SLE SLE1000 User Manual

Page 1
SLE1000
User manual
Version 2.6.4 software
Page 2
Contact Information:
SLE Limited Twin Bridges Business Park 232 Selsdon Road South Croydon Surrey CR2 6PL
Telephone: +44 (0)20 8681 1414
Fax: +44 (0)20 8649 8570
E-mail: Sales/Training enquires: [email protected]
Web site: www.sle.co.uk
Technical enquires: [email protected]
All rights reserved. No part of this publication may be reproduced, stored in any retrieval system, or transmitted in any form or by any means, electronic, mechanical, photocopy, recording or otherwise, without prior permission of SLE. © Copyright SLE 10/08/2016.
Page 2 of 128
Page 3
This manual is for SLE1000 CPAP Therapy
System
How To Use The
SLE1000 CPAP Therapy System
The warnings on pages 19 to 22 must be read and understood before using the SLE1000 CPAP Therapy System. Failure to do so could lead to injury of the patient.
1 SETUP THE SLE1000: Page 70
2 MAKE SURE THAT THE SLE1000’s BACK-UP POWER
SUPPLY IS FULLY CHARGED: Page 77.
3 THE SLE1000 IS READY FOR PATIENT CONNECTION AND
USE. Page: 88
For more information see “Technical Section”, page 106.
For trouble shooting see “Troubleshooting Chart”, page 90.
Page 3 of 128
Page 4
This page is intentionally blank.
Page 4 of 128
Page 5
Contents
1. Introduction .............................................................................................................10
1.1. Device ............................................................................................................10
1.2. Accessories and generic combination devices...............................................11
1.3. Intended use...................................................................................................11
1.4. Intended users................................................................................................11
2. SLE1000 Description ..............................................................................................14
3. Description of symbols ..........................................................................................16
4. User/owner responsibility ......................................................................................18
5. Warnings ..................................................................................................................19
5.1. Operational warnings .....................................................................................19
5.2. Clinical warnings.............................................................................................22
6. User interface ..........................................................................................................24
6.1. Main LCD display ...........................................................................................24
6.2. LED display windows......................................................................................24
6.3. Push and turn knob ........................................................................................25
6.4. Audio paused and boost buttons....................................................................25
6.5. Light bar .........................................................................................................25
7. Operation .................................................................................................................26
7.1. Power up ........................................................................................................26
7.2. Oxygen calibration routine description ...........................................................29
7.3. Power down....................................................................................................30
8. Standby mode .........................................................................................................31
8.1. Exiting Standby mode ....................................................................................31
9. Selecting a parameter, display or calibration function .......................................32
9.1. Upper and lower level functions. ....................................................................32
9.2. Upper level functions, displays and parameters.............................................37
9.3. Showing battery charge level .........................................................................39
9.4. Alarm history ..................................................................................................42
9.5. Set Flow .........................................................................................................44
9.6. Elapsed time counter......................................................................................46
10. Audio mute button ................................................................................................47
10.1. Pre-mute.......................................................................................................47
10.2. Audio paused ...............................................................................................48
11. Boost button ..........................................................................................................50
12. Indicators ...............................................................................................................51
12.1. Cycle indicator bargraph...............................................................................51
12.2. Breath detection ...........................................................................................51
12.3. Averaged value ............................................................................................52
12.4. Measured values..........................................................................................52
12.5. Battery and mains indicators........................................................................53
Page 5 of 128
Page 6
12.6. Power related faults......................................................................................53
13. CPAP & Flow mode ...............................................................................................56
13.1. Standby to Flow mode..................................................................................56
13.2. Setting a flow in flow mode...........................................................................58
13.3. Flow mode to CPAP mode............................................................................61
13.4. Standby to CPAP mode................................................................................63
13.5. CPAP to Flow mode......................................................................................65
14. SLE1000 Set-up .....................................................................................................70
14.1. First use........................................................................................................70
14.2. Rotation of gas inlets ....................................................................................72
14.3. Water trap and particulate filter.....................................................................72
14.4. Standard setup..............................................................................................73
14.5. Disconnection of air and oxygen hoses........................................................76
14.6. Battery charging............................................................................................77
14.7. Battery care...................................................................................................77
14.8. Using the SLE1000 on battery power...........................................................77
14.9. Mains isolation..............................................................................................78
14.10. Extended storage and transportation..........................................................78
15. Functional testing prior to patient use ...............................................................80
15.1. Battery test....................................................................................................81
15.2. Mains disconnect alarm................................................................................81
15.3. Total power fail alarm test.............................................................................82
15.4. High leak and high pressure alarm ...............................................................82
15.5. Apnoea alarm................................................................................................82
15.6. Zero pressure alarm and purge check..........................................................83
15.7. Gas supply alarms........................................................................................83
16. Operational considerations ..................................................................................86
16.1. General.........................................................................................................86
16.2. CPAP............................................................................................................86
17. Basic operation .....................................................................................................88
17.1. Monitoring.....................................................................................................88
17.2. Operation......................................................................................................88
18. Troubleshooting chart ..........................................................................................90
Technical Information 93
19. Cleaning and disinfection .....................................................................................94
19.1. Preparation of a new SLE1000 CPAP driver................................................94
19.2. Cleaning and disinfection of an in-service driver ..........................................94
19.3. Cleaning Method...........................................................................................95
19.4. Disinfection method ......................................................................................95
20. Messages & alarms ...............................................................................................96
20.1. Messages......................................................................................................96
20.2. Alarm messages and sounds........................................................................98
Page 6 of 128
Page 7
21. Technical specification ........................................................................................106
21.1. Operating mode............................................................................................106
21.2. Monitoring.....................................................................................................106
21.3. Indicators......................................................................................................106
21.4. Controls........................................................................................................106
21.5. Alarms ..........................................................................................................107
21.6. Outputs.........................................................................................................107
21.7. Oxygen sensor life........................................................................................107
21.8. Gas supplies.................................................................................................108
21.9. Operating environment.................................................................................108
21.10. Power requirements ...................................................................................109
21.11. Dimensions.................................................................................................109
21.12. Classification ..............................................................................................109
21.13. Environmental storage conditions ..............................................................109
21.14. Product life cycle ........................................................................................109
22. Maintenance schedule ..........................................................................................110
22.1. SLE1000 service requirements ....................................................................110
22.2. Maintenance schedule .................................................................................110
22.3. Preventative Maintenance............................................................................111
23. EMC compliance ...................................................................................................112
23.1. Electromagnetic immunity ............................................................................113
23.2. Recommended separation distances between portable and mobile RF
communications equipment and the SLE1000......................................................115
24. RS232 .....................................................................................................................116
24.1. RS232 output specification:..........................................................................116
25. Pneumatic unit schematic SLE1000 ....................................................................117
26. SLE1000 Overlay ...................................................................................................118
27. Equipment checklist .............................................................................................119
28. Consumables and accessories ...........................................................................122
29. Glossary of abbreviations used in this manual .................................................124
30. Index .......................................................................................................................125
Page 7 of 128
Page 8
This page is intentionally blank.
Page 8 of 128
Page 9
Introduction
Page 9 of 128
Page 10

1. Introduction

1.1 Device

The SLE1000 is a CPAP driver for delivering an air-oxygen treatment gas to patients via nasal prongs or nasal mask. The SLE1000 is intended for providing CP AP to treat infant and neonatal patients with compromised respiratory systems while in an acute or sub-acute hospital environment.
The SLE1000 can be operated from either wall serviced, compressor or bottled pressurized air and oxygen sources in the range 2 to 7 bar.
The SLE1000 has been designed so that it can be used with only a compressed air supply. The SLE1000 automatically corrects the flow rate, to maintain the set CPAP or maintains the
set flow. ‘Typical’ flow-CPAP relationship is shown by the centre line in the nomogram below. This is
the relationship produced by common Nasal Generators.. The line also represents the SLE1000 default calibration settings.
The highlighted area in the nomogram below represents the whole operating range. The SLE1000 will alarm if the patient leak or the exhaust impedance (block) is excessive and causes the automated flow correcting to reach a boundary of the operating zone.
Page 10 of 128
Page 11

1.2 Accessories and generic combination devices

The SLE1000 can be used to drive CPAP through a generic humidified and heated patient circuit, with or without a bacteria filter, to any universal patient interface, including nasal prongs, flow generator, or nasal mask that operate within tolerance of chart 1.

1.3 Intended use

The SLE1000 is intended to provide pressure support up to 12 mbar, to neonates and infants. The device is intended for use in the hospital environment where care is provided by medical professionals and includes use at bedside and for intra hospital transport provided compressed gas is available.

