Innovative technologies for monitoring air quality
Page 2
Haz-Dust
Model EPAM-7500
User’s Guide
Page 3
Licenses, Copyrights and Trademarks
This documentation contains trade secrets and confidential information
proprietary to Environmental Devices Corporation (EDC). The software
supplied with the instrumentation, documentation and any information
contained therein may not be used, duplicated or disclosed to anyone, in
whole or in part, other than as authorized in a fully executed EDC End User
License or with the express written permission of EDC.
Haz-Dust ™ is a registered trademark of Environmental Devices Corporation.
Other trademarks are the property of their respective holders.
ii
Page 4
Safety Notice
Repair of instrumentation supplied by Environmental Devices Corporation
(EDC) should only be attempted by properly trained service personnel, and
should only be conducted in accordance with the EDC system documentation.
Do not tamper with this hardware. High voltages may be present in all
instrument enclosures. Use established safety precautions when working with
this instrument.
The seller cannot foresee all possible modes of operation in which the user
may attempt to utilize this instrumentation. The user assumes all liability
associated with the use of this instrumentation. The seller further disclaims
any responsibility for consequential damages.
iii
Page 5
Warranty
Environmental Devices Corporation (EDC) warrants only non-expendable
products (expendable items include batteries disposable filters and cassettes
or other consumables), parts and labor, for a period of one year from date of
shipment to the original purchaser. The warranty covers only product parts
and labor that failed due to normal operation of the instrument and not due to
abuse or negligence. The product will be delivered in accordance to its
published specifications and free from defects in materials or workmanship.
If a product fails to conform to this one year limited warranty it may be
returned to the factory for repair or replacement of the defected part(s). EDC
must be notified of all returning warranty repair products, either in writing or
by telephone. Shipping and insurance cost will be prepaid by the purchaser.
EDC makes no other express warranty and disclaims any implied warranty of
fitness or merchantability.
iv
Page 6
Table of Contents
Licenses, Copyrights and Trademarks ............................................................................................. ii
Safety Notice ..................................................................................................................................... iii
Warranty .......................................................................................................................................... iv
Chapter 1 Introduction to the Haz-Dust Model EPAM-7500..................................................................
Introduction to the Haz-Dust Model EPAM-7500 ......................................................................... 1-1
Introduction to the EPAM-7500 ................................................................................................... 1-2
Overview of the EPAM-7500 ........................................................................................................ 1-4
Appendix A Menu Screens .............................................................................................. Appendix A
Appendix B NIOSH/OSHA Particulate Air Monitoring Reference ...................................... Appendix B
Appendix C Glossary of Terms .......................................................................................... Appendix C
vi
Page 8
Chapter 1 Introduction to the Haz-Dust Model EPAM-7500
Page 9
Introduction to the Haz-Dust Model EPAM-7500
Introduction
In this chapter
Chapter Overview
This chapter gives a complete overview of the Haz-Dust Model EPAM-7500.
This chapter:
• Introduces and describes EPAM-7500.
• Explains operating principles of the EPAM-7500.
• Identifies features, specifications and components of EPAM-7500.
This chapter contains the following topics.
Introduction to the EPAM-7500 1-2
Overview of the EPAM-7500 1-4
Real-Time Dust Monitoring Principles 1-6
Features 1-7
Specifications 1-8
Components 1-9
Topic See Page
1-1
Page 10
Introduction to the EPAM-7500
Introduc
tion
Figure 1.2. Picture of the Haz
-
Dust EPAM
-
7500
The Haz-Dust EPAM-7500 is a high sensitivity real-time particulate monitor
designed for ambient environmental and indoor air quality applications. This
unit combines traditional filter techniques with real-time monitoring methods.
These techniques combined overcome limitations of all other aerosol
monitoring products.
Continued on next page
1-2
Page 11
Introduction to the Haz-Dust EPAM-7500, Continued
Comparison of
Description of
Advantages of
•
Description of
Advantag
es of
•
The traditional and real-time dust monitoring methods are described below.
methods
Air is drawn by a vacuum pump through a 47mm diameter membrane filter
traditional
method
traditional
method1
real-time
method
EPA FRM Style. The fibers and particles collected on the membrane filter
must be counted or weighed in a laboratory for further analysis.
EPA or OSHA compliance reference method.
• High level of specificity and accuracy.
• Collection of dust particles, which are available for further chemical
analysis.
