The patented* SKC Button Aerosol Sampler is a
reusable fi lter sampler with a porous curved-surface
sampling inlet designed to improve the collection
characteristics of inhalable dust (< 100 μm aerodynamic
diameter
viable analysis.
While closely following the ACGIH/ISO sampling
criteria for inhalable particulate mass at 4 L/min, the
Bu on Sampler's design also minimizes wind sensitivity, provides equal distribution
of particle loading and low intersample variation, and can be used in any orientation.
Sample analysis for inhalable dust is performed gravimetrically using NIOSH
Method 0500. Metals can be analyzed using direct-reading techniques (e.g.,
NIOSH Method 7702). Fungal spore analysis includes microscopy, immunoassay,
or polymerase chain reaction (PCR).
† Very large particles (> 100 μm) are thought to be less inhalable, but may enter the inlets of
†
) including bioaerosols for viable or non-
SKC Button Aerosol
some samplers by projection or sedimentation. The Bu on Aerosol Sampler design minimizes
collection of these particles.
Sampler
Performance Profi le
Flow Rate: 4 L/min
The Bu on Sampler closely follows the ACGIH/ISO inhalability
Construction: Sampling inlet: conductive stainless steel
Body: aluminum
Support screen: stainless steel
Clip: stainless steel and nylon
O-rings: PTFE (inlet) and BUNA-N (body)
Filters: 25 mm select as specifi ed by the method
Analysis: Inhalable dust: Gravimetric (GR)
Fungal spores: Epifl uorescence microscopy, immunoassay, or
Metals: X-ray Fluorescence (XRF)
Autoclavable: Up to approximately 273.2 F (134 C)
Tubing: 1/4-inch ID
* U.S. Patent Nos. 5,954,845 and 5,958,111
curve at 4 L/min. This provides optimum sampling. Sampling
effi ciency is maintained within ± 30% at fl ows ranging from
2 to 5 L/min.
polymerase chain reaction (PCR)
Page 2
Operation
Wear gloves and use forceps when handling fi lters.
1. For gravimetric sampling, condition and weigh a 25-mm fi lter according to
method used.
2. Unscrew the inlet of the sampler counterclockwise until it comes off and
remove the PTFE O-ring.
3. Place the 25-mm fi lter on top of the support screen inside the sampler.
Note: Representative fi lters used for calibration do not require conditioning and
weighing.
4. Place the PTFE O-ring on top of the fi lter.
5. Replace the inlet and very gently turn clockwise until moderately tight.
Caution: Excessive tightening of the inlet section can twist and tear the fi lter.
Collar Clip
Body
Cap
Inlet
Section
PTFE
O-ring
Filter
Support
Screen
BUNA-N
O-ring
Outlet Plug
Calibration
1. Ensure pump had run for 5 minutes before calibrating.
2. Using fl exible tubing, connect the outlet of the
Bu on Sampler to the inlet of a sampling pump.
3. Place the calibration adapter outlet onto the Bu on
Sampler inlet, and connect the calibration adapter
inlet to a primary standard calibrator.
4. Calibrate to 4 L/min following the instructions in the
pump and calibrator operating instructions.
5. Disconnect the calibrator and calibration adapter.
Replace the fi lter used to set the fl ow with a fresh
Button Sampler
with Calibration
Adapter
conditioned and weighed fi lter for sample collection.
Note: If using the Bu on Sampler with protective shield for abrasive blasting
sampling, calibrate the Bu on Sampler without shield using the Calibration
Adapter. Following calibration, push shield onto sampler inlet and sample.
Sampling
1. Ensure the fl ow rate has been calibrated (see Calibration).
2. Using fl exible tubing, connect the outlet of the Bu on Sampler to the inlet of
a sampling pump such as an AirChek
3. For personal sampling, clip the Bu on Sampler onto a worker’s collar or
pocket near the breathing zone. Clip the pump onto the worker’s belt or
place it in a protective pouch. Start the pump and record the start time,
worker location, and fl ow rate.
®
XR5000 Sampler.
Page 3
4. For area sampling, position the Bu on Sampler to avoid cross-dra s and
direct projection of the particles into the inlet.
5. At the end of the sampling period, stop the pump and record the stop time.
Sample Shipment
1. When sampling is complete, very gently unscrew the sampler inlet and use
forceps to remove the PTFE O-ring.
2. Using forceps, carefully remove the fi lter from the sampler.
3. Package for shipment as directed by the method or laboratory.
4. Ship samples with pertinent sampling information to an accredited laboratory
for analysis.
Analysis
Dependent upon the hazard of interest, samples taken with the Bu on Sampler
can be analyzed using gravimetric analysis, analytical techniques such as x-ray
fl uorescence, atomic absorption (AA), or inductively coupled plasma (ICP).
Growth culture, microscopy, and other assays may be used for the determination
of fungal spores, endotoxin, and bacteria.
If gravimetric analysis is used, weigh the fi lter and subtract the preweight from the
postweight. The net weight will be the dust collected.
