SKC 225-9594 User Manual

BioSampler
®
Operating Instructions
SKC Inc.
863 Valley View Road
Eighty Four, PA 15330 USA
Form 37084 Rev 1404
Description
Performance Profi le
Flow Rate: Approximately 12.5 L/min Medium: Water or more viscous liquids such as ViaTrap mineral oil
Collection Effi ciency: Nearly 100% over a
wide range of particle sizes; decreases to approximately 90% at
0.5 μm
Pump: High-volume vacuum
pump capable of achieving sonic fl ow
† May not be suitable for PCR analysis
®
Inlet (aerosol in)
To pump
Figure 1. BioSampler with 20-ml vessel
(5-ml vessel available)
* U.S. Patent Nos. 5,902,385 and 5,904,752
Sonic Nozzles (3)
1
Operation
Calibration
The BioSampler’s tangential nozzles act as sonic orifi ces. Therefore, if the pump maintains a pressure drop of 0.5 atm (15 in Hg) or more across the sampler at normal atmospheric conditions, the fl ow rate will always be the same (about
12.5 L/min) unless one or more of the nozzles become clogged. Figure 2 shows the recommended calibration set up. Calibration is particularly important when sampling in areas with non-standard temperature and pressure.
1. Connect a fl owmeter (bubble or dry-gas) to the inlet of the BioSampler.
2. Connect the outlet of the BioSampler (fi lled with collection liquid or coated) to the inlet of a liquid trap (Cat. No. 225-22) and the outlet of the trap to a vacuum pump with a vacuum pressure gauge or manometer in line.
3. Adjust the pump until sonic fl ow is achieved (vacuum gauge or manometer reads between 15 and 30).
4. Read and record fl ow rate, Q
Flow
Meter
L/min.
s
Pressure drop of
0.5 atm (15 in Hg)
Mano-
meter
Liquid
Trap
Figure 2. Calibration of BioSampler
Exhaust
Vacuum
pump
2
Loading...
+ 6 hidden pages