Dust particles are drawn into the sensor head and are detected once every
second. Dust concentrations are instantaneously calculated and displayed on
the Haz-Dust EPAM-7500’s LCD. All data points are stored in memory for
later analysis.
Immediate estimations of the concentration of a contaminant, permitting
real-time
method1
on-site evaluations.
• Provision of permanent 24-hour records of contaminant concentrations
using continuous monitors.
• Internal audible alarm to warn workers of approaching hazardous
situations.
• Reduction of number of manual filter tests.
• Reduction of number of laboratory analyses.
• Provision of more convincing evidence for presentation at hearings and
litigation proceedings.
• Reduced cost of obtaining individual results.
1
“The Industrial Environment - It’s Evaluation & Control”, U.S. Department of Health & Human Services, CDC,
NIOSH, 1973.
1-3
Page 12
Overview of the EPAM-7500
Ease of use
General
•
Figure 1.4. Diagram showing breathing zones of Inhalable, Thoracic, and Respirable dust
• The user controls all functionality and programming using menus displayed
on a high contrast LCD.
• A 24-hour rechargeable battery capacity.
• Automatic clean air purging of sensor for increased stability and accuracy.
• Internal temperature compensation for ambient use.
The LCD displays real-time concentration in milligram per cubic meter
Information
(mg/m3) in accordance with EPA or OSHA Reference Methods.
• Statistical information of TWA, STEL, Max and Min levels can be viewed
instantly.
• The Haz-Dust is calibrated using Arizona Road Dust (ARD) against NIOSH
method 0600 for Respirable dust with a + 10% accuracy.
• The calibration of the Haz-Dust can be adjusted to compensate for changes
in particle composition and distribution.
particles.
Continued on next page
1-4
Page 13
Overview of the EPAM-7500, Continued
DustComm 7
Series Software
DustComm 7 Series supplied software is designed for more detailed
analysis of sampled data.
Data can easily be exported in comma-delimited ASCII Text Files
importable into spreadsheet programs such as Microsoft Excel and Lotus
1-2-3.
The EPAM-7500 data can be translated to a PC using DustComm 7 Series
Software or Windows Mobile.
The data plots provided with DustComm 7 series enable:
· Detailed statistical analysis.
· Creation of charts.
1-5
Page 14
Real-Time Dust Monitoring Principles
Principles
Figure 1.5. Diagram showing
the principle of near
-
forward light scattering used in the Haz
-
Figure 1.6. Graph illustrating the principle of near
-
forward light scattering.
•
The EPAM-7500 uses the principle of near-forward light scattering of an
infrared radiation to immediately and continuously measure the
concentration in mg/m3 of airborne dust particles.
•
This principle utilizes an infrared light source positioned at a 90-degree
angle from a photo detector.
•
As the airborne particles enter the infrared beam, they scatter the light. The
amount of light received by the photo detector is directly proportional to the
aerosol concentration.
•
A unique signal processes internally and compensates for noise and drift.
This allows high resolution, low detection limits and excellent base line
stability.
Dust..
Light Scattering vs. Gravimetric Filter
1
0.9
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
Response
Relative Optical
00.51
Gravimetric Filter Weight
ARD
ARD
Filter
0.1
ARD = Arizona Road Dust
RTD = Real-Time Dust
1-6
Page 15
Features
Introduction
Real
-
time
•
Functional
•
Operational
Data
display of
features
The EPAM-7500 provides a unique combination of features to provide
superior data quality, ease of use, and flexibility to the user. Below is a
partial list of distinctive features.
Particulate exposure levels.
• TWA, STEL, Min, and Max levels.
• PM 1.0, PM 2.5, PM 4.0, PM 10, or TSP.
• Stored data by date and location code.
• Real-Time graphing capability
• Touch Display
Calibrated to NIOSH methods for lung damaging particles.
• In line concurrent filter samples for gravimetric analysis.
• High sensitivity of 0.001 to 40 mg/m3 (1 µg/m3 – 4000 µg/m3).
• Optional sensitivity of 0.01 to 400 mg/m3
• Interchangeable size-selective sampling inlets.
• Internal air sampling pump.
• Auto purging sensor.
• Easy user access to rechargeable battery and internal filter.
• On-screen programming of sampling and data storage parameters.
features
• Real-time clock.
• User selectable audible alarm.
• In-field zero and span check of instrument calibration.
This chapter describes the steps involved in starting the EPAM-7500 and
configuring its operating parameters.
This chapter contains the following topics.
