Preloaded Coated Filters
Cat. No. 225-9002
SKC coated fi lters are shipped preloaded in the casse es with end plugs and shrink bands in place. No assembly is required; they are ready to use.
Method & Chemical: OSHA 42 (modifi ed*) for 1,6-hexamethylene diisocyanate (HDI), toluene-2,6 diisocyanate (2,6-TDI), and toluene 2,4-diisocyanate (2,4-TDI). OSHA 47 (modifi ed*) for methylene bisphenyl isocyanate (MDI). ASTM D5836 (modifi ed*) for 2,4-TDI and 2,6-TDI only
* OSHA Method 42 specifi es a fi lter loading of 0.1 mg of 1-(2-pyridyl) piperazine. OSHA 47 and
ASTM D5836 specify a fi lter loading of 1 mg of 1-(2-Pyridyl) piperazine; however, Coated Filter 225-9002 (loading of 2 mg to increase stability and shelf-life of fi lters) is acceptable for both sampling methods.
Filter & Coating: Glass fi ber fi lter and cellulose support pad coated with
2.0 mg of 1-(2-pyridyl) piperazine loaded into a three-piece 37-mm casse e. Use in the open-face sampling confi guration.
Prior to Sampling: Store at 39.2 F (4 C) for up to 6 months.
Sample Stability: Submit extracted samples to a laboratory as soon as possible
a er sampling. Store at 39.2 F (4 C) for up to one week if delay is unavoidable.
For information on other available
coated fi lters, request SKC Publication 1219.
How to Use the SKC Preloaded Coated Filters
Do not expose coated fi lters to high temperatures or prolonged periods in direct sunlight.
1. Select a coated fi lter casse e for calibrating pump fl ow rate. A red plug secures the casse e inlet, which is clearly marked “inlet,” and a blue plug secures the outlet. Set up the sampling train for calibration with the representative fi lter casse e in line. For “open-face” sampling, remove the outlet plug and the casse e inlet section then connect the casse e to the sampling train. For details on se ing up a sampling train, refer to the SKC Sample Setup Guide “Sampling Train - Filters and Cyclones” (SKC Publication 1166).
2. Calibrate the pump to the desired fl ow rate using a fi lm fl owmeter or calibrated precision rotameter. For details on calibrating a pump, refer to the SKC Sample Setup Guide “Calibrating a Pump Using a Film Flowmeter” (SKC Publication 1163).
3. Replace the casse e used for calibration with a fresh coated fi lter casse e for sample collection. Attach the cassette to a worker’s collar, as close to the breathing zone as possible, by using a filter cassette holder (SKC Cat. No. 225-1). Sample for the time specifi ed in the method and record the time.
4. Remove the fi lter casse e at the end of the specifi ed sampling period and replace the inlet and both plugs. Recheck the fl ow rate using the same casse e and fl owmeter used for calibration (in Step 2) to ensure that the fl ow rate has not changed by more than 5%.
5. Don disposable gloves. Remove casse e inlet section and use clean forceps to remove fi lter from casse e. Place fi lter in a glass vial containing fi eld extraction solution following the procedure in Appendix B of OSHA Directive CPL 03-00-017 (National Emphasis Program on Isocyanates) at www.osha.gov/OshDoc/Directive_ pdf/CPL_03-00-017.pdf .
6. Submit extracted samples to a laboratory as soon as possible a er sampling. Store at 39.2 F (4 C) for up to one week if delay is unavoidable. Appropriately package and ship samples, blanks, and all pertinent data to a laboratory for analysis. Refrigerated shipment is not required. Follow all applicable hazardous materials shipping restrictions and requirements. See Appendix B of OSHA Directive CPL 03-
00-017 for details.