Smeg WD6010TM User manual [GB]

WD6010
WD7010
WD7015
WD6010 – WD7010 – WD7015 WD – WASHER DISINFECTOR – Hospital sector
- USER MANUAL -
Translation of the original instructions issued in Italian
USER MANUAL
WD6010 – WD7010 – WD7015
Page 2 - 145
Summary
1 INITIAL RECCOMENDATIONS ..................................................................................................................... 7
2 SERVICE LIFE .............................................................................................................................................. 8
3 INTENDED USE AND CLASSIFICATION ........................................................................................................ 8
3.1 INTENDED USE ................................................................................................................................... 8
3.2 WD CLASSIFICATION .......................................................................................................................... 9
3.3 STANDARD 15883 ............................................................................................................................ 10
3.4 DEFINITION: “RESPONSIBLE BODY” OF THE DEVICE ........................................................................ 11
3.5 TYPE OPERATOR AND “SUPERUSER” ............................................................................................... 11
4 PRESENTATION ........................................................................................................................................ 12
4.1 WD6010 and GW6010 - MACHINE WITH SWINGING DOOR ........................................................... 13
4.2 WD7010 - GW7010, WD7015 – GW7015 – MACHINE WITH SLIDING DOORS ................................ 16
5 CONFIGURATIONS AND OPTIONALS ........................................................................................................ 21
5.1 WD6010 ........................................................................................................................................... 22
5.1.1 Standard models ...................................................................................................................... 22
5.1.2 Models equipped with Steam heating system ........................................................................ 23
5.2 WD7010 ........................................................................................................................................... 24
5.3 WD7015 ........................................................................................................................................... 25
6 SAFETY AND HANDLING PRECAUTIONS .................................................................................................. 26
6.1 KEY TO THE SYMBOLS USED IN THE MANUAL AND ON THE MACHINE .......................................... 26
6.2 GENERAL RECOMMENDATIONS ...................................................................................................... 28
6.3 RECOMMENDATIONS FOR TRANSPORT .......................................................................................... 30
6.4 ACCESSING AND REUSING THE DEVICE AFTER AN INCOMPLETE CYCLE ......................................... 32
6.5 USE OF THE RELEASE TROLLEY ........................................................................................................ 32
6.6 DOOR OPENING, DOOR CLOSING .................................................................................................... 33
6.6.1 WD6010 ................................................................................................................................... 33
6.6.2 WD7010 – WD7015 ................................................................................................................. 35
6.7 MANUAL DOOR UNLOCK ................................................................................................................. 37
7 OPERATING PROCEDURES ....................................................................................................................... 38
7.1 Access to detergent compartment .................................................................................................. 38
7.1.1 WD6010 - GW6010 .................................................................................................................. 39
7.1.2 WD7010 - GW7010 – WD7015 – GW7015 .............................................................................. 40
7.2 USE OF DETERGENTS ....................................................................................................................... 40
7.2.1 LIQUID DETERGENT SUCTION SYSTEM .................................................................................... 41
8 DETERGENTS ............................................................................................................................................ 42
9 LOAD PREPARATION FOR WASH CYCLE AND DISINFECTION................................................................... 44
10 BASIC OPERATIONS .............................................................................................................................. 47
USER MANUAL
WD6010 – WD7010 – WD7015
Page 3 - 145
10.1 HOME SCREEN – ID 1.0.0.0 .............................................................................................................. 47
10.1.1 DOOR CONTROL - MACHINES WITH HINGED DOORS ............................................................. 47
10.1.2 DOOR CONTROL - MACHINES WITH SLIDING DOORS ............................................................. 47
10.2 PROGRAMS COMPLETE MENU - ID.1.1.0.0 ..................................................................................... 49
10.3 FAVOURITE PROGRAMS MENU - ID.1.1.0.0 .................................................................................... 49
10.4 PROGRAM START - ID. 1.1.1.0 ......................................................................................................... 49
10.4.1 EXTERNAL PERIMETER BUTTONS AND DATA. ......................................................................... 50
10.4.2 “INTERNAL PERIMETER” SCREEN DATA ................................................................................... 51
10.4.3 SELECTED CYCLE OPTIONS – ID. 1.6.1.0 ................................................................................... 51
10.4.4 CYCLE START – “USER CHECK” FUNCTION ACTIVE .................................................................. 53
10.4.5 NOTES ...................................................................................................................................... 53
10.5 PROGRAM UNDER WAY - 1.1.1.3 .................................................................................................... 53
10.6 SUBPHASES IN PROGRESS ............................................................................................................... 56
10.6.1 PROGRAM IN PROGRESS “DISPLAY EASY”- 3.4.2.1 .................................................................. 56
10.7 COMPLETED PROGRAM - 1.1.1.4 ..................................................................................................... 57
11 SETUP OPERATION - 1.4.0.0 ................................................................................................................ 58
