Snuza Hero MD User Manual [es]

0 (0)

English

INSTRUCTION MANUAL

READ INSTRUCTIONS IN FULL BEFORE USE

HeroMD MNL Type1 V2 Jun16

English

Important: This portable baby breathing monitor may not be able to detect all life-threatening events.

SNUZA is a registered trademark of

Snuza International (Pty) Ltd

Unit 11 Roeland Square

Roeland Street

Cape Town 8001

South Africa

Made in South Africa

Patented

Advena Ltd.

Thorne Widgery House

33 Bridge Street

Hereford, HR4 9DQ, UK

Reg No. 3517275. VAT # GB702590065

Patent nos:

2008212520 | 2,677,990 | EP 2124748 |08709966.9/2124748 | 200300 |2009-548789 | MX/a/2009/008469 | 578560 | 2,009,134,887 | 2009/05652 | 2359364 | 8,461,996

 

 

 

English

CONTENTS

 

 

1.

IMPORTANT – READ FIRST! ..............................................................

1

2.

Introduction ............................................................................................

3

3.

Precautions and Contra-indications .......................................................

5

4.

Parts Identification ..................................................................................

6

5.

Quick Start Guide ...................................................................................

7

6.

Diaper (Nappy) Fitment .........................................................................

8

7.

Fitting your HeroMD (Medical Device) ....................................................

8

8.

Operation .............................................................................................

10

 

8.1

Startup ............................................................................................

10

 

8.2

Monitoring ......................................................................................

11

 

8.3

Clothing .........................................................................................

13

9.

Settings ................................................................................................

13

 

9.1

Audible Tick ....................................................................................

13

 

9.2

Rouse Warning...............................................................................

14

 

9.3

Movement-Rate Warning ...............................................................

14

10. Battery ...................................................................................................

15

 

10.1 Battery Indicator ...........................................................................

15

 

10.2 Battery Life ...................................................................................

15

 

10.3 Replacing the Battery ...................................................................

15

11.

Maintenance .........................................................................................

17

 

11.1 Disposal........................................................................................

17

 

11.2 Cleaning and Storage...................................................................

17

12.

Troubleshooting....................................................................................

18

13.

Disclaimer.............................................................................................

18

14.

Guarantee.............................................................................................

19

15.

Technical Description ...........................................................................

20

English

1. IMPORTANT – READ FIRST!

It is essential that HeroMD’s coloured flexible tip containing the breathing sensor maintains constant contact with baby’s tummy.

ALWAYS:

Use the correct size diaper (nappy). If it is too loose, HeroMD may lose contact.

Fold the top of the diaper over before attaching HeroMD to ensure a firm grip. Use the diaper adjusters to provide a snug fit around the tummy.

Firmly clip HeroMD onto the folded diaper waistband, using the full extent of the clip.

1

English

Ensure that HeroMD sits squarely on the diaper, and that the Breathing Sensor (1) is in contact with baby’s tummy.

Press the ON/OFF button (6) to start monitoring.

Always follow the above procedure to ensure that your baby’s breathing is being monitored, and to prevent false alarms caused by lack of contact with baby’s tummy.

NOTE:

Handle HeroMD with care to prevent damage to clip or sensor.

2

English

2. INTRODUCTION

Intended use:

HeroMD is a breathing monitor for the detection of infant breathing cessation.

Intended user:

This equipment is intended to be operated by the baby’s caregiver. The caregiver must have a good understanding of the instructions and safety measures contained in the user manual.

Usage environment:

Home Safety

HeroMD safely and conveniently monitors your baby's breathing movement while sleeping to provide an early intervention opportunity in the event of any interruption in baby’s breathing.

HeroMD should be clipped onto your baby's diaper (nappy) for total portability. No cords, wires, sensor pads, flat surface or external power are required.

Clip on, switch on and experience peace of mind.

HeroMD detects even the slightest abdominal movement and will alert you if your baby’s breathing rate falls to less than 8 breaths per minute (this warning can be disabled), or if no breathing is detected at all for a period of 20 seconds. HeroMD will first vibrate gently after 15 seconds in an attempt to rouse the baby.

