7151
COMPUTING MULTIMETER
MAINTENANCE MANUAL
Issue 1: June 1984
SOLARTRON
Schlumberger
PartNo. 71510011
Solartron Instruments, Victoria Road, Farnborough
Hampshire, England GU14 7PW Telephone: Farnborough (0252) 544433
Telex: 858245 Solfar G Cables: Solartron Famborough
A division of Schlumberger Electronics (UK) Ltd
Solartron pursues a policy of continuous development and product improvement |
19©84 |
The specification in this document may therefore be changed without notice |
n
7151 COMPUTING MULTIMETER
MAINTENANCE MANUAL
1642g/0072g
CONTENTS
Chapter |
1 |
General |
|
Chapter |
2 |
Calibration Procedures |
|
Chapter |
3 |
Dismantling & Reassembly |
|
Chapter |
4 |
Circuit Descriptions & Diagrams |
|
Chapter |
5 |
Fault |
Diagnosis Guide |
Chapter |
6 |
Parts |
Lists & Component Layout |
1642g/0072g
r
Chapter 1 General
1642g/0072g
f
CHAPTER 1
1.GENERAL
The Solartron 7151 Computing Multimeter performs all common measurement functions, and offers: a library of programs; clock controlled measurements; and a programmable power-on status.
The instrument is suitable for general purpose bench work, or for use within a system where 7151 would be operated via one of its remote control interfaces. The interfaces provided are the IEEE 488 (1978) STD system and the RS232C V24 serial system.
2.SAFETY
The 7151 multimeter has been designed in accordance with the recommendations of IEC 348. To ensure the user's safety, and the continued safe operation of the instrument, it is advisable to fully observe the procedures and specifications given in the Operating Manual (Part No. 71510010).
An Earth wire is provided to ensure the user's safety. Therefore, if an extension mains cable is used, check that the Earth connection is maintained throughout the length of the extension.
When using 7151 on equipment which is capable of delivering high voltages (e.g. inductive circuitry giving high back emf's such as the secondary of a large mains transformer), it is most important that 7151's test leads are disconnected from the equipment before switched it off. This ensures that harmful back-emf's do not reach 7151. Care should always be exercised when handling the input leads, especially where high voltages are known to be present, or where high transients could occur.
Whenever it is likely that the safety of |
the |
instrument |
has been |
impaired |
||||
- e.g. if |
it |
shows visible signs of |
damage, |
if it fails |
to |
perform |
||
correctly, |
or |
if the specifications |
have |
been exceeded |
in |
any way |
- it |
should be made inoperative and referred to a suitable repair depot. Any maintenance, adjustment or repair of the multimeter must be carried out by skilled personnel only, in accordance with the procedures and precautions detailed in this Maintenance Manual (part no. 71510011).
A Wherever this symbol appears on the front or rear panel it is advisable to consult the appropriate section of the Operating Manual for further information.
3.SUMMARY OF OPERATION
A schematic block diagram of the 7151 is shown in Fig. 1.1. 7151 is essentially a voltage measuring instrument which uses the pulse width technique of analog to digital conversion.
All inputs to the instrument are first converted to dc voltages before being passed to the input amplifier. This is simple enough for current (dc) and resistance, but ac inputs also undergo rms conversion to dc.
All inputs are suitably scaled by the input amplifier and passed to the
0072g/1634g |
1.1 |
analog to digital converter (ADC). With no input, the ADC produces two balanced pulse trains of mark space ratio 1:1. When an input is received, the mark-space ratios of the trains respond in an equal and opposite manner, proportional to the size of the input. These trains are then converted to a single end and gated into a reversible counter. The nett result is a pulse count proportional to the measure of the input.
The measuring circuits are controlled by what is termed the 'floating' logic and consists essentially of a 8-bit microprocessor with 'Piggyback' ROM. The other circuits of 7151 are organised in a bus arrangement which is controlled by the 'earthy' logic and consists essentially of another 8-bit microprocessor. Isolated communication between the floating and earthy logic is acheived by opto coupled serial links. It is the earthy logic which is responsible for effective control of measurements, processing, remote control, the real time clock, the displays, and so on.