1.4 Intended users

The SLE1000 is to be used by appropriately trained and authorized personnel only. In this manual they will be referred to as the “User”.
Page 11 of 128
Page 12
This page is intentionally blank.
Page 12 of 128
Page 13
SLE1000 Description
Page 13 of 128
Page 14

2. SLE1000 Description

Light bar
Main LCD display
Battery icon Mains disconnect
Fresh gas port
% Oxygen
Proximal airway port
CPAP Boost button
Alarm mute button
Main control knob
icon
display
Set pressure
display
Set apnoea
time display
Front View
Page 14 of 128
Page 15
Mains power inlet
RS232
Water trap
Air supply
inlet
Oxygen supply
inlet
Mains Fuse Holder
Exhaust port
Connector
Battery
Fuse Holder
Mains power
switch
Equipotential stud
Oxygen cell cover
Rear View
Page 15 of 128
Page 16

3. Description of symbols

Symbol Description
Read manual (Situated on rear panel).
Date of Manufacture (Appears on serial number label).
Do not dispose of as general waste (WEEE directive). (Appears on serial number label).
EEC conformity marking showing compliance with the Medical Devices Directive.
Symbol for degree of protection against electric shock. Type B applied part. (Appears on front panel)
Alarm mute.
CPAP boost.
Indicates the mains power switch.
Equipotential point.
Mains power not present, (when illuminated).
Battery icon
Indicates a warning in the manual.
Indicates a note in the manual.
Page 16 of 128
Indicates a caution in the manual.
Page 17
Operational and Clinical Warnings
Page 17 of 128
Page 18

4. User/owner responsibility

The SLE1000 and the authorised accessories for it are designed to be used in accordance with supplied manuals and instructions. This equipment must be periodically checked, recalibrated, maintained (as per maintenance schedule) and components repaired and replaced when necessary for the equipment to operate safely and reliably.
Parts that have failed, in whole or in part, or exhibit excessive wear, or are contaminated, or are otherwise at the end of their useful life, should not be used and must be replaced immediately with parts supplied by SLE, or parts which are otherwise approved by SLE.
Equipment which is not functioning correctly or is otherwise in need of repair or maintenance must not be used until all necessary repairs and/or maintenance have been completed and a factory authorised service representative has certified that the equipment is fit and ready for use. This equipment, its accessories or component parts should not be modified. The use of non-approved parts or accessories will invalidate the warranty.
The owner/user of this equipment shall have the sole responsibility and liability for any damage or injury to persons or property (including the equipment itself) resulting from operation not in accordance with the operating instructions, or from faulty maintenance not in accordance with the authorised maintenance instructions, or from repair by anyone other than the factory authorised service representative, or from unauthorised modification of the equipment or accessories, or from the use of components or accessories that have been either damaged or not authorised for use with this equipment by the manufacturer.
Page 18 of 128
Page 19

5. Warnings

5.1 Operational warnings

The following warnings must be read and understood before using the SLE1000. Failure to do so could lead to injury or death of the patient.

5.1.1 General

1. The whole of this manual should be read and understood before using the SLE1000. Operators must be suitably trained and clinically authorised for using the SLE1000 with patients.
2. Oxygen - Clinical use. Oxygen is a drug and should be prescribed as such.
3. Oxygen - Fire Hazard. Oxygen vigorously supports combustion and its use requires special precaution to avoid fire hazards. Keep all sources of ignition away when oxygen is in use. Do not use oil or grease on oxygen fittings or where oxygen is used.
4. The SLE1000 is not a dedicated transport device and does not conform to the requirements of BS EN 794-3.
5. Always disconnect the air and oxygen hoses from the wall outlets prior to disconnection from the unit.
6. Do not use in the presence of flammable anaesthetics.
7. Check the condition of the gas supply hoses to the SLE1000. Do not use any hose that shows signs of cracking, abrasion, kinking, splits, excessive wear or ageing. Make sure that the Air or O2 hose has not come into contact with oil or grease.
8. Audible and Visual warning alarms indicate a potentially harmful condition to the patient.
9. The patient must not be connected to the unit in “Standby” mode. The unit does not deliver any therapeutic flow in this modes. In “St andby” mode all p atient and gas alarms are disabled with the exception of the “Low battery” alarm and all system failure alarms.
10. When the SLE1000 is being used on a patient, a suitably trained person must be in attendance at all times to take prompt action should an alarm or other indication of a problem occur.
11. The SLE1000 must not be used as the sole form of patient monitoring. Independent monitoring of blood gases should be used in conjunction with the SLE1000. Breath detection technology used in the apnoea monitor function is based on rate and magnitude of change in the pressure measured in the patient interface (that is shown on the bar graph). The sensitivity of the technology has limitations with the smallest patients (<1000g) and if a leak exists or if the patient opens/breathes by the mouth. The apnoea monitor may be disabled if breath non-detection / over detection results in nuisance Apnoea / High BPM alarms.
12. At low CP AP pressures (<4mbar) it is possible for the unit not to detect a leak condition, even in the presence of large leak.
Page 19 of 128
Page 20
13. An alternative form of ventilation should be available whenever the SLE1000 is in use, in the event of failure.
14. Do not touch the patient and SLE1000 metalwork at the same time to avoid earthing the patient.
15. The SLE1000 is not compatible with MRI equipment or other equipment radiating high EM fields.
16. The SLE1000 contains temperature dependant devices which perform normally in controlled environments in hospitals. However if the SLE1000 has been stored at a temperature different to that in which it will be used, allow the unit to acclimatize before powering up, if difference greater than 10°C allow 2 hours. (See “Environmental storage conditions” on page 109).
17. Any computer connected to the SLE1000 must be specified for medical use.
18. The SLE1000 does not use Latex, nor was it used in its construction.
19. The SLE1000 is not protected against the ingress of water. (IP20)
20. Do not cover the SLE1000 or allow the SLE1000 to become covered by any fabric or curtain. Do not allow the exhaust ports or inlet vents to become obstructed or blocked by positioning the SLE1000 near curtains or fabric.
21. The SLE1000 does not support automatic record keeping.
22. The SLE1000 has no emergency air intake.
23. The SLE1000 functional test routine must be carried out each time the SLE1000 is used on a patient. If any of these tests do not function as described then there is a problem and the SLE1000 must not be used until it is rectified.
24. Failure to comply with the recommended service programs could lead to injury of the patient, operator or damage to the SLE1000. It is the owners responsibility to ensure that the equipment is regularly maintained.
25. The SLE1000 must be connected to a suitably rated and grounded electrical power source.
26. The SLE1000 must not be started on battery power alone.
27. The SLE1000 must have its batteries charged at least every 10 days. The SLE1000 must be connected to a mains power source for 8 hours to ensure a full charge. Failure to charge the batteries will shorten the service life of the batteries which will result in reduced battery life in a mains power fail situation.
28. The SLE1000 can operate on battery power in the event of mains power failure for a maximum of 4 hours (If battery fully charged at start of mains power failure). Whilst on battery power an alternative form of ventilation should be made ready if the mains power supply cannot be restored.
Page 20 of 128
Page 21
29. If the SLE1000 is operating on battery power and allowed to completely discharge the battery, it should be recharged a soon as possible. This is to avoid damaging the batteries by allowing them to remain in a deep discharged state for any period of time. Allowing the batteries to remain in the deep discharged state will shorten the life of the batteries and the amount of time the SLE1000 can remain on battery power.
30. If the SLE1000 is not to be used for a period in excess of 10 days then disconnect the batteries from the power supply. (See “Extended storage and transportation” on
page 78).
31. If the SLE1000 is adversely affected by equipment emitting electromagnetic interference then that equipment should be switched off or removed from the vicinity of the SLE1000. Conversely, if the SLE1000 is the source of interference to other neighbouring equipment, it should be switched off or taken to another location.
32. The functioning of this machine may be adversely affected by the operation of equipment such as high frequency surgical (diathermy) equipment, defibrillators, mobile phones or short-wave therapy equipment in the vicinity.
33. Do not clean the SLE1000 whilst the unit is in operation.
34. Disposal of the Oxygen cell should be in accordance with local regulations for hazardous substances. Do not incinerate. SLE offers a cell disposal service.
35. The SLE1000 contains one sealed lead acid battery. At the end of its useful life this battery should be disposed of in accordance with local authority guidelines. Note: the battery should only be replaced by service personnel.
36. Apart from the battery and O2 cell, the SLE1000 and accessories do not contain any hazardous components therefore no special precautions are required for their disposal.
The SLE1000 should be disposed of in accordance to the local WEEE (Waste Electrical and Electronic Equipment) guidelines.
37. Care should be taken when attaching other equipment as this may affect mechanical stability.
38. In the event of main LCD screen failure remove the patient as soon as possible (under clinical supervision) from the SLE1000 to an alternative form of ventilation. The SLE1000 will continue to deliver CPAP on the set parameters whilst this is arranged.
39. Do not use solvent based cleaning solutions to clean the SLE1000.