Turning the EPAM-7500 on and off. 2-2
Using the Main Menu. 2-3
Setting the Date and Time. 2-4
Setting the Alarm. 2-5
Clearing the Memory. 2-6
Topic See Page
2-1
Page 20
Turning the EPAM-7500 On and Off
Introduction
Power
-On
Power
-
Off
Result:
Power can be supplied to the EPAM-7500 either from its internal battery or
from the provided AC power transformer (Battery Charger).
Note: THE BATTERY SHOULD BE FULLY CHARGED before each use. To
charge battery use the EDC supplied charger only. LED indicator on the
EPAM-7500 will turn green when the battery is fully charged. Charge time is
approximately 22 hours and will run for approximately 24 hours. To check
battery status see Page 5-8.
Plug battery pack into EPAM-7500. When power is connected to the EPAM7500, the instrument will turn itself on. There will be a blinking green LED
on the front of the instrument when the EPAM-7500 begins to boot up.
If the EPAM-7500 does not turn on, press the green I/O power button on the
front of the EPAM-7500.
Press the green I/O power button on the front of the EPAM-7500 to
turn the EPAM-7500 off. Also, pressing the red I/O graphic on the display
can turn the EPAM-7500 off or go to standby mode.
2-2
Page 21
Using the Main Menu
Introdu
ction
Accessing the
Using the menu
The EPAM-7500 graphic menu appears on the graphical display.
Note: See Appendix A for menu option flow charts.
The Main Menu is displayed when the EPAM-7500 is first powered on.
main menu
The EPAM-7500 is operated using the following menu selections.
<ON/OFF>
<Settings>
< Upload To Card >
< Review >
< Monitor >
Selection Function
Turns the EPAM-7500 off or into
standby mode.
Changes the settings of the EPAM-
7500.
Allows the user to upload any data
saved from the internal memory to
the Compact Flash Card.
Allows the user to look at the
statistical data before uploading to
the PC or Compact Flash Card
Runs the EPAM-7500. The EPAM7500 does not start logging when the
monitor graphic is pressed. To start
logging in monitor mode, press the
record graphic.
2-3
Page 22
Setting the Date and Time
Introduction
Date and Time
View settings
Settings
Date/Time
Change s
ettings
Settings
The date and time are pre-set by the factory to Eastern Standard Time and are
maintained by an internal clock. It may be necessary to change the date and
time due to local time zones or daylight savings time.
Note: It is important that the system date and time are correct for accurate
record keeping.
Time is entered and displayed in military time format.
settings
Date is entered and displayed in European format (i.e., 03-DEC-2011).
Follow the steps in the table below to check the unit’s date and time.
Step Action
1 Select
2
Follow the steps in the table below to change the unit’s date and time.
Step Action
1 Select
2 Press on time zone drop down menu and select your timezone.
3 Press on the date drop down menu and select the year, month, and
date.
4 Press on the hour position on the time box.
To... Press...
Increase the value of the selected digit.
Decrease the value of the selected digit.
Change minutes or seconds Press on minutes or
5 Press Green check mark when the correct information has been
entered.
from the Main Menu.
is shown on the first screen and in the first three boxes.
from the Main Menu.
< >
< >
seconds and change the
value as described
above.
2-4
Page 23
Setting the Alarm
Introduction
Alarm settings
Using the alarm
Monitor
Alarm
<
<
An audible alarm can be set to alert the worker of approaching threshold
limits.
The concentration level must be set to the defined agency standard for the
particulate type being sampled.
Note: See Appendix B for a partial listing of the most common dust
particulates and their corresponding concentration levels.
Follow the steps in the table below to set the alarm level.
Step Action
1 Select
2 Select
3 Enter the appropriate concentration level using the table below.
Increase the value of the
Decrease the value of the
5 Click on Green check mark to accept the alarm set point.
To run unit with alarm, after setting alarm click on the alarm
graphic from the Main Menu.
graphic.
To... Press...
Scroll the bar to the
selected digit.
or press on the concentration
number and enter the value you
would like to set. >
Scroll the bar to the
selected digit.
or press on the concentration
number and enter the value you
would like to set. >
graphic. This will remove the red X on the alarm. Press the alarm
again to put the red X back on mute the alarm.
Result: The alarm has been set.
2-5
Page 24
Clearing the Memory
Introduction
Clearing
Upload To Card
The memory of the EPAM-7500 can be cleared at any time.
Note: User can select one data file or many data files at once to delete.
