863 Valley View Road, Eighty Four, PA 15330 USA
Tel: 724-941-9701 Fax: 724-941-1369 e-mail: skctech@skcinc.com
Operating Instructions
Flow Rate Guide
Select the fl ow rate that will provide the desired 50% cut-point to capture the
particle size of interest.
15 L/min is recommended as an optimal fl ow rate.
50% Cut-point (μm)
VersaTrap Spore Trap Cassettes
Cat. Nos. 225-9820 and 225-9821
Flow Rate
30.0 1.5
25.0 1.7
20.0 1.9
15.0 2.3
10.0 2.8
5.0 3.9
VersaTrap
Sampling Interval Guide
Following are recommended sampling times for specifi c dust conditions.
VersaTrap Spore Trap Casse es
1. Select a fl ow rate (see Flow Rate Guide on back) and calibrate pump fl ow rate.
VersaTrap
®
Casse es do not produce signifi cant back pressure, therefore, it is
not necessary to calibrate fl ow with a casse e in line.
2. Remove and retain sealing tabs from casse e inlet and outlet.
3. Connect casse e outlet to pump inlet.
a. Direct method: If pump accommodates direct mounting, insert casse e
adapter into pump inlet and push casse e outlet onto bowl end of
adapter.
b. Indirect method: Use 1/2-in ID tubing to connect casse e outlet to pump
inlet.
4. Turn on sample pump and sample for appropriate sampling time. See
Sampling Interval Guide on back.
5. When sampling is complete, turn off pump, remove casse e, and reseal with
sealing tabs.
6. Place sample casse es, outdoor background samples, non-complaint
area samples, and blanks in a corrugated box and ship to a laboratory for
analysis.
SKC recommends reducing length of sampling time when using a
30 L/min fl ow rate to prevent overloading.
Recommended Sampling Intervals for the VersaTrap Air Sampling Cassette
Typical Environmental
Conditions
Clean “offi ce” or
outdoors (no visible dust)
Indoor environment,
high activity personnel
Indoor environment,
drywall renovation, or
heavy industrial dust
at Typical Collection Flow Rates
Flow rate
Collection
@ 15 L/min
10
minutes
5
minutes
1
minute
Flow rate
Collection
@ 30 L/min
5
minutes
2
minutes
1/2
minute