3

English

If the vibration succeeds in rousing the baby and breathing is detected after such an event, on each detected breath the Breathing Indicator Light (3) will flash red. After three vibration/rouse events, the Rouse Warning will alert you to the fact that your baby’s breathing movements have stopped for 15 seconds on three occasions. The Rouse Warning is characterized by a short beep and flashing of the red Breathing Indicator Light (3) every 2 seconds. Monitoring will continue when the Rouse Warning is active but will require the caregiver to attend to the baby to investigate. The Rouse Warning alarm can be reset by briefly pressing either the ON/OFF or MODE/OFF buttons, or by turning the device off and then back on. The rouse warning feature can be disabled during monitoring by pressing the MODE/OFF button for 2 seconds indicated by a single beep, flashing of the red Breathing Indicator Light (3) and Alarm Indicator Light (4). When the device is turned off and on again, the settings revert to the rouse warning feature being enabled.

4

English

3.PRECAUTIONS AND CONTRA-INDICATIONS

This device is not intended to prevent or monitor a disease or condition such as Obstructive Apnea.

Movement from external sources such as being pushed in a pram or travelling in a vehicle or aircraft may be detected by HeroMD ’s sensor. HeroMD should not be relied upon where external sources of movement are present.

The monitored baby must sleep in his/her own separate bed as the monitor may detect the breathing or movement of any other person sharing the bed with the baby, and will not alarm as long as it detects that movement, even if the baby stops breathing.

Radio transmitters such as cell phones, audio and video baby monitors, cordless phones, walkie-talkies, wireless routers, etc. may cause interference with the monitor’s ability to detect breathing or movement. The monitor should not be placed within 2 meters of other transmitting devices.

No modification of this device is allowed.

A single device should be dedicated for use with one baby only.

Small parts that have become detached for some unlikely reason may result in a choking hazard.

In case your baby develops skin sensitivity where the device makes contact with your baby’s skin, refer to section 8.3. If the skin sensitivity persists, contact us via www.snuza.com.

Keep out of reach of pets, pests and other children as damage to the device might cause it to not perform as intended.

Do not insert anything into the Sound Emission Hole (2).

Do not allow any liquid to enter the Sound Emission Hole.

Avoid bending the breathing sensor unnecessarily (fig 1).

Do not use undue force to bend the clip (fig 2).

Do not pull the Battery Cover Release Clip (10) beyond the Guard Rail (11).

5

English

4.PARTS IDENTIFICATION FRONT

 

 

 

 

 

 

 

 

 

(1)

Breathing Sensor

 

 

 

 

 

 

 

 

 

(2)

Battery Indicator Light

 

 

 

 

 

(3)

Breathing Indicator Light

 

 

 

 

 

(4)

Alarm Indicator Light

 

 

 

 

 

 

 

 

(5)

Sound Emission Hole

 

 

 

 

 

 

 

 

 

(6)

ON/OFF Button

 

 

 

 

 

 

(7)

MODE/OFF Button

 

REVERSE

(12) Applied Part

(8)Diaper/Nappy Clip

(9)Battery Cover

(10)Battery Cover Release Clip

(11)Clip Guard Rail

6

English

5.QUICK START GUIDE

5.1Important: Use the correct diaper size and make sure it fits snugly (see Section 6 Diaper/Nappy Fitment).

5.2Do not switch the monitor on when clipped onto your baby’s diaper.

5.3Place the monitor on a flat, stationary surface. Switch on by briefly pressing the ON/OFF button (6).

5.4A start-up test will illuminate the indicator lights, vibrate the stimulator, play 3 audible tones, perform a battery test, and perform a sensor test.

5.5When the sensor has been tested, a startup tone is heard.

5.6Clip HeroMD onto the diaper so that the Breathing Sensor (1) is in direct contact with the baby’s tummy (see Section 7 Fitting your HeroMD).

5.7The Breathing Indicator Light (3) flashes on each movement. See Section 8.2 Monitoring for a full description.

5.8If breathing is not detected for 15 seconds, the vibrating stimulator will attempt to rouse the baby.

5.9If breathing resumes within 5 seconds after a vibration, the Breathing Indicator Light (3) will flash red to indicate that there has been a 15-second period of no breathing.

5.10After three vibrations, the Rouse Warning is activated – HeroMD beeps every 2 seconds and the Breathing Indicator Light (3) flashes red with every beep.

