Single Polypropylene Fitted Bags
232, 236, 252, and 262 Series Sample Bags
The 232, 236, 252, and 262 Series sample bags feature a single polypropylene  ing that is used for both fi lling the bag and removing the sample for analysis. The  ing contains both a syringe port with PTFE-lined septum and a hose connection and acts as a shut-off valve for the hose connection.
Guidelines for Bag Sampling
1. Ensure that the bag material and  ings are appropriate for the compounds to be sampled.
2. Use only PTFE tubing to connect the sample bag to the pump to prevent sample loss by adsorption on tubing walls.
3. Before using, fl ush the bag thoroughly with purifi ed air or nitrogen.
4. Analyze the sample within 24 to 48 hours. Long-term storage of air­contaminant mixtures in bags is not recommended.
5. Do not ship sample bags by air unless the cargo cabin is pressurized.
6. Infl ate bag properly; see page 3 for details.
FlexFoil Bag Note: Due to the physical properties of foil bag material, minor crimps and creases may appear on the bags during SKC manufacturing and quality testing as well as user handling. These marks only affect the aesthetics of the bag and not its performance. Foil bags exhibiting minor crimps and creases are not subject to replacement or refund as functionality is not affected.
Cautions
Do not use sample bags to sample compounds with boiling points > 249.8 F (121 C).
SKC sample bags are designed for sampling air at atmospheric pressure only. A empting to pressurize the bag can result in bag rupture and sample loss. Do not ship bag samples by air freight in non-pressurized cargo cabin. Bags can burst under such conditions.
All federal and state packaging and transporting regulations apply.
Failure to follow warnings and cautions voids any warranty.
Sample bags are designed for single use only.
Page 2
Operation
¨
000277
1. Flush the bag at least 3 times with purifi ed air or nitrogen before use.
2. To fi ll a bag, connect PTFE tubing from the exhaust port of an air sample pump to the hose connection on the bag (stem protruding from the side of the  ing).
3. To open the shut-off valve, hold the side stem and turn the entire upper portion of the  ing (including the brown syringe port and the white section to which it is a ached) counterclockwise one revolution. Turn on the pump and sample (see Proper Bag In ation).
CAUTION: Do not turn side stem.
Do not use sample bags to sample compounds
with boiling points > 249.8 F (121 C).
4. When sampling is complete, turn off the pump. To close the shut-off valve, hold the side stem and turn the entire upper portion of the  ing (including brown syringe port and the white section to which it is a ached) clockwise until it is snug.
5. To withdraw samples using a needle and syringe, carefully insert the needle into the septum port in the center of the brown cap and pierce the septum.
Do not allow the needle to puncture the bag material when piercing the septum.
2
Page 3
Valves
SKC bag  ing valves are extremely durable but are not intended for use as handles or hanging devices. This type of handling may damage the seal causing leakage. It is considered to be misuse and will void the SKC warranty.
Proper Bag Infl ation
Avoid fi lling any bag more than 80% of its maximum volume. At the end of the sampling period, turn off the pump and close the valve on the bag.
Incorrect
Infl ation
Correct
Infl ation
Filling Bags Using Negative Pressure
The SKC Vac-U-Chamber is a rigid air sample box that allows the direct fi lling of an air sample bag using negative pressure provided by most personal air sample pumps. When using the Vac-U-Chamber, the air sample enters the bag directly, without passing through the pump. This eliminates the risk of contaminating the pump or sample. The Vac-U-Chamber has rigid walls that will not collapse under vacuum conditions, thus eliminating possible errors. All surfaces in contact with samples are constructed of inert materials.
Small Vac-U-Chamber (supplied without pump) with polypropylene fi ttings;
suitable for use with 1-liter bags .................................................................Cat. No. 231-940
Large Vac-U-Chamber (supplied without pump) with stainless steel fi ttings;
suitable for use with 8-liter bags .................................................................Cat. No. 231-939
Extra-large Vac-U-Chamber (supplied without pump) with stainless steel fi ttings;
suitable for use with 10-liter bags ...............................................................Cat. No. 231-944
Jumbo Vac-U-Chamber (supplied without pump) with stainless steel fi ttings;
suitable for use with 25-liter bags ...............................................................Cat. No. 231-946
See Accessories on page 4 for tubing.
SKC sample bags are available in Tedlar®, FluoroFilm FEP, SamplePro FlexFilm, Standard FlexFoil®, and FlexFoil PLUS with fi ttings made of stainless steel, polypropylene, or PTFE in sizes ranging from 0.5 to 100 liters. SKC also manufactures custom sample bags. Contact SKC at 724-941-9701 or skcorder@skcinc.com.
For sample bag stability reports,
visit www.skcinc.com/bags.asp.
3
Page 4
Accessories
PTFE Tubing, ts over bag tting and Grab Air pump exhaust,
3/16-inch ID, 1/4-inch OD ......................................................... 10 feet ....Cat. No. 231-9-23
PTFE Tubing, fi ts inside bag fi tting,
1/16-inch ID, 1/8-inch OD ......................................................... 10 feet ....Cat. No. 231-9-21
PTFE Tubing, fi ts Vac-U-Chamber inlet and 222 pump exhaust,
1/4-inch ID, 5/16-inch OD .......................................................... 10 feet ..... Cat. No. 231-937
50 feet .....Cat. No. 231-924
Twin Port Pocket Pump Tubing Adapter Kit includes two lengths of silicone tubing: 1/8-inch ID, 1/4-inch OD for bag tting and 3/16-inch ID, 3/8-inch OD for pump fi tting; use with
Cat. No. 231-9-23 PTFE tubing (above) .................................................... Cat. No. 231-926