Operating Instructions
SKC All-in-One Low Flow Adapter/Single Tube Holder
Cat. No. 224-27
Introduction
The SKC All-in-One replaces the SKC Single Adjustable Low Flow Tube Holder and CPC for single-tube sampling with AirChek Series and AirLite sample pumps. All required single-tube low fl ow sampling accessories are included in one compact holder. The built-in CPC maintains a constant pressure across the holder’s needle valve for fl ow stability.
Operation
See Notes before calibrating and sampling.
1. Set pump ow rate to 1.5 L/min. See pump operating instructions.
Tube holder with rubber sleeve
Threaded fi tting (fi ts all SKC protective tube covers)
Flow adjust screw (needle valve)
Built-in CPC (for fl ow stability)
Outlet/Tygon tubing to pump
Form 38035 Rev 1803
Larger ID sleeve for larger OD tubes
Pump inlet
2. Attach Tygon tubing to pump inlet.
3. Break tips off the sorbent tube to be used for calibration and insert into rubber sleeve on holder (arrow on tube pointing toward pump).
4. Use a calibrator to calibrate (set) fl ow rate through the sorbent tube. Use a small fl athead screwdriver to turn fl ow adjust screw on holder (counterclockwise to increase fl ow, clockwise to decrease fl ow). See pump operating instructions
for details.
5. Remove tube used for calibration and repeat Step 3 using a clean, labeled tube for sampling.
6. Thread protective tube cover onto holder’s threaded fi tting.
7. Use cover’s collar clip to attach holder on clothing in the breathing zone.
8. Sample for the desired time.
9. When sampling is complete, remove sample sorbent tube and cap. Insert sorbent tube used for calibration into sleeve on holder. Use calibrator to verify fl ow rate has remained within ± 5% of the set fl ow.
All-in-One Low Flow Adapter/Single Tube Holder with built-in
CPC connected to AirChek TOUCH sample pump
Notes
Do not shut off fl ow completely with fl ow adjust screw or use an
oversized screwdriver to adjust fl ow; valve or thread seat damage may result.
Ensure sorbent tube fits snugly in rubber sleeve prior to
sampling to avoid any air leakage. Two sleeves, each a dif ferent inner diameter (ID), are supplied.
Calibrate before each sampling operation using the tube holder
and pump to be used for sampling.
If the pump fl ow faults shortly after the holder tubing is attached
in Step 2, check that the needle valve is open by using a small athead screwdriver to turn the fl ow adjust screw on the holder slightly counterclockwise. If fl ow fault continues, check that the two small holes on the bottom of the built-in CPC are not blocked. See related bullet below.
Two small holes on the bottom of the built-in CPC can become
blocked. Periodically inspect. Clean, if needed, with a small pick and blow particles away with a puff of air.
For more information, see SKC Sample Setup Guides at www.skcinc.com or email skctech@skcinc.com.