0072g/1634g |
1.2 |
|
|
|
A DRIVE |
OHMS |
|
|
|
|
|
|
|
|
|
||
|
|
|
CONTROL |
|
A DRIVE |
|
|
|
|
|
|
|
CONTROL |
|
|
I/P |
o- |
INPUT SWITCHING |
INPUT |
|
|
|
+/-REF |
TERMINALS » |
|
AMP |
|
|
CONVERTER |
REFERENCE |
|
|
|
|
|
||||
|
|
|
|
GAIN |
|
COUNT |
|
|
|
|
|
CONTROL |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
FLOATING |
LOGIC |
RESET |
AUTO-CAL |
|
|
|
|
MEMORY |
|||
|
|
|
|
|
|
|
A.C. |
HOLD OFF |
|
WATCH-DOG |
||
CONDITIONING |
||
ATTEN. & |
||
|
||
|
GAIN CONTROL |
|
|
OPTO COUPLERS |
|
|
2 WIRE |
SERIAL |
|
|
|
|
|
|
|
|
LINK |
|
|
|
|
|
|
|
|
|
|
|
KEYBOARD |
8. |
4 |
ROWS |
KEYBOARD |
& |
|
|
|
|
FRONT/REAR |
SW. |
FRONT/REAR |
SW. |
||
|
|
|
|
|
|
||||
|
|
|
|
DECODERS |
|
|
|
DECODERS |
|
|
HOLD -OFF |
4 COLUMNS |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
WATCH-DPG |
RESET |
EARTHY LOGIC |
|
|
|
|
|
|
|
|
|
|
1 COLUMN |
|
|
|
|
|
|
|
|
|
|
INTERFACE |
8 |
ROWS |
INTERFACE |
SW. |
|
|
A8 TO A15 |
AO/DO TO |
A7/D7 |
DECODERS |
|
|
DECODERS |
|
DATA BUS
ADDRESS BUS
ENABLE
ADDRESS |
CLOCK |
SERIAL |
|
LATCHES |
|||
|
|
At) TO A7
7151 MULTIMETER FUNCTIONAL BLOCK DIAGRAM
TO FLOATING |
FLOATING |
|
POWER |
||
POWER RAILS |
||
SUPPLY |
||
|
TO EARTHY |
EARTHY |
|
POWER |
||
POWER RAILS |
||
SUPPLY |
||
|
DISPLAY
LATCHES DRIVER
DRIVER
INTERFACE BUS |
LEVEL |
CHANGING |
TO |
|
GPIA |
TRANSCEIVERS |
■ REAR |
PANEL |
|
|
GPIB |
SOCKET |
||
|
ANALOG OUTPUT |
|
|
|
|
|
|
TO |
|
|
|
|
REAR |
PANEL |
|
LEVEL |
CHANGING |
AUXILIARY SOCKET |
TRANSCEIVERS
7151 MULTIMETER FUNCTIONAL BLOCK DIAGRAM * FIG 1.1
Calibration Procedures
Chapter
1642g/00'72g
{
I
(
{
1
CHAPTER 2
Setting-Up And Calibration Procedures
General
These procedures enable the instrument to be set-up and calibrated to the factory despatch standards.
The procedures are categorised into the following sections:
1.Setting-up procedures
2.Initial calibration procedures
3.Final calibration procedures
Safety
The instrument must be disconnected from the mains supply when dismantling it to gain access to the preset controls and also when it is being reassembled (see Chapter 4 for dlssembly instructions).
When adjusting preset controls beware of high test voltages, the guard potential on the guard plate and also the mains input supply.
Calibration Method
Owing to the automatic calibration circuits incorporated in 7151, it can only be calibrated by connecting it to a remote controller and then using the appropriate calibration commands. Alternatively, a calibration program can be used which is a much faster method of calibrating 7151. Solartron can supply, on tape cassettes, a calibration program for the more common types of controllers.
The user is advised to re-calibrate 7151 annually.
If the instrument's existing state of calibration is judged to be satisfactory, the user can simply re-write the existing calibration constants by sending the REFRESH command to 7151 once it is in the calibration mode.
