SKC 225-9001 User Manual

Operating Instructions
863 Valley View Road, Eighty Four, PA 15330
Tel: 724-941-9701 Fax: 724-941-1369
www.skcinc.com
Preloaded Coated Filters
Cat. No. 225-9001*
SKC coated fi lters are shipped preloaded in the casse es with end plugs and shrink bands in place. No assembly is required; they are ready to use.
Method & Chemical: OSHA ID 105 for inorganic arsenic; ASTM D4765, NIOSH 7902, NIOSH 7906, and OSHA ID 110 for fl uorides. IRSST* Canadian
Patent # 1,299,144
Prior to Sampling: Store at 71.6 F (22 C) for up to two years.
Sample Stability: No storage stability data is available.
How to Use the SKC Preloaded Coated Filters
1. Select one coated fi lter casse e for calibrating the fl ow rate. A red plug secures the inlet, which is clearly marked “inlet,” and a blue plug secures the outlet. Set up the sampling train for calibration with the representative fi lter casse e in line. For “closed-face” sampling, remove the plugs and connect the casse e to the sampling train. For “open-face” sampling, remove the outlet plug and the casse e inlet section then connect the casse e to the sampling train. For details on se ing up a sampling train, refer to the SKC Sample Setup Guide “Sampling Train - Filters and Cyclones” (SKC Publication 1166).
2. Calibrate the pump to the desired fl ow rate using a fi lm fl owmeter or calibrated precision rotameter. For details on calibrating a pump, refer to the SKC Sample Setup Guide “Calibrating a Pump Using a Film Flowmeter” (SKC Publication 1163).
3. Replace the casse e used for calibration with a fresh coated fi lter casse e for sample collection. A ach the casse e to a worker’s collar, as close to the breathing zone as possible, by using a fi lter casse e holder (SKC Cat. No. 225-1). Sample for the specifi ed time interval and record the time.
4. Remove the fi lter casse e at the end of the specifi ed sampling period and replace both end plugs (and the inlet if necessary). Recheck the fl ow rate using the same casse e and fl owmeter used for calibration (in Step 2) to ensure that the fl ow rate has not changed by more than 5%.
5. Appropriately package fi lter casse es and ship the samples, blanks, and all pertinent data to a laboratory for analysis.
For information on other available
coated fi lters, request SKC Publication 1219.
* Concept of sampling fl uorides in air using double fi lter patented in Canada by IRSST,
Institut de recherche Robert-Sauvé en santé et en sécurité du travail du Quebec (Quebec Occupational Health and Safety Research Institute), 21 APR 92.