

 

 

0.00000V

routine over

the

 

 

 

interface

 

15

IkVDC

+10000.00V &

Do calibration

 

 

0.00V

routine over

the

 

 

 

interface

 

LIMITS COMMENTS

+200V set-up.

+lkV Set-up. The Calibrator LO and the 7151 LO input should be mains grounded. Check that the spark-gap does not operate. Apply for

1 minute and check that the reading does not drift more than 2 bits.

16

 

 

Exit Cal

Mode

 

17

2VDC

+1.00000V

Measure

+2 bits

Linearity. Change

 

 

-1.00000V

 

pos-neg

polarity changing

 

 

 

 

error

over inputs.

0072g/1624g

2.8

Table

2.2

Initial Calibration,

Resistance

 

 

 

 

 

 

 

 

TEST

RANGE

5.

INPUT

ACTION

 

LIMITS

COMMENTS

 

 

1

MODE

DMM between

Measure

current

100±5yA

 

 

 

 

 

 

20kfl

 

 

 

 

 

 

 

 

I-

and LO

from -I.

to current.

 

 

 

 

 

 

 

 

 

 

 

 

 

DMM set

 

 

 

 

 

 

 

 

2

200kfl

As

above

As

above

 

10.0

±

0.5uA

 

 

 

 

 

3

2Mfi

As

above

As

above

 

1.0

±

0,5yA

 

 

 

 

 

4

2k£2

l.OOOOOkfl

Do

calibration

 

 

 

2KQ

range set

up

 

 

and

IQ

using

the interface

 

 

 

 

 

 

 

 

5

20kfi

lO.OOOOkQ

Do calibration

 

 

 

20KQ

 

range

set

 

 

and

IQ

using

the interface

 

 

 

up.

 

 

 

 

6

200kfi

lOO.OOOkQ

Do

calibration

 

 

 

200KQ

range set

 

 

and

IQ

using

the interface

 

 

 

up.

 

 

 

 

7

2Mft

1.00000MQ

Do

calibration

 

 

 

2MQ

range set

up

 

 

and

IQ

using

the interface

 

 

 

O.lyF

in

 

 

 

 

 

 

 

 

 

 

 

 

 

 

parallel will

 

 

 

 

 

 

 

 

 

 

 

 

 

reduce scatter

 

 

 

 

 

 

 

 

 

 

 

 

caused by series

 

 

 

 

 

 

 

 

 

 

 

 

mode

 

interference.

8

20MS2

10.0000MQ

Do

calibration

 

 

 

20MQ

 

range

set

 

 

 

and

IQ

using

the interface

 

 

 

up.

O.lyF

in

 

 

 

 

 

 

 

 

 

 

 

 

 

parallel will

 

 

 

 

 

 

 

 

 

 

 

 

 

reduce scatter.

9

2M$2

DMM

across

Measure

the o/c

+5.2V ±

IV

Q source

 

 

 

 

7150 HI & LO

volts

from Q

 

 

 

positive clamp.

 

 

 

 

 

source.

 

 

 

 

 

 

 

 

 

10

2M£2

240VAC

Apply VHI-VLO

 

 

 

Ohms

overload

test

 

 

/

50 Hz

10

seconds.

 

 

 

 

 

 

 

 

11

2Mfi

1.00000MQ

Check

after test 9

1.00000MQ

Survival check

for

 

 

 

 

 

 

 

 

 

±100

bits

damage after test

 

 

 

 

 

 

 

 

 

 

 

 

10.

 

 

 

 

12

DV

+1

kV

Check

display

± 10

bits

IkV step input

 

 

Auto

applied 5

 

 

 

 

 

 

 

test.

LO and

 

 

 

times

 

 

 

 

 

 

 

GUARD must

LO of

 

 

 

 

 

 

 

 

 

 

 

 

connect to

 

 

 

 

 

 

 

 

 

 

 

 

Cal. and also

to

 

 

 

 

 

 

 

 

 

 

 

 

mains ground.

 

 

 

 

 

 

 

 

 

 

 

 

 

7151

must uprange

 

 

 

 

 

 

 

 

 

 

 

 

withou power

 

 

 

 

 

 

 

 

 

 

 

 

 

restarts. It is

 

 

 

 

 

 

 

 

 

 

 

 

permissible

that

 

 

 

 

 

 

 

 

 

 

 

 

the spark-gap

 

 

 

 

 

 

 

 

 

 

 

 

 

operates.

 

 

0072g/1624g

2.9

Table 2.3

Initial Calibration, Current

 

 

 

TEST

RANGE

&

INPUT

ACTION

 

LIMITS

COMMENTS

1

MODE

 

+1.00000A &

Calibrate over

the

 

1 Ampere Set-up

DC A

 

 

 

 

 

open circuit

bus

 

 

 

2

AC A

 

400 Hz

Calibrate over

the

 

2 Ampere Set-up

 

 

 

1.99999A &

bus

 

 

 

 

 

 

0.19999A

 

 

 

 

3

 

 

 

Exit Cal Mode

 

 

 

5

DC AC

 

+1.99999A

Measure voltage at

0.80 volt

Burden

 

 

 

 

current front

 

 

 

 

 

 

sockets with a

 

 

 

 

 

 

 

DMM.