11.1 SET UP DATE AND TIME – ID.1.4.2.1 ............................................................................................... 60
11.1.1 ENTER PASSWORD – ID.1.5.0.4 ............................................................................................... 61
11.2 LANGUAGE SELECTION – ID. 1.4.2.2 ................................................................................................ 61
11.3 COUNTERS – ID. 1.4.6.1 ................................................................................................................... 62
11.4 ARCHIVE AND PRINTER - ID 1.4.6.2 ................................................................................................. 63
11.5 MACHINE PARAMETERS – ID. 1.4.2.0 .............................................................................................. 66
11.5.1 DETERGENT DOSAGE SETUP (“Chemicals dosing”) – ID. 1.4.2.3 ............................................. 67
11.5.2 PROGRAMS ACTIVATION (“Programs enable”) – ID. 1.4.6.0 - 1.4.6.4 ..................................... 68
11.5.3 PASSWORD – ID. 1.5.0.1 .......................................................................................................... 70
11.5.4 LAN PARAMETERS – ID. 1.4.8.0 – 1.4.8.2 ................................................................................ 71
11.5.5 FULL S/N, MAC, ID (device labels) – ID. 1.4.8.1 ....................................................................... 71
11.6 DIAGNOSTICS OPERATIONS – ID.1.4.3.0 ......................................................................................... 74
11.7 PROGRAMS EDITING – ID. 3.4.2.1 ................................................................................................... 76
11.7.1 Customizing the program name .............................................................................................. 79
11.7.2 Customizing the program icon ................................................................................................ 79
11.7.3 NEW PROGRAM PARAMETER: “PASSING / NOT PASSING” FLAG............................................ 80
11.8 DISPLAY AND SOUNDS – ID. 3.4.2.1................................................................................................. 82
11.9 ABOUT – ID. 1.4.5.0 – 1.4.5.1 .......................................................................................................... 83
11.10 USER PASSWORD – ID. 1503 ........................................................................................................ 83
11.11 HEATING TYPE – ID. 1465 ............................................................................................................ 85
11.12 WATER CONDUCTIVITY – ID.1466 ............................................................................................... 86
11.12.1 Purpose of the accessory ..................................................................................................... 87
USER MANUAL
WD6010 – WD7010 – WD7015
Page 4 - 145
11.12.2 Operation of the machine with the probe active: conductivity check function ................. 87
11.13 TROLLEY CHECK – ID.14A0 ........................................................................................................... 89
11.13.1 OPERATION WITH OPTIONAL ACTIVE – TROLLEY CHECK .................................................... 93
11.14 SPRAY ARMS CHECK - ID 14A0 ..................................................................................................... 95
11.14.1 SPRAY ARMS CHECK – CYCLE IN PROGRESS ........................................................................ 97
11.15 AUTOMATIC OPERATION - ID.146B ............................................................................................. 98
12 ALARMS and WARNINGS ..................................................................................................................... 99
12.1 WARNING – ID 1.1.1.2 ................................................................................................................... 100
12.2 ALARMS – ID 1.1.1.6 ...................................................................................................................... 102
13 CLEANING AND MAINTENANCE ........................................................................................................ 112
13.1 PRELIMINARY WARNING ............................................................................................................... 112
13.2 CLEANING OF THE DEVICE and its parts ........................................................................................ 112
13.3 Cleaning of the Filter Unit.............................................................................................................. 116
13.4 IN CASE THE DEVICE REMAINS UNUSED ....................................................................................... 117
13.5 DEVICE RE-USE AFTER A PERIOD OF IDLENESS .............................................................................. 118
13.6 TIME SCHEDULE FOR ROUTINE MAINTENANCE AND CHECKS ...................................................... 119
13.6.1 DAILY ...................................................................................................................................... 119
13.6.2 WEEKLY .................................................................................................................................. 119
13.6.3 EVERY SIX MONTHS ............................................................................................................... 119
13.6.4 EVERY YEAR ............................................................................................................................ 119
13.7 ELIMINATION OF SMALL TROUBLE ................................................................................................ 120
14 INSTALLATION .................................................................................................................................... 121
14.1 PACKAGE REMOVAL and PREPARATION FOR MOVEMENT ........................................................... 121
14.2 POSITIONING AND LEVELING......................................................................................................... 123
14.2.1 “SINGLE DOOR” MACHINE ..................................................................................................... 125
14.2.2 “DOUBLE DOOR” MACHINE ................................................................................................... 126
14.3 ELECTRICAL WIRING REGULATIONS .............................................................................................. 128
14.4 ACCESS TO THE ELECTRICAL CABLE AND TO THE WATER CHARGING HOSES ............................... 129
14.4.1 ACCESS FROM TOP OF DEVICE .............................................................................................. 129
14.4.2 ACCESS FROM REAR .............................................................................................................. 131
14.5 WATER SYSTEM SETUP .................................................................................................................. 134
