SKC Limited BioLite User Manual

Page 1
BioLite Air Sampling Pump
PRODUCT PHOTO
CENTRED
This manual covers the following model:
228-9610
www.skcltd.com
Operating Instructions
Page 2
Thank you for choosing an SKC product. Your purchase is covered by our warranty, details of which can be found inside the rear cover of this manual.
Product Model Number Product Serial Number Date of Purchase
228-9610
SKC recommends annual servicing of this product. The first service is due one year from the date of purchase, and then at yearly intervals on this date.
Service Date Service Date Service Date
1 5 9
2 6 10
3 7 11
4 8 12
Please note that SKC Ltd are the only authorised service centre in the UK, guaranteeing you access to the full range of genuine SKC replacement parts. For all other areas a full list of SKC approved distributors and service centres can be found at www.skcltd.com
SKC UK service centre - Tel: +44 (0)1258 480188 Fax: +44 (0)1258 480184 Email: info@skcltd.com
Page 2 007-01-002 Issue C www.skcltd.com
Page 3
Contents
Specifications ................................................................................................................................................2
General Information .......................................................................................................................................3
The BioLite Pump ..........................................................................................................................................4
Operating the BioLite Pump ...........................................................................................................................5
Troubleshooting .............................................................................................................................................6
Replacement Parts and Accessories .............................................................................................................7
Warranty Information .....................................................................................................................................8
www.skcltd.com 007-01-002 Issue C Page 1
Page 4
Specifications
• Maximum flow rate: 30.5 litre/min
• Maximum back pressure: 800mbar
• Dimensions: W x H x D 162 x 209 x 338mm
• Weight: 6.05kg
• Enclosure rating - IP20
• Operating temperature: 0°C to +40°C
• Operating humidity: 0 to 95% Rh
• Mains supply: 220/230V ~ 50Hz
• Maximum input current: 0.66A
• Operating noise level: 67 - 71dBA @ 1metre
30
20
10
Flow Rate (litre/min)
0
0 100 200 300 400 500 600 700 800 900 1000
Page 2 007-01-002 Issue C www.skcltd.com
Back Pressure (mbar suction)
Page 5
General Information
1) Care of the BioLite Pump
• Do not block the ventilation grill at the front of the pump or the cooling fan outlet grill at the rear of the pump during operation.
• Do not run the pump with the inlet or outlet ports blocked, or with the flow adjuster fully closed. Failure to comply could lead to overheating and seizure of the pump mechanism. The pump is fitted with an overtemperature trip, which will disable operation of the pump motor. This trip will automatically reset when the pump temperature drops to an acceptable level.
• Do not open the pump casing when the mains supply is connected to the pump.
• A small quantity of carbon dust will be exhausted from the pump outlet in normal operation due to normal
wear of the graphite pump rotor vanes.
• The BioLite pump case is IP20 rated, it is not rated as water or splashproof and therefore must not be used where it is possible for water to enter the pump casing.
Warning - Failure to follow these guidelines will void the product warranty.
2) The WEEE Directive
This product is marked with the crossed out wheelie bin symbol, which identifies that it falls within the scope of the EC Directive 2002/96/EC on waste electrical and electronic equipment (WEEE). At the end of it’s useful life, this product must be disposed of in an environmentally sound way as detailed in the Directive. Please contact your local distributor or SKC Ltd for further details on how to comply with the requirements of the WEEE Directive.
www.skcltd.com 007-01-002 Issue C Page 3
Page 6
The Biolite Pump
Back pressure gauge
Flow adjuster
-15
-10
-20
-0.6
-0.4
-0.8
-25
-0.2
-5
-1 bar inHg
-30 0
Locking nut
Carry handle
On/off switch Cooling fan outlet
Pump outlet
228-9610
HIGHVOLUME SAMPLING PUMP
INPUT:220/230V ~50Hz 0.66A
INLETFUSE RATING- 2A
ENCLOSURERATING -IP20
WEIGHT- 6.05kg MAXIMUMFLOW RATE:30.5 l/min MAXIMUMBACK PRESSURE:0.8 bar
NOTE- PUMPMUST NOTBE RUNWITH
FLOWADJUST VALVEFULLY CLOSED.
S/No 123456
www.skcltd.com
Mains inlet
Accessory mounting pointsPump inlet
Page 4 007-01-002 Issue C www.skcltd.com
Fuse holder
Page 7
Operating the BioLite Pump
1. Insert the mains lead into the IEC mains inlet at the rear of the pump and plug the lead into a suitable mains outlet socket. Both UK 3 pin and Euro 2 pin mains leads are supplied with the unit, select the correct lead for your location.
2. Turn the flow adjuster fully counter-clockwise to ensure that it is fully open. Switch on the mains supply. Switch on the pump using the on/off switch at the rear of the pump. Allow the pump to run for five to ten minutes to stabilise before commencing any flow calibration. Switch the pump off using the on/off switch at the rear of the pump.