5.1.2 Humidifier / Patient Circuit

40. The patient circuit tubing should conform to EN12342 (ISO5367:2000) (non kink requirements).
41. On no account should antistatic or electrically conductive tubing be used.
42. The humidifier used in the patient circuit must be operated and maintained in accordance with its manufacturer’s instructions.

5.1.3 Compressor

43. When using a compressor the user must use the correct start-up routine otherwise the watertrap will bleed air. (See section 14.4.3 on page 75).
Page 21 of 128
Page 22

5.2 Clinical warnings

Failure to take corrective action when the alarms are activated could result in injury to the patient.
There are risks inherent in the use of ventilatory support in the neonatal and infant patient. These may include:
1. Damage to nares.
2. Under- or over- ventilatory support (with consequent abnormalities in blood gases);
3. Incorrect humidification; could cause mobilisation of secretions and airway blockage.
4. Damage to trachea and bronchi;
5. Over- or under- inflation of the lung;
6. Atalectasis;
7. Air leak syndrome (pneumothorax, pneumomediastinum, pneumopericardium. pulmonary interstitial emphysema).
Page 22 of 128
Page 23
User interface
Page 23 of 128
Page 24

6. User interface

1
2
3
4
5
The SLE1000 user interface comprises of one LCD display (1), three LED displays (2), one push and turn knob (3), two push buttons (4) and light bar (5).
The SLE1000 is controlled solely through items 3 and 4.

6.1 Main LCD display (1)

This display shows the measured bpm, mean airway pressure, flow, oxygen concentration and airway pressure bar graph. It also displays the function names, alarm messages, alarm history and battery status.
Note: The functions shown in this display are referred to as upper level functions and display.

6.2 LED display windows (2)

The LED displays show the set values for the Apnoea alarm, CP AP pressure and percentage oxygen.
Note: The functions shown in this display are referred to as lower level functions.
Page 24 of 128
Page 25

6.3 Push and turn knob (3)

The push and turn knob is the sole method of selection and adjustment of driver parameters.

6.3.1 Operation

To select a parameter rotate the knob in either direction (clockwise or anti clockwise). Once the user has started to rotate the % Oxygen parameter will st art to flash, continue to rotate the knob and the user will cycle through the adjustable parameters or selectable displays (dependent on mode the user has selected). Each display will start to flash when selected. To adjust the parameter or display press the knob to activate and rotate to change the setting or message. When the required setting has been set, press the knob to confirm.
Note: The selection will time-out after 8 seconds if the confirm button is not pressed. There is also a 2 second delay between button pushes.

6.4 Audio paused and boost buttons (4)

6.4.1 Audio paused

When an alarm threshold has been crossed an alarm signal is triggered. This consist of two components, visual and audible. Pressing the audio paused button, mutes the audible component of the alarm for 120 seconds.
Pressing the audio paused when no alarms are active, mutes the unit for a period of 120 seconds. For any alarms generated within this period only the visual component will activate.
The unit will display the text “Audio Paused” and will alternate this message with any alarm message in an alarm condition.
Pressing the button for a second time in audio paused mode will revert the unit to normal operation.
Note: Any latching alarms active during audio paused mode are reset on the second press of the button. See “Latching and non latching alarms” on page 98.
6.4.1.1 Audio paused alarm setting
In the flow mode if the button is held down for 4 seconds it sets the High leak and Exhaust Impeded alarm thresholds ±2mbar around the delivered CP AP pressure.

6.4.2 Boost button

This delivers an up to 3 mbar of gas as long as the user holds the button down for a maximum of 30 seconds. After 30 seconds the unit stops the boost, but the user can re-apply the boost by repressing the boost button.

6.5 Light bar (5)

The light bar illuminates in an alarm condition. Red for high priority alarms, Y ellow for medium priority alarms and Blue for low priority and technical alarms.
Page 25 of 128
Page 26

7. Operation

7.1 Power up

1. Push the power button on the rear of the machine for 3 seconds.
Note: If the battery indicator and mains indicator lights on the front panel are both illuminated, the unit cannot be turned on. See “Battery and mains indicators” on page 53.
2. St art of self test by cycling the LED’ s. The unit will sound 2 speaker tones. This is to confirm the speaker operation.
3. The battery low and mains disconnected icons will flash briefly.
4. At 3 in the Apnoea display window, the light bar and audio paused button should become illuminated with red LEDS.
Page 26 of 128
Page 27
5. At 2 in the Apnoea display window the light bar and audio paused button should become illuminated with yellow LEDS.
6. At 1 in the Apnoea display window the light bar only should become illuminated with blue LEDS
7. At 0 in the Apnoea display window the light bar should switch off.
8. The unit will run through a self test routine. It will start with a Pneumatic test.
Note: Air has to be connected for the unit to pass the pneumatic test.
Page 27 of 128
Page 28
9. The unit will then enter Standby mode.
10.The unit will now calibrate the oxygen monitoring system, starting with the 21% point.
This is indicated by “CAL” being displayed in the lower right of the LCD.
11.The unit will now calibrate the oxygen monitoring system, starting with the 100% point.
This is indicated by “cal” being displayed in the lower right of the LCD.
12.The stabilisation period for the oxygen system is indicated by the “--” being displayed in the lower right of the LCD.
Page 28 of 128
Page 29
13.Once the calibration is complete the unit will be in “Standby mode” mode
Note: If the unit is being used only with air, the unit will display “--”. The Oxygen control is locked to 21%. If the user tries to select a percentage of Oxygen greater than 21% the message "O2 Calib Error" is displayed.

7.2 Oxygen calibration routine description

The SLE1000 oxygen calibration routine is as follows: The first 180 seconds after power up the unit calibrates the
system using air. The unit displays the text “CAL” in upper case characters.
The second 90 seconds the unit calibrates the system using 100% oxygen. The unit displays the text “cal” in lower case characters.
Note: The user can when the unit is displaying “cal” in lower case letters set the oxygen percentage.
The following 30 seconds the unit displays “--” as the unit allows the oxygen monitoring system to stabilize.
Note: If the unit is being used with air only the “--” will not be replaced by the measured value of percentage of oxygen. In standby mode the “O2 Calib Error alarm” will not trigger. The user will not be alerted to the calibration failure until either CPAP or Flow mode is selected. The user will then cancel the alarm accepting air only use.
Page 29 of 128
Page 30
Warning: If the unit is connected to Oxygen and the “O2 Calib Error alarm” is triggered when entering either CPAP or Flow mode. Manually calibrate the 100% oxygen point in either CPAP or Flow mode. If the “O2 Calib Error alarm” is re-triggered remove the unit from service.
The unit will then display 21% oxygen and the user can now set required percentage.
Note: In the first 4 hours the SLE1000 performs an adaptive calibration routine. If the unit detects that the measured is lower than the set by 3% then it forces a recalibration of the 100% oxygen. The unit will recheck the difference every 8 minutes within the first four hours of operation.
Note: If the unit detects that the measured and set differ by ± 5% it generates the “High Oxygen” or “Low Oxygen” alarm. In this instance the user can manually recalibrate the 100% oxygen. See “100 % O2 Calibration”
on page 40. This check is not limited to the first four hours of operation.
The unit automatically recalibrates the oxygen every 8 hours unless a calibration error has occurred, (i.e. using the SLE1000 with air only) then the unit calibrates every hour.
Note: If the SLE1000 is started using only air, the user can connect the oxygen supply at any time, but will have to manually calibrate the oxygen system.
Warning: If the oxygen supply fails after calibration whilst set to 21% oxygen the unit will not alarm. The " No Oxygen Supply" alarm will only become active when the user tries to set a percentage of oxygen above 21%.