14.5.1 WATER LOAD ......................................................................................................................... 134
14.5.2 DEMINERALIZED WATER........................................................................................................ 135
14.6 WATER DRAIN ................................................................................................................................ 136
14.7 STEAM SYSTEM CONNECTION (OPTIONAL) ................................................................................... 139
14.8 CHAMBER VENT CONNECTION (AIR AND STEAM) ........................................................................ 140
15 TECHNICAL CHARACTERISTICS ........................................................................................................... 141
16 MANUFACTURER REFERENCES AND AFTER-SALES SERVICE ............................................................. 145
USER MANUAL
WD6010 – WD7010 – WD7015
Page 5 - 145
USER MANUAL
WD6010 – WD7010 – WD7015
Page 6 - 145
10
07/04/2020
NEW PROGRAM PARAMETER: “PASSING / NOT PASSING” FLAG [ The firmware
manages this functionality starting from the Master CPU 11.8.0 version of January 2020. ].
09
19/09/2019
Cable type update.
08
17/05/2019
New WD6010 models provided with steam heating system. New functions ( Trolley check, Spray arms check, Automatic operation). New alarms and warning messages.
05
07/02/2018
Addition of model WD7015.
04
23/10/2017
This manual refers to FW CPU main master version 7.16.0: Some features may be absent in previous versions of the firmware.
01
14/02/2017
This manual refers to FW CPU main master version 6.5.0: Some features may be absent in previous versions of the firmware.
Rev.
Date
Note
USER MANUAL
WD6010 – WD7010 – WD7015
Page 7 - 145
1 INITIAL RECCOMENDATIONS
READ THE PRESENT INSTRUCTION MANUAL CAREFULLY:
This manual constitutes an integral part of the machine.
It must be kept intact and at hand for the entire life cycle of the machine. It is necessary to read the following manual carefully prior to using the device.
Failure to read or thoroughly comprehend the instructions in this manual may cause the device to malfunction and present hazards to the user.
Machine installation, maintenance, and repair operations must be performed by authorized technical staff. Repair operations performed by unauthorized staff, besides leading invalidating the product warranty, can be source of danger for the user. In case of replacement, an original spare part must be used.
If the device was not used in accordance to the instructions presented in this manual, the product may no longer be covered by warranty and the safety of this device could be compromised (ref. par 5.4.4 IEC 61010-1:2010).
The producer declines all liability for any use that differs from what indicated in the present manual.
Excluding possible manufacturing defects, consumable materials (Detergents, Air Filters, Thermal Printer Paper…) are not covered under warranty. The product warranty does not cover faulty parts due to negligence, inappropriate use, or failure to comply with the instructions of the device operations; incorrect installation or maintenance; repair and maintenance operations performed by unauthorized staff or with non-original spare parts; transportation damages; and any circumstance that cannot be ascribed to the device manufacturing defects. Moreover, the warranty does not cover operations related to installation and connection of alimentation and drain systems as well as maintenance operations provided in the instruction booklet. The installation of any accessory on the machine must be performed by authorized technical staff.
To request further information about accessories: contact your trusted seller and/or the
authorized technical assistance, using the contact details provided in this manual (ref. 5.4.4.c IEC61010-2-040:2015)
The content of this manual is for information purposes only. The content and the equipment described here may be subject to change without notification. The colours used in photographs of the product (finishing panels), in diagrams and in screenshots are all purely guideline.
USER MANUAL
WD6010 – WD7010 – WD7015
Page 8 - 145
2 SERVICE LIFE
Service life (or “lifetime”) of the device: 10 years or 50000 cycles (whichever limit is reached first), in conditions of normal use and regular maintenance, following the intervals and controls recommended in
this manual, carried out by technical staff authorized by the manufacturer.
“Service life” is considered the period of time when the device characteristics and performances are not
affected to such a degree that nor the efficacy of the treatment neither the safety of the operators are compromised.
3 INTENDED USE AND CLASSIFICATION
3.1 INTENDED USE
PRODUCT DESCRIPTION: Thermal disinfectors designed to wash and sanitize medical devices that can be reused, as well as other items used in medical, dental, pharmaceutical, veterinary and laboratory sectors.
SERIES
Key characteristics
WD6010
Large instrument washer with forced-air drying system and hinged door, available in both single-door and double-door ("pass-through") versions. Door opening type: hinged. Washing capacity: up to 12 DIN baskets. Max. power 13.0 kW.
WD7010
Large instrument washer with forced-air drying system and sliding door, available in both single-door and double-door ("pass-through") versions. Door opening type: sliding (vertical). Washing capacity: up to 12 DIN baskets. Max. power 13.0 kW.
WD7015
Large instrument washer with forced-air drying system and sliding door, available in both single-door and double-door ("pass-through") versions. Door opening type: sliding (vertical). Washing capacity: up to 18 DIN baskets. Max. power 20.0 kW.
WD INTENDED USE: treatment, through washing and thermal disinfection, of instruments inside hospitals and dental centers to prevent, thanks to thermal disinfection, the spread of infectious agents that could potentially harm personnel and final patients.
The treatment of the device is necessary for the subsequent correct sterilization of the load.
. The device has been validated according to ISO 15883-1, 15883-2, 15883-5 standards . Instruments used for the test: in compliance with paragraphs 15883-2 and 6.3.2.2 “Solid devices,” the following elements have been used to simulate the actual loading of surgical instruments: M12 x 100 EN 10088-2 screws and solid surgical instruments (Examples of “Solid devices”: surgical scissors, dissecting forceps, tissue forceps, retractors). . Validation is carried out using the chemical products recommended in this manual. Using other chemical products is not specified by the manufacturer and may impair the intended use of the device.
The appliance does not sterilize the load.
Sterilization is carried out by other devices (e.g. steam sterilizers); this procedure follows the washing and thermal disinfection phase. The washing cart functions in support of the load and determines the typology of instruments that can be processed in a cycle. (The base device is supplied without washing
USER MANUAL
WD6010 – WD7010 – WD7015
Page 9 - 145