3. Connect the sample train to the pump inlet hosetail. Switch on the pump using the on/off switch. Adjust the flow rate if required using the flow adjuster. Turn the flow adjuster clockwise to reduce the flow rate and anti-clockwise to increase the flow rate. Allow the pump to run for five minutes for the flow rate to stabilise before taking any calibration readings. When the desired flow rate has been set, the flow adjuster can be locked using the locking nut.
4. The back pressure gauge gives a reading of the suction generated by the pump due to the restriction to flow of the sample train. This is important for indicating correct operation of the SKC Biosampler - the Biosampler achieves sonic flow at a suction of 15in Hg (refer to the Biosampler instructions for further details).
www.skcltd.com 007-01-002 Issue C Page 5
Page 8
Troubleshooting
Possible Fault Corrective Action
• Check that mains power is present at the mains socket.
• Ensure that pump on/off switch is in the on position.
Pump does not operate
Pump cannot acheive required flow rate
Pump stops due to over­temperature
Pump cooling fan running, pump not running and produces ‘humming’ noise.
Page 6 007-01-002 Issue C www.skcltd.com
• Check for blown fuse in the UK 3 pin mains plug and replace with an identical fuse if required.
• Check for blown inlet fuse at the rear of the pump and replace with an identical fuse if required.
• Check for leakages in the sample train.
• Pump vanes damaged or excessively worn. Return pump to SKC
Limited for service / repair.
• Check for flow blockages in the sample train and rectify if required. Wait for pump to cool before restarting.
• Check that the flow adjust control is not fully closed. Wait for pump to cool before restarting.
• Check that pump cooling fan is running. If cooling fan is not running return pump to SKC Limited for service / repair.
• Pump rotor vanes have jammed. Switch off pump immediately. Return pump to SKC Limited for service / repair.
Page 9
Replacement Parts and Accessories
Part No. Description
Accessories
225-9593 Biosampler kit with 5ml vessel
225-9595 Biosampler kit with 20ml vessel
228-9611 Biosampler mounting bracket for BioLite pump
Replacement Parts
P228100 Inlet fuse (2A)
P228101 Pump rotor vanes (set of four)
P901108 Hosetail
If the required item is not listed, contact your supplier or SKC sales on +44 (0) 1258 480188.
SKC provide an extensive range of sampling media, including filters, sorbent tubes and impingers. A full selection can be found in the current SKC catalogue and at www.skcltd.com
www.skcltd.com 007-01-002 Issue C Page 7
Page 10
Warranty Information
Limited One Year Warranty
1. SKC warrants that this instrument, and each of its component parts, provided for occupational health and safety and environmental applications is free from defects in workmanship and materials under normal use for a period of one (1) year. This warranty DOES NOT cover any claims due to abuse, misuse, neglect, alteration, or accident, or use in application for which the instrument was either not designed or not approved by SKC, or, due to the buyer’s failure to maintain normal maintenance, improper selection or misapplication. This warranty shall further be void if changes or adjustments to the instrument are made by a person other than an employee of the seller or, if the operating instructions furnished at the time of installation are not complied with.
2. SKC hereby expressly disclaims all warranties either expressed or implied, including any implied warranties of merchantability or fitness for a particular purpose and neither assumes nor authorises any person to assume for it any liability in connection with the sale of these instruments. No description of the goods being sold has been made a part of the basis of the bargain or has created or amounted to an express warranty that the goods will conform to any such description. The Buyer shall not be entitled to recover from SKC any consequential damages; damages to property, damages for loss of use, loss of time, loss of profits or income or any other incidental damages. Nor shall the Buyer be entitled to recover from SKC any consequential damages resulting from defect of the instrument.
3. This warranty extends only to the original purchaser of the warranted instrument during the term of the warranty, the buyer may be required to present proof of purchase in the form of a paid receipt for the instrument.
4. In the event of a defect, malfunction, or other failure of the instrument not caused by any misuse or damage to the instrument while in the possession of the Buyer, SKC will remedy the failure or defect without charge to the buyer. The remedy will consist of service or replacement of the instrument, or refund of the purchase
Page 8 007-01-002 Issue C www.skcltd.com
Page 11
Warranty Information
price, at the option of SKC. However, SKC will not elect refund unless it is unable to provide replacement and repair is not commercially practicable.
5. The terms of this warranty begin on the date the instrument is delivered to the buyer and continue for a period of one (1) year.
6(a) To obtain performance of any obligation under this warranty, the buyer shall return the instrument, freight prepaid to SKC at the following address:-
SKC Limited
11 Sunrise Park
Higher Shaftesbury Road
Blandford Forum
Dorset DT11 8ST
t: 44 (0) 1258 480188
f: 44 (0) 1258 480184
6(b) To obtain further information on the warranty performance contact SKC.
7. This warranty is provided under English law.
8. No other warranty is given by SKC in conjunction with this sale.
The disclaimers and limitations shall not affect the statutory rights of the consumer.
www.skcltd.com 007-01-002 Issue C Page 9
Page 12
SKC Limited 11 Sunrise Park, Higher Shaftesbury Road, Blandford Forum, Dorset DT11 8ST United Kingdom
007-01-002 Issue C March 2009
T: +44 (0) 1258 480188 F: +44 (0) 1258 480184 E: info@skcltd.com W: www.skcltd.com