 

 

 

0072g/1624g

2.10

Table

2.4

Initial Calibration, AC Volts

 

 

 

 

 

 

TEST

RANGE

&

INPUT

ACTION

 

LIMITS

COMMENTS

 

1

MODE

 

s/c

Adjust RV2 for

 

±150 bits

IC15

offset

null

VAC

 

 

 

various

 

minimum @ TP3

 

referred to

adjust.

Use DMM

 

ranges

 

 

 

OV

to monitor

TP3

 

 

 

 

 

 

 

Transformer

to be

 

 

 

 

 

 

 

lamination

 

 

 

 

 

 

 

mains-grounded.

 

 

 

 

 

 

 

The lowest

figure

 

 

 

 

 

 

 

possible

is

 

 

 

 

 

 

 

 

required;

if

 

 

 

 

 

 

 

necessary by

 

 

 

 

 

 

 

error-sharing

 

 

 

 

 

 

 

among

the

ranges.

2

20VAC

 

19.9999V

Note reading

 

 

 

 

 

 

 

 

 

400Hz

 

 

 

 

 

 

 

3

20VAC

 

19.9999V

Adjust CV1 St R10

Value at test Attenuator HF

 

 

 

50kHz

for flat response

12' ±0.010V

trim.

100 bit

 

 

 

 

 

 

 

limit

applies

 

 

 

 

 

 

 

when a dummy lid

 

 

 

 

 

 

 

is fitted.

 

4

0.2VAC

0.199999V &

Calibrate over

the

 

0.2V LF Set-up

 

 

 

0.019999V

bus.

 

 

Fluke

5101.

 

 

 

 

400Hz

 

 

 

 

 

 

 

5

2VAC

 

1.99999V &

Calibrate over

the

 

2V LF Set-up

 

 

 

0.19999V

bus.

 

 

 

 

 

 

 

 

 

400Hz

 

 

 

 

 

 

 

6

20VAC

 

19.9999V S>

Calibrate over

the

 

20V Set-up

 

 

 

 

1.9999V

bus.

 

 

 

 

 

 

 

 

 

400Hz

 

 

 

 

 

 

 

7

200VAC

199.999V S.

Calibrate over

the

 

200V Set-up

 

 

 

 

19.999V

bus.

 

 

 

 

 

 

 

 

 

400Hz

 

 

 

 

 

 

 

8

IkVAC

 

750.00V &

Calibrate over

the

 

1 kV Set-up

 

 

 

 

199.99V

bus.

 

 

 

 

 

 

 

 

 

400Hz

 

 

 

 

 

 

 

9

 

 

 

Exit Cal Mode

 

 

 

 

 

 

10

0.2VAC

30KHz

Check

 

0.199999V ±

 

 

 

 

 

 

 

0.199999V

 

 

.000120V

 

 

 

 

11

0.2VAC

lOKHz

Check

 

0.199999V ±

 

 

 

 

 

 

 

0.199999V

 

 

,000096V

 

 

 

 

0072g/1624g

2.11

Table 2.4 Cent.

 

TEST

RANGE &

INPUT

ACTION

12

MODE

10kHz

Check

2VAC

 

 

1.99999V

 

13

2VAC

30kHz

Check

 

 

1.99999V

 

14

20VAC

30kHz

Check

 

 

19.9999V

 

15

20VAC

10kHz

Check

 

 

19.9999V

 

16

200VAC

10kHz

Check

 

 

199.999V

 

17

200VAC

30kHz

Check

 

 

199.999V

 

18

IkVAC

10kHz

Check

 

 

750.00V

 

19

IkVAC

30kHz

Check

 

 

750.00V

 

20

0.2VAC

s/c

Check

21

2VAC

10Hz

Check

 

 

2.00000V

 

22

2VAC

20Hz

Check

 

 

2.00000V

 

23

2VAC

40Hz

Check

 

 

2.00000V

 

24

2VAC

100Hz

Check

 

 

2.00000V

 

25

2VAC

100kHz

Check

 

 

0.19999V

 

26

20VAC

100kHz

Check

 

 

20.0000V

 

27

200VAC

100kHz

Check

 

 

200.000V

 

LIMITS

1.99999V ±

.00096V

1.99999V ±

.00120V

19.9999V ±

0.0120V

19.9999V ±

.0096V

199.999V +

.096V

199.999V ±

0.120V

750.00V +

0.46V

750.00V +

0.70V

150yV

2.00000V ±

0.01456V

2.00000V ±

0.00416V

2.00000V ±

0.00096V

2.00000V ±

0.00880V

0.199999V ±

0.000880V

20.0000V ±

0.0880V

200.000V ±

0.880V

COMMENTS

s/c zero. Trasnsformer laminations to be mains-grounded.

0072g/1624g

2.12

Loading...
+ 57 hidden pages