7151
COMPUTING MULTIMETER
MAINTENANCE MANUAL
Issue 1: June 1984
SOLARTRON
Schlumberger
Solartron pursues a policy of continuous development and product improvement
The specification in this document may therefore be changed without notice
Solartron Instruments, Victoria Road, Farn bo rough
Hampshire, England GU14 7PW Telephone: Farnborough (0252) 544433
Telex: 858245 Solfar G Cables: Solartron Famborough
A division of Schlumberger Electronics (UK) Ltd
PartNo. 71510011
19©84
n
7151 COMPUTING MULTIMETER
MAINTENANCE MANUAL
1642g/0072g
CONTENTS
Chapter 1
Chapter 2
Chapter 3
Chapter 4
Chapter 5
Chapter 6
General
Calibration Procedures
Dismantling & Reassembly
Circuit Descriptions & Diagrams
Fault Diagnosis Guide
Parts Lists &
Component Layout
1642g/0072g
r
Chapter 1 General
1642g/0072g
f
CHAPTER 1
1. GENERAL
The Solartron 7151 Computing Multimeter performs all common
functions, and offers: a library of
measurements;
The instrument is suitable for general purpose bench work, or for use
within a system where 7151 would
control interfaces. The interfaces provided are the IEEE 488 (1978) STD
system and the RS232C V24 serial system.
2. SAFETY
The 7151 multimeter has been designed in accordance with the
recommendations of IEC 348. To ensure the user's safety, and the
continued safe operation of the instrument, it is advisable to fully
observe the procedures and specifications given in the Operating Manual
(Part No. 71510010).
An Earth wire is provided to ensure the user's safety. Therefore, if an
extension mains cable is used, check that the Earth connection is
maintained throughout the length of the extension.
When using 7151 on
(e.g. inductive circuitry giving high back emf's such as
a large mains transformer), it is most important that 7151's test leads
are disconnected from the equipment before
that harmful back-emf's do not reach 7151. Care should
exercised when handling the input leads, especially where
are known to be present, or where
and a programmable power-on status.
equipment which is capable
programs;
be operated via
high transients could
clock controlled
one of
of delivering
switched it off. This ensures
measurement
its remote
high voltages
the secondary
always be
high
occur.
voltages
of
Whenever it is likely that the safety of the instrument has been impaired
- e.g. if it shows visible signs of damage, if it fails to perform
correctly, or if the specifications have been exceeded in any way - it
should be made inoperative and referred to a suitable repair depot. Any
maintenance, adjustment or repair of the multimeter must be carried out
by
skilled personnel only, in accordance with the procedures and
precautions detailed in this Maintenance Manual (part no. 71510011).
A
Wherever this symbol appears on the front or rear panel it is
advisable to consult the appropriate section of
further information.
3. SUMMARY OF OPERATION
A schematic block diagram of the 7151 is shown in Fig. 1.1. 7151 is
essentially a
technique of analog to digital conversion.
All inputs to the instrument are first converted to dc voltages before
being passed to the input amplifier. This is simple enough for current
(dc) and resistance, but ac inputs also undergo rms conversion to dc.
All inputs are suitably scaled by the input amplifier and passed to the
voltage measuring instrument which uses the pulse width
the Operating Manual for
0072g/1634g
1.1
analog to digital converter (ADC). With no input,
balanced pulse trains of
received, the
opposite
then
manner, proportional to the size of the input. These trains are
converted to a single end and gated into a reversible counter. The
mark-space ratios of the trains respond in an
mark space ratio 1:1. When an input is
the
ADC produces two
equal and
nett result is a pulse count proportional to the measure of the input.
The measuring circuits are controlled by what is termed the 'floating'
logic and
'Piggyback' ROM. The other circuits of 7151 are organised in a
arrangement which is controlled by the 'earthy' logic and consists
essentially
between the floating and earthy logic is acheived by
links. It is the
of
measurements, processing, remote
displays, and so
consists essentially of a 8-bit microprocessor
with
of another 8-bit microprocessor. Isolated communication
opto coupled serial
earthy logic
which is responsible for effective control
control, the real time clock, the
on.
bus
0072g/1634g
1.2
A DRIVE
OHMS
I/P
TERMINALS »
CONTROL
A DRIVE
CONTROL
7151 MULTIMETER FUNCTIONAL BLOCK DIAGRAM
o-
INPUT SWITCHING
A.C.