carts; please contact the manufacturer for additional information on which carts best satisfy your needs). When choosing the most appropriate disinfection treatment: always follow the instructions provided by the instruments’ manufacturer. The operator’s competence and knowledge of the table of programs installed in the machine, as well as the typology of instruments to treat, is essential for the correct choice of the treatment cycle. The device offers options for several cycles, with different values of Time-temperature exposure (Ao parameter) that are determined according to the criticality of the medical devices to be processed.
Shared notion found in the sector’s technical literature: all “semi critical” instruments processed through thermal disinfection which can’t undergo a subsequent sterilization,
should be processed with an Ao at least equal to 3000 (achieved, for example, with a process type at 90°C 5min).
It is necessary to verify that the items to be treated and the selected washing cycle are compatible in respect to the maximum temperatures reached and chemical products
applied.
The elements processed in the machine that are not entitled to a subsequent sterilization must be perfectly dry in order to prevent the building up of bacterial colonies from residual condensation. If necessary, an extra drying phase can be started.
Information provided in compliance with paragraph. 5.4.4.r IEC61010-2-040:2015.
Use of this product for purposes not strictly outlined by this manual are considered improper. The manufacturer declines liability for improper use of this product.
The manufacturer declines liability for any potential damage caused by washing instruments for which the manufacturer has not authorized automatic decontamination through thermal disinfection.
3.2 WD CLASSIFICATION
Instrument washers for hospital use, medical devices in class IIb (in compliance with the classification criteria stated by the Directive 93/42 and later changes and additions, reviewed and amended by directive 2007/47/EC - attachment IX, rule 15).
USER MANUAL
WD6010 – WD7010 – WD7015
Page 10 - 145
3.3 STANDARD 15883
The reference standard for the operating performance of a thermal disinfector is EN ISO 15883.
With reference to paragraph 6 of standard 15883-1, besides being compliant with the device as supplied, at the manufacturer’s expense, the standard requires additional tests for the device as installed, in a process called validation. The user (responsible body) is responsible for the device validation.
VALIDATION
Compliant with the EN ISO 15883-1 standard, validation of the device occurs in three different phases:
1. Installation qualification – IQ
2. Operational qualification – OQ
3. Performance qualification – PQ
Installation qualification QI - Process of obtaining and documenting evidence that equipment has been provided and installed in
accordance with its specification.
Operational qualification OQ - Process of obtaining and documenting evidence that installed equipment operates within predetermined
limits when used in accordance with its operational procedures.
Performance qualification PQ - Process of obtaining and documenting evidence that the equipment, as installed and operated in
accordance with operational procedures, consistently performs in accordance with predetermined criteria and thereby yields product meeting its specification.
Note on the performance qualification
Performance qualification must be executed after the installation qualification and operational qualification phases have been successfully completed. Performance qualification is executed after introducing new or modified items to be cleaned and disinfected, or new loading systems, as long as there is demonstrated equivalence to a reference validated load or to an item or loading system previously validated.
Performance qualification is executed after introducing new procedural parameters (including chemical agents).
Operational qualification should be repeated at preset intervals to check periodically that the appliance operates correctly. It is recommended that the machine performance requalification is carried out every year.
WARNING – USE OF PROBES FOR VALIDATION
The base machine is not equipped with a physical door to insert probes with cables: we recommend the use of wireless probes (datalogger or similar).
USER MANUAL
WD6010 – WD7010 – WD7015
Page 11 - 145
3.4 DEFINITION: “RESPONSIBLE BODY” OF THE DEVICE
Responsible Body: “individual or group responsible for the use and maintenance of equipment”. Definition taken from the standard IEC61010-1:2010, par. 3.5.12. The responsible body should be clearly identified inside the structure where the appliance is operated (i.e. name and responsibility registered on company forms).
3.5 TYPE OPERATOR AND “SUPERUSER”
The machine can be only used by staff members who have been specifically trained.
INTENDED USER PROFILE: the characteristics of the “TYPE OPERATOR:
- staff specifically trained for the use of the appliance.
- staff trained in the thermal disinfection process and in the cycle of treatment of medical devices. It involves staff working in hospitals and dental centers, generally paramedics and nurses.
The installer technician is responsible for training staff when installing the appliance.
The manufacturer declines all liability in case of any malfunction or accident caused by use of the appliance by untrained staff. Staff preparation includes specifications on possible risks in using the appliance as well as training to perform operating procedures in the safest way possible. The installer technician must also communicate USER and SUPERUSER passwords to the responsible body in order to access the configuration parameters. The responsible body is responsible for preserving the passwords in a safe place.
It’s the RESPONSIBLE BODY’s task to make sure that whoever operates the appliance has been adequately
trained on its functioning, safe use and routine controls, and to ensure the continued training of all operators.
Staff training should be checked regularly.
The installer technician is responsible for the correct functioning of the appliance after the installation.
Safety information provided in compliance with 5.4.101.1 IEC61010-2-040:2015.
USER MANUAL
WD6010 – WD7010 – WD7015
Page 12 - 145
4 PRESENTATION
The products covered by this manual are available in two main versions:
1. Single door machine.
2. Double door machine.
In both versions the device is equipped with a touch screen user interface. The touch screen allows the user to interact with the interface by touching icons on the screen to perform actions based on the options available.
DEVICE VERSION
SINGLE DOOR MACHINE: the door to access the tank is used both to load the material to be processed
and to remove the finished load (after wash and thermal-disinfection treatment). In the “single door” version there is only one touch screen for all the user interactive operations: select
and start program, machine parametrization, door lock and unlock.
DOUBLE DOOR MACHINE: the device has two doors to access the tank:
1. Loading door unclean side”: used to load contaminated items to be treated with a wash cycle
and thermal disinfection in the device, in the washing cart.
2. Unload door or “clean side”: used to unload the treated, washed and thermally disinfected
items. Opening this door is only possible after a cycle has been successfully completed.
