SKC 225-367 User Manual

Page 1
Operating Instructions
Tel: 724-941-9701 Fax: 724-941-1369 e-mail: skctech@skcinc.com
863 Valley View Road, Eighty Four, PA 15330 USA
Button Aerosol Sampler
Catalog No. 225-360
The patented* SKC Button Aerosol Sampler is a reusable fi lter sampler with a porous curved-surface sampling inlet designed to improve the collection characteristics of inhalable dust (< 100 μm aerodynamic diameter viable analysis.
While closely following the ACGIH/ISO sampling criteria for inhalable particulate mass at 4 L/min, the Bu on Sampler's design also minimizes wind sensitivity, provides equal distribution of particle loading and low intersample variation, and can be used in any orientation. Sample analysis for inhalable dust is performed gravimetrically using NIOSH Method 0500. Metals can be analyzed using direct-reading techniques (e.g., NIOSH Method 7702). Fungal spore analysis includes microscopy, immunoassay, or polymerase chain reaction (PCR).
Very large particles (> 100 μm) are thought to be less inhalable, but may enter the inlets of
SKC Button Aerosol
some samplers by projection or sedimentation. The Bu on Aerosol Sampler design minimizes collection of these particles.
Sampler
Performance Profi le
Flow Rate: 4 L/min The Bu on Sampler closely follows the ACGIH/ISO inhalability
Construction: Sampling inlet: conductive stainless steel Body: aluminum Support screen: stainless steel Clip: stainless steel and nylon O-rings: PTFE (inlet) and BUNA-N (body)
Filters: 25 mm select as speci ed by the method
Analysis: Inhalable dust: Gravimetric (GR)
Fungal spores: Epifl uorescence microscopy, immunoassay, or
Metals: X-ray Fluorescence (XRF)
Autoclavable: Up to approximately 273.2 F (134 C)
Tubing: 1/4-inch ID
* U.S. Patent Nos. 5,954,845 and 5,958,111
curve at 4 L/min. This provides optimum sampling. Sampling effi ciency is maintained within ± 30% at fl ows ranging from 2 to 5 L/min.
polymerase chain reaction (PCR)
Page 2
Operation
Wear gloves and use forceps when handling fi lters.
1. For gravimetric sampling, condition and weigh a 25-mm fi lter according to method used.
2. Unscrew the inlet of the sampler counterclockwise until it comes off and remove the PTFE O-ring.
3. Place the 25-mm fi lter on top of the support screen inside the sampler.
Note: Representative fi lters used for calibration do not require conditioning and weighing.
4. Place the PTFE O-ring on top of the fi lter.
5. Replace the inlet and very gently turn clockwise until moderately tight.
Caution: Excessive tightening of the inlet section can twist and tear the lter.
Collar Clip
Body
Cap
Inlet Section
PTFE O-ring
Filter
Support Screen
BUNA-N O-ring
Outlet Plug
Calibration
1. Ensure pump had run for 5 minutes before calibrating.
2. Using fl exible tubing, connect the outlet of the
Bu on Sampler to the inlet of a sampling pump.
3. Place the calibration adapter outlet onto the Bu on Sampler inlet, and connect the calibration adapter inlet to a primary standard calibrator.
4. Calibrate to 4 L/min following the instructions in the pump and calibrator operating instructions.
5. Disconnect the calibrator and calibration adapter. Replace the fi lter used to set the fl ow with a fresh
Button Sampler with Calibration Adapter
conditioned and weighed fi lter for sample collection.
Note: If using the Bu on Sampler with protective shield for abrasive blasting sampling, calibrate the Bu on Sampler without shield using the Calibration Adapter. Following calibration, push shield onto sampler inlet and sample.
Sampling
1. Ensure the fl ow rate has been calibrated (see Calibration).
2. Using fl exible tubing, connect the outlet of the Bu on Sampler to the inlet of a sampling pump such as an AirChek
3. For personal sampling, clip the Bu on Sampler onto a worker’s collar or pocket near the breathing zone. Clip the pump onto the worker’s belt or place it in a protective pouch. Start the pump and record the start time, worker location, and fl ow rate.
®
XR5000 Sampler.
Page 3
4. For area sampling, position the Bu on Sampler to avoid cross-dra s and direct projection of the particles into the inlet.
5. At the end of the sampling period, stop the pump and record the stop time.
Sample Shipment
1. When sampling is complete, very gently unscrew the sampler inlet and use forceps to remove the PTFE O-ring.
2. Using forceps, carefully remove the fi lter from the sampler.
3. Package for shipment as directed by the method or laboratory.
4. Ship samples with pertinent sampling information to an accredited laboratory for analysis.
Analysis
Dependent upon the hazard of interest, samples taken with the Bu on Sampler can be analyzed using gravimetric analysis, analytical techniques such as x-ray uorescence, atomic absorption (AA), or inductively coupled plasma (ICP). Growth culture, microscopy, and other assays may be used for the determination of fungal spores, endotoxin, and bacteria.
