SKC 225-317 User Manual

Page 1
Operating Instructions
Tel: 724-941-9701 Fax: 724-941-1369 email: skctech@skcinc.com
863 Valley View Road, Eighty Four, PA 15330 USA
DPM Cassette with Impactor
Catalog No. 225-317
SKC Diesel Particulate Matter Cassettes are designed for sampling in atmospheres where it is necessary to separate DPM from other respirable dust (such as coal dust). Single-use SKC DPM Cassettes are tamper-evident sealed to ensure sample integrity. Each streamlined plastic cassette comprises an impactor with precision sapphire orifice, impaction substrate, and two heat-treated quartz filters. The impactor screens out respirable particles 1.0 μm. Particles less than 1.0 μm are collected on the filter. Samples are analyzed for organic and elemental carbon content using a highly sensitive Evolved Gas Analysis (EGA) technique with thermal-optical analyzer as specified in NIOSH 5040.
www.skcinc.com
Plug
Outlet
Support pad
Heat-treated quartz fi lter
Heat-treated quartz fi lter
Stainless steel ring (defi nes fi lter area)
Impactor assembly
Cassette body
Inlet
Plug
Performance Profi le
Casse e Size: 37 mm Filter: Two heat-treated quartz fi ber
Figure 1. Exploded view of
DPM Cassette
(Cat. No. 225-317)
with cellulose support pad Filter Collection Area: 8.04 cm
2
Impactor: Four-nozzle, precision sapphire orifi ces Flow Rate: Calibrated at 1.7 and 2.0 L/min for sampling of sub-micron
particles
Analysis: For organic and elemental carbon using evolved gas analysis
(EGA) with thermal-optical analyzer (NIOSH Method 5040)
Sampling Equipment
The SKC DPM Casse e is designed for use with an appropriate sample pump and cyclone that pre-selects non-respirable particles*. An SKC Universal XR, AirChek® 2000, or AirChek XR5000 sample pump can provide the specifi ed fl ow rate. Use the SKC Casse e/Cyclone Holder (Cat. No. 225-1) to secure the casse e when used with the GS-1 Cyclone.
* A cyclone is not necessary in all sampling situations. It should be used in se ings where larger particulate is likely to clog
the impactor.
Page 2
Setting Up the Sampling Train with Cyclone and Holder
1. Insert stem of GS-1 Cyclone into large hole in casse e holder.
2. Orient cyclone in holder until the small round nodule on the bowl adapter rim fi ts into the notch in the casse e holder.
3. Remove inlet and outlet plugs from a DPM Casse e, pull back spring-loaded hold-down plate on casse e holder, and gently push inlet of casse e onto top of cyclone.
4. Position small hole in hold-down plate over casse e outlet.
5. Insert Luer adapter on the holder's rubber tubing into casse e outlet.
AIRCHEK
Figure 2. DPM Sampling Train with GS-1 Cyclone,
Cassette Holder, and Universal PCXR4 Pump
SAMPLER
SAMPLE PERIOD — MINUTES
START
HOLD
U
LISTED 124U
5
L
®
INTRINSICALLY SAFE PORTABLE AIR SAMPLING PUMP FOR USE IN HAZARDOUS LOCA TIONS CLASS I, GROUPS A B C D
4
AND CLASS II, GROUPS E F G
FLOW
AND CLASS III, TEMPERATURE
AND
CODE T3C.
BATTERY CHECK
ON
FLOW
ADJ
AIRCHEK SAMPLER MODEL 224-PCXR8
WARNING - SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRIN SIC SAFETY. USE ONLY UL LISTED PORTABLE AIR SAMPLING PUMP BATTERY PACK MODEL P21661
SERIAL NO. SKC INC. EIGHTY FOUR PA 15330
3
2
1
Setting Up the Sampling Train without a Cyclone
1. Remove the plug from the outlet of the casse e (Figure 1).
2. A ach one end of a length of Tygon® tubing to the casse e outlet and the other end to the inlet of an air sample pump capable of maintaining the desired fl ow rate.