Using the Button Sampler as an Abrasive Blasting
Sampler for Heavy Metals
SKC has designed a protective shield for the Bu on Sampler for use when sampling
in abrasive blasting environments. The shield protects the fi lter from shredding
or being overloaded by large particles thrust onto the sampler during
blasting. Tested by the U.S. Military during recent abrasive blasting operations, the
Bu on Sampler with protective shield (Abrasive Blasting Sampler for Heavy
Metals) withstood very high particle concentrations and provided recoverable
samples and meaningful exposure information. The shield is very easy to use,
simply push it onto the sampler inlet and sample (see Sampling).
‡ The Bu on Sampler with shield is not suitable for sampling silica. While silica is commonly found in
abrasive blasting environments, sampling for silica using OSHA Method ID 142 requires a cyclone.
‡
Using the Button Sampler to collect Inhalable
Bioaerosols
The patented SKC Bu on Aerosol Sampler provides superior collection of
inhalable particles including bacteria and fungal spores. Use the Bu on Sampler
with a 25-mm membrane fi lter such as MCE or PVC to collect bioaerosols for
viable or non-viable analysis. Using SKC gelatin fi lters with the Bu on Sampler
increases the survival of stress-sensitive microorganisms during sampling. The
Bu on Sampler is used with a sample pump capable of 4 L/min for personal
inhalable particulate sampling following the ACGIH/ISO criteria.
Note: Gelatin fi lters dissolve when placed on agar.
Page 4
Cleaning
Using mild soapy water, clean all parts of the Bu on Sampler before further sam-
pling. The Bu on Sampler can be brushed and/or wiped with a clean dust-free
tissue, air dried, blown dry, or wiped with isopropyl alcohol. Take particular care
in cleaning the PTFE O-ring.
The Bu on Sampler may also be autoclaved up to approximately 273.2 F (134 C).
Prepare the Bu on Sampler before autoclaving by removing the collar clip from
the sampler using a Phillips head screwdriver. Also remove the BUNA-N O-ring
from the sampler body (see page 2 for exploded view of sampler).
References
Kalatoor, S., Grinshpun, S., Willeke, K., “New Aerosol Sampler with Low Wind Sensitivity and Good Filter
Collection Uniformity,” Atmospheric Environment, Vol. 29, No. 10 (1995), pp. 1105-1112
Hauck, B., Grinshpun, S., Reponen, A., Reponen, T., Willeke, K., Bornschein, R., “Field Testing of New
Aerosol Sampling Method With a Porous Curved Surface as Inlet,” American Industrial Hygiene Association Journal, Vol. 58, No. 10 (1997), pp. 713-719
Aizenberg, V., England, A., Grinshpun, S., Willeke, K., Carlton, G., “Metal Exposures Among Abrasive
Blasting Workers at Four U.S. Air Force Facilities,” Applied Occupational and Environmental Hygiene,
Vol. 15, No. 10 (2000), pp. 766-772
Aizenberg, V., Grinshpun, S., Willeke, K., Smith, J., Baron, P.A., “Performance Characteristics of the Bu on
Personal Inhalable Aerosol Sampler,” AIHA Journal, Vol. 61, May/June 2000, pp. 398-404
Optional Accessories
DescriptionCat. No.
Button Sampler, requires a 25-mm fi lter225-360
Button Sampler Pump Kit includes Button Sampler, standard
XR5000 Sample Pump, single charger with cable, 3 feet of
Tygon tubing, and calibration adapter, requires a 25-mm fi lter 100-240 V
Abrasive Blasting Kit includes Button Sampler and protective
shield, requires a 25-mm fi lter225-367
Button Sampler Calibration Adapter225-361
Protective Shield, for abrasive blasting environments225-366
Filter Transport Case, for 25-mm fi lters, conductive plastic225-67
Replacement O-ring, PTFEP32265
210-4121
Filters (25 mm)
Glass fi ber, 1.0 μm, pk/500225-702
PTFE with PMP support (Tefl o
Polyvinyl chloride (PVC), 5.0 μm, pk/100225-8-04
Mixed cellulose ester (MCE), 1.2 μm, pk/100225-1912
Gelatin,
∆ A pore size of 1.0 μm or larger is recommended for use with the Bu on Sampler due to back pressure
limitations of personal samplers.
∞
Back pressure on PTFE fi lters can vary within the same lot.
# Gelatin fi lters dissolve when placed on agar.
** Maximum operating temperature for the Cat. No. 225-1711 PTFE fi lter with PMP support is
464 F (240 C) based on the PMP support ring.
∆
®
∞**
),
3.0 μm, pk/50225-1711
#
sterilized, pk/50225-9551
Cat. No.
SKC Limited Warranty and Return Policy
SKC products are subject to the SKC Limited Warranty and Return Policy, which
provides SKC’s sole liability and the buyer’s exclusive remedy. To view the complete SKC Limited Warranty and Return Policy, go to h p://www.skcinc.com/
warranty.asp.
Form 3780 Rev 1303
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.