Calibration Source
It is recommended that the calibration source has an accuracy of at least two times better than the accuracy specified for the various 7151 functions. The 7151 specification is given in the Operating Manual and the important percentage accuracies are as follows:-
DC Volts |
0.002% |
DC Current |
0.02% |
AC Volts |
0.05% |
AC Current |
0.05% |
Resistance |
0.002% |
0072g/1624g |
2.1 |
ENTERING CALIBRATION MODE
Insert a shorted 2.5mm jack plug Into the rear panel CAL socket, causing the front panel CAL Indicator to repeatedly flash. The short may be within the plug Itself, or externally via a switch. The plug must remain fitted throughout the calibration, and can be removed after calibration is complete.
Note: Do not switch mains power on or off when the shorting plug Is fitted, otherwise the Internal calibration constants may be altered.
Using the controller, send the command CALIBRATE ON to 7151, putting It Into the calibration mode. The CAL Indicator should then be steady. Also displayed Is the word, 'CAL'. Once the calibration mode has been selected, the following conditions apply:
(a) Three commands cannot be used: TRIG |
'OFF' |
status |
Is |
adopted. |
|
TRACK |
|||||
NULL |
'OFF' |
status |
Is |
adopted, |
|
|
all |
|
nulls being deleted. |
||
(b) Four commands become available: HI |
|
|
|
|
|
LO |
|
|
|
|
|
WRITE |
for |
refreshing existing |
|||
REFRESH |
|||||
|
cal. |
constants. |
|
0072g/1624g |
2.2 |
CALIBRATING MEASUREMENT RANGE
Using the controller, select the function and range to be calibrated by sending the appropriate MODE and RANGE commands.
7151 must then be supplied with two precisely known reference inputs (non-negative) one at approximately nominal full scale (referred to as the Hi point), and one at approximately zero (referred to as the Lo point), in the case of ac ranges the Lo point should not be less than approximately 5% of nominal full scale rather than zero. This ensures that all inputs are within the optimum part of 7151's linear range.
After a reference input is applied, 7151 must be informed of the precise value of the input. This is achieved by using the HI command for a Hi point, and the LO command for a Lo point. These commands must be accompanied by an integer argument number, of up to six digits in length, which expresses the applied input in terms of 5 x 9's count.
An integer value of 200000 corresponds to nominal full scale for any range.
For example, applying 2V on the 2V range, enter 200000 applying 20V on the 20V range, enter 200000 applying 5V on the 200V range, enter 005000
Apply the Hi point input to 7151 for the requisite function/range.
For example, 2.00843V on 2V dc range.
Using the controller, send the HI command to 7151.
For |
example, |
HI200843. |
|
|
|
|
|
|
|
7151 |
responds |
by displaying |
'Hi Pt' for |
about |
1.5 seconds, |
during which |
|||
time |
it measures the |
applied |
reference |
input. |
When finished, the |
It is |
|||
instrument displays |
(and |
outputs) its measured count, e.g. |
214576. |
||||||
of no consequence if |
the |
displayed count differs from the applied |
input. |
Repeat the above procedure for the Lo point. For example, reference = OV (short circuit), and send the LO command. For example LOO (leading zeroes need not be specified).
Having specified the Hi point and Lo point (in any order), send the command WRITE to 7151 (no argument required). This causes the calibration constants for the selected range/function to be calculated and stored in memory. If successful, the message 'Good' is displayed. If unsuccessful, an error message will be displayed and output to the controller.
Repeat the above instructions for each function/range to be calibrated.
0072g/1624g |
2.3 |
RESTORING THE MEASUREMENT FUNCTIONS
Using the controller, send 7151 the command CALIBRATE OFF. The CAL indicator will then flash indicating that the CAL shorting plug is still fitted.
Withdraw the CAL shorting plug. The CAL indicator should then be invisible, the instrument being ready for normal use.
SUMMARY
(a)Insert CAL shorting plug (2.5mm) in rear panel socket.
(b)Select the calibration mode by sending the CALIBRATE ON command.
(c)Select the requisite function and range to be calibrated and perform the calibration sequence. Repeat for each range/function to be calibrated.
(d)De-select the calibration mode by sending the CALIBRATE OFF command.
(e)Remove CAL plug.