In the “double door” version there are two touchscreens: “unclean side” and “clean side”. “Unclean side”: through the interface, the operator can select a wash, start the cycle, block a cycle, and
establish machine parameters.
Clean side”: the operator can only unlock the door to unload items once the cycle has been completed
successfully. From the clean side it is also possible to interrupt a cycle in progress.
A cycle can be started only if the doors are closed.
USER MANUAL
WD6010 – WD7010 – WD7015
Page 13 - 145
4.1 WD6010 and GW6010 - MACHINE WITH SWINGING DOOR
DEVICE VERSION
IMAGE
SINGLE DOOR MACHINE
There is a small lower door to access the detergent compartment and the electricity supply disconnector.
DOUBLE DOOR MACHINE
IMPORTANT: in the double door version, on the load
side only, there is a lower door to access the detergent compartment and the electricity supply disconnector.
“LOADING” SIDE VIEW of the device.
TOUCH SCREEN INTERFACE
DOOR HANDLE
GLASS DOOR TO ACCESS THE WASH TANK
LOWER DOOR TO ACCESS DETERGENT COMPARTMENT
LOWER DOOR HANDLE
USER MANUAL
WD6010 – WD7010 – WD7015
Page 14 - 145
“LOADING” SIDE VIEW of the device: the lower small door to access the detergent compartment has been removed. The main starting switch can only be accessed with the door open.
In the double door version: the disconnector is only on the “unclean” side (“loading side”). USB PORT: Near the disconnector is the USB port, to be used by authorized technicians, for communication and device diagnostics. In the double door version: the USB port is only on the load side.
“UNCLEAN” SIDE (“LOADING” SIDE) VIEW of the device, the lower small door to access the detergent
compartment has been removed. The detergent jerrycans are not visible, but the following can be seen:
- detergent suction lances with level sensor
- air filters for drying system (dryer)
- peristaltic pumps (pumps for liquid detergent dosage)
DETERGENT
JERRYCANS
MAIN STARTING
SWITCH
(DISCONNECTOR)
OPTIONAL PRINTER
DETERGENT
PERISTALTIC
PUMPS
MAIN STARTING
SWITCH
(DISCONNECTOR
)
DRYER FILTERS
DETERGENT
SUCTION LANCES
USER MANUAL
WD6010 – WD7010 – WD7015
Page 15 - 145
“UNCLEAN” SIDE (“LOADING” SIDE) VIEW of the device: the extracted cart is positioned on the door with
stops to limit movement.
CART EXTRACTOR
HOLDER MOUNTED ON
THE DOOR
WASHING CART
WATER MANIFOLD. LOADING SIDE VIEW:
THE MANIFOLD MUST
WASHING CART
USER MANUAL
WD6010 – WD7010 – WD7015
Page 16 - 145
4.2 WD7010 - GW7010, WD7015 – GW7015 – MACHINE WITH SLIDING DOORS
The sliding doors travel up and down to allow the operator access to the washing chamber for loading and unloading of the instruments to be washed and disinfected. Door opening and closure are activated by the user with the aid of the touch screen interface.
WD7010, GW7010
VIEW OF "UNCLEAN" SIDE (LOADING SIDE) of the device.
MAIN SWITCH ON-OFF SWITCH
OPTIONAL PRINTER
SAFETY KEY FOR SOLE USE OF AUTHORISED TECHNICAL STAFF
TOUCH SCREEN INTERFACE
USB PORT
USER MANUAL
WD6010 – WD7010 – WD7015
Page 17 - 145
WD7015, GW7015
VIEW OF "UNCLEAN" SIDE (LOADING SIDE) of the device.
On the WD7015, GW7015 models the touch screen interface is on the side control panel.
MAIN SWITCH ON-OFF SWITCH
OPTIONAL PRINTER
SAFETY KEY FOR SOLE USE OF AUTHORISED TECHNICAL STAFF
TOUCH SCREEN INTERFACE
USB PORT
USER MANUAL
WD6010 – WD7010 – WD7015
Page 18 - 145
Loading the chamber using the multi-purpose trolley.
Loading side opening the door at the bottom gives access to:
- detergent jerrycan compartment (detergent jerrycans and intake nozzle)
- dryer air filters
- peristaltic pumps (liquid detergent dispensing pumps).
WARNING: If the detergent compartment is open, the loading side door cannot be operated.
PERISTALTIC PUMPS
DRYER FILTERS
DETERGENT JERRYCAN COMPARTMENT
USER MANUAL
WD6010 – WD7010 – WD7015
Page 19 - 145
Unloading or "clean" side. Devices:
- "Mushroom" emergency stop button.
- Touch screen display.
Emergency stop button: allows the user on the unloading (or "clean") side to halt the machine. For example, the button can be pressed if the user notices a malfunction that justifies stopping the current cycle at once, or to halt movement of the doors.
The touch screen display on this side of the machine is read-only; the user can view the active screens but not interact with the machine. The only function enabled is door opening/closure. The door can only be opened at the end of the cycle, and only if the wash cycle has been completed successfully. (The user may have halted the drying phase: in this case opening of the clean side door is possible).
WARNING: once the “emergency button” has been pressed, it must be rearmed (by turning it) in order to restart the device. The device should also be switched off and back on with the main switch before restarting normal use.
EMERGENCY STOP BUTTON
USER MANUAL
WD6010 – WD7010 – WD7015
Page 20 - 145
The following optional data communication ports are available on the TOP of the device:
- LAN PORT (optional) – for network communications using software supplied by the manufacturer
- RS232 optional for remote printer (the port can only be provided if the device is not fitted with the thermal panel printer). See the specification for the optional “WD-PRINTEINK” printer. For the connection ports positions, refer to the “Installation Requirements" documents.
USER MANUAL
WD6010 – WD7010 – WD7015
Page 21 - 145
5 CONFIGURATIONS AND OPTIONALS
Symbols meaning
feature present
optional feature, installable on the model.
-
feature not present and not installable on the model.
WARNING – CHEMICAL AGENTS DOSAGE
Chemical agents dosage is performed through peristaltic pumps. For each installed peristaltic pump the following control procedures are strongly encouraged:
1. Chemical dosage check– through flow sensors “flowmeter”;
2. Tank level check – through level sensors
Only when the control systems are installed can the device verify whether the dosage used complies with the default setting and can warn the user of low dosage levels.
Control devices can be also installed upon use by technical staff authorized by the employer. Please contact your trusted dealer for more information and clarifications.
Notes on variants - Power supply:
Unless otherwise specified, the device is set for a power supply of three-phase plus neutral with 400V between phases (400V 3N~). This configuration is also referred to as "standard" (“std”).
USER MANUAL
WD6010 – WD7010 – WD7015
Page 22 - 145
5.1 WD6010
The model is available in different versions, according to the installed elements. Common elements in versions: tank material AISI316, tempered glass doors, appearance and interface (touch screen),
single washing pump with soft start, single dryer engine, floor drain, 3 water pipe connections., Door type “Drop down”. Washing capacity: up to 12 DIN baskets.
5.1.1 Standard models
FEATURES
ID. CODE
860457
860459
860649
860536
860460
860566
860558
860458
860568
860652
Model name
WD6010M
WD6010TM
WD6010M3
WD6010TM6
WD6010
WD6010-6
WD6010-36
WD6010T
WD6010T6
WD6010M6
Door type (Drop down/ sliding)
Drop
down
Drop
down
Drop
down
Drop
down
Drop
down
Drop
down
Drop
down
Drop
down
Drop
down
Drop
down
Doors number (tank access: single/ double)
Single
Single
Singola
Single
Double
Double
Double
Double
Double
Single
Detergent peristaltic pump P1