Follow the steps in the table below to clear the memory of the EPAM-7500.
memory
Step Action
1 Select
2 Select the files you wish to remove.
3 Select the green check button to move the files to the Compact
Flash Card.
4 Connect the Compact Flash Card to a computer and delete the files.
Note: The EPAM-7500 can also be connected to a computer using
a USB cable and files can be deleted from the PC.
graphic from the Main Menu.
Page 25
Chapter 3 - Operating the Haz-Dust EPAM-7500
Page 26
Operating the Haz-Dust EPAM-7500
Introduction
In this chapter
Chapter Overview
This chapter describes and diagrams operation procedures of the EPAM-
7500.
This chapter contains the following topics.
Selecting the Particle Size: 3-2
EPAM-7500 Impactor Instructions 3-3
The inlet system of the EPAM-7500 can be configured to sample TSP, PM
1.0, 2.5, 10.0 µm dust particulates. The following pages detail the selection
process for each of these particle types. Impactor Sleeve holds one optional
Impactor accessory.
Figure 3-1. Picture: A) Impactor Jet, B) Impaction cup, and C) Impactor
Sleeve. For TSP sampling use impactor sleeve without impactor.
Impactor values are engraved on bottom of impactor jet.
3-2
Page 28
EPAM-7500 Impactor Instructions
Impactor Part Number: EPAM-1.0, EPAM-2.5, EPAM-4.0, and EPAM-10
Impactor Sleeve Part Number: IMS-105
The Impactor consists of the cup and the impactor jet. The Impactor cup fits onto the impactor
jet. The cup has three female connectors and the jet has three pins. User can easily remove for
cleaning without the use of any tools.
STEP 1: The completed assembly of the impactor fits into the impactor sleeve. Insert the cup
end of the impactor first into the impactor sleeve. You will need to insert the impactor
approximately one inch or 25mm down into the impactor sleeve.
Impactor Sleeve
Impactor Jet
Sizes 1.0um, 4.0, 2.5um, 10um
Impaction Cup
remove for applying
impactor grease & cleaning
Impactor Assembly
Push down into sleeve 1
inch (25mm)
3-3
Page 29
STEP 2: Place impactor sleeve with impactor inserted into sensor of EPAM-7500 monitor.
STEP 3: In order to change impactor to different size, to clean or to grease impactor cup,
impactor must be removed. To remove impactor from sleeve, push upward with the trim stick
provided with model EPAM-5000, a pencil, or similar from bottom of sleeve.
Push up to remove
3-4
Page 30
1.0um Dust Particulates
Settings
Sample Type
1.0 um
– E
Green
Follow the steps in the table below to select PM-1.0 dust particulates.
Step Action
Install 47mm filter if
desired.
Remove filter holder
by unscrewing
bottom round cover.
1 Select
2 Select
3 Select
4 Select
6 Insert the sampling inlet into the sensor head of the EPAM-7500.
Note: If also collecting concurrent 47mm EPA FRM Style filter
samples place a clean gravimetric filter in the filter cassette. The
flow rate should be checked each time a new gravimetric filter is
used. See page 5-5 for information on checking the flow rate.
7 Attach the filter cassette to the sensor head of the EPAM-7500.
See diagram below.
8 Turn to page 3-7 and follow the instructions to Manual-Zero the
EPAM-7500.
Figure 3-2. Diagram of sensor lid. To unlock sensor lid unscrew pem screw
located on right side of sensor lid to expose the filter cassette. *NOTE: When
closing sensor cover be sure not to tighten pem screw too tightly the spring
inside the screw can break.
from the Main Menu.
.
.
check mark button to accept Changes.
3-5
Page 31
2.5 um Dust Particulates
Settings
Sample Type
2.5 um
– S
Green
Follow the steps in the table below to select PM-2.5 dust particulates.
Step Action
1 Select
2 Select
3 Select
4 Select
6 Insert the sampling inlet into the sensor head of the EPAM-7500.
Note: If also collecting concurrent 47mm EPA FRM Style filter
samples place a clean gravimetric filter in the filter cassette. The
flow rate should be checked each time a new gravimetric filter is
used. See page 5-5 for information on checking the flow rate.
7 Attach the filter cassette to the sensor head of the EPAM-7500.
See diagram below.
8 Turn to page 3-7 and follow the instructions to Manual-Zero the
EPAM-7500.
from the Main Menu.
.
.
check mark button to accept Changes.