5.11If the period of no breathing reaches 20 seconds, the full audible alarm will sound, HeroMD will vibrate and the Alarm Indicator Light (4) will flash red.

5.12The caregiver must attend to the baby.

5.13The alarm can only be deactivated by switching the monitor off.

5.14To switch off the monitor, press both the ON/OFF (6) and MODE/OFF (7) buttons simultaneously. A shutdown tone will be heard and monitoring will stop.

7

English

6.DIAPER/NAPPY FITMENT

Important: Correct diaper fitment is essential for the operation of HeroMD. The thin flap of material at the top of the diaper must be folded over the thicker part before the diaper is closed (this is the section to which HeroMD will be attached). The diaper should fit snugly on the baby with the top of the diaper just below the belly button. Loose fitting diapers will gape, causing the monitor to lose contact with the baby and false alarm. A very ‘full’ or wet diaper may also cause it to gape. If baby sleeps on his or her side, HeroMD should be placed off-center of the navel and the mattress.

7.FITTING YOUR HEROMD

When attaching your Snuza monitor you should put your fingers in between the baby’s diaper and the baby’s skin. Make sure that the monitor is clipped securely onto the diaper and away from the baby’s skin when clipped on in order not to harm the baby.

Ensure that the Battery Cover (9) is secure. Clip your HeroMD all the way onto the diaper to the full extent of the clip, as shown:-

8

English

Make sure that the Breathing Sensor (1) is in contact with the baby’s tummy around the belly button area, although closer to the sides may suit different sleeping positions.

The World Health Organisation (WHO) recommends that babies be placed to sleep on their back unless advised otherwise by a doctor. See http://www.who.int/en/

9

English

8.OPERATION

8.1Startup

HeroMD must not be switched on while clipped onto your baby’s diaper. The monitor must be placed on a flat, stationary surface during the self-test.

Switch on by briefly pressing the ON/OFF button (6).

A startup-test will run which will test all indicator lights, the vibration stimulator, the audible alarm and the battery level. If a failure of any item is detected, the Self-Test Failure Alarm will sound and the device cannot be used.

After the battery test, the Battery Indicator Light will flash 3 times with simultaneous single-tone beep. The colour of the Battery Indicator Light has the following meaning:

 

Green

Battery OK

 

 

 

 

 

 

 

Battery nearing depletion – the monitor can

 

Yellow/Orange

be used but the battery must be replaced as

 

 

soon as possible.

 

 

Battery flat – HeroMD cannot be used,

 

Red

replace battery. The Low Battery Signal will

 

 

immediately commence.

If the battery indicator is either green or yellow/orange, HeroMD will proceed to test the sensor. The sensor self-test can last between 1 and 40 seconds, if the device does not detect zero movement.

The sensor test is indicated by a beep every 2 seconds and flashing of the red Breathing Indicator Light (3) and Alarm Indicator Light (4). The device must be placed on a stationary surface during this test. If the 40 second limit is reached then the self-test Failure Alarm will sound and the device cannot be used.

10

English

If the sensor test passes, a startup tune will be heard with simultaneous flashing of the green Breathing Indicator Light (3) indicating that monitoring has commenced and HeroMD can be positioned on the baby’s diaper/nappy

(see section 7 Fitting your HeroMD).

Warning: Ensure that the battery level is high enough for proper operation (i.e. green or orange light). Otherwise, DO NOT use the monitor. If the battery is too flat to operate, the Battery Indicator Light (2) will flash red accompanied by a short beep every second. Change the battery before further use (see Section 10.3 Replacing the Battery).

To switch OFF, press both the ON/OFF button (6) and the MODE/OFF button (7) simultaneously.

8.2Monitoring

During monitoring, the Breathing Indicator Light (3) provides information about your baby’s breathing, while the Alarm Indicator Light (4) alerts you to potential problems.

If there is no breathing detected then the alarm indicator light will flash red every 3 seconds. If breathing is detected then the Breathing Indicator Light

(3) flashes with each breath:-

 

Green on each

Breathing detected with no prior

 

breath

rouse event.