Acid neutraliser peristaltic pump P2


Optional peristaltic pump P3
        

Optional peristaltic pump P4
        

Check detergent dosage P1 – FM1


Check acid neutraliser dosage P2 – FM2
        

Check detergent dosage P3 – FM3
        

Check detergent dosage P4 – FM4
        

Check tank level P1 – SL1


Check tank level P2 – SL2


Check tank level P3 – SL3
        

Check tank level P4 – SL4
        

INTEGRATED printer
        

LAN connection
        

USB serial port
        

Demineralised water pump (pressure re­launch)
        

Integrated drain pump for wall drain
        

Drying absolute filter HEPA H13
        

Lighting tank kit “LED6010”
        

Vapours condenser
        

Warm water boiler
-  -  - - -
-
Demineralized water boiler
        

Mains frequency
50 Hz
50 Hz
50 Hz
60 Hz
50 Hz
60 Hz
60 Hz
50 Hz
60 Hz
60 Hz
Notes for electrical connection
std
std
230V 3~
380-400V
3N~
std
380-400V
3N~
220-230V
3~
std
380-400V
3N~
380-400V
3N~
Steam heating system
- - - - - - - - -
-
USER MANUAL
WD6010 – WD7010 – WD7015
Page 23 - 145
5.1.2 Models equipped with Steam heating system
FEATURES
ID. CODE
860743
860744
860742
860745
Model name
WD6010MV
WD6010TMV
WD6010V
WD6010TV
Door type (Drop down/ sliding)
Drop down
Drop down
Drop down
Drop down
Doors number (tank access: single/ double)
Single
Single
Double
Double
Detergent peristaltic pump P1


Acid neutraliser peristaltic pump P2


Optional peristaltic pump P3
  
Optional peristaltic pump P4
  
Check detergent dosage P1 – FM1


Check acid neutraliser dosage P2 – FM2
  
Check detergent dosage P3 – FM3
  
Check detergent dosage P4 – FM4
  
Check tank level P1 – SL1


Check tank level P2 – SL2


Check tank level P3 – SL3
  
Check tank level P4 – SL4
  
INTEGRATED printer
  
LAN connection
  
USB serial port
  
Demineralised water pump (pressure re-launch)
  
Integrated drain pump for wall drain
  
Drying absolute filter HEPA H13
  
Lighting tank kit “LED6010”
  
Vapours condenser
  
Warm water boiler
-  -
Demineralized water boiler
  
Mains frequency
50 Hz
50 Hz
50 Hz
50 Hz
Notes for electrical connection
std
std
std
Std
Steam heating system
  
USER MANUAL
WD6010 – WD7010 – WD7015
Page 24 - 145
5.2 WD7010
The model is available in different versions, according to the installed elements. Common elements in versions: tank material AISI316, tempered glass doors, appearance and interface (touch screen), single washing pump with soft start, single dryer engine, floor drain, 3 water pipe connections., Door type: “Sliding”. Washing capacity: up to 12 DIN baskets.
CHARACTERISTICS
ID. CODE
860462
860464
860465
860463
860746
Model name
WD7010M
WD7010TM
WD7010
WD7010T
WD7010T6
Door type (Drop down/ sliding)
Sliding
Doors number (tank access: single/ double)
Single
Single
Double
Double
Double
Detergent peristaltic pump P1

Acid neutraliser peristaltic pump P2

Optional peristaltic pump P3
    
Optional peristaltic pump P4
    
Check detergent dosage P1 – FM1

Check acid neutraliser dosage P2 – FM2
    
Check detergent dosage P3 – FM3
    
Check detergent dosage P4 – FM4
    
Check tank level P1 – SL1

Check tank level P2 – SL2

Check tank level P3 – SL3
    
Check tank level P4 – SL4
    
INTEGRATED printer
    
LAN connection
    
USB serial port
    
Demineralised water pump (pressure re­launch)
    
Integrated drain pump for wall drain
    
Drying absolute filter HEPA H13
    
Lighting tank kit “LED6010”
    
Vapours condenser
    
Warm water boiler
-  -

Demineralized water boiler
    
Steam heating

Mains frequency
50 Hz
50 Hz
50 Hz
50 Hz
60 Hz
Notes for electrical connection
std
std
std
std
STD
USER MANUAL
WD6010 – WD7010 – WD7015
Page 25 - 145
5.3 WD7015
The model is available in different versions, according to the installed elements. Common elements in versions: tank material AISI316, tempered glass doors, appearance and interface (touch screen),
double washing pump with soft start, double dryer engine, floor drain, 3 water pipe connections., Door type: “Sliding”.
Washing capacity: up to 18 DIN baskets.
CHARACTERISTICS
ID. CODE
860481
860482
860479
860480
860687
860689
860737
Model name
WD7015M
WD7015TM
WD7015
WD7015T
WD7015-3
WD7015-6
WD7015T6
Door type (Drop down/ sliding)
Sliding
Doors number (tank access: single/ double)
Single
Single
Double
Double
Double
Double
Double
Detergent peristaltic pump P1