3-6
Page 32
4.0um Dust Particulates
Settings
Sample Type
4.0 um
– SM
Green
Follow the steps in the table below to select PM-4.0 dust particulates.
Step Action
1 Select
2 Select
3 Select
4 Select
6 Insert the sampling inlet into the sensor head of the EPAM-7500.
Note: If also collecting concurrent 47mm EPA FRM Style filter
samples place a clean gravimetric filter in the filter cassette. The
flow rate should be checked each time a new gravimetric filter is
used. See page 5-5 for information on checking the flow rate.
7 Attach the filter cassette to the sensor head of the EPAM-7500.
See diagram below.
8 Turn to page 3-7 and follow the instructions to Manual-Zero the
EPAM-7500.
from the Main Menu.
.
.
check mark button to accept Changes.
3-7
Page 33
10.0um Dust Particulates
Settings
Sample Type
10
.0 um
– M
Green
Follow the steps in the table below to select PM-1.0 dust particulates.
Step Action
1 Select
2 Select
3 Select
4 Select
6 Insert the sampling inlet into the sensor head of the EPAM-7500.
Note: If also collecting concurrent 47mm EPA FRM Style filter
samples place a clean gravimetric filter in the filter cassette. The
flow rate should be checked each time a new gravimetric filter is
used. See page 5-5 for information on checking the flow rate.
7 Attach the filter cassette to the sensor head of the EPAM-7500.
See diagram below.
8 Turn to page 3-7 and follow the instructions to Manual-Zero the
EPAM-7500.
from the Main Menu.
.
.
check mark button to accept Changes.
3-8
Page 34
Auto-Zero
Introduction
Auto
-
Zero
Settings
Auto Zero
Green
Auto-Zero purging feature automatically adjusts for baseline drift due to
severe ambient temperature change.
This feature is a default setting on the EPAM-7500.
The Auto-Zero feature purges the sensor optics with clean air and reestablishes the baseline every 30 minutes as a default.
To deactivate or reactivate the Auto Zero purging feature follow the table
below.
Step Action
1 Select
2 Select blue
3 Select
4 Select desired Auto Zero time.
5 Select
from the Main Menu.
drop down menu.
check button to accept changes.
3-9
Page 35
Sampling
Introduction
Conditions
Sampling
Monitor
Record
Note:
Monitor
Once you have selected a Particle Size and completed the EPAM-7500 is
ready to begin sampling.
Note: It is recommended to trial sample in the environment where you will be
air monitoring for 15 minutes to allow EPAM-7500 to stabilize prior to
beginning first sample.
The following conditions should be met before starting the sampling process.
Condition... For further Information
See Page...
The correct particle size must be selected. 3-2
The correct sampling inlet must be attached. 3-2 to 3-8
The date and time must be checked and/or set. 2-4
The alarm level must be set if sampling with
2-5
the alarm feature.
Follow the steps in the table below to begin the particle sampling.
Step Action
1 Turn unit on.
2 Select
3 Select the
4
See page 3-13 for explanation of location codes.
5 When in
.
graphic to start recording.
, select the information graphic to see the
settings of the EPAM-7500.
3-10
Page 36
Location Codes
Introduction
Maximum
Assigning
The EPAM-7500 assigns a location code to each sampling sequence. The
active location is indicated in information screen (See figure 3.10).
The EPAM-7500 can store a total of greater than 120,000 data points.
location codes
The location code assigned to the site is determined by the date and runs for
location codes
the day (ex. P10NOV2012-000).
3-11
Page 37
Reviewing Stored Data
Graphical
Viewing data
Review
Record Name
Green
Introduction
display
The EPAM-7500 provides extensive capabilities for reviewing internally
stored data and statistics on the graphical display or downloading to a PC
using DustComm 7 Software (Chapter 4).
The following information can be displayed on the graphical display.
Display Description Tag #
Date Date of sampling.
Start Time sampling began.
Stop Time sampling was terminated.
Duration Time of occurrence of reported statistic.
MAXIMUM Sample Highest concentration of dust particles.
MINIMUM Sample Lowest concentration of dust particles.
T.W.A. Time weighted average concentration of dust
Samples How many samples were taken during the logging
S.T.E.L. Short-term exposure limit.
particles.
session.
on the
graphical
display
Follow the steps in the table below to review stored information and statistics.
Step Action
1 Select
2 Select
3 Select
graphic
.
check mark button.