7.3 Power down

Push and hold the power button on the rear of the machine for 3 seconds. Re-press the button to cancel the complete power fail alarm.
Note: The unit can be turned on again after 3 seconds.
Page 30 of 128
Page 31

8. Standby mode

On power up the unit enters “S t andby” mode. In “Standby” mode the unit reduces the flow to less than 2 l/min.
Stand-by mode is indicated in the CP AP LED display by one horizontal dash “-”.
The LCD displays “Standby...”.
Warning: The patient must never be connected to the unit in “Standby” mode. The unit does not deliver any therapeutic flow in this mode and all patient and gas alarms are disabled with the exception of the "Battery Low" alarm.
Note: In standby mode the user can only select the "% Oxygen", "CPAP", "Apnoea" and "Set Flow"

8.1 Exiting Standby mode

The user has three routes to exit standby mode. A. Selecting a CPAP pressure. B. Selecting a flow of 6 l/min (or if a flow has already been set the last set flow). C. Selecting a flow directly.
Note: Please see the “CPAP and Flow mode” chapter for a detailed description on page 55.
Page 31 of 128
Page 32

9. Selecting a parameter, display or calibration function

9.1 Upper and lower level functions.

The following is a the list of selectable parameters using the control knob:
Set Oxygen Level Parameter - lower level function Set CPAP level Parameter - lower level function Set Apnoea Time Parameter - lower level function Set Trigger Sen. (Sen. = Sensitivity) Parameter - upper level function Show Batt. Level (Batt = Battery) Display function - upper level function Start O2 Calib (Calib = Calibration) Calibration function - upper level function Show Alarm Hist. (Hist. = History) Display function - upper level function Set Flow Parameter - upper level function

9.1.1 Selecting lower level functions

Oxygen %, CPAP, Apnoea and Set Flow values (Lower level functions)
Note: Please see the “CPAP and Flow mode” chapter for a detailed
description on page 55.
Note: If the user is in Standby mode, they can only select one of the following parameters % Oxygen, CPAP, Apnoea and Set flow.
The procedure is the same for all the settable parameters.
Push or turn the knob.
The Oxygen parameter will start to flash.
Page 32 of 128
Page 33
Rotate the knob to the left to cycle through the selectable parameters, displays or calibration functions.
The screen will display the title of the parameter that can be adjusted.
Push the knob to adjust the flashing parameter.
The LCD display changes from “Set Apnoea time” to “Adjust Parameter”.
Rotate the knob to change the set parameter value.
Page 33 of 128
Page 34
Push the knob to confirm the setting. The LCD display will return to displaying “Flow Mode”.
Note: If the unit is delivering CPAP then the Cycle indicator bargraph is displayed. See section
12.1 on page 51.
Note: Failure to push the knob to confirm the setting will cause the set value to be discarded after 8 seconds.
Page 34 of 128
Page 35

9.1.2 Selecting upper level functions

Note: For the user to select an upper level function the unit will have to be in either "Flow Mode" or "CPAP" mode.
The procedure is the same for all the settable functions or displays..
The display messages are located in the LCD window.
Push or turn the knob.
Rotate the knob until the upper line of text starts to flash.
Page 35 of 128
Page 36
Push the knob to select.
Page 36 of 128
Page 37

9.2 Upper level functions, displays and parameters

9.2.1 Pressure trigger sensitivity

Push or turn the knob.
Rotate the knob until the text “Set Trigger Sen.” appears and starts to flash.
Push the knob to select. The default value of 3 is
displayed.
Page 37 of 128
Page 38
Rotate the knob to increase or decrease the valve.
The trigger sensitivity has a range of 1 to 6, with the default being 3.
1 being least sensitive. 6 being most sensitive.
Push the knob to select the new sensitivity setting. The unit display returns to the Flow mode.
Note: If the unit is delivering CPAP then the cycle indicator bargraph is displayed. See section
12.1 on page 51.
Page 38 of 128
Page 39

9.3 Showing battery charge level

Rotate the knob to select the “Show Batt. level” and starts to flash.
Push the knob and the battery charge level will be displayed.
Push the knob to return to the default display.
Note: The battery level times out after 8 seconds.
Page 39 of 128
Page 40

9.3.1 100 % O2 Calibration

Push or turn the knob.
Rotate the knob until the text “Start O2 Calib” appears and starts to flash.
Push the knob to select.
Page 40 of 128
Page 41
Rotate the knob to select the 100% Oxygen calibration routine.
Push the knob to start the 100% Oxygen calibration routine. The unit display returns to the Fixed Flow mode but displays the word “cal” in the measured percentage of Oxygen area. (See “Oxygen calibration routine description” on page 29.)
Note: If the unit is delivering CPAP then the Cycle indicator bargraph is displayed. See section 12.1 on page 51.
Note:If the unit has been setup to operate with air only the 100% oxygen calibration will time out after 3 minutes. The unit will return to displaying “--”. No alarms will be generated.
Page 41 of 128
Page 42

9.4 Alarm history

Push or turn the knob.
Rotate the knob until the text “Show Alarm Hist.” appears and starts to flash.
Push the knob to select. The most recent alarm to
be generated will be displayed. This is indicated by -1.
The alarms are tagged -1 to
-25, -1 being the most recent.
Note: As a new alarm is generated the last alarm is deleted and all the existing alarms move up in ranking.
Page 42 of 128
Page 43
Rotate the knob to scroll through the last 25 alarm messages.
Push the knob to return to the default display.
Note: The alarm history times out after 6 seconds.
Note: The alarm history is cleared when the unit is turned off.
Page 43 of 128
Page 44

9.5 Set Flow

Push or turn the knob.
Rotate the knob until the text “Set Flow.” appears and starts to flash.
Push the knob to select. The default flow value of 6
l/min will be displayed in the right hand corner.
Note: If a Flow value has already been set during the session the unit will use this value.
Page 44 of 128
Page 45
Rotate the knob to select the required flow.
As the flow value is changed the text will flash.
Confirm the setting. The unit will enter flow
mode.
Note: Please see the
“CPAP and Flow mode” chapter for a detailed description on page 55.
Page 45 of 128
Page 46

9.6 Elapsed time counter

The SLE1000 has an hour counter which records the total elapsed in use time.
To access the display, press the “Audio paused” and “Boost” buttons together.
The text “Been used for: N h” will be displayed.
Press mute button to cancel the display.
Note: The “Been used for h” does not time out.
Page 46 of 128
Page 47

10. Audio mute button

10.1 Pre-mute

The user has the ability to pre-mute the alarm system.
The user pushes the audio paused button once.
The unit will alternate between the text “Audio Paused” and the cycle indicator bargraph.
The unit will flash the light bar (Blue) when the text “Audio Paused” is displayed.
The pre-mute will last for two minutes.
The user can cancel the pre-mute by re­pressing the audio paused button.
Note: Any alarms that are generated during the pre-mute period will only be able to initiate the visual component (i.e. the alarm bar and alarm message.)
The alarm message will alternate between the alarm message text and the audio paused text.
Page 47 of 128
Page 48

10.2 Audio paused

The user has the ability to mute the alarm system.
When an alarm has been triggered, the user pushes the audio paused button once.
The unit will display the text “Audio Paused”.
Note: The alarm message will alternate between the alarm message text and the audio paused text.
The mute will last for two minutes. The alarm message will alternate with audio paused message.
The user can cancel the mute by re­pressing the audio paused button.
Note: Any latching alarms active during audio paused mode are reset on the second press of the button. See “Latching and non latching alarms” on page 98.
Note: The user needs to be aware that the “Mains disconnect”, “No oxygen supply” and “O2 Calib. Error” alarms will not re-instate. This is because the unit can operate on battery power and on air only.
Page 48 of 128
Page 49
If the alarm condition has not been cleared then the alarm will be re­instated.
If the alarm condition has been cleared the SLE1000 returns to CPAP mode.
Note: The audio mute button has a third function of setting alarm thresholds in flow mode. See page 67.
Page 49 of 128
Page 50

11. Boost button

The boost button delivers an up to 3 mbar above the measured flow.
The boost is maintained whilst the button is pressed, up to 30 seconds.
The boost is delivered at the set Oxygen percentage.
In the main display the text +3 is displayed.
Note: The Boost button will time out after 30 seconds. This is a safety feature. The user can re-apply the boost if required by releasing and repressing the Boost button.
Note: The Boost button has minimal effect between 10 - 12 mbar.
Note: On pressing the boost button the patient trigger detection is suspended for 5 seconds
Warning: When the set pressure is 11 or 12 mbar and the user applies the boost, the delivered pressure might go above 15.5 mbar momentarily and trigger the "High Pressure" alarm. If the alarm is triggered this will cut of the gas supply for 5 seconds. See “Alarm messages and sounds.” on page 98.
Page 50 of 128
Page 51

12. Indicators

12.1 Cycle indicator bargraph

When the SLE1000 is delivering CPAP the LCD displays a airway pressure bargraph. The bargraph is an indication of breath cycling (mechanical or patient).
The bargraph has a span of ±2.5 mbar around the measured mean airway pressure.
The decimal point in the measured mean airway pressure reading denotes the centre point the bargraph
The bargraph gives a visual indicator to breath cycling.