Acid neutraliser peristaltic pump P2

Optional peristaltic pump P3
     
Optional peristaltic pump P4
     
Optional peristaltic pump P5
     
Check detergent dosage P1 – FM1

Check acid neutraliser dosage P2 – FM2
     
Check detergent dosage P3 – FM3
     
Check detergent dosage P4 – FM4
     
Check detergent dosage P5 – FM5
     
Check tank level P1 – SL1

Check tank level P2 – SL2

Check tank level P3 – SL3
     
Check tank level P4 – SL4
     
Check tank level P5 – SL5
     
INTEGRATED printer
     
LAN connection
     
USB serial port
     
Demineralised water pump (pressure re­launch)
     
Integrated drain pump for wall drain
     
Drying absolute filter HEPA H13
     
Lighting tank kit “LED6010”
     
Vapours condenser
     
Warm water boiler
-  -  -
-
Demineralized water boiler
     
Steam heating

Mains frequency
50 Hz
50 Hz
50 Hz
50 Hz
50 Hz
60 Hz
60 Hz
Notes for electrical connection
std
std
std
std
230V 3~
380-400V
3N~
380-400V
3N~
USER MANUAL
WD6010 – WD7010 – WD7015
Page 26 - 145
6 SAFETY AND HANDLING PRECAUTIONS
6.1 KEY TO THE SYMBOLS USED IN THE MANUAL AND ON THE MACHINE
Below are the symbols used on the machine and in this manual, in compliance with par. 5.4.4.e IEC61010-2­040:2015.
SYMBOL
MEANING
Particular attention must be paid when reading this.
This symbol indicates particularly important regulations or recommendations.
“ON” Power
Standardized symbol (5007 IEC 60417-1): it is on the disconnector (main switch) to indicate connection to the electrical supply.
“OFF” Power
Standardized symbol (5008 IEC 60417-1): it is on the disconnector (main switch) to indicate no connection to the electrical supply.
Manufacturer
The symbol appears on the product’s technical dataplate, accompanied by the name and address of the medical device's Manufacturer. (Symbol 5.1.1 ISO 15223-1, ISO 7000 no.3082)
Consult instructions for use.
The symbol appears on the product’s technical dataplate to indicate that users must consult the
instructions for use of the device. (Symbol 5.4.3 ISO 15223-1, ISO 7000 no.1641)
Warning, danger: consult the manual.
This symbol is placed on the specific plate of the product to highlight that it is necessary to read the manual before using the device. This symbol is found next to the safety regulations as well.
Warning, warm surface. This symbol is placed on the machine next to parts that can overheat and present burning hazards – avoid any contact with parts next to the symbol. This symbol is also found in the manual to highlight the safety regulations regarding burning hazards.
Warning, risk of electric shock. The symbol is placed next to live parts – it is necessary to disconnect from the electrical supply source before carrying out any operations on these parts. Avoid any contact with live parts if electricity is on.
Specific notes and regulations for Hospital sector.
USB port – data communication and device setup
Transport, storing and unpacking regulations. Biohazard.
Fire danger.
Take care to avoid all liquid spills. This symbol is affixed near external parts of the device which must be protected from accidental spills and drips.
USER MANUAL
WD6010 – WD7010 – WD7015
Page 27 - 145
SYMBOL
MEANING
The symbol of a “crossed out wheelie bin” (attachment IV directive 2002/96/EC): placed on the specific label and package of the product: at the end of its service life, the product must be sent to the disposal facilities for recovery and recycling, in compliance with the regulations in force in the country where it is installed. Contact the specialized disposal installations. The management of WEEE (Waste of electrical and electronic equipment) is in the hands of manufacturers, whose duty is to plan and manage the collection system; today it is possible to deliver the waste directly to the dealer for free, when purchasing a device of the same type. If purchasing a new equivalent device to replace the previous one, the disposal fees are at the expense of the manufacturer of the new device. At the end of its service life the machine might be contaminated, in particular the tank and water circuit (e.g. end of service life caused by a malfunction which compromised the efficiency of the last thermal disinfection cycle): use proper caution in the disposal operations. By making sure that this product is disposed of properly, the user contributes in preventing any possible negative effects on the environment and health.
The disposed device must be made unusable. Remove the power cable after unplugging it / after disconnecting the cable from the power outlet.
CE marking, IMQ notified body (CE0051 on model series “WD” only, medical devices). The symbol is on the specific label of the machine and in this manual to identify a medical device with
CE certification released by IMQ (“0051” is the identification number of the notified body IMQ).
USER MANUAL
WD6010 – WD7010 – WD7015
Page 28 - 145
6.2 GENERAL RECOMMENDATIONS
MAXIMUM LOADING WEIGHT
Leaning or sitting on the device door when open may cause it to overturn with a
consequent danger for people. The main function of the door is not to support unnecessary weight.
Devices with hinged doors: The maximum weight that can be loaded on the door must never exceed 60 kg.
Apart from the cart and the instrument baskets, the maximum weight of the load to be treated must never exceed:
- WD6010, WD7010: 30kg
- WD7015: 45Kg
To obtain an optimal DRYING, the load to be treated must never exceed:
- WD6010, WD7010: 30kg.
- WD7015: 45kg.
RELEASE TROLLEY
Sliding door devices: The multi-purpose trolley is a necessary device for the safe handling of
washing carts: for inserting the cart in the chamber and removing it.
Cart loading and removal operations performed without the aid of the release trolley constitute "misuse". MALFUNCTIONS
In the event of malfunction (e.g. water leaks or unexpected operation) disconnect the device from the electrical source and turn off the water taps. Refer to the “ALARMS” section of this
manual and, if necessary, contact the closest authorized Service Center.