CONDITIONING
INPUT
AMP
ATTEN. &
GAIN CONTROL
GAIN
CONTROL
FLOATING LOGIC
OPTO COUPLERS
CONVERTER
HOLD OFF
2 WIRE SERIAL
LINK
COUNT
RESET
+/-REF
WATCH-DOG
REFERENCE
AUTO-CAL
MEMORY
TO FLOATING
POWER
RAILS
TO EARTHY
POWER RAILS
FLOATING
POWER
SUPPLY
EARTHY
POWER
SUPPLY
ADDRESS BUS
ENABLE
WATCH-DPG
ADDRESS
LATCHES
At) TO A7
HOLD -OFF
RESET
A8 TO A15
EARTHY LOGIC
AO/DO TO A7/D7
4 COLUMNS
1 COLUMN
CLOCK
KEYBOARD 8.
FRONT/REAR SW.
DECODERS
INTERFACE
DECODERS
4 ROWS
8 ROWS
DATA BUS
SERIAL
KEYBOARD &
FRONT/REAR SW.
DECODERS
INTERFACE SW.
DECODERS
LATCHES
GPIA
DISPLAY
DRIVER
INTERFACE BUS
DRIVER
LEVEL CHANGING
TRANSCEIVERS
TO
■ REAR PANEL
GPIB SOCKET
ANALOG OUTPUT
TO
LEVEL CHANGING
TRANSCEIVERS
REAR PANEL
AUXILIARY SOCKET
7151 MULTIMETER FUNCTIONAL BLOCK
*
DIAGRAM
FIG 1.1
Chapter
Calibration
Procedures
1642g/00'72g
{
I
(
{
1
CHAPTER 2
Setting-Up And Calibration Procedures
General
These procedures enable the instrument to be set-up
factory
despatch
standards
.
and
calibrated to the
The procedures are categorised into the following sections:
1. Setting-up procedures
2. Initial calibration
3. Final calibration
procedures
procedures
Safety
The instrument must be disconnected from the mains supply when
dismantling it to gain access to the preset controls and also
being reassembled (see
Chapter 4 for dlssembly instructions)
when it is
.
When adjusting preset controls beware of high test voltages, the guard
potential on the guard plate and also the mains input supply.
Calibration Method
Owing to the automatic calibration circuits incorporated in 7151, it can
only be calibrated by connecting it to a remote controller and then using
the appropriate calibration commands.
program can be used which is a
Solartron
can
supply, on tape cassettes, a calibration program for the
much faster method of calibrating 7151.
Alternatively, a calibration
more common types of controllers.
The user is advised to re-calibrate 7151 annually.
If the instrument's existing
satisfactory, the user can
constants by sending the REFRESH
state of calibration is judged to be
simply re-write the existing calibration
command to 7151 once it is in the
calibration mode.
Calibration Source
It is recommended that the calibration source has an accuracy of at least
two times better than the accuracy specified for the various 7151
functions. The 7151 specification is given in the Operating
Manual and
the important percentage accuracies are as follows:-
DC Volts
DC Current
AC Volts
AC Current
Resistance
0.002%
0.02%
0.05%
0.05%
0.002%
0072g/1624g
2.1
ENTERING CALIBRATION MODE
Insert a shorted 2.5mm jack plug Into the rear panel
the front panel CAL Indicator to repeatedly flash. The short may be
CAL socket, causing
within the plug Itself, or externally via a switch. The plug must remain
fitted throughout the calibration, and can be removed after calibration
is complete.
Note: Do not switch mains power on or off when the shorting plug Is
fitted, otherwise the Internal calibration
Using the controller, send the command CALIBRATE
Into the
calibration
mode. The
CAL Indicator should then be steady.
Also displayed Is the word, 'CAL'. Once the calibration mode
constants may be
ON to 7151, putting It
altered.
has been
selected, the following conditions apply:
(a) Three commands cannot be used: TRIG
TRACK
NULL
'OFF' status Is adopted.
'OFF' status Is adopted,
all nulls being deleted.
(b) Four commands become available: HI
LO
WRITE
REFRESH
for refreshing
existing
cal. constants.
0072g/1624g
2.2
CALIBRATING MEASUREMENT RANGE
Using the controller, select the function and range to be calibrated by
sending the appropriate MODE and RANGE commands.
