Simpson 896 Operator's Manual

Page 1
Model 896 Multi/Level Calibrator
OPERAT OR’S MANUAL
Page 2
About this Manual
To the best of our knowledge and at the time written, the information con­tained in this document is technically correct and the procedures accurate and adequate to operate this instrument in compliance with its original ad­vertised specifications.
Notes and Safety Information
This Operator’s Manual contains warning symbols which alert the user to check for hazardous conditions. These appear throughout this manual where applicable, and are defined below. To ensure the safety of operating perfor­mance of this instrument, these instructions must be adhered to.
Warning, refer to accompanying documents.
!
Caution, risk of electric shock.
!
This instrument is designed to prevent accidental shock to the operator when properly used. However, no engineering design can render safe an instru­ment which is used carelessly. Therefore, this manual must be read carefully and completely before making any measurements. Failure to follow direc­tions can result in a serious or fatal accident.
Technical Assistance
SIMPSON ELECTRIC COMPANY offers assistance Monday through Friday 7:30 am to 5:00 pm Central Time. To receive assistance contact Technical Support or Customer Service at (847) 697-2260. Internet: http://www.simpsonelectric.com
Warranty and Returns
SIMPSON ELECTRIC COMP ANY w arrants each instrument and other articles manufactured by it to be free from defects in material and workmanship un­der normal use and service, its obligation under this warranty being limited to making good at its factory or other article of equipment which shall within one (1) year after delivery of such instrument or other article of equipment to the original purchaser be returned intact to it, or to one of its authorized service centers, with transportation charges prepaid, and which its examination shall disclose to its satisfaction to have been thus defective; this warranty being expressly in lieu of all other warranties expressed or implied and of all other obligations or liabilities on its part, and SIMPSON ELECTRIC COMPANY nei­ther assumes nor authorizes any other persons to assume for it any other liability in connection with the sales of its products.
2
Page 3
NOTES
3
Page 4
Contents
1. INTRODUCTION................................................................................ 7
1.1 Technical Data ....................................................................................7
1.2 Items And Accessories .......................................................................8
2. CARE OF INSTR UMENT .................................................................. 8
2.1 Unpacking And Inspection ................................................................. 8
2.2 Warranty..............................................................................................8
2.3 Shipping .............................................................................................8
2.4 Care ....................................................................................................8
2.5 Battery Installation Or Replacement ..................................................9
3. OPERATION...................................................................................... 9
3.1 Precautions.........................................................................................9
3.2 Operating Items And Features.......................................................... 10
3.3 Effects Of Atmospheric Pressure And Temperature .........................1 1
3.4 Battery Check ................................................................................... 1 1
3.5 Sound Level Meter Setting ............................................................... 11
3.6 Calibration of 884-2, 886-2 and 899 Sound Level Meters...............1 1
4
Page 5
NOTES
5
Page 6
NOTES
6
Page 7
1. INTRODUCTION
The Simpson 896 Multi-Frequency/Level Calibrator (hereafter referred to as the 896, or the Instrument) is an accurate sound source designed for conve­nient field calibration of the Simpson 884-2, 886-2 or 899 Sound Lev el Meters . The 896 is compact, durably constructed, lightweight, and battery-powered for portable use. The 896 has two sound pressure levels (114 dB or 94 dB) and five output frequencies (125, 250, 500, 1000 and 2000 Hz). This allows versatility of checking the A, B or C weighted frequency response characteristics of the Sound Level Meter. The Instrument’s housing shields the electronic circuitry and provides efficient acoustic coupling between the calibrator transducer and the Sound Level Meter microphone.
!
Never place the operating calibrator tightly against the ear. The high sound pressure level could be harmful, especially to someone with a hearing defi­ciency.
1.1 Technical Data
Table 1-1 lists the technical specifications for the 896 Multi-Frequency/Level Calibrator.
Table 1-1. Technical Data
Acoustic Outputs
Frequencies: 125, 250, 500, 1000 or 2000 Hz.