3. Secure the casse e on a worker by using a cable tie and collar clip.
4. Remove plug from the casse e inlet immediately before sampling.
Page 3
Calibration
Option 1: Using a Calibration Jar (Figure 3)
1. Prepare a DPM Casse e or cyclone/DPM Casse e assembly. If using a cyclone, ensure grit pot remains on cyclone body during calibration.
SKC recommends using the smallest calibration jar possible. To achieve this, do NOT use Cassette Holder 225-1 during calibration.
2. Place DPM Casse e or cyclone/DPM Casse e assembly into an airtight calibration jar that contains an inlet and outlet (SKC Cat. No. 225­111, see Figure 3).
3. Using fl exible tubing, connect inlet of calibrator to inlet of calibration jar.
4. Run tubing from outlet of DPM Casse e or cyclone/DPM Casse e assembly through outlet of calibration jar and to the inlet of a sample pump. If using the SKC 225-111 jar, connect the Luer adapter inside the jar to the cyclone/ casse e outlet. Connect jar outlet to inlet of a sample pump.
5. Turn on pump and calibrate to desired fl ow rate following directions in the pump and calibrator operating instructions.
Figure 3. Calibration Train with Jar
6. A er calibration, disassemble calibration jar, remove the DPM Casse e or cyclone/DPM Casse e assembly, and replace DPM Casse e used for calibration with a fresh (unused) casse e to be used for sampling.
Option 2: Jarless Calibration (Figure 4)
This calibration option is recommended when using a Defender primary standard calibrator to calibrate personal air sample pumps for size­selective particulate sampling using particulate samplers that do not have their own calibration adapters.
Figure 4. Jarless Calibration with DPM Cassette and
AirChek XR5000 Pump
1. Use a length of fl exible tubing to a ach the inlet of the sample pump to the suction port of a Defender calibrator.
2. Use the shortest length of tubing possible to a ach the outlet of the DPM Casse e or cyclone/DPM Casse e assembly to the pressure port of the Defender.
3. Set the Defender to take at least 20 fl ow measurements in order to average out the fl ow variations caused by direct connection of the pump to the calibrator. See calibrator operating instructions for details.
Page 4
Sampling
1. Ensure pump is calibrated to the appropriate fl ow rate (see Calibration).
2. Turn on pump and sample for the appropriate time.
3. When sampling is completed, turn off the pump, remove the DPM Casse e or cyclone/DPM Casse e assembly from the holder, remove the casse e from the cyclone, and immediately plug both the inlet and outlet of the casse e with the provided plugs.
Shipping
Package the casse es for shipment to a laboratory for analysis.
Analysis
Total carbon represents more than 80% of DPM particulate emissions. NIOSH Method 5040 can be used to quantitatively determine the total carbon (organic and elemental) content of the sample.
Ordering Information
Description Cat. No.
SKC DPM Cassette with impactor, contains two 37-mm
heat-treated quartz fi lters, tamper-evident sealed, single
use,
meets NIOSH 5040 specifi cations, average sampler
deposition area is 8.04 cm
Limited shelf-life 225-317 SKC GS-1 Cyclone,* single inlet, 10 mm, conductive plastic 225-105 Filter Cassette/Cyclone Holder, for attaching a DPM or
standard fi lter cassette to a worker’s clothing in the breathing zone. May be used with or without a cyclone, required when
using DPM Cassette with GS-1 Cyclone 225-1 Tubing, Collar Clip, and Cable Tie includes 3 feet Tygon tubing (1/4-inch ID) and alligator clip attached to a nylon cable tie 225-13-8 Multi-purpose Calibration Jar 225-111
2
, pk/10
* A cyclone should be used in se ings where larger particulate is likely to clog the impactor. † Requires 1/4-inch tubing for fi lter casse e/cyclone holder
SKC also off ers DPM casse es without impactors. For more information,
see www.skcinc.com/prod/225-317.asp.
SKC Limited Warranty and Return Policy
SKC products are subject to the SKC Limited Warranty and Return Policy, which provides SKC’s sole liability and the buyer ’s exclusive remedy. To view the complete SKC Limited Warranty and Return Policy, go to h p://www.skcinc.com/warranty. asp.
Form 40081 Rev 1401