4 Review Screen will have the data from the above chart.
3-12
Page 38
Chapter 4– DustComm 7 Software
Page 39
Introduction to the DustComm 7 Software
Introduction
Spreadsh
eet
Data plots
Open a File
Statistics
Notes
Export to Excel
Import into
DustComm 7 is a powerful and flexible Windows application software
package designed for use with the EPAM-7500 Particulate Monitoring
Equipment.
DustComm 7 is a data manipulation tool, enabling detailed analysis and
reporting of sampled data.
DustComm 7 easily translates data into spreadsheet ASCII text files. These
applications
files can be open into spreadsheet programs such as Microsoft Excel
The data plots provided with DustComm 7 enable:
• Detailed statistical analysis.
• The creation of charts.
To open a file in DustComm 7 software, press file, then open. Go to where
the file was saved on the Flash Card and open the file.
DustComm 7 will display a graph of the data at the bottom of DustComm7.
The statistical data is above the graph.
Notes can be added to the DustComm 7. The notes field is on the right side of
the statistical data.
To export the data into excel, press File and select export to CSV.
To import the data into excel, open Microsoft Excel. Press File and Open.
Excel
Open the CSV data file from above. When importing the data, ensure to select
comma delimited data.
4-1
Page 40
File Menu
DustComm 7 Software Screen Shot
File Name
Statistical
Data
Graphed
Data
Notes
4-2
Page 41
Connecting the EPAM-7500 to the PC USB
Introduction
Windows
The EPAM-7500 can also download the data through the USB connector on
the top of the instrument. To download the EPAM-7500 through USB, the PC
would need ActivSync. ActivSync can be downloaded from the Microsoft
website. The address is http://www.microsoft.com/en-
us/download/details.aspx?id=15 to download the ActivSync.
Windows Mobile can also be downloaded on Microsoft’s website if your
computer does not have the software included. The address to download
Windows Mobile is http://www.microsoft.com/en-
us/download/details.aspx?id=14 for a 32-bit system. Windows Mobile can be
downloaded and installed instead of using ActivSync.
Windows Mobile can also be used on the PC with ActivSync to browse the
Mobile
files on the flash card as well as the internal memory.
4-3
Page 42
Windows
Mobile
Once the EPAM-7500 is connected, click on the browse files. You can drag
the files you want onto the desktop and use the DustComm 7 software.
4-4
Page 43
Chapter 5 - Maintenance
Page 44
Maintenance
Introduction
In this chapter
Chapter Overview
This chapter covers the routine maintenance procedures for the Haz-Dust
EPAM-7500.
This chapter contains the following topics.
Checking the Calibration Span. (Optional Accessory). 5-2
Checking the Flow Rate. 5-4
Adjusting the Flow Rate. 5-5
Battery Maintenance. 5-6
Cleaning the Impactors 5-8
Cleaning the Sensor Optics. 5-10
Topic See Page
5-1
Page 45
Checking the Calibration Span
Introduction
When to check
Conditi
ons
The Span Reference Insert Part Number CS-105 is a light scattering device
that provides a constant value (termed a “k” factor).
The Span Reference should be used as a reference to check factory calibration
of the EPAM-7500 Monitor.
The calibration span should be checked under the following conditions:
the calibration
span
• Once a month with normal usage. (Normal use is twice a week).
• If the EPAM-7500 is dropped or otherwise damaged.
• The first time you use the unit to double check the factory calibration.
Note: The EPAM-7500 must be sent into EDC annually for recalibration. In
accordance with ISO procedures.
The following conditions must be met before checking the calibration span.
Condition... For further information
The Environment must be clean. --The Battery must be fully charged. 5-7
Figure 5-1. Inserting CS-105 into Haz-Dust Sensor.
see page...
5-2
Page 46
Checking the Calibration Span, Continued
Checking
Monitor
Follow the steps in the table below to check the Span of the EPAM-7500
Calibration
Span
Monitor.
Note: Failure to follow this procedure in its entirety may cause an incorrect
“k” value reading.
Step Action
1 Turn EPAM-7500 on.
2 Press
3 When the EPAM-7500 starts to begin monitoring, the instrument
will run an auto zero one minute after beginning.
5 Remove Auto Zero filter and impactor sleeve.
6 Insert the span reference into the sensor head (see diagram 5.1).
Allow 2-3 minutes for reading to stabilize.