Sound Pressure Levels: 114 dB SPL or 94 dB SPL
(Ref: 0 dB = 20 mPascals).
Accuracies: SPL: 0.5 dB all frequencies at reference
conditions.
Frequencies: 2% all frequencies.
Distortion: Less than 5%.
Temp. Range: 0°C to 50°C.
Output Temperature
Coefficient: Less than minus 0.02 dB/°C.
Operating Relative
Humidity: 0 to 90%.
Power Requirements: One 9 V (NEDA 1604) battery.
Battery Life: Approx. 120 hours (2 hours per day
operation). Nominal Current Drain: 3 mA. (Refer to appropriate Manufacturer’s data for temperature effects on battery life.)
Reference Conditions: 23°C, 760 mm Hg and 30% to 60% Relative
Humidity.
Weight: 15 oz. (0.426 kg).
Dimensions: 2" wide x 5-7/8" long.
Construction: Aluminum housing with all electronic circuitr y
on a single printed circuit board.
7
Page 8
1.2 Items And Accessories
All items and accessories required to operate the 896 are furnished with the Instrument and listed in Table 1-2.
Table 1-2. Items and Accessories Furnished With the Instrument
Quantity Description Part Number
1 Screwdriver 5-116470 1 9 V Battery , NED A 1604* 1 Operator’s Manual 6-110689
*Available from local retail stores (transistor radio type).
2. CARE OF INSTRUMENT
This section contains unpacking and inspection procedures, warranty, ship­ping, care, and battery installation.
2.1 Unpacking And Inspection
Examine the shipping carton for signs of damage before unpacking. If ship­ping carton is in good condition, unpack and inspect the Instrument for pos­sible damage incurred during shipment. If damaged, notify the carrier and supplier. Do not attempt to use the Instr ument.
If the Instrument appears to be in good condition, read the Operator’s Manual in its entirety . Become f amiliar with the Instrument as instructed in the manual, then check the electrical performance as soon as possible. Check that all items are included with the Instrument (see Table 1-2). Save the shipping carton and packing materials for storing or future shipping of the Instrument.
2.2 Warranty
The Simpson Electric Company warranty policy is printed on the inside front cover of this manual. Read it carefully before requesting a warranty repair. For all assistance, contact the factory directly. Give full details of any prob­lems along with the model number, serial number, and date of purchase.
2.3 Shipping
Pack the Instrument carefully, and ship it prepaid and insured to the proper destination.
2.4 Care
1. Immediately clean any spilled material from the Instrument and wipe dry. If spillage is corrosive, use a suitable cleaner to remove spillage and to neutralize corrosive action.
2. When Instrument is not in use, set the Rotary Switch to the OFF position.
8
Page 9
3. Avoid prolonged exposure or usage in areas subject to temperature and humidity extremes, vibration, mechanical shock, dust, corrosive fumes, and strong electrical and electromagnetic interferences.
4. Be sure the transducer cap is firmly in place. (Remo val of the Transducer cap may invalidate calibration.)
5. If the Instrument has not been used for 30 days, check battery for leak­age and replace if necessary.
6. It is recommended that the Instrument be returned to the factory annually (sooner if required) for a complete overall check, adjustment, and cali­bration.
7. When not in use, store instrument in a room free from temperature ex­tremes, dust, corrosive fumes, mechanical vibration or shock. If storage time is expected to exceed 30 days, remove the battery.
2.5 Battery Installation Or Replacement
1. Set the Rotary Switch of the 896 to the OFF position.
2. Remove the battery cap (Table 3-1, item 1) by turning it counterclock­wise.
3. If replacing the battery, remove the old battery.
4. Snap the battery connector onto the mating terminals of the replacement battery. The connector is polarized to allow it to be connected only one way.
5. Tuck the connector cable neatly under the battery so it will not interfere when the battery cap is replaced, and slide the battery into the batter y compartment. Replace the battery cap.
