SKC Limited IOM Sampler User Manual

Page 1
391-01 “Calidaptor” Calibration Adaptor
This Calidaptor is designed for use with the following sampling heads:
225-70A I.O.M. inhalable
sampling head in plastic with plastic filter cassette
225-76A I.O.M. inhalable
sampling head in stainless steel with stainless steel filter cassette
225-79A I.O.M. inhalable
sampling head in plastic with stainless steel filter cassette
225-50 Seven hole sampling head
225-52 Single hole sampling head
A Calidaptor designed specifically for use with the cowled asbestos heads is also available (part no. 391-05).
The SKC Calidaptor allows the flow setting of a sampling train without the problems of holding the head, pump and adjusting the flow all at the same time.
A hinged bracket and clamping screw hold the sampling head securely against a closed pore foam ring providing a leak proof joint.
The Calidaptor can either be screwed directly into the outlet of SKC rotameters (390, 391 and 392 series only) with an ‘O’ ring providing a leak proof joint, or connected to other flow measuring devices using the supplied screw-in hosetail and a length of flexible tubing.
www.skcltd.com
“Calidaptor” Calibration Adaptor
Operating Instructions
Hinged bracket
Foam ring
O ring
Clamping screw
Calidaptor bowl
Hosetail
007-09-008 Issue D October 2009
Page 1
Page 2
Use with SKC rotameters
The Calidaptor is screwed into the top (outlet port) of the rotameter, tightening enough to just compress the ‘O’ ring sealing it to the rotameter.
Pass the sampling head through the hinged bracket, and rest its inlet against the foam ring. Clamp the sampling head in place with the plastic clamping screw until the foam ring is compressed by approximately 1mm.
Connect the sample head outlet hosetail to the pump inlet with a length of tubing.
Use with other sampling heads
With the Calidaptor fitted to the top of an SKC rotameter, screw the supplied hosetail fitting into the threaded hole inside the Calidaptor bowl. Connect from the hosetail to the inlet of the sampling head with a length of flexible tubing.
Note: If subsequently calibrating an I.O.M. (or similar) sampler, especially if used with a foam insert, ensure that the hosetail is removed from inside the Calidaptor bowl.
Use with other types of calibrator
The Calidaptor can be used as described above, with any other type of flow calibrator, by screwing the supplied hosetail into the threaded hole in the inlet of the Calidaptor. Connect from the hosetail to the flow calibrator with a length of flexible tubing.
SKC Limited 11 Sunrise Park, Higher Shaftesbury Road, Blandford Forum, Dorset DT11 8ST United Kingdom
007-09-008 Issue D October 2009
T: +44 (0) 1258 480188 F: +44 (0) 1258 480184 E: info@skcltd.com W: www.skcltd.com