0072g/1624g |
2.4 |
Setting-Up Procedures
DC Power Supply Checks
Measure the dc supplies on PCB1 and PCB2 at the output pins of the appropriate regulator IC's. Tolerances of the most important supplies, mains voltage 240V, follow:
floating 15V unregulated floating 15V regulated floating 5V unregulated floating 5V regulated earthy 5V unregulated earthy 5v regulated
Display Checks
between 20.7V and 21.6V 15 ± 0.75V
between 8.8V and 9.1V 5 + 0.25V
between 9.5V and 9.8V 5 ± 0.25V
The contrast of the display can be adjusted by means of RV301. Make the digits appear as black as possible but without introducing slurring when a reading changes.
Keyboard Checks
The following sequence exercises all 16 keys.
Key Press
FILT 2 press minimum
K£2
V~
V===
AUTO 2 presses minimum
V
A
LOCAL
NULL 2 presses minimuim 6x92 presses minimum TRACK 2 presses minimum SAMPLE with "HOLD" asserted
COMPUTE
MENU
Display Response
"FILT" on/off
Finish with "FILT" off ma-
mA===
KQ
V~
V===
"AUTO" on/off
Check for downranging Check for upranging
"GPIB nm" where nm is address value.
"NULL" on/off "6x9" on/off "HOLD" on/off
“HOLD" goes out briefly and returns.
"NO PROG" “PROBES?"
0072g/1624g |
2.5 |
Initial Calibration Procedures
Test Equipment
1.General purpose DMM.
2.General purpose oscilloscope
3.Controller, e.g. Commodore PET fitted with BASIC III or BASIC IV firmware.
4.Calibrator, e.g. Fluke 5101 fitted with GPIB interface.
5.ACV Calibrator, e.g. Hewlett-Packard 745.
6.ACV High Voltage Amplifier e.g. Hewlett-Packard 746.
7.Capacitor O.lyF polypropylene attached to a twin 4mm banana plugs (3/4" centres).
Switch on 7151 and allow to warm up for at least one hour before calibration.
The initial calibration procedures are detailed in the following tables:
Table No. |
Initial |
Procedure |
|
2.1 |
calibration, |
DC Volts |
|
2.2 |
Initial |
calibration, |
Resistance |
2.3 |
Initial |
calibration. |
Current |
2.4 |
Initial |
calibration, |
AC Volts |
Please Note: The limits of error expressed in the following tables are those adhered to by the factory for a new instrument. As an instrument 'ages', "components become more noisy or their tolerances increase.
Therefore, when calibrating a used instrument, it may be necessary to accept limits of error that are marginally higher than those listed in these pages. However, the instrument should always conform to the commercial specification (see Operating Manual) after calibration.
0072g/1624g |
2.6 |
Table |
2.1 |
Initial Calibration, DC Volts |
|
|
||
TEST |
RANGE |
& |
INPUT |
ACTION |
LIMITS |
COMMENTS |
|
MODE |
|
|
|
|
|
1Configure the rear panel interface switches and connect the controller to
7151.
Do not insert the calibration Key Jack yet. FRONT/REAR switch to 'FRONT'.
2
3 |
|
|
Power on |
|
|
4 |
|
|
Insert Calibration |
|
|
|
|
|
Key Jack |
|
|
5 |
2VDC |
s/c link |
Adjust RV3. |
± |
100 yV |
|
|
VHI-VLO |
DVM between link |
|
|
|
|
front |
& ROME |
|
|
6 |
2VDC |
s/c |
Check display ^ |
3 |
adjacent |
|
|
VHI-VLO |
'for2scatter |
values |
|
|
|
Front |
|
|
|
7 |
0.2VDC |
o/c |
Check reading |
0 |
± lOOyV |
I/P amp gross offset null.
2V range noise test. The reading may jump every 10 secs at drift-correct.
Input current measurement. Value may be exceeded at drift-correct.
8 |
2VDC |
4V< plus |
Measure C4 with |
|
|
|
overload |
DMM referred |
to |
|
|
< -100V |
ROME |
|
9 |
2VDC |
-4V< minus |
As above |
|
|
|
overload |
|
|
|
|
< -100V |
|
|
10 |
2VDC |
±1,99999V |
Adjust RV1 |
|
|
|
alternatively |
CAL BAL |
|
11 |
2VDC |
+1.99999V & |
Do calibration |
|
|
|
0.00000V |
routine over |
the |
|
|
|
interface |
|
12 |
0.2VDC |
0.199999V & |
Do calibration |
|
|
|
0.00000V |
routine over |
the |
|
|
|
interface |
|
13 |
20VDC |
+19.9999V & |
Do calibration |
|
|
|
0.00000V |
routine over |
the |
|
|
|
interface |
|
+3.90 |
Positive |
input-clamp |
|
+3.05 |
test |
(D6) |
|
-3.05 |
Negative |
input-clamp |
|
-3.90 |
test |
(D26) |
|
+ and - |
Cal. Bal |
Adjustment |
|
equal within Use continuous |
|||
1 bit |
drift-correct (Yl). |
+2V set-up
+2V set-up. Use the calibrator to deliver 0 volt.
