Operating Instructions
Tel: 724-941-9701 Fax: 724-941-1369 email: skctech@skcinc.com
863 Valley View Road, Eighty Four, PA 15330 USA
DPM Cassette with Impactor
Catalog No. 225-317
SKC Diesel Particulate Matter Cassettes are
designed for sampling in atmospheres where
it is necessary to separate DPM from other
respirable dust (such as coal dust). Single-use
SKC DPM Cassettes are tamper-evident sealed
to ensure sample integrity. Each streamlined
plastic cassette comprises an impactor
with precision sapphire orifice, impaction
substrate, and two heat-treated quartz filters.
The impactor screens out respirable particles
≥ 1.0 μm. Particles less than 1.0 μm are collected
on the filter. Samples are analyzed for organic
and elemental carbon content using a highly
sensitive Evolved Gas Analysis (EGA) technique
with thermal-optical analyzer as specified in
NIOSH 5040.
www.skcinc.com
Plug
Outlet
Support
pad
Heat-treated
quartz fi lter
Heat-treated
quartz fi lter
Stainless steel ring
(defi nes fi lter area)
Impactor
assembly
Cassette
body
Inlet
Plug
Performance Profi le
Casse e Size: 37 mm
Filter: Two heat-treated quartz fi ber
Figure 1. Exploded view of
DPM Cassette
(Cat. No. 225-317)
with cellulose support pad
Filter Collection
Area: 8.04 cm
2
Impactor: Four-nozzle, precision sapphire orifi ces
Flow Rate: Calibrated at 1.7 and 2.0 L/min for sampling of sub-micron
particles
Analysis: For organic and elemental carbon using evolved gas analysis
(EGA) with thermal-optical analyzer (NIOSH Method 5040)
Sampling Equipment
The SKC DPM Casse e is designed for use with an appropriate sample pump and cyclone
that pre-selects non-respirable particles*. An SKC Universal XR, AirChek® 2000, or AirChek
XR5000 sample pump can provide the specifi ed fl ow rate. Use the SKC Casse e/Cyclone
Holder (Cat. No. 225-1) to secure the casse e when used with the GS-1 Cyclone.
* A cyclone is not necessary in all sampling situations. It should be used in se ings where larger particulate is likely to clog
the impactor.

Setting Up the Sampling Train with Cyclone and Holder
1. Insert stem of GS-1 Cyclone into large hole in casse e holder.
2. Orient cyclone in holder until the small round nodule on the bowl adapter
rim fi ts into the notch in the casse e holder.
3. Remove inlet and outlet plugs from a DPM Casse e, pull back spring-loaded
hold-down plate on casse e holder, and gently push inlet of casse e onto
top of cyclone.
4. Position small hole in hold-down plate over casse e
outlet.
5. Insert Luer adapter on the holder's rubber tubing into
casse e outlet.
AIRCHEK
Figure 2. DPM Sampling Train with GS-1 Cyclone,
Cassette Holder, and Universal PCXR4 Pump
SAMPLER
SAMPLE PERIOD — MINUTES
START
HOLD
U
LISTED 124U
5
L
®
INTRINSICALLY SAFE
PORTABLE AIR SAMPLING PUMP
FOR USE IN HAZARDOUS LOCA
TIONS CLASS I, GROUPS A B C D
4
AND CLASS II, GROUPS E F G
FLOW
AND CLASS III, TEMPERATURE
AND
CODE T3C.
BATTERY
CHECK
ON
FLOW
ADJ
AIRCHEK SAMPLER
MODEL 224-PCXR8
WARNING - SUBSTITUTION OF
COMPONENTS MAY IMPAIR INTRIN
SIC SAFETY. USE ONLY UL LISTED
PORTABLE AIR SAMPLING PUMP
BATTERY PACK MODEL P21661
SERIAL NO.
SKC INC.
EIGHTY FOUR PA 15330
3
2
1
Setting Up the Sampling Train without a Cyclone
1. Remove the plug from the outlet of the casse e (Figure 1).
2. A ach one end of a length of Tygon® tubing to the casse e outlet and the
other end to the inlet of an air sample pump capable of maintaining the
desired fl ow rate.
3. Secure the casse e on a worker by using a cable tie and collar clip.
4. Remove plug from the casse e inlet immediately before sampling.