 

 

Breathing rate below 8 breaths per

 

 

minute. If the breathing rate

 

Yellow and

increases to over 8 breaths per

 

breathing rate alarm

minute, breathing indicator light will

 

(Dual-tone beep

reset to green (normal) and the

 

every second)

warning will stop. To disable this

 

 

function, see Section 9.3 Breathing

 

 

rate warning.

 

 

There has been at least one incident

 

Red on each breath

of no breathing for 15 seconds,

 

 

where breathing was detected after

11

 

 

 

 

English

 

 

 

 

 

 

 

 

 

 

 

 

 

the vibration (see below).

 

 

The Rouse Warning indicates that

 

Red and Rouse

there has been 3 vibration/rousing

 

 

 

 

Warning Alarm

incidents. To disable this function,

 

 

 

see Section 9.2 Rouse Warning.

 

The Alarm Indicator Light (4) signifies the following:-

 

 

 

 

 

 

 

 

If no movement is detected at all for

 

 

 

15 seconds, HeroMD gently vibrates

 

 

 

in an attempt to rouse the baby. The

 

 

 

Alarm Indicator Light (4) flashes red

 

 

Flashes red with

simultaneously with a beep. If

 

 

breathing movement is detected

 

(( ))

vibrating stimulus

 

again within the next 5 seconds, the

 

 

and 5 short beeps

 

 

Breathing Indicator Light (3) will flash

 

 

 

 

 

 

red on each subsequent breathing

 

 

 

movement. This is an indication that

 

 

 

there has been an incident of no

 

 

 

breathing movement for 15 seconds.

 

 

 

If no breathing movement at all is

 

 

 

detected for a further 5 seconds after

 

 

 

the vibration stimulus (20 seconds in

 

 

 

total), the Alarm Indicator Light (4)

 

))

Flashes red and

flashes red and the high-priority

 

audible alarm sounds. The operator

 

alarm sounds

 

must attend to the baby immediately.

 

 

 

 

 

 

The alarm can be deactivated by

 

 

 

turning HeroMD off completely (press

 

 

 

ON/OFF (6) and MODE/OFF (7)

 

 

 

buttons simultaneously).

 

12

English

NOTE: It is recommended that the parent / caregiver is trained to deal with situations where the baby has, for example, suffered from a prolonged apnea event.

8.3Clothing

HeroMD can be clipped onto the folded waistband of the diaper or the waistband of elasticated pants, as long as the Breathing Sensor (1) touches the baby in the tummy area.

The soft sensor tip on the monitor is made from medical grade silicone and is hypoallergenic. In the unlikely event that a red friction mark is noticed on the baby’s skin where the monitor was positioned, then we recommend that a thin layer of clothing is placed between the baby’s skin and the monitor, but ensure that the Sensor maintains contact. The position of the monitor must also be changed to a different location on the tummy, away from where the red mark was noticed and preferably closer to the side.

The monitor can also be used if baby is dressed in light clothing (for example, a Babygro) over the nappy or pants to which HeroMD is attached.

Too much clothing may attenuate or muffle the alarm. Make sure

the alarm can be heard if baby is covered in clothing and/or blankets.

9.SETTINGS

9.1AudibleTick

This feature causes HeroMD to emit an audible 'tick' on each breathing movement. To enable the Audible tick, ensure that HeroMD is switched ON, then press the MODE/OFF button (7) briefly (less than 1 second). To increase the volume of the tick, press the MODE/OFF button once again for medium volume and again for high volume. To turn off, briefly press the

MODE/OFF button a fourth time – HeroMD will give a louder beep. This 13

English

feature may be used with a baby sound monitor to enable you to ‘hear’ baby’s movements from a distance.

9.2Rouse Warning

During monitoring, if no breathing movement is detected for 15 seconds, HeroMD vibrates gently in an attempt to rouse the baby. If breathing movement resumes within the next 5 seconds, the Breathing Indicator Light

(3)will flash red on each subsequent breathing movement as an indication that there has been an event of no breathing for 15 seconds. With the Rouse Warning enabled, after 3 such vibration and successful rousing incidents, the Breathing Indicator Light (3) will flash red every 2 seconds with a simultaneous audible beep. This alerts you to the fact that your baby has stopped breathing for 15 seconds on three occasions and has then resumed breathing movement after HeroMD has vibrated.