+20V set-up.
0072g/1624g |
2.7 |
Table 2.1 Cent. |
|
|
||
TEST |
RANGE & |
INPUT |
ACTION |
|
|
MODE |
|
|
|
14 |
200VDC +199.999V & |
Do calibration |
||
|
|
0.00000V |
routine over |
the |
|
|
|
interface |
|
15 |
IkVDC |
+10000.00V & |
Do calibration |
|
|
|
0.00V |
routine over |
the |
|
|
|
interface |
|
LIMITS COMMENTS
+200V set-up.
+lkV Set-up. The Calibrator LO and the 7151 LO input should be mains grounded. Check that the spark-gap does not operate. Apply for
1 minute and check that the reading does not drift more than 2 bits.
16 |
|
|
Exit Cal |
Mode |
|
17 |
2VDC |
+1.00000V |
Measure |
+2 bits |
Linearity. Change |
|
|
-1.00000V |
|
pos-neg |
polarity changing |
|
|
|
|
error |
over inputs. |
0072g/1624g |
2.8 |
Table |
2.2 |
Initial Calibration, |
Resistance |
|
|
|
|
|
|
|
|
|||||
TEST |
RANGE |
5. |
INPUT |
ACTION |
|
LIMITS |
COMMENTS |
|
|
|||||||
1 |
MODE |
DMM between |
Measure |
current |
100±5yA |
|
|
|
|
|
|
|||||
20kfl |
|
|
|
|
|
|
||||||||||
|
|
I- |
and LO |
from -I. |
to current. |
|
|
|
|
|
|
|
|
|||
|
|
|
|
|
DMM set |
|
|
|
|
|
|
|
|
|||
2 |
200kfl |
As |
above |
As |
above |
|
10.0 |
± |
0.5uA |
|
|
|
|
|
||
3 |
2Mfi |
As |
above |
As |
above |
|
1.0 |
± |
0,5yA |
|
|
|
|
|
||
4 |
2k£2 |
l.OOOOOkfl |
Do |
calibration |
|
|
|
2KQ |
range set |
up |
||||||
|
|
and |
IQ |
using |
the interface |
|
|
|
|
|
|
|
|
|||
5 |
20kfi |
lO.OOOOkQ |
Do calibration |
|
|
|
20KQ |
|
range |
set |
||||||
|
|
and |
IQ |
using |
the interface |
|
|
|
up. |
|
|
|
|
|||
6 |
200kfi |
lOO.OOOkQ |
Do |
calibration |
|
|
|
200KQ |
range set |
|||||||
|
|
and |
IQ |
using |
the interface |
|
|
|
up. |
|
|
|
|
|||
7 |
2Mft |
1.00000MQ |
Do |
calibration |
|
|
|
2MQ |
range set |
up |
||||||
|
|
and |
IQ |
using |
the interface |
|
|
|
O.lyF |
in |
|
|
||||
|
|
|
|
|
|
|
|
|
|
|
|
parallel will |
|
|||
|
|
|
|
|
|
|
|
|
|
|
|
reduce scatter |
||||
|
|
|
|
|
|
|
|
|
|
|
|
caused by series |
||||
|
|
|
|
|
|
|
|
|
|
|
|
mode |
|
interference. |
||
8 |
20MS2 |
10.0000MQ |
Do |
calibration |
|
|
|
20MQ |
|
range |
set |
|
||||
|
|
and |
IQ |
using |
the interface |
|
|
|
up. |
O.lyF |
in |
|
||||
|
|
|
|
|
|
|
|
|
|
|
|
parallel will |
|
|||
|
|
|
|
|
|
|
|
|
|
|
|
reduce scatter. |
||||
9 |
2M$2 |
DMM |
across |
Measure |
the o/c |
+5.2V ± |
IV |
Q source |
|
|
||||||
|
|
7150 HI & LO |
volts |