12.2 Breath detection

When the SLE1000 registers inspiratory effort from the patient it displays a star character in the upper left line of the LCD display.
Page 51 of 128
Page 52

12.3 Averaged value

12.3.1 BPM

The BPM is an averaged value from a 30 second rolling window.

12.4 Measured values

12.4.1 MAP

The Mean Airway Pressure (MAP) is a measured value in mbar.

12.4.2 Flow

The Flow is a measured value in l/min.

12.4.3 Oxygen

The % oxygen is a measured value of the fraction of inspired O2.
Page 52 of 128
Page 53

12.5 Battery and mains indicators

12.5.1 Battery indicator

The battery icon will illuminate when battery drops to 10% or less of the full charge level. It is accompanied by a audible and visual alarm. The icon will flash whilst the battery is charging (less than 100%).

12.5.2 Mains disconnect indicator

The mains disconnect icon will illuminate when mains power is unavailable. The icon remains illuminated until mains power is restored. It is accompanied by a audible and visual alarm.
Note: If both the battery indicator and mains icon flash when the power button is depressed the battery is depleted, the user will be unable to restart the unit until mains power has been restored.

12.6 Power related faults

12.6.1 With mains power connected

If both the battery and mains indicator lights are flashing, this indicates that there is a hardware fault. The unit should be removed from service.
Note: If the battery icon is illuminated and fixed the battery has failed or has been disconnected.

12.6.2 With mains power disconnected

If both the battery and mains indicator lights are flashing whilst pressing the mains power on button, the main battery is completely discharged. The user will not be able to start the device on battery power . Connect mains power and allow the unit to charge the battery for 8 hours. It is not recommended that the unit be used on battery power when mains power is available. The purpose of SLE1000 battery is to allow continued operation in the event of a mains power failure.
Warning: If the unit has shutdown due both mains power being disconnected and battery depletion, the user will be unable to restart the unit until mains power has been restored.
Page 53 of 128
Page 54
This page is intentionally blank.
Page 54 of 128
Page 55
CPAP & Flow mode
Page 55 of 128
Page 56

13. CPAP & Flow mode

This chapter describes in detail the procedure of selecting either CPAP or Flow mode.

13.1 Standby to Flow mode

Push the knob. The Oxygen parameter will
start to flash.
Rotate the knob to the left to the CPAP parameter.
Push the knob to select the parameter.
The display reverts to "Standby...".
Page 56 of 128
Page 57
Rotate the knob to the right to set the control to the “--”
Push the knob to confirm the setting.
The unit will enter "Flow Mode" mode (set at 6 l/min).
Note: If a Flow value has already been set during the session the unit will use this value.
Page 57 of 128
Page 58

13.2 Setting a flow in flow mode

Push the knob.
Note: The default value of 6 l/min is used.
The Oxygen parameter will start to flash.
Rotate the knob to the right to select "Set Flow".
Page 58 of 128
Page 59
Push the knob to select. The flow will now be displayed in the top right hand corner of the display
Rotate the knob to the right to set the required flow. The text will starts to flash.
Push the knob to confirm the required flow.
The text will change to "Flow mode".
Page 59 of 128
Page 60
The SLE1000 will start to deliver the set flow.
At this stage the light bar will illuminate with a medium priority colour. This is to inform users that the alarm thresholds require setting.
Note: The user should be aware that in “Flow Mode” the trigger thresholds for the “High leak” and “Exhaust Impeded” alarms do not auto track the MAP. The default Flow is 6LPM and the alarm thresholds are default to 2mbar (Leak) and 6 mbar (Exhaust).
When the user increases or decreases the set flow, the MAP will cross one of the previously set thresholds, triggering the appropriate alarm. This alarm will not clear unless the thresholds are set by the user around the new MAP that is being used to achieve the set flow or if the MAP crosses back to within the previously set threshold parameters.
Depress and hold the alarm mute button for 4 seconds. This will set the alarm thresholds around the measured MAP. by ± 2 mbar.
The unit indicates that the alarms have been set by flashing the light bar yellow for 1 second.
Page 60 of 128
Page 61

13.3 Flow mode to CPAP mode

The user is in “Flow Mode”.
Push or turn the knob.
Rotate the knob to the left to select the “Set CPAP level”.
Page 61 of 128
Page 62
Push the knob to adjust the CPAP parameter.
The message “Set CPAP Level” will change to “Adjust Parameter”.
Rotate the knob to the right to adjust CPAP parameter.
Push the knob to confirm the setting.
The SLE1000 will now adjust the pressure.
The message display now changes to the airway pressure bargraph.
Page 62 of 128
Page 63

13.4 Standby to CPAP mode

Push or turn the knob. The Oxygen parameter will
start to flash.
Rotate the knob to the left to the CPAP parameter.
Push the knob to select the parameter.
Page 63 of 128
Page 64
Rotate the knob to the right, Continue past the “--”
Continuing to rotate the knob to set the required CPAP. The text adjust parameter will be displayed.
Push the knob to confirm the setting.
The SLE1000 will now adjust the pressure.
The message display now changes to the airway pressure bargraph.
Page 64 of 128
Page 65

13.5 CPAP to Flow mode

The user is in CPAP mode.
Push the knob.
Rotate the knob to the right to select "Set Flow".
Page 65 of 128
Page 66
Push the knob to select. The flow will now be displayed in the top right hand corner of the display
Rotate the knob to the right to set the required flow . The text will starts to flash.
Push the knob to confirm the required flow.
The text will change to "Flow mode".
Page 66 of 128
Page 67
The SLE1000 will now deliver the set flow.
At this stage the light bar will illuminate with a medium priority colour. This is to inform users that the alarm thresholds require setting.
Note: The user should be aware that in “Flow Mode” the trigger thresholds for the “High leak” and “Exhaust Impeded” alarms do not auto track the MAP. The default Flow is 6LPM and the alarm thresholds are default to 2 mbar (Leak) and 6 mbar (Exhaust).
When the user increases or decreases the set flow, the MAP will cross one of the previously set thresholds, triggering the appropriate alarm. This alarm will not clear unless the thresholds are set by the user around the new MAP that is being used to achieve the set flow or if the MAP crosses back to within the previously set threshold parameters.
Depress and hold the alarm mute button for 4 seconds. This will set the alarm thresholds around the measured MAP. by ± 2 mbar.
The unit indicates that the alarms have been set by flashing the light bar (Yellow) for 1 second.
Page 67 of 128
Page 68
This page is intentionally blank.
Page 68 of 128
Page 69
SLE1000 Set-up
Page 69 of 128
Page 70

14. SLE1000 Set-up

A
B
C
D
E

14.1 First use

Setting up the SLE1000 for the first time. Step 1. Take the SLE1000 out of
its transport carton and remove the protective film.
Step 2. Remove the air and
oxygen inlet port caps.
Step 3. Check that the following
ports are free from any packing material A. Fresh gas port B. Proximal pressure port
C. Air inlet D. Oxygen inlet E. Exhaust port
Page 70 of 128
Page 71
Step 4. The SLE1000 is shipped
75mm
with the main battery disconnected.
Using a screw driver remove the battery fuse holder. (Indicated by arrow)
Step 5. Insert the fuse into the
fuse holder. (T2.0A 250V). (The fuse is located in the accessory pack, Part nº: M0457/ 2AC/001)
Step 6. Insert the fuse holder and
fuse into the receptacle. Do not over tighten.
Step 7. Remove the screw and
washer holding the P clip. Insert the power cable into the socket and secure the cable with the P clip.
Allow 75mm of cable to form the loop.
Page 71 of 128
Page 72

14.2 Rotation of gas inlets

Remove the locking screw to release the air and oxygen inlets
Rotate inlets 90°
Remove the oxygen cell cover plate
The gas inlets on the SLE1000 can be rotated through 90°. The SLE1000 is supplied with the inlets in the horizontal position. The user can rotate the inlets through 90° by releasing the swivel lock.

14.3 Water trap and particulate filter

Each SLE1000 unit is fitted with a water trap and particulate filter in the air side of the pneumatic system. The water trap is auto-draining. The particulate filter will remove particles of more than 5 micrometer (Microns) in size.