Note:- Monitoring will continue while the Rouse Warning is sounding. The Rouse Warning is automatically enabled each time HeroMD is switched ON. To disable this feature, ensure that HeroMD is switched ON, then press and hold the MODE/OFF button (7) for at least 2 seconds until a tone is heard with a brief vibration and a simultaneous flash of the red Breathing Indicator Light (3) and Alarm Indicator Light (4). If you want to enable the warning again, switch HeroMD off completely (press both ON/OFF (6) and MODE/OFF (7) buttons simultaneously), then switch it ON again (the rouse warning is enabled by default when HeroMD is switched on).

9.3Breathing-rate Warning

The Breathing-rate Warning alerts you if there are less than 8 breathing movements per minute. The warning is identified by the Breathing Indicator Light (3) flashing yellow with a simultaneous dual-tone audible beep every second. This feature is automatically enabled each time HeroMD is switched ON. To disable this warning during monitoring, ensure that HeroMD is switched ON, then press and hold the ON/OFF button (6) for more than 2 seconds until there is a brief double-beep with short vibration and a

14

English

simultaneous flash of the yellow Breathing Indicator Light (3) and Alarm Indicator Light (4). If you want to enable the warning again, switch HeroMD off completely (press both ON/OFF (6) and MODE/OFF (7) buttons simultaneously), then switch it ON again. The Breathing-rate Warning is enabled by default when HeroMD is switched on.

10.BATTERY

10.1Battery Indicator

If the battery becomes depleted during monitoring, the Battery Indicator Light (2) will flash as follows:-

3 audible beeps

every 4 seconds and red flash every 2 seconds

The battery is nearing depletion.

Replace battery before further use (see Section 10.3 Replacing the Battery).

10.2Battery Life

The life of the replaceable battery depends entirely on usage. Battery life will vary significantly depending on the average daily duration of use and/or alarm duration and/or number of alarms. Turning on and off unnecessarily shortens the battery life. If the number of alarms is high, check diaper fitment.

Please note that the product guarantee does not apply to the battery. The typical operating time before the battery needs replacement is 4 - 6 months when used on average 8 hours per day.

10.3Replacing the Battery

For all MD models: use a CR2 (CR15270) battery available from most retail outlets (please see our website www.snuza.com for information about batteries and details of your national distributor).

15

Snuza Hero MD User Manual

English

10.3.1 Use your fingernail or a small flat implement. With a pinching action (Figure 2 and Figure 3), gently pull back the Battery Cover Release Clip (10) on the underside of the monitor and lift the Battery Cover at the same time (Figure 4).

Fig 1

Fig 2

Fig 3

Fig 4

10.3.2Remove the depleted battery by gently tapping HeroMD on your hand with the battery cover opening towards the ground.

10.3.3Insert the new battery, negative end first, as shown in Figure 5.

Ensure that the polarity is correct (align the positive end ‘+’ of the battery with the ‘+’ symbol in the battery housing, as indicated). Replace the Battery Cover and ensure that it is correctly closed (Figure 6).

16

English

Fig 5

Fig 6

 

Do not use the monitor without the Battery Cover in place.

10.3.4Test the unit by switching on and ensuring that the Battery Indicator Light (2) flashes green.

10.3.5If the unit starts alarming after the battery has been replaced, simply press the ON/OFF and OFF/MODE buttons simultaneously to turn it off.

NOTE: Incorrect placement of the battery may result in a safety hazard; please follow the instructions as above.

11.MAINTENANCE

11.1Disposal

The device, battery and packaging material must be disposed of in accordance with your country’s laws governing the disposal and recycling of batteries.

11.2Cleaning and Storage

Gently wipe with a damp cloth to remove any dirt. A mild soap may be used for stubborn dirt.

17

English

Do not use any other cleaning agents as they may contain damaging chemicals.

The monitor is splash resistant, not waterproof. Do not immerse in water. If it gets wet, dry it off immediately.

When changing the battery, dirt that is trapped between the battery replacement hatch and the case may be cleaned with a dry cotton bud once the Battery Cover (9) has been removed. Be sure that the rubber seal is free of dirt before replacing the Battery Cover, as this could compromise the seal.