7151 must then be supplied with two precisely known reference inputs
(non-negative) one at approximately nominal full scale (referred to as
the Hi point), and
one at approximately zero (referred to as the Lo
point), in the case of ac ranges the Lo point should not be less than
approximately 5% of nominal full scale rather than zero. This ensures
that all inputs are within the optimum part of 7151's
linear range.
After a reference input is applied, 7151 must be informed of the precise
value of the input. This is achieved by using the HI command for a Hi
point, and the LO command for a Lo point. These commands must be
accompanied by an integer argument number, of up to six digits in length,
which expresses the applied input in terms of 5 x 9's count.
An integer value of 200000 corresponds to nominal full scale for any
range
For example, applying 2V on the 2V range, enter 200000
.
applying 20V on the 20V range, enter 200000
applying 5V on the 200V range, enter 005000
Apply the Hi point input to 7151 for the requisite function/range.
For example, 2.00843V on 2V dc range.
Using the controller, send the HI command to 7151.
For example, HI200843.
7151 responds by displaying 'Hi Pt' for about 1.5 seconds, during which
time it measures the applied reference input. When finished, the
instrument displays (and outputs) its measured count, e.g. 214576. It is
of no consequence if the displayed count differs from the applied input.
Repeat the above
(short circuit), and send the LO command. For example LOO (leading
zeroes need
Having
command
specified the Hi point and Lo point (in any order), send the
WRITE to 7151 (no argument required). This causes the
calibration constants for the selected range/function to be calculated
and
stored in memory. If successful, the message 'Good' is displayed.
If unsuccessful, an error message will be displayed and output to the
procedure for the Lo point. For example, reference = OV
not be specified).
controller.
Repeat the above instructions for each function/range to be calibrated.
0072g/1624g
2.3
RESTORING THE MEASUREMENT FUNCTIONS
Using the controller, send 7151 the command CALIBRATE OFF. The CAL
indicator will then flash indicating that the CAL shorting plug is still
fitted.
Withdraw the CAL
shorting plug. The
CAL indicator should then be
invisible, the instrument being ready for normal use.
SUMMARY
(a) Insert CAL shorting plug (2.
(b) Select the calibration mode by sending the CALIBRATE ON
5mm) in rear panel socket
.
command.
(c) Select the requisite function and range to be calibrated and perform
the calibration sequence. Repeat for each range/function to be
calibrated.
(d) De-select the calibration mode by sending the
CALIBRATE OFF
command.
(e) Remove CAL plug.
0072g/1624g
2.4
Setting-Up
Procedures
DC Power Supply Checks
Measure the dc supplies on PCB1 and PCB2 at the output pins of the
appropriate regulator IC's. Tolerances of the most important supplies,
mains voltage 240V, follow:
floating 15V unregulated
floating 15V regulated
floating 5V unregulated
floating 5V regulated
earthy 5V unregulated
earthy 5v regulated
between 20.7V and 21.6V
15 ± 0.75V
between 8.8V and 9.1V
5 + 0.25V
between 9.5V and 9.8V
5 ± 0.25V
Display Checks
The contrast of the display can be adjusted
by means of RV301. Make the
digits appear as black as possible but without introducing slurring when
a reading changes.
Keyboard Checks
The following sequence exercises all 16 keys.
Key Press
FILT 2 press minimum
Display Response
"FILT" on/off
Finish with "FILT" off
ma-
mA===
K£2
V~
V===
AUTO 2
V
A
LOCAL
presses
minimum
KQ
V~
V===
"AUTO" on/off
Check for downranging
Check for upranging
"GPIB nm" where nm is address
value.
NULL 2 presses minimuim
6x92
TRACK 2
presses
presses
minimum
minimum
SAMPLE with "HOLD" asserted
COMPUTE
MENU
"NULL" on/off
"6x9" on/off
"HOLD" on/off
“
HOLD" goes out briefly and
returns
.
"NO PROG"
“
PROBES?"
0072g/1624g
2.5
Initial
Calibration
Procedures
Test Equipment
1. General purpose DMM.
2. General purpose oscilloscope
3. Controller, e.g. Commodore PET fitted with BASIC III or BASIC IV
firmware.
4. Calibrator, e.g. Fluke 5101 fitted with GPIB interface.
5. ACV Calibrator, e.g. Hewlett-Packard 745.
6. ACV High Voltage
7. Capacitor O.lyF polypropylene attached to a twin 4mm banana
(3/4" centres)
Switch on 7151 and
Amplifier e.g. Hewlett-Packard 746.