from Q |
|
|
|
positive clamp. |
||||||||
|
|
|
|
|
source. |
|
|
|
|
|
|
|
|
|
||
10 |
2M£2 |
240VAC |
Apply VHI-VLO |
|
|
|
Ohms |
overload |
test |
|||||||
|
|
/ |
50 Hz |
10 |
seconds. |
|
|
|
|
|
|
|
|
|||
11 |
2Mfi |
1.00000MQ |
Check |
after test 9 |
1.00000MQ |
Survival check |
for |
|||||||||
|
|
|
|
|
|
|
|
|
±100 |
bits |
damage after test |
|||||
|
|
|
|
|
|
|
|
|
|
|
|
10. |
|
|
|
|
12 |
DV |
+1 |
kV |
Check |
display |
± 10 |
bits |
IkV step input |
|
|||||||
|
Auto |
applied 5 |
|
|
|
|
|
|
|
test. |
LO and |
|
||||
|
|
times |
|
|
|
|
|
|
|
GUARD must |
LO of |
|||||
|
|
|
|
|
|
|
|
|
|
|
|
connect to |
||||
|
|
|
|
|
|
|
|
|
|
|
|
Cal. and also |
to |
|||
|
|
|
|
|
|
|
|
|
|
|
|
mains ground. |
|
|||
|
|
|
|
|
|
|
|
|
|
|
|
7151 |
must uprange |
|||
|
|
|
|
|
|
|
|
|
|
|
|
withou power |
|
|||
|
|
|
|
|
|
|
|
|
|
|
|
restarts. It is |
||||
|
|
|
|
|
|
|
|
|
|
|
|
permissible |
that |
|||
|
|
|
|
|
|
|
|
|
|
|
|
the spark-gap |
|
|||
|
|
|
|
|
|
|
|
|
|
|
|
operates. |
|
|
0072g/1624g |
2.9 |
Table 2.3 |
Initial Calibration, Current |
|
|
|
|||
TEST |
RANGE |
& |
INPUT |
ACTION |
|
LIMITS |
COMMENTS |
1 |
MODE |
|
+1.00000A & |
Calibrate over |
the |
|
1 Ampere Set-up |
DC A |
|
|
|||||
|
|
|
open circuit |
bus |
|
|
|
2 |
AC A |
|
400 Hz |
Calibrate over |
the |
|
2 Ampere Set-up |
|
|
|
1.99999A & |
bus |
|
|
|
|
|
|
0.19999A |
|
|
|
|
3 |
|
|
|
Exit Cal Mode |
|
|
|
5 |
DC AC |
|
+1.99999A |
Measure voltage at |
0.80 volt |
Burden |
|
|
|
|
|
current front |
|
|
|
|
|
|
|
sockets with a |
|
|
|
|
|
|
|
DMM. |
|
|
|
0072g/1624g |
2.10 |
Table |
2.4 |
Initial Calibration, AC Volts |
|
|
|
|
|
|
||
TEST |
RANGE |
& |
INPUT |
ACTION |
|
LIMITS |
COMMENTS |
|
||
1 |
MODE |
|
s/c |
Adjust RV2 for |
|
±150 bits |
IC15 |
offset |
null |
|
VAC |
|
|
||||||||
|
various |
|
minimum @ TP3 |
|
referred to |
adjust. |
Use DMM |
|||
|
ranges |
|
|
|
OV |
to monitor |
TP3 |
|||
|
|
|
|
|
|
|
Transformer |
to be |
||
|
|
|
|
|
|
|
lamination |
|||
|
|
|
|
|
|
|
mains-grounded. |
|||
|
|
|
|
|
|
|
The lowest |
figure |
||
|
|
|
|
|
|
|
possible |
is |
|
|
|
|
|
|
|
|
|
required; |
if |
||
|
|
|
|
|
|
|
necessary by |
|||
|
|
|
|
|
|
|
error-sharing |
|||
|
|
|
|
|
|
|
among |
the |
ranges. |
|
2 |
20VAC |
|
19.9999V |
Note reading |
|
|
|
|
|
|
|
|
|
400Hz |
|
|
|
|
|
|
|
3 |
20VAC |
|
19.9999V |