When not in use, keep your HeroMD in its protective case. Do not store in a damp area.

12.TROUBLESHOOTING

Check incorrect fitment and placing of HeroMD on baby’s diaper (see Section 6 Diaper/Nappy Fitment and Section 7 Fitting your HeroMD).

Battery failure. The Battery Indicator Light (2) will show red (see Section 10.3 Replacing the Battery).

Please visit our website www.snuza.comfor further information.

13.DISCLAIMER

Customers use this monitor at their own risk and Snuza International (Pty) Ltd, its members, employees, agents, contractors, representatives, manufacturers, distributors and associates ("SNUZA") shall not be held liable for any loss, damage or injury suffered by the customer or any other person arising from, or in connection with, the use or manufacture of the monitor, including loss, damage or injury suffered by the customer or any other person as a result of any defects in respect of the monitor, and all warranties, and any other liability from whatsoever cause arising, are hereby excluded and the customer hereby indemnifies Snuza International (Pty) Ltd in respect of any such liability.

18

English

14.GUARANTEE

Snuza International (Pty) Ltd guarantees this product to be free from defects in materials and workmanship for a period of 12 months from the original date of purchase, verified by a sales receipt. Our guarantee liability shall be limited to replacing the unit free of charge and does not include transport costs. Our guarantee extends to all countries in which this product is supplied. By the act of operating this product, the user accepts all resulting liability. We reserve the right to modify the provisions of this guarantee at any time without notice. This guarantee does not cover the battery, incorrect application, tampering, misuse or transport damage.

19

English

15. TECHNICAL DESCRIPTION

 

 

Type BF applied part

 

 

 

 

 

CE marking according to Annex II of MDD

 

 

93/42/EEC

 

 

 

IP52

 

IP52 The enclosure of the monitor is

 

protected against the ingress of dripping

 

 

 

 

water

 

 

Device Serial Number (the first two

 

 

numbers indicate the year of

 

 

manufacture)

 

 

 

 

 

See the Disposal section for proper

 

 

disposal of the monitor

 

 

 

Typical service life

 

HeroMD has an expected service life of 2

 

 

years from the date of manufacture

Environmental Conditions

 

Permissible environmental operating

 

 

conditions:

 

 

· Temperature: 5 ºC TO 40 ºC.

 

 

· Humidity: 15% TO 93%, Non

 

 

Condensing

 

 

· Air pressure: 700HPA TO 1060HPA

 

20

 

 

DeutschEnglish

 

 

 

 

 

 

 

 

Permissible environmental

· 25 °C without relative humidity control;

 

conditions of transport and storage

and

 

after having removed the device

+70°C at a relative humidity up to 93%,

 

from its protective packaging and

non-condensing

 

subsequently between uses:

 

 

 

Power Source:

Internally Powered

 

Battery Specification:

Common name: CR2

 

 

IEC: CR17355

 

 

3.0 Volt

 

 

800 mAh

 

 

Size (mm): 26 x 15 x 15

 

 

Lithium

 

Sound pressure level:

75dBm at 10cm

 

Scan the following QR code to see the electromagnetic compatibility tables:

http://us.snuza.com/wp-content/uploads/2016/07/Snuza-HeroMD- Electromagnetic-Compatibility-Table.pdf

Contact Us:

For assistance with using or maintaining HeroMD or to report unexpected operation or events, contact us via:

Internet:

www.snuza.com/contact.php

Email:

support@snuza.com

Tel:

+27 (0)21 461 8530

Address:

Unit 11 Roeland Square, Roeland Street, Cape Town 8001,

 

South Africa

 

21

Français

Moniteur de respiration portatif

NOTICE D’UTILISATION

LISEZ ENTIEREMENT ET ATTENTIVEMENT LE MANUEL D’INSTRUCTION

AVANT LA MISE EN SERVICE

22

Français

Important : Ce moniteur de respiration portatif peut ne pas détecter certains dangers mortels.

SNUZA est une marque déposée et

Snuza International (Pty) Ltd

Unit 11 Roeland Square

Roeland Street

Cape Town 8001

South Africa

Fabriqué en Afrique du Sud

Breveté

Advena Ltd.

