Operating Instructions
How to Use the SKC Coated Filters
863 Valley View Road, Eighty Four, PA 15330
Tel: 724-941-9701 Fax: 724-941-1369
www.skcinc.com
Coated Filters
Cat. No. 225-9011
Method & Chemical: MDHS 25/3 (U.K.) for Organic Isocyanates; PI, TDI,
HDI, MDI.
Filter & Coating: 25-mm glass fi ber fi lter coated with 1-(2-methoxyphenyl)
piperazine. Filters only, not loaded in casse es
Prior to Sampling: Store at ≤ 39.2 F (4 C) for up to 6 months.
Sample Stability: Store TDI at ≤ 39.2 F (4 C) for up to 90 days. No data is
available for PI, HDI, and MDI.
For information on other available
coated fi lters, request SKC Publication 1219.
1. Set up the sampling train for calibration with the representative fi lter
casse e in line. For “closed-face” sampling, remove the plugs and connect
the casse e to the sampling train. For “open-face” sampling, remove the outlet
plug and the casse e inlet section then connect the casse e to the sampling
train. For details on se ing up a sampling train, refer to the SKC Sample Setup
Guide “Sampling Train - Filters and Cyclones” (SKC Publication 1166).
2. Calibrate the pump to the desired fl ow rate using a fi lm fl owmeter or
calibrated precision rotameter. For details on calibrating a pump, refer to the
SKC Sample Setup Guide “Calibrating a Pump Using a Film Flowmeter” (SKC
Publication 1163).
3. Replace the casse e used for calibration with a fresh fi lter casse e for
sample collection. Attach the cassette to a worker’s collar, as close to the
breathing zone as possible, by using a fi lter casse e holder (SKC Cat. No.
225-1). Sample for the specifi ed time interval and record the time.
4. Remove the fi lter casse e at the end of the specifi ed sampling period and
replace both end plugs (and the inlet if necessary). Recheck the fl ow rate using
the same casse e and fl owmeter used for calibration (in Step 2) to ensure that
the fl ow rate has not changed by more than 5%.
5. Store sample at ≤ 39.2 F (4 C) for up to 90 days. Appropriately package
samples, blanks, and all pertinent data to send to a lab for analysis. Expedite
shipping of samples to laboratory. Friday shipments are not recommended.