SKC 225-9690 User Manual

Operating Instructions
863 Valley View Road, Eighty Four, PA 15330 USA
Tel: 724-941-9701 Fax: 724-941-1369 e-mail: skctech@skcinc.com
Sampling
Sampling Tip: Take a sample from a non-suspect wall cavity to use as a negative control.
1. Remove sealing labels from a new casse e and a ach to adapter without disturbing adapter tubing in wall.
Wall Sampling Guide
with VersaTrap Cassettes
Wall Adapter
QuickTake® 30
Sample Pump
Complete Wall Sampling train.
Calibration and Preparation
1. Calibrate sample pump with representative VersaTrap® Casse e in line to desired fl ow rate (15 L/min is recommended for VersaTrap). Remove casse e a er calibration.
2. Use drill with bit to create a 3/8-inch hole in wall 3 to 6 inches from fl oor.
Caution: Drill slowly to minimize aerosolization of gypsum dust.
3. Insert angled end of VersaTrap Wall Adapter tubing into hole.
4. Prepare a "vacuum" casse e by removing labels from inlet and outlet of a casse e. A ach rectangular end of adapter to casse e inlet. Use "vacuum"
casse e only to collect excess dust.
(inserted in wall)
VersaTrap
Cassette
Flow Rate: 15 L/min Sample Time: 2 minutes
2. Turn on pump and sample for two minutes.
3. Disconnect adapter and casse e. Before replacing sealing labels on casse e, hold it up to the light. If you cannot see through the slide, discard casse e and take another sample. Continue sampling until a usable sample is obtained.
4. Seal casse e inlet and outlet with sealing labels, label casse e, and ship with blanks to a laboratory for analysis. Go to www.skcinc.com for a list of
laboratories that analyze VersaTrap Casse es.
Wall Sampling Considerations
1. Sample location. Total counts may vary in diff erent types of wall cavities.
E.g., a four-foot wall versus a 12-foot plumbing wall will diff er in volume of air within the wall cavity.
2. Moisture level. Excessive moisture in a sample location can prevent
microbial contaminants from becoming airborne.
3. Gypsum dust. If gypsum dust is trapped inside the casse e, accurate
microscopic analysis cannot be performed because the dust will mask mold spores. Following the guidelines in this operating instruction should help minimize the collection of dust when using this device.
Ordering Information
Media and Accessories Cat. No. VersaTrap 37-mm Sampling Cassette
Shelf-life one year from date of manufacture.
VersaTrap Wall Adapters 225-9822/5 QuickTake 30 Sample Pump with Rotameter, 10 to 30 L/min,
includes charger (100-240 V)
† Contains Li-Ion ba eries and is subject to special shipping regulations.
Wall Inspector Kit includes 10 VersaTrap Cassettes, 10 VersaTrap Wall Adapters, Wall Adapter Rod, drill with bit, stud fi nder, moisture meter, and carry case 225-9690
225-9820/10 225-9821/50
228-9530A
5. Turn on pump for 10 seconds. Turn off pump. Remove "vacuum" casse e and set aside for re-use.