14.3.1 Setup

Prior to first use of the SLE1000 the outlet of this water trap needs to be connected to a drain hose and water collection cup (Not supplied). The hose should have an internal diameter of 4mm and be pliable enough to fit over the hose tail bard on the outlet of the water trap.
The other end of the hose should feed any collected water into an open topped collection container that can be easily emptied.

14.3.2 Maintenace

Please refer to the SLE1000 service manual for maintenance instructions and replacement filter instructions.
Page 72 of 128
Page 73

14.4 Standard setup

SLE1000 can be used with air and oxygen gas supplies or air only. For use with air and oxygen start with section 14.4.1 on page 73 and for air only start with section 14.4.2 on page 74.

14.4.1 With Air and Oxygen from a wall supply between 2 and 7 bar

Step 1. Connect the air and
oxygen hoses to the SLE1000.
Step 2. Connect the air and
oxygen hoses to the wall supplies.
Step 3. Connect SLE1000 to
mains power. Do not turn the SLE1000 on.
Proceed to “Functional testing prior to patient use” on page 80.
Page 73 of 128
Page 74

14.4.2 With air only from a wall supply between 2 and 7 bar

Step 4. Connect the air hose to
the SLE1000.
Step 5. Connect the air hose to
the wall supply.
Step 6. Connect SLE1000 to
mains power. Do not turn the SLE1000 on.
Proceed to “Functional testing prior to patient use” on page 80
Page 74 of 128
Page 75

14.4.3 With air only from a medical air compressor

Step 7. Connect the air hose to
the SLE1000.
Step 8. Connect the air hose to
compressor.
Step 9. Turn on the compressor
and allow the pressure to build up above 2 bar.
Note: The water trap / particulate filter may bleed air due to the auto drain mechanism not closing. Block the outlet momentarily to allow the mechanism to close.
Step 10. Connect SLE1000 to
mains power. Do not turn the SLE1000 on.
Proceed to “Functional testing prior to patient use” on page 80
Page 75 of 128
Page 76

14.5 Disconnection of air and oxygen hoses

Warning: The SLE1000 does not produce any pneumatic noise when turned off. Always disconnect the air and oxygen hoses from the wall outlets prior to disconnection from the unit.
Step 1. Disconnect the air and
oxygen hoses to the wall supplies.
Step 2. Disconnect the air and
oxygen hoses to the SLE1000.
Page 76 of 128
Page 77

14.6 Battery charging

14.6.1 Prior to first use

The SLE1000 should be connected to mains power for 8 hours prior to first use to allow the batteries to charge.
After 8 hours confirm that the battery status indicator reads 100%.

14.6.2 Battery charging whilst in use

The SLE1000 will charge the batteries when connected to mains.
Note: The SLE1000 does not have to be turned on for the charging to take place.

14.7 Battery care

The SLE1000 contains 1 sealed lead acid battery that can operate the unit for upto 4 hours in the event of a mains power fail situation. The 4 hour period stated is for a battery that is new or for a battery that has not been allowed to remain in a deep discharged state.
If the SLE1000 has been used on battery power and has been allowed to fully discharge the battery , the battery can be described as being in a deep discharged state. If the unit is stored or placed out of service without recharging the batteries, the deep discharged state will reduce the battery life.
Warning: For a battery that has been allowed to remain in a deep discharged state, the damage is irreversible. On re-charging a damaged battery the service life will be curtailed.

14.8 Using the SLE1000 on battery power.

The SLE1000 is not designed to be started and used on battery power only, as it is a mains powered device with battery backup.
If attempting to start the unit purely on battery power and the battery and mains indicator lights are flashing whilst the power on button is depressed, the unit will not start. Reconnect mains power and leave connected to mains for a minimum of 8 hours to fully recharge the battery.
If the unit starts on battery power and after the pneumatic and oxygen calibration test the low battery (medium priority) alarm becomes active, restore mains power immediately as the unit has only 10 to 20 minutes operating time left. If the low battery (high priority) alarm is triggered the unit will shut down after 1 minute.
Page 77 of 128
Page 78

14.9 Mains isolation

To isolate the SLE1000 electrically from the mains supply, remove the mains plug.

14.10 Extended storage and transportation

14.10.1 Storage

If the user is intending to store the driver for a period greater than 10 days and is unable to charge the batteries during this time as recommended in the “Warnings” on page 19, then the following procedure is recommended. (Refer the ventilator to a qualified technician).

14.10.2 Transportation

If the user is intending to transport the driver then the battery fuse must be removed. (Refer the ventilator to a qualified technician)
14.10.2.1 Fuse removal procedure
Step 1. Charge the batteries for 8 hours. Step 2. Remove the battery fuse by reversing the process described in ’14.1
First use’ on page 70. Step 3. Ensure that the fuse remains with the unit. Step 4. Mark the unit as having the battery disconnected.
Page 78 of 128
Page 79
Functional Testing
Page 79 of 128
Page 80

15. Functional testing prior to patient use

Humidifier supply line
Humidifier
Generator set
Inspiratory supply tube
Proximal airway monitoring tube
Exhaust tube
Optional bacterial filter
Connect only to circuit NOT the SLE1000 nCPAP driver.
Do not make this connection.
Set the driver up as per the standard setup (section 14.4 on page 73). Set the humidifier up as per the manufacturers instructions.
Note: When functional testing requires handling of the generator / patient circuit please ensure that cross contamination does not occur.
Warning: If any of these tests do not function as described and the problem can not be rectified by reference to the trouble shooting chapter on page 89, then the unit should not be used until it has been repaired. Please contact an SLE approved engineer, or SLE.
1. Connect the delivery circuit and flow generator as per the manufacturer’s
instructions. Flow generators and delivery circuits may differ in construction from what is shown below .
Note: DO NOT connect the proximal airway tube to the driver at this time. Connection may cause the pneumatic test to fail. (Gas flow error).
Note: Fit a set of prongs that can be discarded after functional testing.
Page 80 of 128
Page 81

15.1 Battery test.

1. Disconnect the mains plug.
2. Turn on the SLE1000.
Caution: If both the battery indicator and mains icon flash when the power button is depressed the battery is depleted. Connect mains power and allow the unit to charge for a minimum of 8 hours. The user can use the unit, but must be aware that the backup battery is depleted and may not be able to support ventilation in the event of a mains power fail situation, until the recharge cycle is complete. Check the battery status from the CPAP or flow modes.
3. Allow the unit to perform its self test.
4. Confirm that the alarm sounder generated an alarm tone, all the LCD and LED displays cycled through the test as described in section 7.1 on page 26.
Warning: If the user notes that on power up the unit sounder did not produce an alarm tone, or that the LCD or LED did not function as described, then remove the unit from service and route for repair/ investigation.
5. Re-connect mains power.
6. Wait for then unit enter “Standby mode”.

15.2 Mains disconnect alarm

7. Connect the proximal airway tube
8. Set a CPAP pressure of 5 mbar.
Note: If you are using the unit with air only, the “O2 CALIB ERROR” will be triggered. Press the alarm mute button twice to clear the alarm message.
9. Occlude the generator outlet.
10. Remove the mains lead from the wall socket.
Page 81 of 128
Page 82
11. The SLE1000 should display and sound the “MAINS DISCONNECT” alarm. The mains power not connected icon on the front of the driver should Illuminate.
12. Reconnect mains power and cancel the alarm indication.

15.3 Total power fail alarm test.

13. Turn the unit off. The total power fail alarm will now sound.
14. Repress the button to cancel.
15. Disconnect the proximal airway tube.

15.4 High leak and high pressure alarm

16. Turn the unit back on.
17. Allow the pneumatic test to clear and the unit to enter standby mode.
18. Connect the proximal airway tube.
19. Wait until the oxygen system has completed its calibration.
20. Set a CPAP pressure of 5 mbar.
Note: If you are using the unit with air only, the “O2 CALIB ERROR” will be triggered. Press the alarm button twice to clear the alarm message.
21. Remove the occlusion from the generator outlet.
22. The SLE1000 should display and sound the “HIGH LEAK” alarm after approximately 1 minute.
23. Occlude the generator outlet and the exhaust tube.
24. The SLE1000 should display and sound the “HIGH PRESSURE” alarm, the driver should cut the fresh gas flow.
25. Remove the occlusion from the generator outlet and exhaust tube.
26. The SLE1000 will re-instate the CPAP pressure after approximately 5 seconds.
27. The SLE1000 should display and sound the “HIGH LEAK” alarm.
28. Re-occlude the generator.
29. Wait for the alarm to clear.
30. Move directly to the Apnoea alarm test.