If gravimetric analysis is used, weigh the fi lter and subtract the preweight from the postweight. The net weight will be the dust collected.
Using the Button Sampler as an Abrasive Blasting Sampler for Heavy Metals
SKC has designed a protective shield for the Bu on Sampler for use when sampling in abrasive blasting environments. The shield protects the fi lter from shredding or being overloaded by large particles thrust onto the sampler during blasting. Tested by the U.S. Military during recent abrasive blasting operations, the Bu on Sampler with protective shield (Abrasive Blasting Sampler for Heavy Metals) withstood very high particle concentrations and provided recoverable samples and meaningful exposure information. The shield is very easy to use, simply push it onto the sampler inlet and sample (see Sampling).
‡ The Bu on Sampler with shield is not suitable for sampling silica. While silica is commonly found in
abrasive blasting environments, sampling for silica using OSHA Method ID 142 requires a cyclone.
Using the Button Sampler to collect Inhalable Bioaerosols
The patented SKC Bu on Aerosol Sampler provides superior collection of inhalable particles including bacteria and fungal spores. Use the Bu on Sampler with a 25-mm membrane fi lter such as MCE or PVC to collect bioaerosols for viable or non-viable analysis. Using SKC gelatin fi lters with the Bu on Sampler increases the survival of stress-sensitive microorganisms during sampling. The Bu on Sampler is used with a sample pump capable of 4 L/min for personal inhalable particulate sampling following the ACGIH/ISO criteria.
Note: Gelatin lters dissolve when placed on agar.
Page 4
Cleaning
Using mild soapy water, clean all parts of the Bu on Sampler before further sam- pling. The Bu on Sampler can be brushed and/or wiped with a clean dust-free tissue, air dried, blown dry, or wiped with isopropyl alcohol. Take particular care in cleaning the PTFE O-ring.
The Bu on Sampler may also be autoclaved up to approximately 273.2 F (134 C). Prepare the Bu on Sampler before autoclaving by removing the collar clip from the sampler using a Phillips head screwdriver. Also remove the BUNA-N O-ring from the sampler body (see page 2 for exploded view of sampler).
References
Kalatoor, S., Grinshpun, S., Willeke, K., “New Aerosol Sampler with Low Wind Sensitivity and Good Filter Collection Uniformity,” Atmospheric Environment, Vol. 29, No. 10 (1995), pp. 1105-1112
Hauck, B., Grinshpun, S., Reponen, A., Reponen, T., Willeke, K., Bornschein, R., “Field Testing of New Aerosol Sampling Method With a Porous Curved Surface as Inlet,” American Industrial Hygiene Association Journal, Vol. 58, No. 10 (1997), pp. 713-719
Aizenberg, V., England, A., Grinshpun, S., Willeke, K., Carlton, G., “Metal Exposures Among Abrasive Blasting Workers at Four U.S. Air Force Facilities,” Applied Occupational and Environmental Hygiene, Vol. 15, No. 10 (2000), pp. 766-772
Aizenberg, V., Grinshpun, S., Willeke, K., Smith, J., Baron, P.A., “Performance Characteristics of the Bu on Personal Inhalable Aerosol Sampler,” AIHA Journal, Vol. 61, May/June 2000, pp. 398-404
Optional Accessories
Description Cat. No. Button Sampler, requires a 25-mm lter 225-360 Button Sampler Pump Kit includes Button Sampler, standard
XR5000 Sample Pump, single charger with cable, 3 feet of Tygon tubing, and calibration adapter, requires a 25-mm lter 100-240 V Abrasive Blasting Kit includes Button Sampler and protective shield, requires a 25-mm lter 225-367
Button Sampler Calibration Adapter 225-361 Protective Shield, for abrasive blasting environments 225-366 Filter Transport Case, for 25-mm lters, conductive plastic 225-67 Replacement O-ring, PTFE P32265
210-4121
Filters (25 mm) Glass fi ber, 1.0 μm, pk/500 225-702 PTFE with PMP support (Tefl o Polyvinyl chloride (PVC), 5.0 μm, pk/100 225-8-04 Mixed cellulose ester (MCE), 1.2 μm, pk/100 225-1912 Gelatin,
A pore size of 1.0 μm or larger is recommended for use with the Bu on Sampler due to back pressure
limitations of personal samplers.
Back pressure on PTFE fi lters can vary within the same lot. # Gelatin fi lters dissolve when placed on agar. ** Maximum operating temperature for the Cat. No. 225-1711 PTFE fi lter with PMP support is
464 F (240 C) based on the PMP support ring.
®
**
),
3.0 μm, pk/50 225-1711
#
sterilized, pk/50 225-9551
Cat. No.
SKC Limited Warranty and Return Policy
SKC products are subject to the SKC Limited Warranty and Return Policy, which provides SKC’s sole liability and the buyer’s exclusive remedy. To view the com­plete SKC Limited Warranty and Return Policy, go to h p://www.skcinc.com/ warranty.asp.
Form 3780 Rev 1303
Loading...