3. OPERATION
This section contains information required to operate the Instrument in a safe and proper manner. Follow the instructions herein and in the Operator’s Manual for the Sound Level Meter.
3.1 Precautions
!
Do not place the operating Calibrator tightly against the ear. The high sound levels could be harmful, especially to someone with a hearing deficiency.
Check the calibration of the 896 periodically or whenever accur acy is in doubt. Do this by comparing Instrument with another 896 Calibrator or a reference Sound Level Meter known to be accurate. If another 896 or reference Sound Level Meter is unavailable, return the Instrument to the factory for a complete overall check.
Be sure the Sealing Ring in the Transducer Cap is not damaged and that the leak-to-atmosphere opening is not blocked. (Table 3-1, item 5.)
9
Page 10
3.2 Operating Items And Features
All items or features used to operate the calibrator are described in Table 3-1. Become familiar with each before operating the Instrument for the first time.
1
+
94 dB L E V E L
114dB
+
T
F
O
F
T
A
B
T
S
E
T
2
z
H
k
z
1
H
k
0
z
0
H
5
z
H
0
O
z
H
5
F
2
F
5
2
1
Pressure Leak
CAUTION
NEVER PLACE THE CALIBRATOR
AGAINST THE EAR. READ THE
OPERATOR'S MANUAL
Sealing Ring
Figure 3-1. Model 896
Table 3-1. Operational Items or Features
1. Battery Cap: This aluminum cap provides easy access to the Battery Compartment and holds the bat­tery firmly in position.
2. BA TT TEST Indicator: When set to BATT TEST position, a Light­Emitting Diode (LED) indicator will light if the battery is strong enough to provide specified Sound Pressure Levels of 114 or 94 dB SPL.
NOTE: If the LED does not light, replace the battery and retest. DO NOT use the calibrator if it does not pass the battery test. The results will not be accurate.
3. OFF/Batt Test/Frequency
Select Switch: This 8-position rotary switch is used to turn
calibrator on/off, to test the battery or to se­lect one of the five output frequencies .
4. 114 dB/94 dB Level
Select Switch: This 2-position slide switch selects either the
114 or 94 dB calibration levels.
5. Transducer Cap: This cap forms an acoustic coupler between the calibrator and the Sound Level Meter mi­crophone. A small hole in the side of the cap provides a controlled pressure leak to the at­mosphere.
10
4
2
3
5
Page 11
3.3 Effects Of Atmospheric Pressure And Temperature
For any one location, the effects of normal variations of atmospheric pressure are usually negligible. (The effect of elevation is shown in Figure 3-2.) The graph shows the change in Instrument output level compared to that obtained at sea level. For example, if the Instrument is being used at 5000 feet, cali­brate the Sound Level Meter for an indication of 114.58 dB for the frequency of 1000 Hz.
The effects of temperature are less than -0.02 dB/°C (Reference is 23°C).
3.4 Battery Check
Check battery condition by setting the Rotary Switch to the BATT TEST posi­tion. The LED will light, indicating that the batter y is adequate. If it does not light, replace the battery.
(An audible output may be obtained with a weak battery, however, any cali­bration made with the Calibrator will be invalid.)
3.5 Sound Level Meter Setting
1. Set the Rotary Switch to the 110-120 dB range.
2. Set the Response Switch to the SLOW or FAST position.
3. Set the A, B, or C weighting switches to weighting specified for applica-
tion. If no weighting is specified, use C weighting (A weighting only for
884).