.
allow to warm up for at least one hour before
plugs
calibration.
The initial calibration
Table
2.1
No.
2.2
2.3
2.4
procedures are detailed in the following tables:
Procedure
Initial
Initial
Initial
Initial
calibration, DC
calibration, Resistance
calibration. Current
calibration, AC Volts
Volts
Please
Note: The limits of error expressed in the following tables are
those adhered to by the factory for a new instrument. As an instrument
'ages', "components become more noisy or their tolerances increase.
Therefore, when calibrating a used instrument, it may
be
necessary to
accept limits of error that are marginally higher than those listed in
these pages. However, the instrument should always conform to the
commercial specification (see Operating Manual) after calibration.
0072g/1624g
2.6
Table 2.1 Initial Calibration, DC Volts
TEST RANGE & INPUT
ACTION
LIMITS
COMMENTS
MODE
1 Configure the rear panel interface switches and connect the controller to
7151.
Do
not insert the calibration Key Jack yet. FRONT/REAR switch to 'FRONT'.
2
3
4
5
6
7
2VDC
2VDC
0.2VDC
s/c link
VHI-VLO
front
s/c
VHI-VLO
Front
o/c
Power on
Insert Calibration
Key Jack
Adjust RV3.
DVM between link
& ROME
Check display ^
'for2scatter
Check reading
± 100 yV
3 adjacent
values
0 ± lOOyV
I/P amp gross
offset null.
2V range noise test.
The reading may jump
every 10 secs at
drift-correct
Input current
measurement.
may be exceeded at
drift-correct
.
Value
.
10
11
12
13
8
9
2VDC
2VDC
2VDC
2VDC
0.2VDC
20VDC
4V< plus
overload
< -100V
-4V< minus
overload
< -100V
±1,99999V
alternatively
+
1.99999V &
0.00000V
0.
199999V &
0.00000V
+19.
9999V &
0.00000V
Measure C4 with
DMM referred to
ROME
As above
Adjust RV1
CAL BAL
Do calibration
routine over the
interface
Do calibration
routine over the
interface
Do calibration
routine over the
interface
+3.90
+3.05
-3.05
-3.90
+ and -
equal within
1 bit
Positive input-clamp
test (D6)
Negative input-clamp
test (D26)
Cal. Bal Adjustment
Use continuous
drift-correct (Yl)
.
+2V set-up
+2V set-up. Use the
calibrator to deliver
0 volt
.
+20V set-up.
0072g/1624g
2.7
Table 2.1 Cent.
TEST RANGE & INPUT
MODE
14 200VDC +199.999V &
0.00000V
15 IkVDC +10000.
00V &
0.00V
16
17 2VDC +1.00000V
-1.00000V
ACTION
Do calibration
routine over the
LIMITS
COMMENTS
+200V set-up.
interface
Do calibration
routine over the
interface
+lkV Set-up. The
Calibrator LO and the
7151 LO input should
be mains grounded.
Check that the
spark-gap does not
operate. Apply for
1 minute and check
that the reading
does not drift more
than 2 bits.
Exit Cal Mode
Measure
+2 bits
pos-neg
error
Linearity. Change
polarity changing
over inputs.
0072g/1624g
2.8
Table 2.2 Initial Calibration, Resistance
TEST
1
2
3
4
5
6
7
RANGE
MODE
20kfl
200kfl
2Mfi
2k£2
20kfi
200kfi
2Mft
5. INPUT
DMM between
I- and LO
As above
As above
l.OOOOOkfl
and IQ
lO.OOOOkQ
and IQ
lOO.OOOkQ
and IQ
1.00000MQ
and IQ
ACTION
Measure current
LIMITS
100±5yA
COMMENTS
from -I.
DMM set to current.
As above
As above
Do calibration
10.0 ± 0.5uA
1.0 ± 0,5yA
2KQ range set up
using the interface
Do calibration
using the interface
Do calibration
using the interface
Do calibration
using the interface
20KQ range set
up.
200KQ range set
up.
2MQ range set up
O.lyF in
parallel will
reduce scatter
caused by series
mode interference.
10
11
12
8
20MS2
10.0000MQ
and IQ
Do calibration
using the interface
20MQ range set
up. O.lyF in
parallel will
reduce scatter.