Note: Be sure the locating pin on the calibration reference slides
into the locating hole on the sensor head. Also, Push down on the
calibration reference to be sure it is aligned properly. Slightly twist
CS-105 clockwise when positioned in alignment hole. This will
ensure the CS-105 is in the exact position and ensure
reproducibility when “K” value is displayed on the EPAM monitor.
7 Observe the printed “k” value on the calibration reference.
Note: The printed “k” value should match the concentration value
shown on the Haz-Dust LCD to within + 10%.
graphic.
IF THEN
The numbers agree
within + 10%
The two numbers
do not agree within
+ 10%.
8 Remove Span Reference and place in its protective sleeve.
Note: Optical windows on CS-105 cannot have fingerprints or
contamination, please clean using KK-101 Cleaning Kit.
Note: If the CS-105 is purchased from SKC or as an after-market accessory,
the end user must assign “k” value. To assign “k” value repeat steps 1-6 three
times. Take an average of the three numbers you recorded to get your “k”
value.
The EPAM-7500 has passed the span
check test.
1. Repeat the process to rule out error,
then:
2. Call EDC technical support or
return the EPAM-7500 for
recalibration.
5-3
Page 47
Checking the Flow Rate
Introduction
Checking the
Monitor
It is good technique to check the flow rate every time before and after a new
gravimetric filter is used for sampling.
Use the steps in the table below to check the flow rate.
flow rate
Notes:
• When using an impactor, be sure the proper sampling inlet is attached to
the sensor head.
Step Action
1 Attach your airflow calibrator to the EPAM-7500 using the table
below.
If sampling... Then...
PM-10, PM-
4.0, PM-2.5,
PM-1.0 or
TSP
2 Activate the internal sampling pump using the steps below.
Step Action
1 Select
3 Observe the flow rate on your air flow calibrator.
If... Then...
1. Attach one end of the calibration airflow
tubing to flow adapter.
2. Connect the other end of the calibration airflow
tubing to your airflow calibrator.
from the Main Menu.
The flow rate
is 2.0 LPM.
The flow rate
is not 2.0
LPM.
No flow is
present.
The flow rate is properly calibrated. Detach the
airflow calibrator tubing and continue with the
“Selecting the Particle Size” process.
The flow rate must be adjusted. See page 5-7 for
instructions.
See Troubleshooting section on Chapter 6.
5-4
Page 48
Adjusting the Flow Rate
Introduction
Adjusting the
Settings
Lock
white box
HD7
Enter
Check Box
Force Pump On
Pump
Green
The flow rate must be adjusted when it does not equal 2.0 LPM.
Follow the steps in the table below to adjust the flow rate.
flow rate
1 Press
2 Press the button until you are on the last screen.
3 Press the
4 Press the
5 Press
6 Press the
7 Press the
8 Press the or to adjust
9 Record the Flow Rate.
10 Detach the airflow calibrator and calibration airflow tubing.
11 Press
from the main menu.
graphic to unlock the settings.
to enter the security code.
on the keyboard.
button on the keyboard.
next to the
to 2.0 LPM.
check button to accept changes.
.
5-5
Page 49
Battery Maintenance
Introduction
Checking the
Recharging the
The battery pack is a 10.8-volt Li-Ion rechargeable battery. It is important to
check the battery periodically and recharge when necessary.
The battery status can be checked from any screen on the EPAM-7500. The
Battery
battery
graphical icon is on the bottom middle of the screen.
Follow the steps in the table below to recharge the battery using the supplied
charger.
Note: If the battery is too low, the EPAM-7500 will turn off. Charge the
battery for 15 minutes before turning on.
Step Action
1 Plug the battery charger into an electrical outlet.
2 Plug the battery charger into the battery charge jack on the top
panel of EPAM-7500, or into the supplied charging cord.
Results:
• LED indicator on front of the EPAM-7500 should be Red.
• When LED is green battery is fully charged.
Note: Recharging time is approximately 22 hours when the
instrument is not operating.
CAUTION:Do not charge in a hazardous environment. Use only
the EDC approved charger designed for the EPAM-7500.
5-6
Page 50
Battery Maintenance, Continued
Removin
g and
Removing the
The battery pack can be removed and replaced whenever necessary.
replacing the
battery pack
Follow the steps in the table below to remove the battery pack.
battery
Step Action
1 Loosen the two retaining pem screws from the top plate of the
EPAM-7500.
NOTE: These are special inserts and the screws are attached by a
spring to tighten push the screw down and turn to loosen turn the
screw and it will pop on the spring DO NOT pull the screw
because you will break the spring and be unable to secure your
battery pack.