Thorne Widgery House

33 Bridge Street

Hereford, HR4 9DQ, UK

Reg No. 3517275. VAT # GB702590065

Les numéros de brevets :

2008212520 | 2,677,990 | EP 2124748 |08709966.9/2124748 | 200300 |2009-548789 | MX/a/2009/008469 | 578560 | 2,009,134,887 | 2009/05652 | 2359364 | 8,461,996

23

 

 

Français

CONTENU

 

 

1.

IMPORTANT – A LIRE AVANT TOUTE UTILISATION! ....................

25

2.

Introduction ..........................................................................................

27

3.

Précautions et contre-indications .........................................................

28

4.

Identification des Éléments ..................................................................

30

5.

Guide de Démarrage Rapide ...............................................................

31

6.

Installation de la Couche .....................................................................

32

7.

Positionnement de votre HeroMD .........................................................

32

8.

Functionnement ...................................................................................

34

 

8.1 Démarrage......................................................................................

34

 

8.2 Supervision ....................................................................................

35

 

8.3 Vêtements .....................................................................................

37

9.

Réglages .............................................................................................

38

 

9.1 « tic » Sonore .................................................................................

38

 

9.2 l’Avertisseur de Réveil par Stimulation...........................................

38

 

9.3 l’Avertisseur de Mouvement Faible ................................................

39

10. Batterie ..................................................................................................

40

 

10.1 Indicateur d’État de la Batterie .....................................................

40

 

10.2 Durée de Vie de la Batterie ..........................................................

40

 

10.3 Remplacement de Batterie ...........................................................

40

11.

Maintenance .........................................................................................

42

 

11.1 Destruction ...................................................................................

42

 

11.2 Nettoyage et Stockage .................................................................

43

12.

Dépannage ...........................................................................................

44

13.

Avis de non Responsabilité ..................................................................

44

14.

Garantie................................................................................................

44

15.

Description Technique..........................................................................

45

 

24

 

Français

1.IMPORTANT – A LIRE AVANT TOUTE UTILISATION!

Si la procédure ci-dessus est toujours respectée, vous pouvez être sûr que les mouvements abdominaux comportant le détecteur de respiration de votre bébé sont surveillés, et que les fausses alarmes causées par le manque de contact entre le capteur et le ventre du bébé seront évitées.

IL FAUT TOUJOURS :

Utilisez une couche de taille correcte. Si elle est trop large, le contact avec le détecteur HeroMD risque d’être insuffisant.

Pliez la partie supérieure de la couche avant de fixer le HeroMD afin de garantir une fixation ferme. Utilisez le système de réglage de la couche pour permettre une fixation ferme autour du ventre.

Fixez fermement le détecteur HeroMD sur la ceinture de la couche pliée en enfonçant totalement la pince du détecteur HeroMD.

25

Français

Vérifiez que le détecteur Hero est placé au centre de la couche et que le détecteur de respiration (1) est en contact avec le ventre de bébé.

Appuyez sur le bouton ON/OFF (6) pour activer le contrôle.

REMARQUE:

Manipuler l’appareil HeroMD avec précaution pour éviter d’endommager le détecteur ou le clip d’attache.

26

Français

2. INTRODUCTION

Usage prévu :

Snuza HeroMD est un moniteur respiratoire, qui détecte quand un bébé cesse de respirer.

Utilisateur prévu :

Cet appareil est prévu pour être utilisé par la personne s’occupant du bébé. La personne doit avoir une bonne connaissance des instructions et des consignes de sécurité indiquées dans le manuel d’utilisation.

Environnement d’utilisation :

En sécurité à la maison

HeroMD surveille les mouvements respiratoires de votre bébé pendant son sommeil, de façon pratique et sans danger, pour vous permettre d’agir très tôt en cas d’arrêt respiratoire.

HeroMD devrait être fixé sur le rebord supérieur replié de la couche de votre bébé pour une parfaite mobilité. Pas de fils électriques, des câbles, de fréquences radio ou d’alimentations électriques externes ne sont nécessaire.

Fixez le, allumez le et soyez serein.

HeroMD détecte le moindre mouvement abdominal et peut également vous avertir si les mouvements abdominaux le rythme respiratoire de votre bébé a une

27

Loading...
+ 72 hidden pages