9
2M$2
DMM across
7150 HI & LO
2M£2
240VAC
/ 50 Hz
2Mfi
1.00000MQ
Measure the o/c
volts from Q
source
.
Apply VHI-VLO
10 seconds.
Check after test 9
+5.2V ± IV
1.00000MQ
±100 bits
Q source
positive clamp.
Ohms overload test
Survival check for
damage after test
10.
DV
Auto
+1 kV
applied 5
times
Check display
± 10 bits
IkV step input
test. LO
GUARD must
connect to LO of
Cal. and also to
and
mains ground.
7151 must uprange
withou power
restarts. It is
permissible that
the spark-gap
operates.
0072g/1624g
2.9
Table 2.3 Initial Calibration, Current
TEST
1
2
3
5
RANGE
MODE
DC A
AC A
DC AC
& INPUT
+1.00000A &
open circuit
400 Hz
1.99999A &
0.19999A
+1.99999A
ACTION
Calibrate over the
LIMITS
COMMENTS
1 Ampere Set-up
bus
Calibrate over the
2 Ampere Set-up
bus
Exit Cal Mode
Measure voltage at
current front
sockets with
a
0.80 volt
Burden
DMM.
0072g/1624g
2.10
Table 2.4 Initial Calibration, AC Volts
TEST
1
2
3
RANGE &
MODE
VAC
various
ranges
20VAC
20VAC
INPUT
s/c
19.9999V
400Hz
19.9999V
50kHz
ACTION
Adjust RV2 for
minimum @
TP3
Note reading
Adjust CV1 St R10
for flat response
LIMITS
±150 bits
referred to
OV
Value at test
12'
±0.0
10V
COMMENTS
IC15 offset null
adjust.
to monitor TP3
Transformer
Use
DMM
lamination to be
mains-grounded.
The lowest figure
possible is
required; if
necessary by
error-sharing
among the ranges
Attenuator HF
trim. 100 bit
limit applies
when a dummy lid
is fitted.
.
10
4
5
6
7
8
9
0.2VAC
2VAC
20VAC
200VAC
IkVAC
0.2VAC
0.199999V &
0.019999V
400Hz
1.99999V &
0.19999V
400Hz
19.9999V S>
1.9999V
400Hz
199.999V S.
19.999V
400Hz
750.00V &
199.99V
400Hz
30KHz
0.199999V
Calibrate over the
bus.
Calibrate over the
bus.
Calibrate
bus.
Calibrate over the
bus.
Calibrate over the
bus.
Exit Cal Mode
Check
over the
0.
.000120V
199999V ±
0.2V LF
Fluke 5101.
2V LF Set-up
20V Set-up
200V Set-up
1 kV
Set-up
Set-up
11
0072g/1624g
0.2VAC
lOKHz
0.199999V
Check
2.11
0.
199999V ±
,000096V
Table 2.4 Cent.
TEST
12
13
14
15
16
17
18
19
RANGE &
MODE
2VAC
2VAC
20VAC
20VAC
200VAC
200VAC
IkVAC
Ik VAC
INPUT
10kHz
1.99999V
30kHz
1.99999V
30kHz
19.9999V
10kHz
19.9999V
10kHz
199.999V
30kHz
199.999V
10kHz
750.00V
30kHz
750.00V
ACTION
Check
Check
Check
Check
Check
Check
Check
Check
LIMITS
1.99999V ±
.00096V
1.99999V ±
.00120V
19.9999V ±
0.0120V
19.9999V ±
.0096V
199.999V +
.096V
199.999V ±
0.120V
750.00V +
0.46V
750.00V +
0.70V
COMMENTS
20
21
22
23
24
25
26
27
0.2VAC
2VAC
2VAC
2VAC
2VAC
2VAC
20VAC
200VAC
s/c
10Hz
2.00000V
20Hz
2.00000V
40Hz
2.00000V
100Hz
2.00000V
100kHz
0.19999V
100kHz
20.0000V
100kHz
200.000V
Check
Check
Check
Check
Check
Check
Check
Check
150yV
2.00000V ±
0.01456V
2.00000V ±
0.00416V
2.00000V ±
0.00096V
2.00000V ±
0.00880V
0.199999V ±
0.000880V
20.0000V ±
0.0880V
200.000V ±
0.880V
s/c zero.
Trasnsformer
laminations to be
mains-grounded.
0072g/1624g
2.12