The machine can be used only by staff properly trained on its functioning. The machine tank was not designed for the operator to enter inside. The operator must never enter completely inside the tank – this might compromise its safety (rif. 7.102 IEC61010-2-040:2015). If the display (loading or unloading side) remains off or unresponsive even with the main switch correctly set ON: do not touch the display as this may trigger unexpected operation of parts of the device. Switch the device off using the main switch. Wait at least 10 seconds after switching off before switching back on again. Refer to the "ELIMINATION OF SMALL TROUBLES" section.
DOOR OPENING
The door is equipped with safety door-locking mechanism. Do not force the opening manually. It is possible to unlock the door with the touch interface of the device and it is
permitted only when the washing or drying cycle is not in progress.
MACHINE WITH SLIDING DOORS
DO NOT APPROACH THE DOORS during automatic opening or closure.
The machine is fitted with doors which are opened/closed automatically. Before enabling door opening/closure, check that other operators are not in a potentially hazardous position. Never attempt to open the door with a program running. The appliance's control system prevents this for safety reasons. Never attempt to force the door open by hand.
USER MANUAL
WD6010 – WD7010 – WD7015
Page 29 - 145
The tops of the doors are fitted with pressure-sensitive edges: if the sensitive edge encounters an obstacle while the door is in motion, door travel is first stopped and then reversed to prevent crushing hazards. Pressure-sensitive edges:
1. considerably reduce the crushing hazard for the user;
2. reduce the possibility of closing the door with objects obstructing its travel.
The user must still take the greatest care during automatic operation of the doors, and never:
1. allow any incorrectly positioned object to obstruct free travel of the doors;
2. touch moving parts.
The user is unable to open or close the door on the opposite side of the machine. Nor is it possible to open both doors simultaneously. The tank of the appliance is not designed for users to climb into, even for extraordinary maintenance. The user must never climb into the tank - this might put his safety at risk.
DETERGENT COMPARTMENT DOOR
The bottom door, giving access to the detergent compartment, must be closed correctly to enable automatic door opening and closure.
Do not open the bottom door during automatic door operation.
The bottom door must be correctly fitted and closed to allow use of the machine: safety devices are installed to restrict operation of the machine if the door is not correctly positioned.
INTERNAL TANK LIGHTING - OPTIONAL "LED6010” KIT
The appliance's lighting system is classified in the "EXEMPT" risk category (under the IEC 62471 standard): there is no photobiological risk in normal conditions of use.
DEMINERALIZED WATER
The demineralized water connection is required to install the device.
In case demineralized water is not available, it is the user’s responsibility to verify that the
quality of the water supplied to the device does not cause the deposit of mineral salts or other substances.
FLAMMABLE SUBSTANCES
Do not introduce flammable substances to the device. Do not use flammable detergents.
Do not introduce alcohol or solvents such as turpentine, which might cause an explosion. Do not introduce materials dirty with ash, wax or paint.
USER MANUAL
WD6010 – WD7010 – WD7015
Page 30 - 145
CHEMICAL DISINFECTION
Warning: if possible, thermal disinfection is always more advisable than a chemical disinfection process. According to regional regulations, the decontamination process with
chemical agents might be considered invalid by the responsible authorities, and can be used only if the loaded items cannot bear the specific temperature of thermal disinfection. The suggested washing and chemical disinfection cycles are specifically designed for the recommended products and can be inappropriate with other chemical products. Do not use products that were not specified by the manufacturer.
The authority responsible for the device and the decontamination process is also responsible for the selection of the most appropriate treatment cycle.
CLOG WASHING AND DECONTAMINATION
The device where clogs are washed and decontaminated should be assigned to this specific function, in order to avoid mistakes of use. Clogs require specific supports and specific cycles,
which are different from those used to process instruments.
6.3 RECOMMENDATIONS FOR TRANSPORT
The machine is initially packaged: it is positioned and secured on wooden pallets, wrapped in protective nylon for dust and water infiltrations and covered in cardboard. Do not use devices which were damaged during transport. STORING: The device must be preserved in a closed environment, not in the open air, in temperature and moisture conditions similar to those specified for its functioning.
Transport of the machine to the place of its actual installation is at the client’s expense
ad requires a lift truck.
Refer to the product’s technical sheet to know its net weight and gross weight with
packaging.
When using the lift truck the product must be secured with belts to prevent it from overturning in the event of braking.
Unpacking operations:
1. Take the external strapping off the packaging
2. Open and remove the cardboard and the nylon wrapping
3. Do not place the machine on its side
4. Remove the polystyrene corner protectors
5. The machine base is fastened to a pallet to be lifted and transported. When the
machine is placed in the position for actual operation, the pallet holders, made with self-tapping wood screws, must be removed.
6. Place the machine on a horizontal plane with a maximum gradient of 2°; level it
using the adjustable feet.
7. Do not place the machine on a flammable surface.
8. Do not use the door handles to transport the product.
Manual transportation of the product must be avoided as far as possible. The use of a tail lift is recommended. If manual transportation is necessary: involve at least 3 people or more to transport the product. Use a “ramp” to take the device off the pallet more easily. Take care to secure the ramp to the wooden pallet to prevent unexpected repositioning.
Loading...
+ 115 hidden pages