15.5 Apnoea alarm

31. Set the Apnoea time to 10 seconds.
32. After 10 seconds the APONEA alarm will sound and the driver will increase the CP AP level by 3 mbar. this is indicated by “+3” being displayed in the right hand corner of the LCD display.
33. Turn off the Apnoea alarm.
Page 82 of 128
Page 83

15.6 Zero pressure alarm and purge check

Air supply hose
O2 supply hose
34. Disconnect the proximal airway tube from the proximal pressure port.
35. The SLE1000 should display and sound the “Zero Pressure” alarm.
36. Occlude the Proximal Pressure port.
37. The SLE1000 should display and sound the “High Pressure” alarm.
38. Remove the occlusion from the port.
39. The SLE1000 should display and sound the “Zero Pressure” alarm.
40. Reconnect the proximal airway tube to the proximal pressure port.
41. Reset the alarm by pressing the alarm button twice.

15.7 Gas supply alarms

Warning: Disconnect the gas supplies from the wall outlet only. DO NOT unscrew the hose connections from the rear of the SLE1000.

15.7.1 Procedure when using air and oxygen

1. Select fixed flow mode of 6 l/min.
2. Set the % of Oxygen to 25%
3. Disconnect the air supply from wall outlet. The SLE1000 should display and sound the “NO AIR SUPPLY” alarm.
4. Reconnect air supply, cancel the alarm.
5. Disconnect the O
supply from
2
wall outlet. The SLE1000 should display and sound the “NO OXYGEN SUPPLY” alarm.
6. Reconnect oxygen supply, cancel the alarm. The unit will re-calibrate the Oxygen.
7. Set the SLE1000 to “Standby...” mode.
Page 83 of 128
Page 84

15.7.2 Procedure when using an air only

Air supply hose
1. Select fixed flow mode of 6 l/min.
2. Disconnect the air supply from wall outlet. The SLE1000 should display and sound the “NO AIR SUPPLY” alarm.
3. Reconnect air supply, cancel the alarm.
4. Set the SLE1000 to “Standby...” mode.

15.7.3 Procedure when using an air compressor

1. Select a CPAP pressure of 12 mbar.
2. Turn off the compressor. The SLE1000 should display and sound the “NO AIR SUPPLY” alarm.
Note: The user may have to wait for the pressure in the compressor reservoir to drop, before the alarm is generated. As the pressure drops the “Zero Pressure” alarm is generated as the pressure drops before the “No Air Supply” alarm is generated.
3. Turn on the compressor. Occlude the watertrap outlet unit the internal valve closes.
4. Set the SLE1000 to “Standby...” mode.
Functional testing is now complete.
Page 84 of 128
Page 85
Operational Considerations
Page 85 of 128
Page 86

16. Operational considerations

16.1 General

16.1.1 Standby mode

The patient must not be connected to the unit in “Standby” mode. The unit does not deliver any therapeutic flow in this modes. In “Standby” mode all patient and gas alarms are disabled with the exception of the “Low battery” alarm and all system failure alarms.

16.2 CPAP

16.2.1 CPAP setting of 2 to 4 mbar

If the user is using the SLE1000 with small or medium prongs and a CPAP pressure of between 2 and 4 mbar, the unit may not detect the disconnection of the patient from the circuit.
The SLE1000 must not be used as the sole form of patient monitoring. Independent monitoring of blood gases must be used in conjunction with the SLE1000. Breath detection technology used in the apnoea monitor function is based on rate and magnitude of change in the pressure measured in the patient interface (that is shown on the bar graph). The sensitivity of the technology has limitations with the smallest patients (<1000g) and if a leak exists or if the patient opens/breathes by the mouth. The apnoea monitor may be disabled if breath non-detection results in nuisance apnoea alarms.

16.2.2 Setting pressure trigger sensitivity

When delivering CPAP with Apnoea detection turned ON the user may have to adjust the pressure trigger sensitivity. The method of setting the pressure trigger sensitivity is described in’9.2.1 Pressure trigger sensitivity’ on page 37. If the user deems that the BPM rate is high due to reasons other than patient effort then they can decrease the sensitivity by reducing the value from the default of 3 down to 1. The user can increase the sensitivity by increasing the value from 3 to 6. Patient trigger detection determines the delivery of backup breaths in an apnoeaic episode.
Page 86 of 128
Page 87
Operation
Page 87 of 128
Page 88

17. Basic operation

The following chapter describes the basic operation and is not intended to be prescriptive.

17.1 Monitoring

The recommended patient monitoring requirements are:
1. ECG/heart rate.
2. Respiration rate
3. Blood pressure (either by invasive or non invasive means).
4. Oxygen saturation.
5. Blood gas saturation by either transcutaneous carbon dioxide arterial /
capillary blood sampling.
6. Regular cranial ultrasound examinations (in the newborn).
7. Standard nursing care for Intensive Care patients.

17.2 Operation

Step 1 Setup the SLE1000 as described in section 14.4 on page 73. Step 2 Ensure that the SLE1000 has been functionally tested as described in section 15.
on page 80.
Step 3 Ensure the unit has calibrated the oxygen system.
Note: If you are using the unit with air only, the “O2 CALIB ERROR” will be triggered. Press the alarm button twice to clear the alarm message. The unit
will display “- -”. Do not set the % O2 above 21% as this will trigger the “NO O2 SUPPLY” alarm.
Step 4 Set the SLE1000 into the required mode. Step 5 Prepare the patient bonnet and select the prong or mask suitable for the patient. Step 6 Attach the generator circuit to the patient. Step 7 Set the Apnoea alarm time. Step 8 Adjust the % oxygen, (If the unit is connected to air only leave the setting at 21%). Step 9 Monitor the patient condition and adjust parameters accordingly.
Page 88 of 128
Page 89
Troubleshooting Chart
Page 89 of 128
Page 90

18. Troubleshooting chart

Symptom Possible Cause Remedy Alarm message:
High Pressure
Alarm message:
No Air Supply
Alarm message:
No Oxygen Supply
Alarm message:
High Leak
Proximal pressure exceeds 15.5mbar.
Nasal prongs blocked
Filters occluded with water. Exhaust tube blocked.
Air supply is too low (<2.0 bar) to assure normal operation.
Check wall connection. Ensure air pressure is
Oxygen supply is too low (<2.0 bar) to assure normal operation.
Check wall connection. Ensure oxygen pressure is
Unable to attain set CPAP level, and flow has reached maximum level.
Check patient connection.
Check patient circuit.
between 2 and 7 bar.
between 2 and 7 bar.
Alarm message:
Zero Pressure
Alarm message:
Exhaust Impeded
Possibly the circuit or patient interface has become disconnected.
Flow mode selected without alarm thresholds being set.
MAP drops to zero
Proximal pressure tube disconnected Check patient connection.
Pressure exceeds set CP AP, and flow has reached minimum level.
Exhaust tube blocked.
Flow mode selected without alarm thresholds being set.
Check patient connection. Check patient circuit.
Set alarm threshold.
Check patient circuit. Check bacterial filter.
Check patient circuit
Set alarm threshold.
Page 90 of 128
Page 91
Symptom Possible Cause Remedy Alarm message:
Gas Flow Error
Alarm message:
High BPM rate
Unit started with complete circuit attached.
Water in circuit creating circuit noise.
See “High BPM rate” on page 100.
Disconnect proximal airway line and cycle power.
Check patient circuit.
Adjust trigger sensitivity.
Page 91 of 128
Page 92
This page is intentionally blank.
Page 92 of 128
Page 93
Technical Information
Page 93 of 128
Page 94

19. Cleaning and disinfection

All cleaning, disinfection and sterilizing should be carried out under the direction of the appropriate hospital authority.
DO NOT allow moisture to enter the electronic module or its electrical sockets. Electronic malfunction may result.
DO NOT steam autoclave the driver or otherwise subject it to temperatures above 62°C. DO NOT immerse any part of the driver in any liquid

19.1 Preparation of a new SLE1000 CPAP driver

Remove all transit packaging. Inspect the fresh gas port and proximal airway port for any packing material. (Retain packaging for future use).
Remove the protective film. Remove the inlet air and O2 gas port caps. (Retain for future use).
Allow the unit to acclimatize, i.e if the unit has been stored in a cold or humid environment whilst in transit/storage.
Disinfect the driver as per section 19.4 on page 95.