2 Slide the battery plate sideways out of the unit.
3 To re-install reverse above proceeding
Figure 5-3. Diagram of battery being removed from Haz-Dust.
5-7
Page 51
Cleaning the Impactors
Cleaning the
Introduction
Impactors should be disassembled and cleaned and greased at regular
intervals.
Example: You should clean impactors every thirtieth sample or once a month
to start, but heavy loadings are observed on the target disk, as often as
appropriate.
Impactors
Follow the Steps in the table below to clean the impactors.
Step Action
1 Pushing with a pen from bottom remove the impactor through
top of impactor sleeve into the palm of your free hand.
2 Rinse the impactor from top to bottom with a solvent
(hexane, white gas, lantern gas) using a squeeze bottle,
paying particular attention to the impaction target disks. An
acceptable alternative method of cleaning involves the use of
an ultrasonic bath with mild soapy water solution.
Note: The impaction cup should be removed prior to regreasing. This is accomplished by pulling the impaction cup
apart from the impactor jet gently.
3 Let all parts of the impactor air-dry.
4 Prepare a mixture of solvent and impactor grease (Apiezon ®
M, Glisseal ® Ht) or similar grease in a dropper bottle until
thoroughly mixed and of a fluid consistency. Use a 1-inch
length of grease to 30ml of solvent. Vigorously shake the
mixture until an opaque, uniform suspension, free from
grease globs, is obtained.
5-8
Page 52
5 Put two or three drops of the solution on the impaction cup.
The drops should saturate the disk, flowing freely to the edge.
6 Let the impaction cup “dry” by allowing the solvent to
volatilize, leaving a thin film of grease on the impaction cup.
7 Replace the impaction cup onto the impactor jet. Re-insert
the impactor into the impactor sleeve.
5-9
Page 53
Cleaning the Sensor Optics
Introduction
Cleaning the
Figure 5
-
4. Picture of the cleaning kit.
Although the EPAM-7500 has an internal sensor purge it is important to keep
the sensor optics of the EPAM-5000 clean to ensure the integrity of the
optical sensor.
The sensor optics needs to be checked every 2 months or on a weekly basis,
when used in a 2 to 3 mg/m3 T.W.A. environment.
Follow the steps in the table below to clean the sensor optics.
Sensor Optics
5-10
Page 54
Cleaning the Sensor Optics, Continued
Step Action
1 Remove the screws located on sensor cover.
2 Inspect internal sensor. Inspect cavities for residual dust. Blow the
dust away with low-pressure air or wipe with foam tipped cotton
swab.
3 Inspect the glass lens covers for dust.
Note: Use one of the following methods to clean the glass lens.
• Blow the dust away with low pressure air, or,
• Use a small amount of isopropyl alcohol and wipe with cotton
swabs.
CAUTION: Do not spill any alcohol into the internal cavity of
the EPAM-7500.
4 Replace the sensor cover.
5 Tighten the screws snugly into place.
NOTE: If screws are not tightened enough the flow of the EPAM7500 could be off.
5-11
Page 55
Appendix
Page 56
Appendix A Menu Screens
Introduction
This Appendix contains a complete overview of the Menu Screens for the
EPAM-7500.
The following pages show diagrams of the Main Menu Screens.
Note: Selecting Red X button will cancel from any Menu Screen will back
out of the current function.
Monitor
Run
Instrument
Record
Instrument
Settings
Alarm
Review
Unit/Card
Files
Upload To
Card
Compact
Flash Card
Settings
Date/time
Sample type
Auto Zero
Sensor
Settings
Flow Rate
Adjustment
Power
OFF
Suspend
Appendix A
Page 57
Appendix B NIOSH/OSHA Particulate Air Monitoring
Reference
µm Micron, 1/1000 of a meter.
ARD Arizona Road Dust
Inhalable Dust
Particulates
LPM Liters per minute.
mg/m3 Milligrams per cubic meter.
NIOSH National Institute of Occupational Safety & Health
OSHA Occupational Safety & Health Administration.
Respirable Dust
Particulates
STEL Short-term exposure level. Maximum dust
Thoracic Particulates Particulates having a 50% cut point at 10µm.
TWA Time Weighted Average. Average particulate
Particulates having a 50% cut point at 100µm.
Particulates having a 50% cut point at 3.5µm.
concentration over a 15-minute period.
concentration over a period of time.
Appendix C
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.