117 116 115 114 113 112 111 110
896 OUTPUT LEVEL (dBSPL)
109 108
MM
HG
INCHES
HG
FEET ABOVE SEA LEVE;
(THOUSANDS)
Figure 3-2. Effect of Elevation Upon Calibrator Output
760
250 Hz 500 Hz
125 Hz
1K Hz
2K Hz
650
26
4
600
550
24 22 20
8
6
10
500
450
18
12
14
400
16
16
350
14
20
18
750
700
30
28
2
0
1
500 Hz 125 Hz
300
3.6
Calibration of 884-2, 886-2 and 899 Sound Level Meter s
The Simpson 884-2, 886-2 and 899 Sound Level Meters have a condenser microphone and electronic weighting networks in order to obtain random in­cidence response characteristics as specified in ANSI S1.4-1983. To corre­late these random incidence response characteristics to those readings ob-
11
Page 12
tained on the Sound Level Meter when coupled to an 896 Calibrator, correc­tion factors relating the Sound Level Meter random incidence frequency re­sponse to its pressure frequency response (obtained in an acoustic coupler) must be known. These correction factors take into consideration the con­denser microphone random incidence versus pressure frequency response differences and the A, B, or C frequency weighting versus flat frequency re­sponse differences. The former differences are negligible for the frequencies involved when using the 896. The latter differences are listed in Table 3-2.
ycneuqerF
gnithgieW521052005K1K2 A1.61-6.8-2.3-0 2.1+ B2.4-3.1-3.0-0 1.0­C2.0- 0002.0-
Table 3-2. Nominal Differences between A, B, or C Frequency
Weighting Responses and Flat Frequency Response (dB)
Calibration at 1kHz:
1. Pull off the small metal cap covering the calibration adjustment of the Sound Level Meter.
2. If using the windscreen, remove and carefully insert the microphone fully into the acoustic cavity of the calibrator.
3. With one hand, hold the 896 and the Sound Level Meter against the body, while holding the Sound Level Meter in a horizontal plane. (See Figure 3-3.)
4. Set the Rotary Switch of the 896 to the 1 kHz position.
Calibrating an 884-2, 886-2 and 899 SLM
+
94 dB L E V E L
114dB
+
T
O
F
T
A
F
B
T
S
E
T
2
z
H
k
z
1
H
k
0
0
z
5
H
z
0
H
O
z
5
H
2
F
F
5
2
1
CAUTION
NEVER PLACE THE CALIBRATOR
AGAINST THE EAR. READ THE
OPERATOR'S MANUAL
OFF
50
60
BATT TEST
SLOW
FAST
884-2
SOUND LEVEL METER
12
130
OSHA
80
70
(OSHA)
TYPE S2A
Figure 3-3
120
100
110
90
Page 13
5. Set Slide Switch to the 114 dB position.
6. Observe the meter indicator on the Instrument. It should read 114 dB
60.5 dB (when atmospheric and temperature effects are negligible). If it does not, make the necessary adjustment using screwdriver provided with the calibrator.
NOTE: The factors of atmospheric pressure and ambient temperature must be taken into consideration when observing the meter reading and before performing any adjustment (see paragraph 3.4).
7. Set the Rotary Switch of the 896 to the OFF position.
8. Set the Range Switch of the Sound Level Meter to the OFF position.
9. Replace the cap covering the calibration adjustment of the Sound Level Meter.
The 884-2, 886-2 or 899 can also be calibrated at 2000, 500, 250, or 125 Hz in a manner similar to the 1000 Hz calibration; however, the weighting effects must be taken into consideration. ( Refer to Table 3-2.) For example, if cali­bration at 250 Hz in the A weighting is desired, the actual Sound Level Meter indication would be 114 - 8.6 = 105.4 dB SPL. Set the Sound Level Meter to the 100 dB range and the 896 to the 250 Hz and 114 dB SPL and calibrate the Sound Level Meter for 105.4 dB SPL.
13
Page 14
NOTES
14
Page 15
NOTES
15
Page 16
SIMPSON ELECTRIC COMPANY 853 Dundee Avenue
Elgin, IL 60120-3090 (847) 697-2260 FAX (847) 697-2272
Printed in U.S.A. Part No. 06-110689 Edition 4, 10/02
Visit us on the web at: www.simpsonelectric.com
16
Loading...