19.2 Cleaning and disinfection of an in-service driver

Before cleaning or disinfecting the exterior of the driver the following tasks should be performed:
• The mains cable should be disconnected from the mains supply.
• Remove the patient circuit and bacterial filters. Discard any single use items as per appropriate hospital authority guidelines. Reusable items should be processed as per appropriate hospital authority guidelines and the manufacturers instructions.
• Disconnect the gas supplies from the wall outlets.
• Disconnect the Oxygen and Air hoses from the driver and cap the inlet ports.
Warning (General): Do not insert any object in to the gas ports. This action will result in damage to the port. If the user believes there is a foreign object in a gas port, please refer the driver to qualified service personnel for inspection and repair.
Page 94 of 128
Page 95

19.3 Cleaning Method

Note: Cleaning is an essential prerequisite to disinfection and sterilization.
For cleaning use three clean, disposable, absorbent, non-shedding cloths. Wipe clean with the first cloth using a hand hot water/mild general purpose detergent solution (as prescribed by the appropriate hospital authority). Do not overload the cloth with liquid. Remove the water/mild general purpose detergent solution with the second cloth using water only . Do not overload the cloth with liquid. Wipe dry with the remaining cloth. Care should be taken to ensure that the driver gas ports are not blocked by any debris.
Warning: Ensure that the detergent solution does not enter the unit (LCD and LED windows). Do not use any abrasive cleaners these surfaces.

19.4 Disinfection method

Note: Alcohols such as 70% isopropanol have a good activity against bacteria and viruses. They should only be used after all visible surface dirt has been removed from the area to be disinfected.
For disinfection use two clean, disposable, absorbent, non-shedding cloths. Wipe clean with the first cloth using Alcohol (70% isopropanol). Wipe dry with the remaining cloth.
Warning: Disinfectants containing compounds similar to PHENOL or ALKYLAMINES (Glucorrotamine) are unsuitable.
Page 95 of 128
Page 96

20. Messages & alarms

20.1 Messages

The user interface messages displayed on 20 character wide LCD.
User message Description Action
Airway pressure bar graph (non­scaled)
Audio Paused Indicates that user has activated the
Adjust Parameter Second stage of user interacting with
Indicates normal (technical and patient) operation. Tolerable underlying conditions have been acknowledged by the user.
alarm cancel or alarm ‘pre-mute’.
If alarm button is pressed while the ‘Alarm Paused’ message is displayed, then normal ‘un-paused’ mode is immediately resumed.
device, where the parameter is adjusted.
None
Resets any current alarm condition.
Pauses alarm audio for 120 seconds or until button or is pressed again.
Turning the dial will increment/decrement the parameter in focus. Pressing the dial will confirm selection. Times out and reverts to original value after 3 sec of inaction.
Battery N% Display of the battery charge level in
percent
Flow Mode Displayed to indicate that the user is
in flow mode
Pneumatic test Displayed during internal test Clear only when test has
Set Apnoea Time
Set CPAP Level
Page 96 of 128
Indicator that the Apnea Time can be selected.
Indicator that the CPAP Level can be selected.
Clears itself after 10 sec.
Present whilst delivering the prescribed flow.
passed. Clears itself after 5 sec.
If selected clears after 10 sec.
Clears itself after 5 sec. If selected clears after 10 sec.
Page 97
User message Description Action
Set Flow
Indicator that the flow can be selected.
Set Oxygen Level
Indicator that the Oxygen Level can be selected.
Show Batt. level
Indicator that the Battery Level can be selected.
Show Alarm Hist. Indicator that the Alarm History can be
selected.
Set Trigger Sen. Indicator that the Trigger Sensitivity
can be selected.
Start O2 Calib Indicator that the O2 calibartion
routine can be selected.
Clears itself after 5 sec. If selected clears after 10 sec.
Clears itself after 5 sec. If selected clears after 10 sec.
Clears itself after 5 sec. If selected clears after 10 sec.
Clears itself after 5 sec. If selected clears after 10 sec.
Clears itself after 5 sec. If selected clears after 10 sec.
Clears itself after 5 sec. If selected clears after 10 sec.
Been used for: Display of the battery charge level in
percent
SLE1000 Ver. N.N Version display during power-on-self-
test.
Ready... Indicates that power-on-self-test has
completed.
<no message> Indicates that a particular history log
location does not have any stored event (i.e. alarm has not occurred)
The user has to press the alarm mute button to cancel display.
Clears itself on completion of the power-on-self-test.
Clears itself after 2 sec.
None
Page 97 of 128
Page 98

20.2 Alarm messages and sounds.

20.2.1 Alarm messages.

Alarm messages ranked in order of priority. The current message of highest priority ranking is stored in the history log. Simultaneous lower ranking messages are often the consequence of a higher priority, and they are not stored in the history log. The term ‘alarm’, when used in the Action column, means sounding and displaying a conforming audio-visual alarm.
20.2.1.1 Latching and non latching alarms
A latching alarm is an alarm message that once generated will continue to display both the audible and visual elements even though the original alarm condition has cleared.
A non latching alarm is an alarm message that will display both the audible and visual elements only during the alarm condition. If the alarm condition clears the audible and visual elements self cancel.
Alarm message Description Action
Simultaneous ‘Battery Error’ and ‘Mains disconnect’
(blank display)
Total power loss. Battery has emptied or battery fuse has failed, while in battery operating mode. The 1 to 2 l/min bypass air flow will cease once the volume chamber is empty.
Priority: HIGH Type: TECHNICAL
Device appears as switched off. Normal alarm system and history log is inoperable.
Power manager sounds backup sounder continuously, while backup capacitor power exists (approx 10 minutes), or until user presses power ON­OFF button.
Calib Faulty Calibration integrity error
Latching HIGH priority alarm. Cannot be disabled. Priority: HIGH Type: TECHNICAL
Page 98 of 128
Remove the unit from service.
Remove patient to an alternative form of ventilation. Remove the unit from service.
Page 99
Alarm message Description Action
System Failure N
System Failure
Single or multiple sub­system failures.
N = system failure number which can be between 1-255.
Latching HIGH priority alarm. Cannot be disabled. Priority: HIGH
Remove patient to an alternative form of ventilation. Remove the unit from service.
Type: TECHNICAL
Gas Flow Error Start up self test has detected
abnormal pneumatic system response, or lacking gas supply.
Latching HIGH priority alarm. Priority: HIGH Type: TECHNICAL
High Pressure Proximal pressure exceeds 15.5
mbar. Non-latching HIGH priority alarm.
Priority: HIGH Type: TECHNICAL
Self test is repeated and clears itself upon successful result.
If ventilating - remove patient to an alternative form of ventilation. Remove the unit from service.
Cuts fresh gas flow and evacuates pressure from the circuit. 1.5 l/min bypass flow is delivered, to flush CO2 build up. Device will
restart after 5 seconds. Check Patient, check
patient circuit. Adjust ventilation parameters.
If alarm repeats after the above checks remove patient to an alternative form of ventilation. Remove the unit from service.
Apnoea Breath not detected within the user
set apnoea period. Non-latching HIGH priority alarm.
Is reset by a patient breath. Priority: HIGH Ty pe: PATIENT
Triggers a stimulation breath of 3 mbar above set CPAP for 0.3 seconds (only pressure mode).
User has option for disabling the apnoea monitoring function (value shows ‘--’).
Check Patient, check patient circuit. Adjust ventilation parameters.
Page 99 of 128
Page 100
Alarm message Description Action
No Air Supply Air supply is too low (<2.0 bar) to
assure normal operation.
Device will deliver 100% oxygen as set flow.
Connect air supply. If
Non-latching HIGH priority alarm. Suspends high oxygen alarm.
ventilating - remove patient to an alternative form of ventilation. Remove the unit
Priority: HIGH
from service.
Type: TECHNICAL
No Oxygen Supply Oxygen supply is too low (<2.0 bar)
to assure normal operation. Only active if FiO
is set >21%.
2
Non-latching HIGH priority alarm. Suspends low oxygen alarm. Priority: HIGH Type: TECHNICAL
High BPM rate The measured breath rate exceeds
90 BPM for at least 6 seconds when the Apnoea is set in the range of 5 and 90 seconds.
Device continues to deliver air only gas to patient.
User can prevent recurrence by setting FiO
2
to 21%. Connect oxygen supply. If no oxygen present, set
the unit to run on air only by canceling the alarm message.
If ventilating - remove patient to an alternative form of ventilation. Remove the unit from service.
Check Patient, check patient circuit. Adjust Apnoea parameters.
Page 100 of 128
Disabled if Apnoea is OFF (--) High rate could indicate false
breath detection due to pressure signal noise (e.g. from water in breathing circuit). False detection risk masking an apnoea condition.
Non-latching HIGH priority alarm. Priority: HIGH
Loading...