Adjust CV1 St R10 |
Value at test Attenuator HF |
|||||
|
|
|
50kHz |
for flat response |
12' ±0.010V |
trim. |
100 bit |
|||
|
|
|
|
|
|
|
limit |
applies |
||
|
|
|
|
|
|
|
when a dummy lid |
|||
|
|
|
|
|
|
|
is fitted. |
|
||
4 |
0.2VAC |
0.199999V & |
Calibrate over |
the |
|
0.2V LF Set-up |
||||
|
|
|
0.019999V |
bus. |
|
|
Fluke |
5101. |
|
|
|
|
|
400Hz |
|
|
|
|
|
|
|
5 |
2VAC |
|
1.99999V & |
Calibrate over |
the |
|
2V LF Set-up |
|||
|
|
|
0.19999V |
bus. |
|
|
|
|
|
|
|
|
|
400Hz |
|
|
|
|
|
|
|
6 |
20VAC |
|
19.9999V S> |
Calibrate over |
the |
|
20V Set-up |
|
||
|
|
|
1.9999V |
bus. |
|
|
|
|
|
|
|
|
|
400Hz |
|
|
|
|
|
|
|
7 |
200VAC |
199.999V S. |
Calibrate over |
the |
|
200V Set-up |
|
|||
|
|
|
19.999V |
bus. |
|
|
|
|
|
|
|
|
|
400Hz |
|
|
|
|
|
|
|
8 |
IkVAC |
|
750.00V & |
Calibrate over |
the |
|
1 kV Set-up |
|
||
|
|
|
199.99V |
bus. |
|
|
|
|
|
|
|
|
|
400Hz |
|
|
|
|
|
|
|
9 |
|
|
|
Exit Cal Mode |
|
|
|
|
|
|
10 |
0.2VAC |
30KHz |
Check |
|
0.199999V ± |
|
|
|
|
|
|
|
|
0.199999V |
|
|
.000120V |
|
|
|
|
11 |
0.2VAC |
lOKHz |
Check |
|
0.199999V ± |
|
|
|
|
|
|
|
|
0.199999V |
|
|
,000096V |
|
|
|
|
0072g/1624g |
2.11 |
Table 2.4 Cent. |
|
||
TEST |
RANGE & |
INPUT |
ACTION |
12 |
MODE |
10kHz |
Check |
2VAC |
|||
|
|
1.99999V |
|
13 |
2VAC |
30kHz |
Check |
|
|
1.99999V |
|
14 |
20VAC |
30kHz |
Check |
|
|
19.9999V |
|
15 |
20VAC |
10kHz |
Check |
|
|
19.9999V |
|
16 |
200VAC |
10kHz |
Check |
|
|
199.999V |
|
17 |
200VAC |
30kHz |
Check |
|
|
199.999V |
|
18 |
IkVAC |
10kHz |
Check |
|
|
750.00V |
|
19 |
IkVAC |
30kHz |
Check |
|
|
750.00V |
|
20 |
0.2VAC |
s/c |
Check |
21 |
2VAC |
10Hz |
Check |
|
|
2.00000V |
|
22 |
2VAC |
20Hz |
Check |
|
|
2.00000V |
|
23 |
2VAC |
40Hz |
Check |
|
|
2.00000V |
|
24 |
2VAC |
100Hz |
Check |
|
|
2.00000V |
|
25 |
2VAC |
100kHz |
Check |
|
|
0.19999V |
|
26 |
20VAC |
100kHz |
Check |
|
|
20.0000V |
|
27 |
200VAC |
100kHz |
Check |
|
|
200.000V |
|
LIMITS
1.99999V ±
.00096V
1.99999V ±
.00120V
19.9999V ±
0.0120V
19.9999V ±
.0096V
199.999V +
.096V
199.999V ±
0.120V
750.00V +
0.46V
750.00V +
0.70V
150yV
2.00000V ±
0.01456V
2.00000V ±
0.00416V
2.00000V ±
0.00096V
2.00000V ±
0.00880V
0.199999V ±
0.000880V
20.0000V ±
0.0880V
200.000V ±
0.880V
COMMENTS
s/c zero. Trasnsformer laminations to be mains-grounded.
0072g/1624g |
2.12 |